Mtronix Precision Measuring Instruments reserves the right to, without notice, make changes to this manual and to the software and equipment to which this manual refers. Great care has been taken to ensure that this manual is free of errors and inaccuracies.
Mtronix Precision Measuring Instruments reserves the right to, without notice, make changes to this manual and to the software and equipment to which this manual refers. Great care has been taken to ensure that this manual is free of errors and inaccuracies.
Mtronix Precision Measuring Instruments reserves the right to, without notice, make changes to this manual and to the software and equipment to which this manual refers. Great care has been taken to ensure that this manual is free of errors and inaccuracies.
User's Manual Version: March 14, 2008 Copyright 2004 - 2008 by mtronix Precision Measuring Instruments. All rights reserved. This document may not be reproduced in any form, or transferred to any other media, without prior written permission by mtronix Precision Measuring Instruments. mtronix Precision Measuring Instruments reserves the right to, without notice, make changes to this manual and to the software and equipment to which this manual refers. Disclaimer Great care has been taken to ensure that this manual is free of errors and inaccuracies. However, mtronix Precision Measuring Instruments disclaims all responsibility and liability for errors or omissions of any kind, including, but not limited to, any implied warranties of merchantability or fitness for a particular purpose of this manual and the equipment and software described therein. mtronix Precision Measuring Instruments shall not be held responsible for any incidental or consequential damage resulting from the use of this manual, or the equipment and software to which this manual refers. mtronix Precision Measuring Instruments is a member of the OMICRON group. Technical Service and Main Distributor OMICRON electronics GmbH +43 (0)5523 507-0 Oberes Ried 1 +43 (0)5523 507 999 6833 Klaus lwww.omicron.at Austria info@omicron.at mtronix Precision Measuring Instruments +49-(0)30-39889590 Lehrter Str. 16-17 +49-(0)30-39889594 D-10557 Berlin linfo@mtronix.de Germany www.mtronix.de Universal Current Measuring System MI 600 Users Manual CONTENTS 1 GETTING TO KNOW YOUR MEASUREMENT SYSTEM........................................................................5 1.1 MI 600 Sensor Unit.............................................................................................................................6 1.2 MCU 502 Fiber Optics Controller and Control Station PC.....................................................7 1.3 mtronix Software..................................................................................................................................7 1.4 Setting up the Measurement System For Use.............................................................................8 2 THE SOFTWARE WORKSPACE...................................................................................................................9 2.1 Status Displays..................................................................................................................................11 2.2 The Sensor Display...........................................................................................................................12 2.3 The Visualization Display.................................................................................................................13 2.3.1 The System View......................................................................................................................13 2.3.2 The Scope View........................................................................................................................15 2.3.3 The Trend View.........................................................................................................................16 2.4 The Control Panel.............................................................................................................................17 2.4.1 The General Tab.......................................................................................................................17 2.4.2 The Trend Tab...........................................................................................................................21 2.4.3 The Calibrate Tab.....................................................................................................................23 APPENDIX A - KEYBOARD SHORTCUTS.................................................................................................26 Gain-Settings...........................................................................................................................................26 Display-Settings......................................................................................................................................26 APPENDIX B - TECHNICAL SPECIFICATIONS........................................................................................27 Sensor Units.............................................................................................................................................27 Hardware/Software Requirements.....................................................................................................27 System Data.............................................................................................................................................28 3 APPENDIX C - ADVANCED MEASUREMENT INSTALLATION..........................................................29 3 Universal Current Measuring System MI 600 Users Manual LIST OF FIGURES Figure 1 - Measurement System consisting of a laptop, an mtronix MCU 502 fiber optics controller and two mtronix MI 600 sensor units with batteries.............................................................5 Figure 2 Front and back panel of the MI 600 Sensor Unit................................................................6 Figure 3 Front panel of the MCU 502 Fiber Optics Controller.......................................................7 Figure 4 Measurement Setup...................................................................................................................9 Figure 5 - Software Workspace................................................................................................................10 Figure 6 - Sensor Display...........................................................................................................................12 Figure 7 - Sensor Display with sensor information window and warning message.....................12 Figure 8 - System View...............................................................................................................................15 Figure 9 - Scope View.................................................................................................................................15 Figure 11 - The measured quantities display.........................................................................................16 Figure 10 - Trend View................................................................................................................................17 Figure 12 - Control Panel, General Tab..................................................................................................19 Figure 13 - Control Panel, Trend Tab .....................................................................................................23 Figure 14 - Control Panel, Calibrate Tab................................................................................................25 4 Universal Current Measuring System MI 600 Users Manual 1 GETTING TO KNOW YOUR MEASUREMENT SYSTEM The MI 600 universal current measuring system is a high-precision, modular acquisition and analysis system for gauging current and key characteristics of electrical equipment, including dissipation factor and capacitance. Complete electrical insulation between acquisition units and the control station provide superior safety in high-voltage setups. High-resolution digital processing enables exceptional measurement precision. The easy-to-use mtronix software features various real-time visualization and monitoring options, and integrates with multiple mtronix products. The MI 600 Universal Current Measuring System (see Figure 1) consists of the following components: Two MI 600 sensor units (the reference sensor and the test object sensor), an MCU 502 or USB 502 fiber optics controller, and a PC running the mtronix software. Getting To Know Your Measurement System 5 Figure 1 - Measurement System consisting of a laptop, an mtronix MCU 502 fiber optics controller and two mtronix MI 600 sensor units with batteries. Universal Current Measuring System MI 600 Users Manual 1.1 MI 600 SENSOR UNIT The MI 600 sensor units (SUs) are connected to the control station PC by means of a high- speed fiber optical network. This FO network provides complete insulation from the high- voltage conditions to which measurement units may be subjected, and establishes a highly accurate absolute time reference amongst the sensor units. The MCU 502 fiber optics controller drives the FO network and connects it to the control station PC. Each sensor is connected to the MCU 502 fiber optics controller on its own fiber optical strand. In general, the FO network is organized as a ring, i.e. a single optical fiber connects two neighboring measurement units. To support this ring architecture, each measurement unit has two optical connectors: a receive port (red color coded and a transmit port (black color coded). A unit sends data through its transmit port to the next unit in line, which receives it on its receive port. The first unit receives data from the MCU 502 fiber optics controller on its receive port, and the last unit sends its data to the fiber optics controller through its transmit port, which completes the ring structure. In the most common setup, there will be two MI 600 sensor units (the reference and the test object sensor, respectively), one per ring. Optionally, an additional PD unit (MPD 600 or MPD402) for PD measurements may be connected to one of the rings. In addition to the FO connectors, each sensor unit has one TNC input, which receives the current signal. 6 MI 600 Sensor Unit Figure 2 Front and back panel of the MI 600 Sensor Unit. Universal Current Measuring System MI 600 Users Manual 1.2 MCU 502 FIBER OPTICS CONTROLLER AND CONTROL STATION PC The MCU 502 fiber optics controller connects the control station PC to the fiber optics network. A single MCU 502 controller supports two ring networks, and consequently has four optical connectors. A transmit and receive port for ring 1and a transmit and receive port for ring 2. The MCU 502 features a USB 2.0 port, which is used to connect the MCU 502 fiber optics controller to the control station PC. Note that the control station PC must be compatible with USB 2.0, PCs that support USB 1.1 only will not work with the MCU502. Two status LEDs (one green and one red) provide information on the system status. The green LED reflects the status of the FO network(s) connected to the controller. If at least one ring is connected and working properly, the green LED illuminates. The red LED is lit when the software is not using the fiber optics bus, and is switched off when communication between the software and the units is up and running. Figure 3 Front panel of the MCU 502 Fiber Optics Controller. 1.3 MTRONIX SOFTWARE The mtronix software running on the PC receives measurement data from each unit via the FO network for analysis, display, and storage. The software is also responsible for controlling and configuring the sensor units Getting To Know Your Measurement System 7 Universal Current Measuring System MI 600 Users Manual 1.4 SETTING UP THE MEASUREMENT SYSTEM FOR USE Before measurements can be performed using the mtronix Software, the measurement environment must first be set up. To prepare the system for use, follow these steps: 1. Install the sensors in your measurement setup (see ). The reference sensor is installed in the branch with the reference capacitor (the reference branch). Conversely, the test object sensor is installed in the branch with the object whose dissipation factor is to be determined (the test object in the test object branch). 2. Connect the MCU 502 fiber optics controller to a USB port on the computer. (The USB port must be compatible with USB 2.0.) 3. Connect the reference sensor to fiber optics bus 1 on the MCU 502 fiber optics controller. The red color coded connector of the sensor must be connected to the red color coded plug at the fiber optics controller. 4. Connect the test object sensor to fiber optics bus 2 on the USB fiber optics controller in the same manner as the reference sensor. 5. Start the mtronix Software by double-clicking the icon on the desktop. IMPORTANT: The sensors have to be placed isolated to the ground to prevent device damages and avoid ground loops and stray capacitances which are influencing the measurement results. The fiber optical cables for both sensors must be of same length. The measurement cables (TNC cables) must be as short as possible and of same length as well. The software will now detect and initialize the two sensors. Each sensor is equipped with a red and a green LED. As soon as the sensor has power, the red LED will begin to flash. During the initialization phase, the LEDs on the sensors will progress through various stages: First the red LED will change from flashing to steadily lit. When the software has detected the sensor the green LED will illuminate in addition to the red LED, and the software will begin configuring the sensor. Once the configuration process is complete, the red LED will turn off and only the green LED will remain on to indicate that the sensor has been successfully configured. 8 Setting up the Measurement System For Use Universal Current Measuring System MI 600 Users Manual Once both sensors are shown and sensor initialization is complete, the system is ready for use. Getting To Know Your Measurement System 9 Figure 4 Measurement Setup Universal Current Measuring System MI 600 Users Manual 2 THE SOFTWARE WORKSPACE The mtronix Software workspace is divided into four major parts: 1. The sensor display on top C, 2. the visualization display on the left C, 3. the control panel on the right C, and 4. the status displays at the bottom (log area C, system status bar , progress bar , and fiber optics status indicator 1) which indicate the current condition of the software (Refer to Figure 5). 10 The Software Workspace Figure 5 - Software Workspace Universal Current Measuring System MI 600 Users Manual 2.1 STATUS DISPLAYS The following status indicators provide information about the current system status: The log area C displays status and warning messages in user-readable form. The system status bar shows the initialization status of the software by means of 5 LEDs. During startup of the software, the LEDs in the system status bar will be illuminated in succession until all five LEDs are lit. The progress bar visualizes the progress of a certain operation. This is used, for example, to show the initialization progress of newly connected sensors. A message to the left of the progress bar shows what specific operation is being performed (or ready. if no operation is pending). The fiber optics status indicator 1 shows which fiber optics channels are properly connected. Each fiber optics channel corresponds to one sensor. During normal operation, both LEDs must be green. If one of the sensors is off line, or the connection between the sensor and the USB controller is disturbed, the corresponding LED will turn red. Temporary errors on the fiber optics link will cause the LED to turn yellow for the duration of the error. The Software Workspace 11 Universal Current Measuring System MI 600 Users Manual 2.2 THE SENSOR DISPLAY The sensor display shows which sensors have been detected by the software. Each detected sensor is represented by a colored rectangle. The color of the rectangle represents the status of the battery or wall-plug power supply to which the sensor is connected. A green rectangle indicates that the power supply voltage is good, while a red rectangle signifies that the voltage is too low, and the battery should be replaced or recharged to avoid deep discharge. Figure 6 - Sensor Display A sensor can be selected in the sensor display by clicking on the corresponding colored rectangle; the selected sensor then shows a dashed selection frame around its box. Selecting a sensor prompts the software to show the rest of the workspace, which is not visible if no sensor is selected. To perform measurements, both the reference sensor and the test object sensor must be present. Additionally, a sensor must be selected in the sensor view, although it is irrelevant which one. The mtronix Software will automatically select the reference sensor as soon as both the reference sensor and the test object sensor have been detected. To hide the visualization and control panel views, click anywhere on the empty space next to the sensors. Hovering the mouse over a selected sensor rectangle will pop up a sensor information window, which shows the serial number of the sensor, its current supply voltage, and, if the voltage is too low, a warning message (see Figure 7). 12 The Sensor Display Figure 7 - Sensor Display with sensor information window and warning message Universal Current Measuring System MI 600 Users Manual 2.3 THE VISUALIZATION DISPLAY The visualization display is only visible if a sensor is selected in the sensor display. It takes up most of the left side of the workspace. The visualization display is further subdivided in the Scope/System View section C, the Trend View section C, and the Measured Quantities Display (refer to Figure 5). You can switch between the Scope and System views by clicking on the corresponding tab on the right of the Scope/System View area. 2.3.1 The System View The system view visualizes the wiring of the measurement setup in an interactive circuit diagram. The two sensors are each shown as a box with a level indicator and a gain slider. The level indicator shows the input level at the corresponding sensors TNC connector. It consists of 16 LEDs, which illuminate according to the input level. The lower 14 LEDs are green when illuminated, the 15 th LED is yellow, and the 16 th LED is red. If the 16 th (red) LED illuminates, the input is overdriven. Each illuminated LED represents an input level difference of 3.7 dB. The gain slider allows the user to select the gain level that is applied to each sensors input. Moving the slider upwards increases the gain, and hence increases the sensors sensitivity. When a sensor is connected or disconnected, and when the software starts up or shuts down, the gain is automatically set to 0. Note that while the gain slider allows 11 different settings, only the lower 8 can be selected by the user unless a shunt is installed or high-sensitivity mode is selected (see Configuration Panel section below for more information).There are some keyboard shortcuts for often-used gain settings. Pressing the F9 key resets the gain on both sensors to the lowest (safest) setting, while pressing F8 automatically chooses a gain setting that is optimal with respect to the input level on both sensors. Additionally, you can use F2 and CTRL-F2 to manually increase the gain on the reference and test object sensors, respectively. Likewise, pressing F3 and CTRL-F3 will reduce the gain on the corresponding sensor. The Software Workspace 13 Universal Current Measuring System MI 600 Users Manual To change between the system view and the scope view, you can either click on the corresponding tab or press the F4 key. Users may install a shunt in either branch to enable measurements in high-voltage scenarios. If a shunt is present in the measurement setup, it must be enabled in the software by choosing the corresponding shunt option from the drop-down box next to the sensor. Doing so will update the wiring diagram to reflect the presence of the shunt, and the shunts characteristics will be shown in the diagram. Users can add, remove, and edit shunt options from the Calibrate panel (see below). As all calculations in the software are based on the wiring depicted in the system view, it is important to ensure that the actual setup matches the wiring diagram. 14 The Visualization Display Figure 8 - System View Universal Current Measuring System MI 600 Users Manual 2.3.2 The Scope View The scope view displays the current in both branches as a curve over time. The time base (i.e. the time corresponding to one horizontal division), the color of each curve, the vertical scale of each curve (i.e. the current corresponding to one vertical division), the vertical offset of each curve, and the horizontal position (i.e. the time corresponding to the left border) can be adjusted through the control panel. A small colored triangle on the left border of the view shows that vertical position which corresponds to 0 A. In addition, a thick gray line is drawn at each curves zero level. Both the horizontal and the vertical axes are labeled. The time values shown on the horizontal axis are with reference to the zero point of the current curve. Vertical axis labels are shown for both curves; labels for the reference sensors current curve are shown on the left-hand side, while the test object sensors curve is labeled on the right. Alternatively, the user can choose to show the reference voltage rather than the reference branch current in the scope view. Press F6 to toggle between the voltage and current view. The Software Workspace 15 Figure 9 - Scope View Universal Current Measuring System MI 600 Users Manual 2.3.3 The Trend View Figure 10 - Trend View Several measured quantities can be recorded into trend curves. These curves plot the respective quantity over time, and are visible in the trend view. Users can choose which trend quantities to show and can set display properties of each trend curve (including color, vertical scale, vertical offset, and time base). Vertical axis labels can be enabled for a single trend quantity only, but users can choose the quantity for which axis labels should be shown. The horizontal axis plots the actual date and time at which the quantities were measured. Note that the time is given in Universal Coordinated Time (UTC, also known as GMT, or Greenwich Mean Time). The Measured Quantities Display This display area shows the current values of the measured quantities. The following quantities are shown: 1. The current dissipation factor (tan ), 2. The RMS value of the voltage, 3. The RMS value of the current through the test object, 4. The measured frequency of the voltage, 5. The measured capacitance of the test object. If a reliable measurement cannot be performed, the frequency value will be shaded red, as shown in . The following reasons can lead to a measurement not being reliable: 16 The Visualization Display Figure 11 - The measured quantities display Universal Current Measuring System MI 600 Users Manual The input level is too low, The frequency of the voltage changes too quickly (i.e. the frequency drift is too large), One of the sensors is not connected to the USB fiber optics controller, or One of the sensors is not correctly installed in the appropriate branch. The measured quantities display is updated every 0.3 seconds. 2.4 THE CONTROL PANEL The behavior of the software is controlled through the control panel. Like the visualization display, it is only visible if a sensor has been selected in the sensor display. The control panel is divided into three tabs: Settings that affect the measurement itself and the scope view are contained in the General tab, trend view settings are controlled via the Trend tab, and the Calibrate tab allows users to calibrate certain system parameters. 2.4.1 The General Tab The General tab contains display settings for the scope view and general measurement settings. Display settings Display settings can be individually configured for the two sensors. In general, there are two display modes for the scope view: In auto-scale mode, the scale of the current curve will be automatically chosen by the software in such a way that the diagram matches the available dynamics of the sensors input at the selected input sensitivity. This means that the top border of the diagram corresponds to full gain (for the selected input sensitivity). When showing the reference voltage rather than the reference branch current, the auto-scale setting has a different effect: The curve is scaled such that it neatly fits into the diagram, and that the Volts per divisions (V/div) setting is not a rational value. In manual mode, users specify the scale at which the curve is shown. The scale is specified in Amperes per divisions (A/div) when the reference current is shown and in Volts per divisions (V/div) if the reference voltage curve is visible. In manual mode users may additionally specify an offset by which the current curve is displaced from its The Software Workspace 17 Universal Current Measuring System MI 600 Users Manual natural zero point. This is useful for zooming in on a portion of the curve and examining a specific point in time. The Auto scale checkbox determines whether auto-scale mode is active or not. If the box is checked, auto-scale mode is in effect and both the scale and offset fields are unavailable for the corresponding sensor. You can also press F5 to toggle between auto-scale and manual mode for the reference sensor; CTRL-F5 has the same effect for the test object sensor. 18 The Control Panel Figure 12 - Control Panel, General Tab Universal Current Measuring System MI 600 Users Manual A color button provides access to a dialog which allows users to specify the color of each curve. The general tab also allows users to choose whether they want to see the reference current or the reference voltage, which is obtained by numerically integrating the current signal on the reference sensors input. Pressing F6 toggles between the two possibilities. Additionally, the user can choose whether the peak-to-peak voltage or the rms voltage is to be shown in the Measured Quantities display. The shortcut CTRL-F6 toggles this setting. The Timebase field is used to specify the horizontal resolution of the diagram, given in time per horizontal division. Time specifications can be anything from 1 as (atosecond) to several days. While the software understands days (d), hours (h), minutes (m), seconds (s), milliseconds (ms), microseconds (s or us), nanoseconds (ns), picoseconds (ps), femtoseconds (fs), and atoseconds (as), useful timebases will usually be in the range of a few milliseconds to a number of seconds. The horizontal position field allows users to specify the time value corresponding to the left border. Times are based on the curves zero point on a rising slope, so a horizontal position of 0 s means that the zero point of the curve coincides with the left border. The Fit to period checkbox enables automatic horizontal scaling. If checked, the software will automatically choose the scaling such that exactly one full period of the curve is visible, with the horizontal position set to 0. The software features a visual aid for moving the horizontal position of the diagram in the form of a cursor. It consists of a thin vertical line connecting two small handles. Users enable the cursor via the Show cursor checkbox. Once enabled, the cursor (and through it the horizontal position of the diagrams left border) can be moved by dragging the top handle with the mouse. The Anchor to cursor setting specifies how changes to the timebase field affect the horizontal position field. If the Anchor to cursor setting is disabled, the horizontal position will not change when the timebase is adjusted. Enabling the setting causes the software to recalculate the horizontal position every time the timebase is changed. The recalculation is performed such that the zero point of the curve always remains at the same screen position. This is very useful for zooming in on the vicinity of the curves zero point. The Software Workspace 19 Universal Current Measuring System MI 600 Users Manual Measurement Settings The Measurement interval field determines how much data from the sensors input is considered for the calculation of the measured quantities. The longer the time is, the more data is evaluated during dissipation factor calculation, and the less noisy the calculated data will be. Valid values for the measurement interval range from 300 ms to 100 s (1 m 40 s). The Invert test object checkbox allows users to specify that the test object sensor is connected to the high-voltage side of the test object instead of the ground side. Checking the box in such a case causes the software to take the changed wiring into account during capacitance and dissipation factor calculation. The Ranging section allows users to specify the ranging mode of the software. The ranging mode determines if and how the software automatically switches between different input gain levels. The following options exist: Manual mode: No automatic switching between input gains takes place. Users must monitor the input level and set an appropriate gain manually. Semi-automatic mode: The software will automatically set a lower gain if the input is overdriven. No automatic gain adjustment will be performed if the input level is too low. Automatic mode: The software will automatically select the most appropriate input gain for the actual input level. Users should note that during input gain switching, the software will be unable to measure the current dissipation factor for a small amount of time (about 300 ms). If uninterrupted measurement is required, the ranging mode should be set to manual. By pressing F7, the user can cycle through the available ranging options. The High sensitivity setting reconfigures the sensors such that a much higher input sensitivity can be achieved. This enables users to work with small measurement currents and still get accurate measurements. However, enabling high-sensitivity mode will make the sensors much more susceptible to damage from overcurrents and surges. It is advised that users only employ high-sensitivity mode if the measurement currents are too low otherwise. The software automatically turns off high-sensitivity mode upon startup and shutdown to protect the sensors. Pressing CTRL-F7 toggles the high sensitivity option. Note that all of the ranging settings apply to both sensors. The Frequency detection section allows users to specify how the frequency of the input current is to be determined. If the Automatic checkbox is enabled, the software will automatically try to determine the input frequency. This is the preferred mode of operation. 20 The Control Panel Universal Current Measuring System MI 600 Users Manual However, in extreme situations it may be necessary to manually preset the input frequency. This is required only if the input levels of the sensors are extremely low and cannot be increased or if the shape of the current curve significantly varies and cannot be approximated by a sinusoid of a fixed frequency. In those cases, users may specify the input frequency in the supplied field and turn off automatic frequency detection. 2.4.2 The Trend Tab The trend tab provides access to display and recording settings for the trend view. The Display settings section lists all quantities whose trends are recorded by the software. The checkbox on the left of each quantity enables the display of that quantity in the trend view, using the color determined by the color button (located to the right of the quantity name). Users can change the color by clicking on the color button and choosing a different color. The vertical scale of the quantitys curve shown in the trend view can be specified in the Scale field. The scale is given in quantity units per vertical division. For example, the frequency is given in Hertz pr division (Hz/div). The tan delta value does not have a unit, so its scale is given in 1 per division (1/div). In addition to the scale, a vertical offset can be applied to the curve. This is very useful for zooming in on a quantity that lies around a certain level. The Offset field provides access to that vertical offset, which is given in the quantitys units. The radio box to the right of the offset field determines for which trend quantity axis labels are drawn in the trend view. Only one quantity can have its vertical axis labeled at any time; this radio button selects which one does. The Timebase field specifies the horizontal resolution of the trend display, in time per horizontal division. The time specification can be anything from 1 atosecond to several days. The horizontal position setting determines which time value corresponds to the left border of the trend display. Since trend data is associated with a specific time and day, this value is relative to the first data point for which a trend was recorded (the so-called trend reference). This reference is shown below the horizontal position field. For example, if the trend reference is 15:05:06 (UTC), April 30, 2004, then a value of -10 s for the horizontal position means that the left border of the trend view corresponds to 15:04:56 (UTC), April 30, 2004. The Auto rescale setting causes the software to automatically rescale the trend view to show all of the available trend data as soon as the right border is reached. During rescaling, the horizontal position is set to 0, and the timebase is adjusted so that all of the available trend data fits in the first half of the display. The Software Workspace 21 Universal Current Measuring System MI 600 Users Manual The interval setting specifies at what rate the measured quantities are to be recorded into the trend. This option can be used to limit the amount of trend data that is generated during measurement. The minimum interval is 300 ms, because that is the rate at which the measured quantities are updated. Users can specify any interval, but note that the entered value will be rounded to a multiple of 300 ms. The Record settings section contains the settings pertinent to recording the trend data. Checking the Create log file box causes the software to store the recorded trend data into a CSV (Comma-Separated Values) file which is compatible with most spreadsheet applications, including Microsoft Excel. Users can specify the file name for the CSV file in the field to the right of the Create log file box, and may browse for a file using the Browse 22 The Control Panel Figure 13 - Control Panel, Trend Tab Universal Current Measuring System MI 600 Users Manual button. Checking the Create log file setting with a file that is already present will load that file into the display. Users can then zoom into the data contained within the file in exactly the same manner as if the data had been directly recorded during the same session. If trend data is available when a user selects the Create log file option with a new file, that trend data is saved into the new file. Note that trend files should never be directly recorded to floppy disks. Doing so may interrupt the measuring process and thus corrupt the calculated values. If trend files are to be saved to a floppy disk, they should be recorded to a hard disk or other fast media first, and be copied to the floppy disk later using the Windows Explorer. Pressing the record button (the button with the red circle) starts acquisition of trend data. To stop trend acquisition, press the record button again. During acquisition the trend view will show the newly acquired data in real-time. To record single values, use the Record now button (or press F11); this will record a single set of trend quantities. Clicking the Reset Trend button clears the trend data, the trend view, and if Create log file is checked the trend file. 2.4.3 The Calibrate Tab The Calibrate Tab contains settings that calibrate the MI 600 system as part of a user setup. The system delay field represents a time delay inherent in the measurement setup. Such delays can be caused by parasitical elements in the setup, or by using optical fibers with different lengths in the reference and test object branches. Specifying the correct value here enables the software to compensate for the system delay in the dissipation factor calculation. To determine the system delay, install a capacitor with an extremely small dissipation factor in the test object branch. A dissipation factor shown in such a setup will be exclusively caused by system delays. Clicking on Calibrate system delay instructs the software to compensate for that delay in its calculation, and the dissipation factor should consequently be shown as zero. The tan delta offset field allows users to specify a dissipation factor offset that is to be applied to all calculations. This is useful if one element in the setup has a known dissipation factor, and only the aggregate dissipation factor of any unknown elements is to be computed. The offset specified here will be added to any dissipation factor value computed by the software before being shown (and before being plotted in the trend diagram). The Software Workspace 23 Universal Current Measuring System MI 600 Users Manual Finally, the calibrate tab also includes the power supply warning threshold fields for both sensors. The values specified in these fields determine when a power supply warning is generated. As soon as the power supply voltage drops to or below the specified level, the sensor display will shade the sensors rectangle red (rather than green) and the sensor information window will show a power supply warning message. When operating a sensor with a battery, specify a suitable value here to prevent deep discharge of the battery. If a wall- plug power supply is used, specify 7 V here; this is the minimum voltage with which the sensor will function reliably. 24 The Control Panel Figure 14 - Control Panel, Calibrate Tab Universal Current Measuring System MI 600 Users Manual The Calibrate panel is also the place where external shunt options can be added, removed, or edited. Each shunt option is stored in a file which is located in the mtronix Software application folder. Upon start of the mtronix Software all shunt option files are loaded and are consequently available to the user. To edit an existing shunt option, choose one from the Shunt option drop-down box. Doing so will enable the edit fields for R Shunt , R Serial , and tau. These values correspond to the shunts characteristics. To save your changes, click on Save. To add a new shunt option, click Add. A window will open prompting you to enter the shunt options name. Once you have entered a name and clicked on Ok, you can proceed by editing the shunts characteristics in the corresponding edit fields. It is important to note that the shunt option will not be saved to a file until you press Save. Removing a shunt option works by selecting the corresponding entry from the drop-down box and clicking Remove. Once removed, the shunt option is no longer available for the present measurement session. The actual shunt option file will not be removed. Thus, the shunt option will again be available on the next start of the mtronix Software. In order to completely remove the shunt option you must manually remove the corresponding shunt option file. The Software Workspace 25 Universal Current Measuring System MI 600 Users Manual APPENDIX A - KEYBOARD SHORTCUTS GAIN-SETTINGS F2 / CTRL + F2 Manually increase the gain on the reference and test object sensors F3 / CTRL + F3 Manually reduce the gain on the reference and test object sensors F8 Automatically choose a gain setting that is optimal with respect to the input level on both sensors F9 Reset the gain on both sensors to the lowest (safest) setting DISPLAY-SETTINGS F4 Toggle between the system view and the scope view F5 / CTRL + F5 Toggle between auto-scale and manual mode for the reference and the test object sensor F6 Toggle between the reference current and the reference voltage CTRL + F6 Toggle between the peak-to-peak voltage and the rms voltage F7 Cycle through the available ranging options CTRL + F7 Toggle the high sensitivity option Annotation: A + means pressing both keys simultaneously. 26 Appendix A - Keyboard Shortcuts Universal Current Measuring System MI 600 Users Manual APPENDIX B - TECHNICAL SPECIFICATIONS SENSOR UNITS Material Extruded aluminum Dimensions 110 mm (W) x 190 mm (D) x 44 mm (H) Power Supply 9 12 V DC, max. power dissipation: 4 W (standby < 10 mW), external plug-in power supply included (input range 100 240 V, 50 60 Hz) Controls None, all functions fully remote-controlled Indicators 2 x LED: stand-by/power, optical fiber data integrity Fiber optic connectors 2 x ST, using multi-mode fiber 50 / 125 ?m up to 2 km in length Input connector 1 x TNC Temperature 0 C 45 C (operating), -10 C 60 C (storage) Humidity 5 % ~ 80 % non-condensing Input frequency range 5 Hz 50 kHz Input impedance 50 Ohms Input current range 20 A 100 mA rms (without external shunt); external shunts for 4A, 15A and 28A available HARDWARE/SOFTWARE REQUIREMENTS CPU Intel Pentium 4 (?2.5 GHz), Pentium M (?1.5 GHz), Core or Core 2 processor; or AMD Athlon 64 or Turion 64 processor; RAM 512 MB, (1GByte recommended) Hard disk 30 GB, min. UDMA-66 Graphics controller 2D-accelerated XGA with dedicated memory (min. 8 MB, no shared memory controllers), 1024 x 768 pixels resolution Connectivity 1 x USB 2.0 port Operating System Microsoft Windows 2000 Professional, XP or Vista Appendix B - Technical Specifications 27 Universal Current Measuring System MI 600 Users Manual SYSTEM DATA USB fiber optics controller 1 x USB 2, bus-powered 2 x FO Transmit (ST, using multi-mode fiber 50 / 125 m) 2 x FO Receive (ST, using multi-mode fiber 50 / 125 m) Dissipation factor display range 0.001E-5 (absolute) Dissipation factor error (2E-5 + 2% of display value) at 50 Hz Test object capacitance error 0.25 % of display value Frequency error 0.0025 % of display value Voltage error 0.5 % of display value Measurement interval 300 ms 28 Appendix B - Technical Specifications Universal Current Measuring System MI 600 Users Manual 3 APPENDIX C - ADVANCED MEASUREMENT INSTALLATION The following illustration shows how the MI 600 system can be extended to simultaneously measure dissipation factor and partial discharges using the MPD 600 Advanced Partial Discharge Analysis System. Figure 15 Setup for simultaneously measuring dissipation factor and partial discharge Appendix C - Advanced measurement installation 29