The ADYN Amplifier is used to supply and balance measurement devices using strain gauges in
bridge configuration.
Limitation LED
Amp balancing
the strain gauges
bridge
Adjustable gain
potentionmeters
Balancing LED
(red/green)
Balancing
potentiometer
Power Switch
Differential
output (B)
Referenced
output (A)
Sensor
connector
Startup procedure :
Connect a digital voltmeter to the referenced output (A). On output A the voltage is
proportional to the torque (see Page 2 for the torque constants of the different
torquemeters).
Adjust the balancing potentiometer to have the smallest voltage (less that 5 to 10mV)
and the balancing LED should be turned off (no green or red display).
The limitation LED truns red when the ADYN amplifier is overloaded. This does not
mean that the torquemeter is overloaded.
1/2
- Gain 5000
- Sortie A
Applied torque :
N Torquemeter
Nominal Torque
Torque Constant
[Nm]
[Nm/V]
K 416
20
4.32
K 418
20
4.35
K 422
20
4.30
K 423
20
4.32
K 424
20
4.30
K 425
20
4.35
K 426
20
4.34
K 427
20
4.42
K 428
20
4.57
K 429
20
4.45
K 430
20
4.40
2/2