Why System Suitability Tests Are Not a Substitute

for Analytical Instrument Qualification or Calibration


Part I
http://www.chromatographyonline.com/lcgc/article/articleDetail.jsp?id=700826
In this two-part article series I'll be discussing the role of system suitability tests (SSTs) in the
context of analytical instrument qualification (AIQ) and the generation of quality analytical results.
I'll look at this first from the perspective of regulatory requirements based upon the United States
Pharmacopeia (USP) general chapter 1058 (1058) on AIQ (1). As much of this discussion is also
good analytical science I will also look at this from the perspective of ISO 17025 (2), the
interpretation of the ISO 9000 quality management system for testing and calibration
laboratories. I'll use the ISO 17025 term of calibration to be equivalent to AIQ and discuss the
reason why I have taken this approach.
Specifically, I want to discuss the question why SSTs are not a substitute for
AIQ for the initial qualification or requalification of a chromatograph (that is, a
periodic check or operational qualification) under GMP or calibration under ISO
17025. The reason for this is that there is a common misconception in some
laboratories that SSTs can be used to qualify a chromatograph. This is wrong.
Using SSTs as the sole instrument qualification approach will leave any
laboratory exposed to regulatory action or nonconformance as the
chromatographs cannot be demonstrated as being fit for their intended
purposes. Therefore, we will explore the reasons for this from both the
regulatory and quality perspectives.
What Is a System Suitability Test?
An SST is simply a check to see if the chromatographic system (instrument
modules, column, and mobile phase) is capable of undertaking the analysis you
require. It is performed immediately before the samples you have prepared in
the laboratory are committed for analysis. At it's simplest, an SST is the single injection of a
standard solution of reference compounds to see if the separation is "OK." The only criterion
used for assessing if the system is acceptable is a visual check by the chromatographer to
assess if the separation is within expected boundaries. However, there are no measurements
made: Just a visual assessment that things are right.
At the other end of the scale are the formal SSTs described in the pharmacopeias used in
pharmaceutical QC laboratories. Here a minimum of five replicate injections of a standard are
injected and the calculated peak areas and other chromatographic criteria (retention time
window, peak shape, peak tailing, resolution between peaks, and so forth) are objectively
compared with predefined specifications defined by the pharmacopeia (peak area repeatability)
and the laboratory (all other chromatographic criteria). If the SST injections meet all the
predefined criteria then the chromatographic system is deemed acceptable and the samples
committed for analysis

Table I: Selected
U.S. good
manufacturing
practice (GMP)
regulations for
instruments and
equipment
GMP and
Pharmacopeial Regulations
Good manufacturing practice (GMP) regulations from either the United
States or Europe state simply that equipment used in the manufacture of
drug products should be fit for their intended use as can be seen in Table I,
which presents the U.S. regulations (3). Also, shown in Table I are the
requirements for ensuring that scientifically sound specifications for the
qualification work are written and that qualification and calibration of
instruments is documented correctly. Note also that if an instrument fails to
meet its specification it must be taken out of service until the problem is
rectified and retested to show that it meets its specification.
However, for chromatographic equipment these GMP regulations are
expanded by either the USP <621> (4) or the European
Pharmacopoeia (EP) chapter 2.2.46 (5), which have both specified
requirements for SST for chromatographic analysis to demonstrate that a chromatograph is fit for
the analysis it will undertake on the day of analysis. SSTs must be run each time before an
analysis is undertaken and each SST is specific for an individual method with predefined
acceptance criteria for example, precision, peak shape, and resolution from other analytes. If
an SST fails, then the samples cannot be assayed.
The principle of a point-of-use check, such as an SST, is applicable to any analytical instrument
or system and this is performed just before an analysis to demonstrate correct performance, for
example, a check to see if the chromatographic system or an analytical balance works before
committing samples. This point of use check is used in other types of instrumental analysis for
example, balance and pH meter checks so it is not an unusual approach and is good
analytical science.
I presented the main elements of USP <1058> on analytical instrument qualification (AIQ) in an
earlier "Questions of Quality" column (6) that looked at the impact of this on the chromatography
laboratory. This informational general chapter consists of six major sections:
Components of data quality.
Analytical instrument qualification process.
Roles and responsibilities.
Software validation.
Change control.
Instrument categories.
The key sections from USP <1058> that concern our debate of
AIQ versus SST are the components of data quality and the
analytical instrument qualification process. The quotations
fromUSP <1058> mentioned later in this column USP <1058>
come from these two sections. I will focus my comments on the
first section of USP<1058> that discusses the components of
analytical data quality. In my earlier column (6) I skipped over
this section but I would like to go into further detail, as the
content is good analytical science and is applicable to all laboratorie

Table lI: Selection of the
main ISO 17025 sections
covering instrument
qualification

Figure 1: The Data Quality Triangle modified
from USP <1058>.



ISO 17025 Accreditation
OK, so what is ISO 170252 and what is accreditation? First question first, ISO 17025 is an
international standard that both testing and calibration laboratories can work against; for the
purpose of this column I will focus on testing laboratories mainly. A major difference between
ISO 17025 and GMP is that compliance with the requirements of the ISO standard is voluntary
rather than imposed by government agencies. However, the advantage of working to ISO 17025
is that a laboratory is working against an internationally recognized standard. To answer the
second question, accreditation is the process of ensuring that the laboratory has implemented
and complies with the requirements of the ISO 17025 standard and continues to do so. This is
achieved by an independent audit by technical officers from the local accreditation body coupled
with external experts in the specific areas offered by a laboratory. Each country in the European
Union has an accreditation body (A2LA for the United States and UKAS for the United Kingdom)
that is responsible for the accreditation of laboratories in their respective country. However, ISO
17025 goes further than GMP in that the test methods used by a laboratory and the performance
of each one are also covered within the scope of a laboratory's accreditation and are included in
the assessments made by accreditation bodies.
ISO 17025 Requirements for Equipment
When considering ISO 17025 we have to remember that this standard is a specific interpretation
of ISO 9000 for testing and calibration laboratories. In this column I will focus mainly on the
requirements for testing laboratories that use chromatographic equipment to perform their work.
The applicable sections of ISO 17025 are presented in Table II and are discussed and
interpreted in more detail below. Please note that key portions of the standard, as were the U.S.
GMP regulations, are highlighted for the purposes of this column and the full sections should be
interpreted in practice.
The first requirement for equipment is in section 5.5.2 that notes that all laboratories are required
to implement and operate a program for the maintenance and calibration of laboratory
equipment. These calibration checks are only required where the equipment setting can
significantly affect the test or analytical result (for example, oven temperature, gas or liquid flow,
detector wavelength accuracy, and photometric linearity for GC or LC modules). Proper
equipment use combined with periodic servicing and calibration does not ensure that an
instrument, in this case a chromatograph, is performing adequately; there also has to be point of
use checks of overall system performance that can be built into test methods (for example,
chromatographic SSTs). These checks should be documented and should be satisfactorily
completed before the equipment is used or before results are accepted.
Reconciling GMP Regulations and ISO 17025 Requirements
How can we have a common understanding of both the GMP and ISO 17025 requirements so
that we can have a debate that is independent of applicable regulations or requirements? This all
comes down to definitions and what we mean by specific words. So here are the key terms used
in either regulations or standards and let's see how we can reconcile them to help us in the
debate:
Equipment or Instrument? In GMP the term "equipment" is used to cover anything used in
the manufacture of pharmaceutical products that includes the laboratory. Similarly, ISO 17025
uses the term equipment to cover the wide range of equipment, instruments and systems used
in the testing laboratory. This is in contrast to USP <1058>, which uses the term instrument.
However, in practice the two terms are synonymous as we can refer to chromatographic
instruments or chromatographic equipment. I prefer the term instrument and will use that in the
remainder of this article.
Calibration or Qualification? Both GMP and ISO 17025 refer to calibration in contrast to
qualification used by USP <1058>. Are the two terms equivalent? Yes, in my view. Calibration is
a simpler term to understand and has been defined by EU GMP as the set of operations that
establish, under specified conditions, the relationship between values indicated by a measuring
instrument or measuring system, or values represented by a material measure and the
corresponding known values of a reference standard (7). I would argue that the work undertaken
in qualification and calibration of an instrument is essentially the same: Does it work for your
purpose and can you demonstrate that by using calibrated test instruments and traceable
reference standards?
Maintenance? All regulations and guidance refer to these terms so there is full agreement and
thus no need for interpretation

US FDA Warning Letter Citation
An example of how a laboratory can fall foul of a regulator is found in the U.S. Food and Drug
Administration (FDA) warning letters publicly available on the Agency website
( http://www.fda.gov/). Shaanxi Hanjiang Pharmaceutical Group, a Chinese company, received a
warning letter in January 2008 following an inspection of its facilities in September 2007 (8).
Quoting from the warning letter we can see that:
3. Calibration of measuring devices was not traceable to the national standards.
Several balances and thermometers were not traceable to national standards. We acknowledge
your firm's submission of a certificate of metrology <redacted> which expires on 16th April 2008.
However, this certificate is deficient in that it does not include identification of the instrument for
which it is meant for and does not include a statement of traceability.
So we can see the requirement for traceability outlined in ISO 17025 can be mirrored in GMP.
The only problem is that there is no explicitly written requirement for this as you can see in Table
I. However, as you can see in 211.160(b) in Table I there is a need for the laboratory controls to
be "scientifically sound," as traceability of calibration is good science. Therefore, if you have not
calibrated or qualified your instruments with appropriate and traceable reference standards then
the data you produce from those instruments will be poor.
Conclusions
In this article, I've looked at the regulations and quality standards for qualification and calibration
of chromatographic instruments. In part II we'll look at how the data quality triangle can be
applied to this from a practical perspective.
R.D. McDowall is principal at McDowall Consulting, Bromley, Kent, UK. He is also a member
ofLCGC's Editorial Advisory Board.
References
(1) United States Pharmacopeia general chapter <1058> Analytical Instrument Qualification.
(2) ISO 17025 General requirements for the competence of testing and calibration laboratories,
International Standards Organisation, 2005.
(3) Good Manufacturing Practice for Finished Pharmaceuticals (21 CFR 211), 2008.
(4) United States Pharmacopeia general chapter <621> Chromatography.
(5) EP Chromatography chapter 2.2.46 Chromatography.
(6) R.D. McDowall, LCGC Europe 22(4), 204210 (2009).
(7) EU Good Manufacturing Practice Glossary.
(8) Shaanxi Hanjiang Pharmaceutical Group, FDA Warning Letter, January 2008
(http://www.fda.gov/)