INFRARED GAS ANALYZER best suited for measuring exhaust gas in stack
Simultaneous Measurement of 5 components; NOX, SO2, CO, CO2 and O2
Type: ZSU
CO GAS ANALYZER
for Air pollution monitoring
Low concentration CO GAS ANALYZER
Type: ZSE
CO2 CONTROLLER
CO GAS ANALYZER for Automobile tunnel
Type: ZSA
OXYGEN ANALYZER SERIES best suited for combustion control and energy saving
IN-SITU ZIRCONIA OXYGEN ANALYZER
Type: ZSB Type: ZRM Type: ZRY
(Converter)
(Converter)
Type: ZFK
(Detector)
ANALYZER best suited for measuring atmosphere gas in heat treatment furnace
INFRARED system
Single beam type CO, CO2, CH4, SO2, NOX and O2 measuring , 4-component analyzer
Type: ZRJ Type: ZRH
Dual beam type NOX, SO2, CO, CO2 and CH4 measuring , 5-component analyzer
Type: ZKJ Type: ZRF
GAS ANALYZER
He, Ar, and H2 measuring
Type: ZAF
Easy-to-carry PORTABLE
CO, CO2 and CH4 analyzer
Type: ZFY
TYPE
Automobile exhaust gas analyzer for measurement of HC, CO, CO2 and O2
Type: ZKE
Satisfactory products for customers will be delivered under strict quality control.
"Status of manufacturer specified in the Measurement Law" already acquired: Specification No. 391901
2
Is the air really safe? Yes. Because a gas analyzer keeps an eye on pollutant gases.
Simultaneous Measurement of NOX, SO2, CO, CO2 and O2 for stack gas
Type: ZSU
Japanese pattern approval SAS992-1 (SO2 meter) SAC992-1 (CO meter) SAN991-1 (NOX meter) SE981 (O2 meter) SF011 (O2 meter)
FEATURES ..................................................
Measurement by the infrared ray method (dual beam optics) excellent in long-term stability. Hardly affected by unintended gases because interference components' influence is corrected with twin detectors. Standard-equipped with automatic zero/span calibrating function. Space saving design allowing maintenance from the front of each analyzer.
FEATURES ..................................................
Measurement by the infrared ray method (single beam optics) excellent in long-term stability. Provided with CO peak count alarm and self-diagnostic function. Standard-equipped with automatic zero/span calibrating function. Space saving design allowing maintenance from the front of each analyzer.
SPECIFICATIONS .......................................
Measuring object: Exhaust gas of Incinerator and boiler, etc. Measurable components: NOX, SO2, CO, CO2, O2 Measuring system: Dual beam type infrared method
SPECIFICATIONS .......................................
Measuring object: Exhaust gas of incinerator, etc. Measurable components: CO, O2 Measuring system: Single beam type infrared method
(O2: Zirconia method or Paramagnetic method) Measurement range: NOX: 0 to 50 ------- 5000ppm SO2: 0 to 50 ------- 5000ppm CO: 0 to 50 ------- 5000ppm CO2: 0 to 10 / 0 to 20% O2: 0 to 10 / 0 to 25vol% Repeatability: 0.5%FS Linearity: 1%FS Zero/span drift: 2%FS/week (O2: 2%/month) Response time: NOX, CO, CO2, O2: 2 minutes, SO2: 4 minutes (90% response, from the device inlet) Output signal: 4 to 20mA DC Contact output: Auto calibration status, maintenance status, concentration alarm, CO peak count alarm, range identification of each component, etc. Contact input: Auto calibration start, range changeover, pump ON-OFF, etc. Function: Auto calibration, O2 correction, average value calculation, concentration alarm, CO peak count alarm, etc. Display: LCD with back light Recorder: 6-point recorder mounted (option) Power supply : 100V, 110V, 115V, 200V or 230V AC Outer dimensions: 800(W) x 1800(H) x 825(D) mm 3
(O2: Zirconia method) Measurement range: CO: 0 to 200 ------- 2000ppm O2: 0 to 25vol%
Repeatability: Linearity:
0.5%FS 2%/month)
1%FS
Air inlet
Sampling module (1 L/min) Flow meter Membrane filter Gas aspirator Mist filter Gas dryer Flow meter NO2/NO converter Atmosphere Three-way solenoid valve Electronic cooler ZERO NO Pressure regulating valve Positive pressure drain pot Negative pressure drain pot Air SO2 CO2 CO Mist catcher Membrane filter Infrared gas analyzer CO
e C or 15 r m o
Sample gas inlet Rc 3/8
SPECIFICATIONS Temperature of gas used: Max 800C (SUS extraction tube) Max 1500C (SIC extraction tube) System: Electrical heating Material of extraction tube: SUS 316, or SIC + SUS 316 Length of extraction tube: 600 to 2000mm Mounting method: JIS 5K 65A flange Filter: SUS 316 wire mesh Sampling gas outlet: RC1/4 Power supply voltage: 100V AC, 100 VA (for built-in heater)
Drain separator
Infrared gas analyzer NO,SO2 O2 Solenoid valve Pressure regulator O2 meter NO /N2 SO2 /N2 CO2 /N2 CO /N2 O2 /N2
Filter regulator
Standard gas
Exhaust 2
Drain
Opening
Exhaust 1
Simultaneous Measurement of NOX, SO2, CO, CO2 and O2 for stack gas
Type: ZSV
FEATURES ..................................................
The CO sensor adopts measurement by the infrared ray method (single beam optics) excellent in long-term stability. Only CO gas is required as standard gas for calibration. CO zero and O2 span can be automatically calibrated using the air. Space saving design allowing maintenance from the front of each analyzer.
FEATURES ..................................................
A maximum 5 components out of NOX, SO2, CO, CO2, CH4 and O2 can be measured simultaneously. The large LCD screen with clear display assures simple operation. Single-beam infrared gas analyzer having excellent stability is adopted. The separate structure allows the analyzer unite and the pretreatment unit to be placed separately. Measurement can be taken continuously for up to 5 days.
SPECIFICATIONS .......................................
Measuring object: Exhaust gas of incinerators Measurable components: CO, O2 Measuring system: Single beam type infrared method (O2: Galvanic method) Measurement range: CO: 0 to 500 / 1000ppm,
SPECIFICATIONS .......................................
Measuring object: Combustion exhaust gas measurement, etc. Measurable components: NOX, SO2, CO, CO2, CH4, O2 Measuring system: Singlel beam type infrared method (O2: Paramagnetic method or galvanic method) Measurement range: NOX: 0 to 500 ------- 5000ppm SO2: 0 to 500 ------- 1%
0 to 500 / 2000ppm O2: 0 to 25vol% Repeatability: 0.5%FS (O2: 1%FS) Linearity: 2.0%FS Zero drift: 2.0%FS/month (O2: 2.0%FS/week) Span drift: 2.0%FS/week (O2: 2.0%FS/month) Response speed: Within 150 seconds for 90% response from the device inlet Output signal: 4 to 20mA DC Contact output: None Contact input: Pump ON/OFF contact input for driving Function: Auto calibration, O2 correction, average value calculation Indication: LCD with back light Recorder: 6-point recorder mounted (option) Standard gas: Two 3.4L cylinders can be accommodated (option) Power supply: 100V AC 50 or 60 Hz (110 to 240V AC: transformer built in) Outer dimensions: 600(W) x 1580(H) x 675(D) mm
CO: 0 to 200 ------- 100% CO2: 0 to 200 ------- 100% SH4: 0 to 2000ppm ------- 100% O2: 0 to 5 / 25% Repeatability: 0.5%FS Linearity: 2%FS Zero/span drift: 1%FS/day Response time: 50 second for 90% response from the device inlet but 3 minutes for SO2 Output signal: 4 to 20mA DC or 0 to 1V DC Communication function: RS-232C(MODBUS) Functions: Auto calibration, O2 correction, average value calculation Indication: LCD with back light Power supply: 100 to 115V AC or 200 to 240V AC, 50/60Hz Outer dimensions: 365(W) x 574(H) x 514(D) mm 4
FEATURES ..................................................
Excellent stability inherent to infrared ray system. Hardly affected by gases other than monitored. Selectable among 4 ranges; 0 to 10, 20, 50 and 100 ppm.
FEATURES ..................................................
Available in 2 types; infrared ray type and controlled potential electrolysis type. Hardly affected by gases other than monitored. Excellent in long-term stability.
SPECIFICATIONS .......................................
Measuring object: CO concentration of atmospheric air Measuring system: Dual beam type infrared method Measurement range: 0 to10/20/50/100ppm (4-range changeover) Minimum detection sensitivity: 0.5% of full scale Repeatability: 1.0% of full scale Linearity: 1.0% of full scale Zero drift: 1.0% of full scale/week Span drift: 2.5% of full scale/week Response time: Within 120 seconds
SPECIFICATIONS .......................................
Measuring object: CO concentration in automobile tunnels Measuring system: Single beam type infrared method or
(90% response, from the device inlet) Interference influence (Zero point): 1.0% of full scale for moisture (80%RH at 25C) 1.0% of full scale for CO2 0.1%
Automatic calibration: Option Concentration alarm: Option Output signal: 4 to 20mA DC or 0 to 1V DC
calibration status
Contact input: Auto calibration start Power supply: 100V AC 50 or 60Hz
CO2 CONTROLLER
Protected Horticulture
For growing sweet melon and strawberry in greenhouse.
Related products
Building Air-conditioning
For control of room ventilation in a building and save energy due to cooling and heating.
SPECIFICATIONS
Recording method: Built-in memory card Data format: ASCII (Data can be read directly into Excel)
FEATURES ..................................................
With built-in suction pump and filter. Two 1c (SPDT) relay contacts incorporated. Analog output signal incorporated.
output is option)
Input signal: Thermocouples, resistance bulbs, voltage/current Indication accuracy: (0.15%+1 digit) of input range Power supply: 100 to 240V AC Outer dimensions: 160(W) x 144(H) x 185(D) mm Others: Standard-supplied PC support software
SPECIFICATIONS .......................................
Measuring object: CO2 concentration in air Measuring system: Single beam type infrared method Measurement range: 0 to 0.2 ---- 20% Repeatability: 1.0% of full scale Zero drift: 10% of full scale/6 months Response speed: Within 10 seconds (90% response) Concentration output signal: 4 to 20mA DC nonlinear or 0 to
SPECIFICATIONS
Recording method: Ink jet system Recording points: 6 points, max. Recording color: 6 colors Recording paper: Effective width 100mm Z fold, 15m length Measured value intermittent recording: 6 points Printing function: Periodic printing, scale printing, alarm printing,
is option)
Input signal: 4 to 20mA DC, thermocouples, voltages Indication accuracy: (0.3%+1 digit) of measurement range Power supply: 100 to 120V AC or 200 to 240V AC Outer dimensions: 144(W) x 144(H) x 197(D) mm
O2
Detecting unit
FEATURES ..................................................
Automatic calibration and manual/auto blow-down functions are provided. A solenoid valve and other necessary devices are incorporated. Measuring range is freely settable within 2 to 50%. Incomplete combustion level indication is selected when oxygen becomes inadequate.
FEATURES ..................................................
Automatic calibration and manual/auto blow-down functions are provided. Measuring range is freely settable within 2 to 50%. There is no need for gas sampling devices such as gas aspirator and dehumidifier. Incomplete combustion level indication is selected when oxygen becomes inadequate.
SPECIFICATIONS .......................................
Measuring object: Oxygen in combustion exhaust gas Measuring range: 0 to 2-----50vol% O2 (freely settable) Repeatability: 0.5%FS Linearity: 2%FS Response time: Within 10 seconds (From calibration gas inlet) Oxygen concentration output signal: 4 to 20mA DC or 0 to 1V DC Contact output: 4 points (1a relay contact) Upper/lower limit
SPECIFICATIONS .......................................
Measuring object: Oxygen in combustion exhaust gas Measuring range: 0 to 2-----50vol%O2 1(freely settable) Repeatability: 0.5%FS Linearity: 2%FS Response time: Within 7 seconds (From calibration gas inlet) Oxygen concentration output signal: 4 to 20mA DC or 0 to 1V DC Contact output: 4 points (1a relay contact) Upper/lower limit
alarm, maintenance status, blow down status, contact output during calibration Contact input: Auto calibration start, inhibit calibration Indication: Oxygen concentration display (3-digit LED), mode display (3pcs LED), operation/setting display(16digit, 2-line LCD) Combustion efficiency display: Option Communication interface: RS485 (option) Calibration method: Automatic, manual by key operation, input from external contact Power supply: 100, 115 or 230V AC 50/60Hz Construction: Pipe stand type or wall mounting type Cable length from converter part to detector: 20m, max.
alarm, maintenance status, blow down status, contact output during calibration Contact input: Auto calibration start, inhibit calibration Indication: Oxygen concentration display (3-digit LED), mode display (3pcs LED), operation/setting display (16-digit, 2-line LCD) Fuel efficiency indication: Option Communication interface: RS485 (Option) Calibration method: Automatic, manual by key operation, input from external contact Power supply: 100/115 V AC or 200/220/230V AC 50/60Hz Mounting method: Mounted flush on panel or mounted on pipe stand Cable length from converter part to detector: 100m, max.
Sample gas
A flow guide tube skillfully utilizing the flow of the measured gas assures fast response.
Flow guide tube Ceramic filter Heater Flange Reference gas
Measured gas
Output Electrode Zirconia element Standard type 800 to 1500C: with ejector
Various combinations of detector and flow guide tube have been prepared for use under different site conditions.
Note) Blow-down refers to a function for blowing away the dust accumulating in the flow guide tube of detector unit by compressed air.
Application Temperature Gas flow Boiler, etc. 600C or less 5 to 20m/s (General- purpose) Garbage incinerator, 600C or less 5 to 20m/s Sludge incinerator (For anti-corrosion) Heating furnace (General- purpose)
Dust Below 0.2g/Nm3 Below 10g/Nm3 Below 1g/Nm3 Below 10g/Nm3 Below 25g/Nm3 Below 25g/Nm3 800C or less 1m/s or less Below 1g/Nm3 1500C or less 1m/s or less Below 1g/Nm3
Remarks Fuel: Gas, oil Fuel: Coal Small amount of moisture in gas
Adaptable detector/flow guide tube General-purpose With blow-down nozzle For anti-corrosion For anti-corrosion with blow-down nozzle For anti-corrosion high dust For anti-corrosion high dust with cover With ejector With ejector
Mounting method JIS 5K 65A JIS 5K 80A JIS 5K 65A JIS 5K 80A JIS 5K 80A JIS 5K 80A JIS 10K 65A JIS 10K 65A
FEATURES ..................................................
A converter having an immersion-proof structure (corresponding to IP65). Measuring range selectable among 5, 10 and 25%. No need for gas sampling devices such as gas aspirator and dehumidifier. Quick response within 7 seconds.
SPECIFICATIONS .......................................
Measuring system: Zirconia method Measurement range: 0 to 20% Repeatability: 0.5%FS Zero drift: 1%FS/week Span drift: 2%FS/week Response time: Within 20 seconds (From calibration gas inlet) Power supply: 100 to 115V AC or 200 to 240V AC, 50/60Hz Outer dimensions: 170(W) x 140(H) x 190(D) mm
SPECIFICATIONS .......................................
Measuring object: Oxygen in combustion exhaust gas Measuring range: 0-5, 10, 25vol% O2 (Selectable with
changeover pin)
Repeatability: 1%FS Linearity: 2%FS Response time: Within 7 seconds (From calibration gas inlet) Output signal: 4 to 20mA DC Contact output: 1 point (1a relay contact), contact output at
abnormality
Indication: Oxygen concentration display (3- digit LED) Calibration method: Manual calibration with key operation Power supply: 90 to 230V AC, 50/60 Hz Mounting method: Mounted flush on panel Cable length from converter part to detector: 100m, max
Measured gas
Power supply
8
O2
FEATURES ..................................................
Response as high as 2 seconds. Hardly affected by other gases (such as H2 and CO2). A suppression range such as 21 to 100% O2 is settable. Gas-contacting parts are made of corrosion-resisting material Automatic calibration function and communication function can be incorporated (at option).
Exhaust
SPECIFICATIONS .......................................
Measuring principle: Paramagnetic method (Pressure detect
type)
Measurement range: (1) When reference gas is N2
Magnetic field
0 to 2 --------100%O2 (freely settable) (2) When reference gas is air 21 to 23 -----100%O2 or 21to 19 ---------0%O2 (freely settable) (3) When reference gas is 100%O2 100 to 98 -------0%O2 (freely settable) Numbers of range: 2 ranges Repeatability: 0.5%FS (in 10%O2 or higher range) Linearity: 2%FS Response time: Within 2 seconds (From sample gas inlet) Oxygen concentration output signal: 4 to 20mA DC or 0 to 1V DC Moving average time: 0 to 99.9 seconds Contact output: 2 points (1a relay contact) Calibration status, contact output of analyzer error Contact input (Option): Remote range changeover, output hold Indication: LCD display, Oxygen concentration display (4digit),range display(4-digit), message display (24digit, 2-line) Alarm output: Upper/lower limit alarm (Option) Communication interface: RS232 (Option) Automatic calibration method: Zero/span automatic calibration function (option) Power supply: 100 to 240V AC, 50/60 Hz Mounting method: Mounting on 19-inch rack, flush mounting on panel or desk top mounting 9
Mass-flow sensor
Reference gas
+43 +0.01
-0.27 -0.20 +0.30 +0.24 -0.30 -0.26 -0.22 -0.02 -0.02
Sample gas tube Mist filter Drain separator Changeover valve Gas aspirator
Flow meter 0.5L/min Membrane filter 0.3m or less ZKG Oxygen analyzer
Regulation valve
N2
O2
FEATURES ..................................................
Response as high as 15 seconds. Hardly affected by other gases (such as H2 and CO2). The front area measuring 220 x 443 mm is mounted flush with a panel.
Exhaust + drain
SPECIFICATIONS .......................................
Measuring principle: Paramagnetic method (Dumbbell type) Measurement range: 0 to 10,25,50,100%O2 (as specified) Numbers of range: 1 or 2 ranges (as specified) Repeatability: 0.5%FS Linearity: 1.0%FS Zero drift: 2.0%FS/week Span drift: 2.0%FS/week Response time: Within 15 seconds (From calibration gas inlet) Oxygen concentration output signal: 4 to 20mA DC or 0 to 1V DC Indication: 4-digit LED (Oxygen concentration indication) in red Power supply: 100 to 240V AC, 50/60 Hz Mounting method: Mounted flush on panel Outer dimensions: 190(W) x 240(H) x 234(D) mm
Reflecting mirror
Photo detector
+0.15 +0.1
-0.35 -0.25
10
NO SO2 CO Low concentrationCO CO2 CH4 Paramagnetic O2 meter built in H2 He Ar Number of measurable gas components Repeatability Zero drift Span drift Response speed (90% response) 4 to 20mA DC Analog output 0 to 1V DC 0 to 10mV DC Communication interface RS232C Contact output Contact input Indication Automatic calibration Function Upper/lower limit alarm Average calculation O2 correction CO peak count Power supply Outer dimensions (W x H x D) mm
0 to 500ppm -
1, 2
Within 15 seconds / 2 components Within 30 seconds / 4 components Option LCD with back light Option Option Option 100 to 240 V AC 483177493
Within 15 seconds Option Option Digital indication Option 100, 115, 220, 240V AC 483133435
0 to 50ppm 0 to 100ppm 0 to 50ppm 0 to 100ppm 0 to 50ppm 0 to 100ppm 0 to 10ppm 0 to 20ppm 0 to 50ppm 0 to 200ppm 0 to 5% 1 to 5 1, 2 0.5% of full scale Within 2% of full scale/week Within 2% of full scale/week Within 60 seconds Within 25 seconds Option LCD with back light Option Option Option 100 to 240 V AC 483177690 Option Option Digital indication Option Option Option Option 100, 115, 220V AC 483220463
Within 50 seconds -
220835232
SPECIFICATIONS
Measuring object: Combustion gas, heat treat furnace,
SPECIFICATIONS
Measuring object: CO2 in carbonizing furnace and
ceramic furnace Measuring components: CO, CO2, or CH4, any one component Measuring method: Infrared method (Single beam) Measurement range: CO: 0 to 0.5 --------50% CO2: 0 to 0.2 ---- 100% CH4: 0 to1.0 ------ 10% Repeatability: 0.5%FS Linearity: 2%FS Zero/span drift: 2%FS/day Response time: 15 seconds (90% response) Gas sampling: Pump, filter, and flow meter built in Output signal: 4 to 20mA DC Indication: Digital indication Power supply: 100V AC, 50/60 Hz Outer dimensions: 260(W) x 170(H) x 375(D) mm Mass (Weight): Approx. 5kg 11
reforming furnace
Measuring method: Calculate CP value by using CO2
(measuring value), temperature (measuring value or set up value, and CO (set up value) CO2 measuring method: Infrared method (Single beam) CO2 measurement range: 0 to 2% Temperature input: 800 to1000C (K-thermocouple) CO concentration set up: 20 to 35% CO2 repeatability: 0.5%FS CO2 drift: 2%FS/day (both zero and span) Response time: Within 30 seconds (90% response) Gas sampling: Built in pump, filter and flow meter Output signal: 1 to5V DC (0 to 2.0%CP) Indication: Digital indication Power supply: 100V AC, 50/60 Hz Outer dimensions: 260(W) x 170(H) x 375(D) mm Mass (Weight): Approx. 5kg
FEATURES ..................................................
Easy-to-see large LCD is helpful for efficient operation. Measured value output signals are linearizable (at option). Line voltage lies within 100 V to 240 VAC, 50/60 Hz. Zero point and span can be automatically calibrated (at option). Other gases interference is correctable (at option). Commuticable with other system through an RS-232C interface (MODBUS)(at option).
DC constant current DC amplifier Fixed resistor
This thermal conductivity gas analyzer measures gas concentration by utilizing the different thermal conductivities of 2 gas components. In the detector, there are reference and measuring chambers in each of which a thin platinum wire is stretched. The reference chamber is filled with reference gas and through the measuring chamber, sample gas is flowed. Each platinum wire composes a bridge circuit in combination with an external fixed resistor, and it is heated by flowing a constant current. When there is a change in the concentration of the component under measurement, the thermal conductivity of sample gas will change to affect the temperature of the platinum wire in the measuring chamber. The resulting thermal change is taken out as a change in electric resistance, according to which the concentration of measured gas is calculated.
SPECIFICATIONS .......................................
Measuring principle: Meaturement of thermal conductivity Measurement component: He, Ar, H2, CH4, CO2 Measurement range: Refer to Table 1 Repeatability: 1%FS Drift: Within 2% of full scale/week Response time(90% response) : High speed type within 10 sec
Standard type within 60 sec Output signal: 4 to 20mA DC or 0 to 1V DC, 0 to 10 mV DC Display unit: LCD with backlight Mounting: Flush mounting on panel Power supply: 100 to 240V AC, 50/60 Hz Optional specifications: Linearization Relay contact output Contact input Interference gas meatured value input Automatic calibration function Communicating function
0 to 3, 5, 10, 20, 50, 80, 100% 100 to 90, 100 to 80% N2, (CO2, Ar) O2, Air 0 to 5, 10, 20, 30, 40, 50, 80, 100% 100 to 90, 100 to 80% N2, O2, Air, (He) 0 to 10, 20, 50, 80, 100% 100 to 90, 100 to 80% N2, (CO2, Ar, He) 0 to 20, 40, 50, 60, 80, 100% 100 to 80% N2, O2, Air, (He) 0 to 10, 20, 50, 100% 100 to 90, 80%
(Note 1) The parenthesized gases require inquiry. (Note 2) Range ratio stands for maximum value.
Flow checker
1L/min
Span gas
Zero gas
SPECIFICATIONS
Measuring components: HC, CO, CO2, O2 Measurement range:
HC: 0 to 10000ppm CO: 0 to 10% CO2: 0 to 20% O2: 0 to 25% Measuring method: Infrared ray system
x 365(D) mm
Mass (Weight): Approx. 5kg
12
Certain gas absorbs certain range of infrared ray It makes internal pressure in the left detector higher than one in the right detector. By blocking the ray, the pressure in both detectors is equalized. The changing volume of absorbing infrared ray is detected.
Interference
Interference of Moisture Cross sensitivity to moisture: Less than 2 ppm at 2C of satulated H2O
without ICD with gas filter with ICD 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Infrared
There is less fluctuations of indication caused by the time lapsed of light source due to single source and dual beam method.
P+P
P Sample gas
Mass-flow
sensor
Basic Principle
Resistance measurement (Detecting Temp. change by Pressure change)
No moving parts Great Noise proof due to low impedance sensor High sensitive sensor introduces good linearity Impure gas in detector which causes interference can be removed by making full vacuum condition.
Reference Cell
Sample Cell
wind
gap Rotary sector
1mm sq
Distribution cell
Single
point chopper
Measuring gas
Measuring cell
Single point chopper is applied for a rotary sector with excellent S/N ratio. The peculiar shape allows an equal
Temperature of Heat wire
Infrared ray source Chopper
Calm situation
Repeatability: 0.2 to 0.5%FS because fo excellent S/N ratio against conventional double point chopper
High sensitivity / High stability / High vibration proof. Super compact / Non-moving part
13 14
Exhaust
ZKJ main unit Sample gas tube 15C or higher Gas aspirator NO2 /NO converter (NOx measuring time) Electronic gas cooler NO,SO2 Gas conditioner ZERO NO Exhaust SO2 CO2 CO Electromagnetic valve O2 Decompression valve
N2
or
Membrane filter
Flow meter
Sample gas
Membrane filter
Flow meter Electronic dehumidifier Safety drain trap Drain pot Ball valve Drain Standard gas for zero calibration Standard gas for span calibration
Air
NO /N2
SO2 /N2
CO2 /N2
CO /N2
O2 /N2
Drain
Exhaust
SV2
Reference cell
Note 2
Analyzing part
Exhaust
Note 1) Remove dust enough (to protect pump, flow meter, etc.) Note 2) Unnecessary when sample gas is dry.
15
Applications of Analyzer
Application
Incinerator Desulfurization and denitration of exhaust gas General combustion furnaces (including boilers) Marine boiler Diesel generation Automobile exhaust gas For fermentation Incubator Superconduction device
Measured gas
SO2, NOX, CO, CO2, O2 SO2, NOX, O2 SO2, NOX, O2 SO2, NOX, O2 SO2, NOX, O2 CO, HC, CO2, O2 methanol, CO2 CO2, O2 He CO2, O2
Model
ZSQ, ZSU, ZSW ZSU ZSU ZRG, ZRF, ZRM, ZFK ZSU, ZSV ZKE ZRG, ZRH, ZRJ, ZSV ZFP, ZFK, ZRJ, ZSV ZAF ZFP, ZRJ ZAJ, ZAF, ZRG, ZRJ ZRH, ZAF, ZAJ, ZRJ ZRH, ZAF, ZAJ, ZRJ ZRH, ZFK, ZRM, ZRJ ZRH, ZRJ, ZAJ ZRH, ZAF, ZRJ ZRG ZFK, ZRM, ZRY, ZSB ZRH, ZRJ, ZSQ, ZSU ZRG, ZRH, ZRJ ZRG ZRH, ZRG, ZRJ ZRH, ZRJ, ZRY, ZFK ZRJ, ZAJ ZRH, ZRJ, ZAJ ZSU ZRH, ZFX, ZFY, ZRJ ZRH, ZRJ, ZFX, ZFY ZRH, ZRJ, ZFY ZRG ZFP ZFP, ZRH, ZRF, ZRJ ZRH, ZRJ ZRH, ZRJ ZRH, ZAF, ZRJ, ZRF, ZRG ZFK, ZRM ZRG, ZRF ZSA ZSA, ZFP ZFP ZRF, ZRH, ZFP, ZFY, ZAJ, ZRJ ZRJ, ZAF ZRJ, ZAF
Air pollution
Biochemistry (microbes) Ultra-low temperature Storage and maturation of fruits Gas separation
Blast furnace Converter funacer Heat treatment furnace Sintering (pellet equipment) Coke oven (CDQ) Oxygen aspirator Boiler, Heat treatment funacer Detection of explosive gas (hydrocarbon gas and others) Detection of poisonous gas Analysis of gases produced through combustion of new building materials Marine inert gas analysis Tunnel kiln Cement Sewerage incinerator Generating furnace Carburization furnace Annealing furnace Nitration furnace Green house Reserch of photo synthesis Oil refining plant Petrochemical plant Gas production plant Solder furnace Ambient Concentration in tunnel Parking area Building management Building ventilation system Laboratory Gas generation for fuel cell
O2, Ar, He, CO CO, CO2, H2, O2 CO, CO2, H2, O2 CO, CO2, O2 CO, CO2, O2 CO, CO2, H2, O2 N2O, CF3, CHBrCR O2, CO2, CO HC CCR4 CO, CO2, HC, O2 CO2, O2 CO, O2 CO, CO2, O2 SO2, NOX, CO, O2 CO2 CO2,CO, O2 CO2,CO, O2 NH3 CO2 CO2 CO, CO2, CH4 CO, CO2, CH4 CO, H2, He, Ar, O2, CO2 O2 CO, CO2 CO CO, CO2 CO2
Hazard prevention
Environmental safeguard
Various gases
H2, CH4, CO, CO2 CO, CO2, CH4, H2, O2
16
Examples of Application
Most recommended for energy saving in airconditioning of buildings is a CO2 controller !
The CO2 gas concentration in a room is required to be within 1,000 ppm by law in Japan. To meet this, the fresh outdoor air is always taken in. Control of the air intake at an appropriate level will save energy to run the airconditioner for cooling and heating.
Residence district Air-conditioning equipment Damper M2 Damper M1 Ambient air
Measuring CO2 gas Control signal (Contact ON/OFF) CO2 controller ZFP
CO/O2 gas analyzers effective for curbing dioxins emission from waste incineration plants.
Continuous measurement of CO and O2 concentrations in exhaust gas is obliged by antipollution ordinance and dioxin regulation. Gas cooling tower Dust collector Gas extractor NOX removal equipment Gas extractor
Incinerator
ZSB type
ZSU type
Measurement of O2
Measurement of CO, O2
NOx, SO2, CO and O2 gas analyzers indispensable for combustion control and exhaust gas regulation of boilers.
Factories in an area under control of total air pollutant emission due to exhaust gas are obliged to measure NOx and SO2 continuously. Dust collector, bug filter
Gas extractor
Air heater
ZSB type
ZSB type
ZSU type
O2 measurement
Measurement of O2
17
Infrared CO2 gas analyzers essential for measuring CO2 and O2 of heat treatment and reforming furnaces.
Propane and butane gases are brought into contact with a hot catalyzer to generate the gases to be supplied into a carburizing furnace. These gas analyzers measure each concentration of CO2 and O2 gases to be supplied into a carburizing furnace. Source gas Air In iron treatment in a carburizing furnace, the CO2 concentration of the in-furnace atmosphere is an index of furnace work.
Reforming furnace
CO2, O2 analyzer
ZRJ type
Portable CP meter
ZFX type
Infrared CH4, CO2 and O2 gas analyzers optimum for bio-gas measurement.
Alternative current wet refuse Front process facilities Methane fermentation tank Gas sampling equipment Desulfurization device Gas sampling equipment Refiner Fuel buttery
Infrared CO2 and O2 gas analyzer for storage of foodstuffs such as vegetable and fruit.
Foodstuffs can be kept fresh by controlling the CO2 and O2 concentrations properly in a storage house. Storage Carbon dioxide gas generator For growing vegetables, etc. steadily in a greenhouse, the internal CO2 concentration needs to be kept constant. Because this controller incorporates a CO2 sensor and control function, the concentration of CO2 gas emitted from the CO2 generator CO2 installed in a controller ZFP type greenhouse can be controlled.
Change of concentration causes the start of operation CO2 gas generator
Contact signal
Contact signal
CO2, O2 analyzer
ZRJ type
CO2
18
Head Office Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fesys.co.jp/eng
Instrumentation Div. International Sales Dept. No.1 , Fuji-machi, Hino-city, Tokyo,191-8502 Japan Phone : 81-42-585-6201,6202 Fax : 81-42-585-6187 http://www.fic-net.jp/eng