Академический Документы
Профессиональный Документы
Культура Документы
(BIS 543)
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
CT: History
CT founded in 1970 by Sir Godfrey Hounsfield
Engineer with EMI, LTD. First applications were in neuroradiology
Spring 2012
Medical Equipment II CT
BIS 543
Tomography
Tomography is imaging by sections or sectioning, through the use of any kind of penetrating wave.
Basic principle of tomography: superposition free tomographic cross sections S1 and S2 compared The image is perpendicular to the long axis with the projected image P. of the body
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
beam differently
According to individual densities Radiation received by detector
Spring 2012
Medical Equipment II CT
BIS 543
BIS 543
A CT scan image consists of many cells, each assigned a number and displayed as a brightness level on the monitor. Each cell is called a pixel. The pixel is a 2-dimensional representation of a corresponding tissue volume. The tissue volume is called a voxel and is determined by the pixel size X the slice thickness (1-10mm).
Spring 2012
CT: Numbers
Medical Equipment II CT
BIS 543
CT: Numbers
The numerical information stored in each pixel is a CT number or Hounsfield Unit (HU). CT numbers range from -1000 to +1000 for each pixel: Water is defined as 0. -1000 corresponds to air while +1000 corresponds to dense bone. Soft tissues range from 0-100. Fat is -100.
Spring 2012
Medical Equipment II CT
Cont.
BIS 543
CT: Numbers
Enormous waste of information: Actual range of image is 2000 HU's or shades of gray. A video screen can only display 256 shades of gray. Therefore, windowing is needed.
Cont.
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
Cont.
Window width Determines range of CT numbers displayed on an image o Values above this range = white o Values below this range = black Window level Sets the center CT number displayed on the monitor Determines the location on the Hounsfield scale about which the window width will be centered
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
CT: Gantry
Houses the following: CT X-ray tube High voltage generator Detector array Data acquisition system Slip ring Can be angled from 15o-30o -this is useful in imaging the skull Table automatically positions the patient in exact increments.
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
Cont.
BIS 543
Cont.
BIS 543
Cont.
A solid-state CT detector is composed of a scintillator coupled tightly to a photodetector. The scintillator emits visible light when it is struck by x-rays, just as in an x-ray intensifying screen The light emitted by the scintillator reaches the photodetector, typically a photodiode, which is an electronic device that converts light intensity into an electrical signal proportional to the light intensity. solid-state detectors typically have better x-ray absorption efficiency
Spring 2012
Medical Equipment II CT
BIS 543
Cont.
Multiple detector arrays are a set of several linear detector arrays, tightly abutted. The multiple detector array is an assembly of multiple solid-state detector array modules. With multiple detector arrays, slice width is determined by the detectors, not by the collimator (although a collimator does limit the beam to the total slice thickness). To allow the slice width to be adjustable, the detector width must be adjustable. It is not feasible, Therefore, with multislice systems, the slice width is determined by grouping one or more detector units together. To combine the signal from several detectors, the detectors are essentially wired together using computer-controlled switches.
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
BIS 543
Assume that a very simple 2 x 2 image is known only by the projection values Using algebra (N equations in M unknowns), one can solve for the image values in the simple case of a 4-pixel image Simple backprojection starts with an empty image matrix (an image with all pixels set to zero), and the m value from each ray in all views is smeared or backprojected onto the image matrix.
Spring 2012
Medical Equipment II CT
BIS 543
Conventional CT
Axial
o Start/stop
Volumetric CT
Helical or spiral CT
o Continuous acquisition
Spring 2012
Medical Equipment II CT
BIS 543
Spring 2012
Medical Equipment II CT
BIS 543
Produces one crosssectional image Once this is complete patient is moved to next position
Process starts again at the beginning
Spring 2012
Medical Equipment II CT
BIS 543
X-ray tube and detector rotate 360 Patient table moves continuously
With X-rays on
BIS 543
BIS 543
CT: Generations
First Generation (Original EMI Scanner)
Translate-rotate mechanism Linear translation through the patient called a scan. Each scan is followed by a 1O rotation before beginning another scan. This continued until the tube had moved a total of 180O Pencil-thin X-Ray beam. A single paired detector. Scan time is up to 5 minutes per slice [Primary disadvantage of lst generation scanners].
Spring 2012
Medical Equipment II CT
BIS 543
Second Generation
CT: Generations
Major objective was to increase speed. Multiple detectors Incorporated the natural extension of the single detector to multiple detectors. 5-30 detectors now used Fan-shaped beam to be used ,with the multiple detectors. Also a translate-rotate mechanism, but Fewer linear scans are required because 30 detectors gather more data per scan. This allows greater rotary steps. Up to 30o rotation between scans instead of 1o Fewer linear scans result in faster scan times. [Scan times down to 20 seconds].
Spring 2012
Medical Equipment II CT
BIS 543
CT: Generations
Third Generation
Rotate-rotate mechanism. X-Ray tube and detector are rotated concentrically about the patient.
No linear translations
Larger fan-shaped X-Ray beam, 30o-60o. Larger number of detectors used to read the larger fan-shaped beam
BIS 543
Fourth Generation
CT: Generations
Rotate-fixed mechanism.
Only the tube rotates. The detectors are fixed. Beam is fan-shaped as the 3rd generation.. Radiation detection is through a fixed circular array of detectors
[Up to 2400 elements in the detector array].
BIS 543
CT: Generations
Spring 2012
Medical Equipment II CT
BIS 543
Helical Scanners
CT: Generations
Newest CT technology, available since 1990. Slip-ring technology allows continuous rotation of the tube Allows simultaneous translation and data acquisition. Continuous data acquisition can be achieved in a single breath hold. (360o rotation in 1 second or 1 rev/sec) Results in a dramatic increase in speed and consequently throughput. (20-40 seconds for HCT scan vs. 4-6 minutes with conventional CT) Increased 3D longitudinal resolution, optimizing image quality. Position and spacing of axial slices can be chosen after data acquisition. Radiation dose decreased slightly because of decreased repeats due to motion artifacts Image reconstruction takes time and can't be viewed immediately. (Patient can be removed after collection of the raw data)
Spring 2012
Medical Equipment II CT
BIS 543
Helical Scanners
CT: Generations
Cont.
During helical acquisition, the data are acquired in a helical path around the patient Before reconstruction, the helical data are interpolated to the reconstruction plane of interest. Interpolation is essentially a weighted average of the data from either side of the reconstruction plane, with slightly different weighting factors used for each projection angle. Helical scanning allows the production of additional overlapping images with no additional dose to the patient.
Spring 2012
Medical Equipment II CT
BIS 543
BIS 543
Cont.
Spatial resolution Slice thickness Dose Table incrementing light localizer, collimator
Medical Equipment II CT
Spring 2012
BIS 543
Cont.
The five-pins are each made of a different material with known physical and X-ray absorption properties, as shown in the following table :
Material Polyethylene Polystyrene Nylon Plexiglas Water
Spring 2012
Spring 2012
Medical Equipment II CT
BIS 543
Cont.
Neuro-Angiography
Spring 2012
Medical Equipment II CT
Cont.
MasterCut
Panoramic View
MIP 4D-Angio
Renal Arteries
Spring 2012
Medical Equipment II CT
Questions ?
Spring 2012
Medical Equipment II CT
BIS 543