Operators Handbook
Siemens AG 2006
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved
The information contained herein is the property of Siemens Magnet Technology Ltd and may not be copied in whole or in part except with the prior permission of Siemens Magnet Technology Ltd or, if it has been furnished under a contract with another party, as expressly authorised under that contract.
Siemens Magnet Technology Ltd Headquarters Support Centre Wharf Road Eynsham Oxford 0 X 2 9 4BP ENGLAND Tel: +44 (0) 1865 850164 Fax: +44 (0) 1865 850174 Email: mailto:magnet.hsc.med@siemens.com
Revision/Disclaimer
Document Revision Level
This document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed. Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorised by Siemens or one of its affiliates to provide such services. Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorised by Siemens or one of its affiliates before attempting installation or service procedures.
Table of Contents
1
1.1 1.2 1.3
General
General Description Technical Specification Environmental Conditions
4
4 4 5
Indoor use 2
2.1 2.2 2.3 2.4
5 5
5 6 8 8
Equipment Operation
Helium Level Meter Kit Contents Instructions for use Helium Volt Sense Lead Instructions for Cleaning
Equipment Maintenance
REVISION CONTROL SHEET
8
9
1 General
1.1 General Description
The 4018 Helium Level Meter has been specifically designed for measurement of liquid helium levels within Siemens superconducting MRI magnets during transit and preliminary stages of installation i.e.; without electrical power and without the associated electronics connected. It is only to be used on the following products; OR24, OR26, OR35, OR42, OR48, OR50, OR58, OR64, OR70, OR71, OR76, OR91, OR92. If you are in any doubt please contact Siemens Magnet Headquarters Support Centre. The 4018 Helium Level Meter is part of a kit used to complete a circuit with the magnet helium level probe. A propagating current is sent through the magnet probe and the resistance of the probe is measured. The result is fed into the Helium Level Meter and an electronic comparison made between it and known values. The meter uses 10 LED's to display the level of helium liquid in steps of 10%, the meter is calibrated to 5%.
Mains supply voltages not to exceed 10% of the nominal voltage (for either PSU) AC supply cable - Standard IEC320 connector Power supply output connector -5 way DIN plug 180 Helium Level Meter Current output Output connector Probe resistance resistance
10
1 amp 30 volts
2 Equipment Operation
2.1 Helium Level Meter Kit Contents
Part Number 10103563 10102262 10123452 10123451 08132222 10108588 10104512 10104088 10108406 4018 Helium Level Meter Sinpro Electronic SPU50-7 30V PSU PSU lead American PSU lead European 3m Helium probe lead Carry Case Extension lead 4018 HLM Analogue Volt Sense Lead Operators Handbook
Power Supply
Operators Handbook
Extension lead
WARNING
Strong magnetic fields. Risk of personal injury and equipment damage. The DC supply unit must remain as far away as possible from the magnet if it is at field. = > Ensure the leads are at full length before connecting to the helium level meter.
Connect the DC Supply Unit to a suitable mains supply with one of the leads supplied in the kit. Switch on the AC mains supply to the DC Supply Unit. Ensure no LED's are illuminated on the meter display. Ensure the "Normal/Boost" switch is set to "Normal", before applying power to the meter. It is recommended that the "Boost" facility is NOT used when the magnet is at field. Press the "Test" button and after 10 seconds observe the level indicated. NOTE The level displayed may have an error of up to 5% from actual.
If it is known that the magnet has a helium level of less than 95% but the unit continues to display 95%-100% either the probe is faulty or the probe is not propagating correctly. Set the "Normal/Boost" switch to "Boost" and press the "Test" button again. o If a reading is not obtained press the "Test" button one more time. If no reading is obtained then the helium level probe could be faulty.
Once the magnet helium level has been satisfactorily obtained, switch off the AC mains supply and disconnect the helium probe lead from the helium probe connector at the turret. Return all items to the transport case. Ensure that the system lead is re-connected.
Helium Probe Lead and extension Part No.10104512 AC Mains Magnet Turret Connector 'A'
Cabling Diagram
NOTE
Under normal operating conditions of the 4018 meter the volt sense lead should not be required.
o
3 Equipment Maintenance
There are no user replaceable components in this product.
Rev No 03