Академический Документы
Профессиональный Документы
Культура Документы
Bertil Magnusson
Content
an overview of direct XRF measurement will be presented using standard wavelength dispersive and energy dispersive instruments as well as handheld instrument
What we all know about XRF Direct measurement versus normal analysis in laboratory Elemental range Systematic effects Application of direct measurements 1. Quality control in nuclear power plant 2. Analysis of sediment and wood cores 3. Determination of thickness of coating
Few restrictions
< 5 cm No
Today light elements possible in air with short pathlengt example handheld XRF
Zn Ka Measuring depth
S Ka
Si Ka Na Ka
Direct measurement without sample preparation: Systematic effects for inhomogeneous samples!
Direct measurement with XRF 7
Measurement cycle
Client
Decision on result Client Issue
Data presentation
Define issue
Report on measurement
Decision on measurement
Client
Interface
Measuring scientist
Evaluation Analysis
Sampling
Measuring scientist
Direct measurement with XRF 8
11
Client
Decision on result Client Issue
Data presentation
Define issue
Report on measurement
Decision on measurement
Client
Interface
Measuring scientist
Evaluation Analysis
Sampling
Measuring scientist
Direct measurement with XRF 12
Crud sampling
Direct measurement with XRF XRF analysis on active samples Validation with synthetic samples and active test samples
13
Crud* sampling
The crud is scraped or brushed from a defined area of the fuel rod Zircalloy The crud is collected on a filter membrane The crud/filter sample is analysed to determine: Quantity (g/m2) - XRF Elemental composition (g/m2): XRF
crud
ZrO2
This application is presented courtesy to Eva Fredriksson & Pentti Hietala, Westinghouse
Crud* sampling
HOW
C Fuel crud is X sampled Side Orientation between Orientation X theSubbundle spacers 2 Spacer SP normally 5 or 6 Marking levels B 3 D
Sampling level
Top plate
SP 6
7
4
6
1
SP 5
Water reference samples are takenD 10 1 Channel Center both before and 9 Marking 8 7 6 after 4each 2 3 5 4 3 assembly 2 3
1 C
5
A 4
SP 4
SP 3
SP 2
3 1
A B C D E F GH I J 2
Spacer 3
Spacer 2 Spacer 1
SP 1
2
B
1
Direct measurement with XRF 15
Crud sampling
HOW
XRFcontainer
20
Sample excitation from active samples ED-XRF with X-ray tube on&off?
Sample is a filter. Sampling 10 l water from primary cooling circuit in a nuclear power plant. Activity 25 S/cm close
Tube on
Tube off
Spectrum presented courtesy to Kari Moum and Torill Solheim Institutt for Energiteknikk/Halden Reaktor Prosjektet
Direct measurement with XRF 21
Tube on
Tube off
22
Tube on
Tube off
23
Cr Ka
Mn Ka Green is ca 30 ug/cm2
Tube on
Tube off
24
25
26
New scan of water blank with samples and without samples in sample changer
Validation
Instrument calibrated with standardless software 1. Validation with synthetic crud samples prepared from pure metals Low recovery - < 80 % 2. Validation with synthetic crud samples grinded further recovery close to 100 % 3. Validation with active samples analysed with ICP-MS recovery close to 100 % for most elements
28
Validation with synthetic crud samples Fe, Co, Cr, Mn, Ni, Cu, Zn, Pt
Figure 1 Particle size distribution for crud 1. Surface weighted mean 3 m Figure 2 Particle size distribution for crud 1 after sample was further grinded. Surface weighted mean 1.2 m
1 m
Direct measurement with XRF 29
crud samples
2010-08-17
30
Residuals
crud samples
2010-08-17
31
150 Resiidual ug
100
50 0
0 20 40 60 80 100 120
-50
Similar results were obtained 2010 Decision to use active crud samples for calibration
crud samples 2010-08-17
32
Conclusions
Systematic effects yes. Validation with synthetic standards - The sample particles were finer than the synthetic crud standards (before further grinding) Validation with active test samples For most elements the recovery was close to 100 % comparing XRF Uniquant with ICP-MS results The Zn gave 150 % recovery when comparing with ICP-MS. No understanding of this effect so far. The ICP-MS were validated with synesthetic crud and gave excellent results
33
Conclusions
New calibration performed with active crud samples, analysed with ICP-MS, to minimise systematic effects. For elements not found in crud synthetic samples were used. Interference from active samples Increased background with radioactive samples. No characteristic peaks found with WD-XRF Fitness for intended purpose Direct analysis now possible for elemental composition and amount using XRF for elements of interest
Direct measurement with XRF 34
Client
Decision on result Client Issue
Data presentation
Define issue
Report on measurement
Decision on measurement
Client
Interface
Measuring scientist
Evaluation Analysis
Sampling
Measuring scientist
Direct measurement with XRF 35
Applications
Si (blue)
Fe (red)
This application is presented courtesy to Anders Rindby
50 mm
36
Sample is irradiated with an X-ray tube with a well-defined beam. The fluorescence (XRF) and absorption (radiography) are measured. An camera system provides sample images.
37
profiler
Sample
38
820 mm
K profile
Ti profile
Fe profile
As profile
Br profile
Direct measurement with XRF Sample with courtesy of Dr. Bernard Dennielou, Ifremer, France 39
Illustrations/graphs area
Client
Decision on result Client Issue
Data presentation
Define issue
Report on measurement
Decision on measurement
Client
Interface
Measuring scientist
Evaluation Analysis
Sampling
Measuring scientist
Direct measurement with XRF 41
42
Illustrations/graphs area
Illustrations/graphs area
Calibration curve Mo Lb
Intensity versus Mo (mg/dm2)
Illustrations/graphs area
Calibration curve Ca
Intensity versus Ca (mg/dm2)
Illustrations/graphs area
ED-XRF
Illustrations/graphs area
Illustrations/graphs area
Using a general metal calibration The apparent sulfur concentration is used for screening if the product fulfils the specification
Illustrations/graphs area
Illustrations/graphs area
51
53