PURPOSE OF ANALYSIS
SELECTION OF ANALYSER
CALIBRATION
TROUBLE SHOOTING
TOPICS
II
PURPOSE OF ANALYSIS
SELECTION OF ANALYSER
CALIBRATION
PURPOSE OF ANALYSIS
COMPOSITION
SR. NO.
COMPONENT
MIN. %V/V
NOR. %V/V
MAX. %V/V
CHLORINE
85
94
97
HYDROGEN
1.4
1.00
0.2
OXYGEN
2.73
0.964
0.564
CARBON
DIOXIDE
0.541
0.191
0.112
NITROGEN
10.27
3.627
2.12
MOISTURE
0.059
0.004
0.01
SELECTION OF ANALYSERS
SELECTION OF ANALYSER
THERMAL CONDUCTIVITY
SELECTION OF ANALYSER
LAB METHOD
ORSAT APPARATUS
NaOH + Cl2
N a C l + H2O
THERMAL CONDUCTIVITY
DIFFERENT COMPONENTS
GASES
TC factor
Cl2
0.381
H2
6.990
O2
1.052
CO2
0.690
N2
0.996
CO
0.962
SO2
0.377
THERMAL CONDUCTIVITY
SAMPLE
REFERENCE
TYPES OF FILAMENTS
METALLIC WIRES
TUNGSTEN ,NICKEL,PLATINUM
ADVANTAGES:
HIGHER TEMPERATURE
LINEAR OVER A WIDE RANGE
SIMPLE CONSTRUCTION
2
THERMISTOR
ADVANTAGES:
AMBIENT TEMPERATURE
VERY GOOD SENSITIVITY
MEASURING CELL
REFERENCE GAS
FLOWING REFERENCE
USED FOR
SPECIAL TASKS
H2 IN Cl2 ANALYSER
SAMPLE TRANSPORT
TYPE - A SINGLE LINE WITH BYPASS STREAM
SAMPLE CONDITIONING
SAMPLE DISPOSAL
H2 IN Cl2 ANALYSER
SAMPLE CONDITIONING
PHYSICAL PROPERTIES
PRESSURE
CONTROL
FLOW CONTROL
TEMPERATURE CONTROL
AUXILLIARY REACTION
PURPOSE
H2 + C l 2
UV SOURCE
2HCl
UV REACTION ATTACHMENT
CASE MATERIAL
1
2
1
2
UV REACTION ATTACHMENT
CASE MATERIAL
UV REACTION ATTACHMENT
COMPOSITION
SR. NO.
COMPONENT
MIN. %V/V
NOR. %V/V
MAX. %V/V
CHLORINE
3.83
6.61
7.19
HYDROGEN
2.03
3.5
3.8
OXYGEN
19.68
18.72
18.53
CARBON
DIOXIDE
0.41
0.71
0.76
NITROGEN
74.04
70.45
69.71
MOISTURE
0.01
0.01
0.01
TUBULAR FURNACE
FURNACE
MEASUREMENT
Thermal
conductivit
Analyser
Vent
Sample in
Cl2, H2, O2, CO2 etc
Analyser output
H2 concentration
UV
Reaction
HCl, O2, CO2 etc
MEASUREMENT
HYDROGEN CONCENTRATION
H2 IN Cl2 ANALYSER
SAMPLE CONDITIONING
PRESSURE CONTROL
SAMPLE STREAM PRESSURE
OPERATING PRESSURE FOR ANALYSER
FLOW CONTROL
SAMPLE STREAM FLOWRATE
FLOWRATE FOR ANALYSER
- 2.5 BAR
- 0.36 - 1.45 PSI
(0.02 - 0.099 BAR)
- 30 - 60 L/HR
TEMPERATURE CONTROL
SAMPLE STREAM TEMPERATURE
TEMPERATURE FOR ANALYSER
- - 39deg C
- 5 - 50 deg C
H2 IN Cl2 ANALYSER
SAMPLE CONDITIONING
SAMPLING SYSTEM
SAMPLE DISPOSAL
LIMITS
- 150 mmWC to
VENTS
BYPASS STREAM & ANALYSER VENT
50 mmWC
SEPARATE LINES
RELEASING TO ATMOSPHERE
VENDOR PREFERENCE
STATUATORY REQUIREMENT
TREATMENT BEFORE DISPOSAL
WASTE AIR HEADER - BLEACH LYE UNIT
SAMPLE DISPOSAL
VENT
PROVIDE
LIQUID SEAL
ENCLOSURE
CABINET MOUNTING
ANALYSER & SAMPLE HANDLING SYSTEM
PURGING OF PANEL
CORROSSIVE ATMOSPHERE
CALIBRATION OF ANALYSER
PRIMARY CALIBRATION
BY ACTUAL COMPOSITION - MIN , NOR & MAX
CALIBRATION GRAPH
- EMBEDDED IN MICRO-PROCESSOR
- H2 IN N2
ZERO GAS
PURE N2
ATTENTION
CHLORINE LEAK
- 10-12 MONTHS
GASIFIER UNIT
PURPOSE OF ANALYSIS
COMPOSITION
SR.NO
COMPONENT
MIN.
NOR. %
CO
22.89
CO2
2.27
H2
22.93
COS
0.002
CH4
0.2
N2
0.118
Ar
0.183
H2O
51.23
H2S
0.08
10
NH3
0.069
MAX.
Partial oxidation
2 C n H m + n O2
2n CO + m H2
( Moderator )
Partial oxidation
Complete oxidation
>
1290 0 c - 1400 0 c
1400 0 c
2n CO + (m+2n) H2
Crack Reaction :
CmHn
n / 4 CH 4 + m - n / 4 C
CO + H2O
CO2 + H2
SELECTION OF ANALYSER
THERMAL CONDUCTIVITY
ABSORPTION OF IR RADIATION
SELECTION OF ANALYSER
THERMAL CONDUCTIVITY
PHOTOMETRY ( SPECTROMETRY)
PHOTOMETRY
A
A
b
c
= bc
- Absorbance
- Molar Absorptivity
- Thickness of the light absorbing medium
- Concentration of the Absorbing species
PHOTOMETRY
BEER - LAMBERT LAW
In the Process Stream ;
A
= 1 b c1 + 2 b c2 + 3 b c3 + .
A
- Absorbance
1 2 3 - Molar Absorptivities
b
- Thickness of the light absorbing medium
c1 2 3 - Concentrations of the Absorbing species
PHOTOMETRY
INFRARED PHOTOMETRY
IR
UV & Visible
- Excitation of electrons
COMPONENTS
SOURCE
MONOCHROMATOR
SAMPLE CELL
DETECTOR
TRANSMITTER
SOURCE
SOURCE
IR RADIATION
: Heat
MONOCHROMATOR
SAMPLE CELL
The Component which brings the conditioned sample to the optical path of the
Instrument.
CELL BODY - Extn of sample Conditioning
system
SAMPLE CELL
DETECTORS
Thermocouple
Luft / Microphone
Micro-flow Sensor
Pyro- electric
LUFT / MICROPHONE
Capacitance change
Double beam
MICRO- FLOW
Change in flow
Double beam
SINGLE BEAM
SINGLE BEAM IN TIME
DOUBLE BEAM
DOUBLE BEAM IN TIME
MEASUREMENT
= log ( I0 / I ) = b c
MULTICOMPONENT ANALYSER
SAMPLE CONDITIONING
PRESSURE CONTROL
SAMPLE STREAM PRESSURE
- 79 . 4 BAR G
OPERATING PRESSURE FOR ANALYSER - 16 -250 hPa
(0.016 - 0.25 BAR)
FLOW CONTROL
SAMPLE STREAM FLOWRATE
FLOWRATE FOR ANALYSER
- 10 -100 L/HR
TEMPERATURE CONTROL
SAMPLE STREAM TEMPERATURE
- 245 deg C
TEMPERATURE FOR ANALYSER
- 0 to 45 deg C
MEASURING CHAMBER IS MAINTAINED AT A
TEMPERATURE ABOVE AMBIENT.
REFERENCE GAS
MULTICOMPONENT ANALYSER
SAMPLE CONDITIONING
SAMPLE CONDITIONING
ENCLOSURE
ANALYSER HOUSE
ANALYSER
SAMPLE HANDLING SYSTEM
CO DETECTORS
CALIBRATION OF ANALYSER
ZERO GAS
PURE N2
SAMPLE DISPOSAL
VENTS
BYPASS STREAM & ANALYSER VENT
SEPARATE LINES
RELEASING TO ATMOSPHERE ?
REFERENCES
PROCESS ANALYSERS
- D.C.CORNISH
- GARY D. NICHOLS
- D.J.HUSKINS
FLOW VELOCITIES