LOAD TEST ON
3- INDUCTION MOTOR
Name of the
Equipment
3- Variac
2
3
4
Ammeter
Voltmeter
Wattmeter
Tachometer
Connecting Wires
Range
Typ
e
Quantity
415V/(0470)V
(0-10)A
(0-600)V
600V/10A
(09999)rpm
MI
MI
UPF
Digit
al
1
1
2
1
Required
Some
Circuit Diagram:
Precautions:
1. 3- Variac should be in minimum position.
Procedure:
1. Connect the circuit as shown in circuit diagram.
2. Observing the precautions close the TPST Switch and switch ON 440V A.C
supply.
3. Apply the rated voltage to the stator windings of 3- Induction Motor with the
help of 3- Auto-transformer.
4. Note down the readings of wattmeter, ammeter and voltmeter on no-load.
5. Load the Induction Motor in steps using the brake-drum arrangement. At each
step note down the readings of all meters, speed and spring balance readings
till full load current.
Dept of E.E.E
2
Torq
ue
(Nm)
Outp
ut
(W)
(%
)
Formulae:
Torque, Tsh = 9.81 (S1 ~ S2) R
N-m
Input Power, Winp = W1 + W2
W
Output Power, Wout = (2 NTsh)/60
W
%Slip = Ns N X 100
Ns
cos =
Wout
3 VL IL
Efficiency = % = (Output/Input) X 100
Model Graph:
Result:
Dept of E.E.E
Name of the
Equipment
1- Variac
1- Transformer
Ammeter
Voltmeter
Wattmeter
Connecting Wires
Range
Typ
e
Quantity
230V/(0-230)V,
10A
115V/230V,
2kVA
(0-10)A
(0-2)A
(0-150)V
(0-30)V
150V/2A
75V/10A
Cor
e
MI
MI
MI
MI
LPF
UPF
1
1
1
1
1
1
1
Required
Some
Circuit Diagram:
O.C Test:
Dept of E.E.E
Precautions:
1. Connections should be made tight.
2. 1- Variac should be in minimum position.
Procedure:
1. Connect the circuit as shown in circuit diagram for O.C test.
2. Observing the precautions switch ON 1- A.C supply and by using the 1-
variac apply the rated voltage of the primary of the transformer.
3. Note down all the meter readings. Here wattmeter reading gives iron loss.
4. Observing the precautions switch OFF the supply.
5. Connect the circuit as shown in circuit diagram for S.C test.
6. Observing the precautions switch ON 1- A.C supply and by using the 1-
variac apply the rated current to the transformer. (Rated power of the
transformer/Voltage of primary of transformer)
7. Note down all the meter readings, here wattmeter reading gives full-load
copper loss.
8. Observing the precautions switch OFF the supply.
Tabular Columns:
O.C Test:
Vo
Io
Wo = W X
(V)
(A)
M.F (W)
S.C Test:
VSC
ISC
(V)
(A)
WSC = W X
M.F (W)
Efficie
ncy
6
(W)
(W)
%
Regulation
(%)
(W)
(W)
(W)
(%)
%
Regulation
Model Calculations:
Let the transformer be the step-down transformer (O.C Test)
Primary is H.V side and secondary is L.V side
Ro=V1/Iw () where Iw=Iocoso
Xo=V1/I () where Iw=Iocoso
Ro1=WSC/ISC2 ()
ZO1=VSC/ISC
Xo1=( ZO12 - Ro12)
Ro2=K2 Ro1
Xo2=K2 Xo1
where,
K=V2/V1=Transformation Ratio
Calculations to find efficiency:
For example, at 1/4th full load,
Copper Losses=WSC x (1/4)2 (w)
where, WSC=Full Load Copper Losses
Constant Losses=Wo
(W)
Output=(1/4) X VA X cos
(cos may be assumed)
Input=Output + Copper Loss + Constant Loss
Efficiency (%) = (Output/Input) X 100
Efficiency at different loads and power factor can be calculated.
Model Graph:
Dept of E.E.E
Dept of E.E.E
Name of the
Equipment
3- Variac
Ammeter
Voltmeter
Wattmeter
5
6
Range
Typ
e
Quantity
Rheostat
415V/(0470)V
(0-10)A
(0-5)A
(0-600)V
(0-150)V
(0-30)V
600V/10A
300V/10A
100/5A
MI
MC
MI
MI
MC
LPF
UPF
WW
Connecting Wires
1
1
1
1
1
2
2
1
Required
Some
Circuit Diagram:
No-Load Test:
Dept of E.E.E
Precautions:
1. 3- Variac should be in minimum position.
Procedure:
No-Load Test:
1. Connect the circuit as shown in circuit diagram.
2. Observing the precautions close the TPST Switch and switch ON 440V A.C
Supply.
3. Apply the rated voltage to the stator windings of 3- Induction Motor with
the help of 3- Auto-Transformer.
4. Note down the no-load readings of wattmeter, ammeter and voltmeter.
5. Bring back the auto-transformer to its minimum position and switch OFF the
supply.
Blocked Rotor Test:
1. Connect the circuit as shown in circuit diagram.
2. Observing the precautions close the TPST Switch and switch ON 440V A.C
Supply.
3. By varying the variac, apply small voltage until the full load current flows in
the stator windings with the rotor unmoved.
4. Note down the readings and switch OFF the supply after bringing back the
auto-transformer output to minimum position.
Measurement of Stator Resistance:
1. Connect the circuit as per the circuit diagram.
2. Keeping rheostat in maximum resistance position switch ON 220V D.C supply.
Dept of E.E.E
10
Voltage,
VSC (V)
W1
(W)
W2
(W)
No-Load Test:
Rated Voltage=
Io
(A)
Stator Resistance per phase =
Formulae:
G o = Wo
;
Y o = Io
;
3V2
V
Z01 = VSC
;
R01 = WSC
;
2
ISC
3I SC
For Circle Diagram:
cos o = Wo
3 Vo Io
cos SC = WSC
3 VSC ISC
ISN = ISC [Vo / Vsc]
V
(V)
W1
(W)
W2
(W)
Bo = (Yo2 Go2)
X01 = (Z012 R012)
Model Graph:
Dept of E.E.E
11
Result:
Dept of E.E.E
12
PARALLEL OPERATION OF
1- TRANSFORMER
Name of the
Equipment
Range
Type
Quantity
(0-300)V
(0-150)V
115V/230V,
2kVA
(0-20)A
(0-10)A
MI
MI
1
1
Core
MI
MI
Wire
Wound
Knife
1
2
Voltmeter
1- Transformer
Ammeter
Resistive Load
1-, 5kW
SPST Switch
5
Dept of E.E.E
1
1
13
DPST Switch
Wattmeters
Connecting Wires
300V/10A
150V/20A
Knife
UPF
UPF
1
2
1
Required
Some
Circuit Diagram:
Precautions:
1. Load should be in OFF position.
2. SPST Switch is in open position.
Procedure:
1. Connect the circuit as shown in circuit diagram.
2. Observing the precautions switch ON the supply.
3. Two readings of the voltmeter connected across switch K is observed. If it
reads zero which means that the two secondary emfs are opposing each
other, then switch K is closed. If it reads twice the secondary rated voltage
the secondary terminals of any of the transformer are interchanged and then
the switch K is closed to make it read zero reading.
4. The load switch is closed and total load on the transformer is increased in
steps. At each step, the readings of the various meters are noted down. The
total load is increased till the current supplied reaches the maximum load
current of both the transformers when connected parallely.
5. Observing the precautions switch OFF the supply.
Tabular Columns:
From S.C Test:
From O.C Test:
S.N
ISC
Vsc
WSC
Core Loss, Wo (W) =
o.
(A)
(V)
(W)
1
2
3
Dept of E.E.E
14
S.N
o.
IA IB IL
(A (A (A
)
)
)
VL
(
V
)
(
W
)
(
W
)
1
2
3
Formulae:
IA| = IL X ZB
ZA+ZB
W A| = W L X Z B
ZA+ZB
WSC = Isc2R
;
ZSC = ZA = ZB = VSC / ISC
%Reg = VNL VFL X 100
VFL
VNL = No Load Voltage
VFL = Load Voltage at different loads
Output, Wout = VL IL
;
Efficiency, % = Wout X 100
Win
Model Graph:
WL
(
W
)
WA
IA| IB|
|
(A (A
(W
)
)
)
WB
|
(W
)
(
%
)
%
Reg
I B| = IL X ZA
ZA+ZB
;
W B| = W L X Z A
ZA+ZB
= I2R (Since I=IA|+IB|)
;
W Cu
Result:
Dept of E.E.E
15
Name of the
Equipment
Voltmeter
1- Transformer
3
4
Ammeter
DPST Switch
Dept of E.E.E
Range
(0-150)V
115V/230V,
2kVA
(0-2)A
-
Typ
e
MI
Quantity
1
Core
MC
Knife
1
1
16
Wattmeter
300V/2A
400/1.7A
100/5A
Rheostats
Tachometer
(0-9999)rpm
Connecting Wires
LPF
WW
WW
Digit
al
-
1
1
1
1
Required
Some
Circuit Diagram:
Precautions:
1. Field rheostat of the motor must be in minimum position.
2. Armature rheostat of the motor must be in maximum position.
Procedure:
1. Connect the circuit as shown in circuit diagram.
2. Before starting the experiment, decide at which frequency you conduct NoLoad test and calculate the speed and corresponding voltage to maintain the
V/f ratio constant.
3. Observing the precautions switch ON the supply.
4. Run the prime-mover (D.C Motor) at the required speed and adjust the level
of excitation of alternator so as to develop rated transformer primary voltage
and then note down the wattmeter reading after exciting the transformer.
5. In a gradual manner, simultaneously adjust the excitation of the alternator
and its speed so as to maintain a constant V/f ratio.
6. Repeat the same procedure for different frequencies.
7. After completing the experiment, observing the precautions switch OFF the
supply.
Tabular Columns:
Tabular Column-I:
S.N
o.
Dept of E.E.E
Spee
d
N(rp
m)
Freque
ncy
f(Hz)
Volta
ge
V(V)
V/
f
W
i
Wi /
f
17
Freque
ncy
f(Hz)
Hysteresis
Losses,
Wi = Af (W)
Eddy Current
Losses,
We = Bf2 (W)
Total Iron
Losses,
Wi = Wh +
We (W)
1
2
Formulae:
P = 120f
;
N1 = 120f1
N
P
K = V = Rated Voltage
f
Rated frequency
where, f = rated frequency
f1 = selected frequency
N = rated speed
N1 = speed corresponding to f1
Let, Wh = Af
We = Bf2
Wi = Wh + We = Af + Bf2
Wi = A + Bf
f
A oa from graph
B dy = y
dx x
Model Graph:
Result:
Dept of E.E.E
18
SCOTT CONNECTION OF
TRANSFORMERS
Name of the
Range
Type
Quantity
19
Equipment
Voltmeter
3- Variac
Ammeter
(0-600)V
(0-300)V
415V/(0470)V
(0-10)A
Resistive Load
1-, 5kW
Wattmeters
Connecting Wires
MI
MI
2
2
1
3
600V/10A
MI
Wire
Wound
UPF
2
2
Required
Some
Circuit Diagram:
Dept of E.E.E
20
Precautions:
1. 3- Variac should be in minimum position.
2. Load must be in OFF position.
Procedure:
1. Connect the circuit as shown in figure 1.
2. Observing the precautions close the TPST Switch and switch ON 440V A.C
supply.
3. Apply the rated voltage to the primary of the main teaser transformer with
the help of 3- Auto-transformer.
4. Note down the readings of all meters for different load settings till rated full
load current of the transformer.
5. Connect the secondaries of the two transformers as shown in figure 2 to
verify the phase relationship of the two phase voltages.
6. Calculate the efficiency of conversion.
7. Observing the precautions switch OFF 440V A.C supply.
Tabular Columns:
For Figure 1:
W1 W2
I1 I2 I3
(
(
(A (A (A
W W
)
)
)
)
)
Dept of E.E.E
V1
(
V
)
V2
(
V
)
V3
(
V
)
Outp
ut
(W)
Inp
ut
(W)
(
%
)
V
(
V
)
For Figure 2:
V1
I
V2
(
(A
(V
V
)
)
)
21
Result:
Dept of E.E.E
22
EQUIVALENT CIRCUIT OF A
1- INDUCTION MOTOR
23
Dept of E.E.E
24
Precautions:
1. Initially variac must be at zero position.
2. No Loose connections are allowed.
3. Readings should be taken without any parallax errors.
4. Rheostat should be in maximum resistance position.
Procedure:
For No-Load Test:
1. Connect the circuit as shown in circuit diagram.
2. Observing the precautions close the DPST Switch and switch ON 1- A.C
supply.
3. Apply the rated voltage by increasing the variac output gradually (No Load is
connected to the motor).
4. Note down the readings of all the meters.
5. Observing the precautions switch OFF the supply.
For Blocked Rotor Test:
1. Connect the circuit as shown in circuit diagram.
2. For performing the blocked rotor test, block the rotor (i.e., not allowed to
rotate) by applying belt-brake arrangement and increase the voltage slowly
till rated current flows and note down all the meter readings.
Dept of E.E.E
25
Tabular Columns:
No-Load Test
Vo
(V)
Io
(A)
Wo
(W)
Blocked Rotor
Test
VSC
ISC
WSC
(V)
(A)
(W)
I
(A)
V
(V)
Rs
()
Formulae:
Wo
cos o
Vo I o
Vo
I o sin o
V
Z s SC
I SC
Xo
XS
Z S2 R S2 X eq
RS
2
X
x1 x 2 S
2
Vo
Ro
I o cos o
N N
slip S
NS
r1 r2
R S Req
WSC
2
I SC
Dept of E.E.E
26
LOAD TEST ON
1- TRANSFORMER
Dept of E.E.E
27
Name of the
Equipment
Range
Type
Quantity
Core
1
2
1
1
1
1- Variac
1- Transformer
Ammeter
Voltmeter
Wattmeter
230V/(0-230)V,
10A
115V/230V,
2kVA
(0-10)A
(0-150)V
(0-300)V
300V/10A
Resistive Load
1-, 5kW
DPST Switch
MI
MI
MI
UPF
Wire
Wound
Knife
Connecting Wires
1
1
Required
Some
Circuit Diagram:
Dept of E.E.E
28
Precautions:
1. Connections should be made tight.
2. 1- Variac should be in minimum position.
Procedure:
1. Connect the circuit as shown in circuit diagram.
2. Observing the precautions switch ON 1- A.C supply.
3. Apply rated primary voltage of the transformer by properly adjusting the
variac.
4. Note down the readings of all the meters on No-Load.
5. Switch ON the load by closing the DPST Switch and increase the load in steps
and note down the corresponding reading of all meters.
6. Decrease the load on the transformer secondary and then switch OFF the
supply.
Tabular Column:
Input
Input
Voltage Current
(V)
(A)
Input
Power(
W1)
Output
Voltage(
V2)
Output
Current
(I2)
Output
Power(
W2)
Efficie
ncy
(%)
%Regula
tion
(%)
Formulae:
Efficiency, % = Output X100
Input
%Regulation = V02 V2 X 100
V2
Model Graph:
Dept of E.E.E
29
Result:
Dept of E.E.E
30
SUMPNERS TEST
SUMPNERS TEST
Aim:
i.
ii.
Apparatus:
S.
No.
Name of the
Equipment
1- Variac
1- Transformer
Ammeter
Voltmeter
Wattmeter
Dept of E.E.E
Range
230V/(0-230)V,
10A
115V/230V,
2kVA
(0-5)A
(0-20)A
(0-300)V
(0-75)V
300V/5A
75V/20A
Typ
e
Quant
ity
Cor
e
MI
MI
MI
MI
LPF
UPF
2
1
1
2
1
1
1
31
Precautions:
1. The range of voltmeter should be doubled the induced e.m.f of secondary
winding.
Procedure:
1. Connect the circuit as shown in circuit diagram.
2. Connect the primaries of both transformers in parallel after ascertaining the
polarities.
3. Connect the secondaries in series and connect a switch as shown in the
circuit diagram.
4. Initially keep the switch in open circuit.
5. Give supply as per the primary rating of transformers.
6. Note down the readings of all the meters.
7. Close the SPST Switch when voltage Vab = 0 (see circuit diagram).
8. Increase the load till the full load rated current of the transformer flows by
varying the auto-transformer supplying secondaries.
Dept of E.E.E
32
Load
V1
(
V
)
NoLoad
FullLoad
For Unity Power Factor:
W1 (W)
Lo
Iron
ad
Losses
W2 (W)
Copper
Losses
Total
Losses
Outp
ut
(W)
Inp
ut
(W)
Efficie
ncy
(%)
W2 (W)
Copper
Losses
Total
Losses
Outp
ut
(W)
Inp
ut
(W)
Efficie
ncy
(%)
Formulae:
Losses in each transformer =
combined =
Wi Wc
2
VI 1
X 100
VI 1 Wi Wc
VI 1
Efficiency of each Transformer = (%) =
W W X 100
VI 1 i c
2
2
Model Graph:
Dept of E.E.E
33
Result:
Dept of E.E.E
34
REGULATION OF 3-
ALTERNATOR BY
SYNCHRONOUS IMPEDANCE
METHOD
35
Name of the
Equipment
Rheostats
Ammeter
Voltmeter
Tachometer
Connecting Wires
Range
Typ
e
WW
Quantity
100/5A
400/1.7
A
(0-10)A
(0-2)A
(0-5)A
(0-300)V
(0-30)V
(09999)rpm
WW
MI
MC
MC
MI
MC
Digit
al
1
1
1
1
1
1
Required
Some
Circuit Diagram:
Precautions:
Dept of E.E.E
36
S.C Test
If(
A)
ISC
(A)
V
(V)
R
()
E.M.F Method
P.F
cos
Ia
(A
)
Eo (V)
Laggi
ng
Leadi
ng
%
Regulation
Laggi Leadi
ng
ng
P.
F
M.M.F Method
Field
If (A)
Current
If1
If2
Laggi Leadi
(A)
(A)
ng
ng
Formulae:
Dept of E.E.E
37
VOC
I SC
X S Z S2 Ra2
% Re g
Eo V
X 100 (%)
V
where,
Eo = Generated E.M.F of alternator
(or excitation voltage per phase)
V = Full Load rated terminal voltage per phase.
Model Graph:
Result:
Dept of E.E.E
38
DETERMINATION OF Xd & Xq OF
SALIENT POLE SYNCHRONOUS
MACHINE
39
Name of the
Equipment
3- Variac
2
3
4
Rheostat
Ammeter
Voltmeter
Tachometer
Connecting Wires
Range
Typ
e
Quantity
415V/(0470)V
400/1.7A
(0-5)A
(0-150)V
(09999)rpm
WW
MI
MI
Digit
al
1
1
1
1
Required
Some
Circuit Diagram:
Precautions:
1. Connections must be made tight.
2. Field rheostat must be kept in minimum resistance position.
3. Starter must be in OFF position.
Procedure:
1. Connect the circuit as shown in circuit diagram.
2. Start the D.C Motor with the help of starter.
3. Adjust the field rheostat of the D.C Motor so that it runs slightly below the
synchronous speed of synchronous machine.
4. Give A.C Supply to the Alternator with the help of variac so that rated current
flows in the synchronous machine.
Dept of E.E.E
40
Vmi
ax
in
(V)
(V)
Ima
Imi
(A
)
(A
)
Xd
(
Xq
(
)
Formulae:
Vmax
()
I min
V min
()
Quadrature Axis Synchronous Reactance = Xq =
I max
Direct Axis Synchronous Reactance = Xd =
Result:
Dept of E.E.E
41