Академический Документы
Профессиональный Документы
Культура Документы
1 | P a g e
Table
of
Contents:
Abstract
5
Introduction
..
6
Problem
Statement
....
7
Measurable
Objectives
....
7
House
of
Quality
...
7
Figure
1
..
8
Product
Design
Specification
9
Design
Alternatives
..
10
Function
Structure
Diagram
.
10
Figure
2
10
Figure
3
11
Physical
Decomposition
..
11
Figure
4
..
12
Morphological
Chart
..
13
Table
1
...
14
FALL
2014
MATERIAL
..
15
Preliminary
Designs
...
16
Summary
of
Construction
to
Date
.
17
Figure
5
.
17
Materials
of
Construction
..
17
Table
2
..
18
Graph
1
.20
Table
3
.
21
Figure
6
..
21
Graph
2
.
22
Graph
3
.
22
Control
System
.....
23
Figure
7
.
28
Efficiency
Factors
..
28
Figure
8
.
29
Graph
4
.
29
Figure
9
30
Table
4
.31
Table
5
.
31
Figure
10
31
Figure
11
31
Figure
12
32
Calculations
33
Final
Design
...
35
Figure
13
35
Methods
of
Analysis
...
36
Future
Schedule
..
38
2 | P a g e
Figure
14
38
Budget
...
39
Table
6
39
Bill
of
Materials
...
40
Table
7
40
Conclusion
..
41
SPRING
2015
MATERIAL
42
Final
Design
43
Figure
15
43
Figure
16
44
Materials
....
44
Construction
Methods
44
Figure
17
46
Figure
18
46
Control
System
46
List
of
parts
..
49
Table
8
49
Wiring
Schematic
...
50
Figure
19
50
Actual
Circuit
Board...
51
Figure
20
51
Testing
...
51
Figure
21
52
Graph
5
....
52
Graph
6
..
53
Graph
7
..
53
Fluids
Dynamics
54
Figure
22
54
Graph
8
..
55
Graph
9
..
56
Bacterial
Collection
...
56
Figure
23
58
Figure
24
..
58
Figure
25
58
Figure
26
58
Redesign
.....
58
Schedule
.....
59
Figure
27
59
Budget
.....
59
Table
9
60
Bill
of
Materials
......
61
Table
10
61
Conclusion
.....
62
References
..
63
Appendix
1:
Decision
Matrices
....
64
Table
11
..
64
3 | P a g e
Table
12
....
64
Table
13.
..
64
Table
14
..
64
Appendix
2:
Impact
.....
65
Figure
28
65
Figure
29
65
Figure
30
66
Appendix
3:
Impingement
..
67
Figure
31
67
Figure
32
67
Appendix
4:
Impact
+
Impingement
...
68
Figure
33
68
Appendix
5:
Inducted
Fan
..
69
Figure
34
69
Appendix
6:
Materials
..
70
Figure
35
70
Figure
36
71
Figure
37
72
Figure
38
73
Graph
10
74
Appendix
7:
Fluid
Dynamics
...
75
Table
15
.75
Appendix
8:
UAS-Hexacopter
......
76
Table
16
76
Graph
11
.
76
Graph
12
.
77
Table
17
77
Table
18
78
Figure
39
.
79
Table
19
80
Appendix
9:
List
of
FAA
Regulations
............
81
Appendix
10:
Control
System
...
82
Figure
40
82
Figure
41
83
Figure
42
84
Figure
43
85
Figure
44
86
Figure
45
87
Figure
46
88
Appendix
11:
Graphical
Analysis
....
89
Graph
13
89
Graph
14
89
Appendix
12:
Calculations
...
90
4 | P a g e
Abstract:
The
ACROBE
project
consists
of
a
remotely
operated
bacterial
collection
device
designed
for
the
purpose
of
furthering
aerial
bacteria
data
and
aiding
in
the
relatively
new
field
of
bio-precipitation.
Three
current
methods
exist
for
capturing
bio-aerosols,
but
these
methods
do
not
allow
for
the
control
of
variables
such
as
collecting
at
known
altitudes.
The
ACROBE
offers
high
repeatability
and
reproducibility
as
well
as
ease
of
use
for
potential
users.
The
ACROBE
will
rely
on
the
use
of
an
inducted
fan
that
creates
the
devices
vacuum
pressure.
Two
identical
impactors
will
allow
for
increased
sampling
volume
per
flight.
Servo
operated
doors
ensure
the
device
remains
sealed
until
collection
at
target
altitudes.
A
collection
plate
will
hold
industrial
grade
aluminum
foil
coated
with
a
film
layer
of
glycerol,
which
will
reduce
the
bounce
effect
of
the
bacteria.
Various
sensors
will
be
on
board
for
data
collection
of
weather
conditions.
To
ensure
optimal
positioning
of
the
device,
a
lightweight
wind
vane
can
be
used
in
combination
with
a
rotary
encoder.
The
device
will
be
predominately
constructed
of
carbon
fiber
due
to
its
low
density.
The
entire
system
will
be
mounted
on
an
Unmanned
Aerial
System
or
more
specifically
a
Hexacopter.
5 | P a g e
Introduction:
microbes
that
highly
organize
the
water
molecules
so
that
the
freezing
temperature
of
the
water
is
increased
by
7
C
[1].
These microbes
are
called
ice-nucleating
(IN)
bacteria,
and
they
are
known
for
being
plant
pathogens.
In
the
United
States,
nearly
one
billion
dollars
go
to
waste
potentially
due
to
these
ice-nucleating
bacteria
causing
frost
damage
to
the
crops
[2].
The
three
current
methods
for
studying
bacteria
in
the
atmosphere
include
the
HASP
(High
Altitude
Student
Platform),
hailstones
and
ice
core
samples.
The
purpose
of
the
HASP
is
to
carry
twelve
student
payloads
to
approximately
36km
high
with
the
duration
of
each
flight
lasting
between
15-20
hours
[3].
The
HASP,
however,
does
not
allow
the
determination
of
which
altitude
the
collection
of
bacteria
occurred.
Hailstones
are
studied
because
they
have
ice-nucleating
bacteria
at
their
cores.
The
problem
with
this
method
is,
in
order
to
collect
these
IN
bacteria,
one
has
to
wait
for
hail
and
then
wait
more
time
for
the
hailstone
to
melt.
Time
is
a
key
issue
for
studying
bacteria
through
the
use
of
hailstones.
Last,
the
ice
core
samples
are
taken
from
the
North
Pole.
This
method
of
collection
is
inconvenient
due
to
the
location
being
far.
Overall,
the
three
current
methods
allow
for
low
variable
control.
In
order
to
study
IN
bacteria,
a
device
for
microbe
collection
will
be
built
to
allow
scientists
to
learn
more
about
the
bacteria
and
where
it
is
located
in
the
atmosphere.
This
device
is
called
the
ACROBE.
6
|
P a g e
Problem
Statement:
The ACROBE is designed to develop an efficient device that will collect microbes in
Measurable
Objectives:
There are a total of four measurable objectives for ACROBE. After discussing with
Dr.
Brent
Christner,
a
microbiologist
here
at
Louisiana
State
University
who
recently
had
an
article
published
in
Nature
concerning
bio-precipitation,
the
device
should
filter
at
least
1
!
of
air
to
ensure
a
large
enough
sample
size.
Since
the
ACROBE
is
designed
to
determine
at
which
elevation
the
bacteria
exists,
the
ACROBE
should
collect
within
2
of
a
specified
altitude.
Finally,
the
hexacopter
has
the
ability
to
carry
up
to
15
,
so
the
ACROBE
will
weigh
less
than
10
.
House
of
Quality:
For the House of Quality, a scale of 1, 3 and 9 was used to rank the correlation
between
engineering
characteristics
and
customer
requirements
with
1
being
the
least
correlated
and
9
being
the
most
correlated.
The
customer
requirements
were
listed
on
the
vertical
axis
and
engineering
characteristics
on
the
horizontal
axis.
Next,
each
corresponding
box
was
given
a
ranking.
After
tabulating
the
results,
remote
operation
of
the
device
was
found
to
be
the
most
important
engineering
characteristic.
7 | P a g e
Figure
1:
House
of
Quality
for
the
ACROBE.
8 | P a g e
Product
Design
Specification:
The construction materials for the device will mainly be carbon fiber. In order to
make
the
ACROBE
easy
to
repair
and
disassemble,
the
design
will
be
modular.
The
device
will
also
be
designed
for
reproducibility
and
reliability
to
ensure
that
others
who
use
this
will
be
able
to
produce
the
same
outcomes.
Based
on
preliminary
testing
where
a
pressure
sensor
was
placed
in
a
tube
while
the
fan
ran
for
roughly
a
minute
at
approximately
!
20
! ,
the
operation
vacuum
pressure
was
found
to
be
near
0.03
!
or
206.84
! .
Since
the
device
will
be
flown
in
Louisiana,
over
crop
fields,
operation
temperatures
will
be
between
40
F
and
100
F
(4.44
C
and
37.78
C)
due
to
that
being
the
normal,
average
range
of
temperatures
year
round.
The
sampling
time
will
range
from
2.5
to
20
.
The
minimum
of
2.5
is
the
amount
of
time
it
takes
to
collect
exactly
1
3
while
14
minutes
is
the
maximum
amount
of
time
the
UAS-H
can
stay
in
flight.
For
cleaning
purposes,
70%
ethanol
and
2000
ppm
of
NaClO
will
be
used.
Sampling
will
be
done
in
compliance
with
FAA
regulations,
which
state
that
the
UAS-H
cannot
be
flown
over
400
from
the
ground
and
no
closer
than
5
to
an
airport.
9 | P a g e
Design
Alternatives:
Function
Structure
Diagram
The function structure diagram directly below shows that bacteria, electrical energy,
a
control
system
and
air
flow
will
enter
the
ACROBE
in
order
to
achieve
bacterial
collection.
Collected
bacteria,
mechanical
energy,
a
control
system
and
air
flow
will
be
outputs
of
the
device.
Figure
2:
The
above
picture
is
a
simplified
representation
of
the
Function
Structure
Diagram
for
the
device.
Electrical
energy
drives
the
control
system.
In
the
more
complex
function
structure
diagram,
Figure
3,
it
is
shown
that
the
ACROBE
is
driven
by
a
pressure
differential.
A
negative
feedback
exists
where
this
pressure
differential
controls
the
speed
of
the
DC
motor.
10 | P a g e
Figure
3:
This
is
a
more
complex
Function
Structure
Diagram
for
the
collection
device.
Physical
decomposition
The ACROBE can be divided into two main components: the impactor, of which
there
will
be
two,
and
the
electronics.
From
there,
each
of
these
two
components
are
further
decomposed
into
more
subunits.
Refer
to
Figure
4
for
a
more
detailed
description.
11 | P a g e
Figure
4:
Physical
decomposition
for
the
ACROBE.
12 | P a g e
Morphological
Chart
The morphological chart allowed us to determine the best solutions for each
component
of
the
design
before
the
detailed
design
stage
began.
For
bacterial
collection,
impact
was
ranked
the
best,
followed
by
impingement.
Originally,
both
impact
and
impingement
were
the
two
chosen
methods
of
bacterial
collection.
In
a
conversation
with
Dr.
Christner,
he
expressed
his
thoughts
on
impingement.
His
opinion
along
with
not
finding
many
papers
that
praised
impingement,
the
decision
to
completely
remove
impingement
from
the
device
and
to
only
have
two
impactors
was
made.
For
the
platform
that
would
be
used
to
carry
the
device,
a
UAS-H
outnumbered
the
rest
of
the
ideas
due
to
stability
during
flight
and
cost.
In
order
to
be
able
to
control
the
device
from
afar,
remote
and
automated
data
logging
was
the
best
solution.
Next,
for
bacterial
extraction,
agitation
of
the
foil
allows
the
glycerol,
which
contains
bacteria
to
dissolve
into
a
solution
of
choice.
For
the
differential
pressure
source,
an
inducted
fan
was
chosen.
Last,
for
cleaning
purposes,
70%
ethanol
and
NaClO
are
the
best
solutions
for
the
device.
The
strength
of
70%
ethanol,
in
particular,
was
chosen
because
any
percentage
higher
than
70
would
dry
out
the
cell
membrane
rather
than
killing
them.
Having
at
least
30%
of
water
allows
the
organism
to
remain
hydrated
enough
so
that
the
solution
can
penetrate
the
cell.
Once
the
solution
reaches
the
inside
of
the
cell,
it
will
denature
the
proteins
and
dissolve
the
lipid
bilayer
[4].
Ethanol
will
be
used
to
clean
the
outer
surface
of
the
device,
while
the
NaClO
will
be
used
to
bathe
the
components
of
the
device
that
come
into
direct
contact
with
the
bacteria.
13 | P a g e
Table
1:
This
morphological
chart
was
created
based
on
our
decision
matrices,
which
can
be
found
in
the
Appendix
1:
Decision
Matrices.
14 | P a g e
15 | P a g e
Preliminary
Designs
Our original design included both an impact and impingement system. The impact
was
the
first
method
of
bacterial
collection.
A
funnel
containing
a
grid
of
small
rods
would
have
been
coated
with
a
viscous
fluid
to
capture
the
bacteria.
After
further
thought,
the
impact
system
was
changed
to
having
a
coned
funnel
with
a
120
angle
that
would
cause
the
air
to
flow
through
the
inlet
and
come
into
contact
with
a
Trypticase
Soy
Agar
(TSA)
plate.
From
there,
a
hood
was
created
that
contained
two
tubes
to
allow
the
air
to
flow
into
the
impinger
after
the
impaction
stage.
The
purpose
of
this
change
was
based
on
Marple
and
Willekes
method
[5].
This
method
is
used
to
ensure
maximum
collection
efficiency.
The
pressure
differential
caused
unforeseen
problems
with
the
impingement.
Large
bubbles
were
created
which
could
potentially
lead
to
loss
of
bacteria.
To
help
this
problem,
we
created
a
grate
with
holes
with
0.0625
diameter
to
reduce
bubble
size.
The
impinger
was
changed
from
a
cube
to
a
cylinder.
At
the
same
time
in
the
design
process,
the
inlet
tube
was
moved
from
the
side
of
the
impinger
to
the
bottom.
The
geometry
and
position
changes
were
made
to
lower
the
amount
of
liquid
needed.
Even
though
a
grate
was
made,
it
did
not
solve
the
bubble
size
and
bacteria
loss
problem
to
the
extent
that
it
was
needed.
An
inverted
cone
was
placed
in
the
impinger
to
help
control
the
creation
of
bubbles,
their
location
and
their
size.
Original
designs
can
be
found
in
Appendix
2:
Impact,
Appendix
3:
Impingement
and
Appendix
4:
Impact
+
Impingement.
16 | P a g e
Figure
5:
Exploded
view
of
impactor
design.
Materials
of
Construction
Selection
of
materials
for
the
ACROBE
were
chosen
in
order
to
meet
physical
construction
and
biological
criteria.
Materials
were
chosen
based
on
primary
and
secondary
factors.
Primary
factors
include
low
density,
high
corrosion
resistance,
non-
cytotoxicity,
and
low
cost.
Secondary
factors,
such
as
smooth
surface
texture
properites,
17 | P a g e
high
electrical
resistivity,
and
high
strength,
were
considered
only
to
components
of
interest
that
were
constrained
to
unique
design
features.
Shown
below
is
a
table
of
physical
construction
materials
and
the
pertinant
characteristics
of
each:
18 | P a g e
allows
it
to
fill
any
void
spaces
between
two
mating
faces.
Aluminum
6061-T6
may
be
required
if
electrostatic
force
via
induction
become
problematic.
More
specifically,
the
tube
brackets
and
37
tubes
of
0.1
diameter,
shown
in
Figure
5,
will
be
aluminum
in
order
to
electrically
ground
them
to
the
Arduino
common
ground.
6061-T6
is
a
specific
alloy
of
aluminum
that
can
be
easily
machined
and
is
affordable
due
to
its
abundance.
Unfortunately,
6061-T6
will
oxidize
to
aluminum
oxide
(Al2O3)
when
treated
with
sodium
hypochlorite
(NaClO).
This
constraint
can
be
overcome
by
either
applying
a
surface
coating
(anodize)
or
using
marine
grade
aluminum,
which
better
resists
corrosive
conditions
[8].
Nylon
6,6
is
a
thermoplastic,
synthetic
polymer
that
will
be
incorporated
into
the
impactor
design
in
order
to
increase
modularity.
The
three
1.6
inch
nylon
bolts
and
six
hand
grip
style
nuts
shown
above
allow
for
complete
disassembly
of
the
hood
and
collection
disk
from
the
collector
base.
Other
material
options
for
bolts
and
hand
grip
style
nuts
come
with
a
better
strength
(secondary
factor)
but
at
the
cost
of
excess
weight
(primary
factor).
For
this
reason,
nylon
bolt
and
nut
assemblies
were
chosen
[9].
In
general,
construction
materials
were
down
selected
with
non-cytotoxicity
as
first
priority
followed
by
low
density.
Since
the
ACROBE
is
being
attached
to
a
hexacopter
platform,
the
weight
of
the
ACROBE
must
be
minimized
to
allow
for
longer
collection
times.
In
order
to
illustrate
the
dependence
of
weight
in
achieving
maximum
flight
times,
the
graph
below
was
generated:
19 | P a g e
Graph
1:
Overall
flight
time
of
the
UAS-H
(RCTx800
+
ACROBE)
is
dependent
on
the
weight
of
the
ACROBE
itself,
which
is
currently
unknown.
Assumptions
made
in
calculations
include
80%
battery
capacity,
constant
current
drain
(hovering
flight
only),
and
constant
battery
cell
voltage.
An
expected
ACROBE
weight
of
600g
yields
a
max
flight
time
of
15
minutes.
Three
major
biological
materials
were
selected
to
collect,
culture,
and
inhibit
the
growth
of
collected
bio-aerosols.
The
ability
to
encourage
or
disallow
cell
growth
before,
during
and
after
collection
was
a
chief
concern
in
biological
material
selection
because
there
is
an
inherent
time
delay
between
bio-aerosol
collection
and
analysis.
A
reduction
in
possible
change
in
sample
size,
due
to
post-collection
growth,
is
desired
in
order
to
maintain
representative
aero-environment
data.
Collected
bacteria
will
be
preserved
on
the
ACROBEs
collection
plate
via
a
film
layer
of
glycerol,
which
has
been
shown
in
preliminary
experiments
to
not
allow
for
cell
growth,
yet
sustain
cell
viability.
Glycerol
is
a
polar
compound,
making
it
miscible
in
other
aqueous
polar
solutions,
such
as
PBS
and
the
final
biological
material-
R2A.
Phosphate
Buffered
Saline
(PBS)
is
also
able
to
maintain
cell
viability
without
allowing
cell
reproduction;
however,
PBS
is
much
less
viscous
and
will
continue
to
flow
at
even
film
thicknesses.
If
quantification
tests
are
being
conducted,
PBS
will
be
used
as
the
solvent
for
the
bacteria.
Reasoners
2A
is
a
nutrient
broth
that
selects
20 | P a g e
for
and
cultures
slow
growing
bacteria,
like
the
prominent
bio-aerosol
Pseudomonas
syringae.
Substrate
(mg/L)
203.3
148.0
80.0
63.3
Time
6
8
11
13
1/S
(L/mg)
0.0049
0.0068
0.0125
0.0158
(1/hr)
0.1335
0.1054
0.0800
0.0690
max
(1/hr)
7.4889
8.4092
9.4912
10.4921
1/
(hr)
Km
(mg/L)
av
(1/hr)
td (hr)
0.1545
38.96
0.097
7.148
In
the
scenario
that
collected
bacteria
will
be
cultured
in
order
to
produce
a
larger
sample
size,
the
quantification
of
growth
rate,
cell
concentration
vs.
absorbance,
doubling
time,
half-saturation
coefficient,
and
substrate
utilization
of
P.
syringae
is
desired.
To
determine
this,
the
following
experimental
process
was
conducted,
which
yielded
the
relevant
biological
kinetics:
21 | P a g e
10
1/
(hr.)
8
6
4
2
0
-0.03
-0.02
-0.01
0
1/[S]
(L/mg)
0.01
0.02
5E+09
4.5E+09
4E+09
3.5E+09
3E+09
2.5E+09
2E+09
5
9
10
Time
(hr.)
11
12
13
14
Graph
3:
Cell
concentration
data
obtained
by
correlating
absorbance
data
with
cell
plating
results.
22 | P a g e
Control
Systems
The microbial collection device will optimally be used 400 in the atmosphere
during
collection.
Therefore,
the
device
needs
to
be
controlled
remotely
and
must
have
a
receiver
on
board
to
communicate
with
a
transmitter.
Furthermore,
the
device
needs
to
remain
sealed
until
sampling
begins.
Because
of
this,
the
two
entrance
and
one
exit
doors
will
need
to
be
servo
operated.
A
control
system
will
allow
the
servos
to
open
and
close
once
the
device
reaches
the
correct
height.
The
device
will
operate
at
multiple
temperatures,
varying
from
40
F
to
100
F.
Weight
will
be
one
of
the
most
important
factors
since
the
device
will
be
attached
to
a
hexacopter.
All
equipment
must
be
as
light
as
possible.
The
purpose
of
the
device
is
to
allow
scientists
to
study
IN
bacteria
effectively.
Consequently,
the
device
will
need
to
be
outfitted
with
a
range
of
sensors
for
weather
conditions
and
data
collecting
capability.
One
goal
of
the
ACROBE
is
to
have
high
collection
efficiency.
Moderately
high
winds
could
potentially
ruin
efficient
collection
if
the
device
is
not
facing
the
wind.
Therefore,
the
device
should
sense
which
direction
the
wind
is
blowing
and
allow
the
intake
to
face
that
direction.
The
on-board
fan
that
creates
the
vacuum
pressure
will
also
need
to
be
controlled
remotely
so
it
can
turn
on
and
off
during
testing.
Furthermore,
the
speed
of
the
fan
needs
to
be
monitored
and
controlled
for
continuous
efficient
collection.
Cost
is
a
key
factor,
so
the
cost
of
the
controls
must
remain
low.
The
batteries
must
last
at
least
20
to
allow
for
flight
time
to
and
from
the
collection
sight.
23 | P a g e
All
sensors
must
be
easily
detachable
as
well
because
the
device
will
be
modular
and
be
cleaned
frequently.
In
short,
the
constraints
can
be
listed
and
simplified:
o Cost
o Weight
o Temperature
o Sterile
o Efficient
Collection
o Battery
Life
o Ease
of
Use
o Data
Collection
o Remotely
Controlled
Arduino
boards
are
built
for
a
specific
project,
but
the
Arduino
Uno
and
Arduino
Mega
can
be
used
for
many
different
types
of
projects.
Between
these,
the
Uno
does
not
have
as
much
computing
power
or
I/O
pins.
However,
the
Uno
is
cheaper
and
lighter.
After
finding
all
the
sensors
needed
for
the
collection
device,
it
was
apparent
that
they
could
not
all
fit
and
be
used
on
the
Uno.
The
Arduino
Mega
would
need
to
be
used.
The
Arduino
Mega
was
chosen
as
the
microcontroller
for
the
microbial
collection
device.
The
Uno
is
based
on
the
Atmega2560
microcontroller.
It
has
16
analog
input
pins
and
54
digital
I/O
pins.
The
digital
I/O
pins
contain
15
pins
capable
of
PWM.
This
board
provides
enough
space
for
all
the
sensors
needed.
It
has
a
clock
speed
of
16
MHz
and
a
RAM
of
8
.
It
also
operates
on
5V
with
a
DC
current
requirement
of
40
mA
on
all
I/O
pins.
The
hardware
on
board
can
talk
Serial
(UART),
I2C
and
SPI.
Outside
the
chip,
the
Arduino
Mega
board
contains
voltage
regulators,
capacitive
components,
and
I/O
connectors.
The
code
written
in
the
integrated
development
environment
(IDE)
is
the
only
code
that
runs
on
the
chip.
There
is
no
interpreter,
operating
system,
or
firmware.
On
the
other
hand,
the
Raspberry
Pi
is
a
single
board
computer,
or
SBC.
On
board
it
contains
a
32
bit
microprocessor
and
supports
video,
audio,
USB
host,
ethernet,
SD
card,
HDMI
cord,
and
24 | P a g e
GPI
headers.
These
assets
mean
the
Raspberry
Pi
has
more
in
common
with
a
computer
than
an
Arduino.
Instead
of
writing
code
to
control
the
hardware
directly,
the
programs
written
are
actually
run
within
an
operating
system.
The
operating
system
is
typically
Linux.
Both
the
Arduino
and
Raspberry
Pi
contain
a
CPU,
timers,
memory,
as
well
as
I/O
pins.
The
key
difference
is
in
the
I/O
pins.
As
a
microcontroller,
the
Arduino
has
high
I/O
capability,
able
to
drive
external
hardware
directly.
On
the
other
hand,
the
Raspberry
Pi
is
a
microprocessor,
which
tend
to
have
weak
I/O,
needing
transistors
to
drive
most
hardware.
However,
microprocessors
are
effective
processors
containing
better
brain
power
than
a
microcontroller.
The
Raspberry
Pi
has
a
700
MH
clock
speed,
over
40
times
faster
than
the
Arduino.
It
is
also
based
on
a
32-bit
architecture,
compared
to
the
8-bit
of
the
Arduino.
The
Raspberry
pi
has
256,000
times
more
RAM
than
an
Arduino.
The
Arduino
has
more
I/O
pins
and
can
also
utilize
8
times
more
current
in
each
pin
than
the
Raspberry
Pi.
Finally,
the
Raspberry
Pi
typically
operates
on
Linux,
while
the
Arduino
has
no
operating
system.
The
Arduino
was
ultimately
chosen
for
ACROBE
for
its
capability
of
controlling
many
sensors
with
more
I/O
capability,
and,
its
simple,
effective
interface,
while
remaining
cost
friendly.
Modern wind direction sensors on the market are heavy, expensive, and not
designed
to
interface
with
the
Arduino.
The
final
design
selected
for
wind
sensing
was
the
rotary
encoder.
A
rotary
encoder
will
be
placed
in
between
the
wind
vane
and
intake.
Specifically,
the
EMS22P50-B28-LS6
from
Bourns.
It
has
low
friction
and
high
sensitivity.
The
rotary
encoder
will
determine
the
relative
position
of
the
intake
to
the
vane.
A
negative
feedback
loop
will
be
created
between
the
rotary
encoder
and
a
continuous
rotation
servo
attached
to
the
intake.
When
the
vane
is
not
lined
up
with
the
servo,
the
servo
will
turn.
25 | P a g e
Once
lined
up,
the
servo
will
stop.
A
potentiometer
could
have
been
used
to
determine
the
position
of
the
wind
vane,
however,
it
could
increase
friction
on
the
wind
vane.
The
more
friction
the
wind
vane
has,
the
less
sensitive
it
will
be
to
wind
speed.
The
hexacopters
position
also
needs
to
be
known,
and
a
potentiometer
cannot
be
used
for
this.
A
digital
compass
sensor
could
have
been
used,
but
many
of
them
would
have
to
be
used
to
orientate
the
position
of
the
vane,
intake,
and
hexacopter.
The speed of the fan drawing air through the device is imperative for efficient
collection
of
bacteria.
The
speed
of
the
air
moving
through
the
inlets
must
be
in
between
!
20-30
! .
The
final
design
will
have
a
pressure
sensor,
specifically
the
MPX4115a,
located
in
a
pitot
tube
near
the
end
of
the
system
to
measure
the
pressure
when
the
air
speed
in
the
!
inlets
is
20-30 ! .
The
pressure
value
taken
will
then
be
used
in
a
negative
feedback
loop
to
drive
the
fan.
If
the
pressure
ever
reaches
over
or
under
a
normal
value,
the
fan
will
either
reduce
or
increase
its
speed.
The
fan
will
be
controlled
by
an
S-12
ESC.
The
ESC
speaks
to
the
Arduino
using
servo
language,
and
thus
can
be
controlled
using
PPM
from
the
servo
library.
The
ESC
is
lightweight
and
contains
the
power
requirements
to
operate
the
DC
motor.
The
pressure
sensor
is
a
rugged,
temperature
compensated
and
cost
effective
sensor
that
functions
within
the
specified
pressure
and
temperature
range.
to
have
when
studying
cells
that
live
in
ambient
weather.
Humidity
is
related
to
temperature
change,
so
many
breakout
boards
will
read
humidity
as
well
as
temperature.
The
HIH
6130
breakout
board
from
Sparkfun
was
chosen.
It
operates
from
2.3
to
5.5
volts
and
is
temperature
compensated
from
5
to
50
C.
It
also
has
a
compensated
humidity
range
of
10%
to
90%
relative
humidity.
The
HIH
6130
speaks
to
the
Arduino
through
the
26 | P a g e
I2C
bus.
The
HIH
6130
will
be
placed
in
a
protective
location
from
debris
but
will
remain
open
to
the
environment
for
data
acquisition.
The solution for data acquisition needed to be lightweight and cost effective. Most
DAQs
are
extraordinarily
expensive,
heavy
and
consume
copious
amounts
of
space.
Adafruit
manufactures
an
SD
card
shield
that
outmatches
all
of
the
criteria.
Data
can
be
sent
to
the
lightweight
and
portable
SD
card,
ready
for
acquisition
upon
placing
the
card
into
a
computer.
The
shield
is
placed
on
top
of
the
Arduino,
saving
space
and
weight.
Two batteries will be used for the entire system. An 11.1V, 3S Lipo battery will
power
a
12A
ESC
and
the
DC
motor
used
to
drive
the
fan
that
causes
suction.
The
other
batteries
will
be
two
14.4V,
4S
Lipo
battery.
This
battery
will
power
the
six
20A
ESC
used
to
control
the
hexacopters
rotors.
In
addition,
the
5V
BEC
onboard
the
ESC
will
run
to
the
RX
receiver.
The
receiver
will
then
turn
on
the
Arduino
when
commanded
from
the
transmitter.
27 | P a g e
Wiring Schematic:
A cross sectional plane through the middle of the device is shown below in Figure 8.
It
is
color
coordinated
depicting
low,
medium
and
high
velocities,
respectively
blue,
green
and
red.
In
addition,
the
picture
confirms
that
air
impacts
with
the
plate
before
going
28 | P a g e
around
the
plate
and
exiting
through
the
outlet.
A
velocity
curve
through
the
middle
of
one
of
the
tubes
is
also
shown
in
Graph
4.
The
graph
shows
that
our
simulation
produced
maximum
velocities
at
the
center
of
the
tube.
The
maximum
velocity
is
approximately
!"
1095
!
(27.8
! ).
Figure
8:
A
cross
sectional
photo
displaying
velocity
flow
throughout
the
device.
Graph
4:
A
graph
of
the
velocity
at
the
middle
of
one
of
the
small
inlet
tubes.
29 | P a g e
Figure 9 below displays a cross-section of the 37 small inlet tubes in the design. The
design
is
symmetrical
across
both
the
x
and
y
axes,
giving
rise
to
tubes
that
are
spatially
equivalent,
which
are
color
coordinated.
Table
4
shows
the
average
Reynolds
numbers
and
velocities
of
each
set
of
equivalent
tubes.
Autodesk
Simulation
CFD
provided
the
flow
rate
through
each
small
tube
and
using
= ,
the
average
velocity
per
tube
was
calculated.
The
main
trend
within
this
data
is
that
velocity
and
Reynolds
number
increase
slightly
moving
toward
the
center.
The
percent
standard
deviations
were
calculated
for
each
set
of
tubes
for
both
velocity
and
Reynolds
number.
The
minimum
and
maximum
values
are
displayed
in
Table
4.
These
values
show
that
the
standard
deviations
are
fairly
small.
Figure
9:
Cross
sectional
photo
of
the
37
inlet
tubes,
which
are
color
coated
to
relate
spatially
equivalent
tubes.
30 | P a g e
Table
4:
Average
Reynolds
numbers
and
velocities
of
the
spatially
equivalent
tubes.
Table
5:
Range
of
percent
standard
deviations
for
Reynolds
numbers
and
velocities
for
spatially
equivalent
tubes.
picture
of
this
animation
is
below
and
to
the
left.
On
the
right,
velocity
vector
field
is
shown.
After
observing
the
velocity
vector
field,
airflow
behaves
as
predicted,
except
for
the
whirlwind
that
occurs
in
the
corners
of
the
device
in
front
of
the
collection
plate.
Possible
redesign
to
reduce
this
could
be
to
chamfer
those
edges
or
bring
the
collector
plate
closer
to
the
37
inlet
tubes.
For efficient bacterial collection, certain requirements are needed. First, the S/W
ratio
should
be
between
1
and
5.
The
S
value
is
the
distance
between
the
collection
plate
and
the
exit
of
the
inlet
tubes.
The
W
is
the
width
of
each
inlet
tube.
Second,
the
throat
length
must
be
greater
than
or
equal
to
the
nozzle
diameter
with
nozzle
diameter
also
being
W.
Third,
the
Reynolds
number
must
be
between
500
and
3000.
Fourth,
the
velocity
!
Figure
12:
Streamlines
and
particle
trajectories
for
a
typical
impactor.
32 | P a g e
Calculations:
50%collection
diameter
(Yao
and
Mainelis,
2007)
!" =
9/! ! ! * !"
= .
The
d50
equation
was
used
to
calculate
the
d50
specific
to
the
ACROBE.
At
2
m,
the
device
should
collect
50%
of
all
bacteria
with
this
diameter.
2 m
is
a
little
more
than
average
size
of
P.
syringae;
however,
we
can
theoretically
collect
up
to
16
times
the
minimum
1
m3
of
air
volume.
For
this
reason,
we
can
justify
collecting
less
than
50%
of
P.
syringae.
If
any
bacteria
are
collected
with
a
larger
diameter
than
2
m,
more
than
50%
of
bacteria
collected.
= . .
Volumetric
Flow
Rate
=
= .
33 | P a g e
Given
a
Reynolds
number
range
of
500
to
3000
based
on
Marple
and
Willekes
method,
we
calculated
the
smallest
diameter
for
the
inlet
tubes.
Using
3000
as
the
Reynolds
!
number
and
20
!
for
velocity
in
the
above
equations,
we
calculated
that
the
minimum
diameter
must
be
approximately
0.1
inch.
Using
this
diameter,
the
volumetric
flow
rate,
per
!
!!
!
!"!
!!"#
!"!"#
!
= 0.0608! 10
0.0608!
= 1!
1
=
Diameter
of
outlet
pipe
necessary
to
equal
of
37
tubes
!"! = = 37 5.067110!! ! = 1.87510!! !
!"! = 1.87510!! ! =
= .
!"# !
4
.
= .
Using
the
volumetric
flow
rate,
0.06078
m3
of
air
would
filter
through
per
tube,
per
ten
minutes.
In
order
to
filter
the
wanted
1
m3,
it
was
calculated
that
17
tubes
were
needed.
By
doubling
the
number
of
tubes
to
34,
we
can
increase
the
amount
of
air
filtered
through
34 | P a g e
the
device.
We
chose
to
have
37
tubes
in
each
impactor
in
order
to
create
symmetry
about
the
x
and
y
axes.
Final
Design:
The
final
design
includes
two
impactors
that
will
be
placed
on
top
of
the
hexacopter.
The
hexacopter
will
also
hold
the
Arduino
system,
a
wind
vain
and
a
control.
In
Figure
13,
you
can
see
the
ACROBE
mounted
on
top
of
the
hexacopter.
On
the
sides
of
the
hexacopter
are
two
covered
compartments;
one
housing
a
control
and
the
other
housing
the
electronics
for
the
device.
35 | P a g e
Methods
of
Analysis:
Following
bacterial
collection
using
the
ACROBE,
lab
analysis
is
required
to
determine
total
concentration
and
species
of
the
bacteria,
as
well
as,
the
viability
status.
To
prepare
the
bacteria
captured
in
glycerol
for
lab
analysis,
a
transfer
to
either
Phosphate
Buffered
Saline
(PBS)
or
Reasoners
2A
Agar
(R2A)
is
required.
PBS
is
a
balanced
salt
solution
used
often
in
biological
research.
The
main
function
is
to
maintain
pH
and
osmotic
balance,
as
well
as,
provide
cells
with
water
and
inorganic
ions.
PBS
can
be
used
to
maintain
cells
for
the
short
term
in
a
viable
condition
while
the
cells
are
manipulated
outside
the
regular
growth
environment
[10].
R2A
is
a
low
nutrient
medium.
When
R2A
use
is
combined
with
lower
incubation
temperature
and
longer
incubation
time,
it
stimulates
the
growth
of
stressed
bacteria.
Nutritionally
rich
media
support
the
growth
of
fast
growing
bacteria,
and
can
suppress
slow
growing
or
stressed
bacteria
[11].
36 | P a g e
37 | P a g e
Future Schedule:
Figure
14:
The
Gantt
chart
contains
deadlines
for
both
Fall
2014
and
Spring
2015
semesters.
By creating Gantt chart, the progress of this project has remained on track. Next
semester,
the
ACROBE
project
will
continue
moving
forward
in
order
to
build
and
test
while
redesigning
as
needed.
By
the
beginning
of
February,
we
would
like
to
have
all
control
system
parts
working
individually
to
ensure
that
all
components
function
properly
before
construction
of
the
final
design
begins.
Also,
all
other
materials
should
have
arrived
by
this
time.
The
beginning
of
April
is
when
the
final
prototype
should
be
finalized
and
built.
If
time
permits,
bacterial
collection
and
analysis
will
be
performed
as
additions
to
the
goal
of
the
project.
38 | P a g e
Budget:
Material
Propellor
Items
APC
13x10
APC
13x10
Adhesive
Epoxy
PVC
1-inx
5ft
Sch-40
PI
3/4
in
x
5
ft
Sch40
PVC
2
in
x
2ft
PVC
SCh
40
SOL
2
in
sch
40
cap
socket
1/2
in
x
2
ft
PVC
Sch
40
1/2
in
Sch40
cap
1
in
sch40
elbow
(3)
1
in
sch40
coupling(2)
3/4
in
sch40
elbow
(3)
3/4
in
sch40
adapter
3/4
in
sch40
adapter
1
in
sch40
adapter
1
in
x
3/4
in
adapter
8x10
clear
lexan
(3)
Unit
Fan
ss12
brushless
12
amp
esc
lipo
2s
7.4V
1200
30c
ammo
20-40
3500KV
brshless
mtr
hyperflow
370
ep
df
w/o
motor
Polycarbonate
Polycarbonate
tubing
1/4
ID
Polycarbonate
tubing
3/4
ID
Polycarbonate
tubing
7/8
ID
Polycarbonate
tubing
1"
ID
Polycarbonate
tubing
1/2"
ID
polycarbonate
tubing
3/4
ID
actron
vacuum
and
fuel
pressure
Vacuum
kit
Sensors
Temperature
and
Humidty
Sensor
ABS
Pressure
Sensor
(x3)
Total
Price
$7.99
$7.99
$15.99
$3.09
$1.79
$4.14
$2.08
$3.03
$0.97
$1.98
$0.92
$1.41
$1.86
$0.47
$0.77
$1.56
$12.72
$34.99
$16.99
$42.99
$21.99
$20.25
$33.65
$16.11
$13.67
$25.09
$29.14
$17.84
$29.95
$48.27
$423.68
Table 6: The budget containing costs from both Spring and Fall of 2014.
39 | P a g e
The above table shows that the total cost thus far is $423.86, but due to already
having
some
of
the
materials,
we
have
only
spent
$286.75
of
the
total
budget
of
$1,100.
At
the
end
of
this
current,
fall
semester,
about
$800
is
still
available
for
use
in
the
spring.
Bill
of
Materials:
Item
MPX
4115a
HH6130
Sparkfun
Arduino
Mega
EMS22P50-B28LS6
SD
Card
Shield
Adafruit
Futaba
ASD
Sensor
Futaba
BLS153
7.4
V
2S
LiPo
1200
mAh
Futaba
R7008SB
S.Bus2
2.4
GHz
FASSTest
Rx
CF
Rod
1"
x
60"
CF
Rod
.686"
x
72"
CF
Rod
.1"
x
24"
West
Systems
105/206
Epoxy
West
Systems
423
Graphite
Powder
CF
Plate
300mm
x
100mm
x
2mm
Frekote
700
Neoprene
with
Adhesive
54"
x
1'
x
.125"
6061-T6
Aluminum
4"
x
4"
x
.125"
Du-Bro
993
Nylon
Bolt
3"
x
.25"
Aluminum
Foil
18"
x
25'
x
.001"
Teknova
R2A
Broth
Premix
(500g)
Growcells
MRGF-6235-12
10X
PBS
(12L)
2000
ppm
NaHClO3
(1L)
SEOH
R2A
Agar
Media
(500g)
20
pk.
Petri
Dishes
100mm
x
13mm
Carbon
Fiber
Fabric
3K
50"
x
36"
Heathrow
Scientific
HS86655
Stainless
Steel
Large
Bacterial
Cell
Spreader
Total
Quantity
3
1
1
1
1
1
3
1
Combined
Price
$48.27
$29.95
$37.52
$37.80
$19.95
$179.97
$360.00
$16.99
1
1
1
2
1
1
2
1
1
2
1
1
1
1
1
1
10
1
$139.98
$75.99
$55.99
$4.90
$62.61
$19.03
39.99
$20.50
$20.00
$23.66
$2.36
$16.95
$176.57
$292.78
$14.76
$724.28
$75.52
$79.50
$50.85
40 | P a g e
The
Bill
of
Materials
above
includes
all
materials
needed
to
complete
the
ACROBE.
The
non-bolded
items
are
the
materials
that
are
needed
but
that
we
do
not
need
to
purchase
do
to
already
having
them
in
privately
owned
possession.
The
bolded
items
are
the
materials
that
still
need
to
be
purchased.
We
will
spend
a
total
of
$724.28
in
the
spring,
remaining
a
under
the
$1,100
budget.
However,
for
someone
who
plans
to
build
the
ACROBE
with
no
previously
owned
materials
should
expect
to
spend
about
$2,000.
Conclusion:
To end, the ACROBE is an ergonomic and easy to use design. It will allow for efficient
bacterial
collection
with
high
repeatability
as
well
as
reproducibility.
Multiple
runs
will
be
allowed
in
a
day.
Since
we
are
operating
at
high
altitudes,
the
device
will
be
fully
automated
allowing
control
of
the
ACROBE
from
a
distance.
Being
fully
automated,
the
device
is
essentially
a
fire
and
forget
system.
The
control
system
will
allow
for
data
collection
with
the
sensors
placed
within
the
device.
Because
the
ACROBE
will
be
able
to
collect
at
specified
altitudes,
it
is
the
first
of
its
kind.
41 | P a g e
42 | P a g e
Final
Design:
This semesters design has three major and impactful changes from the previous
design.
The
first
change
made
was
having
one
continuous
door
for
both
intakes,
rather
than
one
door
per
intake.
This
change
allows
for
one
servo
to
be
needed
instead
of
two
separate
servos
to
control
the
opening
and
closing
of
the
doors.
The
next
change
we
made
was
adding
flanges
to
seal
the
collectors
simultaneously
with
ease.
Last,
the
collection
disk
is
coated
with
silicone
spray
instead
of
glycerol.
This
change
was
made
due
to
the
glycerol
not
remaining
on
the
collection
plate
without
sliding
off
and
forming
beads.
43 | P a g e
Figure
16:
Picture
of
our
final
design
for
the
ACROBE.
Materials:
The physical and biological materials have not changed from last semester. Please
refer to page 17 for a complete list of the materials using for the ACROBE.
Construction
Methods:
A
variety
of
modern
construction
methods
were
used
in
both
the
prototype
design
and
the
revised,
final
design.
These
methods
include
3D
printing,
laser
cutting,
and
water
jet
CNC.
In
prototype
construction,
the
main
goal
was
to
produce
of
viable
model,
which
could
prove
the
mechanical
concepts
of
the
ACROBE,
such
as
the
rotation
and
fluid
flow
requirements.
Additionally,
the
prototype
was
needed
in
order
to
affirm
the
hexacopters
ability
to
fly
with
an
elevated
load
and
center
of
mass.
Because
the
prototype
was
not
required
to
withstand
chemical
treatment,
the
main
construction
methods
used
were
laser
44 | P a g e
cutting
and
3D
printing.
The
materials
for
these
methods
were
plywood
and
ABS
plastic,
respectively.
Polycarbonate
tubing
was
also
heavily
featured
in
the
prototype
design
as
a
pipe
flow
medium.
Transitioning
to
the
final
design,
it
saved
both
money
and
time
to
utilize
some
prototype
parts
and
pieces
as
a
mold
for
final
parts.
The
weight
of
the
prototype
ACROBE
was
approximately
1.11
kg
(2.45
lbs),
which
is
an
acceptable
weight
considering
our
design
objective
of
10
lbs.
By
using
molds
and
a
lighter
composite
carbon
fiber
material,
the
weight
of
the
ACROBE
was
reduced
25%
to
0.83
kg
(1.83
lbs).
Water
jet
CNC
machining
was
used
to
precision-make
aluminum
and
carbon
fiber
plate
parts.
The
CNC
machine
enabled
us
to
elevate
both
the
quality
of
construction
and
expanded
design
options
to
more
complex
geometries
that
would
otherwise
be
extremely
difficult
to
accomplish
with
traditional
methods.
Anodizing
the
aluminum
parts
was
necessary
for
long-term
successful
operation
of
the
ACROBE,
since
2,000
ppm
bleach
will
oxidize
bare
aluminum.
Basic
steps
for
anodizing
go
as
follows:
1.
2.
3.
4.
5.
6.
7.
8.
9.
45 | P a g e
Figure
17:
The
above
figure
is
a
picture
of
our
prototype.
Control
System:
The Arduino Mega was chosen as the microcontroller for the microbial collection device.
The Mega is based on the Atmega2560 microcontroller. It has 16 analog input pins, and 54
digital I/O pins. The digital I/O pins contain 15 pins capable of PWM. This board provides
enough space for all the sensors needed. It has a clock speed of 16 MHz, and a RAM of 8 kb. It
also operates on 5V with a DC current requirement of 40 mA on all I/O pins. The hardware on
board can talk Serial (UART), I2C, and SPI. Outside the chip the Arduino Mega board contains
voltage regulators, capacitive components, and I/O connectors. The code written in the IDE is
the only code that runs on the chip. There is no interpreter, operating system, or firmware. On
the other hand, the Raspberry Pi is a single board computer, or SBC. On board it contains a 32
bit microprocessor and supports video, audio, USB host, ethernet, SD card, HDMI cord, and GPI
headers. These assets mean the Raspberry Pi has more in common with a computer than an
46 | P a g e
Arduino. Instead of writing code to control the hardware directly, the programs written are
actually run within an operating system. The operating system is typically Linux. Both the
Arduino and Raspberry Pi contain a CPU, timers, memory, as well as I/O pins. The key
difference is in the I/O pins. As a microcontroller, the Arduino has high I/O capability, able to
drive external hardware directly. On the other hand, the Raspberry Pi is a microprocessor, which
tend to have weak I/O, needing transistors to drive most hardware. However, microprocessors
are good at processing, containing better brain power than a microcontroller. The Raspberry Pi
has a 700 MH clock speed, over 40 times faster than the Arduino. It is also based on a 32-bit
architecture, compared to the 8-bit of the Arduino. The Raspberry pi has 256,000 times more
RAM than an Arduino. The Arduino has more I/O pins and can also utilize 8 times more current
in each pin than the Raspberry Pi. Finally, the Raspberry Pi typically operates on Linux, while
the Arduino has no operating system. The Arduino was ultimately chosen for ACROBE for its
capability of controlling many sensors with more I/O capability, and, its simple, effective
interface, while remaining cost friendly.
Modern wind direction sensors on the market are heavy, expensive, and not designed to
interface with the Arduino. The final design selected for wind sensing was the rotary encoder.
A rotary encoder will be placed in between the wind vane and intake. Specifically, the
EMS22P50-B28-LS6 from Bourns. It has low friction and high sensitivity. The rotary encoder
will determine the relative position of the intake to the vane. A negative feedback loop will be
created between the rotary encoder and a continuous rotation servo attached to the intake. When
the vane is not lined up with the servo, the servo will turn. Once lined up, the servo will stop. A
potentiometer could have been used to determine the position of the wind vane, however, it
could increase friction on the wind vane. The more friction the wind vane has, the less sensitive
47 | P a g e
it will be to wind speed. The hexacopters position also needs to be known, and a potentiometer
cannot be used for this. A digital compass sensor could have been used, however, many of them
would have to be used to orientate the position of the vane, intake, and hexacopter. A camera
may have to be used if the information from the digital compass cannot be sent remotely. A
camera sending live feed is not optimal because of the cost and added weight.
The speed of the fan drawing air through the device is imperative for efficient collection
of bacteria. The speed of the air moving through the inlets must be in between 20-30 m/s. The
final design will have a pressure sensor, specifically the MPX4115a, located in a port near the
end of the system to measure the pressure when the air speed in the inlets is 20-30 m/s. The
pressure value taken will then be used in a negative feedback loop to drive the fan. If the
pressure ever reaches over or under a normal value, the fan will either reduce or increase its
speed. The fan will be controlled by a S-12 ESC. The ESC speaks to the Arduino using servo
language, and thus can be controlled using PPM from the servo library. The ESC is lightweight
and contains the power requirements to operate the DC motor. The pressure sensor is a rugged,
temperature compensated, and cost effective sensor that functions within the specified pressure
and temperature range.
Humidity is important for studying bioprecipitation. Temperature is also good data to
have when studying cells that live in ambient weather. Humidity has to be temperature
compensated, so many breakout boards will read out humidity as well as temperature. The HIH
6130 breakout board from Sparkfun was chosen. It operates from 2.3 to 5.5 volts, and is
temperature compensated from 5 to 50 C. It also has a compensated humidity range of 10 to
90% relative humidity. The HIH 6130 speaks to the Arduino through the I2C bus. The HIH
48 | P a g e
6130 will be placed in a protective location from debris, but will remain open to the environment
for data acquisition.
The solution for data acquisition needed to be lightweight and cheap. Most DAQs are
extraordinarily expensive, heavy, and consume copious amounts of space. Sparkfun
manufactures a SD card data logger that outmatches all of the criteria. Data can be sent to the
lightweight and portable SD card, ready for acquisition upon placing the card into a computer.
The breakout board is placed on top of the Arduino, saving space and weight.
Two batteries will be used for the entire system. A 2S Lipo battery will power a 12A
ESC and the DC motor used to drive the fan that causes suction. The other batteries will be two
14.4V, 4S Lipo battery. This battery will power the six 20A ESC used to control the hexacopters
rotors. As well as this, the 5V BEC onboard the ESC will run to the RX receiver. Also, another
5V BEC will run from a separate ESC to power the CR servo. The receiver will then turn on the
Arduino when commanded from the transmitter.
List of Parts
Part
# of
Units
MPX4115a
HIH6130
Company
Manufacturer
Freescale
Semiconductor
Honeywell
Company of
Purchase
Digikey
Combined
Price
48.27
Amazon
29.95
Arduino Mega
Arduino
Amazon
37.52
EMS22P50-B28LS6
OpenLog
Bourns
Mouser
37.8
Sparkfun
Sparkfun
19.95
Futaba S148
Futaba
Adafruit
14
Futaba BLS153
Futaba
ServoCity
360
Total
547.49
Actual Price
187.49
Table
8:
The
total
price
is
everything
needed
for
construction
of
the
control
system.
The
actual
price
only
includes
the
prices
of
equipment
that
we
had
to
buy.
49 | P a g e
Wiring Schematic
Figure
19:
Schematic
of
control
system.
Microcontroller
featured
in
the
center.
DC
motor
is
powered
by
11.1V
Lipo
battery.
Power
source
is
continuous
9
volts
from
the
switch.
50
|
P a g e
Figure
20:
Physical
image
of
completed
circuit
board,
showing
data
logger,
temperature
and
humidity
sensor,
voltage
regulator,
low-pass
filter,
and
quick
disconnects
for
the
ESC
and
CR
servo.
Testing
The wind vane needed to be tested for its mechanical ability to face the wind direction.
The inherent ability for the wind vane to point towards wind direction was tested by placing the
wind vane onto the rotary encoder, and placing into a wind tunnel. In order to have efficient
collection, the wind vane needed to point within 15 degrees of wind direction.
51 | P a g e
Figure
21:
Demonstrates
the
wind
vanes
direction
from
wind
at
three
different
wind
speeds.
Red
shows
the
angles
the
wind
vane
needed
to
stay
within.
Green
shows
the
wind
vane
when
starting
from
90
degrees
to
the
left,
and
purple
shows
the
wind
vane
when
from
starting
90
degrees
to
the
right.
The MPX pressure sensor was connected to the exhaust through a port. It is used to
measure the differential pressure caused by airflow. When the airflow speeds up through the
exhaust, the air becomes turbulent. The turbulent flow causes the pressure sensor to read errant
values. When these values are sent to the ESC, the motors speed fluctuates. A running average
was created in the code to smooth out the values sent to the ESC.
Graph
5:
Demonstrates
the
wide
standard
deviation
bars
caused
by
no
smoothing
with
high
throttle.
52 | P a g e
Graph
6:
Demonstrates
the
reduced
error
bars
created
by
using
a
bin
average.
Graph
7:
Demonstrates
significantly
reduced
error
bars
when
a
moving
average
is
used.
53 | P a g e
Fluid
Dynamics:
The
ACROBE
was
analyzed
using
Autodesk
Simulation
CFD.
When
the
small
inlet
!
tubes
had
a
velocity
of
20
! ,
the
outlet
velocity
was
found
to
be
approximately
10
! .
Using
this
value
for
the
outlet
velocity,
an
anemometer
was
used
to
monitor
outlet
velocity
on
the
!
ACROBE,
and
the
inducted
fan
motor
increased
until
the
velocity
was
10
! .
The
internal
pressure
at
this
time
was
recorded
and
used
to
calibrate
the
pressure
sensor
that
drives
!
the
inducted
fan.
This
ensured
that
the
velocity
in
the
inlet
tubes
was
the
desired
20
! .
Figure
22:
Fluid
flow
analysis
of
ACROBE.
54 | P a g e
Graph
8:
The
velocity
profile
through
one
of
the
small
inlet
tubes
of
ACROBE,
which
shows
that
the
average
velocity
throughout
most
of
the
tube
is
approximately
21
m/s.
In
order
to
determine
where
to
place
ACROBE
relative
to
the
hexacopter,
we
attached
an
anemometer
at
various
heights
above
the
hexacopter,
as
wind
speeds
below
the
hexacopter
were
much
greater
than
above.
The
graph
shown
is
an
average
of
7
data
points
per
bar,
each
at
different
angular
positions
on
the
same
concentric
circle.
The
y
axis
is
the
vertical
wind
velocity
and
the
x
axis
is
the
vertical
distance
above
the
main
deck
cover.
It
was
found
that
4
above
the
main
deck
cover
had
the
least
amount
of
prop
wash.
This
is
where
ACROBE
was
mounted.
55 | P a g e
Graph
9:
Graph
showing
the
vertical
wind
speed
with
respect
to
distance
above
the
main
prop
deck.
Bacterial
Collection:
Following
bacterial
collection
with
the
ACROBE,
the
aluminum
foil
on
which
the
bacteria
were
collected
was
removed
from
the
ACROBE
carefully.
To
remove
the
foil,
gloves
were
worn
and
the
foil
was
touched
only
on
the
edges
to
avoid
removing
any
bacteria.
The
ACROBE
allows
for
two
impact
collections
and
both
were
used
for
separate
purposes.
The
first
foil
is
stamped
onto
an
R2A
plate
to
create
a
press
plate
of
the
original
bacterial
concentration.
The
second
foil
was
transferred
into
40
of
R2A
solution
and
shaken
for
two
hours
to
transfer
the
bacteria
from
the
foil
to
the
R2A
solution.
With
the
R2A-
bacteria
broth
from
the
second
foil,
we
made
serial
dilutions
and
spread
plates.
We
planned
to
create
serial
dilutions
for
up
to
1x10-8.
Each
serial
dilution
tube
contained
0.9
of
R2A
and
0.1
of
the
bacteria-R2A
broth
would
later
be
added
in
for
dilution.
We
transferred
56 | P a g e
0.1mL
of
the
original
R2A-
bacteria
broth
into
a
tube
contain
0.9mL
of
R2A.
This
tube
was
labeled
1x10-1.
After
using
the
vortex
to
mix
thoroughly,
we
transferred
0.1
from
the
tube
labeled
1x10-1
to
a
tube
labeled
1x10-2.
We
continued
this
all
the
way
until
we
reached
1x10-8.
After
creating
all
serial
dilutions,
we
used
a
cell
spreader
and
Bunsen
burner
to
create
spread
plates
starting
from
1x10-3
until
1x10-8.
Finally
we
Para
filmed
all
plates
and
allowed
multiple
days
for
bacterial
growth
at
room
temperature.
Although looking under a microscope at a hemocytometer is not the best method for
quantification,
we
did
for
our
project.
This
served
as
another
method,
along
with
the
serial
dilutions
and
the
press
plate,
to
have
data.
We
performed
four
trials
total.
Trial
1
was
done
here
at
E.B.
Doran
for
2.5
.
The
incubation
time
was
for
about
50
.
Trials
2-4
were
done
in
Maringouin,
LA
which
is
about
20
west
of
Baton
Rouge.
For
these
three
trials,
the
temperature
and
humidity
sensors
were
active
during
the
flights.
Trial
2
had
a
!"##$
cell
concentration
count
of
5.4X
108
!" ,
a
collection
time
of
2.5 ,
an
incubation
time
of
26.5
,
a
temperature
reading
of
30.4C
1.6,
and
a
relative
humidity
of
63.3%
!"##$
3.8.
Trial
3
had
a
cell
concentration
of
4.7x108
!" ,
a
collection
time
of
2.5
,
an
incubation
time
of
24
,
a
temperature
reading
of
27.3C
0.5,
and
a
relative
humidity
of
77.7%
1.6.
Last,
trial
4
had
a
cell
concentration
of
9.9x108
/,
a
collection
time
of
5
,
an
incubation
time
of
24
,
a
temperature
of
25.2C
0.2,
and
a
relative
humidity
of
84.9%
2.6.
We
do
know
that
in
order
to
have
statistically
correct
bacterial
analysis
we
needed
to
have
collected
three
times
within
the
same
conditions.
Although
we
did
not
do
this,
when
comparing
trials
3
and
4,
there
is
about
two
times
the
cell
concentration
on
trial
4
than
trial
3
with
trail
4
having
twice
the
flight
time.
This
data
shows
that
our
bacterial
collection
was
precise.
57 | P a g e
Redesign:
Since the group is in the process of patenting the ACROBE, we have decided not to
give
specifics
to
the
design.
However,
we
do
know
that
we
need
to
have
a
more
efficient
hexacopter
that
can
have
better
batteries
to
have
a
longer
flight
time
and
carry
a
heavier
load.
58 | P a g e
Schedule:
Budget:
Vendor
Amazon
Hobby
Towne
Lowes
Figure
27:
This
Gantt
Chart
contains
deadlines
for
Spring
2015
semester.
Items
Polycarbonate
tubing
1/4
ID
Polycarbonate
tubing
3/4
ID
Polycarbonate
tubing
7/8
ID
Polycarbonate
tubing
1"
ID
Polycarbonate
tubing
1/2"
ID
Polycarbonate
tubing
actron
vacuum
and
fuel
pressure
kit
polycarbonate
tubing
3/4
ID
Polycarbonate
tubing
Temperature
and
Humidty
Sensor
RC
Lipo
Battery
Voltage
Checker
Epoxy
nuts,
bolts,
balsa
woodetc
Wood
monocoat
1-inx
5ft
Sch-40
PI
3/4
in
x
5
ft
Sch40
PVC
Price
$20.25
$33.65
$16.11
$13.67
$25.09
$19.71
$17.84
$29.14
$29.74
$29.95
$22.99
$5.60
$15.99
$22.30
$39.51
$28.27
$3.09
$1.79
59
|
P a g e
Digi-Key
Pololu
The
Felt
Store
The
Composite
Store
Hobby
King
Dollar
Tree
Hobby
Lobby
Online
Metals
Adafruit
Mouser
Spark
Fun
Radio
Shack
Total
with
Shipping
$4.14
$2.08
$3.03
$0.97
$1.98
$0.92
$1.41
$1.86
$0.47
$0.77
$1.56
$12.72
$48.27
$39.86
$17.90
$29.94
$8.70
$43.35
$21.32
$5.00
$57.17
$49.44
$68.10
$1.00
$1.00
$1.00
$1.00
$6.99
$67.63
$29.13
$16.89
$28.96
$17.96
$5.43
$1,054.70
60 | P a g e
Bill
of
Materials:
Item
Temperature
and
Humidity
Sensor
Pressure
Sensor
Continuous
Rotation
Servo
Carbon
Fiber
Tubing
(4
ft)
Carbon
Fiber
Tubing
(2
ft)
Resin
Hardener
Epoxy
Rotary
Encoder
Woven
Carbon
Fiber
Sheet
Slip
Ring
Arduino
Mega
Proto
PCB
Sparkfun
Logger
SD
Card
Carbon
Fiber
Fabric
RC
Lipo
Battery
Arduino
Mega
22
ga.
Insulated
Wire
(10
ft)
Neodymium
Magnets
Futaba
Receiver
Electronic
Switch
2.5
mm
nuts/bolts
Resistors
and
Capacitors
Quantity
Total
1
$29.95
2
1
2
2
1
1
1
1
5
1
1
1
1
2
1
1
1
8
1
1
20
$32.18
$17.90
$8.70
$5.00
$43.35
$21.32
$15.99
$39.86
$68.10
$29.13
$8.45
$28.96
$17.96
$57.17
$22.99
$37.03
$6.99
$5.00
$69.98
$7.39
$10.00
$10.00
2
each
61 | P a g e
Conclusion:
In
conclusion,
the
scope
of
this
senior
design
project
was
able
to
be
completed
underneath
the
umbrella
of
Biological
Engineering.
The
engineering
principles
learned
throughout
the
course
of
our
undergraduate
years
here
at
LSU
helped
us
achieve
our
goals.
We
were
able
to
accomplish
our
goals/objectives
of
making
sure
the
ACROBE
is
modular,
easy
to
use,
capable
of
multiple
runs
in
a
day,
and
be
autonomous.
62 | P a g e
References:
[1]
Christner,
Brent
C.
"Cloudy
with
a
Chance
of
Microbes."
News
Magazine
of
the
American
Society
for
Microbiology
Feb.
2012:
70-75.
Web.
20
Nov.
2014.
[2]
Smith,
A.,
and
R.
Katz,
2013:
U.S.
Billion-dollar
Weather
and
Climate
Disasters:
Data
Sources,
Trends,
Accuracy
and
Biases.
Natural
Hazards.
[3]
Granger,
Doug.
"High
Altitude
Student
Payload
(HASP)."
LSU
Space
Sciences
Group.
Louisiana
State
University,
n.d.
Web.
[4]
Journal
of
Microbiological
Methods,
Volume
107,
December
2014,
Pages
161-168
N.C.
[5]
Whyte,
W.,
G.
Green,
and
A.
Albisu.
"Collection
Efficiency
and
Design
of
Microbial
Air
Samplers."
Journal
of
Aerosol
Science
38.1
(2007):
97-110.ScienceDirect.
Web.
15
Sept.
2014
[6]
Carbon
Fiber
Fabric."
3K
Plain
Weave
Carbon
Fiber.
The
Composite
Store
(CST),
2014.
Web.
02
Dec.
2014.
[7]
"West
Systems
Laminating
Epoxy
Resin."
West
System
Epoxy.
The
Composite
Store
(CTS),
2014.
Web.
02
Dec.
2014.
[8]
"6061
Aluminum."
6061
Aluminum,
Aluminium,
Alumina
Bobco
Metals
LLC.
Bob
Co.
Metals,
2014.
Web.
02
Dec.
2014.
[9]
"Nylon
Nuts,
Bolts
&
Washers."
IndustrySearch.
Industry
Search,
2014.
Web.
02
Dec.
2014.
[10]
Dulbecco,
R.
et
al.
(1954):
Plaque
formation
and
isolation
of
pure
lines
with
poliomyelitis
viruses.
In:
J.
Exp.
Med.
vol.
99
(2),
pp.
167-182.
[11]
Reasoner,
D.
J.,
and
E.
E.
Geldreich.
1979.
A
new
medium
for
the
enumeration
and
subculture
of
bacteria
from
potable
water.
[12]
Boundless.
Viable
Cell
Counting.
Boundless
Microbiology.
Boundless,
14
Nov.
2014.
[13]
Wolochow,
H.
The
Membrane
Filter
Technique
for
Estimating
Numbers
of
Viable
Bacteria:
Some
Observed
Limitations
with
Certain
Species.
Applied
Microbiology
6.3
(1958):
201206.
Print.
[14]
Sauer
S.,
Kliem
M.
2010.
Mass
spectrometry
tools
for
the
classification
and
identification
of
bacteria.
Nat.
Rev.
Microbiol.
8:7482.
63 | P a g e