Академический Документы
Профессиональный Документы
Культура Документы
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to
this laboratory to perform the following calibrations1:
I. Dimensional
Parameter/Equipment
Range
CMC2, 3 ()
Comments
Micrometers
Up to 1 in
(9 + 2L) in
Gage blocks
Calipers
Up to 6 in
(2.6 + 60L) in
Gage blocks
Dial Indicators
Up to 1 in
120 in
Gage blocks
RF Connector Gages
Type N
0.002 in
Super micrometer
Range
100 mV
1V
5V
10 V
15 V
CMC2, 4, 7 ()
2.8 V/V
0.7 V/V
0.8 V/V
0.5 V/V
0.6 V/V
Comments
Calibration of reference
multimeters with Fluke
5720A characterized
calibrator
Page 1 of 45
Parameter/Equipment
Range
CMC2, 4, 6, 7 ()
Comments
DC Voltage System
Generate, Fixed Points5
(cont)
19 V
100 V
1 kV
0.6 V/V
0.6 V/V
0.8 V/V
Calibration of reference
multimeters with Fluke
5720A characterized
calibrator
DC Reference
Calibration Measure
& Generate, Fixed
Points5
10 V reference
1.018 V reference
1.00 V reference
0.3 V/V
0.7 V/V
0.7 V/V
Direct transfer
techniques performed
utilizing Fluke
732A/732B
DC Voltage Measure
& Generate
(0 to 0.02) V
(20 to 120) mV
120 mV to 1.2 V
(1.2 to 12) V
(12 to 120) V
120 V to 1.05 kV
4 V/V + 0.21 V
3 V/V + 0.1 V
0.7 V/V + 0.2 V
0.5 V/V + 0.5 V
0.8 V/V + 20 V
1.9 V/V + 150 V
Direct measurements or
transfer techniques
using the Fluke 4950
(1.2 to 40) kV
0.1 %
Direct measurements
using Ross VD120
voltage divider/Fluke
8846A
3 V/V + 0.3 V
1 V/V + 1 V
0.6 V/V + 2.5 V
0.6 V/V + 5 V
1 V/V + 50 V
1.2 V/V + 0.2 mV
33 /
4 /
0.3 /
0.4 /
0.4 /
0.4 /
0.5 /
1.4 /
2.2 /
3.2 /
18 /
44 /
0.02 %
0.03 %
0.09 %
DC Voltage Measure
0.01
0.1
1
10
100
1 k
10 k
100 k
1 M
10 M
100 M
1 G
100 G
1 T
10 T
HP 3458A
characterized DMM
Page 2 of 45
Parameter/Equipment
Range
CMC2, 4, 6, 7 ()
Comments
(0.0001 to 0.01)
(0.01 to 0.1)
(0.1 to 1)
(1 to 10)
(10 to 100)
100 to 1 k
(1 to 10) k
6
5 /
1 /
0.5 /
0.5 /
0.5 /
0.7 /
Fluke 5450A/8508A
MI 6010 bridge,
working standard
resistors
(10 to 100) k
100 k to 1 M
(1 to 10) M
(10 to 100) M
100 M to 1 G
0.7 /
1.6 /
1.9 /
3.6 /
26 /
Ratio measurements
using MIL bridge or
1:1 measurements with
Fluke 8508A
Resistance Generate
& Measure, Fixed
Points
2 G
4 G
8 G
10 G
100 G
1 T
0.76 %
0.99 %
0.67 %
0.64 %
0.69 %
1.1 %
Fluke 5320A,
QuadTech 1865
megohmmeter
DC Current Generate
& Measure
(1 to 10) A
(10 to 100) A
100 A to 1 mA
(1 to 10) mA
(10 to 100) mA
100 mA to 1 A
(1 to 10) A
(10 to 18) A
(18 to 20) A
6 A/A + 10 pA
6 A/A + 60 pA
6 A/A + 1 nA
6 A/A + 10 nA
6 A/A + 0.1 A
6 A/A + 1.8 A
5 A/A + 5 A
55 A/A + 20 A
80 A/A
0.25 % + 50 mA
0.28 % + 75 mA
0.3 % + 50 mA
0.3 %
(20 to 50) A
(50 to 80) A
18 A/A
26 A/A
Fluke 6105A
(80 to 120) A
28 A/A
Fluke 52120A
transconductance
amplifier
Resistance Generate
& Measure
Generate Only
DC Current Generate
& Measure
Page 3 of 45
Parameter/Equipment
Range
CMC2, 4, 6, 7 ()
Comments
100 A
1 mA
10 mA
100 mA
300 mA
1A
10 A
1.3 A/A
1.8 A/A
1.9 A/A
1.1 A/A
4.4 A/A
2.4 A/A
6.2 A/A
Comparison to DC
reference shunts and
characterized voltmeter
Generate
100 A
1 mA
10 mA
100 mA
300 mA
1A
10 A
2.1 A/A
2.1 A/A
2.1 A/A
1.7 A/A
5 A/A
2.7 A/A
11 A/A
(295 to 299) K
(295 to 299) K
(295 to 299) K
0.04 K
0.04 K
0.04 K
Fluke 8520/PRT
reference and
thermocouple
(0.5 to 4) nF
(4 to 40) nF
(40 to 400) nF
(0.4 to 4) F
(4 to 400) F
40 F to 40 mF
0.005 % + 3 pF
0.03 % + 3 pF
0.014 %
0.0091%
0.014 %
0.024 %
Automated 9100
calibration system
0.18 %
0.08 %
0.067 %
0.06 %
0.035 %
0.05 %
0.03 %
0.02 %
0.05 %
0.02 %
0.05 %
0.03 %
0.02 %
Electrical Calibration of
Thermocouple Reference
Junction
Type J
Type K
Type T
Capacitance Generate
& Measure
Page 4 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage
Generate Fixed
Points
1 mV
Reference multimeter
calibration
2 mV
0.02 %
AC divider
Fluke 5790A or 792A
100 kHz
300 kHz
500 kHz
1 MHz
0.022 %
0.03 %
0.036 %
0.13 %
80 V/V
100 kHz
300 kHz
500 kHz
1 MHz
90 V/V
0.021 %
0.026 %
0.055 %
10 mV
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
10 kHz
20 kHz
30 kHz
50 kHz
100 kHz
300 kHz
500 kHz
1 MHz
0.012 %
0.023 %
0.016 %
0.01 %
0.011 %
0.019 %
0.024 %
0.017 %
0.017 %
0.018 %
0.021 %
0.034 %
0.028 %
0.066 %
0.062 %
Calibration of reference
multimeters
20 mV
10 Hz, 20 Hz
100 Hz, 1 kHz, 10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
1 MHz
60 V/V
40 V/V
40 V/V
50 V/V
85 V/V
0.02 %
0.025 %
0.04 %
6 mV
Page 5 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Generate,
Fixed Points (cont)
60 mV
10 Hz
20 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
1 MHz
0.01 %
67 V/V
47 V/V
38 V/V
48 V/V
53 V/V
72 V/V
0.014 %
0.027 %
0.035 %
0.041 %
100 mV
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
3 kHz
10 kHz
20 kHz
30 kHz
50 kHz
60 kHz
100 kHz
300 kHz
500 kHz
1 MHz
50 V/V
72 V/V
48 V/V
26 V/V
30 V/V
41 V/V
34 V/V
39 V/V
31 V/V
33 V/V
47 V/V
47 V/V
0.012 %
88 V/V
0.016 %
0.021 %
0.033 %
Calibration of
reference multimeter
200 mV
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
100 Hz
300 Hz
1 kHz
3 kHz
10 kHz
35 V/V
23 V/V
36 V/V
23 V/V
33 V/V
18 V/V
33 V/V
18 V/V
32 V/V
18 V/V
Calibration of 5790A
600 mV
10 Hz
20 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
26 V/V
35 V/V
11 V/V
12 V/V
11 V/V
11 V/V
22 V/V
Calibration of transfer
standard Fluke 5790A
Page 6 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Generate,
Fixed Points (cont)
600 mV
100 kHz
300 kHz
500 kHz
1 MHz
12 V/V
95 V/V
76 V/V
83 V/V
Calibration of transfer
standard Fluke 5790A
1V
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
100 Hz
300 Hz
1 kHz
3 kHz
10 kHz
20 kHz
30 kHz
50 kHz
60 kHz
100 kHz
300 kHz
500 kHz
1 MHz
85 V/V
33 V/V
40 V/V
41 V/V
40 V/V
9 V/V
14 V/V
10 V/V
23 V/V
11 V/V
9 V/V
26 V/V
18 V/V
84 V/V
11 V/V
70 V/V
90 V/V
75 V/V
Calibration of reference
multimeters or 5790A
transfer standard.
2V
10 Hz
20 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
1 MHz
25 V/V
20 V/V
8 V/V
6 V/V
7 V/V
7 V/V
14 V/V
11 V/V
90 V/V
90 V/V
75 V/V
5V
1 kHz
0.03 %
6V
10 Hz
20 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
1 MHz
20 V/V
18 V/V
9 V/V
8 V/V
8 V/V
7 V/V
21 V/V
11 V/V
75 V/V
75 V/V
75 V/V
Calibration of Fluke
5790A
Page 7 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Generate,
Fixed Points (cont)
10 V
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
3 kHz
10 kHz
20 kHz
30 kHz
50 kHz
60 kHz
100 kHz
300 kHz
500 kHz
1 MHz
40 V/V
43 V/V
32 V/V
29 V/V
28 V/V
17 V/V
18 V/V
24 V/V
17 V/V
18 V/V
19 V/V
18 V/V
41 V/V
20 V/V
0.01 %
0.013 %
0.041 %
19 V
1 kHz
18 V/V
20 V
10 Hz
20 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
500 kHz
1 MHz
25 V/V
21 V/V
10 V/V
10 V/V
10 V/V
11 V/V
11 V/V
16 V/V
85 V/V
90 V/V
76 V/V
60 V
10 Hz
20 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 kHz
26 V/V
22 V/V
12 V/V
12 V/V
14 V/V
12 V/V
31 V/V
16 V/V
71 V/V
100 V
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
52 V/V
31 V/V
31 V/V
31 V/V
31 V/V
20 V/V
20 V/V
Calibration of reference
multimeters
Calibration of Fluke
5790A
Calibration of reference
multimeters
Page 8 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Generate,
Fixed Points (cont)
100 V
3 kHz
10 kHz
20 kHz
30 kHz
50 kHz
60 kHz
100 kHz
16 V/V
19 V/V
20 V/V
21 V/V
21 V/V
32 V/V
29 V/V
200 V
10 Hz
20 Hz
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
36 V/V
23 V/V
18 V/V
13 V/V
13 V/V
14 V/V
16 V/V
21 V/V
500 V
55 Hz
300 Hz
1 kHz
3 kHz
10 kHz
30 kHz
35 V/V
30 V/V
25 V/V
25 V/V
25 V/V
65 V/V
600 V
100 Hz
1 kHz
10 kHz
20 kHz
50 kHz
100 kHz
21 V/V
17 V/V
17 V/V
18 V/V
37 V/V
55 V/V
700 V
50 kHz
100 kHz
71 V/V
0.018 %
1000 V
40 Hz
55 Hz
100 Hz
300 Hz
1 kHz
3 kHz
10 kHz
20 kHz
30 kHz
50 kHz
100 kHz
29 V/V
27 V/V
22 V/V
25 V/V
18 V/V
22 V/V
18 V/V
19 V/V
56 V/V
42 V/V
54 V/V
1V
Calibration of reference
multimeters
Calibration of Fluke
5790A
Calibration of reference
multimeters or 5790A
Page 9 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Generate,
Fixed Points (cont)
3V
2 MHz, 4 MHz,
8 MHz, 10 MHz
0.5 %
Calibration of reference
multimeters or 5790A
(1 to 2.2) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.6 %
0.6 %
0.6 %
0.6 %
0.7 %
1.2 %
2.3 %
2.4 %
(2.2 to 22) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.21 %
0.2 %
0.19 %
0.21 %
0.28 %
0.57 %
1.1 %
1.2 %
(22 to 220) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.08 %
0.05 %
0.04 %
0.06 %
0.13 %
0.19 %
0.26 %
0.48 %
(0.22 to 2.2) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.05 %
0.02 %
0.01 %
0.02 %
0.03 %
0.08 %
0.2 %
0.31 %
(2.2 to 22) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
0.05 %
0.02 %
0.01 %
0.02 %
0.02 %
AC Voltage Generate,
Ranges
Page 10 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Generate,
Ranges
(cont)
(2.2 to 22) V
0.06 %
0.2 %
0.31 %
(22 to 220) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
0.05 %
0.02 %
0.01 %
0.02 %
0.03 %
(220 to 1000) V
(10 to 50) Hz
50 Hz to 1 kHz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 30) kHz
(30 to 50) kHz
(50 to 100) kHz
0.04 %
0.01 %
0.02 %
0.02 %
0.07 %
0.07 %
0.26 %
(1 to 2.2) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.026 % + 1 V
0.026 % + 1 V
0.028 % + 1 V
0.029 % + 1.6 V
0.032 % + 1.9 V
0.05 % + 3.1 V
0.063 % + 4.7 V
0.19 % + 4.7 V
(2.2 to 7) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.014 % + 1 V
0.011 % + 1 V
0.011 % + 1 V
0.011 % + 1.6 V
0.018 % + 1.9 V
0.048 % + 3.1 V
0.059 % + 4.7 V
0.085 % + 4.7 V
(7 to 22) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
70 V/V + 1 V
70 V/V + 1 V
50 V/V + 1 V
60 V/V + 1.6 V
0.01 % + 1.9 V
0.024 % + 3.1 V
0.029 % + 4.7 V
0.049 % + 4.7 V
AC Voltage Measure
Parameter/Range
Frequency
Comments
Page 11 of 45
CMC2, 4, 5, 6, 7 ()
AC Voltage Measure
(cont)
(22 to 70) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.012 % + 1.2 V
0.011 % + 1.2 V
60 V/V + 1.2 V
80 V/V + 1.6 V
0.014 % + 1.9 V
0.029 % + 3.1 V
0.038 % + 4.7 V
0.051 % + 4.7 V
(70 to 220) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
60 V/V + 1.2 V
50 V/V + 1.2 V
30 V/V + 1.2 V
50 V/V + 1.6 V
90 V/V + 1.9 V
0.014 % + 3.1 V
0.013 % + 4.7 V
0.025 % + 4.7 V
(220 to 700) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
50 V/V
50 V/V
30 V/V
50 V/V
80 V/V
0.013 %
0.017 %
0.023 %
700 mV to 2.2 V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
40 V/V
40 V/V
30 V/V
30 V/V
30 V/V
0.01 %
0.01 %
0.011 %
(2.2 to 7) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
40 V/V
40 V/V
30 V/V
30 V/V
30 V/V
0.01 %
0.01 %
0.011 %
Page 12 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Measure
(cont)
(7 to 22) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
90 V/V
40 V/V
30 V/V
30 V/V
30 V/V
80 V/V
80 V/V
0.011 %
(22 to 70) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
40 V/V
40 V/V
30 V/V
40 V/V
40 V/V
0.01 %
0.032 %
0.092 %
(70 to 220) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
90 V/V
50 VV
30 VV
40 V/V
50 V/V
0.011 %
0.053 %
(220 to 700) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
0.01 %
60 V
40 V/V
30 V/V
40 V/V
(700 to 1000) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
90 V/V
50 V/V
30 V/V
40 V/V
60 V/V
10 Hz
20 Hz
30 Hz
40 Hz
52 A/A
56 A/A
56 A/A
51 A/A
AC Current Generate,
Fixed Points
100 A
Calibration of reference
multimeters
Page 13 of 45
Parameter/Range
Frequency
CMC2, 4, 6, 7 ()
Comments
AC Current Generate,
Fixed Points (cont)
100 A
55 Hz
300 Hz
1 kHz
5 kHz
10 kHz
20 kHz
30 kHz
51 A/A
68 A/A
57 A/A
67 A/A
79 A/A
0.012 %
0.019 %
1 mA
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
5 kHz
10 kHz
20 kHz
30 kHz
73 A/A
21 A/A
21 A/A
22 A/A
22 A/A
24 A/A
24 A/A
22 A/A
23 A/A
80 A/A
80 A/A
10 mA
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
5 kHz
10 kHz
20 kHz
30 kHz
66 A/A
37 A/A
44 A/A
28 A/A
36 A/A
36 A/A
29 A/A
38 A/A
55 A/A
0.01 %
0.01 %
100 mA
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
5 kHz
10 kHz
20 kHz
30 kHz
44 A/A
37 A/A
44 A/A
27 A/A
37 A/A
35 A/A
27 A/A
27 A/A
29 A/A
0.01 %
0.01 %
300 mA
10 kHz
20 kHz
30 kHz
84 A/A
0.01 %
0.015 %
Calibration of reference
multimeters
Page 14 of 45
Parameter/Range
Frequency
CMC2, 4, 6, 7 ()
Comments
AC Current Generate,
Fixed Points (cont)
1A
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
5 kHz
53 A/A
41 A/A
51 A/A
32 A/A
43 A/A
44 A/A
32 A/A
37 A/A
10 A
10 Hz
20 Hz
30 Hz
40 Hz
55 Hz
300 Hz
1 kHz
5 kHz
10 kHz
20 kHz
0.015 %
0.011 %
0.012 %
65 A/A
87 A/A
87 A/A
66 A/A
67 A/A
69 A/A
0.022 %
300 A
10 Hz, 20 Hz
45 Hz, 1 kHz
5 kHz
10 kHz
30 kHz
0.01 %
75 A/A
0.02 %
0.025 %
0.05 %
3 mA
(10, 20) Hz
45 Hz, 1 kHz
5 kHz
10 kHz
30 kHz
0.01 %
65 A/A
85 A/A
0.01 %
0.015 %
Calibration of reference
multimeters
AC Current Measure,
Fixed Points5
MFC calibration
Page 15 of 45
Parameter/Range
Frequency
CMC2, 4, 6, 7 ()
Comments
AC Current Generate
& Measure
(1 to 200) A
10 Hz
(10 to 20) Hz
(20 to 30) Hz
(30 to 40) Hz
(40 to 55) Hz
(55 to 300) Hz
300 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) kHz
(20 to 30) kHz
75 A/A + 5.4 nA
78 A/A + 5.4 nA
78 A/A + 5.4 nA
74 A/A + 5.4 nA
74 A/A + 5.4 nA
87 A/A + 5.4 nA
78 A/A + 5.4 nA
0.012 % + 5.4 nA
0.031 % + 5.4 nA
0.034 % + 7.1 nA
0.053 % + 12 nA
200 A to 2 mA
10 Hz
(10 to 20) Hz
(20 to 30) Hz
(30 to 40) Hz
(40 to 55) Hz
(55 to 300) Hz
300 Hz to 1 kHz
86 A/A + 21 nA
50 A/A + 21 nA
50 A/A + 21 nA
50 A/A + 21 nA
50 A/A + 21 nA
51 A/A + 21 nA
51 A/A + 21 nA
200 A to 2 mA
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) kHz
(20 to 30) kHz
76 A/A + 21 nA
0.02 % + 21 nA
0.031 % + 51 nA
0.051 % + 100 nA
(2 to 20) mA
10 Hz
(10 to 20) Hz
(20 to 30) Hz
(30 to 40) Hz
(40 to 55) Hz
(55 to 300) Hz
300 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) kHz
(20 to 30) kHz
80 A/A + 210 nA
58 A/A + 210 nA
62 A/A + 210 nA
53 A/A + 210 nA
57 A/A + 210 nA
58 A/A + 210 nA
53 A/A + 210 nA
82 A/A + 210 nA
0.021 % + 210 nA
0.032 % + 510 nA
0.051 % + 1.1 A
(20 to 200) mA
10 Hz
(10 to 20) Hz
(20 to 30) Hz
(30 to 40) Hz
(40 to 55) Hz
(55 to 300) Hz
300 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) kHz
(20 to 30) kHz
63 A/A + 2.1 A
58 A/A + 2.1 A
63 A/A + 2.1 A
52 A/A + 2.1 A
58 A/A + 2.1 A
57 A/A + 2.1 A
52 A/A + 2.1 A
78 A/A + 2.1 A
0.02 % + 2.1 A
0.032 % + 5 A
0.051 % + 10 A
Direct measurements or
transfer techniques
using a Fluke 4950
Page 16 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Current Generate
& Measure (cont)
(2 to 20) A
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) kHz
0.031 % + 210 A
0.061 % + 210 A
0.11 % + 510 A
Direct measurements or
transfer techniques
using a Fluke 4950
(20 to 80) A
16 Hz
40 Hz
50 Hz
60 Hz
120 Hz
180 Hz
450 Hz
600 Hz
1.2 kHz
(1.8 to 4.8) kHz
(4.8 to 6.0) kHz
52 A/A
37 A/A
37 A/A
36 A/A
49 A/A
49 A/A
49 A/A
50 A/A
50 A/A
60 A/A
60 A/A
Fluke 6105A
(80 to 120) A
(16 to 450) Hz
850 Hz
1 kHz
(1.2 to 6) kHz
49 A/A
51 A/A
0.011 %
0.011 %
Fluke 52120A
transconductance
amplifier and Fluke
6105A
500 mA to 1 A
(10, 20) Hz
(40, 400) Hz
(1, 5, 10) kHz
0.015 %
90 A/A
90 A/A
(1 to 2) A
10 Hz
(20, 40, 400) Hz
(1, 5) kHz
10 kHz
0.012 %
0.011 %
0.011 %
0.012 %
(2 to 3) A
(10, 20) Hz
(40, 400) Hz
(1, 5, 10) kHz
0.015 %
0.010 %
0.010 %
(3 to 5) A
(10, 20) Hz
(40, 400) Hz
(1, 5, 10) kHz
0.018 %
0.011 %
0.011 %
(5 to 10) A
0.014 %
95 A/A
0.015 %
(10 to 20) A
0.017 %
0.013 %
0.016 %
AC Current Generate
& Measure5
Page 17 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Current Generate
& Measure
(5 to 10) mA8
(10, 20) Hz
(40, 400) Hz
0.013 %
75 A/A
(5 to 10) mA8
75 A/A
90 A/A
(10 to 20) mA
10 Hz
20 Hz
(0.04, 0.4, 1) kHz
(5, 10, 20) kHz
30 kHz
0.013 %
75 A/A
60 A/A
75 A/A
90 A/A
(20 to 30) mA
(10, 20) Hz
(0.04, 0.4, 1) kHz
(5, 10, 20) kHz
30 kHz
0.015 %
95 A/A
95 A/A
0.012 %
(30 to 50) mA
(10, 20) Hz
(0.04, 0.4, 1) kHz
(5, 10, 20) kHz
30 kHz
0.017 %
0.01 %
0.01 %
0.011 %
(10, 20) Hz
(0.04, 0.4, 1) kHz
(5, 10, 20) kHz
30 kHz
0.014 %
85 A/A
85 A/A
0.01 %
(10, 20) Hz
(0.04, 0.4, 1) kHz
(5, 10, 20) kHz
30 kHz
0.013 %
80 A/A
80 A/A
95 A/A
(200 to 300) mA
(10, 20) Hz
(0.04, 0.4, 1) kHz
(5, 10) kHz
(20, 30) kHz
0.015 %
90 A/A
90 A/A
0.015 %
(300 to 500) mA
(10, 20) Hz
(0.04, 0.4, 1) kHz
(5, 10, 20) kHz
30 kHz
0.016 %
0.01 %
0.01 %
0.012 %
(45 to 440) Hz
(45 to 440) Hz
0.32 % + 50 mA
0.35 % + 90 mA
AC Current Generate
(20 to 150) A Turns
(150 to 1000) A Turns
Parameter/Range
Frequency
CMC2, 4, 6, 7 ()
Page 18 of 45
Phase Generate
(0 to 360)
1 Hz to 6.25 kHz
(6.25 to 200) kHz
10 m
20 m
(5 to 10) Hz
10 Hz to 50 kHz
(50 to 100) kHz
(100 to 500) kHz
0.20
50 m
[50 + 2(fkHz 50)] m
[150 + 3.5(fkHz 100)] m
1 kHz
220 H/H
220 H/H
0.11 %
1 kHz
25 F/F
10 MHz
90 fF
50 fF
40 fF
70 fF
Standard capacitors
10 Hz to 10 kHz
10 Hz to 10 kHz
10 Hz to 3 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
(10 to 600) Hz
(10 to 300) Hz
(10 to 150) Hz
(10 to 120) Hz
(10 to 80) Hz
(0 to 50) Hz
(0 to 20) Hz
(0 to 6) Hz
0.38 % + 10 pF
0.38 % + 10 pF
0.38 % + 10 pF
0.19 % + 10 pF
0.19 % + 0.1 nF
0.19 % + 0.1 nF
0.19 % + 0.3 nF
0.19 % + 1 nF
0.19 % + 3 nF
0.19 % + 10 nF
0.3 % + 30 nF
0.34 % + 0.1 F
0.34 % + 0.3 F
0.34 % + 1 F
0.34 % + 3 F
Fluke 5520A
Phase Measure
(0 to 360)
(-180 to +180)
Inductance Measuring
Equipment
1 mH
100 mH
10 H
Capacitance Measure
(1 to 1000) nF
Andeen Hagerling
2500A option E
Capacitance Generate
5 pF
20 pF
29 pF
50 pF
Capacitance Generate
(0.19 to 0.3999) nF
(0.4 to 1.0999) nF
(1.1 to 3.2999) nF
(3.3 to 10.9999) nF
(11 to 32.9999) nF
(33 to 109.999) nF
(110 to 329.999) nF
330 nF to 1.09 999 F
(1.1 to 3.29 999) F
(3.3 to 10.9999) F
(11 to 32.9999) F
(33 to 109.999) F
(110 to 329.999) F
330 F to 1.09 999 mF
(1.1 to 3.2999) mF
Page 19 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
Capacitance Generate
(cont)
(0 to 2) Hz
(0 to 0.6) Hz
(0 to 0.2) Hz
0.34 % + 10 F
0.7 % + 30 F
1 % + 0.1 mF
Fluke 5520A
(100 to 300) pF
(300 to 600) pF
(0.6 to 1.0) nF
(1 to 3.0) nF
(10 to 30) nF
(100 to 300) nF
(1.0 to 3.0) F
(10 to 30) F
(100 to 300) F
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
100 Hz
100 Hz
100 Hz
100 Hz
0.7 %
0.4 %
0.17 %
0.11 %
0.06 %
0.06 %
0.06 %
0.08 %
0.11 %
Fixed Points
3.3 nF
33 nF
0.33 F
3.3 F
33 F
300 F
330 F
1100
1 kHz
1 kHz
100 Hz
100 Hz
100 Hz
50 Hz
50 Hz
50 Hz
0.13 %
0.08 %
0.08 %
0.08 %
0.08 %
0.11 %
0.13 %
0.29 %
(5 to 10) mV
50 kHz to 10 MHz
0.2 %
10 mV to 5.5 V
50 kHz to 10 MHz
0.1 %
5 mV
1%
1.1 %
1.2 %
1.3 %
1.4 %
9.9 mV
0.8 %
0.9 %
1%
1.2 %
(3.3 to 10.9999) mF
(11 to 32.9999) mF
(33 to 110) mF
Capacitance Measure,
Fluke 55XX Series
Calibrators
AC Voltage Flatness
Measure
Fluke 5790 AC
standard
Fixed Points
Page 20 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Flatness
Measure (cont)
10 mV
0.8 %
0.9 %
1%
1.2 %
1.3 %
0.8 %
0.9 %
1.0 %
1.2 %
1.3 %
0.7 %
0.8 %
0.9 %
1%
1.2 V
0.6 %
0.65 %
0.7 %
0.95 %
1%
0.85 %
0.75 %
0.7 %
1%
5.5 V
0.85 %
0.75 %
0.7 %
0.75 %
0.7 %
(10 to 50) Hz
50 Hz to 100 kHz
100 kHz to 2 MHz
(2 to 10) MHz
(10 to 15) MHz
(15 to 20) MHz
(20 to 26) MHz
(26 to 30) MHz
0.038 %
0.021 %
0.03 %
0.05 %
0.065 %
0.08 %
0.13 %
0.15 %
AC Voltage Flatness
Generate & Measure
(1 to 2.2) mV
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
Page 21 of 45
AC Voltage Flatness
Generate & Measure
(2.2 to 7) mV
(10 to 50) Hz
50 Hz to 100 kHz
(100 to 500) kHz
500 kHz to 2 MHz
(2 to 6) MHz
(6 to 10) MHz
(10 to 15) MHz
(15 to 17) MHz
(17 to 20) MHz
(20 to 23) MHz
(23 to 26) MHz
(26 to 28) MHz
(28 to 30) MHz
0.026 %
0.012 %
0.016 %
0.02 %
0.033 %
0.04 %
0.055 %
0.062 %
0.07 %
0.1 %
0.11 %
0.13 %
0.14 %
(7 to 22) mV
(10 to 50) Hz
50 Hz to 100 kHz
(100 to 500) kHz
500 kHz to 1 MHz
(1 to 2) MHz
(2 to 4) MHz
(4 to 8) MHz
(8 to 9) MHz
(9 to 10) MHz
(10 to 12) MHz
(12 to 15) MHz
(15 to 17) MHz
(17 to 20) MHz
(20 to 23) MHz
(23 to 26) MHz
(26 to 28) MHz
(28 to 30) MHz
0.022 %
0.01 %
0.015 %
0.018 %
0.019 %
0.03 %
0.035 %
0.036 %
0.04 %
0.05 %
0.052 %
0.06 %
0.065 %
0.09 %
0.11 %
0.12 %
0.13 %
(22 to 70) mV
(10 to 50) Hz
50 Hz to 100 kHz
(100 to 700) kHz
700 kHz to 1 MHz
(1 to 2) MHz
(2 to 4) MHz
(4 to 6) MHz
(6 to 10) MHz
(10 to 12) MHz
(12 to 15) MHz
(15 to 17) MHz
(17 to 20) MHz
(20 to 23) MHz
(23 to 26) MHz
(26 to 28) MHz
(28 to 30) MHz
0.021 %
0.009 %
0.015 %
0.017 %
0.018 %
0.027 %
0.029 %
0.035 %
0.047 %
0.05 %
0.055 %
0.06 %
0.09 %
0.11 %
0.12 %
0.13 %
Page 22 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Flatness
Generate & Measure
(cont)
(70 to 220) mV
(10 to 50) Hz
50 Hz to 100 kHz
(100 to 500) kHz
500 kHz to 2 MHz
(2 to 4) MHz
(4 to 6) MHz
(6 to 8) MHz
(8 to 10) MHz
(10 to 12) MHz
(12 to 15) MHz
(15 to 17) MHz
(17 to 20) MHz
(20 to 23) MHz
(23 to 26) MHz
(26 to 28) MHz
(28 to 30) MHz
0.02 %
0.009 %
0.014 %
0.017 %
0.025 %
0.027 %
0.031 %
0.032 %
0.045 %
0.046 %
0.055 %
0.06 %
0.09 %
0.1 %
0.11 %
0.12 %
(220 to 700) mV
(10 to 50) Hz
50 Hz to 100 kHz
(100 to 500) kHz
500 kHz to 1 MHz
(1 to 2) MHz
(2 to 4) MHz
(4 to 6) MHz
(6 to 10) MHz
(10 to 12) MHz
(12 to 15) MHz
(15 to 17) MHz
(17 to 20) MHz
(20 to 23) MHz
(23 to 26) MHz
(26 to 28) MHz
(28 to 30) MHz
0.016 %
0.008 %
0.014 %
0.016 %
0.017 %
0.024 %
0.026 %
0.03 %
0.042 %
0.045 %
0.05 %
0.06%
0.09 %
0.1 %
0.11 %
0.12 %
700 mV to 2.2 V
(10 to 50) Hz
50 Hz to 100 kHz
(100 to 500) kHz
500 kHz to 1 MHz
(1 to 2) MHz
(2 to 4) MHz
(4 to 6) MHz
(6 to 9) MHz
(9 to 10) MHz
(10 to 12) MHz
(12 to 15) MHz
(15 to 17) MHz
(17 to 20) MHz
(20 to 23) MHz
0.016 %
0.0075 %
0.013 %
0.016 %
0.017 %
0.023 %
0.026 %
0.028 %
0.03 %
0.04 %
0.045 %
0.05 %
0.06 %
0.09 %
Page 23 of 45
Parameter/Range
Frequency
CMC2, 4, 5, 6, 7 ()
Comments
AC Voltage Flatness
Generate & Measure
(cont)
700 mV to 2.2 V
0.1 %
0.11 %
0.12 %
(2.2 to 7) V
(10 to 50) Hz
50 Hz to 100 kHz
(100 to 500) kHz
500 kHz to 2 MHz
(2 to 6) MHz
(6 to 9) MHz
(9 to 10) MHz
(10 to 12) MHz
(12 to 15) MHz
(15 to 17) MHz
(17 to 20) MHz
(20 to 23) MHz
(23 to 26) MHz
(26 to 28) MHz
(28 to 30) MHz
0.015 %
0.007 %
0.013 %
0.015 %
0.025 %
0.027 %
0.03 %
0.04 %
0.045 %
0.05 %
0.06 %
0.09 %
0.1 %
0.11 %
0.13 %
Parameter/Equipment
Range
CMC2, 4, 7 ()
Comments
Electrical Calibration of
Thermocouple
Indicators
Type B
(600 to 800) C
(800 to 1000) C
(1000 to 1550) C
(1550 to 1820) C
0.44 C
0.34 C
0.3 C
0.33 C
Type C
(0 to 150) C
(150 to 650) C
(650 to 1000) C
(1000 to 1800) C
(1800 to 2316) C
0.3 C
0.26 C
0.31 C
0.5 C
0.84 C
Type E
(-250 to -100) C
(-100 to -25) C
(-25 to 350) C
(350 to 650) C
(650 to 1000) C
0.5 C
0.16 C
0.14 C
0.16 C
0.21 C
Fluke 5520A
Page 24 of 45
Parameter/Equipment
Range
CMC2, 4, 7 ()
Comments
Electrical Calibration of
Thermocouple
Indicators (cont)
Type J
(-210 to -100) C
(-100 to -30) C
(-30 to 150) C
(150 to 760) C
(760 to 1200) C
0.27 C
0.16 C
0.14 C
0.17 C
0.23 C
Type K
(-200 to -100) C
(-100 to -25) C
(-25C to 120) C
(120 to 1000) C
(1000 to 1372) C
0.33 C
0.18 C
0.16 C
0.26 C
0.4 C
Type N
(-200 to -100) C
(-100 to -25) C
(-25 to 120) C
(120 to 410) C
(410 to 1300) C
0.4 C
0.22 C
0.19 C
0.18 C
0.27 C
Type R
(0 to 250) C
(250 to 400) C
(400 to 1000) C
(1000 to 1767) C
0.57 C
0.35 C
0.33 C
0.4 C
Type S
(0 to 250) C
(250 to 1000) C
(1000 to 1400) C
(1400 to 1767) C
0.47 C
0.36 C
0.37 C
0.46 C
Type T
(-250 to -150) C
(-150 to 0) C
(0 to 120) C
(120 to 400) C
0.63 C
0.24 C
0.16 C
0.14 C
Type U
(-200 to 0) C
(0 to 600) C
0.56 C
0.27 C
(-200 to -80) C
(-80 to 0) C
(0 to 100) C
(100 to 300) C
(300 to 400) C
(400 to 630) C
(630 to 800) C
0.04 C
0.05 C
0.05 C
0.04 C
0.04 C
0.05 C
0.05 C
Fluke 5520A
Electrical Calibration of
Thermistor & RTDs
Pt 385, 100
Fluke 525B
Page 25 of 45
Parameter/Equipment
Range
CMC2, 4, 7 ()
Comments
Electrical Calibration of
Thermistor & RTDs
(cont)
Pt 3926, 100
(-200 to -80) C
(-80 to 0) C
(0 to 100) C
(100 to 300) C
(300 to 400) C
(400 to 630) C
0.04 C
0.04 C
0.04 C
0.04 C
0.04 C
0.05 C
Pt 3916, 100
(-200 to -190) C
(-190 to -80) C
(-80 to 0) C
(0 to 100) C
(100 to 260) C
(260 to 300) C
(300 to 400) C
(400 to 600) C
(600 to 630) C
0.03 C
0.04 C
0.04 C
0.04 C
0.04 C
0.04 C
0.04 C
0.05 C
0.05 C
Pt 385, 200
(-200 to -80) C
(-80 to 0) C
(0 to 100) C
(100 to 260) C
(260 to 300) C
(300 to 400) C
(400 to 600) C
(600 to 630) C
0.38 C
0.38 C
0.39 C
0.39 C
0.43 C
0.43 C
0.5 C
0.5 C
Pt 385, 500
(-200 to -80) C
(-80 to 0) C
(0 to 100) C
(100 to 260) C
(260 to 300) C
(300 to 400) C
(400 to 600) C
(600 to 630) C
0.15 C
0.15 C
0.16 C
0.17 C
0.17 C
0.18 C
0.19 C
0.19 C
Pt 385, 1000
(-200 to -80) C
(-80 to 0) C
(0 to 100) C
(100 to 260) C
(260 to 300) C
(300 to 400) C
(400 to 600) C
(600 to 630) C
0.07 C
0.08 C
0.08 C
0.08 C
0.09 C
0.09 C
0.09 C
0.09 C
Ni 120, 120
(-80 to 0) C
(0 to 100) C
(100 to 260) C
0.02 C
0.02 C
0.02 C
Fluke 525B
Page 26 of 45
Parameter/Equipment
Range
CMC2, 4, 6. 7 ()
Comments
Electrical Calibration of
Thermistor & RTDs
(cont)
Cu 427, 10
(-100 to 260) C
0.38 C
YSI 400
(15 to 50) C
0.007 C
(0 to 6.6) V
(0 to 130) V
0.25 % + 40 V
0.05 % + 40 V
50
(0 to 6.6) V
(10 Hz to 10 kHz)
0.25 % + 40 V
1 M
(0 to 130) V
(10 Hz to 10 kHz)
0.1 % + 40 V
5 mV to 5.5 V
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
1.5 % + 100 V
2 % + 100 V
4 % + 100 V
5 % + 100 V
(2 to 5) ns
10 ns
(20 to 50) ns
100 ns to 20 ms
50 ms to 5 s
2.5 s/s
2.5 s/s
2.5 s/s
2.5 s/s
(25 + 1000t) s/s
1 kHz to 10 MHz
Frequency
CMC2, 4, 6, 7 ()
Fluke 525B
Oscilloscopes9
DC:
50
1 M
Fluke 5520A/SC1100
Square Wave:
Frequency
Parameter/Range
Fluke 5520A/SC1100
t = time in seconds
Comments
AC Resistance
Generate & Measure
(25 to 500) m
(1 to 180)
500 1.8 k
60 Hz
60 Hz
60 Hz
0.31 %
0.12 %
0.08 %
Calibration of Fluke
5320A (ground bond
resistance decade
f=60Hz)
10 k to 5M
60 Hz
0.2 %
Calibration of Fluke
5320A load
Page 27 of 45
Parameter/Range
Frequency
CMC2, 4, 6, 7 ()
Comments
AC Resistance Generate
& Measure (cont)
1
5
10 to 1 k
10 k
100 k
2 M
100 M
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
0.085 %
0.027 %
0.025 %
0.027 %
0.025 %
0.07 %
3%
1
5
10 to 100 k
2 M
100 Hz
100 Hz
100 Hz
100 Hz
0.25 %
0.055 %
0.05 %
0.2 %
Frequency
CMC2, 4, 6, 7 ()
RF Power
1 mW
50 MHz
0.15 %
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
0.002 dBm
0.002 dBm
0.002 dBm
0.002 dBm
0.017 dBm
0.017 dBm
0.018 dBm
0.023 dBm
0.03 dBm
0.03 dBm
Comments
Tegam 1830 HP
478H76
RF Flatness Measure
50 :
(+24 to +20) dBm
Measurement of 96x0
series RF generators
Page 28 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
50 :
(+20 to 0) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.002 dBm
0.002 dBm
0.002 dBm
0.002 dBm
0.011 dBm
0.011 dBm
0.012 dBm
0.019 dBm
0.027 dBm
0.027 dBm
0.027 dBm
0.04 dBm
0.04 dBm
0.058 dBm
0.058 dBm
0.066 dBm
0.067 dBm
0.12 dBm
0.12 dBm
(0 to -10) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.002 dBm
0.002 dBm
0.002 dBm
0.002 dBm
0.024 dBm
0.023 dBm
0.023 dBm
0.03 dBm
0.029 dBm
0.034 dBm
0.034 dBm
0.046 dBm
0.046 dBm
0.065 dBm
0.065 dBm
0.072 dBm
0.072 dBm
0.087 dBm
0.12 dBm
Measurement of 96x0
series RF generators
Page 29 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
50 :
(-10 to -20) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.003 dBm
0.003 dBm
0.003 dBm
0.003 dBm
0.024 dBm
0.023 dBm
0.023 dBm
0.03 dBm
0.029 dBm
0.034 dBm
0.034 dBm
0.046 dBm
0.046 dBm
0.065 dBm
0.065 dBm
0.072 dBm
0.072 dBm
0.087 dBm
0.12 dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.003 dBm
0.003 dBm
0.003 dBm
0.004 dBm
0.022 dBm
0.021 dBm
0.021 dBm
0.028 dBm
0.027 dBm
0.032 dBm
0.033 dBm
0.045 dBm
0.045 dBm
0.062 dBm
0.062 dBm
0.07 dBm
0.07 dBm
0.085 dBm
0.12 dBm
Measurement of 96x0
series RF generators
Page 30 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
50 :
(-30 to -40) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.004 dBm
0.004 dBm
0.004 dBm
0.005 dBm
0.022 dBm
0.021 dBm
0.021 dBm
0.028 dBm
0.028 dBm
0.033 dBm
0.033 dBm
0.045 dBm
0.045 dBm
0.062 dBm
0.062 dBm
0.07 dBm
0.07 dBm
0.085 dBm
0.12 dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.004 dBm
0.004 dBm
0.004 dBm
0.006 dBm
0.024 dBm
0.023 dBm
0.023 dBm
0.029 dBm
0.029 dBm
0.034 dBm
0.034 dBm
0.046 dBm
0.046 dBm
0.063 dBm
0.063 dBm
0.07 dBm
0.07 dBm
0.086 dBm
0.12 dBm
Measurement of 96x0
series RF generators
Page 31 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
50 :
(-48 to -58) dBm
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.024 dBm
0.024 dBm
0.03 dBm
0.03 dBm
0.03 dBm
0.03 dBm
0.034 dBm
0.034 dBm
0.038 dBm
0.038 dBm
0.049 dBm
0.05 dBm
0.066 dBm
0.066 dBm
0.078 dBm
0.078 dBm
0.092 dBm
0.12 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.042 dBm
0.042 dBm
0.042 dBm
0.035 dBm
0.04 dBm
0.039 dBm
0.043 dBm
0.043 dBm
0.064 dBm
0.078 dBm
0.11 dBm
0.14 dBm
0.15 dBm
0.18 dBm
0.18 dBm
0.2 dBm
Measurement of 96x0
series RF generators
Page 32 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
50 :
(-68 to -78) dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.045 dBm
0.038 dBm
0.038 dBm
0.038 dBm
0.042 dBm
0.041 dBm
0.045 dBm
0.045 dBm
0.065 dBm
0.08 dBm
0.11 dBm
0.14 dBm
0.15 dBm
0.18 dBm
0.18 dBm
0.2 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.11 dBm
0.11 dBm
0.11 dBm
0.092 dBm
0.064 dBm
0.064 dBm
0.067 dBm
0.067 dBm
0.077 dBm
0.1 dBm
0.13 dBm
0.16 dBm
0.18 dBm
0.18 dBm
0.19 dBm
0.2 dBm
Measurement of 96x0
series RF generators
Page 33 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
50 :
(-88 to -98) dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
0.15 dBm
0.11 dBm
0.076 dBm
0.061 dBm
0.063 dBm
0.063 dBm
0.065 dBm
0.065 dBm
0.086 dBm
0.11 dBm
0.13 dBm
0.15 dBm
0.17 dBm
0.19 dBm
0.12 dBm
0.21 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
0.21 dBm
0.13 dBm
0.13 dBm
0.12 dBm
0.1 dBm
0.01 dBm
0.11 dBm
0.12 dBm
0.21 dBm
0.21 dBm
0.23 dBm
0.24 dBm
0.26 dBm
0.26 dBm
0.28 dBm
Measurement of 96x0
series RF generators
Page 34 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
50 :
(-108 to -118) dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
0.19 dBm
0.19 dBm
0.19 dBm
0.17 dBm
0.16 dBm
0.16 dBm
0.16 dBm
0.39 dBm
0.34 dBm
0.41 dBm
0.43 dBm
0.43 dBm
0.44 dBm
0.45 dBm
0.49 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
0.47 dBm
0.47 dBm
0.47 dBm
0.21 dBm
0.21 dBm
0.21 dBm
0.26 dBm
0.48 dBm
0.48 dBm
0.48 dBm
0.49 dBm
0.49 dBm
0.49 dBm
0.49 dBm
0.55 dBm
Measurement of 96x0
series RF generators
Page 35 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
75 :
(+18 to +11) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
0.002 dBm
0.002 dBm
0.002 dBm
0.002 dBm
0.023 dBm
0.023 dBm
0.023 dBm
0.029 dBm
0.029 dBm
0.029 dBm
0.031 dBm
0.045 dBm
0.045 dBm
0.066 dBm
(+11 to 0) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.002 dBm
0.002 dBm
0.002 dBm
0.002 dBm
0.02 dBm
0.02 dBm
0.02 dBm
0.027 dBm
0.027 dBm
0.027 dBm
0.029 dBm
0.044 dBm
0.044 dBm
0.064 dBm
0.068 dBm
0.13 dBm
0.13 dBm
Measurement of 96x0
series RF generators
Page 36 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure,
(cont)
75 :
(+0 to -10) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.002 dBm
0.002 dBm
0.002 dBm
0.003 dBm
0.02 dBm
0.02 dBm
0.02 dBm
0.027 dBm
0.027 dBm
0.027 dBm
0.029 dBm
0.044 dBm
0.044 dBm
0.065 dBm
0.068 dBm
0.13 dBm
0.13 dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.003 dBm
0.003 dBm
0.003 dBm
0.003 dBm
0.021 dBm
0.021 dBm
0.021 dBm
0.027 dBm
0.027 dBm
0.027 dBm
0.029 dBm
0.044 dBm
0.044 dBm
0.065 dBm
0.068 dBm
0.11 dBm
0.11 dBm
Measurement of 96x0
series RF generators
Page 37 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
75 :
(-20 to -30) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.003 dBm
0.003 dBm
0.003 dBm
0.004 dBm
0.019 dBm
0.019 dBm
0.019 dBm
0.026 dBm
0.026 dBm
0.026 dBm
0.028 dBm
0.043 dBm
0.043 dBm
0.062 dBm
0.067 dBm
0.11 dBm
0.11 dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.004 dBm
0.004 dBm
0.004 dBm
0.005 dBm
0.02 dBm
0.02 dBm
0.02 dBm
0.026 dBm
0.026 dBm
0.026 dBm
0.028 dBm
0.043 dBm
0.043 dBm
0.062 dBm
0.067 dBm
0.11 dBm
0.11 dBm
Measurement of 96x0
series RF generators
Page 38 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
75 :
(-40 to -43) dBm
0.001 MHz
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.005 dBm
0.005 dBm
0.005 dBm
0.005 dBm
0.02 dBm
0.02 dBm
0.02 dBm
0.026 dBm
0.026 dBm
0.026 dBm
0.028 dBm
0.043 dBm
0.043 dBm
0.062 dBm
0.067 dBm
0.11 dBm
0.11 dBm
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.043 dBm
0.043 dBm
0.043 dBm
0.046 dBm
0.046 dBm
0.046 dBm
0.049 dBm
0.049 dBm
0.049 dBm
0.05 dBm
0.06 dBm
0.06 dBm
0.088 dBm
0.09 dBm
0.13 dBm
0.13 dBm
Measurement of 96x0
series RF generators
Page 39 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
75 :
(-55 to -65) dBm
0.02 MHz
0.075 MHz
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.045 dBm
0.045 dBm
0.045 dBm
0.047 dBm
0.047 dBm
0.047 dBm
0.05 dBm
0.05 dBm
0.05 dBm
0.052 dBm
0.061 dBm
0.061 dBm
0.089 dBm
0.09 dBm
0.13 dBm
0.13 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.046 dBm
0.049 dBm
0.049 dBm
0.049 dBm
0.052 dBm
0.052 dBm
0.052 dBm
0.053 dBm
0.063 dBm
0.063 dBm
0.09 dBm
0.091 dBm
0.13 dBm
0.13 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.067 dBm
0.069 dBm
0.069 dBm
0.069 dBm
0.071 dBm
0.071 dBm
0.071 dBm
0.072 dBm
0.079 dBm
0.079 dBm
0.13 dBm
0.13 dBm
0.15 dBm
0.15 dBm
Measurement of 96x0
series RF generators
Page 40 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
75 :
(-85 to -95) dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.07 dBm
0.072 dBm
0.072 dBm
0.072 dBm
0.074 dBm
0.074 dBm
0.074 dBm
0.075 dBm
0.082 dBm
0.082 dBm
0.13 dBm
0.13 dBm
0.16 dBm
0.16 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.13 dBm
0.14 dBm
0.13 dBm
0.13 dBm
0.13 dBm
0.13 dBm
0.13 dBm
0.22 dBm
0.22 dBm
0.22 dBm
0.23 dBm
0.23 dBm
0.24 dBm
0.24 dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.44 dBm
0.44 dBm
0.44 dBm
0.46 dBm
0.46 dBm
0.47 dBm
0.47 dBm
Measurement of 96x0
series RF generators
Page 41 of 45
Parameter/Range
Frequency
CMC2, 4, 7 ()
Comments
RF Flatness Measure
(cont)
75 :
(-115 to -125) dBm
0.1 MHz
0.3 MHz
1 MHz
10 MHz
20 MHz
100 MHz
125 MHz
300 MHz
750 MHz
1000 MHz
1400 MHz
2000 MHz
2500 MHz
3000 MHz
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.25 dBm
0.44 dBm
0.44 dBm
0.44 dBm
0.46 dBm
0.46 dBm
0.47 dBm
0.47 dBm
Measurement of 96x0
series RF generators
50
0.06 dB
0.04 dB
0.03 dB
0.04 dB
0.1 dB
0.18 dB
0.2 dB
75
0.06 dB
0.04 dB
0.03 dB
0.06 dB
0.1 dB
VSWR
Page 42 of 45
IV. Mechanical
Parameter/Equipment
Gauge Pressure
(Pneumatic)
Nitrogen
Gauge Pressure10
(Pneumatic)
Air
Absolute Pressure
(Pneumatic)
Nitrogen
Gauge Pressure
(Hydraulic)
Range
CMC2, 6 ()
Comments
0.0016 % + 0.31 Pa
Fluke PG7601
differential mode
(0 to <7) kPa
(7 to 9) kPa
(9 to 380) kPa
(380 to 720) kPa
(720 to 7,000) kPa
(7 to 10) MPa
0.0077 % + 0.55 Pa
0.009 % + 0.4 Pa
0.0016 % + 0.06 Pa
0.0017 % + 0.13 Pa
0.0017 % + 1.2 Pa
0.01 % but not less than
0.3 kPa
Ruska 7250xi
(0 to 750) Pa
(750 to 2833) Pa
(2833 to 7500) Pa
0.0075 % + 53 mPa
0.009 % + 37 mPa
0.0091 % + 14 mPa
Fluke 7250LP
Differential
(0.7 to 9) kPa
(9 to 380) kPa
(380 to 720) kPa
(720 to 7000) kPa
(7000 to 7100) kPa
(7.1 to 10) MPa
0.0077 % + 15 Pa
0.0016% + 0.21 Pa
0.0017% + 0.24 Pa
0.0017% + 1.2 Pa
0.0017% + 10 Pa
0.01 % but not less than
0.3 kPa
Ruska 7250xi
Fluke PG7601
(0 to 70) MPa
Fluke RPM4
Fluke PG7601
Fluke RPM4
Fluke RPM4
V. Thermodynamics
Parameter/Equipment
CMC2 ()
Range
Comments
Temperature Measure
& Measuring
Equipment
(233 to 773) K
0.05 K
Super-thermometer
(1595A) /PRT
reference thermometer
Humidity Measuring
Equipment
(15 to 90) % RH
1 % RH
Thunder Scientific
2500
Page 43 of 45
Parameter/Equipment
Infrared Radiation
Thermometry
Measuring Equipment
CMC2 ()
Range
(223 to 573) K
(573 to 1073) K
(1073 to 2973) K
0.3 K
1.8 x 10-3 Tm + 0.58 K
1.8 x 10-3 Tm + 0.59 K
Comments
CMC2 ()
Range
Comments
Tektronix 11801 digital
sampling oscilloscope
with SD-22 sampling
head
4 mV to 4 V
7.3 ps
5 parts in 1011
______________________________________________
1
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a
laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration
and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 %
level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a
specific calibration performed by the laboratory may be greater than the CMC due to the behavior of
the customers device and to influences from the circumstances of the specific calibration.
In the statement of CMC, L is the numerical value of the nominal length of the device measured in
inches.
The uncertainties shown relate to voltages and frequencies that lie within 10 % of the specified values.
For intermediate points between the discrete current/voltage/frequency pairs, the uncertainty will be the
greatest of the 4 adjacent voltage frequency/current frequency pairs, plus 10 %.
The measurands stated are generated with the Fluke 4900, 5500, 5700 and 700 series of instruments.
This capability is suitable for the calibration of the devices intended to measure the stated measurand in
the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a
fraction of the reading plus a fixed floor specification.
Page 44 of 45
In the statement of CMCs, percentages are percent of reading, unless otherwise indicated.
The uncertainty quoted on an A2LA certificate will include that of the reference standard plus any
uncertainties attributable to the device under test during calibration.
CMC values shown are for direct characterization and calibration of reference standard contributors
traceable to NIST calibrated AC/DC difference (current) standards and shunts (ref. BIPM KCDB. Pages
23 and 67).
The contributions from the best existing device are not included in the CMC claim for this parameter.
10
Page 45 of 45
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005 General Requirements for
the Competence of Testing and Calibration Laboratories. This laboratory also meets the requirements of ANSI/NCSL Z540-1-1994 and any
additional program requirements in the field of calibration. This accreditation demonstrates technical competence for a defined scope and the
operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqu dated 8 January 2009).
Presented this 6th day of June 2012.
_______________________
President & CEO
For the Accreditation Council
Certificate Number 2166.01
Valid to April 30, 2014
For the calibrations to which this accreditation applies, please refer to the laboratorys Calibration Scope of Accreditation.