Preparation

Quick Reference Guide

Timer
(Not Provided)

Test pouch
(Test Device)

Disposable Gloves
(Not Provided)

Micropipette & Disposable Tip


(Not Provided)

SD BIOLINE

HBsAg WB
Caution:
Each test can be used ONLY ONE TIME.
Do not use the test more than once.

Test Procedure
1

Check Point !
Desiccant

YYYY/MM/DD

Check the expiry date.


If expiry date has passed, use another kit.
Check the desiccant.
If the color is green, use another kit.

Put on the gloves.

Open the test pouch and write the patients name on


the test device.

Prepare the serum/plasma/whole blood sample.


Yellow Color

Green Color

100l

Overview of the Device

Add drawn sample into the sample well (S marked).

HBsAg WB
HBsAg WB

Use the micropipette to collect sample (100l).

Wait 20 minutes after adding sample.


Dont read test results after 20 minutes.
Reading too late can give false results.

Control Line
Result
Window

Interpretation

Test Line

POSITIVE RESULT
HBsAg WB
HBsAg WB

NEGATIVE RESULT
HBsAg WB
HBsAg WB

The presence of two color bands (T band and C band) within the results window no matter which
band appears first indicates a positive result.

www.standardia.com

Tel : +82-31-899-2800 Fax : +82-31-899-2840


E-mail : sales@standardia.com

HBsAg WB
HBsAg WB

The presence of only one band (C band) within the results


window indicates a negative result.

INVALID RESULT
HBsAg WB
HBsAg WB

Sample
Well
HBsAg WB
HBsAg WB

No C line in the result window.


The specimen should be retested using a new test kit.

QU-01FK10WEN
Rev. 1402-00