Chapter 9
Maintenance

Document No: DD+DIS231.10E

This document is separately available. Order number: DD+DIS231.10E

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
02-2012 Printed in Germany
Agfa Company Confidential
Document Node ID: 30879915
eq_09_maintenance_e_template_v08
Copyright 2012 Agfa HealthCare N.V.


CR 30-X 2nd
Type 5175 / 200 / 205 / 220 / 225
CR 30-Xm
Type 5179 / 100

Purpose of this Document

This document contains all routines and tests to be carried out during maintenance.
It describes all must maintenance periodical steps in chronological suitable order.


Document History

Edition.
Revision
Release
Date
Changes
compared to previous Version 1.0
1.1 02-2012 Added information about CR 30-Xm.


Referenced Documents

Document Title
DD+DIS230.10E CR 30-Xm / CR 30-X 2nd Service Manual,
Chapter 3.5 - Replacements / Repair Procedures
DD+DIS230.10E CR 30-Xm / CR 30-X 2nd Service Manual,
Chapter 3.2 - Tools and auxiliary means
DD+DIS230.10E CR 30-Xm / CR 30-X 2nd Service Manual,
Chapter 3.6 - Adjustments and Calibrations


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Manufacturer

Agfa HealthCare N.V.


Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2012 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium,
or its affiliates.



WARNING:
Improper operation or service activities may cause damage or injuries.


INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see Agfa HealthCare Library > General Info > Agfa HealthCare > Publications >
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.


IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare or one of its affiliates
or who are otherwise authorized by Agfa HealthCare or one of its affiliates to provide
such services.

NOTE:
To verify the latest version of a chapter of the Service Manual refer to the Checklist for
Completeness in the Agfa HealthCare Library.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 2 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 3 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential

LIST OF CONTENTS

1 GENERAL INFORMATION ON PREVENTIVE MAINTENANCE.............................................5
1.1 Maintenance Frequency ...........................................................................................................5
1.2 Required Time...........................................................................................................................5
1.3 Required Documents ................................................................................................................5
1.4 Required Software ....................................................................................................................5
1.5 Required Tools..........................................................................................................................6
1.6 Required Cleaning Material ......................................................................................................7
1.7 Required Lubrication Material ...................................................................................................7
2 DIAGNOSTICS.........................................................................................................................8
2.1 Questioning of the Customer ....................................................................................................8
2.2 Analyzing the Info Counter........................................................................................................8
2.3 Documenting the Technical Image Quality of the System........................................................9
2.3.1 Preparing the Image Plate ........................................................................................................9
2.3.2 Erasing the Cassette...............................................................................................................11
2.3.3 Exposing the Cassette............................................................................................................12
2.3.4 Exposing the Cassette at GenRad Modality...........................................................................12
2.3.5 Scanning the Flatfield Image ..................................................................................................14
3 MAINTENANCE POWER OFF............................................................................................15
3.1 Checking of Power and Network Cable..................................................................................15
3.2 Checking of the Covers...........................................................................................................16
3.3 Checking of the Cables...........................................................................................................16
3.4 Checking of the IP Guide Plate...............................................................................................17
3.5 Checking of the four Belt Drives .............................................................................................17
3.6 Performing a mechanical Check of the Drawer Slider ............................................................18
3.7 Performing a visual Check of the Transport Rollers ...............................................................18
3.8 Performing a visual Check of all Stepper Motors....................................................................18
3.9 Cleaning the Inside .................................................................................................................19


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 4 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential
3.10 Cleaning the Erasure Unit.......................................................................................................19
3.11 Cleaning the Prescan Antistatic Brush ...................................................................................20
3.12 Greasing the Spindle at the Cassette Fixation Unit ................................................................21
3.13 Reassembling of the removed Parts.......................................................................................21
3.14 Checking the Cassette Condition............................................................................................22
4 MAINTENANCE POWER ON..............................................................................................22
4.1 Checking the Safety Switch ....................................................................................................23
4.2 Performing a Stall Calibration .................................................................................................23
4.3 Performing Test Cycles...........................................................................................................24
5 TECHNICAL IMAGE QUALITY CHECK OF THE SYSTEM...................................................25
5.1 Exposing the Cassette............................................................................................................25
5.2 Cleaning of the Scan Line and the Mirror ...............................................................................25
5.3 Scanning the Flatfield Image ..................................................................................................26
5.4 Checking the Image at the Lightbox or Viewing Station.........................................................27
6 COMPLETION OF THE MAINTENANCE...............................................................................29
6.1 Resetting the Maintenance Indicator ......................................................................................29
6.2 Creating a Backup of Device specific Data.............................................................................29
6.3 Performing the Customer Conversation .................................................................................29
7 MAINTENANCE CHECKLIST ................................................................................................30



DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
1 General Information on Preventive Maintenance

To ensure quality and functional reliability of the system all the points listed below
(minimum maintenance points) must be carried out.


IMPORTANT:
Check if it is necessary to include country specific regulations as additional
maintenance points.


1.1 Maintenance Frequency

The maintenance has to be carried out (whatever comes first):
Every 30.000 image plate cycles or
Every 2 years

1.2 Required Time


REQUIRED TIME:
Approximately 2 hours

1.3 Required Documents

For a list of required documents see the front page of this document.
Recommendation:
Download the complete Service Manual from the Agfa HealthCare Library:
<Computed Radiography CR Digitizers CR 30-Xm / CR 30-X 2nd>

1.4 Required Software

Make sure that the latest digitizer software is available on the Service PC to be
prepared for a software upgrade during the preventive maintenance. The latest
software can be downloaded from the Agfa HealthCare Library:
<Computed Radiography CR Digitizers CR 30-Xm / CR 30-X 2nd
Software>
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 5 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
1.5 Required Tools


Following required tools are part of digitizer delivery:


TX 20
TX 45


Limit Pattern CD ROM CR 30-X 2nd
Spare part number*: CM+9 5175 9160 1


Limit Pattern CD ROM CR 30-Xm
Spare part number*: CM+604948 00


1.5 mm Copper Filter
Spare part number*: CM+9 5155 1015 2
2.0 mm AL Filter
Spare part number*: CM+9 5148 1090 0



USB Flash Drive
Checked to be virus free
Spare part numbers*:
CR 30-Xm with CIRMXXXX**: CM+6045180
CR 30-X 2nd with CIR_XXXX** : CM+6037431

* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the current revision of the spare part is delivered.
** XXXX = Version number


Following required tools are not part of digitizer delivery:



Network Cable
Commercially available, Acquire locally


Lint free cloth
Commercially available, Acquire locally


Flash light
Commercially available, Acquire locally
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 6 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 7 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential
1.6 Required Cleaning Material

Following cleaning material is required:

Description Order Number*
Vacuum cleaner (220/240 V, 50-60 Hz) CM+9 9999 0895 0
Or a comparable vacuum cleaner:
Commercially available, Acquire locally
Dirt bags for vacuum cleaner (10 x) CM+9 9999 0896 0
CR Phosphor Plate Cleaner (2 bottles) 10+9999911970
Prosat wipes 10+9 9999 1219 0
Polynit wipes 10+9 9999 1273 0

Lint-free cloth Commercially available: Acquire locally
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the current revision of the spare part is delivered.

1.7 Required Lubrication Material

Description Order Number*
Isoflex Topas NB52 (50 ml) CM+9 9999 9247 0
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the current revision of the spare part is delivered.


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 8 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential
2 Diagnostics

2.1 Questioning of the Customer

(1) Ask the customer for any problem that appeared since the last maintenance.


2.2 Analyzing the Info Counter

(1) On the NX workstation log-off the current user and log-in as "crservice".
(2) Open the Internet Explorer.
(3) Enter the digitizer IP address in the address window.
If the IP address is not known: Look it up in the Configuration Tool of the
NX workstation.
(4) Select in the service menu: <Reporting Info Counter>
(5) Check the image plate cycles since last maintenance and the last maintenance
date. Enter these values in the maintenance checklist (see section 7).
(6) Check the errors since last maintenance.
(7) Check whether service documents provides information on error solutions
e.g. by HW modifications, software upgrades etc. If applicable implement the
provided solution.



DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

2.3 Documenting the Technical Image Quality of the System

2.3.1 Preparing the Image Plate


(1) Select the largest available cassette.
If the largest available cassette is a Full Leg / Full Spine (FLFS) Cassette, use
this cassette.


Genrad Cassette
FLFS
FLFS
FLFS Cassette


Figure 1

(2) Open the cassette with the
dedicated key or a pen.

517509ag.cdr


Figure 2

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 9 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(3) Open the shutter.
(4) Turn the cassette around, so that
the black tube side is above.
(5) Let the black tray and image plate
slide out carefully onto the table.
517509ad.cdr
1
1
2


Figure 3


(6) Put the black tray with the image
plate onto the cassette.
(7) Check the image plate for any
contaminations or signs of damage
(e.g. dust particles, scratches).


Figure 4


(8) If the image plate is contaminated clean it with the CR screen cleaner and
a soft lint-free cloth.

If the image plate is damaged take another cassette and inform the customer.


IMPORTANT for cleaning:
Do not put the CR Phosphor Plate Cleaner directly on the image plate.
Always put the CR Phosphor Plate Cleaner on the lint-free cloth.

(9) Wait approximately 10 minutes until the surface is dry.


Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 10 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
517509af.cdr
1
2
Black tube side of the cassette
White phosphor side

(10) Verify that the white phosphor side
is oriented to the black tube side
of the cassette.
(11) Put the image plate back into the
cassette so that the shutter does
not scratch over the image plate.
(12) Insert the key or a pen into the
cassette.
(13) Close the shutter.
(14) Remove the key.
Figure 5




2.3.2 Erasing the Cassette


(1) Press the Erase button.

The Digitizer switches to erase
mode. The status indicator is
continuously lighting up in blue.
5175_c01_006.ai


Figure 6

(2) Put the cassette in the digitizer:
Erasing starts.

5175_c01_005.ai

Figure 7


(3) Remove the cassette with the erased image plate from the cassette output.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 11 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.3.3 Exposing the Cassette
2.3.3.1 Exposing the Cassette at GenRad Modality


(1) Mount the 1,5 mm Cu-filter at the
modality.
(2) Place the Genrad cassette with the
black side facing up.
(3) Ensure that the entire image plate
is fully exposed. The collimated
field must be larger than the image
plate.

Black side
facing up
Cu filter
Collimated
area
5175_c36_001.cdr
1,3 m



Figure 8

(4) Select the following exposure parameters to obtain a dose of 10 Gy:
12 mAs
75 kVp
1.3 m (51,2) distance
Large focus
(5) Expose the cassette.
(6) Turn the cassette 180 (black side suntil facing up).
(7) Expose the cassette again.
After the second exposure the cassette has been exposed in total
with a dose of approximately 20 Gy.
(8) Remove the Cu-filter from the X-ray device.

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 12 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.3.3.2 Exposing the Cassettes at a Mammography Modality


(1) Select the movdality according to
following preference list, if more
than one Mammography modality
is available on site:
1. Siemens Mammomat
2. GE DMR
3. Instrumentarium
(2) Mount the Al-filter at the modality.
(3) Insert the Mammo cassette with
the black side facing up.



Figure 9

(4) Select the following exposure parameters:
200 mAs
28 kVp
Mo source / MO filter
(5) Expose the cassette.
(6) Repeat the exposure for the other cassette format, if available.
(7) Remove the Al-filter from the X-ray device.

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 13 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.3.4 Scanning the Flatfield Image

(1) Insert the exposed cassette in the digitizer.
(2) At the NX workstation select the examination type:
<System Diagnosis Flatfield>

IMPORTANT:
By selection of examination type "System Diagnosis Flatfield", the
correct settings for image processing of the flatfield are used. Any other
examination type may give a result which cannot be compared with the
limit pattern.
This exam type is only available, when working as user crservice.

(3) Identify the cassette.
(4) Wait until the digitizer has finished the scan process.
(5) Print the image or send it to the archive (depending on what is available).
(6) Keep the image as reference image "before maintenance".
(7) Put the scanned cassette aside. It is used again for checking the image quality
at the end of the preventive maintenance.


Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 14 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3 Maintenance Power Off

The maintenance power off is split up into the following steps:

# Step Section
1 Checking of power and network cable. 3.1
2 Checking of the covers. 3.2
3 Checking of the cables. 3.3
4 Checking of the IP guide plate. 3.4
5 Checking of the four belt drives. 3.5
6 Performing a mechanical check of the drawer slider. 3.6
7 Performing a visual check of the transport rollers. 3.7
8 Performing a visual check of all Motors. 3.8
9 Cleaning the inside. 3.9
10 Cleaning the erasure unit. 3.10
11 Cleaning the prescan antistatic brush. 3.11
12 Greasing the spindle at the cassette fixation unit. 3.12
13 Reassembling of the removed parts. 3.13
14 Checking the cassette condition. 3.14


NOTE:
When performing a visual check of the digitizer make sure that the components are
not damaged. If damage is visible replace the defective part.
For more information of the replacement refer to the CR 30-Xm / CR 30-X 2nd
Service Manual, chapter 3.5 Replacements / Repair Procedures.

3.1 Checking of Power and Network Cable

(1) Switch off the digitizer.
(2) Unplug the mains and network cable at the digitizer.
(3) Check the condition of both cables for:
Damages at insulation
Symptoms of bending or squeezing
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 15 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3.2 Checking of the Covers


CAUTION:
Parts underneath the top cover can be damaged.
Carefully remove the top cover.

(1) Remove all covers of the digitizer. For instructions refer to CR 30-Xm /
CR 30-X 2nd Service Manual, chapter 3.5 Replacements / Repair Procedures.
(2) Check all covers for damages (e.g. cracks).


3.3 Checking of the Cables

(1) Check the condition of all cables
inside of the digitizer.
(2) Have a close look to the cables
that are moved when the cassette
unit is opened. See Figure 10.

Figure 10

(3) Remove the screw which fixates
the Handling Control Board and
lift the board slightly up.
(4) Check the flexible cable of the
Chip Card Reader for any signs
of damage.

Figure 11

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 16 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3.4 Checking of the IP Guide Plate


CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.

(1) Remove the Erasure Unit and IP guide plate. Refer to CR 30-Xm / CR 30-X 2nd
Service Manual, chapter 3.5 Replacements / Repair Procedures.

(2) Check the condition of the IP
guide plate. It may not be bent
or damaged (e.g. scratches).


Figure 12


3.5 Checking of the four Belt Drives

belts_cassette_unit.cdr
Roller Clamping Motor Belt
Toothed Belt for
Opener Motor M3
Drawer Motor Belt
Transport Roller Motor Belt
(1) Check the condition of the 4 belt
drives in the cassette unit.

Especially check for:
Abrasion in the area of the
belts
Worn out belts (e.g. broken
steel cores)
Worn out gears

(2) Move the belts forward and
backward manually when checking.
Figure 13

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 17 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

3.6 Performing a mechanical Check of the Drawer Slider

(1) Move the slider backward: It must
move smoothly.
(2) Put the slider back to home position
(i.e. attached to the magnets)


Figure 14

3.7 Performing a visual Check of the Transport Rollers

(1) Check that the surface of the
transport rollers in the cassette
unit is free from defects.

Figure 15

3.8 Performing a visual Check of all Stepper Motors


NOTE:
The following conditions listed below are prerequisites to perform a successful
stall calibration. For more information refer section 4.2 Performing a Stall Calibration.

(1) Check the following conditions of the stepper motors which are using a
stall detection (home/end position):
Smooth-running of the motor drive
Functionality of the mechanism for detection of the home/end position
No barriers in the movement available
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 18 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 19 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential
3.9

Cleaning the Inside

REQUIRED TOOLS:
Vacuum cleaner

(1) Clean the inside with a vacuum cleaner.
Especially check for dust in the IP transport unit, and clean this thoroughly
cleaner.
3.10
with the vacuum


Cleaning the Erasure Unit

REQUIRED TOOLS:
Lint-free cloth



uring operation the erasure unit is hot: Risk of burns.
th the Erasure Unit.
the relevant sticker on the Erasure Unit.

WARNING:
D
Avoid contact wi
Observe
200
c


CAUTION:
Sensitive surface. Reduced image quality possible due to clouded reflector.
Do not touch the reflector of the erasure unit by hand.
Do not use cleaning agents to clean the reflector.
Only use a lint-free cloth

for cleaning.


(1) Check the reflector, i.e. the lower side of the erasure unit.
(2) Remove dust or any other particles with a lint-free cloth.






DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 20 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential
3.11 Cleaning the Prescan Antistatic Brush
(1) emove the Optic Module.
efer to CR 30-Xm / CR 30-X 2nd
ual, chapter 3.5
rocedures.
c

R
R
Service Man
Replacements / Repair
P
(2) Check the Prescan Antistati
Brush for contamination by using
a flash light.

Figure 16

(3) If dust is visible, clean the area
before the Prescan Antistatic
Brush with a small vacuum
cleaner nozzle.

Figure 17





DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3.12 Greasing the Spindle at the Cassette Fixation Unit


IMPORTANT:
Be careful when putting grease at the spindle. When grease gets into the digitizer
it can affect the image quality or the functionality of the digitizer.

(1) Put some grease (Isoflex Topas
NB 52) at the spindle in the area
of the spindle nut.
5175_c09_002.ai

Figure 18


3.13 Reassembling of the removed Parts

(1) Reassemble the removed parts:
Optic Module
IP Guide Plate
Erasure Unit
Covers

(2) Connect the mains and network cable.
(3) Switch on the digitizer.


Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 21 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3.14 Checking the Cassette Condition

(1) Check the following items of the cassettes:
Outside of the cassette:
No signs of wear or defects like cracks, broken layers, buckling of the
cassette, split off parts, worn or loose protection edges
Proper function of the shutter

Inside of the cassette:
No signs of wear or defects like detached fleece, cracks or split off parts
No signs of wear or defects of the tray like cracks, split off parts especially
on the rear edges
No contamination or signs of wear or defects of the IP like any kind of
particles, traces of glue, twisted edges, buckling of the IP, split off parts
of the layer or scratches

(2) If any of the above listed symptoms are detected, separate the defective
cassettes or IPs and instruct the operator / customer about the possible risks
of using defective cassettes or IPs.


NOTE:
It is strongly recommended to replace affected cassettes or IPs, if any of the above
listed symptoms are detected.



4 Maintenance Power On

The maintenance power on is split up into the following steps:

# Step Section
1 Checking the safety switch 4.1
2 Performing a stall calibration for all motors 4.2
3 Performing test cycles for each used cassette format 4.3

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 22 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
4.1 Checking the Safety Switch

(1) Pull out the cassette unit approximately 1 cm (0.4 in). The safety switch has
to switch off the digitizer.
(2) Close the cassette unit.


4.2 Performing a Stall Calibration


NOTE:
Only perform a stall calibration, if a specific error exists.


(1) On the NX workstation log-off the current user and log-in as "crservice".
(2) Open the browser.
(3) Enter the digitizer IP address in the address window.


NOTE:
If the IP address is not known: Look it up in the Configuration Tool of the
NX workstation.


(4) Enter username (default: mega) and password (obtained during the classroom
training).
(5) Select in the service menu: <Maintain and calibrate Stall
Calibration>
(6) Wait until the stall calibration is finished. This takes approximately 5 minutes.
For more information to the Stall Calibration refer to CR 30-Xm / CR 30-X 2nd
Service Manual chapter 3.6 Adjustments and Calibrations.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 23 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
4.3 Performing Test Cycles

(1) Insert the cassette in the digitizer.


NOTE:
Perform the test cycle with every used cassette format.

(2) In the Service GUI select: <Analysis and Repair Endurance Run
Cycle>
(3) Keep the default settings.
(4) Select START.
(5) Check for proper IP transport and abnormal noise.
(6) Repeat this test cycle for all other cassette formats used at the site.


Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 24 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
5 Technical Image Quality Check of the System

The technical image quality check is split up into the following steps:

# Step Section
1 Expose the cassette. 5.1
2 Clean the Scan Line 5.2
3 Scan the flatfield image 5.3
4 Checking the Image at the Lightbox or Viewing Station 5.4


5.1 Exposing the Cassette

Use the same cassette as in the beginning of the preventive maintenance.
Refer to the instructions in section 2.3 of this document.


5.2 Cleaning of the Scan Line and the Mirror

(1) Check the condition of the cleaning
brush (e.g. brush is not dirty or
damaged).
(2) Clean the scan line by wiping with
the cleaning brush a few times
over the whole scan length.

The last movement must be
continuous from rear to front.


Figure 19



Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 25 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
5.3 Scanning the Flatfield Image

(1) Insert the exposed cassette in the digitizer.
(2) At the NX workstation select the examination type:
<System Diagnosis Flatfield>


IMPORTANT:
By selection of examination type "System Diagnosis Flatfield", the
correct settings for image processing of the flatfield are used. Any other
examination type may give a result which cannot be compared with the
limit pattern.

(3) Identify the cassette.
(4) Wait until the digitizer has finished the scan process.
(5) Print the image or send it to the archive (depending on what is available).


Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 26 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
5.4 Checking the Image at the Lightbox or Viewing Station


NOTE:
Checking the images and comparing to the limit patterns may only happen on
lightboxes and viewing stations that comply with the specified installation and
configuration conditions for diagnostic reading. For details please refer to the
respective instructions of the viewing stations / light boxes.


NOTE:
The slow scan direction is always parallel to the short side of the image plate
(except 35 x 35 cm cassettes). Due to the hanging protocol* the appearance of the
saved flatfield might differ from the physical orientation of the cassette as it has
entered the digitizer.
* The hanging protocol (DICOM expression) defines the format and presentation how the images are
displayed on the monitor.


43 cm
35 cm
24 cm
30 cm 30 cm
15 cm
517509ab.cdr
18 cm
24 cm
Slow scan
direction

Figure 20


(1) Check the image quality of the flatfield for the following artifacts:
Stripes in fast scan or slow scan direction
Large area inhomogeneities
Unacceptable number of white dots.


Evaluation:


(2) If the scanned flatfield does not show one of the artifacts, no further action
is required.

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 27 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.


(3) If the scanned flatfield shows one of the artifacts compare the flatfield with the
limit pattern. Compare the image with the limit pattern set. For detailed
instructions see enclosure document which is part of the limit pattern set.
The limit pattern set is in scope of the delivery of the digitizer or is available in
Mednet GSO Library. It is also available as spare part:
Limit Pattern Set for verification CR 30-X 2nd:
Spare part number* CM+9 5175 9160 1
Limit Pattern Set for verification of CR 30-Xm:
Spare part number* CM+604948 00


IMPORTANT:
If a FLFS cassette for flatfield exposure is used, the image will be slightly darker
in the upper and lower image area where the backscatter protection is removed
(approximately 1 cm each). These darker zones also have to be used for image
quality evaluation in slow scan and fast scan direction.

Flatfield with Genrad Cassette
Flatfield with FLFS Cassette
Effect exaggerated


FLFS
FLFS
10 mm
10 mm
Backscatter
Protection
Genrad Cassette FLFS Cassette


Figure 21

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 28 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E Maintenance




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
6 Completion of the Maintenance

The completion of the maintenance is split up into the following steps:

# Step Section
1 Resetting the Maintenance Indicator 6.1
2 Creating a Backup of Device specific Data 6.2
3 Performing the Customer Conversation 6.3
6.1 Resetting the Maintenance Indicator

(1) In the digitizer service menu select: <Maintenance Indicator>
(2) Select: <Reset of Maintenance Indicator>


NOTE:
Resetting the maintenance indicator also resets the relative counters.


6.2 Creating a Backup of Device specific Data

(1) In the service menu select: <Backup & Restore>
(2) Leave default settings.
(3) Select <Complete Backup> to perform a backup.
(4) Select <Save to PC Hard disk>.
(5) Select <Logout>.
(6) Copy the Backup ZIP file from The NX Workstation to the digitizer USB Flash
Drive and the Service PC.


6.3 Performing the Customer Conversation

(1) Fill out the Maintenance checklist and hand it out to the customer.
(2) Explain the results of the maintenance to the customer.

Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 29 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E
Maintenance
Checklist




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
7 Maintenance Checklist

Digitizer Serial Number: Image Plate Cycles and months since last
maintenance:


The maintenance has to be carried out:
Every 30.000 Image Plate Cycles or
Every 2 years


NOTE:
Maintenance must be carried out according to the maintenance instructions:
DD+DIS231.10E

Diagnostics OK Not OK Solved
Questioning of the customer
Analyzing the Info Counter
Documenting the Technical Image Quality of the System

Maintenance points - Power off OK Not OK Solved
Checking of Power and Network Cable
Checking of the Covers
Checking of the Cables
Checking of the IP Guide Plate
Checking of the four Belt Drives
Performing a mechanical Check of the Drawer Slider
Performing a visual Check of the Transport Rollers
Performing a visual Check of the Stepper Motors
Cleaning the Inside
Cleaning the Erasure Unit
Cleaning the Prescan Antistatic Brush
Greasing the Spindle at the Cassette Fixation Unit
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 30 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential


DD+DIS231.10E
Maintenance
Checklist




DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X 2nd Type 5175 / 200 / 205 / 220 / 225 Chapter 9 / Page 31 of 31
02-2012 CR 30-Xm Type 5179 / 100 Agfa Company Confidential
Maintenance points - Power off OK Not OK Solved
Reassembling of the removed Parts
Checking the Cassette Condition

Maintenance points - Power on OK NOT OK Solved
Checking the Safety Switch
Performing a Stall Calibration
Performing Test Cycles

Technical Image Quality Check of the System OK NOT OK Solved
Exposing the Cassette
Cleaning of the Scan Line and the Mirror
Scanning the Flatfield Image
Checking the Image at the Lightbox or Viewing Station

Completion OK NOT OK Solved
Resetting the Maintenance Indicator
Creating a Backup of Device specific Data
Performing the Customer Conversation

Remarks:








............................................................. .........................................................
Date / Signature Service Technician Date / Signature Customer