AUDIT CRITERIA
161 Thorn Hill Road Warrendale, PA 15086-7527

Issued 2006-11 Revised 2012-4 Superseding AC7114S Rev. F

TO BE USED ON OR AFTER July 1, 2012 Nadcap SUPPLEMENTAL AUDIT CRITERIA FOR NONDESTRUCTIVE TESTING (NDT) SUPPLIERS ACCREDITATION PROGRAM

Editorial change made to 3.2.1 & 4.1.1 on 30-Apr-12

SCOPE
These audit criteria are to be used as a Nadcap User supplement to PRI AC7114 for suppliers with Nondestructive Testing Facilities that are seeking Nadcap NDT accreditation. Only the questions that pertain to User supplemental criteria will be found within this checklist. Only those suppliers, who are approved by or have specific written authorization from, a particular user, shall be permitted to be audited to that Users supplemental criteria.

GENERAL INFORMATION
In completing this assessment, auditors are instructed to respond with a "Yes or No" to address compliance with each statement of requirement. (For any negative responses, the auditor must clearly indicate in the NCR if the No" reflects noncompliance with respect to existence, adequacy, and/or compliance. Existence relates to evidence of a documented procedure or policy, adequacy relates to the completeness of the procedure or policy, and compliance relates to evidence of effective implementation). The auditor shall only apply the questions in this checklist to the suppliers who specify the end Users identified herein. U1 - Honeywell U2 - P&W U3 - Rolls Royce Plc U4 - Lockheed Martin U5 - Eurocopter U6 - Rolls Royce Corporation U7 - MTU U8 - Airbus U9 - Hawker Beechcraft U10 - GE Aviation U11- The Boeing Company U14 - SAFRAN U16 - Bell Helicopter U17 - Sikorsky Aircraft U18 - Hamilton Sundstrand U19 - Triumph/Vought

Each question in this checklist must be addressed only in so far as it pertains to each applicable User Member who is represented at that question, and only needs to be answered once. All negative responses require a Nonconformance Report (NCR) with the auditor indicating the specific User requirement that was violated. Not Applicable (N/A) responses do not require an explanation as they indicate that no User Member requirement is addressed at that question or that the process identified is not utilized.

PRI operating procedures provide that "This report is published by PRI to advance the state of technical, engineering, and quality sciences. The use of this report is entirely voluntary, and its applicability and suitability for any particular use, including any patent infringement arising there from, is the sole responsibility of the user." PRI values your input. To provide feedback on this document, please contact the appropriate commodity staff engineer. (Contact information is located at http://www.eauditnet.com under Contact Us.) Copyright 2012 Performance Review Institute. All rights reserved. t-frm-15 3-Jan-12

PRI

AC7114S Rev. G

-2-

3
3.2

INFORMATION
Quality Systems Approvals 3.2.1
Is there evidence of a current approval related to quality system registered to AS9100 / EN9100, AC7004, or ISO17025 (NDT Test Houses only)?

YES NO N/A

U1 U5

Is there evidence the quality system meets AS/EN/JISQ 9100, or for independent NDT labs ISO 17025, or AC7004? Is there evidence the quality systems meets the correct revision of AS/EN9100 or ISO17025?

4
4.1

NDT QUALITY SYSTEM REQUIREMENTS


Customer Requirements 4.1.1 Has a Level 3 been given accountability for identifying and assuring implementation of customer NDT requirements for the following? YES NO N/A

U1

Does SCA Level 3 identify and assure implementation of Honeywell NDT requirements? (See SPOC 165.7 section 4.1)

4.3

Internal NDT Audits 4.3.1 Procedure Is there a procedure requiring an annual internal audit of the NDT System and does it address a review of the NDT inspection process, the associated documentation, personnel certifications, NDT process procedure adequacy and equipment? YES NO N/A

U1 U18

Is there evidence the quality system meets AS/EN/JISQ 9100, or for independent NDT labs ISO 17025, or AC7004? (See SPOC Manual, Section 1.12) If the supplier uses a Corporate (not located on site) or Contract Level 3, does the procedure address internal audits to be performed at least every other month? (See HSM17, section 7.4). 4.3.2 Internal Audit Scope Does the annual internal NDT audit include a review of the NDT inspection process, the associated documentation (planning, travelers and inspection certifications), personnel certifications (for existence and completeness), NDT process procedure adequacy and equipment? YES NO N/A

U1

Are internal NDT audits performed by the Honeywell approved NDT SCA? (See SPOC 165.7, Section 4.1) If the supplier uses a Corporate (not located on site), or Contract Level 3 as their Honeywell approved Level 3/SCA, is an internal NDT audit conducted semi-annually at each assigned site that they support using either a Honeywell approved or Nadcap audit checklist. Audit results (i.e., Audit Report) shall be sent to the Honeywell NDT CAA. (See SPOC 165.7, Section 5) 4.3.3 Internal Auditor Qualification Are the personnel performing internal NDT audits knowledgeable in the YES NO N/A

PRI

AC7114S Rev. G
method(s) they are auditing and are they approved by the Responsible Level 3?

-3-

U1

Are internal audits performed by the Honeywell approved NDT SCA? Is there evidence that internal NDT auditors have completed, as a minimum, the required formal NDT classroom training equivalent to a Level 2 in the applicable method, objective evidence of 8 hours of auditor training (AS9100, ISO, etc.) and two training audits in each method they are auditing? (See SPOC 165.7, Section 2.1)

4.4

Performance Review 4.4.2 Performance Review Scope Is there a documented annual review of processing and inspection of hardware for each certified individual who is qualified to process hardware? YES NO N/A

U1

U2

If the supplier uses a Corporate (not located on site) or Contract Level 3 as their Honeywell approved Level 3/SCA, is the internal audit conducted semi-annually using either a Honeywell or Nadcap checklist? (see SPOC 165.7 section 5) Is there evidence the customer-certified Level 3 has performed the annual over inspection / audit of all certified Level 1 and Level 2 inspectors? (See NDTQ para 6.1.6.8 and 6.1.6.12) 4.4.3 Review Performance Is this review performed by the Level 3 or designee (minimum of Level 2 qualified) in the method being reviewed and records retained? YES NO N/A

U1

Are NDT performance reviews performed by the Honeywell approved NDT SCA? (See SPOC 165.7, Section 4.1) If the supplier uses a Corporate (not located on site) or Contract Level 3 as their Honeywell approved Level 3/SCA, is the NDT performance review conducted semi-annually using either a Honeywell approved or Nadcap audit checklist? (See SPOC 165.7, Section 5)

5
5.1

PERSONNEL RECORDS NAS-410 / EN4179


Written Practice 5.1.1 Written Practice Approval Is the training, qualification and certification of NDT personnel written practice approved by the Responsible Level 3? YES NO N/A

U1 U3 U6

Is the procedure approved by a Honeywell approved Level 3 (SCA)? (See SPOC 165.7, Section 4.1) Is the procedure approved by a Rolls Royce approved Level 3? (See RPS908(2.3)) Is the procedure approved by a Rolls Royce approved level 3?

5.1.5

Personnel Duties and Responsibilities

PRI

AC7114S Rev. G
Does the written practice address the required personnel duties and responsibilities?

-4YES NO N/A

U1 U3 U6 U7 U8

U10

Are there provisions to prevent any NDT Level 1 from interpreting / evaluating results? ( SPOC 165.7 section 6.5) Are there provisions to prevent any NDT Level 1 from interpreting / evaluating results? (See RPS 908 para 2.6) Are there provisions to prevent any NDT Level 1 from interpreting / evaluating results for PT and MT? Are there provisions to prevent any NDT Level 1 from interpreting / evaluating results? Are there provisions to prevent any NDT Level 1 from interpreting / evaluating results? Except where allowed by A1083 Annex D para D4.2 (ASI 6-10 para 4.1 & A1083 Annex D Clause D4.2) Are there provisions to prevent any NDT Level 1 from interpreting / evaluating results for PT and MT? 5.1.6 Training Program Does the written practice address the required training requirements? YES NO N/A

U1

Does the Procedure outline the required formal training hours as called out in SPOC Manual, Section 165.7 without allowances for waving by the responsible level 3? (See SPOC 165.7, Section 6.4) 5.1.9 Examination Requirements Does the written practice address the examination requirements? YES NO N/A

U1

Are the examination practices as outlined in SPOC Manual, Section 165.7? (See SPOC 165.7, Section 6) 5.1.13 Re-examinations Does the written practice address the requirements for re-examination of those personnel who have failed their certification/re-certification examinations or have had their certifications revoked? YES NO N/A

U1 5.2

Are the re-certification requirements as outlined in SPOC Manual, Section 165.7? (See SPOC 165.7, Section 6) Personnel Records 5.2.12 Equivalent Training Is documentation on file regarding equivalent training accepted by the Responsible Level 3 to meet qualification requirements? YES NO N/A

U1

Is there evidence that the required minimum formal training hours as called out in SPOC 165.7, section 6.4 have been obtained and equivalent training not accepted by the responsible Level 3? (See SPOC 165.7, Section 6.4) Examinations 5.3.1 Administration and Grading

5.3

PRI

AC7114S Rev. G

-5-

Is there evidence that the qualification examinations have been administered and graded by the Responsible Level 3 or Examiner? U1 If the supplier uses a National Aerospace Non-Destructive Testing Board (NANDTB), is only the general examination accepted for qualification and certification/recertification of NDT personnel? (See SPOC 165.7, Section 6.8) Have all examinations been administered by a customer approved Level 3, and not an NANDTB? (See NDTQ para 6.1.6.6) Have all examinations been administered by a customer approved Level 3, and not an NANDTB? (See SS8813, Rev 3 Paragraph 3.4.2) 5.3.5 Exam Schedule for Level 1 and Level 2 Are the Level 1 and Level 2 personnel re-certified at a frequency not to exceed five years? U1 Are the Level 1 and Level 2 personnel re-certified at a frequency not to exceed three years? (See SPOC 165.7, Section 6.1) 5.3.7 Level 3 Re-certification by Points When points are utilized for Level 3 re-certification, is the required documentation on file? U1 Does the procedure for qualification and certification require the Level 3 to re-certify by examination? (See SPOC 165.7, Section 6.3) Does the procedure for qualification and certification require the Level 3 to re-certify by examination? (See HSM17, section 7.4) 5.3.8 Level 3 Re-certification by Examination When examinations are utilized for Level 3 re-certification, is the required documentation on file? U1 U18 Are general, specific and practical examinations utilized for Level 3 re-certification, and is the required documentation on file? (See SPOC 165.7, Section 6.3) Are general, specific and practical examinations utilized for Level 3 re-certification, and is the required documentation on file? (See HSM17, section 7.4). 5.3.10 Examination Scoring System Does the examination scoring system for qualification examinations conform to the requirements of the governing standard? U2 Is there evidence that the candidate for certification has achieved a minimum grade of 80% on the general and specific qualification examinations; and the candidate must achieve a minimum grade of 90% for the processing portion of the practical as demonstrated to the satisfaction of the Certifier? (See NDTQ para 7.2.3) 5.3.12 Specific Examination Questions Do the examinations have, at least, the minimum number of questions and do the specific examination questions reflect the specifications, codes, equipment,

YES NO N/A

U2 U17

YES NO N/A

YES NO N/A

U18

YES NO N/A

YES NO N/A

YES NO N/A

PRI

AC7114S Rev. G
operating procedures and test techniques used within the facility?

-6-

U2

For Level 2, is there evidence the specific examinations contain a minimum of 10 questions relative to Pratt & Whitney acceptance standards? (See NDTQ para 6.4.3.2) 5.3.14 Level 1 Practical Examinations Do the Level 1 practical examinations provide evidence that at least one test sample per technique, with a minimum of two test configurations for each method, were processed? Practical examination for all NDT methods and levels of certification shall be comprised of the equipment, procedures, standards and hardware of the facility the candidate is seeking certification (and/or recertification for). (See SPOC 165.7, Section 6.6) 5.3.15 Level 2 Practical Examinations Do the Level 2 practical examinations provide evidence that at least one test sample per technique, with a minimum of two test configurations for each method, were processed? Practical examination for all NDT methods and levels of certification shall be comprised of the equipment, procedures, standards and hardware of the facility the candidate is seeking certification (and/or recertification for). (See SPOC 165.7, Section 6.6) 5.3.17 Level 2 Practical Examinations If the Level 3s duties include processing and/or accepting hardware, does the practical examination provide evidence that at least one test sample per technique (as defined by the written practice) with a minimum of two test samples of differing configurations for each method were evaluated? Practical examination for all NDT methods and levels of certification shall be comprised of the equipment, procedures, standards and hardware of the facility the candidate is seeking certification (and/or recertification for). (See SPOC 165.7, Section 6.6) Review of Suppliers Level 3 Personnel 5.5.1 Internal Level 3 Assignment Does the supplier have an internal Level 3 in all methods being audited? YES NO N/A YES NO N/A

U1

YES NO N/A

U1

YES NO N/A

U1

5.5

U1

U2

U3

U6

Is there evidence that the customer-approved Level 3 has obtained evidence that they have been approved as a Supplier Certifying Agent (SCA) for each applicable NDT method for their facility? If applicable, was there evidence that the supplier notified the user within 5 working days of an NDT Level 3 (SCA) leaving the supplier's employment? If applicable, in the event of an NDT Level 3 (SCA) leaving the supplier's employment, was there evidence that the supplier qualified a new NDT Level 3 (SCA) within ninety calendar days? (See SPOC 165.7, Section 1.3) Is there evidence that the Level 3 is a full time employee of the company, and certified by the user (contractors maybe certified on a case by case basis. Evidence should be a letter) for each method? (See NDTQ, para 7.1.2) Is there evidence of a user approval letter on file for the Rolls Royce designated Level 3 which is specific to the supplier and the designated Level 3? If an independent NDT test house, is the Level 3 a full-time employee? (See RPS 906 Section B) Is there evidence of a user approval letter on file for the Rolls Royce designated Level

PRI

AC7114S Rev. G
3 which is specific to the supplier and the designated Level 3? If an independent NDT test house, is the Level 3 a full-time employee? Is there evidence the responsible Level 3 has been agreed by the customer? (See ASI 6-10 para 1.3 & A1083 Annex D Clause D4.5.2) Is there evidence that the supplier is approved and listed in D1-4426 for providing Level 3 services if provided internally? (See D1-4426 User Instructions / Processor Requirements - Section 5.12) Note The following D1-4426 approval codes apply to NAS 410 in regards to the applicable NDT methods: 849A PT 848A MT 851A UT 850A RT Is there evidence that the Level 3 is a full time employee of the company, and certified by the user (contractors maybe certified on a case by case basis. Evidence should be a letter) for each method? (See SS8813, Rev 3 Paragraph 3.4.2) 5.5.3 Qualification / Certification Record for Outside Agents Does the facility have documentation on file to verify the adequacy of the qualification / certification for the outside Level 3?

-7-

U8 U11

U17

YES NO N/A

U1

U8 U11

If an outside Level 3 is used as the suppliers NDT SCA (Source Certifying Agent); is there evidence that the Level 3 has been approved by Honeywell? (See SPOC 165.7, Section 5) Is there evidence the responsible Level 3 has been agreed by the customer? (See ASI 6-10 para 1.3 & A1083 Annex D Clause D4.5.2) If used, is the outside agency approved and listed in D1-4426 to the appropriate codes? (See D1-4426 Section 5.12)