Relative and differential pressure transmitter type 694

Pressure range -0.5 ... 0.5 mbar / 0 ... 1 50 mbar

Type 694 pressure transmitters are available with or without integral digital display. 010 V / 420 mA outputs are available as well as square. Individually ranged sensors ensure optimum accuracy and long term stability of measurement, and the variety of options means great flexibility, not only in climate control, but also in industrial and medical fine pressure measurement.

Compact construction Fast, easy mounting. Housing incorporates integral bracket for wall or ceiling mounting. Snap-on cover with a single screw Available with or without LCD display Available with or without square root extraction Available zero point and full scale adjustable Attractive price / performance ratio

Huba Control type 694 - Technical data subject to change - Edition 09/2011

1/5

Technical overview
Pressure range Relative and differential Operating conditions Medium Temperature Tolerable overload on one side Rupture pressure Application at over pressure range Application at low pressure range ambient temperature 70 C Medium / ambient Storage No condensation -0.5 ... 0.5 mbar / 0 ... 1 50 mbar

Air and neutral gases 0 ... +70 C -10 ... +70 C P1 = 100 mbar P1 = -4 mbar 2 x overload 1.5 x overload P2 = 4 mbar P2 = -100 mbar

Materials in contact with medium Sensor Diaphragm Housing Electrical overview Output 2 wire 4 ... 20 mA 0 ... 10 V 3 wire 0 ... 20 mA 4 ... 20 mA Polarity reversal protection Dynamic response Response time Load cycle Protection standard Without cover With cover Display LCD display Setting range Zero point Full scale Electrical connection Screw terminals for wire and stranded conductors up to 1.5 mm2 Cable gland with built-in strain relief PG11 Pressure connection Connection pipe Mounting instructions Factory adjustment: Installation arrangement Mounting Tests / Admissions ETL Electromagnetic compatibility Weight Without display With display Packaging Single packaging in cardboard

Ceramic Al2O3 (96%) Silicone Polycarbonat PC

Power supply Load Current consumption 1) voltage - 11 V 11.0 ... 33 VDC <supply 0.02 [Ohm] < 20 mA A 13.5 ... 33 VDC / 24 VAC 15% > 10 kOhm < 10 mA 13.5 ... 33 VDC / 24 VAC 15% < 400 Ohm < 30 mA 13.5 ... 33 VDC / 24 VAC 15% < 400 Ohm < 30 mA Short circuit proof and protected against polarity reversal. Each connection is protected against crossover up to max. supply voltage.

< 20 ms < 10 Hz

IP 00 IP 54 IP 65

3 digit

10% fs 40 ... 100% fs

6.2 mm

Vertical, with pressure connections downwards 2) Horizontal with cover downwards 3) Horizontal with cover upwards 4) Mounting bracket (integrated in case)

CE conformity acc. EN 61326-2-3

~ 90 g ~ 100 g

Accuracy
Parameter Tolerance zero point Tolerance zero full scale Resolution Total of linearity, hysteresis and repeatability Long therm stability acc. to DIN EN 60770 TC zero point 4) TC zero point 4) TC sensitivity 4) TC sensitivity 4)
- no additional root-extracted errors
1) 4)

Unit max. max. max. typ. max. typ. max. % fs % fs % fs % fs % fs % fs/10K % fs/10K % fs/10K % fs/10K

0.5 mbar 1.0 1.0 0.2 3.0 1.0 0.2 1.0 0.3 0.6

0 ... 1 mbar 1.0 1.0 0.2 2.0 1.0 0.2 1.0 0.3 0.6

0 ... 3 mbar 0.7 0.7 0.1 1.0 1.0 0.2 0.5 0.2 0.5

0 ... 5 mbar 0.7 0.7 0.1 1.0 1.0 0.1 0.4 0.1 0.5

0 ... 10 - 50 mbar 0.7 0.7 0.1 0.6 1.0 0.1 0.4 0.1 0.2

Test conditions:
2) 5)

25 C, 45% rF, Power supply 24 VDC TC z.p. / TC z.p. 0 70 C


3)

At nominal pressure Signal approximately ~13 Pa lower than actual pressure

At types forcible TC = Temperature coefficient

Signal approximately ~13 Pa higher than actual pressure

2/5

Huba Control type 694 - Technical data subject to change - Edition 09/2011

With root-extracted output (2 100% pressure) Absolute error: (% of full scale) 0 1 mbar max. 0.6
pFS p

TC zero point: (% fs) 03 50 mbar


pFS p

+ 1.5

max. 0.3

+ 1.5

max. 0.6

pFS p

+ 1.5

Order code selection table


Pressure range 1) mbar (hPa) Pa mmWS (mmH2O) inH2O -0.5 0.5 -50 50 -5 5 -0.2 0.2 0 1 0 100 0 10 0 0.4 0 3 0 300 0 30 0 1.2 0 5 0 500 0 50 0 2 0 10 0 1000 0 100 0 4 0 16 0 1600 0 160 0 6.4 0 25 0 2500 0 250 0 10 0 50 0 5000 0 500 0 20 mbar (hPa) Pa mmWS (mmH2O) inH2O Linear Linear Full scale and zero point adjustable by customer Square root extracted Square root extracted Full scale and zero point adjustable by customer 0 ... 10 V 13.5 33 VDC / 24 VAC 15% 0 ... 20 mA 13.5 33 VDC / 24 VAC 15% 4 ... 20 mA 13.5 33 VDC / 24 VAC 15% 4 ... 20 mA 11.0 33 VDC Without In pressure unit chosen above In % fs Connection pipe 6.2 mm without pressure orifice Connection pipe 6.2 mm pressure orifice on P1 Connection pipe 6.2 mm pressure orifice on P2 Connection pipe 6.2 mm pressure orifice on P1 and P2 IP 54 without IP 54 with connection kit (metal), 90 angled including tube 2 m long (Fig. 1) IP 54 with connection kit (plastic), straight including tube 2 m long (Fig. 2) IP 65 without IP 65 with connection kit (metal), 90 angled including tube 2 m long (Fig. 1) IP 65 with connection kit (plastic), straight including tube 2 m long (Fig. 2) Indicate W and state range on order (e.g.: W0 ... + 8mbar/OUT1...6V)

694. 9 X X X X X X X X X
3 1 1 1 1 1 1 1 1 1 2 3 4 5 6 7 0 2 3 1 1 2 4 3 1 3 4 5 1 0 1 2 1 2 3 4 0 1 2 3 4 5 W

Pressure unit

Output signal / adjustment

1 1

Output / power supply 2)

Display 3 digit Pressure connection / Pressure orifices

Acessories / Connection Kit

Pressure range variation

Accessories 3)
Connection kit for vent duct (metal), 90 angled Connection kit for vent duct (plastic), straight DIN-rail mounting adaptor (Fig. 3) Calibration certificate including tube 2 m long (Fig. 1) including tube 2 m long (Fig. 2) Order number 104312 100064 112854 104551

1)

Other pressure range on request

2)

Other output or power supply on request

3)

Accessories supplied loose

Huba Control type 694 - Technical data subject to change - Edition 09/2011

3/5

Dimensions in mm / Electrical connections

3 2 1
2 wire

3 2 1
3 wire

Accessories
Fig. 1 Fig. 2 Fig. 3

92

97 EN 60 715-TH 35-7.5

4/5

Huba Control type 694 - Technical data subject to change - Edition 09/2011

28

35.2

6 11.6

51

Huba Control AG Headquarters


Industriestrasse 17 5436 Wrenlos Telefon +41 (0) 56 436 82 00 Telefax +41 (0) 56 436 82 82 info.ch@hubacontrol.com

Huba Control AG Niederlassung Deutschland


Schlattgrabenstrasse 24 72141 Walddorfhslach Telefon +49 (0) 7127 23 93 00 Telefax +49 (0) 7127 23 93 20 info.de@hubacontrol.com

Huba Control SA Succursale France


Rue Lavoisier Technople Forbach-Sud 57602 Forbach Cedex Tlphone +33 (0) 387 847 300 Tlcopieur +33 (0) 387 847 301 info.fr@hubacontrol.com

Huba Control AG Vestiging Nederland


Hamseweg 20A 3828 AD Hoogland Telefoon +31 (0) 33 433 03 66 Telefax +31 (0) 33 433 03 77 info.nl@hubacontrol.com

Huba Control AG Branch Office United Kingdom


Unit 13 Berkshire House County Park Business Centre Shrivenham Road Swindon - Wiltshire SN1 2NR Phone +44 (0) 1993 776667 Fax +44 (0) 1993 776671 info.uk@hubacontrol.com

www.hubacontrol.com

Вам также может понравиться