CALIBRATION CERTIFICATE

MSMED Regd. No. 24-019-22-01281 &VUDA REGD. NO. UDA/PLAN-05/233/2005

Calibration certificate No. : Ges/VAD-130402-06/LG-01/13-14

Client

O.P. Goyal

Site Name

Equipment under calibration

Load Gauge

Make

Masta

Serial No

100/579

Range

0-1000 KN

Least Count

5 KN

Ram Dia.

165 mm

Condition on Receipt

Good

Issue Date: 04/04/2013

Summary of Measurements: The Load Gauge was calibrated against Proving Ring upto full Scale by direct
comparison method. The observed results along with the estimated accuracy in the measurements are given on
page 2.

Date of Calibration

: 04/04/2013

Recommended Date of Next calibration : 03/04/2014

CHECKED BY

APPROVED BY

Note (1)
: This Certificate / Report shall not be reproduced except in full.
(2) The measurement results quoted in this Certificate / report refers only to the items submitted and
apply at the time of measurement.
(3) This Calibration / Test has been carried out using measurement standards and istruments whose
calibration is traceable to National / international standards

Page 1 / 2

CALIBRATION CERTIFICATE
Measuring Equipment used :
The FMD Proving Ring No. 08242 of full scale Range 1000 KN Externally calibrated by NATIONAL
COUNCIL FOR CEMENT AND BUILDING MATERIALS, BALLABGARH. vide certificate no.
11090407/D/5/C-178, Dated- 14/11/2013 and Valid up to 26 Months from the date of calibration.

Measurements Results: The calculated reading generated by FMD Proving Ring and the indicated Load by the
Gauge under calibration is given below.

Calibration certificate No. : Ges/VAD-130402-06/LG-01/13-14

Load Gauge Calibration


Load Gauge Reading, kN

Proving Ring
Reading

Relative
Repeatability
Error %

0'0

180'0

360'0

AVG.

0.0

0.0

0.0

0.0

0.0

--

100.0

104.2

104.0

104.6

104.3

0.62

200.0

199.2

198.7

199.4

199.1

0.32

300.0

299.6

298.0

297.3

298.3

0.77

400.0

398.7

398.4

397.8

398.3

0.23

500.0

498.2

497.8

498.5

498.2

0.14

600.0

598.7

598.1

597.5

598.1

0.20

700.0

697.5

697.2

697.8

697.5

0.09

800.0

797.3

798.1

797.5

797.6

0.10

900.0

897.1

897.5

898.1

897.6

0.11

CHECKED BY

APPROVED BY

Page 2 / 2

Issue Date: 04/04/2013

Page 1 / 2

Load Gauge Calibration

Page 2 / 2

Proving
Ring
Reading

Load
Gauge
Reading,
kN

Error %

0.0

0.0

100.0

100

48.9

213.91

104.60199

4.60199

200

93.2

213.91

199.36412

-0.63588

--

300

136.8

213.91

292.62888

-7.37112

104.3

4.25

400

184.2

213.91

394.02222

-5.97778

200.0

199.1

-0.46

500

229.2

213.91

490.28172

-9.71828

300.0

292.6

-2.48

600

274.4

213.91

586.96904 -13.03096

400.0

393.5

-1.62

700

320.2

213.91

684.93982 -15.06018

500.0

489.4

-2.13

800

368.6

213.91

788.47226 -11.52774

600.0

586.0

-2.33

900

413.2

213.91

883.87612 -16.12388

700.0

685.8

-2.03

800.0

788.0

-1.49

900.0

884.3

-1.74

CALIBRATION CERTIFICATE
MSMED Regd. No. 24-019-22-01281 &VUDA REGD. NO. UDA/PLAN-05/233/2005

Calibration certificate No. : Ges/VAD-130418-05/LG-01/13-14

Client

O.P. Goyal

Site Name

Equipment under calibration

Load Gauge

Make

Masta

Serial No

100/579

Range

0-1000 KN

Least Count

5 KN

Ram Dia.

165 mm

Condition on Receipt

Good

Issue Date: 18/04/2013

Summary of Measurements: The Load Gauge was calibrated against Proving Ring upto full Scale by direct
comparison method. The observed results along with the estimated accuracy in the measurements are given on
page 2.

Date of Calibration

: 18/04/2013

Recommended Date of Next calibration : 17/04/2014

CHECKED BY

APPROVED BY

Note (1)
: This Certificate / Report shall not be reproduced except in full.
(2) The measurement results quoted in this Certificate / report refers only to the items submitted and
apply at the time of measurement.
(3) This Calibration / Test has been carried out using measurement standards and istruments whose
calibration is traceable to National / international standards

Page 1 / 2

CALIBRATION CERTIFICATE
Measuring Equipment used :
The FMD Proving Ring No. 08242 of full scale Range 1000 KN Externally calibrated by NATIONAL
COUNCIL FOR CEMENT AND BUILDING MATERIALS, BALLABGARH. vide certificate no.
11090407/D/5/C-178, Dated- 14/11/2013 and Valid up to 26 Months from the date of calibration.

Measurements Results: The calculated reading generated by FMD Proving Ring and the indicated Load by the
Gauge under calibration is given below.

Calibration certificate No. : Ges/VAD-130418-05/LG-01/13-14

Load Gauge Calibration


Load Gauge Reading, kN

Proving Ring
Reading

Relative
Repeatability
Error %

0'0

180'0

360'0

AVG.

0.0

0.0

0.0

0.0

0.0

--

100.0

103.2

103.5

103.8

103.5

0.58

200.0

199.8

198.5

199.4

199.2

0.65

300.0

299.6

298.6

297.5

298.6

0.70

400.0

398.9

398.6

397.8

398.4

0.28

500.0

499.1

497.9

498.8

498.6

0.24

600.0

598.8

598.7

598.2

598.6

0.10

700.0

698.1

697.2

698.8

698.0

0.23

800.0

799.3

798.9

797.9

798.7

0.18

900.0

897.7

897.1

897.5

897.4

0.07

CHECKED BY

APPROVED BY

Page 2 / 2

Issue Date: 18/04/2013

Page 1 / 2

100

48.9

213.91

104.60199

4.60199

200

93.2

213.91

199.36412

-0.63588

300

136.8

213.91

292.62888

-7.37112

400

184.2

213.91

394.02222

-5.97778

500

229.2

213.91

490.28172

-9.71828

600

274.4

213.91

586.96904 -13.03096

700

320.2

213.91

684.93982 -15.06018

800

368.6

213.91

788.47226 -11.52774

900

413.2

213.91

883.87612 -16.12388

avg

Page 2 / 2

50.1

49.4

49.6

49.7

213.9

107.2

105.7

106.1

93.4

92.8

93.2

93.1

213.9

199.8

198.5

199.4

136.6

136.8

136.8

136.7

213.9

292.2

292.6

292.6

183.9

183.8

183.6

183.8

213.9

393.4

393.2

392.7

228.60

228.60

228.90

228.7

213.9

489.0

489.0

489.6

273.9

273.8

373.6

307.1

213.9

585.9

585.7

799.2

320.6

320.8

320.8

320.7

213.9

685.8

686.2

686.2

368.2

368.4

368.4

368.3

213.9

787.6

788.0

788.0

412.8

413.8

413.6

413.4

213.9

883.0

885.2

884.7