Definition of emissivity
this detected radiation, the pyrometer calculates the temperature according to Plancks
radiation law. The amount of energy emitted
= r / b
transmitted
background
radiation
target
object
Pyrometer
* obj
* back
* amb
reected
ambient
radiation
TECHNICAL REPORTS
an object. The total amount of radiation detected by a pyrometers sensor is the sum of several components, as shown in
the following equation.
= * obj + * amb + * back
= Emissivity factor
= Reflectivity factor
= Transmissivity factor
Material
Wavelength
glass
4,8 m
3,43 m
7,9 m
4,27 m
4,5 m
pyrometer will measure at a depth below the glass surface, detecting thermal energy from within the object.
1=++
There will be no transmission of radiation through opaque
objects; thus the transmissivity factor will not be applicable.
1=+
ment.
sight path, the tube diameter should measure at least six times
have distinct wavelength ranges in which the radiation characteristics are good. In order to accurately measure the tempera-
wavelength.
temperature and which wavelength the emissivity value is applicable. Furthermore, the stated emissivity values were obtained
considerably.
TECHNICAL REPORTS
Vergleichsstrahler
Messobjekt
just the pyrometer accordingly, provided that the contact thermometer measurement is very precise.
tures
sequently,
comparison
the
pyrometers
temperature
Because
the
reading.
influence
of
16
4.5 - 4.9 m
8 -14 m
14
12
1.9 - 2.5 m
10
8
1.1 - 1.7 m
0.78 -1.06 m
0.63 - 0.67 m
2
0
0
500
1000
1500
2000
2500
3000
Temperature [C]
Fig. 4 At 1% change in emissivity, depending on temperature and wavelength.
TECHNICAL REPORTS
of 0.67 m.
it is recommendable to consider both the two spectral temperature values as well as the temperature reading based on the
the method which is best suited for his application and adjust
Two-colour pyrometers
measurements irrespective
of emissivity?
Conclusion
When pointing out the attributes of a pyrometer, brochures
often draw special attention to an instruments specified meas-
and thus the temperature, will still remain constant. This is only
Author
Dipl.-Ing. Albert Book
albert.book@keller-msr.de
www.keller-msr.de/pyrometer
TR_008_201604_en