Вы находитесь на странице: 1из 6

Impedance Measurements

Using the HP 4291A and the


Cascade Microtech Prober

HP Product Note 4291A-3


HP 4291A RF Impedance/
Material Analyzer

Introduction
There is an increasing demand
and a need to make impedance
measurements using a prober.
One example would be making
impedance characteristic
evaluations of a narrow pitch
transmission line at high
frequency in order to evaluate the
design of RF circuits and IC
packages. Another example is
performing a precise capacitance
characterization of dielectric thin
film up to the GHz range. In this
product note, a new solution for
this type of measurement is
described in which the HP 4291A
RF impedance/material analyzer
and the Cascade Microtech prober
are used.
HP 4291A Capabilities System Composition Cascade Microtech Products:

The HP 4291A 1MHz-1.8GHz RF The HP 4291A can accurately Summit series 9000 probe station
Impedance/Material Analyzer can measure the impedance of a (1 ea.)
measure impedances covering a narrow pitch transmission line
Positioner, microscope for probe
wide frequency range with great using a Cascade Microtech prober.
station (1 ea.)
measurement accuracy by using This system contains the
the newly developed RF-IV following equipment: Adapter kit for installation of
method (basic measurement HP 4291A test station
accuracy: 0.8%). Impedances that Hewlett-Packard Products:
(No. 106-768, 1 ea.)
differ greatly from 50 ohms are HP 4291A RF Impedance/Material
considered difficult to measure Analyzer (1 ea.) Cable with SMA (M)-(F)
using the reflection coefficient connector (No. HF-502-6, 6 inches
method used in most Network APC7-SMA Adapter long, 50 ohm, 1 ea.)
Analyzers. For example, inductors (HP P/N 1250-1746, 1 ea.)
in nH, capacitors around 1 pF, Probe (1 ea.), see table 1.
etc., can be measured accurately
using the HP 4291A. Also, Impedance standard substrate
regarding Q (quality factor), (choose suitable one for probe,
Q=100 can be measured at 1GHz 1 ea.)
with an accuracy of 15% (typical
value). In addition, a precise, There are major 3 different types
stable probe system can be easily of Cascade Microtech probe as
constructed using a standard shown in table 1. The suitable
1.8 m connection cable as the test probe for the measurement should
station. This station can be be chosen from the table.
brought close to the probe without
affecting the measurement Table 1. Selection of Cascade Microtech Probe
accuracy. Moreover, because
parasitic elements like the Fine-pitch
residual inductance of the probe Compliance Probe Wafer Probe Microprobe
(EPC series) (ACP, WPH series) (FPM-1X)
(which is considered to be one of
the major error factors), can be
removed by using the
compensation function. These
capabilities make it possible to
evaluate the electrical
characteristics of the DUT
correctly.
Electrode Distance Fixed, Fixed, Variable, 0-18 mm.
Selectable between Selectable between
150-1250 um. 150-250 um.
Main Features Suitable for a device Contacts with an Can change electrode
mounted on a PC 11-15° angle. Suitable distance. Possible to
board (or similar for on-wafer contact material which
mount) because it Measurement. Easy has up to several mm
contacts a device at a probing by spring-type difference in level.
45° angle. Can probe tip.
contact non-flat
material using a
spring-type probe tip.
Measurement Best Best Good
Stability

2
System Selection and H H µH
µH
10
m 1m 100 10
Measurement Procedures 100K

Select the system and make the 1


µH
10K
10
measurement according to the fF

following procedures: 10
0n
H
1K
100
f F
1. Choosing the HP 4291A
test head |Z|,R,X
100
10n
H

[Ω] 1pF

Good impedance measurement 1nH


accuracy can be obtained by 10 10p High Z Test Head
F

choosing an appropriate test head


p H Low Z Test Head
for the HP 4291A from 2 types of 1 100
100
test heads depending on the pF

impedance values being 0.1


1nF
measured. For example, chose a 1M 10M 100M 1G

FREQ [Hz]
suitable test head (see Figure 1) to
achieve an accuracy of 10% at
Figure 1. HP 4291A Accuracy (10%)
different frequencies. See the
HP 4291A data sheet for more
details.
2. Installation of the 3. Setting Up the HP 4291A The NOP, which is the number of
HP 4291A test station points for one sweep, can be set
Set the following measurement to a maximum of 801. When data
Fix the adapter kit to the left side conditions: of a higher density is desired (for
of the probe station for the example, when evaluating the
HP 4291A's test station Start/stop frequency (when using
steep frequency characteristic
installation. Then place the test frequency sweep)
precisely), set the number of
station vertically on the adapter points higher. If a short
kit. At that time, place the test Test signal level
measurement time is desired,
head and DUT as close as Number of points (NOP) decrease the number of points.
possible. Fix steadily to the test Usually, it is set to more than 201.
station. Number of points averaging
Point averaging reduces the
Parameter for compensation kit random error of noise by
averaging the data mathematically
The test signal level is usually set when calibrating, performing
as large as possible. For example, compensation, and measuring,
use 1 Vrms (for a measurement according to the specified
under 1 GHz) or 0.5 Vrms (for a averaging time.
measurement over 1 GHz) so that
the measurement can be made More than 32 averaging is usually
under conditions where the recommended in order to remove
signal-to-noise ratio is the largest. as much noise as possible.

3
Table 2. Compensation Kit Parameter However, an interpolation error
will occur. On the other hand, in
Input Value "USER" calibration, the
Value of Impedance Ideal Value measurement can be done with
HP 4291A Compensation
Standard Substrate (FPM) good accuracy and without any
(FPC, ACP, WRH) interpolation error. This is true
OPEN : CONDUCT(G) 0 0 because the internal data for
CAP.(C) Copen 0 calibration is calculated at the
SHORT : TRESIST.(R) 0 0
same points as the frequency
points. Perform the "USER"
INDUCT.(L) Lshort 0
calibration when accurately
LOAD : RESIST.(R) 50 50 measuring components or narrow
INDUCT.(L) Lterm 0 pitch transmission lines. (Notice
that it is necessary to perform
The compensation kit parameter There are two different types of "USER" calibration each time the
is the actual impedance parameter calibration procedure called frequency setting is changed.)
of the impedance standard "FIXED" calibration and "USER"
substrate for compensation. It's calibration. In "FIXED" 5. Connecting the HP 4291A and
set in the HP 4291A in order to calibration, the measurement is the prober
perform more accurate made at the specified frequency
compensation in step 6. Because points in the analyzer for the See figure 2. Connect the
these parameters (Copen, Lshort, calibration and the internal data is APC7-SMA adapter to the test
and Lterm) are dependent on the calculated for the whole span head. Then connect the cable
probe used and the distance frequency. Points other than the between the APC-SMA adapter
between the electrodes, they are calibration measurement points and the Cascade Microtech
supplied together with the probe. are calculated by interpolation. prober.
(However, as there are no The specification of the
parameters supplied for the FPM measurement accuracy will be
probe, the values under ideal satisfied even though the
condition are substituted.) Set the measurement frequency or the
parameters according to table 2. test signal level is changed.
Once the input values are
registered as USER COMPEN KIT,
the appropriate data will be
chosen the next time.
+3$
4. Calibration of the HP 4291A

The measurement accuracy of the


HP 4291A is applicable when
calibration is done on the APC-7 7HVW6WDWLRQ

terminal test head. Calibration


must be performed with the 7HVW+HDG

calibration kit (0 ohm, 0s, 50 ohms


and Low-loss Capacitor) attached 7HVW6WDWLRQ
 Ω 60$&DEOH
to the HP 4291A. 6WDQG

$3&60$$GDSWHU

3UREH+HDG
'87

6WDJH

Figure 2. Connection between the HP 4291A and the Cascade Microtech Probe

4
6. Open/Short/Load 7. Measurement of the DUT Summary
compensation
Place the DUT on the metallic Impedance measurements of
The residual impedance of the stage of the probe station. Move narrow pitch transmission lines
cable and probe from the end of the probe to the measurement and small capacitors on a wafer
the test head can be removed by position and make sure the probe can be performed up to 1.8 GHz
OPEN/SHORT/LOAD is brought into contact with the with a good accuracy by using a
compensation. Perform OPEN device. Choose the applicable combination of the HP 4291A
compensation when the probe is measurement parameter on the impedance/material analyzer and
not touching the DUT. Perform HP 4291A and make a the Cascade Microtech prober.
SHORT compensation by using measurement. One example of an
the short standard substrate Ls-Rs measurement of a
placed on the stage of the probe transmission line is shown in
station. Perform LOAD figure 3.
compensation in the same way by
using the 50 ohms on the standard
substrate. (Note: though there is a
function to set appropriate
electrical length for different
types of test fixtures in the
HP 4291A, it is not necessary to
set the electrical length as phase
shift is also compensated for after
load compensation. Set fixture to
[none] mode.)

Figure 3. Ls-Rs Measurement Example of Transmission Line

5
For more information on Hewlett-Packard
Test and Measurement products,
applications, or services please call your
local Hewlett-Packard sales office. A
current listing is available via the Web
through AccessHP at http://www.hp.com. If
you do not have access to the internet,
please contact one of the HP centers listed
below and they will direct you to your
nearest HP representative.

United States:
Hewlett-Packard Company
Test and Measurement Organization
5301 Stevens Creek Blvd.
Bldg. 51L-SC
Santa Clara, CA 95052-8059
1 800 452 4844

Canada:
Hewlett-Packard Canada Ltd.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(905) 206 4725

Europe:
Hewlett-Packard
European Marketing Centre
P.O. Box 999
1180 AZ Amstelveen
The Netherlands

Japan:
Hewlett-Packard Japan Ltd.
Measurement Assistance Center
9-1, Takakura-Cho, Hachioji-Shi,
Tokyo 192, Japan
Tel: (81 426) 56 7832
Fax: (81 426) 56 7840

Latin America:
Hewlett-Packard
Latin American Region Headquarters
5200 Blue Lagoon Drive
9th Floor
Miami, Florida 33126
U.S.A.
(305) 267 4245/4220

Australia/New Zealand:
Hewlett-Packard Australia Ltd.
31-41 Joseph Street
Blackburn, Victoria 3130
Australia
1 800 629 485

Asia Pacific:
Hewlett-Packard Asia Pacific Ltd
17-21/F Shell Tower, Times Square,
1 Matheson Street, Causeway Bay,
Hong Kong
Fax: (852) 2506 9285

Data subject to change


Copyright © 1996
Hewlett-Packard Company
Printed in U.S.A. 9/96
5965-5054E

Вам также может понравиться