Академический Документы
Профессиональный Документы
Культура Документы
Purposes:
To maintain or improve ventilation, &
tissue oxygenation.
Indications:
1- Acute respiratory failure due to:
anesthesia as well as post cardiac arrest patients often require ventilatory support until they have recovered from the effects of the anesthesia or the insult of an arrest.
Ventilation indicated
> 35 <5 < 15 <-20
Normal range
10-20 5-7 65-75 75-100
Ventilation indicated
Normal range
Positive-pressure ventilators.
Negative-Pressure Ventilators
Early negative-pressure ventilators
were known as iron lungs.
Positive-pressure ventilators
Positive-pressure ventilators deliver
gas to the patient under positivepressure, during the inspiratory phase.
3- High-Frequency Ventilators
1- Volume Ventilators
The volume ventilator is commonly used in
critical care settings.
respiratory rate, inspiratory time, and tidal volume are selected for the mechanical breaths.
2- Pressure Ventilators
The use of pressure ventilators is
increasing in critical care units.
selected gas pressure to the patient early in inspiration, and sustains the pressure throughout the inspiratory phase.
3- High-Frequency Ventilators
High-frequency ventilators use small
tidal volumes (1 to 3 mL/kg) at frequencies greater than 100 breaths/minute.
accomplishes oxygenation by the diffusion of oxygen and carbon dioxide from high to low gradients of concentration.
used to achieve lower peak ventilator pressures, thereby lowering the risk of barotrauma.
1- Volume-cycled ventilator
Inspiration is terminated after a
preset tidal volume has been delivered by the ventilator.
2- Pressure-cycled ventilator
In which inspiration is terminated
when a specific airway pressure has been reached.
3- Time-cycled ventilator
In which inspiration is terminated
when a preset inspiratory time, has elapsed.
Ventilator mode
The way the machine ventilates the
patient
A- Volume Modes
1- Assist-control (A/C) 2- Synchronized intermittent mandatory ventilation (SIMV)
own, but the ventilator assists by delivering a specified tidal volume to the patient. Client can initiate breaths that are delivered at the preset tidal volume.
Disadvantages:
Hyperventilation,
drastically from the tidal volume set on the ventilator, because the tidal volume is determined by the patients spontaneous effort.
Weaning is accomplished by
gradually lowering the set rate and allowing the patient to assume more work
B- Pressure Modes
1- Pressure-controlled ventilation (PCV) 2- Pressure-support ventilation (PSV) 3- Continuous positive airway pressure (CPAP) 4- Positive end expiratory pressure (PEEP) 5- Noninvasive bilevel positive airway pressure ventilation (BiPAP)
breathing gas flows under constant pressure into the lungs during the selected inspiratory time. The flow is highest at the beginning of inspiration( i.e when the volume is lowest in the lungs). As the pressure is constant the flow is initially high and then decreases with increasing filling of the lungs.
blocking agents are frequently indicated, because any patientventilator asynchrony usually results in profound drops in the SaO2. are used. The unnatural feeling of this mode often requires muscle relaxants to ensure patientventilator synchrony.
reverses this ratio so that inspiratory time is equal to, or longer than, expiratory time (1:1 to 4:1). with pressure control to improve oxygenation by expanding stiff alveoli by using longer distending times, thereby providing more opportunity for gas exchange and preventing alveolar collapse.
monitor for the development of hyperinflation or auto-PEEP. Regional alveolar overdistension and barotrauma may occur owing to excessive total PEEP.
airway and intrathoracic pressures rise, potentially resulting in a decrease in cardiac output and oxygen delivery. Therefore, the patients hemodynamic status must be monitored closely.
cause barotrauma & Severe ARDS
volume and respiratory rate must be monitored closely to detect changes in lung compliance.
CPAP allows the nurse to observe the CPAP can be used for intubated and
nonintubated patients.
Frequency ( F) Minute Volume (VE) I:E Ratio (Inspiration to Expiration Ratio) Sigh
100% until arterial blood gases can document adequate oxygenation. can be dangerous ( oxygen toxicity) but it can protect against hypoxemia
Usually the FIO2 is adjusted to maintain an SaO2 Oxygen toxicity is a concern when an FIO2 of
Signs and symptoms of oxygen toxicity :1- Flushed face 2- Dry cough 3- Dyspnea 4- Chest pain 5- Tightness of chest 6- Sore throat
The amount of air inspired and expired Usual volume selected is between 5 to 15
ml/ kg body weight)
Respiratory alkalosis may be required to In this case, the tidal volume and
respiratory rate are increased ( hyperventilation) to achieve the desired alkalotic pH by manipulating the PaCO2.
should not hyperventilated. Instead, the goal should be restoration of the baseline PaCO2.
carbonic acid load, and lowering their carbon dioxide levels rapidly may result in seizures.
Sigh
A deep breath. A breath that has a greater volume than the tidal
volume.
what level of peak pressure may cause damage. In general, keeping peak pressures below 30 is desirable.
Pressure Limit
On volume-cycled ventilators, the pressure
limit dial limits the highest pressure allowed in the ventilator circuit.
inspiration is terminated.
Once the high pressure limit is reached, Therefore, if the pressure limit is being
constantly reached, the designated tidal volume is not being delivered to the patient.
Sensitivity(trigger Sensitivity)
The sensitivity function controls the
amount of patient effort needed to initiate an inspiration negative force) decreases the amount of work the patient must do to initiate a ventilator breath.
amount of negative pressure that the patient needs to initiate inspiration and increases the work of breathing.
sensitivity are -1 to -2 cm H2O The more negative the number the harder it to breath.
the airway, often with resultant dried secretions, mucus plugging and less ability to suction out secretions. may increase circuit resistance and interfere with spontaneous breathing.
Humidifier should not be overfilled as this As air passes through the ventilator to the
patient, water condenses in the corrugated tubing. This moisture is considered contaminated and must be drained into a receptacle and not back into the sterile humidifier.
resistance is developed in the circuit and PEEP is generated. In addition, if moisture accumulates near the endotracheal tube, the patient can aspirate the water.
are responsible for preventing this condensation buildup. The humidifier is an ideal medium for bacterial growth.
I- Airway Complications
1- Aspiration
2- Decreased clearance of secretions 3- Nosocomial or ventilator-acquired pneumonia
IV- Artificial Airway Complications A- Complications related to Endotracheal Tube:1- Tube kinked or plugged 2- Rupture of piriform sinus 3- Tracheal stenosis or tracheomalacia 4- Mainstem intubation with contralateral (located on or affecting the opposite side of the lung) lung atelectasis 5- Cuff failure 6- Sinusitis 7- Otitis media 8- Laryngeal edema
B- Complications related to Tracheostomy tube:1- Acute hemorrhage at the site 2- Air embolism 3- Aspiration 4- Tracheal stenosis 5- Erosion into the innominate artery with exsanguination 6- Failure of the tracheostomy cuff 7- Laryngeal nerve damage 8- Obstruction of tracheostomy tube 9- Pneumothorax 10- Subcutaneous and mediastinal emphysema 11- Swallowing dysfunction 12- Tracheoesophageal fistula 13- Infection 14- Accidental decannulation with loss of airway
Nursing Interventions
1-Maintain airway patency & oxygenation 2- Promote comfort 3- Maintain fluid & electrolytes balance 4- Maintain nutritional state 5- Maintain urinary & bowel elimination 6- Maintain eye , mouth and cleanliness and integrity:7- Maintain mobility/ musculoskeletal function:-
Nursing Interventions
8- Maintain safety:9- Provide psychological support 10- Facilitate communication 11- Provide psychological support & information to family 12- Responding to ventilator alarms /Troublshooting ventilator alarms 13- Prevent nosocomial infection 14- Documentation
Responding To Alarms
If an alarm sounds, respond immediately
because the problem could be serious.
the alarm.
Assess the patient first, while you silence If you can not quickly identify the
problem, take the patient off the ventilator and ventilate him with a resuscitation bag connected to oxygen source until the physician arrives. respond to every ventilator alarm.
Ventilator malfunction is a
potentially serious problem. Nursing or respiratory therapists perform ventilator checks every 2 to 4 hours, and recurrent alarms may alert the clinician to the possibility of an equipmentrelated issue.
When
device malfunction is suspected, a second person manually ventilates the patient while the nurse or therapist looks for the cause. If a problem cannot be promptly corrected by ventilator adjustment, a different machine is procured so the ventilator in question can be taken out of service for analysis and repair by technical staff.
endotracheal tube (ETT) Patient biting the ETT Water in the ventilator tubing. ETT advanced into right mainstem bronchus.
Oxygen alarm
Apnea alarm
Temperature alarm
Methods of Weaning
1- T-piece trial, 2- Continuous Positive Airway Pressure (CPAP) weaning, 3- Synchronized Intermittent Mandatory Ventilation (SIMV) weaning, 4- Pressure Support Ventilation (PSV) weaning.
1- T-Piece trial
It consists of removing the patient from
the ventilator and having him / her breathe spontaneously on a T-tube connected to oxygen source.
number of breaths delivered by the ventilator to allow the patient to increase number of spontaneous breaths
No mandatory (ventilator-initiated)
pressure support is gradually decreased based on the patient maintaining an adequate tidal volume (8 to 12 mL/kg) and a respiratory rate of less than 25 breaths/minute.
PSV weaning is indicated for :- Difficult to wean patients - Small spontaneous tidal volume.
(PEEP) 5 cm H2O F < 25 / minute Vt 5 ml / kg VE 5- 10 L/m (f x Vt) VC > 10- 15 ml / kg PEP (positive expiratory pressure) > - 20 cm H2O ( indicates patients ability to take a deep breath & cough),
factors; treated as indicated, Major electrolytes within normal range, Hematocrit >25%, Core temperature >36C and <39C, Adequate management of pain/anxiety/agitation, Adequate analgesia/ sedation (record scores on flow sheet), No residual neuromuscular blockade.
Role of nurse before weaning:1- Ensure that indications for the implementation of Mechanical ventilation have improved 2- Ensure that all factors that may interfere with successful weaning are corrected:- Acid-base abnormalitie - Fluid imbalance - Electrolyte abnormalities - Infection - Fever - Anemia - Hyperglycemia - Protein - Sleep deprivation
Role of nurse before weaning:3- Assess readiness for weaning 4- Ensure that the weaning criteria / parameters are met.
5- Explain the process of weaning to the patient and offer reassurance to the patient.
6- Initiate weaning in the morning when the patient is rested. 7- Elevate the head of the bed & Place the patient upright 8- Ensure a patent airway and suction if necessary before a weaning trial,
9- Provide for rest period on ventilator for 15 20 minutes after suctioning.
10- Ensure patients comfort & administer pharmacological agents for comfort, such as bronchodilators or sedatives as indicated.
11- Help the patient through some of the discomfort and apprehension.
12- Support and reassurance help the patient through the discomfort and apprehension as remains with the patient after initiation of the weaning process.
13- Evaluate and document the patients response to weaning.
Role of nurse during weaning:1- Wean only during the day. 2- Remain with the patient during initiation of weaning. 3- Instruct the patient to relax and breathe normally. 4- Monitor the respiratory rate, vital signs, ABGs, diaphoresis and use of accessory muscles frequently. If signs of fatigue or respiratory distress develop.
Diaphoresis
Tidal volume 5 mL/kg, Sustained minute SaO2 < 90%, PaO2 < 60 mmHg, decrease
2- Decanulate or extubat
2- Documentation
Good Luck