AR #79
PCB Piezotronics
Depew NY 14043
January, 2003
With the advent of high-resolution digital acquisition and the Three steps must be completed before a facility can
evolution of sophisticated test strategies, test engineers are credibly claim proficiency to perform primary calibration of
demanding higher accuracy from their measuring accelerometers. First, the equipment and technical staff
equipment. For vibration applications, arguably the most must in place. In terms of equipment, the vibration exciters
important links in the measurement chain are the choice of are of particular importance, especially at high frequency
accelerometer and the calibration of its sensitivity. where maintaining uniform piston-like motion free of
distortion can be a challenge. Because of the exciter’s
Accelerometer calibration is classified as either primary or importance, a considerable investment in shaker technology
secondary. Primary calibration can provide the highest has been made by PCB Piezotronics. Three shakers are
measurement accuracy, but is more costly than secondary now used in the calibration facility: an air-bearing long
calibration and is thus not performed on routine or stroke shaker (6 inch peak-peak) for low frequency
production basis. In a secondary, or comparison excitation, an air bearing shaker with beryllium armature for
calibration, the unit under test (UUT) is mounted to a back- calibration to high frequency (20 kHz), and a specially
1
1
T E C H N I C A L I N F O R M AT I O N AR_79
designed air bearing shaker with alumina armature for Over the past 10 years quadrature detection strategies
calibration of large-size working standard accelerometers to have gained widespread use and acceptance for
high frequency. Second, the facility must demonstrate accelerometer calibration. This can be attributed, in part, to
proficiency through comparison with a reputable NMI. In the availability of low-cost high-speed digital signal
PCB’s case, a transfer standard has been successfully acquisition.
calibrated at both PTB (Physikalisch-Technische
Bundesanstalt, Germany) and NIST (United States). Both homodyne and heterodyne quadrature detection
Finally, the organization’s technical competency must be strategies have been used successfully in calibration. The
audited by an unbiased organization. In PCB’s case, the interferometer described in this paper generates the
quadrature pair using a heterodyne laser in conjunction with
primary calibration facility has been accredited to ISO/IEC
17025-1999 by the American Association of Laboratory a quadrature phase demodulator. Heterodyne lasers have
Accreditation (A2LA). an advantage over homodyne lasers in that a single
photodetector is required. This results in improved linearity
CALIBRATION BY LASER INTERFEROMETER of the quadrature pair, enabling more accurate
measurement at high frequency where the armature
In primary calibration by laser interferometery, reference displacement may be smaller than a fringe. Another
motion is determined through a direct measure of the base advantage of the heterodyne approach is that the non-
quantities length and time. The UUT is mounted to the linearity in quadrature pair and the influence of this non-
armature of an exciter that provides linear sinusoidal motion
linearity on overall measurement uncertainty is easily
at a stable and known frequency f. UUT voltage, u, is
characterized. An approach for quantifying this influence is
measured with a precision voltmeter and armature described in this paper.
displacement, x, is measured with a HeNe laser
interferometer. UUT sensitivity is calculated as: Laser interferometer technology is well established and its
influence on the overall calibration uncertainty is generally
u small. At high frequency, calibration error is dominated by
S= ⋅ 9.80665 [Volts/g] (1) the influence of non-uniform exciter motion. The exciter
(2πf ) 2 x must produce linear piston-like sinusoidal motion free of
There are number of suitable schemes for measuring distortion and cross axis motion components [4]. This can
armature motion using laser interferometers. A good be a challenge to achieve over wide frequency range and
overview of the methods can be found in ref [1]. The with a variety of UUTs that may range from a few grams to
underlying foundation for all the methods is that HeNe laser hundreds of grams in weight. Because of the exciter’s
is stable and has an accepted wavelength equal to a importance, a considerable investment in improved shaker
constant value of 632.81 nm at standard laboratory technology has been made and three shakers are now used
temperatures and pressures. The ISO standard in PCB’s calibration facility: an air-bearing long stroke
organization describes three primary calibration schemes shaker (6 inch peak-peak) for low frequency excitation, an
known as Methods 1, 2, and 3 [2]. Methods 1 and 2 utilize air bearing shaker with beryllium armature for calibration to
a Michelson interferometer with a single photodetector. high frequency (20 kHz), and a specially designed air
Method 3 utilizes an interferometer with a pair of bearing shaker with alumina armature for calibration of
photodetector signals in quadrature. large-size working standard accelerometers to high
frequency.
Of the three methods, Method 1 (the fringe counting
method) is simplest in concept and practice. For each INTERFEROMETER THEORY
interval of armature displacement equal to 1/2 a HeNe The basis for the laser interferometer is the constructive
wavelength, an interferometer fringe will be produced. A and destructive interference of two coherent light waves,
fringe is detected by a photodector producing a signal that one being a reference beam with a fixed path and the
changes linearly from low to high. A precision frequency second beam reflected from the moving surface to be
counter counts the number of fringes per excitation period, measured. Consider the reference and reflected waves in
providing a measure of armature peak-to-peak
Fig 1 displaced by a distance φ m . In a single pass
displacement. Quantization of the displacement limits the
usefulness of fringe counting to low frequency (less than interferometer this is related to target displacement x(t)
1000 Hz) where the armature displacement is large relative through:
to the HeNe wavelength [2]. A second limitation of the
fringe counting method is phase of the UUT cannot be φ m = 2 x (t ) (2)
determined.
Or in terms of phase between the reference and reflected
The problem of fringe quantiziation can be overcome by waves:
using a quadrature detection scheme, such as Method 3,
enabling interpolation of displacement between fringes and 2π 4π
resolution equal to a small fraction of wavelength of light. In
ϕ= φm = x(t ) (3)
λ λ
quadrature detection, two photodetector signals time-shifted
by a ¼ wavelength are acquired via a PC-based digital The interferometer’s photodetector responds to intensity of
acquisition board. The displacement is reconstructed by the interfering light. The following cosine relationship
computing the unwrapped inverse tangent of the quadrature between target displacement and light intensity can be
pair. The method has been proven accurate to high derived:
frequency (10 kHz [2]) and provides phase calibration.
2
2
T E C H N I C A L I N F O R M AT I O N AR_79
3
3
T E C H N I C A L I N F O R M AT I O N AR_79
CALIBRATION EQUIPMENT this writing PTB has the capability of measuring sensitivity
to the lowest uncertainty of any NMI. The base-reference
A schematic of the accelerometer calibration equipment is PCB and NIST data is compared to the PTB data in Fig 11.
shown in Fig 7. The block labeled “laser” contains the Over the frequency range from 5 Hz to 15 kHz, the PCB
HeNe laser, optics, Bragg cell, and photodectors. The laser and PTB sensitivities are found to agree to within the PTB
is mounted on a granite table for isolation from stated uncertainty. The PCB top-reference data is
environmental vibration. An air bearing shaker provides compared to the PTB data in Fig 12. Over the frequency
uniform distortion-free sinusoidal excitation of the range from 5 Hz to 15 kHz the sensitivities are found to
accelerometer. A PC-based DAQ acquires four signals: agree to within the PTB stated uncertainty.
accelerometer output, the quadrature photodetector pair I
and Q, and an analog velocity signal. Table 2: X353M295 Transfer Standard
Nominal Sensitivity 10 mV/g
Calibration of sensitivity is performed as follows. The
Mounting 10-32 female
accelerometer is excited sinusoidally at the desired
frequency and amplitude level. The vibrometer, I and Q Sensing element Inverted quartz shear
pair, and the accelerometer voltage are acquired. Power ICP® constant current
Displacement is determined from the I and Q photodetector Dynamic range ±500 g
pair using Eq 8. Accelerometer sensitivity is calculated Weight 9 grams
using Eq 1. Case titanium hermetic
Connector 10-32 Microdot
A number of safegaurds are built into the calibration Mounting torque 20 inch-lb
procedure to ensure the reasonableness of the calibration
Frequency response (±10%)
results. For example, in addition to the I and Q 0.7 Hz to 11 kHz
Base Reference
computation, armature acceleration is determined via an 0.7 Hz to 20 kHz
Top Reference
analog vibrometer. This is accomplished through the
measure of the Doppler shift of the heterodyned
photodetector signal using commercially available analog Table 3: X353M295 Base-Reference Calibration, 100 and
hardware (Polytec PI). This hardware is labeled 160 Hz. Stated uncertainty in p arentheses.
“vibrometer controller” In Fig 6. Because of the analog Frequency NIST PTB PCB
electronics, the velocity signal is not a primary (Hz)
measurement, and thus is not used in calibration of Sensitivity Sensitivity Sensitivity
accelerometer sensitivity. However, the vibrometer output (mV/g) (mV/g) (mV/g)
provides a “sanity check” of sensitivity and is useful for 100 10.22 10.26 10.25
quick analysis of an accelerometer’s frequency response. (±1.0%) (±0.2%) (±0.2%)
Sensitivity computed from the vibrometer and the I and Q
160 10.19 10.26 10.26
pair generally agree to within 0.2%.
(±1.0%) (±0.2%) (±0.2%)
PROFICIENCY TESTING
A specially designed transfer standard, Model X353M295, CONCLUSION
was tested at PCB’s primary calibration facility, PTB in
Germany, and NIST in the United States. Two PCB Piezotronics’ new accelerometer calibration facility,
proficiencies were demonstrated: top-reference utilizing a heterodyne laser interferometer, has been
measurement for calibration of “back-to-back “ working accredited by A2LA. A transfer standard was tested at the
standards and base-reference measurement for calibration new facility and at national metrology laboratories in
of transfer standards (Fig 8). In many ways the base- Germany (PTB) and the United States (NIST). Calibration
reference calibration is more difficult because shaker data at all three facilities were found to agree to within the
vibration modes, such as armature rocking and umbrella stated uncertainties.
modes, have a greater influence on measurement error. REFERENCES
The X353M295 utilizes an inverted element allowing both
top-reference and base-reference measurements. [1] von Martens, H.J., "Current state and trends of ensuring
traceability for vibration and shock measurements," Metrologia, Vol.
Accelerometer specifications are provided in Table 2. 36, pp 357-373, 1999.
Base-reference calibration was performed at NIST at the [2] ISO Standard ISO 16063-11:1999, Methods for the calibration of
reference frequencies of 100 Hz and 160 Hz (Table 3). vibration and shock transducers-Part 11: Primary vibration
NIST states a measurement uncertainty of 1% at these calibration by laser interferometery.
frequencies. The PCB and NIST calibrations were found to
[3] Link, A., H-J von Martens, and W. Wabinski, “New method for
agree to within the NIST stated uncertainty.
absolute shock calibration of accelerometers,” Proc. of SPIE, Third
PTB performed base-reference and top-reference International Conference on Vibration Measurements by Laser
Techniques,” pp. 224-235, 1998.
calibrations of the transfer standard over the frequency
range 5 Hz to 15 kHz. The PTB, PCB, and NIST base- [4] Payne, B., “Laser Interferometer and Reciprocity Calibration of
reference sensitivities are plotted as a function of frequency Accelerometers Using the NIST Super Shaker,” Proc. of SPIE,
in Fig 9 (NIST data is only at 100 Hz and 160 Hz). The Third International Conference on Vibration Measurements by
PCB and PTB top-reference calibrations are plotted in Fig Laser Techniques,” pp. 187-194, 1998.
10. The PTB measurement uncertainty at the 100 Hz and ICP® is a registered trademark of PCB Piezotronics
160 Hz reference frequencies is only 0.2%. At the time of
4
4
T E C H N I C A L I N F O R M AT I O N AR_79
FIGURES
REFERENCE HeNe
WAVE
φm
TARGET
x(t )
HeNe Laser
REFLECTED HeNe
WAVE x(t)
Bragg Cell
u Bragg = sin 2ω c t
Photo
detector
INTERFEROMETER
Reference INTERFEROMETER
x (t ) Mirror uHeNe
uc = sin ω c t
Mixer
HeNe Laser Splitter Filter I (ti ) = sin ϕ (ti )
RF LP ADC
TARGET
Phase
Filter
Q(ti ) = cos ϕ (ti )
LO Shift 0°
LP ADC
90°
Photodetectors
DEMODULATOR ACQUISITION
I ( ti ) = sin ϕ (t i ) Q( ti ) = cos ϕ (t i )
5
T E C H N I C A L I N F O R M AT I O N AR_79
I vs. Q
0.08
0.15
0.06 Phase Measured Harmonic fit
0.1
0.04
0.05 0.02
Q (V olts )
0 0.00
-0.05 -0.02
-0.04
-0.1
-0.06
-0.15
-0.08
LASER
0.04
0.02
INTERFEROMETER
I
Q (V o lts)
0 I/Q
DEMODULATOR Q
-0.02
RF
LO
-0.04
-0.06 VIBROMETER
CONTROLLER
-0.08 -0.06 -0.04 -0.02 0 0.02 0.04 0.06 0.08 UUT
I (Volts)
Acceleration Velocity
RF = -10 dBm
Fig 4: Typical I vs Q Lissajous figure for two levels of RF. DAQ
SHAKER
COMPUTER
6 ARMATURE
5
Fig 7: Equipment for heterodyne accelerometer calibration.
Measured Phase (radian)
4
LASER
3
LASER
2
UUT
1
ARMATURE
0
0 2 4 6
True Phase (Radian)
TOP- BASE-
REFERENCE REFERENCE
6
6
T E C H N I C A L I N F O R M AT I O N AR_79
12.5
11.5
11.0
10.5
10.0
1 10 100 1000 10000 100000
Frequency (Hz)
10.7
10.6
PTB Calibration
PCB Calibration
10.5
Sensitivity (mV/g)
10.4
10.3
10.2
10.1
1 10 100 1000 10000 100000
Frequency (Hz)
7
T E C H N I C A L I N F O R M AT I O N AR_79
2.0
1.5
PCB Calibration
1.0
NIST Calibration
Deviation From PTB (%)
0.5
0.0
-0.5
-1.0
-1.5
-2.0
1 10 100 1000 10000 100000
Frequency (Hz)
Fig 11: Base-reference deviation from PTB calibration data, Model X353M295 SN 002. Heavy lines indicate PTB stated
uncertainty.
2.0
1.5
PCB Calibration
1.0
Deviation From PTB (%)
0.5
0.0
-0.5
-1.0
-1.5
-2.0
1 10 100 1000 10000 100000
Frequency (Hz)
Fig 12: Top-reference deviation from PTB calibration data, Model X353M295 SN 002. Heavy lines indicate PTB stated
uncertainty.
8