Вы находитесь на странице: 1из 4

For Detection of Platelet Dysfunction

The PFA-100 System measures the complex process of primary hemostasis and aids in the rapid detection of platelet dysfunction. It is the first commercially available in vitro testing system that incorporates high-shear flow to provide a realistic hemodynamic environment to initiate and monitor primary hemostasis. PFA-100 Test Cartridges Collagen/Epinephrine (Col/EPI) primary screening cartridge. Collagen/ADP (Col/ADP) differentiates major platelet dysfunction from Aspirin-related dysfunction. The self-contained, disposable cartridges include all functional elements required for testing, while assuring consistency within each lot, as well as limiting test variability.

Clinical Applications Helps to detect inherited, acquired and drug-induced platelet dysfunction Detects von Willebrand disease Assesses the anti-platelet effect of Aspirin Evaluates platelet dysfunction in children Evaluates platelet dysfunction in multiple clinical settings such as high-bleeding-risk surgery, high-risk pregnancy and menorrhagia

PFA-100 Platelet Function Analyzer

Answers for life.

PFA-100 Platelet Function Analyzer Specifications

Test System Principle: Measuring channel: Incubation category: Incubation temperature: Incubation warm up-time: Sample Handling Type of sample: Sample mechanism: Barcode identification: Citrated whole blood Manual pipette dispensing (800 L) Optional external barcode reader for sample identification Reagent Handling Type: Reagent recognition: Additional reagent: Throughput Single test: Duplicate test: Software Interface: Host connection: Patient ID: Stored control results: Stored patient results: Software update: LCD display Unidirectional, external PC RS-232 25 pin D port Maximum 12 numeric characters Last 50 results Last 20 results PCMCIA Card Col/EPI or Col/ADP: 6 tests/hour Col/EPI or Col/ADP: 10 tests/hour Single disposable self-contained test cartridge Positive optical coding reagent cartridge identification Trigger solution High-shear force dynamic flow system Single measuring channel Two incubation wells 37.9 C 1.0 C <15 minutes (room to operating temperature)

Printer Printer: Type: Printer paper: Power Supply Operating voltage: Mains frequency: Current (maximum): Power consumption: Fuse: Note: 100 130 V / 60 Hz, 200 250 V / 50 Hz 50 60 Hz 1.5 A @ 100 V, 0.75 A @ 200 V 110 W (maximum) Sloblo, 250 V, 1.5 A Main supply voltage should not exceed recommended ranges Built-in dot matrix printer, sound level printing 77 dBA Ribbon Epson #ERC-22 black or #ERC-09 purple 58 mm wide paper roll

Environmental Conditions Operating temperature: Non-operating storage: Relative humidity: Altitude (maximum): Operating humidity: Non-operating storage humidity: Heat emission: Dimensions Dimensions: Weight: Remarks The PFA-100 Platelet Function Analyzer is available for sale in the US and complies to the IVD Directive 98/79/EC. 384 mm x 229 mm x 360 mm (W x D x H) 10.9 kg 18 - 29 C -29 - 50 C 20 - 90 % 2000 meters 20 - 90 % RH 0 - 95 % RH 94 kcal/h (375 BTU/h)

Ordering Information Catalog No. B4170-00 B4170-20 B4170-21 B4170-50 Product Description PFA-100 Platelet Function Analyzer Dade PFA-100 Collagen/EPI Test Cartridges Dade PFA-100 Collagen/ADP Test Cartridges Dade PFA-100 Trigger Solution Quantity 1 unit 1 x 20 pieces 1 x 20 pieces 3 x 11 mL

For more information please contact your local sales representative or visit our website.

* Siemens is the exclusive distributor of Sysmex CA Systems in Western Hemisphere and select countries worldwide.

This brochure is for use outside the United States. Product availability may vary from country to country and is subject to varying regulatory requirements. Please contact your local representative for availability.

Siemens Healthcare Diagnostics Inc. 1717 Deerfield Road Deerfield, IL 60015-0778 USA

2008 Siemens Healthcare Diagnostics Inc. Order No. A91DX-0700923-C1-4A00 Order No. 131663 E 08/06-S1 Printed in USA Dade, PFA-100 and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. All other trademarks and brands are the property of their respective owners.

www.siemens.com/diagnostics