Академический Документы
Профессиональный Документы
Культура Документы
ReaPan B27
Manufactured by
2
ReaPan B27
1. INTENDED USE
4. REAGENT
5
tubes.
Precautions
1. Warning: The ReaPan B27 reagent contains Sodium azide.
Sodium azide is harmful if swallowed. Wear suitable protective
clothing. If swallowed, seek medical advice immediately. Contact
with acids liberates toxic gas. Azides should be flushed with
large amounts of water during disposal to avoid deposits in lead
or copper plumbing.
2. Warning: All blood specimens are considered biohazards.
Handle them as if they are capable of transmitting infection and
dispose off with proper precautions in accordance with
governmental regulations.
3. The addition of the precise volume of blood is critical to obtain
correct results. Use a calibrated pipette and operate according
to the manufacturer’s instructions.
4. The ReaPan B27 reagent and the control beads are in dried
form. It is absolutely critical to ensure vigorous vortexing after
addition of blood sample to ensure solubilization of the reagent
and the beads.
6
Storage and Handling
1. Store the reagent at room temperature in a dry place. Do not
use the reagent after the expiry date on the label.
2. Do not freeze the ReaPan B27.
3. The ReaPan B27 reagent is light sensitive. Do not expose to
direct light either during storage or when mixed with blood.
5. INSTRUMENT
The ReaPan B27 reagent has been tested on the FACScan™ and
FACSCalibur™ systems manufactured by Becton-Dickinson. ReaMetrix
recommends running this reagent on these instruments. The instrument
should be calibrated for setting photomultiplier tube voltages,
fluorescence compensation, and checking instrument sensitivity
according to the manufacturer’s guidelines.
7
6. SPECIMEN COLLECTION
Important!
7. PROCEDURE
Reagent Provided
ReaPan B27 reagent
Reagents and materials required but not provided
1. Blood collection tube
8
2. Calibrated pipettes
3. Vortex mixer
4. ReaLyse Solution
5. Sheath fluid (BD FACSFlow™ Catalog No. 340398 or
equivalent)
6. Calibrite Beads
Assay Protocol
This protocol assumes that the flow cytometer has been setup according
to the manufacturer’s instructions (For example, In the case of
FACScan™, the instrument should be setup and calibrated using
CaliBrite™ beads and FACSComp™ software). Open the CellQuest™
Pro software and connect to the cytometer.
• FL1 – 530 ± 30 nm
• FL2 – 585 ± 42 nm
• FL3 – 670 nm LP (BD FACS Calibur™) or 650 nm LP (BD
FACScan™)
Open a new document (File>New document) and select the option to
view instrument settings:
1. Draw two dot-plot (View-1 and View-2) and four histogram (View -3,
View-4, View-5, View-6) views –
• View-1 (Dot Plot): CD3 PE-Dy649 fluorescence (FL3) versus
side scatter (SSC – Linear scale).
• View-2 (Dot Plot): FL2 versus FL1 of events gated as beads
from the first view to capture the fluorescent reference beads.
The beads have a high FL1 and FL2 Fluorescence Intensity
• View-3 (Histogram): HLA-B27-PE fluorescence (FL2) versus
counts of events gated as CD3 positive (R1) from View-1.
• View-4, View-5, and View-6 (Histogram): Fluorescent bead
intensity (on FL1, FL2 & FL3) obtained from R3 of View 2
versus Counts.
• Set the number of events to capture at 4000 for R-3 gate
(View 2) before data acquisition. When 4000 beads are
acquired in the gate, acquisition stops.
11
• Select the histogram statistics option for View-3, View-4 and
View-5 & View-6
2. Adjust the Side-scatter and FL3 voltage settings on View 1 to
acquire CD3 positive (R1) and bead populations (R2) as
shown. Ensure that both the beads and the lymphocyte
populations are acquired and can be visualized as separate
from the background.(Note: Multi-colored gating can be used
to prevent unintentional thresholding out beads)
3. Adjust the FL2 voltage settings to position R3 gated beads to
600±10 channels (Histogram stats – M1 – Median
Fluorescence Intensity) on View
4. Similarly adjust FL1 and FL3 voltage gain to set the R3 gated
beads on FL1 at a MdFI of 600±10 and 400±10 channels
respectively. (The above steps are critical for the accuracy of
the test results and must not be skipped)
5. Obtain the R1-gated HLA-B27 population on View 3 to record
the fluorescence intensity on histogram stats (of all events).
6. Enter the K value for the particular batch of reagent from the
label on the pouch and calculate RMX value as shown.
12
MdFIsample – K x MdFIbeads on FL2
RMX Value* =
MdFIbeads on FL2
*K is a constant unique to each reagent lot given on the Reagent Pouch Label
If RMX Value ≤0, then sample is HLA B27 negative
If RMX Value >0, then sample is HLA B27 positive
View 1 View 2
Dot plot views: View-1: CD3-PE-Dy649 fluorescence (FL3 - log scale) versus side-scatter
(SSC – Linear scale) to capture CD3+ (R1) and fluorescence beads (R2) View-2: FL1
versus FL2 (both on log scale) and R2 gate applied to obtain a clean reference beads
population (R3)
13
View 3
14
8. LIMITATIONS
1. The ReaPan B27 reagent has only been validated with K3EDTA
treated whole blood.
2. It is the responsibility of the user to ensure that the flow cytometer
instrument is calibrated according to manufacturer’s instructions.
3. It is important that the gain settings are adjusted to position the
fluorescent beads as described in Section 6.0. The R-1 gate must be
visually checked to ensure adequate T-lymphocyte gating.
4. The test result bears a direct relation to the intensity of the
Phycoerythrin (PE) in the reagent. Exposure of reagent to direct light
or use of expired reagents may prove detrimental to the accuracy of
the test results.
5 The HLA-B27 test result is not indicative of the presence or absence
of a disease condition. HLA-B27 expression along with symptomatic
presentation and clinical history of the patient should be carefully
considered by the physician in making a diagnosis (7).
15
9. WARRANTY
10. REFERENCES
16
2. Michael J. Warzynski. HLA-B27 TESTING: MEAN CHANNELS, QC,
AND SURVEYS. Cytometry (Communications in Clinical Cytometry)
30:208–211 (1997)
5. Michael T. Seipp, Maria Erali, Rae Lynn Wies, and Carl Wittwer. HLA-
B27 Typing: Evaluation of an Allele-Specific PCR Melting Assay and
Two Flow Cytometric Antigen Assays. Cytometry Part B (Clinical
Cytometry) 63B:10–15 (2005)
17
6. W.H.B.M. Levering, H. Wind, H. Hooijkaas, K. Sintnicolaas, B.
Brando, J.W. Gratama. Flow cytometric screening for HLA-B27 on
peripheral blood lymphocytes. J Biol Regul Homeost Agents 2003;
17: 241-6
18
10. Wilfried Levering External quality assessment in flow cytometry.
Educational aspects and trends toward improvement. Doctoral
Thesis
11. Wilfried H.B.M. Levering, Rene van den Beemd, Jeroen G. te Marvelde,
Wil A.M. van Beers, Herbert Hooijkaas, Kees Sintnicolaas, and Jan W.
Gratama. External Quality Assessment of Flow Cytometric HLA-B27
Typing. Cytometry (Communications in Clinical Cytometry) 42:95–105
(2000)
19
Manufactured by ReaMetrix India Pvt. Ltd.
50-B, II Phase, Peenya Industrial Area
Peenya, Bangalore 560058, India
Ph: +91-80-28378693/5, Fax: +91-80-41172451
E-mail: info@reametrix.com
www.reametrix.com
Rev No. 2.0, 21-Apr-09
20