Lung Cancer Detection Using an Electronic Nose

Anna Folinsky California Institute of Technology March 10, 2005

Abstract Electronic nose systems have been used both in and out of laboratory settings for a wide variety of applications. One of the larger general initiatives is towards using them in biomedical applications, notably the detections of analytes which may be correlated with disease states. This proposal outlines a system to detect biomarkers found in the breath which are associated with lung cancer, one of the leading causes of death in the USA. Use of standard polymer/carbon black sensors from our lab will be augmented with novel sensing technologies from our lab, in an eort to obtain the necessary sensitivity and discriminatory power needed for the task. These should be able to dierentiate diseased and healthy states, and potentially separate dierent stage cancer patients.

Introduction
Lung Cancer
It has long been known that certain diseases produce volatile compounds that can be smelled on the breath, or elsewhere on the body. Two of the oldest, most common examples are the scents of ketones on the breath of diabetic patients, and the smell of freshly baked bread on

the skin of patients with typhoid. 1 More recently, GC studies have shown quantitatively that many diseases produce distinct patterns of volatile compounds that could potentially be used as biomarkers for these diseases 25 . Among these, lung cancer has been shown by several GCMS studies to display elevated levels of several volatile organic compounds (VOCs) on the breath, mostly C4 to C20 monomethylated alkanes, in addition to certain benzene derivatives
6,7

, although there is some variaton between studies as to precisely which compounds are

most indicative. On one occasion, Phillips et al identied styrene, 2-methylheptane, and decane and three of the most important compounds for discrimination 8 ; in another, butane, 3-methyltridecane, and 4-methyloctane 2 were most informative. Lung cancer is one of todays leading health problems. It was the third leading cause of death in the USA in 2002, behind only heart and cerebrovascular diseases. Estimates indicate that 172,570 new cases will be reported in 2005, and 163,510 deaths from lung cancer are expected the same year. The current 5-year survival rate for lung cancer is 15%, but this rate rises to 49% if the cancer is discovered when it is still localized. Unfortunately, only 16% of lung cancer cases are discovered while they are still localized 9 . There is clearly a great need to improve early detection ability for this disease. Bronchoscopy, biopsy, and sputum cytology are the current most common ways to diagnose lung cancer, but these methods can occasionally miss tumors, and are dependent on tumor size 10 . There are also some reports of new methods for earlier detection, such as uorescence bronchoscopy 11 , spiral CT scanning 12,13 , PCR sputum assays 14 , and using computers to aid in the analysis of chest radiographs 12 . All of these, however, are expensive and time consuming. A non invasive breath test would have great potential as a widespread screen. II

Electronic Nose Systems


The vapor sensing method we pursue in our lab involves the use of arrays of sensors. No sensor is designed to respond specically towards an individual compound. Instead, each sensor is broadly responsive to a variety of odorants. Each analyte produces a distinct ngerprint from the array of broadly cross-reactive sensors (Fig.1). Pattern recognition algorithms can then be used to obtain information on the identity, properties, and concentration of the exposed vapor 1518 . In this respect, our system resembles that used in the mammalian olfactory system, in which each olfactory receptor responds to a wide variety of odorants 19 , and our array of sensors may be seen as analogous to the array of receptors in the nasal epithelium. Due to this similarity, our system is sometimes designated as an electronic nose.

Figure 1: Dierentiation between odorants: (a) an array of broadly-cross reactive sensors in which each individual sensor responds to a variety of odors; (b) pattern of dierential responses across the array produces a unique pattern for each odorant or odor.

A variety of signal transduction mechanisms have now been implemented to construct electronic nose systems. Surface acoustic wave devices (SAWs) 20 , metal oxide sensors 21 , conducting organic polymers 22 , polymer coated quartz crystal microbalances (QCMs) 23 , polymer coated micro-machined cantilevers 24 , thin lm capacitors 25 , and polymer composite

III

Figure 2: Swelling induced resistance changes of conducting composite vapor sensors chemically sensitive resistors 26 have all been used. Except for metal oxide sensors, they all share the trait that the analyte in the vapor phase is sorbed into a lm coated onto the transducer, and the dierential sorption of the various analytes into an array of such lms generates the distinct pattern produced by the devices. Dierences then result from varying the sorptive coating, general convenience of use and ability to reduce noise. Additionally, some variance is found in the complexity of the required data analysis and data reduction algorithms. Our focus to date has been on sensors prepared from composites of conductive carbon black particles interspersed with insulating polymers. These are low power, inexpensive, lightweight, and allow for simple transduction circuitry as well as simple data analysis methods. The sorption of a vapor induces swelling in the polymer, which reduces the number of conductive pathways in the lm, and increases the resistance of the lm, which is readily measured (Fig. 2). At small swellings, the sensors are highly reversible and respond rapidly to the presence of

IV

analyte vapors, demonstrated reproducibly over thousands of cycles over a period of months, to a variety of organic vapors. The thin lms have been used to obtain response times in essentially real time, with rise and fall times of less than 100ms in most cases and less than 20 ms in some systems 27 (limited by the diusivity of small molecules permeating through low glass transition, rubbery, polymeric lms at room temperature). This proposal will be centered upon this technology, to be augmented by newer sensor substrates from our lab.

Proposed Studies
Rationale
This proposal will use the various sensor systems we have been developing to detect lung cancer biomarkers in patients breath. GC studies showing that several VOCs were elevated in the breath of lung cancer patients were able to use discriminant analysis on the alveolar gradient of the GC-MS information to distinguish between healthy and cancerous patients
2,3

. Certain compounds of interest (styrene, decane, and isoprene among them) are found

at 1-20 parts per billion (ppb) in healthy breath, but are seen at levels from 10-100 ppb in cancerous patients 28 . Yu and coworkers used a GC column coupled with a SAW sensor and a neural network to discriminate between healthy and diseased patients 29 , with preconcentration. Another study used an array of coated QCM sensors, with no preconcentration 30 , to detect lung cancer in patients prior to surgical tumor removal. They made no eort to quantify the analytes detected, nor correlate them to stage of disease. However, they did show that the signature V

returned to a healthy pattern approximately one month after removal of the tumor in a pair of cases, providing further evidence that the analytes are related to the disease. Coated QCMs are comparable in sensitivity to SAW devices, however, carbon black / polymer sensors have been shown to be more sensitive towards alkanes and related VOCs 31 ; therefore the initial step of detection should be easily obtainable. Our lab has already demonstrated the ability to detect and discriminate between all the low molecular weight straight-chain alkanes, as seen in Fig. 3 and to quantify their concentrations based on the amplitude of recorded patterns 32 . Separately, it has been shown that patterns for alkanes and those for a variety of aromatics can be dierentiated 33 . We have also shown that even very similar alkane mixtures can be discriminated 26 . A binary mixture of n-heptane at P/Po = 0.0011 (64 ppm) and n-hexane at P/Po = 0.00090 (192 ppm) has been dierentiated from a binary mixture of n-heptane at P/Po = 0.00090 (58 ppm) and n-hexane at P/Po = 0.0011 (235 ppm) with 95% correct discrimination, using an optimal subset of a chemiresistive detector array. Similarly, a mixture of 1-propanol at P/Po = 0.0025(72 ppm) and 2-propanol at P/Po = 0.0025 (150 ppm) could be dierentiated with 98% correct classication rate from a mixture of 2-propanol at P/Po = 0.0027 (164 ppm) and 1-propanol at P/Po = 0.0023 (64 ppm).

Current Methods
Human breath is composed mostly of water vapor, with all other analytes existing only as minor substituents. This fact implies the need to deal not only with the low levels of the analytes of interest but also with the high background level of water. While the level of water is considered a primary obstacle for commercial sensor arrays, whose polar, VI

Figure 3: Principal components data from a 20-detector array exposed 5 times to of the labeled analytes, each at 0.005 - 0.03 P/Po , containing 99% of the total variance. The ellipsoids contain 99% of the data for each analyte. All presentations were in each set randomized over all repetitions inherently conducting polymer sensors (made from inherently conductive materials, such as polyaniline, polypyrrole, polythiophene, etc) are highly sensitive to water vapor, our approach of using composites of inorganic conductors and sorptive insulating organic phases allows development of chemiresistive sensors that are relatively insensitive to water vapor 19 . There are also a pair of newer sensing substrates being tested, expected to yield improved sensor classication. The rst of these is based on composites of homogenous or blended organic nonvolatile molecules with conductors such as carbon black 34 . These sensors have

VII

been constructed from either pure compounds or mixtures of moderate length monomeric organic molecules used as binders, mixed with carbon black. The sensors show fast response time, good reversability, and high stability. Furthermore, they show the ability to discriminate and classify both similar and dierent types of analytes, even at low concentrations in air, compared to similar sensors based on polymer binders. They also allow an even broader choice of substrates, and also allow us to achieve a higher density of functional groups in the thin lms than we are allowed by using the polymer composites. Principal component analysis of the sensor array responses showed clear distinctions between dierent types of vapors allowing easy classication. This approach is particularly applicable to the development of biosensors where the ability to detect low concentrations of specic types of compounds is required. The second approach uses ligand-capped Au nanoparticles. Thiol capped gold nanoparticles in the range of 2-10 nm have been synthesized and tested as sorption based detectors for dierent analytes 35 . Thiols investigated varied by chain length, polarity, and functional group. Eects of chemical functionalization of the thiol on sensor sensitivity , specicity, and stability were studied. Sensor specicity varied with the thiol functional group. Most thiolate gold nanoparticles chemiresistors showed good stability over three months, with longer chain length thiols having better stability than short chain lengths, and a high ability to classify and discriminate analytes according to their polarity and vapor pressure. In specic cases, the selectivity and discrimination of such arrays were superior to arrays of carbon black/polymer composite sensors. Recent work in our lab has shown that such sensors can dierentiate between mixtures of long chain alkanes, at very low concentrations, in the presence of saturated water vaVIII

por. Two mixtures of nonane and hexadecane were tested in background ows of saturated water vapor at room temperature, to roughly simulate human breath (human breath is at 34.5o C saturated water vapor; however, the dierence is minor for these demonstration purposes). Standard exposures consisted of a background stream of water-vapor saturated air which contained one of two predetermined mixtures of nonane and hexadecane. The low concentration mixture had nonane at P/Po = 0.0010 (4 ppm) and hexadecane at P/Po = 0.010 (40 ppb) in saturated water vapor, and was representative of the key alkane VOC components of a healthy patients breath, while a higher concentration mixture of nonane at P/Po = 0.0050 (20 ppm) and hexadecane at P/Po = 0.050 (200 ppb) in saturated water vapor was used to be representative of the key elevated concentration VOC components that have been reported as diagnostic of lung cancer. Sixty exposures to each breath type were performed at ows of 5 L/min and data were obtained for carbon black/polymer composites, carbon black/monomer composites, and ligand capped Au nanoparticles. Other than selection for sensors we knew empirically to provide generally good responses, the sensors were not optimized for the task. Figure 4 presents the principal component plots for the normalized data of the three sensor classes used in the preliminary experiment. The principal component plots are simply transformations of the multivariate responses of the sensor array into orthogonal directions which statistically capture the most variance between the data, and therefore allow convenient visualization of the dierentiation ability of the sensor array for the analytes of interest, through a reduction in dimensionality of the data. As shown in Fig 4, discrimination between cancerous and healthy breath was possible with each class of sensor arrays. The best discrimination ability was exhibited by the carbon black/monomer composite sensors, which IX

Figure 4: Principal componenent plots indicating the ability to distinguish between healthy and cancerous patients. (a) 6 sensors of poly(vinylstearate),poly(ethylene-co-vinyl acetate), and poly(ethylene vinyl alcohol), all 40 wt% carbon black (CB). (b) 4 sensors of 2-5 nm Au colloids, capped with hexanethiol or 6-mercapto-1-hexanol. (c) 8 sensors of lauric acid, tetracosane, and tetracosanoic acid, with 30 wt% dioctyl phthalate, and tetracosanoic acid, all with 75 wt% CB can be attributed either to the larger number of those sensors, or to the unique properties of these sensors. For the other two classes, increasing the diversity of chemiresistors (in an array of sensors) and modifying the physical properties of their building blocks is expected to give better discrimination. Since our sensors operate solely by detecting changes, we will be able to detect these analytes even in the presence of a constant background of other components of an analyte mixture (as has been previously demonstrated in a variety of situations for nerve agent simulants at low concentrations in the presence of many dierent X

background ambients 36 ). This, and the prior data, demonstrate the ability of our sensors to discriminate between levels, and mixtures, of straight chain alkanes that have been reported to be signatures of lung cancer in breath, even in the presence of saturated water vapor, and without preconcentration.

Experimental
We will receive 4 L samples of lung air in inert Tedlar bags from our clinical partner, Dr. David E. McCune, Chief, Clinical Studies Service, and Director of Clinical Trials at the Madigan Army Medical Center, Tacoma, Washington. The Army is one of the largest healthcare providers in the USA, and this could facilitate ready expansion of the number of testing sites if such is warranted further in the study. Vapor sampling will be done by extended breath sampling, in which the patient will breathe into the collection apparatus for 10-30 min. The rst two minutes of breath sample will be discarded, due to likely contamination of upper respiratory air. The later deep lung air will be retained for testing purposes. These samples will be collected with a straw, or other suitable tube in the patients mouth, that is connected to the collection bag. These subjects will be diverse, and background information will be obtained from patients regarding such factors as age, sex, race, smoking history, etc, which will allow us both to provide for diversity in our data, and also to match each patient with a control subject. We will take these samples back to Caltech and ow them into a home-made chamber containing our sensors, at a relatively slow ow rate (approx. 100-300 mL/min), to obtain sensor response data, and will also use these samples in a parallel GC-MS study.

XI

The targeted VOCs are diverse enough that comparison studies will be needed to validate the results we will obtain with our chemiresistor sensor arrays. We will do this via a GC study in parallel with acquiring the sensor response data. A portion of each sample will be retained for use with our GC-MS system, coupled with an automatic headspace sampler. As standard headspace/gas sampling techniques have sensitivities only in the ppt-ppm range 37 , we will use purge and trap thermal desorption or solid phase micro-extraction methods to allow GC-MS to attain the necessary ppb detection levels to provide the information on the VOC level and composition of our samples. We will run several parallel sensor studies to both validate the responses and to optimize the testing method. The initial method will involve owing the as-received samples through our sensor chambers and a comparison of the breath sample results with the response to a clean background of laboratory air. We will only use sensors having minor sensitivity to water (as those used in Fig. 4). Additionally, in parallel analyses, we will remove the water from the samples by running the analyte ow through a desiccant chamber before presenting the ow to the sensors. Both methods will be compared critically for their performance with the breath samples. Preconcentration will also be used on the samples to produce higher concentrations of the VOCs to be analyzed. Another key step will be to create articially the VOC sample concentrations indicated by the GC-MS, most likely through successive dilutions of the saturated vapor phase of the pure VOCs, in order to create the extremely low concentration levels needed. That mixture will then be exposed to the sensors to determine whether the breath sample response pattern is indeed due to the VOCs detected by the GC-MS method, or whether the sensor pattern arised from dierent volatile breath biomarkers. The rst set of samples will be taken from patients with known conditions, to XII

allow us to calibrate and perfect the sampling and signal processing methods. Subsequent samples will be analyzed by personnel who are blind as to whether the patient has lung cancer (as diagnosed by conventional methods) or not. Ideally, we would be able to use this data to also classify patients on the basis of stage of their cancer. This would clearly be a great aid in diagnosing lung cancer at the clinically more useful earlier stages. Some information suggests this may not be possible, however. One study suggests that the elevated VOC levels are comparable in all stages of the cancer, because they are not associated with the size of the cancerous tumor, but instead are a signature of oxidative stress related to a change in body chemistry as a result of development of the cancerous condition 2 . With ndings consistent with this hypothesis, a study in India was performed on 108 patients with abnormal chest radiographs who were scheduled for bronchoscopy 38 . The breath of these patients was assayed by GC-MS. Lung cancer was conrmed histologically in 60 patients. A combination of 22 VOCs, predominantly alkanes, alkane derivatives, and benzene derivatives, discriminated between patients with and without lung cancer, regardless of stage (all p < 0.00003). For stage 1 (initial) 39 lung cancer, a combination of 22 VOCs had 100% sensitivity and 81.3% specicity. Cross validation of the combination correctly predicted the diagnosis in 71.7% of the patients with lung cancer and 66.7% of those without lung disease. Sensors with good shelf life and homogeneous response can be fabricated in batch form in our lab. When using spray coating, batches of sensors typically have 10% dierence in their baseline resistances and 15% variation in their response during a usage period over several months 19 . Data analysis will be performed initially using standard chemometric methods such as principal components analysis and linear discriminant analysis, which are easily XIII

implemented using commercial software packages, such as MATLAB, with algorithms that already exist and have been extensively previously used in our lab. Although these methods have proven to be quite useful for analyte discrimination, more sophisticated algorithms (e.g. supervised or unsupervised neural networks, or a variety of non-linear methods) will be employed if necessary. However, if, as suggested, the biomarkers are entirely due to static levels of production due to oxidative stress, that is also of great interest. While we wont be able to dierentiate, we will still be able to detect at early stages, which would still be enormously benecial as an early detection screen.

References
[1] Cone, T. E. Pediatrics 1968, 41(5), 993. [2] Phillips, M.; Cataneo, R. N.; Cummin, A. R. C.; Gagliardi, A. J.; Gleeson, K.; Greenberg, J.; Maxeld, R. A.; Rom, W. N. Chest 2003, 123(6), 21152123. [3] Phillips, M.; Cataneo, R. N.; Ditko, B. A.; Fisher, P.; Greenberg, J.; Gunawardena, R.; Kwon, C. S.; Rahbari-Oskoui, F.; Wong, C. Breast J 2003, 9(3), 18491. [4] Kaji, H.; Hisamura, M.; Saito, N.; Murao, M. Journal of Chromatography 1978, 145(3), 464468. [5] Olopade, C. O.; Zakkar, M.; Swedler, W. I.; Rubinstein, I. Chest 1997, 111(4), 862865. [6] Gordon, S. M.; Szidon, J. P.; Krotoszynski, B. K.; Gibbons, R. D.; Oneill, H. J. Clinical Chemistry 1985, 31(8), 12781282. [7] Preti, G.; Labows, J. N.; Kostelc, J. G.; Aldinger, S.; Daniele, R. Journal of Chromatography-Biomedical Applications 1988, 432, 111. [8] Phillips, M.; Gleeson, K.; Hughes, J. M. B.; Greenberg, J.; Cataneo, R. N.; Baker, L.; McVay, W. P. Lancet 1999, 353(9168), 19301933. [9] Jemal, A.; Murray, T.; Ward, E.; Samuels, A.; Tiwari, R. C.; Ghafoor, A.; Feuer, E. J.; Thun, M. J. CA Cancer J Clin 2005, 55(1), 1030. [10] Schreiber, G.; Mccrory, D. C. Chest 2003, 123(1), 115S128S. [11] Banerjee, A. K.; Rabbitts, P. H.; George, J. Thorax 2003, 58(3), 266271. XIV

[12] Henschke, C. I.; Yankelevitz, D. F. Radiologic Clinics of North America 2000, 38(3), 487. [13] Kaneko, M.; Eguchi, K.; Ohmatsu, H.; Kakinuma, R.; Naruke, T.; Suemasu, K.; Moriyama, N. Radiology 1996, 201(3), 798802. [14] Boiselle, P. M.; Ernst, A.; Karp, D. D. American Journal of Roentgenology 2000, 175(5), 12151221. [15] Geladi, P.; Kowalski, B. R. Analytica Chimica Acta 1986, 185, 117. [16] Kowalski, B. R.; Bender, C. F. Analytical Chemistry 1972, 44(8), 1405. [17] Duda, R. O.; Hart, P. E. Pattern classication and scene analysis; Wiley: New York,, 1973. [18] Burns, J. A.; Whitesides, G. M. Chemical Reviews 1993, 93(8), 25832601. [19] Lewis, N. S. Accounts of Chemical Research 2004, 37(9), 663672. [20] Ballantine, D. S.; Rose, S. L.; Grate, J. W.; Wohltjen, H. Analytical Chemistry 1986, 58(14), 30583066. [21] Srivastava, R.; Dwivedi, R.; Srivastava, S. K. Sensors and Actuators B-Chemical 1998, 50(3), 175180. [22] Partridge, A. C.; Jansen, M. L.; Arnold, W. M. Materials Science & Engineering CBiomimetic and Supramolecular Systems 2000, 12(1-2), 3742. [23] Fu, Y.; Finklea, H. O. Analytical Chemistry 2003, 75(20), 53875393. [24] Lang, H. P.; Baller, M. K.; Berger, R.; Gerber, C.; Gimzewski, J. K.; Battiston, F. M.; Fornaro, P.; Ramseyer, J. P.; Meyer, E.; Guntherodt, H. J. Analytica Chimica Acta 1999, 393(1-3), 5965. [25] Willing, B.; Kohli, M.; Muralt, P.; Oehler, O. Infrared Physics & Technology 1998, 39(7), 443449. [26] Burl, M. C.; Sisk, B. C.; Vaid, T. P.; Lewis, N. S. Sensors and Actuators B-Chemical 2002, 87(1), 130149. [27] Briglin, S. M.; Lewis, N. S. Journal of Physical Chemistry B 2003, 107(40), 11031 11042. [28] Personal communication. Wang, P. 2005. [29] Detection volatile organic compounds in breath as markers of lung cancer using a novel electronic nose. Yu, H.; Xu, L.; Cao, M.; Chen, X.; Wang, P.; Jiao, J.; Wang, Y. 2003. [30] Natale, C. D.; Macagnano, A.; Martinelli, E.; Paolesse, R.; DArcangelo, G.; Roscioni, C.; Finazzi-Agro, A.; DAmico, A. Biosensors and Bioelectronics 2003, 18(10), 1209 1218. XV

[31] James, D.; Scott, S. M.; Ali, Z.; OHare, W. T. Microchimica Acta 2005, 149(1-2), 117. [32] Severin, E. J.; Doleman, B. J.; Lewis, N. S. Analytical Chemistry 2000, 72(4), 658668. [33] Sisk, B. C.; Lewis, N. S. Sensors and Actuators B-Chemical 2003, 96(1-2), 268282. [34] Array-based vapor sensing using chemically sensitive, carbon black - monomeric organic molecule resistors (submitted). Gao, T.; Lewis, N. S. [35] Briglin, S. M.; Gao, T.; Lewis, N. S. Langmuir 2004, 20(2), 299305. [36] Matzger, A. J.; Vaid, T. P.; Lewis, N. S. In Proc. SPIE Vol. 3710, p. 315-320, Detection and Remediation Technologies for Mines and Minelike Targets IV, Abinash C. Dubey; James F. Harvey; J. Thomas Broach; Regina E. Dugan; Eds., pages 315320, 1999. [37] http://sisweb.com/reference/applnote/app-39.htm. [38] http://www.indiandoctors.com/iaso/oncoinfo.htm. [39] http://www.cancer.org/docroot/CRI/content/CRI 2 4 3X How is lung cancer staged 26.asprnav=cri.