Вы находитесь на странице: 1из 22

SIEMENS

AKTIENGESELLSCHAFT
Medical Solutions
Inde
x
Type No. / Board-No.

Surve sheet
to wiring diagram

Product:

Sheet
No. Revision-No.

1A
1

Type No.:

Polydoros IT

000
00

X2287
Author
Department

nd.Mittlg.Nr.

Date

02.08.05

Schnapperelle
CO PTS

Print Nr. COPL-136.844.03.01.02

X2287 - A - 57 60 074
Type No.

00

00

00

00

00

00

00

Print No.

00

10

00

Type
No.

11

00

12

00

13

00

14

00

15

00

16

00

17

00

18

00

18

00

19

00

20

00

21

00

Inde
x

01

Part number

Rev
.

Part number

Rev
.

Part number

Rev
.

Print No.
Type No.

Inde
x

Inde
x

RevisionNo.000
Dept.: CO PTS

Date:
Checked:
Als Betriebsgeheimnis anvertraut/Disclosed as trade secret, Alle Re chte
vorbehalten/ All rights reserved, Uebersichtsblatt zum Schaltplan / Surve sheet
to wiring diagram, Erzeugnis / Produkt, Sach-Nr./ Part No, Blatt-Nr./ Sheet no.,
Ausfhrungs-Nr./ Version No., Typen-Nr./ Type No., nd.-Mittlg.-Nr./ Change
nitice No., Datum / Date, Ersteller / Issuer, Abtlg./ Department, Index / Index,
Geprft / Checked, U-Art / Doc.type, U-Stand / Doc.status, Druck Nr./ Print No.,
Ersatz fr / Replacement for

Part-No.

57 60 074
Replacement for:

Art

EFS
-

Teil

- 01S -

Version

00

Type-No.

- X2287
Sheet 01A

s
A

Table of contents

Only for
internal
use
appoimted.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Page
1A

Cover page

Overview of pages

Functional Overview Polydoros IT

Overview of the Generator

External connections

Schematic Polydoros IT 30 / 55

Schematic Polydoros IT 65 / 80

Power supply

D500 Circuit board 77 59 819

D500 Circuit board 57 60 165

10

LEDs / Jumpers D500

11

Test points D500

12

D501 Circuit board 77 59 835

13

D501 Circuit board 57 60 181

14

LEDs / switches D501

15

D470 Filament

16

D115 Rotation Anode Circuit RAC

17

D510 Inverter IT 30/55

18

D510 Inverter IT 65/80

19

Connection Single Tank

20

Connection Optitop Tube

21

Option Energy Storage Unit

Sheet No. 1
Description

RevisionNo.00

Overview of pages

Part-No

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

D500 Generator

D510 kV Inverter

Controller board
kV Control
4

Main Inverter
Control

MCB *1
+
Dongle *2
RAC Control

I_Roh Control
I FilamentControl

FIL Inverter
Control

UDC
D511

Monitoring

Resonant
circuit

D501 Interface

Only for
internal
use
appointed.
All rights
reserved.

D470 Filament inverter

24V dc
Supply

AEC
IONTOMAT

D115 Rotating Anode


Circuit

Copyright
SIEMENS AG
All rights
reserved.

I/O Interface

Tube

Control
Console
*3

Author:
Date:
Schnapperelle
CO PTS
02.08.05

External
connections
HVT
M16
Linefilter

Transformer
optional
1

additional in STTA

L1 L2 L3

*1 Inclusive SW Generator and OGP Data for Sta ndard desk


*2 SW Optionen
*3 Touch panel with SD Card for SW and OGP's
A

Sheet No. 2

RevisionNo.00
Description

Functional Overview Polydoros IT

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
C

Overview of the Generator Cabinet

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Overview of the Control Consoles

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Control console (standard) /


organ console (option)

Touch panel
1

Sheet No. 3

RevisionNo.00
Description

Overview of the Generator

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
C

X4

V_P15V

F2 (5A)

6
5

F4 (1A)

-15V

ANA_GND

+5V

GND

ANA_GND

Workstation

24V

2
4

PREP

STANDBY

CONS_IN_2
CONS_IN_1

XRAY

24V, 5W

24V

Raster_In
Xray_On

K8 24V

Select_G2
K12

HK_G2

24V

GND24

5
24V, 5W

STANDBY

Only for
internal
use
appointed.
All rights
reserved.

24V, 5W

K14 24V
PLANI_TIME_E

1
2

Copyright
SIEMENS AG
All rights
reserved.

PLANI_TIME_C
24V
DOOR_STATE_IN
24V

TUBE: TEMP

3
4

T_MAX_TUBE_IN

TUBE: PRESSURE

Energie Storage
Unit:READY

7
8

ESU_READY

24V

Schnapperelle
CO PTS
02.08.05

Ready_In_G1

24V
PLANI_END_S_IN ( Plani_Time_A )
PLANI_STARTPOS_IN ( Plani_Time_B )
PLANI_8GRAD_SYNC_IN ( Plani_Time_C )
PLANI_30GR_SYNC_IN ( Plani_Time_D )
RASTER_IN (Grid)
PLANI_HK (Exposure_Request)

V_24V_1
GND24

Dose Area Product_Unit


(KermaX)
Iontomat_Chamber A

X14

Ready_In_G2

X33

V_24V_1

24V, 5W

24V
GND

1
2
3
4

Xray_End_PF1
Xray_End_PF2
GND24

X22

Iontomat_Chamber B

24V

G2 OEM

Xray_End

X34

V_24V_1
Gen. Purpuse In

X23

1
2
3
4
5
6
7
8

HK_G2
Grid_G2
HK_G2_PF1
HK_G2_PF2

24V

G1 OEM

X31

GND24
Select_G2

24V

KK2.G1/G2

1
2
3
4
5
6
7
8

HK_G1
Grid_G1
HK_G1_PF1
HK_G1_PF2
V_24V_1

24V

X30

Select_G1

24V

KK 1B.G1/G2

18
20

GND24

HK: Exposure_Request
VK: Prep
PF: Potential Free

GND24

P_MAX_TUBE_IN

1
2
3
4
5
6
7
8

Author:
Date:

PLANI_TIME_D

Signal description:

24V

5
6

X10

14
15
16
17

PLANI_TIME_B

X8

Control
Console

X25

PLANI_TIME_A

DOOR

13

9
12
14
16
20

VK

PREP

13

2
5
7
8

HK_G1

1
24V, 5W

7
14

X24

GND24
Select_G1

X7

14

S2
Generator ON

Workstation

X19 W515

CONS_OUT_2

K114

230V, 100W

CONS_OUT 1

K112

230V, 100W

D501
Interface
Board

K108

230V, 100W

S1
Generator
OFF/ON

K113

1
XRAY

W503

F3
(0.5A)

X6

230V, 100W

Sensor T3

2
1

VCC

F1 (5A)

X3 W503

+15V

X1

Power
Supply

Gen. Purpuse Out


GND24

1
2
3
4

not used

Sheet No. 4

RevisionNo.00
Description

External connections

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
A

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 5
Description

RevisionNo.00
A

Schematic Polydoros IT 30 / 55

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
A

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 6
Description

RevisionNo.00
A

Schematic Polydoros IT 65 / 80

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

RevisionNo.00

Description

T1

Clamp

4-5-6

1-2-3

T 3,15A

F2

400VA

T3

F33
T
6,3A
F31
T 0,8A

Schnapperelle
CO PTS
02.08.05

T 3,15A

F1

Author:
Date:

L3

L1

480V

440V

Voltage

P 15V A

Part-No.

Art

Teil

Version

57 60 074

EFS

01S

00

C
Poti
5V

(1) (3)

T 1,25A

F3

+24V

X21

(6)

X4

T 1A

T 5A
(5)

F4

F2

-15V

X13

+15V

+15V (7)
GND

X12

T 5A

F1
(8)

+5V

GND
(7)
(9)

+2,5V
+3,5V

D501

X17

ribbon cable 16 pole

ribbon cable 50 pole

X3

X2

D115

D470

+15V (47)
-15V (13)
GND

15V / 50kHz

15V / 50kHz

X20

X21

X1

+15
V
+5V
-15V
GND

X4

(4)

(3)

(2)

(1)

Power
Supply

X3
230V AC

23VAC

X3

X2

X16

+2,5V
+3,5V

Only for
internal
use
appointed.
All rights
reserved.

X1

N 15V A
VCC

+5V (1)
GND (2)
+15V (3)
-15V (4)

X1

X2

D500

Connection

X1

+15V (1)
GND (2)

X10

X10

Copyright
SIEMENS AG
All rights
reserved.

+5V (25)
GND
+15V (30)
-15V (26)

D-SUB plug 15 pole

D510

s
C

Sheet No. 7

Power supply

X2287

Type-No.

s
B

S3
Hardware
configuration
X11

V4

LEDs
Workstation
Selection

OFF

ON
4

LEDs
Dominat

X16

V6

X4

X5

V45

V3

V49
V54

V7

Test Points
IONTOMAT

LEDs
Flex

D300
MCB3

ON

V21

Dongle D502 (Option)

Only for
internal
use
appointed.
All rights
reserved.

S1
Service

(further info see page "Dongle" )

OFF

D300/X11
PIN 3 A-B for Download SW
from SPC

Test points
Tube unit current

Service PC or Canon PC
connection

VCC

Ground
P_15_A

D500 part no.: 77 59 819

Copyright
SIEMENS AG
All rights
reserved.

X1

X15
V230

2
Test point
Starter,
kV controller

X302 3 2 1
Jumper
Hardware
configuration
1

OFF ON

X12

S4
Ignition

Test point
mA_Meas

X110

X 311

X186

Test point
Voltage
not on partno.:
5760165

X210

X13

X221

Schnapperelle
CO PTS
02.08.05

2
X14

X301
Jumper
Hardware
configuration

X100

X150

Author:
Date:

N_15_A

123

X 319

P_15_D

X300 1 2 3
Jumper
Hardware
configuration

Test points
Heating

V67

S2
mA_Measureswitch
1

X10

Sheet No. 8
Description

RevisionNo.00
A

D500 Circuit board 77 59 819

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

LEDs
Dominat

S3
Hardware
configuration
X11

V45

LEDs
Workstation
Selection

OFF

ON

V4

X16

V6

X4

X5

V3

V49

V54

V7

Test Points
IONTOMAT

LEDs
Flex

D300
MCB3

Dongel D502 (Option)

Only for
internal
use
appointed.
All rights
reserved.

S1
Service

(further info see page Dongle)

OFF

D300/X11
PIN 3 A-B for Download SW
from SPC

ON

V21

Test points
Tube unit current

Service PC or Canon PC
connection

Ground
Copyright
SIEMENS AG
All rights
reserved.

D500 part no.: 57 60 165


X1

X15

Test points
Heating

2
Test point
Starter,
kV controller

X110

Test point
mA_Meas

OFF ON

X186

S4
Ignition

X210

X13

X221

X150

Schnapperelle
CO PTS
02.08.05

X14

X100

Author:
Date:

V67

S2
mA_Measureswitch

1
X12

X10

Sheet No. 9
Description

RevisionNo.00
A

D500 Circuit board 57 60 165

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

LEDs

FLEX
Logistical component not programmed; no function of board is possible
Flashes when D500 is ready
Heating inverter start signal
RAC start signal
Inverter start signal
Radiation display (UTact>20kV)
Radiation request
RAC -Error
Error in the primary inverter
Error in heating inverter
Error with power supply voltages
Over-voltage error (+/- kV > 82kV)
Service switch switched on; generator goes into service mode
Pre-heating (both focuses are heated) active
Small focus active

V7
V8
V9
V10
V11
V12
V13
V14
V15
V16
V17
V18
V19
V20
V21

red
green
green
green
green
yellow
yellow
red
red
red
red
red
yellow
yellow
yellow

Flex Fault
Cycle LED
Start FIL Inv
Start RAC_Inv
Start Inv
XRAY_ON
SE
RAC_Inv_Fault
Inv_Fault
FIL_Inv_Fault
Supply_Volt
UT_MAX
Service Mode
Pre Head
Small Focus

V4
V6
V3

yellow
yellow
yellow

Dominat
If inverting is active: left dominant selected, otherwise the right one (Change after prep.)
Middle dominant selected (Change after prep.)
If inverting is active: right dominant is selected, otherwise the left on (Change after prep.)

V45
V49
V54

yellow
yellow
yellow

AP 0
AP 1
AP 2

Workstation Selection
Workstation Selection 0: off for Channel A (table), on for Channel B (Bucky W.S.)
Workstation Selection 1: always off
Workstation Selection 2: always on

V67

yellow

mA_Meas

mA_Meas
mA_Meas switch in position 2

V230
V231
V232
V233
V234

yellow
yellow
yellow
yellow
yellow

3,3V
P15V_A
P15V_D
N15V_A
VCC

3,3V
P15V_A
P15V_D
N15V_A
VCC

Voltage
Only for
internal
use
appointed.
All rights
reserved.

Switches and Jumper


Copyright
SIEMENS AG
All rights
reserved.

S1

S2

S4

1
Author:
Date:

2, 3, 4

Schnapperelle
CO PTS
02.08.05

5
6
7
8

X300
X301
1

X302

S1 Service
Normal condition
Generator goes into the service mode after a restart

OFF
ON

mA_Meas
Position 1 LED V67 off -> Normal mode
Position 2 LED V67 on -> tube current can be measured directly between 1 and 2 at test point X221
OFF
ON

Ignition
Ignition pulses not transmitted to inverter; no high voltage
Normal condition

OFF
ON
ON
OFF
ON
ON
OFF
ON
ON

S3 Hardware configuration
With Single Tank
With high voltage transformer
Must be ON
With power storage unit
No power storage unit
Must be ON
With MDM Detector
No MDM Detector
Must be ON

1 and 2
2 and 3
1 and 2
2 and 3
1 and 2
2 and 3

POLYDOROS
POLYDOROS
POLYDOROS
POLYDOROS
POLYDOROS
POLYDOROS

Jumper (not on board partno.: 57 60 165)


IT 65/80
IT (S) 30/55
IT 65/80
IT (S) 30/55
IT 65/80
IT (S) 30/55

On Board D300
D300/X11 PIN 3 A-B Just for Download SW from SPC/ see Update Instructions

Sheet No. 10
Description

RevisionNo.00
A

Leds / Jumpers D500

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

new
old
designation

X170
4

X120

X319
X317
X315
X313
X305
X307
X309
X311

A_GND
Vin
Vout
Fout
MP4
MP3
MP2
Start_Inv

VION
DL_IN

SWR

Test Points
Iontomat
Analog ground
Input of Iontomat U/F converter
Iontomat U/F converter voltage output Iontomat drift measurement capability
Iontomat U/F converter possible frequency

not used
not used
not used
Inverter start signal; for triggering during troubleshooting; low active
Voltage Test point (not on board partno.: 57 60 165)
Ground
Analog Ground
Digital Ground
+ 3,3V (+/- 5%) not adjustable
VCC + 5V range (5,0 V - 5,2 V) adjustable
+ 15V Analog (+/- 5%) not adjustable
- 15V Analog (+/- 5%) not adjustable
+ 15V Digital (+/- 5%) not adjustable

GND
A_GND
GND10
VCC3.3
VCC
P_15V_A
N_15V_A
P_15V_D

Tube current (IT)


3

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

X243

ITact

X242
X241
X240

ITnom
Threshold
A_GND

Actual value of tube current:


< 8mA: 1V= 1mA
< 80mA: 1V=10mA
> 80mA: 1V=100mA
Ref. value of tube current (as a reference for falling load)
Threshold DAC output for debug
Analog ground

X150

A_GND
F_Rot
I_Max
I_Load

Starter, kV controller (RAC)


Analog ground
Tube unit remanence signal
Starter max. current
Stator current / 1V = 10A

F_ROT
I_ANL

X100

Analog ground
U/F converter controller output
Inverter control frequency (OutA or OutB) / range: 38-100kHz
Control frequency OutA / range: 19-50 kHz

X210

I_Load
I_Last
I_Load nom
I_Load act
A_GND

RMS-value of Inverter oscillation current / 1V = 50A


Oscillating current reference value of controller: / 1V = 50A
Oscillation current actual value / 1V = 50A
Analog ground

UZ_IST

X110

UDCact
Ucon
Upre
A_GND

Intermediate circuit voltage / 1V = 100V (565V +/-10%)


Controller output with pre-control
Controller output without pre-control
Analog ground

X186

P_UT
N_UT
UTact
A_GND

KV+_ACT
KV-_ACT
KV_IST

Positive tube unit voltage / 1V = 10kV


Negative tube unit voltage / 1V = 10 KV
Composite tube unit voltage / 1V = 20kV
Analog ground

I_Fact
Y_LIM
I_Load

F _RMS

Author:
Date:

X140

F_INV
1

A_GND
U_VFO
F_INV
OutA

Schnapperelle
CO PTS
02.08.05

I_T_ACT

2 MAS
1 MAS

X221

RevisionNo.00

Heating (FIL)
RMS value of heating current / 1V = 1.5A (Stand By ~2A) depends on Tube type
Limited controller output of heating controller
Oscillation current actual value / 1V = 1.5A
Control frequency of heating inverter (OutA or OutB) / range: 30-100kHz;
Signal is low active
mA_Meas
tube current between X220.1 and X220.2 / mA_Meas Switch has to be in Position 2
tube current between X220.1 and X220.2 / mA_Meas Switch has to be in Position 2

All possible test points that are labeled on the board are listed
in the tables.
Sometimes the test points are labeled, but there is no signal
present on the test points.
Only the test points defined under "Designation old" are
relevant for service.
Otherwise the test points are explicitly listed in the documents.
A

Sheet No. 11
Description

Test points D500

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

X24

X150

D470

X111

X22

X21

X14

X25

X31

Main
Contactors
X19 Console

Test point
Voltage

V24 Power Fail 2


V24 Power Fail 1

ON

F1
F2

X3

V33

X1

X2

V32

X4

OFF

F3

D501 part no.: 77 59 835

S1

S2

X33
X112

V24

F4

V34

Schnapperelle
CO PTS
02.08.05

Ground

X102

Author:
Date:

V 31

X20

X10

Copyright
SIEMENS AG
All rights
reserved.

1
Only for
internal
use
appointed.
All rights
reserved.

X8

X34

X40

X5

X17

X30

X23

1 2

X110

Iontomat C

Iontomat A

X7

X6

Iontomat B

Sheet No. 12
Description

RevisionNo.00

D501 Circuit board 77 59 835

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

4
X7

X14

D470

X21

X25

X22

X23

X24

Iontomat A

X6

Iontomat B

Main
Contactors

X17

X30

X40

X5

X31
X8
X19 Console

S2

X3

F2
V34

Author:
Date:

Ground

X102

F4

Schnapperelle
CO PTS
02.08.05

D501 part no.: 57 60 181

F1

V32
V33

X1

X2

ON

F3

OFF

X4

S1

V31

V24_Power fail

X20

X33
X112

Copyright
SIEMENS AG
All rights
reserved.

X10

Only for
internal
use
appointed.
All rights
reserved.

Sheet No. 13
Description

RevisionNo.00

D501 Circuit board 57 60 181

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

LEDs (not on board partno.: 5760181)


4

V69
V70
V71
V72
V73
V127
V128

red
green
green
green
green
green
red

V_24V Power Fail_1


V_24V_D
VCC
V_N15V
V_P15V
V_24V_2
V_24V Power Fail_2

24V permanent Power


+5V VCC Power
-15V Power
+15V Power
24V for 80kW conductor Power

Switches
S1

OFF
ON

The generator cannot be switched on. Required only in a service situation.


The generator can be switched on. Normal switch position.

S2

When working in the generator cabinet, it is sometimes helpful to switch on the generator
with the S2 switch. In addition, the generator cabinet can be completely initialized without
The control console connected for troubleshooting if the switch is pressed and held.

Test Points (not on board partno.: 5760181)

Only for
internal
use
appointed.
All rights
reserved.

Voltage

Copyright
SIEMENS AG
All rights
reserved.

X123
X121
X119
X117
X115
X113

GND24
V_24V_2
V_24V_Dauer
V_P15V
V_N15V
VCC

24V Supply for 80 kW conductors


24 Volt permanent (+/- 5%) not adjustable
+ 15V Analog (+/- 5%) not adjustable
- 15V Analog (+/- 5%) not adjustable
VCC + 5V range (5,0 V - 5,2 V) adjustable

X206
X204
X202
X200

GND
MP3
MP2
MP1

not used
not used
not used

Author:
Date:

External Connection

Schnapperelle
CO PTS
02.08.05

X6
X7
X8
X14
X19
X22_A
X23_B
X24
X25
X30
X31
X32

Lamps 230V AC
Lamps 24V DC
Door contact / Tube temp
Dose Area Product Unit
Control Console
IONTOMAT A
IONTOMAT B
KK1B. G1/G2
KK2.G1/G2
G1 OEM
G2 OEM
Xray_End

KermaX
Bucky table
Bucky wall Stand
Tomo unit

Sheet No. 14
Description

RevisionNo.00

LEDs / Switches D501

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
C

I_max_FIL

X3.5
Driver A

X3.1

X6.6/5 RTN_FIL

D500 X13
Driver B

X6.3 FIL_Small

X3.3
X6.1 FIL_Large
L= 49H

I_max_FIL

L= 2*3,3H

High voltage
acquisition
D532 / D530
.X1 FIL Small
.X3 FIL Large

X3.5
X3.13

Test point: X140:


FIL / I_load

X3.15

Swing current
acquisition

X4.1 UDC_CEN
X4.3 UDC_NEG

D510 X3

X4.6 UDC_POS

Snubber circuit
VCC
X1.1 / X2.1
UHF

D115 X21
D501 X17

Only for
internal
use
appointed.
All rights
reserved.

X1.3 / X2.3

GND

Copyright
SIEMENS AG
All rights
reserved.

Standby:
Test point D500.X140: FIL / I_load
(1V  1,5A)
2

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 15
Description

RevisionNo.00

D470 Filament

Part-No.

Art

57 60 074

EFS

Version

Teil

01S

00
C

Type-No.

X2287

s
B

D510.X3
4
X5
U_DC +
Driver 4

X20.41
Driver 5
X20.39
X20.43

ANL
5

Gate5

Driver 3

D500.X12
X20.37
X20.45

Test point: X150:


RAC / I_load

X21

D470.X1
3

ANL
3

X21

ANL
1

ANL
4

Gate4

X7

X8

X9

terminal
blocks
AN_ I

Driver 2
ANL
2

Gate2

AN_0

Gate3

Driver 0

Driver 1
X20.35

X6
U_DC -

ANL
0

Gate0

AN_II

Gate1

+15
V
GND

Current
acquisition

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Author:
Date:
Schnapperelle
CO PTS
02.08.05

*Option
2 phase Stators

Sheet No. 16
Description

RevisionNo.00

D115 Rotation Anode Circuit RAC

Part-No.

Art

Teil

Version

57 60 074

EFS

01S

00
C

Type-No.

X2287

s
C

D511

D500.X10
Monitoring U_Bmax, I_Load, U CE,
P15V, N15V, VCC, auxiliary supply
(50kHz)

L_par

Heatsink

X10
X20

X21

X10

X23

X22

X34

X35

Snubber circuit

U_DC +
I_LOAD
U_Bmax

Driver B

Driver A
C_s

L1
L2
L3

Out_B

Out_A

Driver A

L1

Driver B

U_DC -

Charging

X44

L3

X46

X32

X33

D500
X14
3

Only for
internal
use
appointed.
All rights
reserved.

Test Points
X210: I_Load
X186: P_UT
X186: N_UT
X186: UTact

X13

X7

X240: IT_act

X15

D530

Copyright
SIEMENS AG
All rights
reserved.

kV +
kV -

HVTransformer
2

2
Operating point 81kV 100mA
Test Point D500.X210: I_Load

Operating point 100kV 500mA


Test Point D500.X210: I_Load

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 17
Description

RevisionNo.00

D510 Inverter IT30/55

Part-No.

Art

57 60 074

EFS

Version

Teil

01S

00
C

Type-No.

X2287

s
B

D511

D500.X10
Monitoring U_Bmax, I_Load, U CE,
P15V, N15V, VCC, auxiliary supply
(50kHz)

TF 1

TF 2
L_par

Heatsink

X10
X20

X21

X10

X23

X22

X34

X35

Snubber circuit

U_DC +
X40

I_LOAD
U_Bmax

Driver B

Driver A
C_s

L1
L2
L3

Out_B

Out_A

Driver A

Driver B

X42

L1

U_DC -

Charging

X44

L3

X46

X32

X33

D500
X14

K10

Test Points
X210: I_Load
X186: P_UT
X186: N_UT
X186: UTact

X240: IT_act
Only for
internal
use
appointed.
All rights
reserved.

X13

X7

X40

X4

X42

X5

X34

X1

X44

X2

CBuffer

X15

D580
80kW Opt ion

D530

D501

Copyright
SIEMENS AG
All rights
reserved.

CS

D501. X34

kV +
kV -

HVTransformer
2

Operating point 81kV 100mA


Test Point D500.X210: I_Load

Operating point 99kV 800mA


Test Point D500.X210: I_Load

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 18
Description

RevisionNo.00
A

D510 Inverter IT 65/80

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
B

Single Tank Tube


Assembly

GENERATOR
4

Shield

D115.X8

D115.X7

II

D115.X9

M16 PE

D470.X6

D500.X14

PE

heating

X1 / 3

Temp.

temparature / pressure switch

X7

pressure
3

X11

X32

D510

X33

X13

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 19
Description

RevisionNo.00

Connection Single Tank

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

s
C

OPTITOP

GENERATOR
4

Shield

K3.R3

K5.1

II

K3.R5

M16 PE

PE

D501.X8.5
D501.X8.6

Oil pressure switch

pressure

D501.X8.3
D501.X8.4

temparature switch

Temp.

kV +
HV- Tank

kV -

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 20
Description

RevisionNo.00

Connection Optitop-Tube

Part-No.

Art

57 60 074

EFS

Version

Teil

01S

00
C

Type-No.

X2287

s
A

Only for
internal
use
appointed.
All rights
reserved.

Copyright
SIEMENS AG
All rights
reserved.

Author:
Date:
Schnapperelle
CO PTS
02.08.05

Sheet No. 21

RevisionNo.00
Description

OptionEnergy Storage Unit

Part-No.

Art

57 60 074

EFS

Teil

Version

01S

00
C

Type-No.

X2287

Вам также может понравиться