Вы находитесь на странице: 1из 4

Evaluating Temperature Characteristics

using a Temperature Chamber


and the HP 4291A

Product Note 4291A-2


HP 4291A RF Impedance/
Material Analyzer

HP 4291A Overview

The HP 4291A RF analyzer


(Figure 1) provides impedance
and material measurements up to
1.8 GHz. This analyzer provides a
complete measurement solution
for a wide range of applications,
from evaluating electronic
materials used in RF components
to evaluating various impedance
parameters of components.

Figure 1. HP 4291A RF Impedance/Material Analyzer with SMD fixtures

Introduction

In the research and development end-product reliability and


of electronic components and performance. This note
materials, development cycle time introduces an efficient and highly
must be reduced. In accordance reliable method for evaluating
with this trend, more emphasis is temperature characteristics using
being placed on improving the a combination of the HP 4291A
efficiency of temperature RF Impedance/Material Analyzer
characteristics evaluation, which and a Tabai Espec temperature
plays a major role in ensuring chamber.
Table 1. HP 4291A Specification Summary time required for software
Frequency Range 1 MHz to 1.8 GHz development and other integration
Measurement Parameters |Z|, θz, |Y|, θy, R, X,G, B, Cp, Cs, Lp, Ls, Rp, Rs, D,
tasks.
Q,|Γ|, θΓ, Γx, Γy
Overview of the Tabai
Basic Accuracy 0.8%
Espec Temperature
Impedance Range 0.1 ohm to 50 kohm
Chamber
Oscillator Level 0.2 mV to 1 V (<= 1 GHz) 0.2 mV to 0.5 V (> 1 GHz)
DC Bias (opt 001) 0 V to 40 V, 0 to 100 mA The Tabai Espec temperature
chamber, model SU-240-Y, is a
New capabilities for evaluating 3. HP-IB compact, bench-top unit. The
RF components chamber provides high
The HP-IB is included as standard
equipment and is indispensable performance levels, (for example:
Highly accurate impedance temperature range and
for configuring an automatic
measurement in the RF band distribution), which are equivalent
measurement system.
A wide frequency range, from to those available with standard,
1 MHz to 1.8 GHz, is available, 4. Optional HP IBASIC larger-sized temperature
making it possible to measure the With optional HP IBASIC (option chambers. The SU- 240-Y is
impedance of RF components at 1C2), the temperature chamber designed to integrate easily with
operating frequencies. Impedance can be controlled via the HP-IB. the HP 4291A. This is specifically
can be measured with a basic using the controller built into the advantageous for the research and
accuracy of 0.8 percent, and Q analyzer. development of electronic
(quality factor) with an accuracy components and materials
of 15 percent for measurements of 5. Application Program Available requiring evaluation over a range
Q = 100 at 1 GHz. of temperatures.
An application program for
Analyzer functions ease system controlling the temperature of the Chamber features
configuration and integration chamber and evaluating the
temperature characteristics based High performance with compact
The HP 4291A includes the on the measured results from the dimensions
following features to simplify HP 4291A is included with the
configuring a temperature This unit has a temperature range
optional high temperature test
characteristic testing system. of -30°C to +150°C, which is
heads.
(See Figure 2.) equivalent to a standard chamber.
These features make system The upper limit temperature of
1. Standard 1.8m cable configuration with the +150°C is available with the
A 1.8 m cable makes it possible to temperature chamber extremely standard unit. This high
extend the test station to the easy, and significantly reduce the performance also extends to the
temperature chamber without
increasing the measurement error.

2. High temperature test heads


The high temperature test heads
(option 013: for high impedance,
option 014: for low impedance)
can be used within the range of
-55°C to 200°C, maintaining high
accuracy.

Figure 2. Analyzer-Temperature chamber system configuration

2
temperature distribution and Table 2. Tabai Espec Chamber SU-240-Y Specification Summary
temperature adjustment range, Temperature and Humidity Control Balanced Temperature Control Humidity Control
accommodating full System (BTC system)
environmental test requirements Refrigeration System Mechanical single-stage condensation method
in a compact size. Temperature Range -30°C to +150°C
Compact size saves space Temperature Control Stability ±0.3°C (-30°C to 100°C) ±0.5°C (101°C to 150°C)
The compact design is beneficial Temperature Uniformity <1> ±0.5°(-30°C to 100°C) ±0.8°C (101°C to 150°C)
for laboratories with a lack of Temperature Heat-up time within 70 minutes from -30°C to 150°C
space. The freezer uses a rotary Temperature Pull-down time within 70 minutes from 20°C to -30°C
compressor, reducing size and
Lowest Attainable Temperature <2> -30°C
weight even further.
Capacity 21 liters
HP-IB as standard Inside Dimensions W=30 cm x H=30 cm x D=24 cm (excluding
HP-IB, a standard feature, is used protrusion)
for connecting to the HP 4291A Outside Dimensions W=44 cm x H=69 cm x D=74 cm (excluding
(using the HP IBASIC function) or protrusion)
to an external controller, allowing <1>: At 23°C ambient without DUT <2>: Ambient below 30°C without DUT
remote control of the chamber
functions. System configuration Option 1C2: HP IBASIC, HP-HIL
keyboard/cable
Measuring port To configure a system using the Temperature characteristics can
A measuring port (Figure 3) is temperature chamber, the be evaluated more efficiently and
available for installing the high following options should be added reliably by using the HP 4291A to
temperature test head of the to the HP 4291A: control the functions of the
HP 4291A, eliminating the need to chamber. A system configuration
Option 013: High temperature test
create additional measurement for this purpose is shown in
head for high-impedance
cable access holes. Figure 4.
measurement or Option 014: High
temperature test head for Measurement procedures
low-impedance measurement
The following are general
procedures for evaluating
temperature characteristics using
the HP 4291A and the Tabai Espec
temperature chamber:
1. Configure an automatic
measurement system as
Figure 3. Measuring Port of the Tabai described previously
Espec temperature chamber SU-240-Y
2. Load the application program
Thermocouple for measurement Using the I BASIC function of the
of temperature around the DUT HP 4291A, load the application
A thermocouple is available for program that comes with the high
measuring the temperature around temperature test head for the
the DUT. Using this facility, HP 4291A. The application
temperature characteristics can program controls settings,
be evaluated accurately by measurements, and analysis
Figure 4. System configuration interactively.
monitoring the temperature near
the DUT.

3
H

3. Setup the HP 4291A/ 6. Graphically analyze the For more information on Hewlett-Packard
Test and Measurement products,
temperature chamber measurement results applications, or services please call your
local Hewlett-Packard sales office. A
The program is first used to setup After the measurement is current listing is available via the Web
the HP 4291A and temperature complete, use the program to through AccessHP at http://www.hp.com. If
chamber. For the HP 4291A, the graphically display the you do not have access to the internet,
please contact one of the HP centers listed
parameters for measurement and measurement results as illustrated below and they will direct you to your
sweep must be defined. For the in Figure 6 to evaluate the nearest HP representative.
temperature chamber, the upper parameters of interest.
United States:
limit/lower limit/step values of Hewlett-Packard Company
temperature as well as the Test and Measurement Organization
repeating pattern of temperature 5301 Stevens Creek Blvd.
Bldg. 51L-SC
control must be set. (Figure 5). Santa Clara, CA 95052-8059
1 800 452 4844

Canada:
Hewlett-Packard Canada Ltd.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(905) 206 4725
Figure 5. Repeating patterns of
temperature control Europe:
Hewlett-Packard
To ensure highly accurate Figure 6. Evaluating the temperature European Marketing Centre
characteristics of a capacitor P.O. Box 999
measurements, the HP 4291A 1180 AZ Amstelveen
requires further calibration at the Conclusion The Netherlands
APC-7 measurement terminal as
Japan:
well as OPEN/SHORT/LOAD By using the HP 4291A RF Hewlett-Packard Japan Ltd.
compensation in the test fixture. Impedance/Material Analyzer in Measurement Assistance Center
These should be carried out at the 9-1, Takakura-cho, Hachioji-shi,
combination with the Tabai Espec Tokyo 192, Japan
standard temperature of the temperature chamber SU-240- Y, Tel: (81) 426 48 3860
temperature characteristics. (The an automatic measurement Fax: (81) 426 48 1073
standard temperature is normally system can be easily configured. Latin America:
23 degrees ±5 degrees.) Temperature characteristics of Hewlett-Packard
various electronic components Latin American Region Headquarters
4. Connect the DUT 5200 Blue Lagoon Drive
and materials can then be 9th Floor
Connect the DUT to the test efficiently and accurately Miami, Florida 33126
fixture positioned in the evaluated over frequencies from U.S.A.
(305) 267 4245/4220
temperature chamber. 1 MHz to 1.8 GHz.
Australia/New Zealand:
5. Make the measurement For more information about the Hewlett-Packard Australia Ltd.
Tabai Espec chamber, contact: 31-41 Joseph Street
When the measurement has Blackburn, Victoria 3130
started, the program automatically Tabai Espec Corp.
Australia
131 347 ext. 2902
controls the temperature chamber 3-5-6, Tenjinbashi,
and HP 4291A based on the Kita-Ku, Osaka 530, Japan Asia Pacific:
settings made in step 3. Phone: 06-358-4785/4741 Hewlett-Packard Asia Pacific Ltd
17-21/F Shell Tower, Times Square,
Temperature characteristic data is Fax: 06-358-4786/5500 1 Matheson Street, Causeway Bay,
collected. Hong Kong
ESPEC Corp. (America) (852) 2599 7070
425 Gordon Industrial Court, S.W.
Grand Rapids, MI 49509 USA
Phone: 616-878-0270,
Toll Free: 1-800-537-7320
Fax: 616-878-0280 © Copyright 1993
Hewlett-Packard Company
Data subject to change
Printed in U.S.A. 12/93
5962-6922E

Вам также может понравиться