Академический Документы
Профессиональный Документы
Культура Документы
Prescription and Objectives: According to the protocol GP was planned to 60Gy over 30 total
fractions. 100% of the PTV_6000 volume was to receive 95% of the dose. The organs at risk
included Esophagus, heart, lungs and esophagus. Spinal cord constraints were to be kept at a
maximum dose less than 50.5Gy. The V20 of the sum lung volume was to be less than 30%. The
esophagus and heart were to meet all qualifications as outlined by the protocol. (Table 3)
Beam Arrangement and Planning Technique: A 9 field IMRT technique was generated in
order to meet the constraints of the protocol. The beams chosen were dependent upon organs at
risk based on the positioning of the tumor. Beam angles 175 and 350 first were setup first
because those angles oppose one another at the optimal position to spare the cord. After the first
two angles were positioned, seven more equidistant beams were positioned between angles 175
and 350. These nine angles, [175, 196, 218, 240, 26, 284, 306, 328 and 350] allow for
diversification of dose. All beam angle entrances were kept through the right lung in order to
avoid entrance through the contra-lateral lung as much as possible.
The PTV_6000 optimization technique was set to 100% of the prescription dose since 100% of
the volume was to get 95% of the dose. After setting up the tumor volume dose constraint, spinal
cord constraints follow. The upper limit constraint of the spinal cord is set to 4400cGy. Then
the Spinal Cord margin volume, which is defined as a 1cm axial margin and 3cm posterior
margin, was used to guide the dose to the area around the cord. The Lung V20 constraint for the
sum lung was set to 25%-30%. Then multiple upper constraints are set to help lower the dose to
the lung. The final organs constrained are the Esophagus and the heart. These constraints were
based off the limits within the protocol and then adjust accordingly. Lastly, the normal tissue
optimizer (NTO) was set. The NTO was set to start at 0.8cm distance from the target volume.
The start dose was set at 100%, the end dose at 65% and a 0.15 fall off. Constraints were
adjusted as the plan actively optimized.
Plan Outcome: Upon final review of the plan, in order to receive the outlined objective for the
PTV_6000, I normalized the plan to 98.5%. 100% of the PTV_6000 volume received 95% of
the dose. All organs at risk objectives were met as outlined in the protocol. (See Table 3 and
Figure 3)
GTV_6000
Table 3: OR Objectives
Figure 3: DVH