The procedure entails removal and fitting of the Digital PCB, the fitting of
the extra cover required and the fitting of the CI and Solids & DCI modules.
Initial Procedures
1) If the system is not already out of vacuum, vent the system.
3) If necessary, remove the transit locking bolt at the rear of the instrument.
2) For the connectors that have locating nuts, use a 5mm nut driver to remove
them.
3) Remove all the screws that hold the PCB to the unit.
3) Using the screws that held the previous PCB, fasten the PCB down into the
unit.
5) Fasten the cover to the rear of the instrument in the orientation shown.
Extra Cover
Locating Nuts
Connector for
Digital CI Valve
Connector for
Solids & DCI
Probe Cable