Академический Документы
Профессиональный Документы
Культура Документы
TRUFiL
TF/UM/27.1/01
Uncertainty calculations
For Electronic Weigh Balance
Date 11.01.2015
Calibrator Used
Name of the equipment Equipment no
Make
-
Laboratory Condition
Temperature - 26C
Humidity
36 %
Document Number
TRUFiL
TF/UM/27.1/01
Uncertainty calculations
For Electronic Weigh Balance
Type A
Measurement Data in grams.
(Measurement of 10 gms of standard Weight)
Serial No.
1
2
-0.00022
0.00038
3
4
5
10.0012
10.0015
10.0016
0.00018
-0.00012
-0.00022
Mean Reading
Standard Deviation
( x - X)
x = (X)/5.
{ (x-X) / 4}
[ { (x-X) 2 /4}]/5
(x - X)
4.84 x 10
1.44 x 10
3.24 x 10
1.44 x 10
4.84 x 10
10.00138 Gms.
= 2.68 x 10
= 1.19 x 10
Document Number
TRUFiL
TF/UM/27.1/01
Uncertainty calculations
For Electronic Weigh Balance
Type B
Include all uncertainties which are related to other factors including calibration
uncertainties and another factor.
Calibration Uncertainty At
Calibration Uncertainty of 10 gm
weight at 95% Confidence Level at
Normal Distribution i.e.(Ub = B*2)
c) Least count of Electronic Weigh
Balance is 0.0001 gms then For
Rectangular Distribution B/1.732
Value (B)
0.00004
Ub
0.00002
(Ub)2
4 x 10
0.0001
5.77 x 10-5
3.32 x 10-9
3.72 x 10-9
= 2.60 x 10
Result =
= 2.60 x 10 gms.
10.00138 Gms
Tested By -
Approved By Calibration-In-Charge
Technical Manager