Вы находитесь на странице: 1из 136

NUFLO

TM
MC-III EXP
Flow Analyzer
User Manual
Manual No. 9A-50165003, Rev. 04

Important Safety Information
Symbols and Terms Used in this Manual
!
WARNING: This symbol identifes information about practices or circumstances that can lead to per-
sonal injury or death, property damage, or economic loss.
CAUTION: Indicates actions or procedures which if not performed correctly may lead to personal
injury or incorrect function of the instrument or connected equipment.
Important: Indicates actions or procedures which may affect instrument operation or may lead to an
instrument response which is not planned.
Symbols Marked on Equipment
Attention! Refer to manual

Protective (earth) ground


Technical Support Contact Information
Cameron
Measurement Systems Division
14450 John F. Kennedy Blvd.
Houston, TX 77032
Phone: 1-800-654-3760; 281-582-9500
Fax: 281-582-9599
NuFlo and MC-III are trademarks of Cameron.
Modbus is a registered trademark of the Modbus Organization, Inc.
Windows is a registered trademark of Microsoft Corporation.
Acrobat Reader is a registered trademark of Adobe Systems Incorporated.
2013 Cameron International Corporation (Cameron). All information contained in this publication is
confdential and proprietary property of Cameron. Any reproduction or use of these instructions, drawings, or
photographs without the express written permission of an offcer of Cameron is forbidden.
All Rights Reserved.
Printed in the United States of America.
Manual No. 9A-50165003, Rev. 04
August 2013
i
MC-III EXP Flow Analyzer Table of Contents
Contents
Section 1Introduction ................................................................................................................................... 1
Operation ............................................................................................................................................................ 1
Product Identifcation .......................................................................................................................................... 2
Key Product Features ........................................................................................................................................ 5
LCD Display ................................................................................................................................................. 5
Keypad ......................................................................................................................................................... 6
Interface Software ........................................................................................................................................ 6
Power SupplyCSA-Approved Units .......................................................................................................... 7
Power SupplyATEX-Approved Units ........................................................................................................ 7
Multipoint Linearization ................................................................................................................................ 8
Gas Volume Correction ................................................................................................................................ 8
Input Options ................................................................................................................................................ 8
Output Options ............................................................................................................................................. 8
Flow Log Archival ........................................................................................................................................ 8
Event Log Archival ...................................................................................................................................... 8
Password-Protected Security ....................................................................................................................... 8
Optional Features ............................................................................................................................................... 9
Explosion-Proof Reset/Control Switch ......................................................................................................... 9
RS-485 Communications Adapter ................................................................................................................ 9
USB Communications Adapter (CSA-Approved Units Only) ....................................................................... 9
RS-485 IS Barrier (CSA-Approved Units Only) .......................................................................................... 10
Commonly Used Functions .............................................................................................................................. 10
Reading Totals ........................................................................................................................................... 10
Saving Totals to Memory ............................................................................................................................ 10
Resetting the Total ..................................................................................................................................... 10
Viewing Daily and Hourly Logs ...................................................................................................................11
Saving and Uploading Confguration Files ..................................................................................................11
Exporting Log Data .....................................................................................................................................11
Saving Log Data in a Report .......................................................................................................................11
Section 2Installation ................................................................................................................................... 13
Hazardous Area Installations (ATEX) ............................................................................................................... 13
Conditions for Safe Use (ATEX)................................................................................................................. 14
Mounting Options (ATEX) .......................................................................................................................... 14
Hazardous Area Installations (CSA) ................................................................................................................. 16
Conditions for Safe Use (CSA) .................................................................................................................. 16
Mounting Options (CSA) ............................................................................................................................ 17
Installing the MC-III EXP .................................................................................................................................. 19
Field Wiring....................................................................................................................................................... 20
Internal Power Supply ................................................................................................................................ 21
External Power Supply ............................................................................................................................... 21
Turbine Flowmeter (TFM) Input ................................................................................................................. 22
Pulse Input ................................................................................................................................................. 22
Remote Reset Input ................................................................................................................................... 23
Pulse Output ............................................................................................................................................. 24
Analog (4-20 mA) Rate Output ................................................................................................................. 25
Flowmeter Frequency Output .................................................................................................................... 26
RS-485 Output ........................................................................................................................................... 27
ii
Table of Contents MC-III EXP Flow Analyzer
Section 3Confguration and Operation via Keypad ................................................................................. 29
Entering a Calibration Factor ........................................................................................................................... 30
Entering a Calculated Divisor ........................................................................................................................... 31
Setting Input Type and Sensitivity .................................................................................................................... 32
Confguring the Total Display ............................................................................................................................ 33
Confguring the Rate Display ............................................................................................................................ 34
Confguring the 4-20 mA Rate Output .............................................................................................................. 35
Confguring the Pulse Output ........................................................................................................................... 37
Entering the Slave Address .............................................................................................................................. 38
Entering the Baud Rate .................................................................................................................................... 38
Section 4Confguration and Operation via Software ............................................................................... 39
Installing the Software ...................................................................................................................................... 39
User Documentation .................................................................................................................................. 39
Connecting the MC-III to a Computer ............................................................................................................... 40
Running the Software ....................................................................................................................................... 40
Automating Functions on Software Startup ............................................................................................... 41
Changing Autorun Settings ........................................................................................................................ 42
Express Connect Option ............................................................................................................................ 42
Changing the Communications Port .......................................................................................................... 43
Software Connection in Multi-Device Network ........................................................................................... 43
Setting Log Download Preferences .................................................................................................................. 45
Confguring the MC-III EXP .............................................................................................................................. 46
Confguration Wizard ................................................................................................................................. 48
MC-III Main Screen .................................................................................................................................... 50
Buttons and Tools....................................................................................................................................... 51
System Setup ................................................................................................................................................... 54
Time/Date Synchronization ....................................................................................................................... 54
Contract Hour ............................................................................................................................................. 54
LCD Contrast Adjustment .......................................................................................................................... 54
Security Setup ............................................................................................................................................ 55
Firmware Version Number ......................................................................................................................... 55
Serial Number ............................................................................................................................................ 55
Communications Port ....................................................................................................................................... 55
Slave Address ............................................................................................................................................ 55
Baud Rate ................................................................................................................................................. 56
Bus Delay ................................................................................................................................................... 57
Bus Timeout ............................................................................................................................................... 57
Software Communication Options ............................................................................................................. 57
Wellsite Information .......................................................................................................................................... 57
Turbine Input .................................................................................................................................................... 58
Volume Display .......................................................................................................................................... 58
Rate Display ............................................................................................................................................... 59
Input Type and Sensitivity Confguration .................................................................................................... 59
Cut-Off Thresholds ..................................................................................................................................... 59
Calculation Period ...................................................................................................................................... 59
K-Factor Entry .................................................................................................................................................. 60
K-Factor Units ............................................................................................................................................ 60
K-Factor Type............................................................................................................................................. 60
K-Factor Backup ........................................................................................................................................ 61
Gas Volume Correction (Supercompressibility Calculation) ...................................................................... 62
4-20 mA Output ................................................................................................................................................ 64
Enabling 4-20 mA Output .......................................................................................................................... 65
4-20 mA Output Testing.............................................................................................................................. 65
iii
MC-III EXP Flow Analyzer Table of Contents
Pulse Output ..................................................................................................................................................... 67
Confguring Pulse Output ........................................................................................................................... 67
Pulse Output Testing .................................................................................................................................. 68
Saving and Uploading Confguration Files ....................................................................................................... 69
Saving a Confguration File ........................................................................................................................ 69
Uploading a Confguration File ................................................................................................................... 70
Advanced Access ............................................................................................................................................ 72
Section 5Flow Logs and Event Logs ........................................................................................................ 73
Auto-Save Log Formats.................................................................................................................................... 73
Log Directory and Filenames............................................................................................................................ 74
Flow Archive ..................................................................................................................................................... 74
Downloading Flow Logs ............................................................................................................................. 75
Viewing Trend Charts ................................................................................................................................. 76
Printing/Saving a Report ............................................................................................................................ 77
Viewing a Saved Report............................................................................................................................. 79
Exporting Flow Logs ................................................................................................................................. 80
Event Archive.................................................................................................................................................... 81
Downloading Event Logs ........................................................................................................................... 82
Printing/Saving a Report ............................................................................................................................ 83
Exporting Event Logs ................................................................................................................................ 83
Section 6 - MC-III EXP Maintenance .............................................................................................................. 85
Lithium Battery Replacement ........................................................................................................................... 85
Alkaline Battery Replacement .......................................................................................................................... 87
Circuit Assembly Replacement ......................................................................................................................... 89
Keypad Replacement ....................................................................................................................................... 91
Firmware Update .............................................................................................................................................. 92
MC-III Parts List (CSA) ..................................................................................................................................... 93
MC-III Parts List (ATEX) ................................................................................................................................... 94
Appendix AMC-III EXP Hardware Options ...............................................................................................A-1
Explosion-Proof Reset/Control Switch.............................................................................................................A-1
Installation .................................................................................................................................................A-1
Operation ..................................................................................................................................................A-2
Explosion-Proof Communications Adapter ......................................................................................................A-3
Assembly of the Plug Connector ...............................................................................................................A-3
Connecting to the Communications Adapter .............................................................................................A-4
Disconnecting the Communications Adapter ...........................................................................................A-5
Installation of Adapter (when purchased separately from an MC-III EXP) ................................................A-5
USB Communications Adapter ........................................................................................................................A-6
User-Supplied Hardware ...........................................................................................................................A-7
Using the Adapter .....................................................................................................................................A-7
Covering the Adapter ................................................................................................................................A-7
Adapter Kit Installation ..............................................................................................................................A-7
Appendix BSoftware Program Options ...................................................................................................B-1
General Options ..............................................................................................................................................B-1
Autorun Options...............................................................................................................................................B-2
Communications Options ................................................................................................................................B-3
Express Connect Option ...........................................................................................................................B-4
Auto-Negotiate Option ..............................................................................................................................B-4
Enable Auto-Negotiated Baud Rate Option ..............................................................................................B-4
iv
Table of Contents MC-III EXP Flow Analyzer
Toggle RTS Line Option ............................................................................................................................B-4
Enable Modbus Address Support for Firmware Versions 1.06 through 1.08 ............................................B-4
Timing Parameters ....................................................................................................................................B-4
Downloading Options ......................................................................................................................................B-5
Advanced Options ...........................................................................................................................................B-6
Clear EEPROM .........................................................................................................................................B-6
Automatic Data Logging ............................................................................................................................B-6
Appendix CLithium Battery Information ..................................................................................................C-1
Lithium Battery Disposal .................................................................................................................................C-1
Transportation Information ..............................................................................................................................C-1
Material Safety Data Sheet..............................................................................................................................C-1
Appendix DCommunications Protocol ....................................................................................................D-1
Introduction ......................................................................................................................................................D-1
Supported Commands.....................................................................................................................................D-1
Data Types ......................................................................................................................................................D-1
Registers ........................................................................................................................................................D-2
System Confguration ................................................................................................................................D-3
Product Code (register 1000) ....................................................................................................................D-4
Firmware/Register Table Version Numbers (registers 1001, 1002) ..........................................................D-4
Manufacture Date/Sales Date (registers 1003, 1004) ...............................................................................D-4
Slave Address (register 1009) ...................................................................................................................D-4
Baud Rate (register 1010) .........................................................................................................................D-4
Real Time ..................................................................................................................................................D-5
Input Confguration ....................................................................................................................................D-5
Output Confguration .................................................................................................................................D-8
Holding Registers (16-bit Mode) ...............................................................................................................D-9
Holding Registers (16-bit Mode)Least Signifcant Word First .............................................................. D-11
Base Units/Confgured Units ................................................................................................................... D-11
Conversion Factors ................................................................................................................................. D-11
Polling Registers .....................................................................................................................................D-12
Pointer/Daily/Event Pointer (registers 17001 through 17006) .................................................................D-12
Real Date (registers 17007, 17008) ........................................................................................................D-12
Real Time (registers 17009, 17010) ........................................................................................................D-12
Totals ......................................................................................................................................................D-12
Pulse Output Pulses (register 17077) .....................................................................................................D-12
Analog Output Current (register 17079) ..................................................................................................D-12
Calculated K-Factor (register 17081) ......................................................................................................D-12
Control Register ......................................................................................................................................D-15
Wellsite Parameters ................................................................................................................................D-15
Log Data ........................................................................................................................................................D-15
Enron Registers ......................................................................................................................................D-16
Enron Hourly/Daily Record Format .........................................................................................................D-16
Enron Event Record Format ...................................................................................................................D-17
Reset Status ...........................................................................................................................................D-17
Log Capacity ...........................................................................................................................................D-17
1
MC-III EXP Flow Analyzer Section 1
Section 1Introduction
The NuFlo MC-III EXP Flow Analyzer (Figure 1.1) packs a full spectrum of gas and liquid measurement
functionality, high-speed performance, and log archive and retrieval capabilities in an easy-to-use, explosion-
proof totalizer. Commonly used operations can be accessed from the six-button keypad on the front of the
instrument or from the dynamic interface software, allowing you to calibrate and confgure the unit quickly
and easily.
When the MC-III EXP is installed in accordance with the provisions in this manual (Section 2, page 13), it is
approved by CSA for use in hazardous areas in Canada and the United States, and meets the ATEX/IECEx
standards for use in hazardous areas worldwide.
Operation
The MC-III EXP calculates and displays instantaneous fow rates and accumulated totals based on a turbine
fowmeter input signal. The MC-IIIs microprocessor circuitry counts the pulses generated by a companion
fowmeter, converts that data into volume and rate values in accordance with calibration settings, and displays
the totalized data on a two-line liquid crystal display (LCD). The eight-digit top readout indicates total fow
volume; the six-digit bottom readout indicates fow rate.
Figure 1.1MC-III EXP fow analyzer (shown with CSA-approved pipe reducer/union, left, and the ATEX/
IECEx-approved tube adapter, right)
2
Section 1 MC-III EXP Flow Analyzer
With the press of a single key, totals are saved to nonvolatile memory, minimizing the risk of data loss even if
a power outage occurs.
Up to 384 daily logs, 768 hourly logs, and 345 event logs can be archived and accessed quickly on demand.
Offering a variety of user-confgurable display options, input and output options, RS-485 Modbus


compatibility, fow logging, and turbine fowmeter linearization, the MC-III EXP is one of the most versatile
totalizers on the market. For specifcations, see Table 1.1, page 3.
An optional explosion-proof reset switch further expands the units versatility, making it possible to reset the
total and view daily fow logs without opening the enclosure.
Product Identifcation
Each device is labeled with a serial tag that identifes the product by part number and serial number and
identifes the devices input power requirements and safety classifcation. The CSA-approved device is
marked as shown in Figure 1.2. The ATEX-approved device is marked as shown in Figure 1.3. A description
of the electrical protection afforded by SIRA certifcation and associated safety markings are also shown in
Figure 1.3.
PART
SERIAL
167018
Figure 1.2Serial tag for CSA-approved device
Equipment Group II, Category 2 (hazardous conditions
are likely to occur in normal operation occasionally (>10<1000 hours/year)
Explosive Atmosphere: Gas, Dust
Flameproof for explosive gas environments other than
mines; temperature class
CE marking and number of notified
body responsible for production
Explosion-proof marking
Certification numbers
Temperature tested for dust and suitable for use in
Zone 21 area; ingress protection: dust-tight and protected against
the effects of low-pressure jets of water from all directions;
maximum surface temperature: 85C
NUFLO
TM
MC-III EXP FLOW ANALYZER
PART
SERIAL
FOR INSTALLATION & OPERATION INFO,
SEE MANUAL PART NO. 9A-50165003.
WARNING: DO NOT OPEN WHEN
AN EXPLOSIVE ATMOSPHERE
MAY BE PRESENT.
II 2 GD
AAAA
Ex d IIC T6 Gb (-40C to 70C) or
Ex tD A21 IP66 T85C (-40C to 70C)
Sira 10ATEX1264X
IECEx SIR10.0127X
INPUT POWER: 6 to 30V @ 6 mA
LOCATION OF MANUFACTURER
Figure 1.3Serial tag for ATEX-approved device
3
MC-III EXP Flow Analyzer Section 1
Table 1.1MC-III EXP Specifcations
Electrical Safety
Classifcation
Approved by CSA for US and Canada
Class I, Div. 1, Groups B, C, D (explosion-proof)
Type 4 enclosure
T6 temperature class
Approved by SIRA to
ATEX 10ATEX 1264X
IECEx SIR10.0127X
E
II 2 GD
Ex d IIC T6 Gb (-40C to 70C) or Ex tD A21 IP66 T85C (-40C to 70C)
System Power Internal power supply
3.6 VDC, D-size lithium battery (2-year typical life)
Alkaline battery pack containing 3 C-size industrial-grade batteries (CSA
only)
External power supply (6 to 30 VDC) with internal battery backup (reverse
polarity protected)
Loop-powered (4-20 mA) with internal battery backup
(reverse polarity protected)
Loop power: 8 to 30 VDC
Load resistance: 1100 ohms @ 30 VDC; 200 ohms @ 12 VDC
Operating Temperature Lithium-Powered: -40C to 70C (-40F to 158F)
Alkaline-Powered (CSA only): -18C to 55C (0F to 130F)
LCD contrast is reduced below -20C (-4F)
LCD Display 8-digit Total (volume) display (7-segment characters)
6-digit Rate display (11-segment characters for easy-to-read prompts)
0.3 character height
Adjustable contrast and update period
User-selectable units of measurement (Total):
Preprogrammed units: BBL, GAL, LIT, M
3
, CF, SCF, any unit x 1000
User-defned units
User-selectable units of measurement (Rate):
Preprogrammed units: BBL, GAL, LIT, M
3
, CF, SCF (per DAY, HR,
MIN, SEC), any unit x 1000 (per DAY, HR, MIN, SEC)
User-defned units
Keypad 6-key membrane switch
Communications/
Archive Retrieval
RS-485 Modbus

communications with transfer speeds up to 115.2K (allows


full download in less than 1 minute)
Logging 384 daily logs
768 hourly logs
345 event logs
4
Section 1 MC-III EXP Flow Analyzer
Table 1.1MC-III EXP Specifcations
Inputs Turbine Meter Input
Confgurable sensitivity adjustment via front panel
Sensitivity adjustment range: 20 mV P-P to 40 mV P-P
Frequency range: 0 to 3500 Hz
Remote Reset Input
Optically isolated
3.0 to 30 VDC
Pulse duration > 3 seconds to reset
Explosion-Proof Reset Switch (option)
Press and hold > 3 seconds, then release to reset total
Press and hold < 1 second, then release to view daily log data
Pulse Input
Optically isolated
3.0 to 30 VDC
Frequency range: 0 to 3500 Hz
Outputs Analog Output
4-20 mA, loop-powered (two-wire)
16-bit resolution
Accuracy: 0.1% of full scale @ 25C, 50 PPM/C temperature drift
Loop power: 8.0 to 30 VDC
Zero and full-scale engineering values confgurable from front panel
RS-485 Communications
Baud rates: 300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600 and up
to 115.2K
Volumetric Pulse Output
Solid-state relay
Output rating: 60 mA max. @ 30 VDC, on-state drop = 1.4 VDC @ 50 mA,
0.25 VDC @ 10 mA
Confgurable pulse width (duration): 10 to 60,000 ms
Amp & Square (Flowmeter Frequency) Output
Open-drain transistor output of turbine meter input signal
Output rating: 50 mA @ 30 VDC, on-state drop = 0.3 VDC @ 50 mA,
0.1 VDC @ 10 mA
(Analog output and amp & square outputs cannot be used simultaneously.)
Modbus

RTU mode Modbus

supports 16-bit and 32-bit holding registers. For more


information, see Appendix D, page D-1.
Enron Modbus

Flow log parameters (time stamp, period total, period run time, and supply
voltage) and download method are Enron-compatible.
5
MC-III EXP Flow Analyzer Section 1
Table 1.1MC-III EXP Specifcations
System Requirements Operating System - Windows XP, Windows Vista

, or Windows 7
Computer/Processor - 1 GHz or faster Pentium-compatible CPU
Memory - 128 MB of RAM
Hard Disk Space - 21 MB for program fles, 30 MB for Adobe Reader,
adequate space for data fles
Drive - CD-ROM for install
Display - 800 x 600 (SVGA), 16-bit (thousands of colors) color display or
greater
Browser - Internet Explorer 6 or later
Internet Connection - for web links, tech support
Communications Port - physical or virtual RS-232 compatible serial port
Key Product Features
This section presents an overview of key features of the MC-III EXP. Many of these features are discussed in
more detail in Section 3, page 29, and Section 4, page 39 (confguration procedures) and Section 5, page 73 (fow
log archival).
Key features discussed here include:
LCD display
keypad
interface software
power supply
calibration options
input options
output options
fow log archival
password-protected security
LCD Display
The liquid crystal display (Figure 1.4, page 6) provides a simultaneous indication of accumulated total (top
readout) and fow rate (bottom readout). The eight-digit total display uses 7-segment characters to form
numbers and letters, which results in a combination of uppercase and lowercase letters. The six-digit fow rate
display uses 11-segment characters to form numbers and letters for improved readability. When the keypad
is used to calibrate the MC-III EXP, the name of the menu option selected appears in the lower (rate) display,
and settings are entered in the top (total) display.
Flow volume can be measured in barrels, gallons, liters, cubic meters, cubic feet, standard cubic feet or other
user-defned units. A multiplication factor is also available for indicating fow volume in terms of 1,000 units.
The unit of measure for the Total readout and the decimal point position are selected by the operator during
calibration. If a user-defned unit is used, none of the preprogrammed volume units will be visible on the
display during operation.
Flow rate can be measured in a wide variety of preprogrammed units, or other user-defned units. The fow
rate unit of measure is selected in two steps: (1) a volume unit is chosen and (2) a time-base unit (per day, per
hour, per minute, or per second) is chosen. Users can choose any combination of preprogrammed volume and
time units in establishing the fow rate engineering unit (for example, gallons per hour, gallons per day, or
6
Section 1 MC-III EXP Flow Analyzer
gallons per minute). Also, the volume unit used for the fow rate can be different from the volume unit used to
read Total volume. The unit of measure for the Rate readout and the decimal point position are selected by the
operator during calibration.
The daily index (Day) display is a two-digit number for selecting a daily archive log for viewing. The
number shown here represents the number of days that have passed since the log was saved. For example, an
entry of 01 would yield yesterdays log. An entry of 05 would yield the log generated 5 days prior. Up to 99
consecutive daily logs can be viewed using the keypad. In addition, up to 384 daily logs, 768 hourly logs, and
345 event logs can be viewed through the interface software.
The LCD contrast can be adjusted with the interface software (see Section 4, page 39).
00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
00
Day
Total
Rate
Figure 1.4LCD display showing location of the Total, Rate, and daily index (Day) displays. Only the units of
measurement selected for displaying total and rate will be visible during normal operation.
Keypad
The six-button keypad allows users to perform a basic confguration of the instrument. Figure 1.5, page
7, summarizes the functions that can be accessed with each button. Most parameters can be confgured in
seconds by selecting one of the three menu keys (K-Factor, Output, or Display), navigating settings with the
arrow buttons, and saving the selections with the Enter key.
Section 3, page 29, contains procedures for confguring the MC-III EXP using the keypad. Icons of the six
buttons provide a pictorial reference to help guide users through each step of confguration.
Important: Some confguration parameters are accessible only through the interface software. See
Section 4, page 39, for instructions on confguring the instrument using the software.
Interface Software
Developed within the familiar Windows environment, the MC-III interface software is an intuitive and
easy-to-use application that provides access to all controls for setting up and operating the instrument. The
interface tailors the controls to the users needs, providing three options for confguring the instrument:
a Confguration Wizard, which steps through the most common confguration tasks
individual confguration menus for accessing specifc settings
an advanced menu that gives host programmers access to Modbus

registers
For instructions on installing the interface software and entering confguration parameters via the interface
software, see Section 4, page 39.
7
MC-III EXP Flow Analyzer Section 1
LOG TEST SAVE
ENTER
K-FACTOR
MENU
OUTPUT
MENU
DISPLAY
MENU
DISPLAY
MENU
K-FACTOR
MENU
OUTPUT
MENU
RESET TOTAL
LOG TEST SAVE
ENTER
Day
Total
Rate
Calibrate instrument
Set flowmeter input sensitivity
Select pulse input
Enable/disable/set up 4-20 mA output
pulse output
Enter slave address
Enter baud rate
Enable/disable/set up
Set engineering units for Total readout
Set engineering units for Rate readout
Set decimal placement for both readouts
During calibration, navigates
between submenus and scrolls
through menu selections
During operation, accesses
daily volume archive
During calibration, incrementally
changes digits and decimal point
position, on/off settings
During operation, tests temperature
and system voltage
During calibration, saves
calibration settings
During operation, saves
Total to nonvolatile memory
Note: To zero the Total
readout, press the LOG (left
arrow) and SAVE (ENTER)
keys simultaneously.
Figure 1.5MC-III EXP keypad functions
Power SupplyCSA-Approved Units
The CSA-approved MC-III EXP is shipped with either a lithium battery or an alkaline battery pack. Alter-
nately, the MC-III EXP may be powered by an external power source; in this case, the lithium battery or alka-
line battery pack provides a backup power supply, signifcantly extending battery service life. A low-power
microprocessor enables the MC-III EXP to operate approximately 2 years on a single lithium battery.
Wiring diagrams for connecting an external power supply are provided in Section 2, page 13.
Power SupplyATEX-Approved Units
The ATEX-approved MC-III EXP is shipped with a lithium battery. Alternately, the MC-III EXP may be
powered by an external power source; in this case, the lithium battery provides a backup power supply,
signifcantly extending battery service life. A low-power microprocessor enables the MC-III EXP to operate
approximately 2 years on a single lithium battery.
Wiring diagrams for connecting an external power supply are provided in Section 2, page 13.
8
Section 1 MC-III EXP Flow Analyzer
Multipoint Linearization
The MC-III interface software supports up to 12 calibration points in addition to single-point calibration based
on the K-factor provided with the turbine fowmeter. See K-Factor Type, page 60, for more information.
Gas Volume Correction
Gas turbine meters are calibrated in actual cubic feet (ACF), and measure gas in actual cubic feet. In some
applications, a user may beneft from referencing gas measurements back to standard conditions by measuring
in terms of standard cubic feet (SCF). The MC-III EXP makes this process quick and easy, by using fxed
average values for the fowing gas temperature and fowing gas pressure. See Gas Volume Correction, page 62,
for more information.
Input Options
The fowmeter signal can be obtained from a magnetic pickup or a pre-amplifer device. The sensitivity of the
fowmeter input may be adjusted with the instrument keypad or the interface software. See Section 2, page 13,
for wiring diagrams. See Section 3, page 29, and Section 4, page 39, for confguration procedures.
Output Options
The MC-III EXP standard circuitry provides:
a scaled pulse output representing an increment in volume for each pulse
a loop-powered 4-20 mA output representing the fow rate
a fowmeter frequency output for use with remote equipment to derive fow rate and volume
an RS-485 output for communication with interface software or other telemetry equipment
When the 4-20 mA rate output feature is used, the MC-III EXP is powered by the current loop, and the lithium
battery or alkaline battery pack is used as a backup supply.
The pulse output and 4-20 mA output features should be turned off when not required for reduced current
consumption. See Section 2, page 13, for wiring diagrams. See Section 3, page 29, and Section 4, page 39, for
confguration procedures.
Flow Log Archival
The MC-III EXP saves up to 384 daily logs and 768 hourly logs in nonvolatile memory. By connecting with
the interface software, users can download the logs for viewing and/or printing in tabular format or in a trend
chart.
Users can also export daily and hourly logs to a spreadsheet. For more information, see Section 5, page 73.
Event Log Archival
The MC-III EXP saves up to 345 user event logs. Event logs are generated to track user changes such as K-
Factor changes, input setting changes, power-on reset and watch-dog reset, fow cut-off and frequency cut-
off.
By connecting with the interface software, users can download the logs for viewing and/or printing in tabular
format. In addition to showing old and new values, each event log is time-stamped, and includes the register
associated with the change. For more information, see Section 5, page 73.
Password-Protected Security
A keypad security access code prevents unauthorized personnel from altering the calibration or accumulated
volume data in the instrument. The security feature may be disabled if this protection is not required.
9
MC-III EXP Flow Analyzer Section 1
Password-protected security access is enabled using the interface software. When this feature is enabled,
the user will be prompted for a password when attempting to enter any menu from the keypad. For more
information, see Security Setup, page 55.
Optional Features
Explosion-Proof Reset/Control Switch
An optional external reset/control switch allows the user to zero the total and view daily logs instantaneously
without opening the enclosure or accessing the interface software. The switch can be mounted in either of
the conduit openings near the top of the MC-III EXP enclosure. For installation and wiring instructions, see
Explosion-Proof Reset/Control Switch, page A-1.
Important: The switch is approved by CSA and SIRA (ATEX) for use in hazardous areas. The CSA-
approved model is Part No. 9A-30054001. The ATEX-approved model is Part No. 9A-
30054002.
Zeroing the Total. To reset the total with an explosion-proof switch, press and hold the switch for 3 seconds,
then release it.
Important: When resetting the total with the explosion-proof reset switch, do not release the switch
prematurely. Releasing the switch too quickly (before 3 seconds have passed) will
activate the daily log view function and WILL NOT reset the volume to zero.
Viewing Daily Logs. To view daily logs with the explosion-proof switch, press the switch for 1 second
and release it. Subsequent press-and-release actions increment the daily index, allowing the user to select
a specifc daily total. Additionally, when the log viewing function times out, the accumulated total is
automatically saved to memory.
Saving the Total. When the daily log display times out, the MC-III EXP will automatically save the
accumulated total.
RS-485 Communications Adapter
An explosion-proof communications adapter provides an RS-485 connection for connecting a laptop or PC
to the instrument without removing the instrument cover. When the adapter is ordered with an MC-III EXP,
it is factory installed. It may be relocated to either conduit opening in the instrument housing. When ordered
separately, the adapter is easily installed using the instructions provided in this manual. For installation and
wiring instructions, see Explosion-Proof Communications Adapter, page A-3.
Important: The RS-485 communications adapter is approved by CSA and SIRA (ATEX) for use in
hazardous areas. The CSA-approved model is Part No. 9A-90017004. The ATEX-approved
model is Part No. 9A-90017008.
USB Communications Adapter (CSA-Approved Units Only)
The USB communications adapter allows an MC-III EXP to be connected directly to a USB port of a laptop
or PC. A user-supplied universal USB cable is required. The adapter is factory-installed when purchased with
an MC-III EXP. It is also available as a kit with an installation CD for upgrading communications in a feld
unit. See USB Communications Adapter, page A-6.
10
Section 1 MC-III EXP Flow Analyzer
Important: The USB communications adapter is approved by CSA for use in hazardous areas. The
CSA-approved model is Part No. 2295524-01. The adapter kit, which includes an installa-
tion CD, is Part No. 2295634-01.
RS-485 IS Barrier (CSA-Approved Units Only)
The RS-485 output of a CSA-approved MC-III is approved for intrinsically safe installations. An intrinsically
safe barrier assembly with a conduit seal enables the RS-485 output to be used as an intrinsically safe output.
In such installations, only the turbine fowmeter input may be used. See Figure 2.18, page 28, for an IS barrier
assembly wiring diagram.
Commonly Used Functions
While the functions of the MC-III EXP are too numerous to mention, some of the most commonly used
functions are detailed in this section. They include:
reading the rate and accumulated total
saving totals to memory
resetting the total
viewing daily and hourly logs
saving and uploading confguration fles
exporting log data to spreadsheet
saving log data in a report
Reading Totals
Current totals can be viewed from the LCD on the front of the MC-III EXP or from the interface software
(MC-III Main screen). The software calculates the fow total and updates the LCD display every 4 seconds, by
default. The user can adjust the calculation period with the interface software. See Calculation Period, page 59,
for more information.
Saving Totals to Memory
Hourly and daily totals are automatically saved to nonvolatile memory. A user may save an accumulated total
at any time by opening the enclosure and pressing ENTER (SAVE) on the keypad. In the event of a power
failure, the last saved total will be displayed on the LCD when power is restored. See also Explosion-Proof
Reset/Control Switch, page 9.
Note: Always save the accumulated total before replacing batteries.
Resetting the Total
Totals can be reset to zero using the keypad, the interface software, an explosion-proof switch mounted in the
top of the enclosure (if equipped), or a pulse from an external device.
To reset the total with the keypad, press LEFT ARROW (LOG) and ENTER (SAVE) keys simultaneously.
To reset the total with the interface software, double-click the NuFlo MC-III icon on the computer desk-
top and wait for the software to connect to the instrument; then select MC-III Main from the Device
Autorun Options screen, and click on the Reset Flow Total button in the lower right corner of the Main
screen.
11
MC-III EXP Flow Analyzer Section 1
To reset the total with an explosion-proof switch, press and hold the switch for 3 seconds, then release it.
See also Explosion-Proof Reset/Control Switch, page 9.
To reset the total with an external pulse generator located in safe area, confgure the pulse to be active for
3 seconds. See also Remote Reset Input, page 23.
Viewing Daily and Hourly Logs
Each day, as the user-defned contract hour passes, a daily fow total is saved to nonvolatile memory. Hourly
logs are also automatically saved. A total of 384 daily logs and 768 hourly logs are accessible for viewing and
exporting using the interface software. See Section 5, page 73, for details.
Up to 99 daily fow log totals can be viewed from the LCD. Hourly fow log totals are accessible only through
the interface software.
To view daily fow totals from the LCD, perform the following steps:
1. Press the LEFT ARROW (LOG) key on the keypad. The words Daily Volume Archive will scroll across
the bottom of the LCD and the day index will display 01. The daily fow total recorded at the last con-
tract hour will appear at the top of the LCD.

The index number represents the number of days previous to the current date. For example, yesterdays
totals are read by entering an index of 01; totals from two days previous are read by entering 02.
2. Press UP ARROW (TEST) to increment the index (01, 02, 03...); press LEFT ARROW (LOG) to decre-
ment the index (01, 99, 98...).
3. Press ENTER (SAVE) to exit the Daily Volume Archive menu. (After 2 minutes of inactivity, the Daily
Volume Archive menu will timeout and the total readout will be restored automatically.)
The optional explosion-proof reset switch also allows users to view daily log totals. For more information, see
Explosion-Proof Reset/Control Switch, page 9.
Saving and Uploading Confguration Files
The MC-III interface software allows users to save an unlimited number of confguration fles to their
computer. In the event that a confguration setting gets changed unintentionally or a user simply wants to
restore the settings he used previously, the user can upload the confguration fle and resume operation within
minutes. The upload function also allows a confguration fle to be loaded quickly into multiple devices. The
default directory for saving confguration fles is C:\NuFlo log data\MC-III. However, the user can specify a
different location, if desired.
For complete information, see Saving and Uploading Confguration Files, page 69.
Exporting Log Data
Flow logs and event logs can be directly exported to an .xls or .csv fle. For complete information, see
Exporting Flow Logs, page 80, and Exporting Event Logs, page 83. The default directory for exported logs is C:\
NuFlo log data\MC-III. However, the user can specify a different location, if desired.
Saving Log Data in a Report
Flow logs can be saved in a report format that can be loaded back into the software for viewing or printing
at a later time. For complete information, see Printing/Saving a Report, page 83. The default directory for log
reports is C:\NuFlo log data\MC-III\<WELL NAME>. However, the user can specify a different location.
12
Section 1 MC-III EXP Flow Analyzer
13
MC-III EXP Flow Analyzer Section 2
Section 2Installation
The MC-III EXP is fully assembled at the time of shipment and ready for mounting to a fowmeter. Field
wiring enters the MC-III EXP enclosure through either of two threaded conduit openings in the top of the
enclosure. When not in use, the openings are sealed with pipe plugs (Figure 2.1). Ground screws are provided
both inside and outside the enclosure.
The device is shipped with recessed pipe plugs installed facing right and left, and the pipe union facing
downward. If the device is equipped with an explosion-proof reset switch or a communications adapter, only
one plug is required. The switch may be installed in the right or left conduit opening.
Pipe plugs
Ground
screws
Figure 2.1Enclosure entry and ground screw locations
Installation requirements and meter mounts vary with certifcation regulations and meter types. For ATEX
installation requirements, see Hazardous Area Installations (ATEX) below. For CSA installation requirements,
see Hazardous Area Installations (CSA), page 16.
Cameron recommends that operators confgure the device prior to mounting if the instrument is to be installed
in a hazardous area. The enclosure must be opened to confgure the device, either via keypad controls or via
software.
!
WARNING: Once the instrument is mounted in a hazardous area, the cover should not be re-
moved unless the area is void of combustible gas and vapors.
Hazardous Area Installations (ATEX)
The ATEX-certifed MC-III EXP is fully compliant with European ATEX Directive 94/9/EC, Annex II,1.0.6.
The following instructions apply to equipment covered by certifcate numbers 10ATEX1264X and IECEx
SIR10.0127X :
The device may be used in the presence of fammable gases and vapours with apparatus groups IIA, lIB
and IIC and with temperature classes T1, T2, T3, T4, T5, and T6.
The device is certifed for use in ambient temperatures in the range -40C to 70C and should not be used
outside this range.
14
Section 2 MC-III EXP Flow Analyzer
The device has not been assessed as a safety-related device (as referred to by Directive 94/9/EC Annex II,
clause 1.5).
Installation shall be carried out by suitably trained personnel in accordance with the applicable code of
practice (EN 60079-14 within Europe).
Repair of this equipment shall be carried out by Cameron or in accordance with the instructions in this
manual.
If the instrument is likely to come into contact with aggressive substances, the user is responsible for tak-
ing suitable precautions to prevent it from being adversely affected, thus ensuring that the type of protec-
tion is not compromised.
Aggressive substances may include, but are not limited to, acidic liquids or gases that may attack met-
als, or solvents that may affect polymeric materials.
Suitable precautions may include, but are not limited to, regular checks as part of routine inspections
or establishing from the materials data sheet that it is resistant to specifc chemicals.
Conditions for Safe Use (ATEX)
In accordance with EN60079-0, Clause 16.5, all cable and cable glands must be rated for 80C.
Do not remove the cover unless the area is void of combustible gas and vapors.
Before removing the pipe plug to install a communications adapter, or disassembling the union nut and
blanking plug in a pre-installed communications adapter, make sure the area is nonhazardous. Reinstall
the pipe plug in accordance with the instructions in this manual.
An ATEX-approved standoff tube shall be used with a Barton 7000 Series turbine pickup coil boss or a
NuFlo turbine pickup coil boss in accordance with certifcate SIRA 03ATEX1474U. See the MC-III Parts
List (ATEX), page 94, for part numbers.
Replace the lithium battery only with the ATEX-approved battery supplied by Cameron (Part No. 9A-
90099000).
The lithium battery that powers the MC-III EXP is a sealed unit; however, should a lithium battery
develop a leak, toxic fumes could escape upon opening the enclosure. Ensure that the instrument is in a
well-ventilated area before opening the enclosure to avoid breathing fumes trapped inside the enclosure.
Exercise caution in handling and disposing of spent or damaged batteries.
Mounting Options (ATEX)
The ATEX-approved MC-III EXP may be mounted directly to a fowmeter, or mounted remotely on a pole
and connected to a fowmeter with explosionproof signal cable. The type of meter mount required varies with
the meter in use. See Figures 2.2 through 2.4 for installation dimensions.
Mount the device to a NuFlo turbine meter. A 3/4-in. to 1-in. standoff tube is required (Figure 2.2, page 15).
Mount the device to a Barton 7000 Series turbine meter. A 3/4-in. to 3/4-in. stainless steel standoff tube is
required (Figure 2.3, page 15). (The 3/4-in. to 1-in. reducer supplied with the Barton 7000 Series meter is
not required when the standoff tube is used.)
Mount the device to a fat, vertical surface.
Remote-mount the device to a 2-in. pole using a NuFlo hardware kit (Figure 2.4, page 16). Pole mount-
ing may be preferred where limited space or pipe vibration prohibits direct-mount installation. For best
results, install the device as near as possible to the meter to minimize the length of explosionproof signal
cable that must be used for communications. Stranded, shielded cable is recommended.
15
MC-III EXP Flow Analyzer Section 2
5.71 (145)
8.57
(217.7)
approx.
4.94
(125.5)
5.00 (127)
Figure 2.2Direct mount to NuFlo turbine meter with 3/4-in. to 1-in. standoff tube (ATEX)
5.71 (145)
4.94
(125.5)
9.29
(236.1)
approx.
5.71 (145)
Figure 2.3Direct mount to a Barton 7000 series fowmeter with 3/4-in. to 3/4-in. standoff tube (ATEX)
16
Section 2 MC-III EXP Flow Analyzer
5.00 (127)
4.94
(125.5)
7.45
(189.3)
6.56
(16.7)
6.95
(17.65)
Figure 2.4Remote mount on a 2-in. pole using a NuFlo hardware kit (Part No. 9A-30028004)
Hazardous Area Installations (CSA)
The CSA-approved MC-III EXP has been certifed as explosion proof for Class I, Division I, Groups B, C,
and D for use in the US and Canada.
Conditions for Safe Use (CSA)
The device may be used in the presence of fammable gases and vapours with apparatus groups B, C, and
D and with temperature classes T1, T2, T3, T4, T5, and T6.
The device is certifed for use in ambient temperatures in the range -40C to 70C and should not be used
outside this range.
Do not remove the cover unless the area is void of combustible gas and vapors.
Before removing the pipe plug to install a communications adapter, or disassembling the union nut and
blanking plug in a pre-installed communications adapter, make sure the area is nonhazardous. Reinstall
the pipe plug in accordance with the instructions in this manual.
All feld wiring must conform to the National Electrical Code, NFPA 70, Article 501-4(b) for installations
within the United States or the Canadian Electric Code for installations within Canada. State and local
wiring ordinances may also apply. All feld wiring must be rated for temperatures of 90C or higher, and
have a wire range of 22 to 14 AWG. Terminal block screws must be tightened to a minimum torque of 5
to 7 in-lb. to secure the wiring within the terminal block. Only personnel who are experienced with feld
wiring should perform these procedures.
Do not use Tefon

tape on threads of the union, reducer, or pipe plugs. Use of Tefon

tape will void the


explosion-proof rating of the instrument.
17
MC-III EXP Flow Analyzer Section 2
Replace the lithium battery only with the CSA-approved battery supplied by Cameron (Part No. 9A-
100005111).
Replace the optional alkaline battery pack only with the CSA-approved battery pack supplied by Cameron
(Part No. 9A-50099002).
The lithium battery that powers the MC-III EXP is a sealed unit; however, should a lithium battery
develop a leak, toxic fumes could escape upon opening the enclosure. Ensure that the instrument is in a
well-ventilated area before opening the enclosure to avoid breathing fumes trapped inside the enclosure.
Exercise caution in handling and disposing of spent or damaged batteries.
Mounting Options (CSA)
The CSA-approved MC-III EXP is available in either of two enclosure styles. The standard enclosure shown
on the cover of this manual and in Figure 2.5, page 18, features integral mounting holes on either side of the
enclosure, allowing it to be mounted to a fat bulkhead.
The alternate enclosure (Figure 2.6, page 18) does not have the bulkhead mounting holes on the sides. Another
distinguishing feature is the bright red ring that frames the display window. The alternate enclosure is slightly
taller than the standard enclosure.
The MC-III EXP may be mounted directly to a fowmeter, or mounted remotely on a pole and connected to a
fowmeter with explosionproof signal cable. The type of meter mount required varies with the meter in use.
See Figure 2.5 through Figure 2.7 for installation dimensions.
Mount the device to a NuFlo turbine meter or a Barton 7000 Series turbine meter. A CSA-approved pipe
reducer and union connector is required (Figure 2.5, page 18). Figure 2.6, page 18, shows the same installa-
tion with the alternate MC-III enclosure.
If mounting to the Barton 7000 Series meter, a 3/4-in. to 1-in. reducer supplied with the meter is used to
achieve the 3/4-in. connection required to mate the union to the meter.
Mount the device to a fat, vertical surface. The standard enclosure is ideal for a bulkhead mount; the
alternate enclosure is not recommended for bulkhead mounting.
Remote-mount the device to a 2-in. pole using a NuFlo hardware kit (Figure 2.7, page 19). Pole mount-
ing may be preferred where limited space or pipe vibration prohibits direct-mount installation. For best
results, install the device as near as possible to the meter to minimize the length of explosionproof signal
cable that must be used for communications. Stranded, shielded cable is recommended.
18
Section 2 MC-III EXP Flow Analyzer
5.71 (145)
6.73
(171.1)
1.75
(44.5)
5.00 (127)

Figure 2.5Direct mount (standard MC-III EXP enclosure) to NuFlo turbine meter with pipe reducer and
union (CSA)

5.00 (127.0)
7.05
(179.1)
1.63 (41.4)
4.55 (115.6)
Figure 2.6Direct mount (alternate MC-III EXP enclosure) to NuFlo turbine meter with pipe reducer and
union (CSA)
19
MC-III EXP Flow Analyzer Section 2
5.00 (127)
4.94
(125.5)
7.45
(189.3)
6.56
(16.7)
6.95
(17.65)
Figure 2.7Remote mount on a 2-in. pole using a NuFlo hardware kit (Part No. 9A-30028004)
Installing the MC-III EXP
If the device is to be installed in a hazardous area, confgure the device prior to mounting it. See Sections 3
and 4 for instructions.
1. Make sure the fowmeter and magnetic pickup are properly installed as follows:
a. Install the turbine fowmeter in the fow line.
b. Lightly grease the threads on both ends of the magnetic pickup, taking care to keep grease off of the
connector contacts. If the connector is plastic, apply grease only to the end that threads into the meter.
c. Install the magnetic pickup in the fowmeter.
4. Position the MC-III EXP above the fowmeter. If installing an ATEX-approved MC-III EXP with a tube
adapter, make sure the signal cable is disconnected from the MC-III circuit board. This will prevent the
cable from being damaged from twisting when the tube adapter is threaded onto the meter in step 6.
5. Plug the MC-III EXP cable connector into the magnetic pickup and hand-tighten the knurled nut on the
connector.
6. Screw the MC-III EXP onto the fowmeter threads surrounding the magnetic pickup with the display fac-
ing the desired direction.
7. Tighten all pipe connections.
Caution: Do not use Tefon

tape on threads of the union, reducer, or pipe plugs. Use of Tefon


tape will void the explosion-proof rating of the instrument.
20
Section 2 MC-III EXP Flow Analyzer
Field Wiring
The MC-III EXP supports connections for the following: internal power (battery), external power, turbine
fowmeter input, pulse input, remote reset input, pulse output, 4 to 20 mA output, fowmeter frequency (amp
& square) output, and RS-485 output.
Wiring diagrams for each connection are provided below.
Caution: For a CSA-approved MC-III EXP installed in a hazardous location, all feld wiring must
conform to wiring methods for explosion-proof installations as defned in the
National Electric Code for installations within the United States or as specifed in the
Canadian Electric Code for installations within Canada. State and local wiring
ordinances may also apply.
To wire the MC-III EXP for operation, complete the following feld connections:
1. Unscrew the cover of the enclosure and remove it.
2. Using a small standard blade screwdriver, remove the two #4-40 7/8 screws located to the right and
left side of the display.
3. Lift the display/keypad assembly from the enclosure, making sure the circuit assembly does not contact
the enclosure.
4. Connect the lithium battery or alkaline battery pack to the J1 connector marked BATTERY near the top
of the circuit assembly.
5. Connect wiring for external power, if appropriate. See Figure 2.8, page 21.
6. Connect the fowmeter or pulse input wiring to terminal block TB1. See Figure 2.9, page 22, and Figure
2.10, page 22.
7. Connect wiring for the remote reset input to terminal block TB1, if appropriate. See Figure 2.11, page 23
and Figure 2.12, page 23.
8. Connect wiring for output signals, if appropriate. See Figure 2.13, page 24, through Figure 2.17, page 27.
9. Place the circuit assembly over the standoffs and fasten with the two #4-40 7/8 screws, ensuring that
all connector wiring is inside the enclosure and in no position where it may be damaged when the enclo-
sure cover is replaced.
10. Recalibrate the MC-III EXP (if necessary).
11. If external and internal power supplies were removed, reset the clock to ensure that the time stamps in the
log data are accurate. The clock is reset using the MC-III interface software. See Time/Date Synchroniza-
tion, page 54.
12. Replace the enclosure cover.
21
MC-III EXP Flow Analyzer Section 2
Internal Power Supply
The MC-III EXP is shipped with one of three internal power supplies:
a CSA-approved 3.6V lithium battery
an ATEX-approved 3.6V lithium battery
a shrink-wrapped alkaline battery pack containing three C-size industrial-grade batteries (available only
with CSA-approved devices)
Low-power microprocessor technology enables the MC-III EXP analyzer to operate approximately 2 years on
a single lithium battery. The lithium battery is strongly recommended for use in extreme temperatures (below
-20C).
Users can power the instrument from an external power supply or a 4-20 mA current loop, and use the lithium
or alkaline battery (where applicable) as a backup power supply. The use of an alternate power source extends
battery life and helps ensure that timekeeping and volume accumulation will not be interrupted during a
power failure.
External Power Supply
The MC-III EXP can be connected to a remote power supply by a two-conductor cable (Figure 2.8). The
power supply and cable must be capable of supplying 6 to 30 VDC @ 10 mA. This capability is available
only if the 4-20 mA rate output is not used.
A Zener diode (Part No. 1.5KE33CA) must be installed for CE approval.
Caution: When using the amp & square output with the external power supply, make sure the
power supplies for both features share a common negative (-) terminal or that they are
totally isolated from each other, since both share a common negative (-) connection.
T
B
3
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-
3
0
V
D
C
POWER
SUPPLY
6 to 30
VDC
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
TB3 MARKINGS:
GND = SIGNAL GROUND
Figure 2.8External power supply wiring
22
Section 2 MC-III EXP Flow Analyzer
Turbine Flowmeter (TFM) Input
The TFM input provides the turbine fowmeter input signal generated by a magnetic pickup, enabling the MC-
III EXP to calculate and display instantaneous fow rates and accumulated totals.
6
-3
0
V
D
C
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TURBINE
MAGNETIC PICKUP
VMAX = 3.9 V
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
A
B
Figure 2.9Flowmeter input wiring
Pulse Input
The pulse input provides an optically isolated input in systems where a preamplifer is inserted between the
sensor and the MC-III EXP.
T
B
3
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
EXT POWER
6
-3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
PULSE INPUT
3 TO 30 VDC
Figure 2.10Pulse input wiring
23
MC-III EXP Flow Analyzer Section 2
Remote Reset Input
The remote reset input allows the operator to reset the accumulated volume on the MC-III EXP to zero
without opening the enclosure. This input is optically isolated. The input is shown connected in two ways,
with a power supply and switch in a safe area (Figure 2.11), and with a pulse generator in a safe area (Figure
2.12). The remote reset input can also be controlled with an optional explosion-proof switch mounted in
the top of the enclosure. See Explosion-Proof Reset/Control Switch, page A-1, for installation and wiring
instructions.
The reset input, reset pulse, or local reset switch must be active for 3 seconds to clear the total. (The
explosion-proof switch can also be used to view daily logs and to save the accumulated volume to memory
without opening the enclosure. See Explosion-Proof Reset/Control Switch, page 9, for more information.)

TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
6
-3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
POWER SUPPLY
3 to 30 VDC
Figure 2.11Reset input wiring (when power supply and reset switch are in a safe area)
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
6
-3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
RESET PULSE
3 TO 30 VDC
Figure 2.12Reset pulse input wiring (when pulse generator is in a safe area)
24
Section 2 MC-III EXP Flow Analyzer
Pulse Output
The pulse output is a solid-state relay. Each pulse represents a user-defned volume. Because the circuit is
isolated, it can be used in conjunction with any other feature on the MC-III EXP. A two-conductor cable from
the MC-III EXP to the remote location is required (Figure 2.13). The maximum current rating of the pulse
output circuit is 60 mA at 30 VDC.
A Zener diode (Part No. 1.5KE56CA) must be installed for CE approval.
For information on confguring the pulse output with the keypad, see Confguring the Pulse Output, page 37.
For information on confguring the pulse output with the interface software, see Confguring Pulse Output,
page 67.
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-
3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
Leave this
end of shield
disconnected.
POWER SUPPLY
5 to 50 VDC
*
Resistor may be included in pulse readout
device. Size the resistor to limit the current
to 60 mA.
*
PULSE READOUT
DEVICE
Figure 2.13Pulse output wiring
25
MC-III EXP Flow Analyzer Section 2
Analog (4-20 mA) Rate Output
The 4-20 mA rate output provides a linear current output that represents fow rate. This output requires a two-
conductor cable connected to an 8 to 30 VDC power supply (voltage required is dependent on loop resistance)
and a current readout device located in the remote location. The 4-20 mA rate output current loop also powers
the MC-III EXP. The internal battery (lithium or alkaline) provides a power supply backup to maintain
timekeeping accuracy and to continue accumulating volume in the event that the 4-20 mA current loop fails.
Figure 2.14 shows the minimum required voltage to power the instrument for a given loop resistance. In
addition, the mathematical relationship between loop voltage and load resistance is given. For example, if a
power supply voltage of 24 volts is available to power the current loop, the maximum load resistance would
be 800 ohms.
Caution: The 4-20 mA rate output and the fowmeter frequency output circuits are not isolated
from each other and cannot be used simultaneously. When the 4-20 mA output option is
used, do not connect external power to TB3.
For information on confguring the 4-20 mA rate output with the keypad, see Confguring the Rate Display,
page 34. For information on confguring the 4-20 mA rate output with the interface software, see 4-20 mA
Output, page 64.
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
POWER SUPPLY
8 to 30 VDC
LOAD
Resistor may be
included in readout
device.
4-20 mA and
flowmeter frequency
(amp & square)
cannot be used
simultaneously.
*
*
LOOP SUPPLY VOLTAGE (VDC)
L
O
A
D

R
E
S
I
S
T
A
N
C
E

(
O
H
M
S
)
OPERATING
REGION
Figure 2.144-20 mA rate output wiring
26
Section 2 MC-III EXP Flow Analyzer
Flowmeter Frequency Output
The fowmeter frequency (amp & square) output provides an open drain transistor output at the turbine meter
frequency, which may be used to provide fow rate and/or total information to peripheral equipment. The
output requires a two-conductor cable from the MC-III EXP to the remote frequency readout device requiring
50 mA or less and a 5 to 30 VDC power supply (Figure 2.15).
Caution: The fowmeter frequency output and 4-20 mA rate output are not isolated from each other
and cannot be used simultaneously.
Caution: When using the fowmeter frequency output and powering the device from an
external power supply, make sure both power supplies share a common negative (-)
terminal or are totally isolated from each other.
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-
3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
Leave this
end of shield
disconnected.
POWER SUPPLY
5 to 30 VDC
FREQUENCY
READOUT DEVICE
*
Resistor may be included in frequency
readout device. Size the resistor to limit
the current to 50 mA.
4-20 mA and flowmeter frequency
(amp & square) cannot be used
simultaneously.
*
TB3 MARKINGS:
GND = SIGNAL GROUND
A&S = AMP & SQUARE
Figure 2.15Flowmeter frequency (amp & square) output wiring
27
MC-III EXP Flow Analyzer Section 2
RS-485 Output
The RS-485 output is required for communication with the interface software. Wiring diagrams are provided
for a permanent connection (Figure 2.16), as well as for temporary laptop connections using an RS-485 to RS-
232 converter (Figure 2.17).
The RS-485 output may also be used to install an external communications adapter for connecting a
laptop without opening the enclosure. See Explosion-Proof Communications Adapter, page A-3, and USB
Communications Adapter, page A-6, for details.
Additionally, in a CSA-approved device, the RS-485 output is approved for intrinsically safe installations. An
intrinsically safe installation requires an intrinsically safe barrier assembly with a conduit seal (Figure 2.18,
page 28). In such installations, only the turbine fowmeter input may be used.
A
B
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-3
0
V
D
C
RS-485
Communications
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
Figure 2.16RS-485 output (permanent connection)
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
TD(B)
TD(A)
GND
RS-232
9 - PIN
CONNECTOR
Part No.
9A-101283116
TX+
TX-
Figure 2.17RS-485 output (connection to laptop with 9-pin converter)
28
Section 2 MC-III EXP Flow Analyzer
Figure 2.18Intrinsically safe RS-485 connection, based on control drawing 9A-50053010 (applicable to
CSA-approved device only)
29
MC-III EXP Flow Analyzer Section 3
Section 3Confguration and Operation via Keypad
Calibration of the MC-III EXP is a simple matter of entering necessary parameters into the instrument.
The process for calibrating the MC-III EXP depends on how the instrument will be used. The keypad can
be used to calibrate the MC-III EXP for liquid or gas measurement using preprogrammed units, or for liquid
measurement using a calculated divisor. When preprogrammed units are used, the MC-III EXP automatically
calculates the divisor for volume calculation and the rate multiplier for fow rate calculation, making
calibration quick and easy.
Important: For compensated gas measurement, or gas or liquid measurement requiring multipoint
(2- to 12-point) linearization, calibration must be performed with the software user inter-
face (see Section 4, page 39).
Because the keypad is protected beneath the lid of the instrument, the enclosure must be opened to access the
keypad. For this reason, it is important to calibrate the instrument before installing it in a hazardous area.
!
WARNING: To prevent ignition of hazardous atmospheres, do not remove the cover unless the
area is void of combustible gas and vapors.
The following parameters can be confgured using the six-button keypad on the front of the instrument:
calibration factor and unit
units of measurement for Total and Rate displays
decimal position for Total and Rate displays
input type and sensitivity level
4-20 mA output
pulse output scale factor
slave address
baud rate
30
Section 3 MC-III EXP Flow Analyzer
Entering a Calibration Factor
When the volume is to be expressed in barrels (BBL), gallons (GAL), liters (LIT), cubic meters (M
3
), or cubic
feet (CF), and the fow rate is to be expressed in barrels, gallons, liters, cubic meters or cubic feet per day, per
hour, per minute, or per second, the MC-III EXP calculates the divisor automatically; only the turbine meter
calibration factor is required.
When the volume is to be expressed in a unit other than the preprogrammed units listed above, a calculated
divisor must be entered at the K-Factor prompt. See Entering a Calculated Divisor, page 31.
To Enter a Calibration Factor:
Enter the K-Factor menu. Press K-FACTOR MENU.
K-FACTOR
MENU
00000000
M
3
BBL
GAL
LIT
CF
CF
X1000
X1000
Select the unit of measure associated
with the calibration factor (typically,
pulses per gallon).
Press UP ARROW until the correct
unit is displayed.
TEST
Enter the decimal point position for
the calibration factor.
Press LEFT ARROW to select the
decimal point.
LOG
00000000
M
3
BBL
GAL
LIT
CF
CF
X1000
X1000
.
Press UP ARROW to change the
position of the decimal point.
TEST
Press LEFT ARROW to save the
decimal point position and proceed
with entering a calibration factor.
LOG
Enter the calibration factor, starting
with the last digit and entering digits
right to left.
Press UP ARROW until the last digit
of the calibration factor is displayed.
TEST
00000000
M
3
BBL
GAL
LIT
CF
CF
X1000
X1000
Press LEFT ARROW to select the
next digit to the left.
Repeat using UP and LEFT arrows to
enter all remaining digits.
LOG
Press ENTER.
SAVE
ENTER
The Input Sensitivity menu will appear following the entry of the calibration factor. See Setting Input Type
and Sensitivity, page 32, for the input sensitivity confguration procedure.
31
MC-III EXP Flow Analyzer Section 3
Entering a Calculated Divisor
When registering the volume in units other than cubic meters, cubic feet, barrels, gallons, or liters, a
calculated divisor must be entered in the K-Factor menu instead of the turbine meter calibration factor.
Important: When a calculated divisor is used, the units for both the volume display and the volume
portion of the fow rate display should be set to USER (no units visible on the display).
The divisor and divisor decimal point must be determined, then entered in the K-Factor menu.
The formula for calculating the divisor is:
Divisor = FC CON
Where:
FC = meter factor in pulses per gallon (P/G)
CON = the conversion factor for number of gallons per unit volume of desired measure
To Enter a Calculated Divisor:
Enter the K-Factor menu. Press K-FACTOR MENU.
K-FACTOR
MENU
Select no unit of measure. Press UP ARROW until no unit is
displayed (as shown).
TEST
00000000
M
3
BBL
GAL
LIT
CF
CF
X1000
X1000
Enter the decimal point position for the
calculated divisor.
Press LEFT ARROW to select the
decimal point.
LOG
00000000
M
3
BBL
GAL
LIT
CF
CF
X1000
X1000
.
Press UP ARROW to change the
position of the decimal point.
TEST
Press LEFT ARROW to save the
decimal point position and proceed
with entering a calculated divisor.
LOG
Enter the calculated divisor, starting
with the last digit and entering digits
right to left.
Press UP ARROW until the last digit
of the calculated divisor is displayed.
TEST
00000000
M
3
BBL
GAL
LIT
CF
CF
X1000
X1000
Press LEFT ARROW to select the
next digit to the left.
Repeat using UP and LEFT arrows to
enter all remaining digits.
LOG
Press ENTER.
SAVE
ENTER
The Input Sensitivity menu prompt will appear immediately following the entry of the calibration factor. See
Setting Input Type and Sensitivity, page 32, for the input sensitivity confguration procedure.
32
Section 3 MC-III EXP Flow Analyzer
Setting Input Type and Sensitivity
The fowmeter signal can be obtained from a magnetic pickup or a pre-amplifer device.
The input sensitivity of the MC-III EXP is measured in millivolts (mV) peak-to-peak. This is the threshold
value at which the circuitry responds to a signal. If the input signal is less than this value, the MC-III EXP will
not count the electrical pulses as a valid turbine meter signal. If the input signal is equal to or greater than this
value, the electrical pulses received at the input will be counted. Care must be taken to ensure that the input
sensitivity is high enough to reject any electrical noise on the signal line, but not so high that fowmeter pulses
are missed. The input sensitivity of the MC-III EXP may be set to low, medium, or high when the input will
be provided by a turbine fowmeter.
If the input will be provided in the form of a pulse from a pre-amplifer or other device (rather than directly
from a turbine fowmeter), the Pulse In (pulse input) setting in the Input Sensitivity menu should be
selected.
See Section 2, page 13, for input wiring diagrams.

To Set Input Type and Sensitivity:
Enter the K-Factor menu. Press K-FACTOR MENU.
K-FACTOR
MENU
Locate the input sensitivity setting. Press ENTER, repeatedly if
necessary, to bypass calibration
factor.
SAVE
ENTER
If the input is the turbine meter, select
a sensitivity setting: low, medium, or
high.
If the input is provided by a different
device, such as a pre-amplifer, select
Pulse in (for pulse input).
Press UP ARROW to select a
sensitivity setting.
TEST
Select low, medium,
high, or Pulse In
Press ENTER. SAVING will appear
in the bottom display.
SAVE
ENTER

33
MC-III EXP Flow Analyzer Section 3
Confguring the Total Display
The Total display can be confgured for measuring volume in any of fve preprogrammed units, any
preprogrammed unit times 1,000, or a user-defned unit.
Users can specify a decimal point position ranging from 0.1 to 0.0001 of a unit.
The device will display volumes up to 10,000,000, however, a maximum of eight digits (including decimals)
are visible at a time. Therefore, once the volume (including decimals) grows beyond eight digits, digits on
the leading edge of the value will be hidden from view. To increase visibility of the whole numbers, move the
decimal one or more places to the right.
Important: Before enabling or disabling the 1000 unit option for displaying fow totals, clear the
total display to zero. If the total is not cleared, the accumulated total display will be inac-
curate. To clear the display, press LEFT ARROW (LOG) and ENTER (SAVE) keys simulta-
neously.
To Confgure the Total Display:
Enter the Display menu. Press DISPLAY MENU.
DISPLAY
MENU
Select the unit of measure in which
volume will be displayed.
Press UP ARROW until the correct
unit is displayed.
NoteIf a calculated divisor was
entered, select user-defned (no units
visible).
TEST
00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
To read the volume in terms of
thousands of units (ex. 1.0 = 1,000
bbl), continue pressing UP ARROW
until both the unit of choice and the
X1000 option are displayed.
TEST 00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
Enter the decimal point position. Press LEFT ARROW three times,
or until the decimal point in the Total
display begins blinking.
LOG
00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
.
Press UP ARROW to change the
position of the decimal point.
TEST
Press ENTER. SAVING will appear
in the bottom display.
SAVE
ENTER
34
Section 3 MC-III EXP Flow Analyzer
Confguring the Rate Display
The Rate display comprises two parts: a volume unit and a time-base unit. The volume portion of the Rate
display can be confgured in one of fve preprogrammed engineering units or in a user-defned unit (for
use with a calculated divisor). To display the volume portion in multiples of 1,000 units, continue scrolling
through the volume unit options until the 1000 entry is displayed, along with the appropriate unit of
measure.
The time-base portion of the Rate display can be confgured in one of four preprogrammed engineering units:
per day, per hour, per minute, or per second.
Users can specify a decimal point position, ranging from no decimal up to 0.001 of a unit.
To Confgure the Rate Display:
Enter the Display menu. Press DISPLAY MENU.
DISPLAY
MENU
Enter the volume factor used to
express fow rate (Ex.: bbl in bbl/day).
Press LEFT ARROW. The volume
unit of measure in the bottom portion
of the LCD window should begin
blinking.
LOG 00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
Press UP ARROW until the desired
volume unit of measure is displayed.
To read rate in terms of thousands of
units, press UP ARROW until both
the unit and X1000 are displayed.
TEST
NoteIf a calculated divisor was
entered, select user-defned (no units
visible).
Enter the time-base factor used to
express fow rate (Ex. /day in bbl/day).
Press LEFT ARROW. The time-base
factor in the bottom portion of the
LCD window should begin blinking.
LOG 00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
Press UP ARROW until the desired
time-base factor is displayed.
TEST
Enter the decimal point position. Press LEFT ARROW twice, or until
the decimal point in the Rate display
begins blinking.
LOG 00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
.
Press UP ARROW to change the
position of the decimal point.
TEST
Press ENTER. SAVING will appear
in the bottom display.
SAVE
ENTER
Caution: An overrun error occurs when the number of digits in the fow rate value exceeds the
display capacity of the device. To clear this error, reduce the number of decimal places
(for example, display rate in tenths rather than hundredths), or enable the 1000 unit
setting for the rate display, as described in the procedure above, to display rate in terms
of thousands of units.
Figure 3.1Overrun error
35
MC-III EXP Flow Analyzer Section 3
Confguring the 4-20 mA Rate Output
Caution: Before performing any 4-20 mA confguration, ensure that all peripheral equipment con-
nected to the 4-20 mA current loop is either disconnected or disabled. Confguring and
testing the 4-20 mA output feature on the MC-III EXP with the peripheral equipment in
operation may cause false alarms or erroneous operation of the peripheral device or as-
sociated equipment.
The MC-III EXP has a 4-20 mA output feature that represents fow rate. Zero and full-scale values can be
confgured to represent any fow rate range within the range of the fowmeter.
The low (or zero) setting is the fow rate value that will produce a 4-mA output. The high (or full-scale)
setting is the fow rate value that will produce a 20-mA output. Typically, the high value is greater than the
low value, as shown in Figure 3.2.

Figure 3.24-20 mA output confguration
See Analog (4-20 mA) Rate Output, page 25, for instructions on installing and wiring the 4-20 mA rate output.
Flow rates between the minimum and maximum rate setpoints will result in an output of current between
4 mA and 20 mA according to the following calculation:
[ ]
[ ]
MAX MIN
OUT CURR MIN
I I
I RATE Low I
High Low

= +

Where:
I
OUT
= output current
I
MAX
= maximum current output (20 mA)
I
MIN
= minimum current output (4 mA)
High = programmed fow rate that produces a 20-mA output
Low = programmed fow rate that produces a 4-mA output
RATE
CURR
= fow rate
36
Section 3 MC-III EXP Flow Analyzer
To Confgure the 4-20 mA Output:
Enter the Output menu. Press OUTPUT MENU.
OUTPUT
MENU
Enable or disable the 4-20 mA output. Press UP ARROW to toggle output to
on or off.
TEST
Toggles between
off and on
Press ENTER. If the 4-20 mA output
is enabled, the LCD will display the
Low fow rate and the decimal point
in the numeric display will begin
blinking.
SAVE
ENTER
Enter the decimal point position for the
Low fow rate. The words 4-20 mA
LOW should appear at the bottom of
the LCD.
Press UP ARROW to change the
position of the decimal point.
TEST
00000000
/MIN
CF
.
Enter the Low (4-mA) fow rate.
Digits are entered from right to left.
Press LEFT ARROW until the
rightmost digit in the top display
begins blinking.
LOG
00000000
/MIN
CF
Press UP ARROW to change the
digit.
TEST
Press LEFT ARROW to select the
next digit to the left.
Continue pressing UP ARROW and
LEFT ARROW alternately to enter all
remaining digits.
LOG
Press ENTER. The LCD will display
the High fow rate and the decimal
point in the numeric display will begin
blinking.
SAVE
ENTER
Enter the decimal point position for the
High fow rate. The words 4-20 mA
HIGH should appear at the bottom of
the LCD.
Press UP ARROW to change the
position of the decimal point.
TEST
00000000
/MIN
CF
.
Enter the High (20-mA) fow rate.
Digits are entered from right to left.
Press LEFT ARROW until the
rightmost digit in the top display
begins blinking.
LOG
00000000
/MIN
CF
Press UP ARROW to change the
digit.
TEST
Press LEFT ARROW to select the
next digit to the left.
Continue pressing UP ARROW and
LEFT ARROW alternately to enter all
remaining digits.
LOG
Press ENTER.
SAVE
ENTER
The Pulse Output menu prompt will appear immediately following the entry of 4-20 mA parameters. See
Confguring the Pulse Output, page 37, for the pulse output confguration procedure.
37
MC-III EXP Flow Analyzer Section 3
Confguring the Pulse Output
The pulse output feature of the MC-III EXP is typically disabled to reduce current consumption. When the
pulse output is not needed, Cameron recommends that this feature be disabled.
If the pulse output feature is required, the user will be prompted to enter a pulse output scale factor, which is
the volume increment that will cause a pulse output to occur.
If the pulse output is confgured via interface software instead of via the keypad Output menu, the user can
also specify a pulse length (pulse width), which determines the length of each output pulse in milliseconds
(ms). This parameter is confgurable only with the interface software. See Pulse Output, page 67, for more
information.
See Pulse Output, page 24, for instructions on installation and feld wiring of the pulse output feature.
To Confgure the Pulse Output:
Enter the Output menu. Press OUTPUT MENU.
OUTPUT
MENU
Locate the Pulse Output setting. Press ENTER repeatedly until the
words Pulse Output appear in the
lower display.
SAVE
ENTER
Enable or disable the pulse output. Press UP ARROW to toggle output to
on or off.
TEST
Toggles between
off and on
Press ENTER. If the pulse output is
enabled, the LCD will display Pulse
Output Scaler and the decimal point
in the numeric display will begin
blinking.
SAVE
ENTER
Enter the decimal point position for the
output scale factor.
Press UP ARROW to change the
position of the decimal point.
TEST
00000000
M
3
BBL
GAL
LIT
CF X1000
...
.
Enter the pulse output scale factor.
Digits are entered from right to left.
Press LEFT ARROW until the
rightmost digit in the top display
begins blinking.
LOG
00000000
M
3
BBL
GAL
LIT
CF X1000
...
Press UP ARROW to change the
digit.
TEST
Press LEFT ARROW to select the
next digit to the left.
Repeat using UP and LEFT arrows to
enter all remaining digits.
LOG
Press ENTER.
SAVE
ENTER
The Slave Address menu prompt will appear immediately following the entry of pulse output parameters. See
Entering the Slave Address, page 38, for information on entering the slave address.
38
Section 3 MC-III EXP Flow Analyzer
Entering the Slave Address
The slave address is a setting used in Modbus

communications. It is a number that ranges from 1 to 65535,


excluding 252 to 255 and 64764. If the Modbus

request message contains the matching address, the


device will respond to the request. In network arrangements, the device must have a unique slave address.
For more information about Modbus

communications, refer to the protocol manual section. If Modbus


communications are not used, leave the slave address at the factory setting (1).
To Enter the Slave Address:
Enter the Output menu. Press OUTPUT MENU.
OUTPUT
MENU
00000000
...
00000000
...
00000000
...
00000000
...
Locate the Slave Address setting. Press ENTER until the words Slave
Address appear in the lower display.
SAVE
ENTER
00000000
...
The rightmost digit in the top display
will begin blinking.
Enter the slave address.
(range: 1 to 65535, excluding 252 to
255 and 64764)
Press UP ARROW until the correct
digit is displayed.
TEST
Then press LEFT ARROW to select
the next digit to the left.
Repeat using UP and LEFT arrows to
enter all remaining digits.
LOG
Press ENTER.
SAVE
ENTER
The Baud Rate menu prompt will appear immediately following the entry of the slave address. See Entering
the Baud Rate below for the baud rate entry procedure.
Entering the Baud Rate
The baud rate is the number of bits per second that are on the serial port. This setting must match the setting
of the master device polling the MC-III EXP or the serial port. This only applies to the Modbus

communica-
tions; if Modbus

communications are not used, leave the baud rate at the factory setting (9600).
To Enter the Baud Rate:
Enter the Output menu. Press OUTPUT MENU.
OUTPUT
MENU
Locate the Baud Rate setting. Press ENTER repeatedly until the
words Baud Rate appear in the
lower display.
SAVE
ENTER
115200
Enter the baud rate. Press UP ARROW until the correct
baud rate is displayed.
TEST
Press ENTER. SAVING will appear
in the bottom display.
SAVE
ENTER
39
MC-III EXP Flow Analyzer Section 4
Section 4Confguration and Operation via Software
The MC-III interface software provides easy access to all confguration parameters and is the quickest
way to confgure the device. While basic parameters can be confgured with the keypad (see Section 3), the
interface software is the only means of accessing every confgurable parameter. For example, multipoint
calibration for liquid or gas measurement and gas volume correction parameters for measuring gas in standard
cubic feet are confgurable only with the interface software.
Table 4.1, page 49, provides a quick reference for confguration methods.
Installing the Software
The MC-III interface software is stored on a CD that is shipped with each instrument, and it installs directly
onto a PC or laptop. To install the interface software on a computer, perform the following steps:
1. Place the MC-III CD in your CD-ROM drive. The MC-III install screen will load automatically.
2. Click Install MC-III on the introductory screen.
3. Click on Next to initiate the installation, and follow the instructions on the screen. An MC-III icon
(Figure 4.1) will appear on the desktop of the users PC or laptop when installation is complete.
To access the interface software, click on the MC-III icon on your desktop or go the Start menu and select
Programs>NuFlo>MC-III 2.1.0>MC-III 2.1.0.
Figure 4.1MC-III desktop icon
User Documentation
Click the Help menu in the task bar at the top of the Welcome screen (Figure 4.2, page 40) to access
documentation for installing, confguring, and operating the MC-III EXP fow analyzer. This documentation
includes:
a read me fle for general reference (includes contact information for technical assistance)
an electronic copy of the user manual
an electronic copy of the quick-start guide for installing and confguring the MC-III EXP
a COM troubleshooting guide for addressing communications errors (this guide will automatically display
on screen when a communication error is detected)
40
Section 4 MC-III EXP Flow Analyzer
Connecting the MC-III to a Computer
Before attempting to run the interface software, verify that the Scanner is connected to a computer with an
RS-485 to RS-232 converter or an RS-485 to USB converter.
See Figure 2.17, page 27, for converter wiring instructions.
See Explosion-Proof Communications Adapter, page A-3, for instructions on installing and wiring an RS-
485 communications adapter.
See USB Communications Adapter, page A-6, for instructions on installing and wiring the USB commu-
nications adapter. A USB driver supplied on the MC-III software installation CD must be installed on
the computer before the USB adapter will be operational. A customer-supplied universal USB cable is
required to connect the NuFlo USB adapter to a computer.
Running the Software
Once the software is installed and the computer is connected to the MC-III, the user may proceed with
running the software and confguring the device.
To connect to the MC-III device via software, perform the following steps:
1. Click on the NuFlo MC-III icon on the desktop, or select Start>Programs>NuFlo>MC-III 2.1.0>MC-III
2.1.0. The Welcome screen (Figure 4.2) will appear and a Select COM Port window will prompt the user
to select the communications port to be used to connect with the MC-III EXP (Figure 4.2).
2. Select a port from the dropdown list and click OK. If the NuFlo USB communications adapter is in
use, the COM port will appear as NuFlo USB Adapter. The software will attempt to connect with the
MC-III EXP. This automatic launch feature is referred to as an express connect.
3. If the internal time and date programmed into the instrument differs from the time and date displayed on
the computer by more than 60 seconds, the software will prompt the user to synchronize the instrument
time/date to match the computer time/date (Figure 4.3, page 41). This ensures that the instrument displays
correct time, regardless of time zone differences.
Figure 4.2Select COM Port window
41
MC-III EXP Flow Analyzer Section 4
Figure 4.3Prompt for authorizing time and date synchronization
The time difference setting used to generate a synchronize prompt is user-confgurable. For more information,
see General Options, page B-1.
Automating Functions on Software Startup
When the software connects with the instrument, the Device Autorun Options screen appears (Figure 4.4).
From this screen, users can confgure the instrument, download logs, or upload previously saved confguration
settings.
A checkbox at the bottom of the Device Autorun Options screen allows the user to bypass the Device Autorun
Options screen and perform a selected function each time the software connects to the instrument.
Example: To download all logs on connection, select Download all logs from the Device Autorun Options
screen and check the Always do... checkbox. Each time the software computer connects to the instrument,
the software will bypass the Device Autorun Options screen, open the Flow Archive screen, and download all
logs. See Section 5, page 73, for more information on fow logs.
Figure 4.4Device Autorun Options screen
42
Section 4 MC-III EXP Flow Analyzer
Changing Autorun Settings
To change an autorun setting that is, to initiate an autorun action, to terminate an autorun action, or to
change to a different autorun actionwithout exiting to the Welcome screen, choose Options/Program
Options from the menu bar and select Autorun from the dropdown list (Figure 4.5).
Figure 4.5Options/Program Options screen for changing autorun settings
Express Connect Option
By default, the MC-III EXP interface software automatically attempts to connect to the instrument when a
user clicks on the desktop application icon. For more information on disabling the express connect feature
and changing communications options, see Express Connect Option, page B-4.
Canceling Express Connect
To temporarily cancel the express connect function without exiting the software program, click in the LCD
area of the device on the Welcome screen or choose File>Cancel Express Connect from the menu bar (Figure
4.6, page 43). The software will stop all attempts to connect to the device. The express connect function will
be restored when the user exits the program and attempts to re-open the software.
To reattempt a connection to the device after canceling the express connect function, click the LCD on the
Welcome screen again, or choose File>Express Connect from the menu bar.
43
MC-III EXP Flow Analyzer Section 4
Figure 4.6Cancel Express Connect option
Changing the Communications Port
The computer will attempt to connect to the MC-III EXP via the port that the user selects the frst time
he connects to the instrument. However, if a network dictates that the computer connect through another
communications port (actual or virtual), the user must specify the appropriate COM port. To change the COM
port setting, click on Options>Select COM Port (Figure 4.7), and select the appropriate COM port from the
list of ports detected.
Figure 4.7COM port selection menu
Software Connection in Multi-Device Network
In most cases, the automatic launch (or express connect) feature is a convenient time-saver. However, if
multiple devices are daisy-chained together in a network, this feature can interfere with the users ability to
connect to a specifc device. The MC-III allows the user to temporarily disable the express connect feature
and query the system for all connected devices. From the resulting list of devices, the user can then connect to
the appropriate slave address.
44
Section 4 MC-III EXP Flow Analyzer
To select a device from a network of instruments, perform the following:
1. From the Welcome screen, choose File>Cancel Express Connect (Figure 4.6, page 43).
2. Click File>Discover Modbus

Slaves.
3. Enter a range of addresses you want to search (Figure 4.8).
4. Enter the baud rate established for network communications.
5. Adjust the time-out setting, if necessary.
6. Click Search. The software will scan all addresses specifed and display all instruments connected.
7. Click on the appropriate address to connect to a device.
Note: In daisy-chain confgurations, verify that each device has a unique slave address.
Figure 4.8Automatic detection of all connected devices
A user who knows the address of a specifc device may also connect to it by clicking File>Connect Device at
Address (Figure 4.9).
Figure 4.9Slave address selection screen
45
MC-III EXP Flow Analyzer Section 4
Setting Log Download Preferences
The MC-III EXP automatically saves daily fow logs on the contract hour, and hourly fow logs around the
clock. The instrument also automatically creates an event record each time a user change is made. After
downloading these logs from the instrument, the instrument saves the data in a fle. By default, the logs are
saved as .csv fles.
The fle format of the auto-save fow log and event log fles is user-confgurable. Users can confgure the
instrument to download the auto-save fow and event logs as Excel (.xls) fles by selecting Options>Program
Options>Downloading from the menu bar (Figure 4.10).
Figure 4.10Log fle type selection
46
Section 4 MC-III EXP Flow Analyzer
Confguring the MC-III EXP
The MC-III EXP offers three methods for confguring parameters:
The Confguration Wizard (Figure 4.11) condenses the confguration process into nine easy-to-follow
steps (compensated gas measurement and multipoint linearization are not supported in the wizard).
The MC-III Main screen (Figure 4.12, page 47) gives users complete access to all confguration parameters.
The Advanced screen (Figure 4.13, page 47) provides access to Modbus

registers that can be modifed to


change confguration parameters (recommended only for system confgurators and host programmers).
If the desired confguration settings have been saved previously, the confguration fle can also be uploaded,
eliminating the need to re-input confguration data. See Saving and Uploading Confguration Files, page 69, for
details.
Figure 4.11Confguration Wizard (for novice users)
47
MC-III EXP Flow Analyzer Section 4
Figure 4.12MC-III Main confguration option for complete access to confgurable parameters
Figure 4.13Advanced Access confguration option for system confgurators and host programmers
48
Section 4 MC-III EXP Flow Analyzer
Confguration Wizard
The Confguration Wizard is ideal for frst-time users who want to perform a basic calibration for liquid or gas
measurement. By following the instructions on nine screens, even a frst-time user can successfully confgure
the MC-III EXP.
See Table 4.1, page 49, for a list of settings that are confgurable with the keypad, the software Confguration
Wizard, and the software MC-III Main Screen.
Important: The Confguration Wizard does not allow multipoint calibration or gas volume
correction. To select these parameters, the user must access the MC-III Main screen.
Caution: Do not use the Confguration Wizard to change individual settings. When the Confgura-
tion Wizard is used, system default settings are restored before the new settings entered
in the Calibration Wizard are written to the program. Previous confguration settings are
saved automatically in C:\NuFlo log data\MC-III\Confg Backup Files.
Using the Pencil Tool
The data entry felds in the Confguration Wizard are highly intuitive, and most can be completed by clicking
in the feld and entering new data or selecting a checkbox. For data felds that require the user to enter a value,
such as a K-factor, and designate a decimal position, a pencil icon provides a link to a data-entry window.
To enter a new value, click on the pencil icon (Figure 4.14). A new window will appear, allowing the user
to enter a new value, and designate the appropriate decimal position using a slide bar. This method ensures
that the decimal point setting and data-entry resolution displayed in the software and on the instrument LCD
correlate.
This tool is used to enter the calibration factor, fow rates for 4-20 mA output, and the scale factor for pulse
output.
Figure 4.14Pencil tool icon and data-entry window
49
MC-III EXP Flow Analyzer Section 4
Table 4.1Menus for Confguring Parameters
Confgurable Parameter Instrument
Keypad
Confguration
Wizard
Menu Screen
(accessed from
MC-III Main)
For
information,
see
Well Name Step 1 Wellsite
Information
page 57
LCD Contrast System Setup page 54
Security Setup System Setup page 55
Contract Hour Step 2 System Setup page 54
Volume Display (units,
decimal location)
Yes Step 5 Turbine Input page 58
Rate Display (units, decimal
location)
Yes Step 5 Turbine Input page 59
Input Type Yes Step 3 Turbine Input page 59
Input Sensitivity Yes Turbine Input page 59
Input Cut-off Thresholds Turbine Input page 59
Calculation Period (seconds) Turbine Input page 59
K-Factor Yes Step 4 K-Factor Entry page 60
Multipoint K-Factor K-Factor Entry page 60
Gas Volume Correction K-Factor Entry page 62
4-20 mA Output Yes Step 6 4-20 mA Output page 65
Pulse Output Yes Step 7 Pulse Output page 67
Slave Address Yes Step 8 COM Port page 55
Baud Rate Yes Step 8 COM Port page 56
Bus Delay COM Port page 57
Bus Timeout COM Port page 57
50
Section 4 MC-III EXP Flow Analyzer
MC-III Main Screen
The MC-III Main menu screen (Figure 4.15) is the confguration hub of the MC-III EXP. From the Main
screen, users can
review all current instrument settings and fow readings
check the log archive and instrument status
access all submenus for changing confgurable parameters
reset the total to zero
Figure 4.15MC-III Main screen provides direct access to all menu functions using the scrolling menu bar at
the left side of the screen.
Data Display and Updates
Each time the software polls the instrument, a small icon appears briefy in the lower left corner of the screen.
If the Live Updates checkbox in the lower right corner of the screen is checked, the instrument settings and
the LCD displays will automatically update with each poll.
Live Updates Checkbox
The Live Updates option is enabled by default. However, in the event that slow communication speeds are
used or the system times out frequently, the user can uncheck the Live Updates checkbox and periodically
query the system by clicking on the LCD graphic on the Main screen.
Reset Flow Total Button
The fow total can be reset to zero at any time by clicking on the Reset Flow Total button.
Scroll Bar
A scroll bar at the left side of the screen (Figure 4.15) gives the user easy access to any of seven confguration
menus and two log archive screens (see page 52 and page 53). Using these menus, the user can change all
confgurable parameters of the instrument. Once the instrument has been confgured, the Main screen access
should always be used for changing individual parameters while retaining all other confguration settings.
51
MC-III EXP Flow Analyzer Section 4
Buttons and Tools
Apply and OK Buttons
Changing parameters on a submenu screen involves selecting the proper screen from the scroll bar, selecting
information from dropdown menus or entering data in data felds, and saving the data using the Apply or
OK button. Apply allows the user to apply the changes without leaving the submenu screen. OK allows
the user to apply the changes and returns the user to the Main screen.
Refresh and Cancel Buttons
To update the display of settings on any submenu screen by forcing a reread of confgured parameters, press
the Refresh button. The Refresh button clears any entry that has not been applied, therefore, pressing
Refresh is also a convenient way to erase an incorrect entry. A Cancel button allows the user to close out
of a submenu screen without saving new settings and returns the user to the Main screen.
Pencil Tool for Data Entry
The data-entry felds in the MC-III Main sub-menu screens are highly intuitive, and most can be completed
by clicking in the feld and entering new data, or selecting a checkbox. For data felds that require the user to
enter a value, such as a K-factor, and designate a decimal position, a pencil icon provides a link to a data-entry
window.
To enter a new value, click on the pencil icon (Figure 4.16). A new window will appear, allowing you to enter
a new value, and designate the appropriate decimal position using a slide bar. This method ensures that the
decimal point setting and data-entry resolution displayed in the software and on the instrument LCD correlate.
This tool is used to enter the calibration factor, fow rates for 4-20 mA output, and the scale factor for pulse
output.
Figure 4.16Pencil tool icon and data-entry window
52
Section 4 MC-III EXP Flow Analyzer
Confguration Submenus (See Scroll Bar on MC-III Main Screen)

Wellsite Information
System Setup Communications Port
Turbine Input
K-Factor Entry 4-20 mA Output
53
MC-III EXP Flow Analyzer Section 4
Confguration Submenus (contd)
Pulse Output
Flow Archive
Event Archive
54
Section 4 MC-III EXP Flow Analyzer
System Setup
The frst of the submenus on the Main screen scroll barSystem Setupallows users to adjust time/date,
set the desired contract hour, enable or disable the password-protected security option, and adjust the LCD
contrast (Figure 4.17).
Figure 4.17System Setup screen
Time/Date Synchronization
The MC-III EXP includes a real-time clock for timekeeping and log time stamps. Upon connecting with
the instrument, the interface software prompts the user to synchronize the instrument time and date with
the time and date displayed on the computer if the time difference exceeds 60 seconds. To change the clock
synchronization threshold to a value other than 60 seconds, see General Options, page B-1.
A user can synchronize the instrument and computer clocks at any time by clicking the Sync to Computer
Time button on the System Setup screen.
Contract Hour
Hourly fow totals are logged in the archive on the hour. The contract hour determines the exact time when
the daily fow total is logged. The contract hour is easily adjusted by selecting the appropriate hour from the
dropdown list on the System Setup screen.
LCD Contrast Adjustment
LCD contrast is temperature-dependent and may require adjustment for optimum readout visibility. A slide
bar allows users to quickly adjust the contrast of the LCD.
55
MC-III EXP Flow Analyzer Section 4
Security Setup
Setting a keypad security code will prevent unauthorized personnel from altering calibration data or resetting
totals, and the security function is recommended to preserve data integrity of the system. The MC-III EXPs
keypad security feature is disabled at the factory.
To enable the security protection, check the Keypad Lock checkbox on the System Setup screen, and enter
a four-digit lock code in the feld below the checkbox. (Do not use 0000.) The security feature will be
activated the next time a user attempts to access the keypad.
Firmware Version Number
As frmware is upgraded, frmware version numbers change. The frmware version number shown may assist
technical support personnel in troubleshooting an operational issue.
Serial Number
The serial number of your MC-III EXP can be used to track the history of your unit to include warranty status.
Communications Port
The Communications Port screen (Figure 4.18) allows users to change the settings that are required for
Modbus

communication.
Figure 4.18Communications Port screen
Slave Address
The slave address allows the MC-III EXP to communicate with other devices via Modbus

. It is a number
that ranges from 1 to 65535, excluding 252 to 255 and 64764. If a Modbus

request message contains the


matching slave address, the device will respond to the request. In network arrangements, the slave device
must have a unique address. For more information about Modbus

communications, refer to the protocol


manual section.
56
Section 4 MC-III EXP Flow Analyzer
To change the slave address, select a number using the plus and minus buttons on the screen, or using the
page-up and page-down keys on your computer keyboard. If Modbus

communication is not used, leave the


slave address at the factory setting (1).
Baud Rate
The baud rate is the number of bits per second that are on the serial port. Baud rates supported by the MC-III
EXP range from 300 to 115.2K. This setting must match the setting of the master device polling the MC-III
EXP or the serial port. This only applies to the Modbus

communications; if Modbus

is not used, set the baud


rate to 9600.
To change the baud rate, slide the selector bar to the appropriate setting. The default setting is 9600 baud.
Note that the 9600 baud rate is available in both the low-power and high-power modes. When the instrument
is powered by batteries alone, the low-power mode is recommended.
Auto-Negotiate Option
When connected devices are capable of switching baud rates automatically or when devices are directly
connected to an MC-III EXP, a user may confgure the MC-III interface to automatically negotiate the fastest
baud rate during device connection. This negotiated baud rate does not replace the confgured baud rate,
but rather provides a temporary boost of baud rate while the computer is connected to the device for faster
downloads. After the software disconnects, the device reverts to the confgured baud rate.

By default, this function is disabled. To enable automatic baud rate negotiation on device connect, choose
Options/Program Options from the menu bar, select Communications from the dropdown options list, check
the Attempt auto-negotiation... checkbox, and click OK (Figure 4.19). To initiate an immediate search
for the fastest baud rate without exiting tto the Welcome screen, check the Enable auto-negotiated baud
rate checkbox, click the Test Now button to the right of the checkbox, and click Yes at the prompt in the
Confrm dialog box. See also Auto-Negotiate Option, page B-4.
Figure 4.19Communications Options screen
57
MC-III EXP Flow Analyzer Section 4
Bus Delay
Bus delay is the amount of time (in milliseconds) that passes before the MC-III EXP attempts to take control
of the RS-485 bus and transmit a message back to the requesting device. The MC-III EXP responds very
quickly to incoming Modbus

requestsin some cases, too quickly. A confgurable delay allows the user to
control this response time. A setting of 10 ms is normally suffcient, but may require adjustment when the
MC-III EXP is communicating with a radio or other end device that responds more slowly.
Bus Timeout
Bus timeout is the amount of time (in milliseconds) that must pass to cause the MC-III EXP to reset its
internal Modbus

message handler. The default value of 50 ms is normally suffcient.


Software Communication Options
The default timing parameters in the MC-III EXP support communication with most radios and Modbus


devices. However, if the MC-III EXP is used with a device that has a very slow response time, it may be
necessary to adjust the timing parameters to enhance performance. For information about these parameters,
see Timing Parameters, page B-4.
Wellsite Information
The Wellsite Information screen (Figure 4.20) allows users to enter information that distinguishes the wellsite,
such as company name, well name, and site location. While most of the felds on this screen are optional, a
well name should be assigned.
The well name (limited to 20 characters in length) will also serve as the name of the folder where fow logs
generated by the MC-III software will be archived. If a well name is not entered on this screen, logs will be
saved in a folder titled NO_WELLNAME.
Figure 4.20Wellsite Information screen
58
Section 4 MC-III EXP Flow Analyzer
Turbine Input
The Turbine Input screen (Figure 4.21) allows users to confgure the displays for volume and rate, select the
type of input to be used (turbine input or pulse input), determine cut-off thresholds for measuring fow, and
determine the display update frequency.
Figure 4.21Turbine Input screen
Volume Display
The Volume display can be confgured to display totals in one of fve preprogrammed engineering units or in a
user-defned unit (for use with a calculated divisor).
Important: If volume is to be measured in user-defned units, the K-Factor unit must also be
entered as pulses per unit. For more information, see Entering a Calculated Divisor, page
31.
To display the volume in multiples of 1,000 units, check the Multiplier checkbox.
The device will display volumes up to 10,000,000, however, a maximum of eight digits (including decimals)
are visible at a time. Therefore, once the volume (including decimals) grows beyond eight digits, digits on
the leading edge of the value will be hidden from view. To increase visibility of the whole numbers, move the
decimal one or more places to the right.
Important: Before enabling or disabling the 1000 unit option for displaying fow totals, clear
the total display to zero. If the total is not cleared, the accumulated total displayed will
represent a combination of two different units and will not provide an accurate fow read-
ing.
59
MC-III EXP Flow Analyzer Section 4
Rate Display
The Rate display comprises two parts: a volume unit and a time-base unit. The volume portion of the Rate
display can be confgured in one of fve preprogrammed engineering units or in a user-defned unit (for use
with a calculated divisor). To display the volume portion in multiples of 1,000 units, check the Multiplier
checkbox.
The time-base portion of the Rate display can be confgured in one of four preprogrammed engineering units:
per day, per hour, per minute, or per second. Users can specify decimal point position, from no decimal up to
0.001 of a unit, using the slide bar on the Turbine Input screen.
Important: If the volume portion of the Rate display is to be measured in user-defned units, the
K-Factor unit must also be entered as pulses per unit. For more information, see Enter-
ing a Calculated Divisor, page 31.
Caution: An overrun error occurs when the number of digits in the fow rate value exceeds the
display capacity of the device. To clear this error, reduce the number of decimal places
(for example, display rate in tenths rather than hundredths), or enable the 1000 unit
setting for the rate display using the Multiplier checkbox described above to display
rate in terms of thousands of units.
Figure 4.22Overrun error
Input Type and Sensitivity Confguration
The MC-III EXP accepts either a turbine input or a pulse input. When the turbine input is selected, the user
can also specify one of three sensitivity levels (low, medium, or high) to help minimize noise interference in
the signal reception. A low sensitivity setting will detect a signal of approximately 20 mV peak-to-peak,
while a high sensitivity setting will detect only signals that exceed 40 mV, peak-to-peak.
When a pulse input is selected, no sensitivity setting is required.
Cut-Off Thresholds
Periods of very low or interrupted fow can lead to inaccuracies in indicated fow totals. To help prevent these
inaccuracies, the user can establish cut-off thresholds to determine the minimum frequency and/or fow rate at
which a signal will be recorded.
The fow rate cut-off is a setpoint for the accumulation of fow time in the hourly and daily records, and can
be applied to either a turbine input or pulse input.
The frequency cut-off is reserved for the turbine input only. When the low-frequency cut-off is confgured, the
MC-III EXP will ignore inputs that are less than the user-entered value.
Calculation Period
The fow calculation period is the number of seconds that are allowed for calculating the fow condition. This
setting is easily adjusted with the plus and minus buttons on the screen, or the page-up and page-down
keys on your computer keyboard. Increasing this setting value results in fow rates with greater resolution.
This setting is also the time allowed for processing of display updates.
60
Section 4 MC-III EXP Flow Analyzer
K-Factor Entry
The K-Factor Entry screen (Figure 4.23) allows users to calibrate the MC-III EXP using a single calibration
factor from a turbine fowmeter or multipoint linearization. Users can also confgure the MC-III EXP to
compensate for the effect of pressure, temperature, and compressibility on gas volume measurements.
Figure 4.23K-Factor Entry screen
K-Factor Units
K-Factor entry consists of selecting the K-Factor unit and type (linear or multipoint). In selecting the unit,
users can choose from fve preprogrammed units (pulses per gallon, barrel, cubic meter, cubic foot, or liter). If
a calculated divisor is to be used, pulses per unit should be selected.
K-Factor Type
For linear calibration, the user need only select the Linear checkbox and enter the K-factor of the turbine
meter. To enter the K-factor value, click on the pencil icon to open a data-entry window (Figure 4.24). Enter
the digits, left to right, and adjust the slide bar to move the decimal point to the proper position.
Figure 4.24Data-entry screen for linear K-factor
61
MC-III EXP Flow Analyzer Section 4
For multipoint calibration, the user selects the Multipoint checkbox, and then enters the number of calibration
points he desires using the plus and minus buttons on the screen or the page-up and page-down keys on
a computer keyboard. Up to 12 calibration points may be entered. Frequency and factor data-entry felds are
automatically activated for the number of calibration points selected. The user can then enter the appropriate
data for each point.
K-Factor Backup
Each time a calibration factor is entered or changed, the software prompts the user to copy the previous
settings to backup memory (Figure 4.25). In the event that an error is made while entering confguration
parameters, the user can restore the last saved set of confguration settings to the instrument.
Figure 4.25Prompt to copy previous calibration settings to backup
Select yes to save the previous calibration factor to memory. The date and time of this backup is displayed
next to the Retrieve K-Factor Backup button on the K-Factor Entry screen (Figure 4.26, page 62). Only the
last saved backup can be retrieved.
To retrieve the last saved K-Factor backup, perform the following steps:
1. Press the Retrieve K-Factor Backup button. A confrmation screen will appear with a prompt to proceed
or terminate the retrieval.
2. Press yes to retrieve the last backup.
3. Click OK or Apply to save the retrieved settings to the confguration fle. Because the retrieval of
these settings represents a change to the current K-Factor settings, you will again be prompted to copy the
previous settings to backup.
62
Section 4 MC-III EXP Flow Analyzer
Gas Volume Correction (Supercompressibility Calculation)
Gas turbine meters are calibrated in actual cubic feet (ACF), and measure gas in actual cubic feet. In some
applications, a user may beneft from referencing gas measurements back to standard conditions by measuring
in terms of standard cubic feet (SCF). The MC-III EXP makes this process quick and easy, by using fxed
average values for the fowing gas temperature and fowing gas pressure.
By default, this option is disabled.
To access this feature, check the Enable volume correction checkbox in the Gas Volume Correction section
of the K-Factor Entry screen. The STANDARD annunciator on the LCD will illuminate, and all other Gas
Volume Correction data felds will be activated (Figure 4.26).
Figure 4.26Gas volume correction settings on K-Factor Entry screen
Atmospheric pressure is set at 14.73 psia by default, but can be adjusted to local conditions.
To compensate the gas measurement to standard conditions, complete the following steps:
1. Enter the atmospheric pressure.
2. Enter the base pressure and unit. The base pressure is the pressure to which the gas will be compensated
(typically, 14.73 psia).
3. Enter the base temperature and unit. The base temperature is the temperature to which the gas will be
compensated (typically, 60F).
4. Enter the working pressure and unit. The working pressure is the user-entered pressure that represents the
average working (fowing) pressure.
5. Enter the working temperature and unit. The working temperature is the user-entered temperature that
represents the average working (fowing) temperature.
63
MC-III EXP Flow Analyzer Section 4
6. Enter a known compressibility factor, or press Calculate compressibility from gas comp. to view a
selection of gas compositions (Figure 4.27).
To automatically calculate the compressibility factor, enter the gas composition and click on
Calculate. The software will calculate fowing gas compressibility in accordance with the
AGA-8 detail method.
The K-Factor Entry screen will reappear, and the calculated super-compressibility factor will be
displayed.
Figure 4.27Gas Composition screen
64
Section 4 MC-III EXP Flow Analyzer
4-20 mA Output
Caution: Before performing any 4-20 mA calibration, ensure that all peripheral equipment con-
nected to the 4-20 mA current loop is either disconnected or disabled. Calibrating and
testing the 4-20 mA output feature on the MC-III EXP with the peripheral equipment in
operation may cause false alarms or erroneous operation of the peripheral device or as-
sociated equipment.
The 4-20 mA Output screen (Figure 4.28) allows users to confgure the MC-III EXP to provide a 4-20 mA
output representing fow rate. Zero and full-scale values can be confgured to represent any fow rate range
within the range of the fowmeter. A test mode function is also included, allowing a user to calibrate and/or
verify the output received by an end device.
Figure 4.284-20 mA Output screen
The low (or zero) setting is the fow rate value that will produce a 4-mA output. The high (or full-scale)
setting is the fow rate value that will produce a 20-mA output. Typically, the high value is greater than the
low value.
Flow rates in-between the minimum and maximum rate setpoints will result in an output of current between
4 mA and 20 mA according to the following calculation:
[ ]
[ ]
MAX MIN
OUT CURR MIN
I I
I RATE Low I
High Low

= +

Where:
I
OUT
= output current
I
MAX
= maximum current output (20 mA)
65
MC-III EXP Flow Analyzer Section 4
I
MIN
= minimum current output (4 mA)
High = programmed fow rate that produces a 20-mA output
Low = programmed fow rate that produces a 4-mA output
RATE
CURR
= fow rate
Figure 4.294-20 mA output confguration
Enabling 4-20 mA Output
By default, the 4-20 mA output option is disabled. To enable this feature, perform the following steps:
1. Check the Enable checkbox.
2. Click on the pencil icon next to the Low Flow Rate feld to launch a data-entry window (Figure 4.30) and
enter the low or zero fow rate value that will produce a 4-mA output. Adjust the decimal position,
if necessary, using the slide bar. A fow rate less than or equal to this minimum setpoint will result in an
output of 4 mA.
3. Click on the pencil icon next to the High Flow Rate feld and enter the high or full-scale fow rate
value that will produce a 20-mA output. Adjust the decimal position, if necessary, using the slide bar. A
fow rate equal to or greater than this maximum setpoint will result in an output of 20 mA.
Figure 4.30Data-entry screen for 4-20 mA low and high values
4-20 mA Output Testing
The MC-III software allows users to test the 4-20 mA output by entering an output current value and reading
the live value generated in response.
66
Section 4 MC-III EXP Flow Analyzer
To use the test mode, press the Test 4-20 mA Output button at the bottom of the 4-20 mA Output screen.
The test page will appear (Figure 4.31), and the 4-20 mA Output option will be temporarily disabled.
To test the output, enter a value in the Output Current (mA) feld, and press Set Output.
Figure 4.314-20 mA output test screen
67
MC-III EXP Flow Analyzer Section 4
Pulse Output
The Pulse Output screen (Figure 4.32) allows users to confgure the MC-III EXP to provide a pulse output
representing increments in volume. A test mode function is also included, allowing a user to calibrate and/or
verify the output received by an end device.
Figure 4.32Pulse Output screen
Confguring Pulse Output
By default, the pulse output option is disabled to reduce current consumption. To conserve power, do not
enable this feature unless it will be used.
To enable the pulse output option, perform the following steps:
1. Check the Enable checkbox.
2. Enter the pulse duration value using the plus and minus buttons on the screen, or the page-up and
page-down keys on a computer keyboard. The pulse length (pulse width) determines the length of each
output pulse in milliseconds (ms).
3. Click on the pencil icon next to the scale feld to open a data-entry window (Figure 4.33, page 68), and
enter the scale factor. Adjust the decimal position, if necessary, using the slide bar. The scale factor sets
the volume increment that will cause a pulse output to occur. The unit for this factor is determined by the
unit set for the volume display.
NoteThe scale factor can be any number; it is not limited to powers of 10 (1, 10, 100, etc.).
In applications where high fow rates may occur for extended periods, pulses can be lost due to the
instruments inability to register the fow rate at the maximum output frequency. To avoid the loss of pulses,
users can raise the scale factor or shorten the pulse duration. The software calculates the maximum fow
rate that can be reached before pulse caching is likely to occur, and displays the information on the Pulse
68
Section 4 MC-III EXP Flow Analyzer
Output screen. In applications where short bursts of high fow rates are common (separators, tank loading,
dump cycles, etc.), the displayed fow rate can exceed the maximum fow rate for short periods without
compromising the accuracy of the total represented by the pulse output.
Figure 4.33Data-entry screen for pulse output scale factor
Pulse Output Testing
The MC-III software allows users to test the pulse output by writing a value to the pulse accumulator. Such
tests can aid in troubleshooting problems with hardware compatibility, threshold settings, and pulse duration
settings.
To use the test mode, press the Test Pulse Output button at the bottom of the Pulse Output screen. The test
page will appear (Figure 4.34), and the Pulse Output option will be temporarily disabled.
To test the output, enter a value in the Pulse Count feld, and press Set Output. The value in the Live Value
feld will decrement to zero in accordance with the programmed pulse duration.
Figure 4.34Pulse output test screen
69
MC-III EXP Flow Analyzer Section 4
Saving and Uploading Confguration Files
The MC-III EXP allows a user to save confguration settings in a fle that is stored on the users computer and
can be uploaded to the MC-III EXP as needed. Confguration fles are easily identifed by the .mc3 extension.
The default flename begins with confg and uses the following structure:
File Format Filename Structure
.mc3 confg<date (YYMMDD)> <time (hhmmss)>.mc3
Example: confg050209133922.mc3 is a confguration fle that was
downloaded on February 9, 2005 at 1:39:22 p.m.
Saving a Confguration File
To save current confguration settings for future reference, perform the following steps:
1. Select File>Save Confguration File (Figure 4.35). A Save MC3 Confguration File screen (Figure 4.36)
will appear, prompting the user to save the fle to the computer hard drive. The default directory is C:\Nu-
Flo log data\MC-III\<WELL NAME>, however the user can specify a new location, if desired.
Figure 4.35File dropdown menu allows user to save a confguration fle
Figure 4.36Default directory for confguration fles
70
Section 4 MC-III EXP Flow Analyzer
2. Click Save. An Information window will appear to confrm that the fle has been saved.
The wellname folder will bear the well name that appears on the Wellsite Information screen. If no well
name is entered on this screen, the folder will be named NO_WELLNAME (Figure 4.36, page 69).
Uploading a Confguration File
To upload a confguration fle, perform the following steps:
1. From the Main screen, select File>Return to Welcome Screen (Figure 4.37).
2. Click on the LCD to connect to the MC-III EXP.
3. Click Upload a confguration (Figure 4.38, page 71) and click Do It Now. A Load MC3 Confguration
File window will appear (Figure 4.39, page 71), displaying the contents of the MC-III folder.
4. Click on the appropriate well name folder, select the confguration fle of your choice, and click Open.
A Progress window will appear momentarily, followed by an Information window confrming the
upload.
Figure 4.37File dropdown menu allows user to return to the Welcome screen
71
MC-III EXP Flow Analyzer Section 4
Figure 4.38Upload a Confguration option
Figure 4.39Default directory for confguration fles
72
Section 4 MC-III EXP Flow Analyzer
Advanced Access
System confgurators and host programmers can also access the software through the Advanced Access menu
(Figure 4.40). This option provides access to the Modbus

registers for the MC-III EXP. A series of tabs on


the right side of the screen (real time, input confguration, output confguration...) allows the user to view a
specifc group of registers, or to view all registers in one list.
Important: The Advanced Access option is not required for the routine operation of the MC-III EXP,
and novice users should not access this portion of the software unless instructed to do
so by Cameron technical support personnel.
Data retrieved from the MC-III EXP while in Advanced access mode can be logged to a fle by enabling the
Automatically log data polls checkbox on the Advanced Options screen (Figure B.6, page B-6). The fle will
include values that are retrieved from the instrument when the user clicks the Get All Data button or enables
Auto Polling on the Advanced Access screen. See also Automatic Data Logging, page B-6.
Figure 4.40Advanced Access screen

73
MC-III EXP Flow Analyzer Section 5
Section 5Flow Logs and Event Logs
The MC-III EXPs fow archive expands the users ability to track fow volume over time by allowing the
user to view fow data in tabular and trend formats, to save or print log data in reports, and to export log data
into a spreadsheet. Additionally, an event archive allows users to identify changes to confguration parameters
and other settings for troubleshooting operational issues.
A user can download logs from the Flow Archive and Event Archive screens as needed, or confgure the
software to automatically download the logs upon connection with the instrument. See Automating Functions
on Software Startup, page 41, for more information.
When the download all logs or download only new logs option is selected from the Device Autorun
Options menu and the Always do... checkbox is checked, the software will download both fow logs and
event logs each time the software connects to the MC-III EXP (Figure 5.1).
Figure 5.1Device Autorun Options screen allows automated log downloads with every connection
Auto-Save Log Formats
The MC-III EXP automatically saves daily fow logs on the contract hour, and saves hourly fow logs around
the clock. The instrument also automatically saves an event log each time a user change is made. After
downloading these logs from the instrument, the software saves the data in a fle. By default, the logs are
saved as .csv fles.
The fle format of the auto-save fow log and event log fles is user-confgurable. To save logs in Excel (.xls)
format, click Options>Program Options in the task bar, select Downloading from the dropdown options list,
and check the XLS-MS Excel File checkbox (Figure 5.2, page 74).
74
Section 5 MC-III EXP Flow Analyzer
Figure 5.2Log download options
Log Directory and Filenames
Auto-save log fles are saved in C:\NuFlo log data\MC-III\<WELL NAME>. The wellname folder will bear
the well name that appears on the Wellsite Information screen. If no well name is entered on this screen, the
folder will be named NO_WELLNAME.
Log flenames incorporate the serial number of the MC-III EXP, the date, the time, and a capital letter
indicating the log type, as shown below.
File Format Filename Structure
.xls or .csv <Serial number>_<date (YYYYMMDD)>_<time (hhmmss)>_<log
type indicator> where the log type indicator is
H for hourly, D for daily, or E for event
Example: 16_20050302_102031_H.xls is an hourly log fle from the
MC-III EXP serial number 16 that was downloaded on March 2, 2005
at 10:20:31 AM.
Flow Archive
The MC-III EXP automatically saves daily fow logs on the contract hour, and hourly fow logs. The Flow
Archive screen (Figure 5.3, page 75) allows users to download, view, and print trend charts, and export daily
and hourly logs. On the Flow Archive screen, users can view up to 384 daily logs and 768 hourly logs in
tabular format (note the Daily Logs and Hourly Logs tabs at the bottom of the screen), or in a trend chart (see
Daily Trend and Hourly Trend tabs).
75
MC-III EXP Flow Analyzer Section 5
Figure 5.3Flow Archive screen
Downloading Flow Logs
To download fow logs, click the Download button on the Flow Archive screen, and select either Download
All Flow Logs or Download Only New Flow Logs (Figure 5.4). (Download Only New Flow Logs will display
only fow logs that have been created since the last download was performed.) Tabular views of the archive
record (Figure 5.5, page 76) include a time stamp showing the exact time each log was recorded, as well as
fow volume, run time, and supply voltage. Trend views (Figure 5.6, page 77) show the fow volume and time
stamp only.
Figure 5.4Menu for downloading fow logs from the Flow Archive screen
76
Section 5 MC-III EXP Flow Analyzer
Figure 5.5Downloaded fow logs (tabular view)
Viewing Trend Charts
In the daily and hourly trend views, fow volumes are charted on a grid, with fow record numbers forming
the horizontal axis, and fow volumes shown on the vertical axis. A graphical line defned by two yellow
endpoints represents the range of archived logs available for viewing. A yellow triangle marks the frst record,
and a yellow circle marks the fnal record. The arrowhead cursor is paired with dotted circle. As the user drags
the cursor across the chart, the circles position corresponds with the record displayed at the top of the screen
(Figure 5.6, page 77). The circles snap function allows the user to pinpoint transitional points within the trend
display; the greater the zoom, the more precise the navigation.
NoteThe interactive cursor cannot be turned off.
To zoom in, left-click and drag the mouse to the right and down.
To zoom out, drag the mouse up and to the left.
To reposition the chart within the viewing window, right-click the chart and move the mouse slowly in
any direction.
To exit the page and return to the Main screen, press OK.
77
MC-III EXP Flow Analyzer Section 5
Figure 5.6Flow Archive screen, trend view
Printing/Saving a Report
A fow log can be printed and/or saved as a table (Figure 5.7) or a trend chart (Figure 5.8, page 78).
To print a log, select the appropriate tab at the bottom of the Flow Archive screen, and press the Print
button in the upper right corner of the Flow Archive screen. (If the display is tabular, the button will read
Print Table; if the display is a chart, the button will read Print Chart.) A Print Preview screen will appear,
displaying the image to be printed.
To print the report, select File>Print from the task bar (Figure 5.7).
Figure 5.7Print preview, tabular view
78
Section 5 MC-III EXP Flow Analyzer
Figure 5.8Print preview, trend view
To save the report, perform the following steps:
1. Select File>Save from the task bar (Figure 5.8). A Save Report As window will open (Figure 5.9).
2. Enter a flename. The flename extension will be .rps.
3. Click Save. By default, log archive reports are saved in C:\NuFlo log data\MC-III\<WELL NAME>.
However, the user can specify a new location, if desired. The wellname folder will bear the well name
that appears on the Wellsite Information screen. If no well name is entered on this screen, the folder will
be named NO_WELLNAME.
Figure 5.9Save Report As screen
79
MC-III EXP Flow Analyzer Section 5
Viewing a Saved Report
To view a saved report, perform the following steps:
1. From the Flow Archive screen, click Print Table (or Print Chart for trend chart views) in the upper
right corner of the screen.
2. Select File>Load from the task bar at the top of the page (Figure 5.10). A Load Report window will open
(Figure 5.11).
3. Click on the report you wish to view. A preview of the report will appear in the right portion of the win-
dow.
4. Click Open.
Figure 5.10Menu for loading a saved report
Figure 5.11Load Report screen for viewing saved fow log reports
80
Section 5 MC-III EXP Flow Analyzer
Exporting Flow Logs
Flow logs can be directly exported to an .xls or .csv fle for ease in viewing and distributing. To export a fle,
perform the following steps:
1. Click the Export Data button near the top of the Flow Archive screen and select the .xls or .csv format
(Figure 5.12).
2. When the Export Daily (or Hourly) Log Data... window appears, click Save. By default, exported
logs are saved in C:\NuFlo log data\MC-III\<WELL NAME>, however the user can specify a new loca-
tion, if desired (Figure 5.13, page 81). The wellname folder will bear the well name that appears on the
Wellsite Information screen. If no well name is entered on this screen, the folder will be named NO_
WELLNAME.
Important: Users can also specify the fle format to be used for all auto-save logs (.csv or .xls). To
change the auto-save format, follow the procedure described in Downloading Options,
page B-5.
Figure 5.12Dropdown menu for exporting log data
81
MC-III EXP Flow Analyzer Section 5
Figure 5.13Default directory for exported log fles
Event Archive
The Event Archive screen (Figure 5.14) allows users to download, view, export, and print up to 345 user event
logs. Event logs are generated to track user changes such as K-Factor changes, input setting changes, power-
on and watch-dog resets, and the fow and frequency cut-off settings.
Figure 5.14Event Archive screen
82
Section 5 MC-III EXP Flow Analyzer
Downloading Event Logs
To download event logs, click the Download button on the Event Archive screen, and select either
Download All Event Logs or Download Only New Event Logs, (Figure 5.15). Download Only New Event
Logs will display only fow logs that have been created since the last download was performed. Tabular views
of the archive record (Figure 5.16) include a time stamp showing the exact time each log was recorded, as
well as the name of the register changed, and the old and new values assigned to that register.
Figure 5.15Menu for downloading event logs from Event Archive screen
Figure 5.16Downloaded event logs
83
MC-III EXP Flow Analyzer Section 5
Printing/Saving a Report
To print an event log, press the Print Table button in the upper right corner of the Event Archive screen. A
Print Preview screen will appear, displaying the image to be printed.
To print the report, select File>Print from the task bar.
To save the report, perform the following steps:
1. Select File>Save from the task bar. A Save Report As window will open.
2. Enter a flename. The flename extension will be .rps.
3. Click Save. By default, log archive reports are saved in C:\NuFlo log data\MC-III\<WELL NAME>,
however the user can specify a new location, if desired.The wellname folder will bear the well name
that appears on the Wellsite Information screen. If no well name is entered on this screen, the folder will
be named NO_WELLNAME.
The procedure for printing and saving an event log report is nearly identical to the procedure for printing and
saving a fow log report. See the fow log screen captures, pages 77 and 78, for reference.
Exporting Event Logs
Event logs can be directly exported to an .xls or .csv fle for ease in viewing and distributing. To export a fle,
perform the following steps:
1. Click the Export Data button near the top of the Event Archive screen and select the .xls or .csv format
(Figure 5.17, page 84.
2. When the Export Event Logs... window appears, click Save. By default, exported logs are saved in
C:\NuFlo log data\MC-III\<WELL NAME>, however the user can specify a new location, if desired. The
wellname folder will bear the well name that appears on the Wellsite Information screen. If no well
name is entered on this screen, the folder will be named NO_WELLNAME.
Important: Users can also specify the fle format to be used for all auto-save logs (.csv or .xls). To
change the auto-save format, follow the procedure described in Downloading Options,
page B-5.
84
Section 5 MC-III EXP Flow Analyzer
Figure 5.17Dropdown menu for exporting log data
85
MC-III EXP Flow Analyzer Section 6
Section 6 - MC-III EXP Maintenance
The MC-III EXP is designed to provide many years of service with minimal maintenance. Batteries require
periodic replacement, and battery life depends on whether battery power is the primary or secondary power
source.
All confguration settings are stored in nonvolatile memory; therefore, confguration settings will not be lost
in the event of battery failure.
The circuit assembly or keypad may also require replacement over the life of the instrument. Procedures are
provided in this section.
Lithium Battery Replacement
The MC-III EXP uses a lithium battery with a life expectancy of approximately 2 years. Due to the fat
discharge characteristics of the lithium battery, it is diffcult to determine how much life remains in a battery
at any given time. To preserve confguration and accumulated volume data, replace the battery at 2-year
intervals.
!
WARNING: CSA-APPROVED DEVICE. To prevent ignition of hazardous atmospheres, do not
remove the cover unless the area is void of combustible gas and vapors. Replace the battery
pack only with battery part number 9A-100005111.
!
WARNING: ATEX-APPROVED DEVICE. To prevent ignition of hazardous atmospheres, do not
remove the cover unless the area is void of combustible gas and vapors. Replace the battery
pack only with battery part number 9A-90099000.
!
WARNING: The lithium battery that powers the MC-III EXP is a sealed unit; however, should a
lithium battery develop a leak, toxic fumes could escape upon opening the enclosure. Ensure
that the instrument is in a well-ventilated area before opening the enclosure to avoid breathing
fumes trapped inside the enclosure. Exercise caution in handling and disposing of spent or
damaged batteries. See additional information on lithium batteries in Appendix C, page C-1 of this
manual.
Important: Press the ENTER/SAVE key on the keypad before replacing the lithium battery to save
the accumulated total to nonvolatile memory. Once the battery is replaced and power is
restored to the unit, the last saved accumulated total will be displayed in the LCD.
The lithium battery is secured inside the enclosure by a velcro strap and connected to a connector (J1) near the
top of the circuit assembly.
To replace a lithium battery in the MC-III EXP, perform the following steps:
1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclo-
sure.
2. Using a small standard blade screwdriver, remove the two #4-40 7/8 screws located to the right and
left side of the display (Figure 6.1, page 86).
86
Section 6 MC-III EXP Flow Analyzer
Remove screws
to access battery
compartment
Figure 6.1 Removal of faceplate/keypad assembly
3. Lift the display/keypad assembly from the enclosure, making sure the circuit assembly does not contact
the enclosure.
4. Loosen the velcro strap, disconnect the battery from the J1 connector on the circuit assembly, and remove
the battery from the enclosure (Figure 6.2).
Figure 6.2 Removal/replacement of lithium battery
87
MC-III EXP Flow Analyzer Section 6
5. Install the new battery in the enclosure in the same position as the original battery, and secure the Velcro
tightly around the battery.
6. Connect the replacement battery to the J1 connector.
7. Place the circuit assembly over the standoffs and fasten with the two #4-40 7/8 screws, ensuring that
all connector wiring is inside the enclosure.
8. Replace the enclosure cover, threading it onto the enclosure in a clockwise direction.
Important: The interruption of power to the MC-III EXP will cause the internal clock time to be inac-
curate. Reset the time via the interactive software. See Time/Date Synchronization, page 54.
Alkaline Battery Replacement
The alkaline battery pack will require periodic replacement. The dimming of the LCD may be an indicator
that battery power is diminishing, and the battery pack should be replaced.
!
WARNING: To prevent ignition of hazardous atmospheres, do not remove the cover while
circuits are alive. Under normal conditions, the MC-III EXP poses no hazard when opened in a
safe area. Replace the battery pack only with battery pack part number 9A-30099007.
Important: Press the ENTER/SAVE key on the keypad before replacing the alkaline battery to save
the accumulated total to nonvolatile memory. Once the battery is replaced and power is
restored to the unit, the last saved accumulated total will be displayed in the LCD.
The shrink-wrapped alkaline battery pack is secured inside the MC-III EXP enclosure by a Velcro strap and
connected to a connector (J1) near the top of the circuit assembly.
To replace the battery pack, perform the following steps:
1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclo-
sure.
2. Using a small standard blade screwdriver, remove the two #4-40 7/8 screws located to the right and
left side of the display (Figure 6.3, page 88).
88
Section 6 MC-III EXP Flow Analyzer
Remove screws
to access battery
compartment
Figure 6.3Removal of faceplate/keypad assembly
3. Lift the display/keypad assembly from the enclosure.
4. Loosen the Velcro strap, disconnect it from the J1 connector on the circuit assembly, and remove the bat-
tery from the enclosure (Figure 6.4).
Figure 6.4Removal/replacement of alkaline battery pack
89
MC-III EXP Flow Analyzer Section 6
5. Install the new shrink-wrapped battery pack in the same position as the original battery pack as shown in
Figure 6.4, page 88 and secure the Velcro strap to hold the pack tightly in place.
6. Connect the replacement battery pack to the J1 connector.
7. Place the circuit assembly over the standoffs and fasten with the two #4-40 7/8 screws, ensuring that
all connector wiring is inside the enclosure.
8. Replace the enclosure cover, threading it onto the enclosure in a clockwise direction.
Important: The interruption of power to the MC-III EXP will cause the internal clock time to be inac-
curate. Reset the time via the interactive software. See Time/Date Synchronization, page 54.
Circuit Assembly Replacement
!
WARNING: To prevent ignition of hazardous atmospheres, do not remove the cover while
circuits are alive. Under normal conditions, the MC-III EXP poses no hazard when opened in a
safe area.
Important: Static electricity can damage a circuit board. Handle new boards only by their edges,
and use proper anti-static techniques (such as wearing anti-static wrist strap or touching
metal to establish an earth ground) prior to handling a board.
Important: If possible, record the accumulated total and all confguration settings before replacing
the circuit board. This information can be recorded by hand or captured by saving a con-
fguration fle that can be reloaded into the unit after the circuit board is replaced. (See
Saving and Uploading Confguration Files, page 69)
To replace the circuit assembly, perform the following steps:
1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclo-
sure.
2. Using a small standard blade screwdriver, remove the two #4-40 7/8 screws located to the right and
left side of the display.
3. Lift the display/keypad assembly from the enclosure.
4. Record the locations of all cable connections to the circuit assembly.
5. Using a small standard blade screwdriver, remove all wiring from terminal blocks TB1, TB2, and TB3,
and J2, ensuring that all wiring that is connected to powered circuits is insulated with tape.
6. Unplug the battery cable from connector J1 on the circuit assembly.
7. Remove the two #4-40 5/16 screws fastening the circuit assembly to the keypad assembly (Figure 6.5,
page 90.
90
Section 6 MC-III EXP Flow Analyzer

Figure 6.5Disassembly of circuit board/keypad assembly
8. Disconnect the keypad ribbon cable from the J3 connector on the LCD side of the circuit assembly as fol-
lows:
a. Grasp the black clip between a thumb and forefnger (see white arrows in Figure 6.6 below).
b. Squeeze both sides of the clip and gently pull to release the clip from the plastic connector that holds
it in place. DO NOT PULL on the ribbon cable. When the black plastic clip is properly disengaged,
the cable will release freely.
Figure 6.6To release the ribbon cable from the connector, press in on the side tabs of the connector (white
arrows) and gently pull forward (black arrow).
91
MC-III EXP Flow Analyzer Section 6
9. Remove the original circuit assembly from the enclosure.
10. Connect the ribbon cable of the keypad to the J3 connector on the LCD side of the circuit assembly as
follows:
a. Insert the end of the ribbon cable into the plastic clip.
b. While holding the ribbon cable in place, press the black plastic clip into the connector until it snaps.
11. Mount the circuit assembly to the keypad with the two #4-40 5/16 screws.
12. Reconnect the battery cable to J1 on the circuit assembly.
13. Reconnect all wiring to terminal blocks TB1, TB2 and TB3 (and J2, if applicable).
14. Mount the display/keypad assembly to the enclosure with the two #4-40 7/8 screws.
15. Recalibrate the MC-III EXP, then replace the enclosure cover.
16. Re-establish power to the peripheral circuitry.
Keypad Replacement
!
WARNING: To prevent ignition of hazardous atmospheres, do not remove the cover while
circuits are alive. Under normal conditions, the MC-III EXP poses no hazard when opened in a
safe area.
To replace the keypad of the MC-III EXP, perform the following steps:
1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclo-
sure.
2. Using a small standard blade screwdriver, remove the two #4-40 7/8 screws located to the right and
left side of the display.
3. Lift the display/keypad assembly from the enclosure.
4. Remove the two #4-40 5/16 screws fastening the circuit assembly to the keypad (Figure 6.5, page 90.
5. Disconnect the keypad ribbon cable from the J3 connector on the LCD side of the circuit assembly as fol-
lows:
a. Grasp the black clip between a thumb and forefnger (see white arrows in Figure 6.6, page 90.
b. Squeeze both sides of the clip and gently pull to release the clip from the plastic connector that holds
it in place. DO NOT PULL on the ribbon cable. When the black plastic clip is properly disengaged,
the cable will release freely.
6. Remove the circuit assembly.
7. Connect the ribbon cable of the keypad to the J3 connector on the LCD side of the circuit assembly as
follows:
a. Insert the end of the ribbon cable into the plastic clip.
b. While holding the ribbon cable in place, press the black plastic clip into the connector until it snaps.
8. Mount the circuit assembly to the keypad with the two #4-40 5/16 screws.
92
Section 6 MC-III EXP Flow Analyzer
9. Mount the display/keypad assembly to the enclosure with the two #4-40 7/8 screws.
10. Recalibrate the MC-III EXP if necessary and replace the enclosure cover.
Firmware Update
Before installing new frmware or moving the instrument to a new site, it is recommended that the user
clear the EEPROM, which will erase all confguration settings and log data. To clear the EEPROM, choose
Options/Program Options from the menu bar, select Advanced from the dropdown options list, and click on
the Clear EEPROM button (Figure 6.7).
Technical assistance is typically required for upgrading frmware. Contact Camerons Measurement Systems
Division to arrange an upgrade.
!
WARNING: Clicking the Clear EEPROM button will delete all confguration settings and log
records in the MC-III EXP memory.
Figure 6.7EEPROM clearing and automatic data log option

93
MC-III EXP Flow Analyzer Section 6
MC-III Parts List (CSA)
CAUTION: EXPLOSION HAZARDSUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY
FOR CLASS I, DIVISION 1 IN THE UNITED STATES AND CANADA. USE OF SPARE PARTS
OTHER THAN THOSE IDENTIFIED BY CAMERON VOIDS CSA CERTIFICATION. CAM-
ERON BEARS NO LEGAL RESPONSIBILITY FOR THE PERFORMANCE OF A PRODUCT
THAT HAS BEEN SERVICED OR REPAIRED WITH PARTS THAT ARE NOT AUTHORIZED
BY CAMERON.
Quantity Part Number Description
1 9A-30166003 Assembly, Switchplate
1 9A-50160002 Circuit Assembly
1 9A-90017002 Cable Assembly, Flowmeter
1 9A-100005111 Battery - Lithium, 3.6 V (CSA-approved units
only)
1 9A-50099002 Battery Pack, Alkaline (3 C), Shrink-wrapped
(CSA-approved units only)
1 9A-100002605 Desiccant Packet
1 9A-30054001 Assembly, Explosion-Proof Reset/Control
Switch, with Extension, Fits in. Female Pipe
Thread, for Standard Enclosure, CSA
1 9A-50053008 Assembly, Reset Switch, Explosion-Proof, for
Alternate Enclosure (Red Rim on Cover), CSA
1 9A-50053009 IS Barrier, RS-485, CSA
1 9A-90017004 Cable Assembly, 3/4 in. NPT Explosion-Proof
Union, 2-Pin Connector, 10 in., for External
RS-485 Communications, CSA
1 2296692-01 Kit, Cable Assembly, USB, Software Driver CD
1 9A-101283116 Converter, RS-485 to RS-232, 9-Pin, with 20-ft
Cable
1 2296650-01 Converter, RS-485 to USB, Port-Powered, with
20-ft Cable
1 9A-100017622 Union, 1-in.
1 9A-30028004 Kit, Pole Mount, Plated Steel
1 9A-30028005 Kit, Pole Mount, Stainless Steel
1 9A-50074001 Assembly, Installation Software CD and CD
Pocket Folder, MC-III
1 9A-50165003 Manual, User
1 2296694-01 Manual, Quick-Start
1 9A-50053010 Control Drawing, Intrinsically Safe RS-485
Connection, MC-III EXP
94
Section 6 MC-III EXP Flow Analyzer
MC-III Parts List (ATEX)
CAUTION: EXPLOSION HAZARDSUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY
FOR ATEX AND IECEX APPROVALS. USE OF SPARE PARTS OTHER THAN THOSE
IDENTIFIED BY CAMERON VOIDS ATEX/IECEX CERTIFICATION. CAMERON BEARS NO
LEGAL RESPONSIBILITY FOR THE PERFORMANCE OF A PRODUCT THAT HAS BEEN
SERVICED OR REPAIRED WITH PARTS THAT ARE NOT AUTHORIZED BY CAMERON.
Quantity Part Number Description
1 9A-30166003 Assembly, Switchplate
1 9A-50160002 Circuit Assembly
1 9A-90017005 Cable Assembly, Flowmeter, ATEX
1 9A-90099000 Battery, Lithium, Intrinsically Safe, Current
Limited, ATEX
1 9A-100002605 Desiccant Packet
1 9A-30054002 Assembly, External Explosion-Proof Reset/
Control Switch, with Extension, Fits 3/4 in.
Female Pipe Thread, ATEX
1 9A-90017008 Cable Assembly, 3/4 in. NPT Explosion-Proof
Brass Union, 2-Pin Connector, 10 in., for
External RS-485 Communications, ATEX
1 9A-101283116 Converter, RS-485 to RS-232, 9-Pin, with 20-ft
Cable
1 9A-30025001 Tube, Standoff, Stainless Steel, for Barton 7000
Series Turbine Meters,1.18 in. Hex X 3.56 in.
long with 3/4 in. NPT Male & Female Ends,
ATEX *
1 9A-30025006 Tube, Standoff, Stainless Steel, for NuFlo
Turbine Meters, 4.00 in. long with 3/4 in. NPT
Male & 1 in. NPT Female Ends, ATEX *
1 9A-30028004 Kit, Pole Mount, Plated Steel
1 9A-30028005 Kit, Pole Mount, Stainless Steel
1 9A-50074001 Assembly, Installation Software CD and CD
Pocket Folder, MC-III
1 9A-50165003 Manual, User
1 2296694-01 Manual, Quick-Start
* Standoff tube adapters are available in lengths up to 18 inches on request.
A-1
MC-III EXP Flow Analyzer Appendix A
Appendix AMC-III EXP Hardware Options
Explosion-Proof Reset/Control Switch
An external explosion-proof reset/control switch (Figure A.1) allows the user to zero the total and manually
select which parameter is displayed on the LCD and view daily logs instantaneously without removing the
instrument cover or connecting the instrument to a laptop computer.
Caution: This switch is available in two models. The CSA-approved model has a stainless steel
adapter, and the ATEX-approved model has a brass adapter.
Figure A.1Explosion-proof reset/control switch (CSA-approved stainless steel adapter shown)
The reset/control switch assembly mounts in either threaded conduit opening in the instrument housing.
The CSA-approved assembly is available in two variations for use with the standard enclosure shown above,
and the alternate enclosure shown in Figure 2.6, page 18. The alternate enclosure is identifed by a red ring
around the face of the cover. Both parts are included in the MC-III Parts List (CSA), page 93.
Installation
If the switch is ordered with an MC-III, it will be installed prior to shipment.
To add a switch to an existing MC-III, connect the leads to connector J2 on the main circuit board (Figure
A.2, page A-2).
A-2
Appendix A MC-III EXP Flow Analyzer
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-3
0
V
D
C
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
Remote switch
connector
(Part No. 9A-50053008)
Figure A.2Explosion-proof switch wiring
Operation
Zeroing the Total. To reset the volume total with an explosion-proof switch, press and hold the switch for at
least 3 seconds, then release it.
Important: When resetting the volume total with the explosion-proof reset switch, do not release the
switch prematurely. Releasing the switch too quickly (before 3 seconds have passed) will
activate the daily log view function and WILL NOT reset the volume to zero.
Viewing Daily Logs. To view daily logs with the explosion-proof switch, press and release the switch. (Do not
depress the switch for more than 1 second.) The words Daily Volume Archive will scroll across the bottom
of the LCD and the day index (Figure A.3) will display 01. The daily fow total recorded at the last contract
hour will appear at the top of the LCD.

The index number represents the number of days previous to the current date. For example, totals from two
days previous are read by pressing the switch until 02 appears in the Day index feld of the display.
Subsequent press-and-release actions increment the daily index, allowing the user to select a specifc daily
total.
00000000
000000
M
3
BBL
GAL
LIT
M
3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
00
Day
Total
Rate
Figure A.3Daily index (Day) displays
A-3
MC-III EXP Flow Analyzer Appendix A
Additionally, when the log viewing function times out, the accumulated total is automatically saved to
memory.
Saving the Total. When the daily log display times out, the MC-III EXP will automatically save the
accumulated total.
Explosion-Proof Communications Adapter
The explosion-proof communications adapter provides an RS-485 connection for connecting a laptop or PC to
the instrument without removing the instrument cover. When the adapter is ordered with an MC-III EXP, it is
factory installed. The installed adapter is comprised of an RS-485 adapter socket, a blanking plug, and a union
nut (Figure A.4). It may be relocated to either conduit opening in the instrument housing.
Caution: This adapter is available in two modelsa plated steel CSA-approved model and a brass
ATEX-approved model.
An RS-232 to RS-485 converter cable (available from Camerons Measurement Systems Division) is required
for connecting the adapter to a laptop computer. See MC-III Parts List (CSA), page 93, and MC-III Parts List
(ATEX), page 94, for converter cable options.

Adapter socket
Union nut
Blanking plug

Figure A.4Components of explosion-proof communications adapter
Assembly of the Plug Connector
A plug connector that mates with the RS-485 adapter socket is shipped with the device (uninstalled). This
plug connector should be wired to an RS-485 converter cable as shown in Figure A.5, page A-4, and stored with
the cable when the COM adapter is not in use.
A-4
Appendix A MC-III EXP Flow Analyzer
PLUG CONNECTOR
(LEFT AND RIGHT WIRE POSITIONS ARE VALID ONLY
WHEN PLUG IS FACING SOCKET IN POSITION SHOWN)
COM ADAPTER SOCKET

RS-485
CONVERTER
RS-232 9-PIN
CONNECTOR
TD(A)
TD(B)
GND
Part No. 101283116
GND
+12V
Figure A.5Wiring of plug connector to 9-pin RS-232 to RS-485 converter cable, Part No. 9A-101283116
Connecting to the Communications Adapter
Take the following precautions when disassembling the communications adapter.
!
WARNING: When a hazardous area is present, ensure the union nut and blanking plug are
properly ftted in the conduit opening. The explosion-proof rating applies only when the union
nut and blanking plug are secured in place. When the union is broken, the device is no longer
explosion-proof.
!
WARNING: Before disassembling the union nut and blanking plug, make sure the area is non-
hazardous.
To connect a PC or laptop to the communications adapter, perform the following steps:
1. Unscrew the union nut to expose the connector socket shown in Figure A.4, page A-3. (A blanking plug will
be removed with the union nut. Store the union nut and blanking plug in a safe place. They will need to be
reinstalled when the adapter is not in use.)
2. Wire the plug connector to an RS-485 converter cable, if it is not already attached (Figure A.5).
3. Insert the plug connector into the adapter socket.
4. Connect the converter cable to the PC or laptop.
A-5
MC-III EXP Flow Analyzer Appendix A
Disconnecting the Communications Adapter
To disconnect the adapter, perform the following steps:
1. Remove the plug connector (with converter cable attached) from the socket.
2. Place the blanking plug inside the union nut (removed in step 1).
3. Screw the union nut onto the union half to cover the socket. Hand-tighten to ensure a snug connection.
Important: Do not disconnect the plug connector from the RS-232 to RS-485 converter cable when it
is not in use. For best results, store the plug connector with the converter cable.
Installation of Adapter (when purchased separately from an MC-III EXP)
!
WARNING: If the communications adapter is ordered separately from the MC-III EXP Flow Ana-
lyzer, the conduit openings in the enclosure will be sealed with brass or stainless steel plugs.
Do not remove the plug from the enclosure to install the adapter unless the area is known to be
non-hazardous.
To install a communications adapter purchased separately from an MC-III EXP, perform the following steps:
1. Thread the cable of the adapter through a conduit opening in the instrument housing and screw the adapter
into place.
2. Connect the adapter cable to the RS-485 port on the MC-III EXP main circuit board (Figure A.6).
3. Connect the plug connector to an RS-485 converter cable (Figure A.5, page A-4) if applicable.
Red
Black
TB1
P
U
L
S
E
I
N
P
U
T
R
E
S
E
T
I
N
P
U
T
T
F
M
A&S
GND
T
B
3
EXT POWER
TB2
R
S
4
8
5
S
L
A
V
E
4
-
2
0
O
U
T
P
U
L
S
E
O
U
T
6
-3
0
V
D
C
Communications
Adapter
J1
J2
RESET
SWITCH
B
A
T
T
E
R
Y
Figure A.6Circuit board connection for communications adapter
A-6
Appendix A MC-III EXP Flow Analyzer
USB Communications Adapter
The NuFlo USB Adapter (Figure A.7) allows a user to connect a computer to the MC-III using a standard
off-the-shelf USB connector cable for quick and easy downloads without opening the MC-III enclosure.
Caution: This adapter is CSA-approved for use in hazardous areas. It does not have ATEX/IECEX
approval.
The USB adapter is comprised of a USB adapter socket, a blanking plug, and a union nut (Figure A.8). A CD
includes the software for installing the driver.
When the USB connection is ordered with an MC-III, the USB adapter is pre-installed at the factory. No
feld wiring is required. If the USB adapter is purchased as a kit, see Adapter Kit Installation, page A-7, for
installation instructions.
Important: Do not connect the USB adapter to a computer until a USB driver is installed using the
CD provided. MC-III software cannot connect to an MC-III without these fles. See Using
the Adapter, page A-7.
Figure A.7NuFlo USB adapter
Figure A.8NuFlo USB adapter components
A-7
MC-III EXP Flow Analyzer Appendix A
User-Supplied Hardware
To connect the USB adapter to an MC-III using MC-III software, a user-supplied universal serial bus USB
A/B cable is required (Figure A.9).
Figure A.9User-supplied USB A/B cable
Using the Adapter
The MC-III software CD supplied with the NuFlo USB Adapter contains the drivers required to enable USB
communications for an MC-III when the NuFlo USB Adapter is installed.
For step-by-step installation instructions, insert the CD in your computer and click the Install USB Driver
link on the Welcome screen. The software will install the appropriate driver that is compatible with your
computers operating system.
When the software is fully installed, the adapter can be connected to the computer and used to connect to the
MC-III via MC-III software (Tools/Select COM Port).
!
WARNING: When a hazardous area is present, ensure the union nut and blanking plug are
properly ftted in the conduit opening. The explosion-proof rating applies only when the union
nut and blanking plug are secured in place. When the union is broken, the device is no longer
explosion-proof.
!
WARNING: Before disassembling the union nut and blanking plug, make sure the area is non-
hazardous.
Covering the Adapter
When the USB port is not in use, nest the blanking plug inside the union nut and screw the union nut onto the
adapter to cover the USB socket. Hand-tighten to ensure a snug connection.
Adapter Kit Installation
If the NuFlo USB adapter is purchased as a kit, install it in the MC-III EXP according to the steps below.
The USB adapter is comprised of a USB adapter socket, a blanking plug, and a union nut (Figure A.8, page A-6).
The blanking plug and union nut are connected to the adapter only when the USB port is not in use.
1. Remove the plug from a conduit opening in the MC-III EXP enclosure.
2. Thread the cable of the adapter through the conduit opening and screw the adapter into place.
A-8
Appendix A MC-III EXP Flow Analyzer
3. Connect the adapter cable to the RS-485 port on the MC-III EXP main circuit board (Figure A.6, page A-5).
4. Connect the plug connector to an RS-485 converter cable (Figure A.5, page A-4), if applicable (black wire to
negative terminal).
TB2
4
-
2
0

O
U
T
R
S
-
4
8
5
S
L
A
V
E

+
U
S
B
C
O
N
N
E
C
T
O
R
P
U
L
S
E
O
U
T
Figure A.10Wiring of NuFlo USB adapter (required only when purchased as a kit)
5. If the USB port will not be used immediately, nest the blanking plug inside the union nut and screw the
union nut onto the adapter to cover the USB socket. Hand-tighten to ensure a snug connection.
Important Before using the USB adapter for the frst time, it is necessary to install a USB driver on
the computer used to run MC-III software. The driver is supplied on the MC-III software
CD provided with the USB adapter kit. For step-by-step installation instructions, insert
the CD in your computer and click the Install USB Driver link on the Welcome screen.
The software will install the appropriate driver that is compatible with your computers
operating system.
B-1
MC-III EXP Flow Analyzer Appendix B
Appendix BSoftware Program Options
In addition to the standard confguration and fow log menus, the MC-III interface software includes a
Program Options menu that allows users to customize the way the software functions, handles log data, and
communicates.
To access the Program Options menu, select Options in the menu bar; Program Options is the last entry in the
dropdown list (Figure B.1).
Figure B.1Program Options menu location
General Options
The General Options screen allows a user to change the threshold setting used to generate a user prompt for
synchronizing the internal clock with the time displayed on the users computer.
The MC-III EXP includes a real-time clock for timekeeping and log time stamps, and the internal time and
date is preset at the factory. However, if the factory-set time and date are inaccurate for the user (for example,
if the user is in a different time zone), the program will automatically detect the difference between the
devices internal time and the date and the time/date displayed on the users personal computer and prompt
the user to authorize time and date synchronization.
By default, the MC-III prompts a user to synchronize the clock display if the time difference is 60 seconds or
more. To change the time difference required to generate a synchronize prompt, enter a new threshold value
(seconds) in the General Options screen (Figure B.2, page B-2) and click OK.
B-2
Appendix B MC-III EXP Flow Analyzer
Figure B.2Threshold setting for internal clock synchronization
Autorun Options
When the software connects with the MC-III EXP, the Device Autorun Options screen appears, prompting the
user to select any of six actions, including navigating to one of three confguration screens, downloading fow
and event logs, and uploading a confguration fle. A checkbox at the bottom of the screen allows the user to
automate the selected function each time the software connects to the instrument.
The Autorun Options screen in the Program Options menu allows a user to terminate all autorun functions, or
change the autorun selection.
To change an autorun setting that is, to initiate an autorun action, to terminate an autorun action, or to
change to a different autorun action at the next connectionselect Autorun from the Program Options menu
(Figure B.3, page B-3), select the appropriate checkboxes, and click OK. Verify that the Always do...
checkbox is checked if the selected function is to be automated upon each software connection.
B-3
MC-III EXP Flow Analyzer Appendix B
Figure B.3Menu for changing autorun option selection
Communications Options
The Communications Options screen (Figure B.4) allows users to customize communication parameters for
optimizing performance, depending on the communication path.
Figure B.4Menu for changing communication parameters
B-4
Appendix B MC-III EXP Flow Analyzer
Express Connect Option
By default, the MC-III interface software automatically attempts to connect to the instrument when a user
clicks on the desktop application icon. This feature is controlled by the Always attempt Express Connect on
program startup checkbox setting in the Communications Option screen (Figure B.4, page B-3).
When multiple devices are daisy-chained together in a network, the express connect function can interfere
with the users ability to connect to a specifc device. To disable the Express Connect function, deselect the
Always attempt... checkbox and click OK. The software will now attempt to connect with the instrument
only when the user clicks on the LCD image on the Welcome screen or selects File>Express Connect. For
more information, see Software Connection in Multi-Device Network, page 43.
To cancel the express connect function without disabling the autorun option, the user can toggle between
connect/disconnect by clicking in the LCD area of the device on the Welcome screen.
Auto-Negotiate Option
When connected devices are capable of switching baud rates automatically or when devices are directly
connected to an MC-III EXP, a user may confgure the MC-III interface to automatically negotiate the fastest
baud rate during device connection. This negotiated baud rate does not replace the confgured baud rate,
but rather provides a temporary boost of baud rate while the computer is connected to the device for faster
downloads. After the software disconnects, the device reverts to the confgured baud rate.
By default, this function is disabled. To enable automatic baud rate negotiation on device connect, check the
Attempt auto-negotiation. check box in the Communications Option screen (Figure B.4, page B-3) and click
OK. To initiate an immediate search for the fastest baud rate without exiting to the Welcome screen, check
the Enable auto-negotiated baud rate checkbox, click the Test Now button to the right of the checkbox,
and click Yes at the prompt in the Confrm dialog box.
Enable Auto-Negotiated Baud Rate Option
The Enable auto-negotiated checkbox enables usage of the baud rate that has been automatically selected
or negotiated by the MC-III interface, either upon connection with the instrument or when the Test Now
button was last clicked. See also Auto-Negotiate Option above.
Toggle RTS Line Option
Some RS-485 to RS-232 converter modules require a request-to-send (RTS) line for switching between a
receive and transmit state. This checkbox allows the software to toggle this line, ensuring that incoming
data will be readily transmitted and that the converter will not stall in a receive mode after a period of
inactivity (see Figure B.4, page B-3). By default, this option is enabled. To disable the option, deselect the
checkbox and click OK.
Enable Modbus Address Support for Firmware Versions 1.06 through 1.08
Operators with MC-III units programmed with frmware versions 1.06 through 1.08 must check the checkbox
labeled Enable Modbus Address Support for Firmware Versions 1.06 through 1.08 to ensure seamless
Modbus communication with their units. Operators using any other frmware version with the MC-III should
leave the checkbox unchecked.
Timing Parameters
Presend Delay
On the Communications Options screen (Figure B.4, page B-3), a user can enter a presend delay (milliseconds)
to help ensure a successful connection between the computer and the MC-III EXP when the instrument is
networked with a radio or other low-speed device. The computer transmits a request to send, which turns
B-5
MC-III EXP Flow Analyzer Appendix B
the device on, and then waits the specifed length of time before attempting to transmit data. By default, this
parameter is set to zero.
Number of Retries
By default, the computer will make two attempts to connect to the MC-III EXP before providing an error
message. In situations where timeouts are anticipatedfor example, due to a slow-speed computerthe
user can increase the length of time allowed for a connection by inputting a larger number in the Number of
Retries feld. See Figure B.4, page B-3.
Timeout Period Type
A user can select an optimal timeout, which is based on the response time of the MC-III EXP to various
requests for data at all baud rates during extensive testing, or a fxed timeout determined by the user. By
default, the optimal period type is enabled. When a communication error occurs, the optimal timeout
allows the software to respond much faster than the more conservative fxed timeout. The optimal setting
is recommended when the software is connecting directly to the instrument. The fxed timeout type is
recommended when data is being transmitted through a radio or other network device (serial to TCP/IP) and
there are recognizable delays in the communication system. See Figure B.4, page B-3.
Timeout Period
When a fxed timeout period type is selected, the user enters the timeout period value (in milliseconds) in
the Timeout Period feld. When an optimal timeout period type is selected, the user can use this feld to
enter a percentage by which the optimal timeout period is extended. This extension can be advantageous
when connecting to a slow computer. See Figure B.4, page B-3.
Downloading Options
The MC-III EXP automatically saves daily fow logs on the contract hour, and hourly fow logs around the
clock. The instrument also automatically saves an event log each time a user change is made. By default, the
logs are saved as .csv fles. The Downloading Options screen (Figure B.5) allows a user to change the format
of the auto-save fow logs and event logs to Excel (.xls) fles.
Figure B.5Menu for changing the auto-save log fle format
B-6
Appendix B MC-III EXP Flow Analyzer
To save the logs in .xls format, check the XLS-MS Excel File checkbox (Figure B.5, page B-5). Additional
checkboxes allow for changes in the appearance of downloaded log data.
Users can also export fow data to a .csv or .xls fle as needed. For complete information, see Exporting Flow
Logs, page 80.
Advanced Options
Clear EEPROM
The Clear EEPROM function will erase all confguration settings and log data. This task is typically
performed when updating to a new frmware version or before moving the instrument to a new site. To clear
the memory, click on the Clear EEPROM button (Figure B.6).
Technical assistance is typically required for upgrading frmware. Contact Camerons Measurement Systems
Division to arrange an upgrade.
!
WARNING: Clicking the Clear EEPROM button will delete all confguration settings and log
records in the MC-III EXP memory.
Figure B.6Menu for clearing EEPROMs and enabling automatic data logs
Automatic Data Logging
Data retrieved from the MC-III EXP while in Advanced access mode can be logged to a fle by enabling
the Automatically log data polls checkbox on the Advanced Options screen. To specify the flename and
location of the fle, click the folder icon on the right of the Data logging flename feld. The fle will include
values that are retrieved from the instrument when the user clicks the Get All Data button or enables Auto
Polling on the Advanced Access screen. See also Advanced Access, page 72.
C-1
MC-III EXP Flow Analyzer Appendix C
Appendix CLithium Battery Information
Lithium Battery Disposal
Once a lithium battery is removed from a device and/or is destined for disposal, it is classifed as solid waste
under EPA guidelines. Depleted lithium batteries are also considered to be hazardous waste because they meet
the defnition of Reactivity, as per 40 CFR 261.23(a)(2), (3) and (5). This document describes how the lithium
reacts violently with water, forms potentially explosive mixtures with water, and when exposed to certain pH
conditions, generates toxic cyanide or sulfde gases.
Federal law requires that depleted lithium batteries be sent to a fully permitted Treatment, Storage and
Disposal Facility (TSDF) or to a permitted recycling/reclamation facility.
Important: Do not ship lithium batteries to Cameron. Cameron facilities are not permitted recycling/
reclamation facilities.
Caution: Profling and waste characterization procedures must be followed prior to shipping a
lithium battery to a disposal site. It is the shippers responsibility to comply with all ap-
plicable federal transportation regulations (see below).
Transportation Information
!
WARNING: The MC-III EXP Flow Analyzer contains lithium batteries. The internal component
(thionyl chloride) is hazardous under the criteria of the Federal OHSA Hazard Communica-
tion Standard 29 CFR 1920.1200. Before shipping a lithium battery or equipment containing a
lithium battery, verify that the packaging and labeling conforms with the latest version of all
applicable regulations.
The transport of the lithium batteries is regulated by the United Nations, Model Regulations on Transport of
Dangerous Goods, (special provisions 188, 230, and 310), latest revision.
Within the US the lithium batteries and cells are subject to shipping requirements under Part 49 of the Code
of Federal Regulations (49 CFR, Parts 171, 172, 173, and 175) of the US Hazardous Materials Regulations
(HMR), latest revision.
Shipping of lithium batteries in aircraft is regulated by the International Civil Aviation Organization (ICAO)
and the International Air Transport Association (IATA) requirements in Special Provisions A45, A88 and A99,
latest revision.
Shipping of lithium batteries on sea is regulated the International Maritime Dangerous Goods (IMDG)
requirements in special provisions 188, 230 and 310, latest revision.
Shipping of lithium batteries on road and rail is regulated by requirements in special provisions 188, 230 and
310, latest revision.
Material Safety Data Sheet
For a link to the current MSDS for the lithium batteries used to power the MC-III EXP Flow Analyzer, see the
Cameron website: www.c-a-m.com.
C-2
Appendix C MC-III EXP Flow Analyzer
D-1
MC-III EXP Flow Analyzer Appendix D
Appendix DCommunications Protocol
Firmware Version 2.03
Register Table Version 1
Introduction
The communications protocol for the MC-III is in accordance with Modicon, Inc. RTU Mode Modbus

as
described in Modicon Modbus Protocol Reference Guide, PI-MBUS-300 Rev. J, June 1996. All registers are
implemented as 4X or holding registers. Reading of registers is implemented via function code 03H (Read
Holding Registers). Writing to registers is implemented via function code 10H (Preset Multiple Registers).
The instrument provides Enron Modbus

compliant downloads for hourly, daily and event records. For details
on Enron Modbus

, refer to Specifcations and Requirements for an Electronic Flow Measurement Remote


Terminal Unit for Enron Corp., Dec. 5, 1994.
Supported Commands
The Modbus

functions supported by the MC-III are as follows:


Function Code
(Hex)
Description
03 Read Holding Registers
05 Preset Boolean
(for Enron event record acknowledgement)
10 Preset Multiple Registers
For the read holding and preset multiple registers, the instrument supports the full 250 bytes of data in a
message. This corresponds to 125 registers in 16-bit holding register size and 62 registers in 32-bit holding
register size.
Data Types
Various data types are implemented in the MC-III. The following table lists the formats and the numbers of
bytes and registers associated with each type.
Data Type Byte Count Register Count
Floating Point (FP) 4 2
Floating Point (FP32) 4 1
Unsigned Word (U16) 2 1
Packed ASCII (PA) 2 1
The Floating Point (FP) type follows the IEEE-754 format and consists of 32 bits contained in two registers.
It is utilized for parameters and values that are not integers. For example, Instantaneous Flow Rate is a
parameter that is a foating-point data type. It can be interpreted by reading two registers, starting with register
17014.
D-2
Appendix D MC-III EXP Flow Analyzer
The Floating Point-32 bit (FP32) data type follows the IEEE-754 format and consists of 32 bits contained in a
single register.
The standard word ordering for multiple register data types, such as foating-point numbers or long integers, is
for the most signifcant word to appear frst in the message. For master devices requiring the least signifcant
word to appear last, a subset of registers is provided on page D-11.
The Unsigned Word (U16) type is used for 16-bit integers and fts into one register.
The Packed ASCII (PA) type contains two bytes that are two unsigned characters. Generally, multiple Packed
ASCII types are arranged consecutively for implementing strings. For example, the devices well name is a
string of 10 unsigned characters that is implemented as 5 Packed ASCII registers. Here is an example of a
well name that contains the string, Test Well 413.
Register Hexadecimal # ASCII Characters
210 54 65 Te
211 73 74 St
212 20 57 <SPACE>W
213 65 6C el
214 6C 20 l<SPACE>
215 34 31 41
216 33 FF 3<UNUSED>

Unused characters at the end of each string will report 0xFF hexadecimal.
Registers
Each register has an Access type: read-only or read-write, as described below.
Access Type Description
Read Only (RO) Register can only be read.
Read/Write (R/W) Register can be read and written.
The registers are grouped into Modbus

map blocks based on function. The MC-III contains the following


map functions.
Function Starting Register
System Confguration 1000
Real Time 1200
Input Confguration 2000
Output Confguration 4000
Holding Register (16-bit) 8000 and 17001
Holding Register (32-bit) 7000
Control Registers 70
Wellsite Parameters 200
D-3
MC-III EXP Flow Analyzer Appendix D
Note All registers cited in this document refer to the address of the register that appears in the actual Mod-
bus

message. For example, register 17001 has an address of 0x4268 Hexadecimal in the message.
Note Two sets of 16-bit holding registers are provided (starting at 17001 and 8000). The 8000 set should
be used with master devices that require the holding registers to be in the range of 40000 to 49999.
System Confguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
1000 03E8 Product Code U16 RO 64
1001 03E9 Register Table Version Number U16 RO 1
1002 03E
Firmware Version Number
(Version = register value divided by 100)
(Example: 100 = Version 1.00)
U16 RO
1003 03EB Manufacture Date (MMYY) U16 RO
1004 03EC Sales Date (MMYY) U16 RO
1005 03ED Serial Number 1 U16 RO
1006 03EE Serial Number 2 U16 RO
1007 03EF
Power Mode
0 = High Power
1 = Low Power
U16 R/W 1
1008 03F0
Internal System Sample Rate
Number of seconds to measure temperature
and battery voltage
U16 R/W 3600
1009 03F1
Slave Address
[1-65535, not including 252-255 or 64764]
U16 R/W 1
1010 03F2
Baud Rate
Low Power High Power
0 - 300 5 - 9600
1 - 600 6 - 19200
2 - 1200 7 - 38400
3 - 2400 8 - 57600
4 - 4800 10 - 115200
5 - 9600
U16 R/W 5
1011 03F3
Bus Delay mS of delay before transmitting data
(Delay = register value multiplied by 10 mS)
U16 R/W
1
(10 ms)
1012 03F4
Bus Timeout mS of delay before resetting
Modbus


(Timeout = register value multiplied by 10 mS)
U16 R/W
5
(50 mS)
1013 03F5
Contract Hour
[0-23]
U16 R/W 8
1014 03F6
Lock Code
[0001-9999]
U16 R/W 0
1015 03F7
Lock Code Enable
0 - Disabled
1 - Enabled
U16 R/W 0
D-4
Appendix D MC-III EXP Flow Analyzer
System Confguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
1016 03F8
LCD Contrast
[0-31]
U16 R/W 6
Product Code (register 1000)
The Product Code is a read-only parameter used for identifcation. This parameter is set at the factory and it
will always read 0x40 hexadecimal (64 decimal).
Firmware/Register Table Version Numbers (registers 1001, 1002)
The Firmware Version and Register Table Version numbers are unsigned 16-bit integers. These parameters are
set at the factory and are read-only. To determine the frmware version number, read the appropriate register
and divide the value read by 100. The general format for frmware version numbers is A.BC. For example, the
frmware register number is read as 0xA7 hexadecimal. This represents 167 and a frmware version of 1.67.
Manufacture Date/Sales Date (registers 1003, 1004)
These parameters are set at the factory and are read-only. These registers are formatted as MMYY. For
example, a value of 0905 represents the date September 2005.
Slave Address (register 1009)
The Modbus

slave address is an unsigned word (U16) data type that has a range of values from 1 to 65535,
not including 252 through 255 or 64764. The slave address is confgured from a laptop or via the keypad and
is stored in nonvolatile memory.
Important: All MC-III devices will respond to addresses 252 through 255 and 64764, and the
response will always be at a baud rate of 9600 baud. For this reason, addresses 252
through 255 and 64764 are reserved and should never be used.
When the slave address is written, the response message will be at the current address. After the response
message is transmitted, the MC-III will change to the new slave address.
Baud Rate (register 1010)
The baud rate is an unsigned word (U16) data type that sets the data rate on the serial port. When the baud rate
is written, the response message will be at the current baud rate. After the response message is transmitted, the
MC-III will change to the new baud rate.

D-5
MC-III EXP Flow Analyzer Appendix D
Real Time
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
1200 04B0
Year
(Real Year = register value plus 2000)
U16 R/W
1201 04B1 Month [1-12] U16 R/W
1202 04B2 Day [1-31] U16 R/W
1203 04B3 Hour [0-23] U16 R/W
1204 04B4 Minute [0-59] U16 R/W
1205 04B5 Second [0-59] U16 R/W
This block of registers is used to set the instruments internal clock. To set the time, it is recommended that all
registers be written in one message.
The time and date can also be read in the holding register groups as foating-point data.
Input Confguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
2000 07D0
Volume Units
Units
0 = No Units
1 = Gallons
2 = Barrels
3 = Cubic Meters
4 = Cubic Feet
5 = Liters
Units 1000
16 = No Units
17 = Gallons
18 = Barrels
19 = Cubic Meters
20 = Cubic Feet
21 = Liters
U16 R/W 2
Standard Units
32 = No Units
33 = Gallons
34 = Barrels
35 = Cubic Meters
36 = Cubic Feet
37 = Liters
Std. Units 1000
48 = No Units
49 = Gallons
50 = Barrels
51 = Cubic Meters
52 = Cubic Feet
53 = Liters
2001 07D1
Volume Decimal Point Location
1 = Tenths
2 = Hundredths
3 = Thousandths
4 = Ten Thousandths
U16 R/W 1
D-6
Appendix D MC-III EXP Flow Analyzer
Input Confguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
2002 07D2
Rate Volume Units
Units
0 = No Units
1 = Gallons
2 = Barrels
3 = Cubic Meters
4 = Cubic Feet
5 = Liters
Units 1000
16 = No Units
17 = Gallons
18 = Barrels
19 = Cubic Meters
20 = Cubic Feet
21 = Liters
U16 R/W 2
Standard Units
32 = No Units
33 = Gallons
34 = Barrels
35 = Cubic Meters
36 = Cubic Feet
37 = Liters
Std. Units 1000
48 = No Units
49 = Gallons
50 = Barrels
51 = Cubic Meters
52 = Cubic Feet
53 = Liters
2003 07D3
Rate Time Base
0 = Second
1 = Minute
2 = Hour
3 = Day
U16 R/W 3
2004 07D4
Rate Decimal Point Location
0 = No digits to the right of the decimal point
1 = Tenths
2 = Hundredths
3 = Thousandths
U16 R/W 2
2005 07D5
Calculation Period
(sec)
U16 R/W 4
2006 07D6
Turbine Meter Input Confguration
0 = Pulse Input
1 = TFM - LOW
2 = TFM - MED
3 = TFM - HIGH
U16 R/W 1
2007 07D7
Low Frequency Cut-Off Threshold
(Hz)
U16 2
2008 07D8
Factor Type
1 = 1-Point K-Factor
2 = 2-Point K-Factor

12 = 12-Point K-Factor
U16 R/W 1
D-7
MC-III EXP Flow Analyzer Appendix D
Input Confguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
2009 07D9
Factor Units
0 = No Units
1 = Gallons
2 = Liters
3 = Barrels
4 = Cubic Feet
5 = Cubic Meters
U16 R/W 1
2010 07DA
Factor Decimal Point Location
0 = No digits to the right of the decimal point
1 = Tenths
2 = Hundredths
3 = Thousandths
4 = Ten Thousandths
U16 R/W 2
2011 07DB Calibration Frequency [1] (Hz)* FP R/W 1.00
2013 07DD Calibration Frequency [2] (Hz)* FP R/W 1.00
2015 07DF Calibration Frequency [3] (Hz)* FP R/W 1.00
2017 07E1 Calibration Frequency [4] (Hz)* FP R/W 1.00
2019 07E3 Calibration Frequency [5] (Hz)* FP R/W 1.00
2021 07E5 Calibration Frequency [6] (Hz)* FP R/W 1.00
2023 07E7 Calibration Frequency [7] (Hz)* FP R/W 1.00
2025 07E9 Calibration Frequency [8] (Hz)* FP R/W 1.00
2027 07EB Calibration Frequency [9] (Hz)* FP R/W 1.00
2029 07ED Calibration Frequency [10] (Hz)* FP R/W 1.00
2031 07EF Calibration Frequency [11] (Hz)* FP R/W 1.00
2033 07F1 Calibration Frequency [12] (Hz)* FP R/W 1.00
2035 07F3
Calibration Factor [1]*
(pulses per confgured factor unit)
FP R/W 900.00
2037 07F5
Calibration Factor [2]*
(pulses per confgured factor unit)
FP R/W 1.00
2039 07F7
Calibration Factor [3]*
(pulses per confgured factor unit)
FP R/W 1.00
2041 07F9
Calibration Factor [4]*
(pulses per confgured factor unit)
FP R/W 1.00
2043 07FB
Calibration Factor [5]*
(pulses per confgured factor unit)
FP R/W 1.00
2045 07FD
Calibration Factor [6]*
(pulses per confgured factor unit)
FP R/W 1.00
2047 07FF
Calibration Factor [7]*
(pulses per confgured factor unit)
FP R/W 1.00
2049 0801
Calibration Factor [8]*
(pulses per confgured factor unit)
FP R/W 1.00
D-8
Appendix D MC-III EXP Flow Analyzer
Input Confguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
2051 0803
Calibration Factor [9]*
(pulses per confgured factor unit)
FP R/W 1.00
2053 0805
Calibration Factor [10]*
(pulses per confgured factor unit)
FP R/W 1.00
2055 0807
Calibration Factor [11]*
(pulses per confgured factor unit)
FP R/W 1.00
2057 0809
Calibration Factor [12]*
(pulses per confgured factor unit)
FP R/W 1.00
2059 080B
Standard Measurement Correction Factor
(to be confgured by software only)
FP RO 1.00
2061 080D
Low Flow Rate Cut-Off Threshold
(in terms of confgured fow rate units)
FP R/W 1.00
*Calibration frequencies and calibration factors are associated in pairsi.e., Calibration Frequency [1] and
Calibration Factor [1] and must be sorted in ascending order, based on calibration frequency values, prior
to writing the values to the instrument. Therefore, Calibration Frequency [1] should be the smallest of all
calibration frequencies.
Output Confguration
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access Default
4000 0FA0
Pulse Output Source
0 = Test Mode/Disabled
1 = Enabled
U16 R/W 0
4001 0FA1
Pulse Output Duration
(time is register value multiplied by 10 ms)
U16 R/W
10
(100 mS)
4002 0FA2 Pulse Output Decimal Point Location U16 R/W 2
4003 0FA3
Pulse Output Scale Factor
(in terms of confgured volume units)
FP R/W 1.00
4005 0FA5
Analog Output Source
0 = Test Mode/Disabled
1 = Enabled
U16 R/W 0
4006 0FA6 Analog Output Decimal Point Location U16 R/W 2
4007 0FA7
Analog Output Low Value
(in terms of confgured fow rate units)
FP R/W 0.00
4009 0FA9
Analog Output High Value
(in terms of confgured fow rate units)
FP R/W 1700.00
Important: Two sets of 16-bit holding registers are provided (starting at 17001 and 8000). The
8000 set displayed in parentheses in the Holding Registers table on page D-9 should be
used with master devices that require the holding registers to be in the range of 40000 to
49999.
D-9
MC-III EXP Flow Analyzer Appendix D
Holding Registers (16-bit Mode)
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
17001
(8000)
4269
(1F40)
Hourly Pointer
[1 to 768]
FP RO
17003
(8002)
426B
(1F42)
Daily Pointer
[1 to 384]
FP RO
17005
(8004)
426D
(1F44)
Event Counter
[1 to 345]
FP RO
17007
(8006)
426F
(1F46)
Real Date
(MM/DD/YY)
FP RO
17009
(8008)
4271
(1F48)
Real Time
(HH:MM:SS)
FP RO
17011
(8010)
4273
(1F4A)
Grand Total
(in terms of confgured volume units)
FP RO
17013
(8012)
4275
(1F4C)
Instantaneous Flow Rate
(in terms of confgured fow rate units)
FP RO
17015
(8014)
4277
(1F4E)
Daily Total
(in terms of confgured volume units)
FP RO
17017
(8016)
4279
(1F50)
Daily Run Time
(seconds)
FP RO
17019
(8018)
427B
(1F52)
Hourly Total
(in terms of confgured volume units)
FP RO
17021
(8020)
427D
(1F54)
Hourly Run Time
(seconds)
FP RO
17023
(8022)
427F
(1F56)
Polling Total
(in terms of confgured volume units)
FP RO
17025
(8024)
4281
(1F58)
Polling Run Time
(seconds)
FP RO
17027
(8026)
4283
(1F5A)
Previous Day Total
(in terms of confgured volume units)
FP RO
17029
(8028)
4285
(1F5C)
Previous Day Run Time
(seconds)
FP RO
17031
(8030)
4287
(1F5E)
Previous Hour Total
(in terms of confgured volume units)
FP RO
17033
(8032)
4289
(1F60)
Previous Hour Run Time
(seconds)
FP RO
17035
(8034)
428B
(1F62)
Previous Polling Total
(in terms of confgured volume units)
FP RO
17037
(8036)
428D
(1F64)
Previous Polling Run Time
(seconds)
FP RO
17039
(8038)
428F
(1F66)
Polling Index
[0-65535]
FP RO
17041
(8040)
4291
(1F68)
Internal Temperature
(C)
FP RO
D-10
Appendix D MC-III EXP Flow Analyzer
Holding Registers (16-bit Mode)
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
17043
(8042)
4293
(1F6A)
Supply Voltage
(VDC)
FP RO
17045
(8044)
4295
(1F6C)
Battery Voltage
(VDC)
FP RO
17047
(8046)
4297
(1F6E)
Grand Total
(base unit)
FP RO
17049
(8048)
4299
(1F70)
Instantaneous Flow Rate
(base unit)
FP RO
17051
(8050)
429B
(1F72)
Daily Total
(base unit)
FP RO
17053
(8052)
429D
(1F74)
Daily Run Time
(seconds)
FP RO
17055
(8054)
429F
(1F76)
Hourly Total
(base unit)
FP RO
17057
(8056)
42A1
(1F78)
Hourly Run Time
(seconds)
FP RO
17059
(8058)
42A3
(1F7A)
Polling Total
(base unit)
FP RO
17061
(8060)
42A5
(1F7C)
Polling Run Time
(seconds)
FP RO
17063
(8062)
42A7
(1F7E)
Previous Day
(base unit)
FP RO
17065
(8064)
42A9
(1F80)
Previous Day Run Time
(seconds)
FP RO
17067
(8066)
42AB
(1F82)
Previous Hour
(base unit)
FP RO
17069
(8068)
42AD
(1F84)
Previous Hour Run Time
(seconds)
FP RO
17071
(8070)
42AF
(1F86)
Previous Polling Total
(base unit)
FP RO
17073
(8072)
42B1
(1F88)
Previous Polling Run Time
(seconds)
FP RO
17075
(8074)
42B3
(1F8A)
Polling Index
[0-65535]
FP RO
17077
(8076)
42B5
(1F8C)
Pulse Output Pulses FP R/W
17079
(8078)
42B7
(1F8E)
Analog Output Current FP R/W
17081
(8080)
42B9
(1F90)
Calculated K-Factor
(in terms of confgured factor unit)
FP RO
D-11
MC-III EXP Flow Analyzer Appendix D
Note: The registers below should be used with master devices that require the foating point numbers to
be presented with the least signifcant word frst. Their counterparts with foating point numbers in
reverse word orderregisters 17011, 17013, 17027 and 17015are listed in the previous table.
Holding Registers (16-bit Mode)Least Signifcant Word First
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
550 226
Grand Total
(in terms of confgured volume units)
FP RO
552 228
Instantaneous Flow Rate
(in terms of confgured fow rate units)
FP RO
554 22A
Daily Total
(in terms of confgured volume units)
FP RO
556 22C
Previous Day Total
(in terms of confgured volume units)
FP RO
Base Units/Confgured Units
The holding register area provides two blocks of registers. The frst block is based on the confgured units
of measurement. The confgured volume units will follow the settings in the Volume Unit register (2000).
The confgured rate units will follow the settings in the Rate Volume units (2002) and the Time-Base register
(2003).
For example, if the settings for the volume unit is 2 (BBL), all of the total parameters such as Grand Total
(17011), Daily Total (17015), etc. will be in terms of barrels. The values in these registers will change if the
user changes the units of measurement that appear on the LCD.
The second block is in terms of base units. These register values are independent of the confgured volume
units in the Volume Unit, Rate Volume Unit, and Time-Base registers. For base units, the time base is always
measured in seconds, and the volume and rate volume units are measured in gallons. If the standard correction
factor has been enabled, the volume unit is standard gallons. The standard gallon is calculated as the gallon
measurement times the Standard Measurement Correction factor (2059). If the x1000 unit is enabled, the
volume unit is gallons times 1000.
Conversion Factors
For base units, the volume unit and rate volume unit is measured in gallons.To convert gallons to another unit
of measure, use the conversion factors listed in the table below. For example, to convert gallons to barrels,
multiply gallons times 0.02380952381.
To Convert From Gallons To... Multiply By...
No units 1.00000000000
Barrels 0.02380952381
Cubic meters 0.00378541178
Cubic feet 0.13368055556
Liters 3.78541178400
D-12
Appendix D MC-III EXP Flow Analyzer
Polling Registers
The MC-III EXP device stores the volume accumulated since the last polling sequence in a set of polling
registers (17023). In addition, the instrument monitors the number of seconds of fow time in each polling
interval (17025). There is also an index to the number of polls requested (17039).
The polling sequence is started by writing to the Control Register. Writing the Control Register 1 (70) with
a value of 20000 decimal transfers the polling total (17023) and polling run time (17025), to the previous
polling total (17035) and previous polling run time (17037) registers, increments the polling index (17039)
register and resets the polling total and polling run time registers.
Pointer/Daily/Event Pointer (registers 17001 through 17006)
These registers provide an index of the last record that was stored in the log data. These values start at 1 and
increment with each newly created log. When the maximum number of records is reached, the pointer resets
at 1 and starts incrementing again.
Real Date (registers 17007, 17008)
This register is a foating-point representation of the date, formatted as MMDDYY. For example, a value of
91005 represents a date of September 10, 2005.
Real Time (registers 17009, 17010)
This register is a foating-point representation of time, formatted as HH:MM:SS. For example, a value of
180205 represents a time of 6:02:05 PM.
Totals
The instrument provides Grand Total (17011), Daily Total (17015), Hourly Total (17019) and a Polling Total
(17023). It also provides Previous Day Total (17027), Previous Hour Total (17031), and Previous Polling
Total (17035). These totals are available in terms of confgured units and base units. Refer to Base Units/
Confgured Units, page D-11, for details.
Pulse Output Pulses (register 17077)
When the pulse output is enabled (see Pulse Output Source, register 4000), this register indicates the
current number of output pulses to be sent to the output relay, based on the confgured duration. In a typical
application, this register will generally indicate a small number, then quickly count down to zero. In a
situation where pulses are being cached, the pulses register will indicate a large value. When the pulse output
is disabled, a user can use this register to write a desired number of pulses to be output, for use in testing end
devices. Users can also perform this output test from the software Pulse Output confguration screen. For
details, see Pulse Output Testing, page 68.
Analog Output Current (register 17079)
When the analog output is enabled (see Analog Output Source, register 4005), this register indicates the
calculated output current based on the low and high values and the instantaneous fow rate. When the analog
output is disabled, a user can use this register to write a desired output current for use in testing end devices.
Users can also perform this output test from the software 4-20 mA Output confguration screen. For details,
see 4-20 mA Output Testing, page 65.
Calculated K-Factor (register 17081)
This register represents the K-factor that was used for the most recent calculation interval. This number
is calculated based on the factor type (register 2008), the instantaneous frequency from the fowmeter,
calibration frequencies (registers 2011 through 2033), and calibration factors (registers 2035 through 2057).
D-13
MC-III EXP Flow Analyzer Appendix D
The unit of measurement for this register is pulses per confgured factor unit. Once the K-factor is calculated
based on frequency and K-factor points, this value is multiplied by 1000 if the user chooses to read the
volume in units x 1000, and it is divided by the Standard Measurement Correction Factor (register 2059).
Holding Registers (32-bit Mode)
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
7000 1B58
Hourly Pointer
[1 to 768]
FP32 RO
7001 1B59
Daily Pointer
[1 to 384]
FP32 RO
7002 1B5A
Event Counter
[1 to 345]
FP32 RO
7003 1B5B
Real Date
(MM/DD/YY)
FP32 RO
7004 1B5C
Real Time
(HH:MM:SS)
FP32 RO
7005 1B5D
Grand Total
(in terms of confgured volume units)
FP32 RO
7006 1B5E
Instantaneous Flow Rate
(in terms of confgured fow rate units)
FP32 RO
7007 1B5F
Daily Total
(in terms of confgured volume units)
FP32 RO
7008 1B60
Daily Run Time
(seconds)
FP32 RO
7009 1B61
Hourly Total
(in terms of confgured volume units)
FP32 RO
7010 1B62
Hourly Run Time
(seconds)
FP32 RO
7011 1B63
Polling Total
(in terms of confgured volume units)
FP32 RO
7012 1B64
Polling Run Time
(seconds)
FP32 RO
7013 1B65
Previous Day Total
(in terms of confgured volume units)
FP32 RO
7014 1B66
Previous Day Run Time
(seconds)
FP32 RO
7015 1B67
Previous Hour Total
(in terms of confgured volume units)
FP32 RO
7016 1B68
Previous Hour Run Time
(seconds)
FP32 RO
7017 1B69
Previous Polling Total
(in terms of confgured volume units)
FP32 RO
7018 1B6A
Previous Polling Run Time
(seconds)
FP32 RO
D-14
Appendix D MC-III EXP Flow Analyzer
Holding Registers (32-bit Mode)
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
7019 1B6B
Polling Index
[0-65535]
FP32 RO
7020 1B6C
Internal Temp
(C)
FP32 RO
7021 1B6D
Supply Voltage
(VDC)
FP32 RO
7022 1B6E
Battery Voltage
(VDC)
FP32 RO
7023 1B6F
Grand Total
(base unit)
FP32 RO
7024 1B70
Instantaneous Flow Rate
(base unit)
FP32 RO
7025 1B71
Daily Total
(base unit)
FP32 RO
7026 1B72
Daily Run Time
(seconds)
FP32 RO
7027 1B73
Hourly Total
(base unit)
FP32 RO
7028 1B74
Hourly Run Time
(seconds)
FP32 RO
7029 1B75
Polling Total
(base unit)
FP32 RO
7030 1B76
Polling Run Time
(seconds)
FP32 RO
7031 1B77
Previous Day
(base unit)
FP32 RO
7032 1B78
Previous Day Run Time
(seconds)
FP32 RO
7033 1B79
Previous Hour
(base unit)
FP32 RO
7034 1B7A
Previous Hour Run Time
(seconds)
FP32 RO
7035 1B7B
Previous Polling Total
(base unit)
FP32 RO
7036 1B7C
Previous Polling Run Time
(seconds)
FP32 RO
7037 1B7D
Polling Index
[0-65535]
FP32 RO
7038 1B7E Pulse Output Pulses FP32 R/W
7039 1B7F Analog Output Current FP32 R/W
7040 1B80
Calculated K-Factor
(in terms of confgured factor unit)
FP32 RO
D-15
MC-III EXP Flow Analyzer Appendix D
Control Register
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
70 0046 Control Register #1 U16 R/W
The Control register allows specifc functions to be implemented via the communication port. The following
table shows the value to be written to the control register to implement the desired function.
Code Function
20000
Transfers the polling total and polling run time
to the previous polling total and previous polling
run time registers, increments the polling index
register, and resets the polling total and polling run
time registers.
30000 Resets grand total
40000 Loads factory defaults
40040 Creates processor reset
Wellsite Parameters
Register
(Decimal)
Register
(Hex)
Description
Data
Type
Access
200 00C8 Company Name [20 characters] PA RO
210 00D2 Well Name [20 characters] PA RO
220 00DC Site Location [20 characters] PA RO
300 012C Site ID [20 characters] PA RO
310 0136 Tag Name [10 characters] PA RO
315 013B Legal Description [34 characters] PA RO
400 0190 Sensor Model Number [20 characters] PA RO
410 019A Sensor Serial Number [20 characters] PA RO
420 01A4 Measurement Type [20 characters] PA RO
500 01F4 User Note [64 characters] PA RO
These registers provide access to the ASCII characters that are used in the Wellsite Information screen of the
software. These registers are read-only; if these parameters need to be changed, the user must use the interface
software program.
Log Data
The MC-III provides Enron Modbus

compliant downloads. For detailed instructions on downloading hourly,


daily and event data, refer to Specifcations and Requirements for an Electronic Flow Measurement Remote
Terminal Unit for Enron Corp. If an Enron host is not available or is too cumbersome to implement, the Enron
records are individually addressed in another Modbus

map. Contact Cameron technical support for details.


D-16
Appendix D MC-III EXP Flow Analyzer
The following registers are used for hourly, daily and event log registers. Hourly and daily records comprise
fve values (registers). See the Enron Hourly/Daily Record Format, page D-16. Note that event records are
downloaded one at a time.
Enron Registers
Register
(Decimal)
Register
(Hex)
Description Data Type Access
32 0020 Enron Modbus

Event Log Register


Refer to Enron
Event Record
Format
RO
700 02BC Enron Modbus

Hourly Log
Refer to Enron
Hourly/Daily
Record Format
RO
701 02BD Enron Modbus

Daily Log
Refer to Enron
Hourly/Daily
Record Format
RO
7000 1B58
Hourly Pointer
[1 to 768]
FP32 RO
7001 1B59
Daily Pointer
[1 to 384]
FP32 RO
7002 1B5A
Event Counter
[1 to 345]
FP32 RO
Enron Hourly/Daily Record Format
Parameter Data Type
Date
(MMDDYY)
FP32
Time
(HH:MM:SS)
FP32
Total
(base units)
FP32
Flow Time
(number of seconds of fow)
FP32
Supply Voltage
(VDC)
FP32
D-17
MC-III EXP Flow Analyzer Appendix D
Enron Event Record Format
Parameter Data Type
Status U16
Address U16
Time
(HH:MM:SS)
FP32
Date
(MMDDYY)
FP32
As-Found FP32
As-Left FP32
The Event Log status is a 16-bit value used to indicate the reset source in the event log. The following table
indicates the reset status.
Reset Status
Status Code Value
Normal 0
Power-On Reset 3
Watch-Dog 8
Software Reset 16
Flash Memory Error 32
Log Capacity
Log Type Capacity
Hourly Logs 768
Daily Logs 384
Event Logs 345
D-18
Appendix D MC-III EXP Flow Analyzer
WARRANTY - LIMITATION OF LIABILITY: Seller warrants only title to the products, soft-
ware, supplies and materials and that, except as to software, the same are free from defects
in workmanship and materials for a period of one (1) year from the date of delivery. Seller
does not warranty that software is free from error or that software will run in an uninterrupted
fashion. Seller provides all software as is. THERE ARE NO WARRANTIES, EXPRESS
OR IMPLIED, OF MERCHANTABILITY, FITNESS OR OTHERWISE WHICH EXTEND
BEYOND THOSE STATED IN THE IMMEDIATELY PRECEDING SENTENCE. Sellers li-
ability and Buyers exclusive remedy in any case of action (whether in contract, tort, breach
of warranty or otherwise) arising out of the sale or use of any products, software, supplies,
or materials is expressly limited to the replacement of such products, software, supplies,
or materials on their return to Seller or, at Sellers option, to the allowance to the customer
of credit for the cost of such items. In no event shall Seller be liable for special, incidental,
indirect, punitive or consequential damages. Seller does not warrant in any way products,
software, supplies and materials not manufactured by Seller, and such will be sold only with
the warranties that are given by the manufacturer thereof. Seller will pass only through to
its purchaser of such items the warranty granted to it by the manufacturer.
RUSSI A
+603. 5569. 0501
ms-kl @c-a-m.com

Вам также может понравиться