METROLOGY LAB LABORATORY MANUAL




Name: ..
Reg. No:
Branch: .
Year & Semester:


Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 2
METROLOGY LAB EXPERIMENTS
1. Measurement of lengths, heights, diameters by Vernier Calipers, Micrometers etc.
2. Measurement of bores by internal micrometers and dial bore indicators.
3. Linear measurement using dial gauge, slip gauge, and calibration of dial gauge
4. Use of gear teeth, Vernier Calipers and checking the chordal addendum and chordal height
of spur gear.
5. Angle and Taper measurements by Bevel protractor, Sine bars, etc.
6. Machine tool alignment test on the Lathe.
7. Machine tool alignment test on milling machine.
8. Tool makers microscope and its application.
9. Use of spirit level in finding the flatness of surface plate.
10. Thread measurement by two wire/three wire method or Tool makers microscope.
11. Surface roughness measurement by Taly Surf.
12 .To calibrate the profile projector using given samples which dimensions is measured by
micrometre as standard
13. Surface Wear Resistance Test using Electro Spark Coating Device.
CNC:
1. Machining of simple components on CNC lathe
2. Machining of simple components on CNC Milling
3. Inspection of quality and dimensional practice using Coordinate Measuring Machine





Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 3

Cycle-1-Experiments

Cycle-1-Experiments
SNO. NAME OF THE EXPERIMENT
1 Measurement of lengths, heights, diameters by Vernier Calipers, Micrometers
2 Measurement of bores by internal micrometers and dial bore indicators.
3 Linear measurement using dial gauge, slip gauge, and calibration of dial gauge
4 Use of gear teeth, Vernier Calipers and checking the chordal addendum and chordal
height of spur gear.
5 Angle and Taper measurements by Bevel protractor, Sine bars
6 Machine tool alignment test on the Lathe.

Cycle -2 experiments

Cycle-2-Experiments
SNO. NAME OF THE EXPERIMENT
1 Machine tool alignment test on milling machine.
2 Tool makers microscope and its application
3 Use of spirit level in finding the flatness of surface plate
4 Thread measurement by two wire/three wire method or Tool makers microscope
5 Surface roughness measurement by Taly Surf
6 To calibrate the profile projector using given samples which dimensions is measured
by micrometre as standard



Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 4
Experiment No. 1 Date: / / 20
MEASUREMENT OF LENGTHS, HEIGHTS, DIAMETERS BY
VERNIER CALIPERS, MICROMETERS
Aim:
To measure the diameters of the given work piece at various sections using Vernier Calipers.
Equipment Required:
1. Vernier Calipers with Least Count = 1mm/50 OR 0.02mm
2. Work piece of various cross sections with different diameters.
Construction:-
Vernier consists of 2 scales one fixed and other movable. The fixed scale known as the main
scale is calibrated on L shaped frame and carriers a fixed jaw. The movable vernier scale
slides over the main scale and carriers a measuring tip when the jaws are closed the zero of
vernier and main scale coincide. An adjustment is provided to lock the sliding scale.
Principle:
Vernier Calipers is the most commonly used instrument for measuring outer and inner
diameters. It works on the principle of Vernier Scale which is some fixed units of length (Ex:
49mm)divided into 1 less or 1 more parts of the unit(Ex: 49mm are divided into 50 parts).The
exact measurement with up to 0.02mm accuracy can be determined by the coinciding line
between Main Scale and Vernier Scale.
Total Reading = M.S.R + L.C X V.C
Where:
M.S.R Main Scale Reading
L.C Least Count
V.C Vernier Coincidence
Use:-
These are used for both internal and external measurement, its generate used for measuring
by closing the jaws on work surface and taking readings from main scale is examined to certain
which of its division coincide and added to the main scale reading.
Following are the constructional parts of vernier caliper:
(1) Fixed scale and movable scale: The Vernier Caliper consists of two scales: one is fixed and
the other is movable.
(2) Fixed and movable jaw: The fixed scale is called as main scale which is calibrated on L-
shaped frame and carries a fixed jaw. The movable scale, called vernier scale slides over the
main scale and carries a movable jaw. The movable jaw as well as the fixed jaw carries
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 5
measuring tip. When the two jaws are closed the zero of Vernier scale coincides with the zero
of main scale. For precise setting of the movable jaw an adjusting screw is provided.
(3) Lock nut: An arrangement is provided to lock the sliding scale on the fixed main scale.
(4) Graduated beam: Main scale markings are there on graduated beam.
(5) Blade or Depth probe: Measures depth.
Least count: The smallest value that can be measured by the instrument is known as its least
count.
Least count of Vernier Caliper:
There are two methods to find the least count of Vernier caliper
(a) First Method (Principle of Vernier)
Length of 49 divisions on main scale = Length of 50 divisions on Vernier scale
It means it follows that for the same length if there is n division on main scale then there should
be n+1 division on
Vernier Scale for the same distance.
Value of smallest division on main scale = 1 mm and
Value of smallest division on Vernier scale = 49/50 = 0.98 mm
Least count = Value of smallest division on main scale
value on smallest division on Vernier scale
= 1 0.98
= 0.02 mm
(b) Second Method
Least Count = Value of smallest division on Main Scale
Total no. of divisions on Vernier Scale
Smallest division on Main scale = 1 mm Total no. of divisions on Vernier Scale = 50 markings
So for this type of Vernier Caliper L.C. = 1/50 = 0.01mm
Least count= One division of main scale reading/ No. division on scale .mm
49 MSD=50 VSD
1 MSD=1mm
1 VSD= 49 MSD/50 VSD
L.C= 1MSD- 1VSD
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 6
= 1- 49/50
=0.02 mm

Reading a vernier caliper:


Formulae for calculating total reading with the help of Vernier caliper is
Total reading=MSR + (VSR X LC) mm
Here 12.84mm is the total reading. If 12.84mm is the total reading then the main scale reading
is 12. We know that the usual least count of V.C. is 0.02 mm.
12.84 =12+ (0.02Xvernier division coinciding with main scale division)
12.84-12 =0.02Xvernier division coinciding with main scale division
0.84 =0.02Xvernier division coinciding with main scale division
42
nd
division of vernier scale exactly coincides with main scale division
12.84=12+ (0.02X42) =12.84mm


Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 7
Procedure:
1. The Least Count is to be determined.
L.C = (Minimum Main Scale Reading) / (No. of Vernier Scale Divisions)
2. The work piece is placed between the jaws of Vernier Calipers correctly.
3. The reading on Main scale which is just behind the first Vernier Scale Division is noted as
Main Scale Reading.
4. The Division on Vernier Scale which coincides with the line on Main Scale is noted down
as Vernier Coincidence.
5. The Diameter can be calculated using the given Formula.
Precautions:
1. Make sure the Vernier Calipers are clean.
2. Clean the measuring faces with paper or cloth.
3. Make sure the work piece axis is perpendicular to the Vernier Calipers.
Calculations:
Length of the specimen:
S No

Main Scale
Reading
VSR VSR X LC TR= MSR + (VSR
X LC) mm
1


2


3


4


5


Diameter of the specimen:
S No

Main Scale
Reading
VSR VSR X LC TR= MSR + (VSR
X LC) mm
1


2


3


4


5


Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 8
Thickness of the specimen:
S No

Main Scale
Reading
VSR VSR X LC TR= MSR + (VSR
X LC) mm
1
2
3
4
5

Result:
The experiment is conducted on the vernier callipers for measuring for the physical quantities
of the given specimen.
1. Length of the specimen - --------------------------- mm
2. Diameter of specimen- ------------------------------ mm
3. Inner diameter of the hollow cylinder- ----------- mm
4. Outer diameter of the specimen- - ---------- ------mm
5. Thickness of the specimen- ------------------------ mm
B) MICROMETER:

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 9
Procedure:
1. The micrometers is checked for zero error.
2. The given component is held between the faces of the anvil and spindle.
3. The spindle is moved by rotating the thimble until the anvil and spindle touches the
cylindrical surface of the component.
4. Fine adjustment is made by ratchet .the main scale reading and thimble scale reading are
noted.
5. Two are more reading are taken at different places of the component.
6. The readings are tabulated and calculated.
TABULATION:
Sl. No Main scale
reading(MSR)
mm
Vernier scale
Division
(VSD)
Vernier scale
reading(VSR) =VSD X
Least count in mm
Total reading
(M.S.R+V.S.R)
In mm






RESULT:
The diameter of the given cylindrical component is determined to an accuracy of 0.01mm.
The result is checked with digital micrometre.
The diameter of the given cylindrical component is____________mm.
Viva questions:
(1) Define (a) Metrology (b) Least count (c) Engineering Metrology (d) Measurand.
(2) What is the scope of metrology in industries?
(3) State different precision linear measuring instruments.
(4) Which linear measurements can be measured by linear measuring instruments?
(5) What is the "Vernier principle"? Explain it with suitable example.
(6) Draw neat sketch of vernier caliper.
(7) List out constructional parts of vernier caliper.
(8) Is it possible to set the dimension 15.73 mm on Vernier Caliper having of least count
0.02mm? Why?
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 10
(9) Is it possible to set the dimension of 15.72 mm on vernier caliper having least count of
0.02mm? Why? If yes then show the dimension of 15.72mm on vernier caliper of least count
0.02mm.
(10) Show the following readings on vernier caliper of least count of 0.02mm least count: (a)
6. 84mm (b) 10.28mm.
(11) Differentiate between A type, B type and C type of vernier calipers.
(12) List out two applications of vernier caliper.
(13) State whether the following statements are true or false. Correct the false statements:
(a) Vernier Caliper has a provision of ratchet for ensuring correct measuring pressure.
(b) Step height can be measured by vernier caliper.
(c) B type of vernier caliper is used for marking purpose.
(d) Vernier Caliper obeys Abbes Principle of Alignment.
(e) Vernier Caliper is an example of Line standard.


















Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 11






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 12






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 13
Experiment No. 2 Date: / / 20
MEASUREMENT OF BORES BY INTERNAL MICROMETERS AND
DIAL BORE INDICATORS.
Aim:-
To determine inside diameter and bore diameter is a given hollow specimen
Apparatus:-
Inside micro meter, hollow specimen
Micro meter:-
It is one of the most common and most popular form of measuring instrument for precious
measurement with 0.001mm accuracy are also available.
Principle:-
Micro meter works on the principle of screw and nut. When screw is turned through nut one
revolutions it advances by one pitch distance i.e., one revolution of screw corresponds to a
linear moment of a distance equal to the pitch of the thread
L.C= Pitch of the spindle/ No of divisions on the spindle
Procedure:-
1. Select the micro meter with a desired range depending upon the size of the work piece to be
measured.
2. The next step is to check it for zero error. In case of 0.25mm micrometre, the zero error is
checked by contracting the faces of fixed anvil and the spindle.
3. The barrel has graduation, in travels of 1mm above the reference line
4. For measuring the dimension, hold work b/w faces of the anvil the spindle by rotating then
touches the work piece
5. Take the thimble reading with coincides with the reference line on the sleeve.
Total reading = MSR + (PSR X LC) mm
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 14



1. CD Left jaw (2) Right jaw (3) Contact point (4) Clamping knob (5) Sleeve (6) Thimble
(7) Ratchet stop
Sleeve 22.5mm
Thimble 37mm
Reading 22.87mm
Precautions:-
1. First clean the micro meter by wiping off dirt, fit, dust grit off it.
2. Clean them with a piece of cloth or paper
3. Set zero readings on instrument before measuring.



Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 15
Inner diameter of the spicemen-1
S No

Main Scale
Reading
(mm)
VSR
(mm)
VSR X LC TR= MSR + (VSR
X LC) mm
1
2
3
4
5

Inner diameter of the spicemen-2
S No

Main Scale
Reading(mm)
VSR(mm) VSR X LC TR= MSR + (VSR
X LC) mm
1
2
3
4
5

Theory:-
Bore gauge, is generally used to determine the bore diameter of components. Bore gauge
consists of following parts.
1. Dial gauge
2. Vertical column
3. Arrangement of anvil and washer
4. Movable spindle
B) DIAL BORE INDICATIORTHEORY AND DESCRIPTION:
Dial bore indicator consists of measuring head and guide is attached with extension rod
&collars for specific dimension chosen from the table in the instrument box, holder is
assembled to the measuring head and dial indicator is fixed inside the holder during tightening.
The condition is initially 1 kgf is applied to the dial indicator for getting exact reading.
PRINCIPLE: Dial bore indicator is works on comparator principle.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 16
PROCEDURE:
1) Once approximate bore is finding out by using inside micro meter.
2) Chose the same little more size extension rod & collar if necessary select and fit.
3) Keep the dial bore indicator into the specimen bore.
4) Repeat same procedure to get the bore diameter at different positions of specimen
Least count = 0.01mm



Sample calculations:-
Least count (LC) =0.01mm
Anvil size = 45mm
Washer size = 45mm
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 17
Total indicator Reading = 14 X 0.01= 0.145mm
Total Reading= (Anvil size + Washer size)-(Dial indicator for Reading)
= (45+4.5)-(0.145) =49.355mm
Calculation Total reading:-
Bore diameter = (Anvil size + Washer Size) - (Dial indicator Reading).
Inner Diameter of the specimen-1:
S No

Anvil
Size
Washer size Dial indicator
reading
TR= Anvil size+
Washer size Dial indicator
reading(mm)
1
2
3
4
5

Inner Diameter of the specimen-2:
S No

Anvil
Size
Washer
size
Dial indicator
reading
TR= Anvil size+
Washer size Dial indicator
reading(mm)
1
2
3
4
5

Result:-
The experiment is used to find the inner diameter/bore diameter of the hollow specimen of
given specimen
The inner diameter of the hallow specimen is ---------------- mm
The bore diameter of the given specimen is .mm

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 18






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 19
Experiment No. 3 Date: / / 20
LINEAR MEASUREMENT USING DIAL GAUGE, SLIP GAUGE, AND
CALIBRATION OF DIALGAUGE
Aim: - Linear Measurement Using Dial Gauge, Slip and Calibration of Dial Gauge.
Apparatus: - Dial Gauge, Slip Gauge, Dial Gauge Indicator
Theory:-
The different component of dial gauge indicator is shown in fig. It consist of plunger,
removable contact pt., stem a transparent glass cover, calibrated dial pointer, bezel camp or
bezel locking nut. Revolution counter in order to counter in order to count the no of revolution
of a pointer, dust proof cap etc.
It consists of a plunger which slides in bearing and carries a rack with it. The rack is meshed
with pinion (P1) again pinion (P2) and gear (G2) are on same spindle [because of which
magnification is taking place]. The gear (G1) is meshed with (P1) again pinion (P2) and gear
(G2) are on spindle basically gear (G2) is meshed with pinion (P3) on whose spindle pointer is
attached. The pinion (P3) is meshed with gear (G3) on which a light is here spring is attached
in order to guide the movement of plunger rack guide is provided and to bring the plunger to
its initial position a light coil spring is attached to plunger.
Any linear displacement given causes rack to move upward during this upward movement as
rack is meshed with pinion (P1) and gear (G1) rotate by some amount but as the no of teeth on
gear (G1) is more compared to that of pinion (P2) which is meshed with it rotate more time.
Let us say if there is 100 teeth on gear G1 and 10teeth on pinion P2 the 1st stage of
magnification is 100/10=10 times again.
Therefore overall magnification can be calculated,
(G1 X G2)/ (P2 X P3)
EG: (100x100/10x10) = 100
In this way dial indicator works you can take diff readings by keeping standard and object.
Then comparison can be made.
The magnification is about 250-1000.
Dial gauge -
Dial gauges divided in two categories, type1 &type2for general engineering purpose depending
upon the movement of the plunger. These are manufactured in two grades, grade a and grade
b, with total plunger movement or lift of 3,5 and 10mm. Type1 dial gauge has the plunger
movement parallel tip the plane of dial and type 2 has the plunger movement perpendicular to
the plane of dial.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 20
Indicator gauge-
Dial indicator has been used with several auxiliary devices for a wide variety of length
measurement. Obviously dial indicator can be used for carrying the needed complimentary
function, resulting in a single tool, it is known as indicator gauge. It must be remembered that
indicator gauge are always comparator type measuring instrument and require the use of a
setting gauges for establishing the basic measuring position
Slip gauge:
Slip gauges with three basic forms are commonly found. These are rectangular square with
center hole, and square without center hole. Rectangular forms is the more widely used because
rectangular block are less expensive to manufacture, and adopt themselves better to application
where space is restricted or excess weight is to be provided. For certain application squarely
gauges, through expensive, are preferred. Due to their large surface area, they wear longer and
adhere better to each other when touch to high stack.



Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 21


PROCEDURE FOR SLIP GAUGES:
1. The slip gauges are cleaned by using cloth
2. The thickness of the given MS plate is determined to the nearest 0.1mm size by using micro
meter.
3. The slip gauges are selected to built up required dimension.
4. Required combination of slip gauges is built up by wringing.
5. The built up gauges is wrung with the wringing faces of the measuring jaws.
6. The jaws and the built up gauges are held in the holder.
7. The given MS plate is placed between the flat surfaces of the measuring jaws and the
thickness of the plate is measured.
OBSERVATION:
Slip gauge
For linear measurement least count = (0.2/200)=0.001 mm/div
For standard dimension
For specimen
Main scale reading =MSR = -------
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 22
Circular scale reading =CSR= ---------
Total reading (TR) = MSR + (CSR * LC)
Total Height of specimen = -----------
RESULT:
The thickness of the given MS plate is measured by using slip gauges.
The thickness of the given MS plate is _______________ mm.
PROCEDURE FOR DIAL GUAGE:
1. The slip gauges are built up to the given weight of the component.
2. Dial gauge with stand is placed on the surface plate.
3. The built up gauge is placed under the plunger.
4. The indicator is set to zero.
5. The built up gauge is removed.
6. The given machined component is placed under the plunger.
7. The variation in the height of the component is noted from the reading of the dial.
TABULATION: component height =____ mm.
Sl.
No

Dial reading on built up slip
Gauges in div.

Dial reading on
Component in div.

Variation of height
In mm





Calculation of Dial Gauge:-
Slip Gauge Reading Dial Gauge Reading
= 50+ = 0.2 + .
= 51.001 mm = 0.2 +
= 51.003 mm = 0.2 + .
Difference for the both should be equal
RESULT:
The height of the machined component is checked with standard dimensioned Component
(slip gauges) using dial gauge.
The variation in height is ________________ mm.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 23






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 24






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 25






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 26
Experiment No. 4 Date: / / 20
USE OF GEAR TEETH, VERNIER CALIPERS AND CHECKING THE
CHORDAL ADDENDUM AND CHORDAL HEIGHT OF SPUR GEAR.
Aim: To measure the tooth thickness of a given spur gear
Instruments Required: Gear vernier, Vernier caliper, Spur gear
Theory:
The tooth thickness is defined as the length of the arc of the pitch circle between opposite faces
of the same tooth. Most of the time a gear vernier is used to determine the tooth thickness. As
the tooth thickness varies from top to bottom, any instrument for measuring on a single
tooth.Gear tooth micro meter is used to measure the thickness of gear tooth at pitch line. It is
similar to simple micro meter but gear tooth micro meter having flanks at the end of anvil and
spindle. The flanks of the micro meter. Gives the thickness of gear tooth at pitch line.
Principle:-
Gear tooth micro meter works on the principle of screw and when screw is turned throughput
for one revolution it advances by one pitch distance i.e., one revolution of screw corresponds
to a linear moment of a distance equal to the pitch of thread.
Least Count (LC) = Pitch of the spindle screw/ No of divisions of the spindle (mm)

Terminology for Spur Gears:
Pitch surface: The surface of the imaginary rolling cylinder (cone, etc.) that the toothed gear
may be considered to replace.
Pitch circle: A right section of the pitch surface.
Addendum circle: A circle bounding the ends of the teeth, in a right section of the gear.
Root (or dedendum) circle: The circle bounding the spaces between the teeth, in a
right section of the gear.
Addendum: The radial distance between the pitch circle and the addendum circle.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 27
Dedendum: The radial distance between the pitch circle and the root circle.
Clearance: The difference between the dedendum of one gear and the addendum of the
mating gear.
Face of a tooth: That part of the tooth surface lying outside the pitch surface.
Flank of a tooth: The part of the tooth surface lying inside the pitch surface.
Circular thickness (also called the tooth thickness): The thickness of the tooth
measured on the pitch circle. It is the length of an arc and not the length of a straight
line.
Tooth space: The distance between adjacent teeth measured on the pitch circle.
Backlash: The difference between the circle thickness of one gear and the tooth space
of the mating gear.
Circular pitch p: The width of a tooth and a space, measured on the pitch circle.
Diametral pitch P: The number of teeth of a gear per inch of its pitch diameter. A
toothed gear must have an integral number of teeth. The circular pitch, therefore, equals
the pitch circumference divided by the number of teeth. The diametral pitch is, by
definition, the number of teeth divided by the pitch diameter. That is,

and

Hence

p = circular pitch
P = diametral pitch
N = number of teeth
D = pitch diameter
That is, the product of the diametral pitch and the circular pitch equals .
Module m: Pitch diameter divided by number of teeth. The pitch diameter is usually
specified in inches or millimeters; in the former case the module is the inverse of
diametral pitch.
Fillet: The small radius that connects the profile of a tooth to the root circle.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 28
Pinion: The smallest of any pair of mating gears. The largest of the pair is called simply
the gear.
Velocity ratio: The ratio of the number of revolutions of the driving (or input) gear to
the number of revolutions of the driven (or output) gear, in a unit of time.
Pitch point: The point of tangency of the pitch circles of a pair of mating gears.
Common tangent: The line tangent to the pitch circle at the pitch point.
Line of action: A line normal to a pair of mating tooth profiles at their point of contact.
Path of contact: The path traced by the contact point of a pair of tooth profiles.
Pressure angle : The angle between the common normal at the point of tooth contact
and the common tangent to the pitch circles. It is also the angle between the line of
action and the common tangent.
Base circle: An imaginary circle used in involute gearing to generate the involutes that
form the tooth profiles.

It should be noted that M is a chord AC, but the tooth thickness is specified as an arc
distance ADC. Also h is the distance EB and this is slightly greater than the addendum ED.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 29

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 30

FORMULA USED:
Depth= (Zm/2) (1+2/Z-COS (90/Z)
Width=Zm x sin (90/Z)
Outer diameter of gear = (Z+2) m
Where,
Z-no of gear tooth, m-module
PROCEDURE:
1. Find the zero error in the horizontal scale and vertical scale of the given gear tooth vernier.
2. Find outer diameter of the given gear by using vernier caliper.
3. Count the no of tooth on the given gear.
4. Calculate the depth of pitch circle from the top circle.
5. Calculate the module (m) of the gear.
6. Similarly calculate the theoretical width by substituting and no of gear tooth in the formula.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 31
7. The vertical gear tooth vernier is made of point the calculate the depth value.
8. Now the gear tooth, i.e. kept in between in the two jaws of the gear tooth vernier.
9. Observe the main scale reading and vernier scale coincidence of the horizontal scale.
10. Repeat the observation of different position of the same tooth and calculate the average.
Least count: Horizontal scale=0.02mm
Vertical scale =0.02mm

MODEL CALCULATION:
Module = outer dia/ (Z+2)
Depth = (Zm/2) (1+2/Z-COS (90/Z))
Width = Zm x sin (90/Z)
Deviation =theoretical value-actual value
RESULT:
Thus the thickness of the gear tooth of the given spur gear is calculated using gear tooth vernier.
Depth of the gear tooth = mm
Width of the gear tooth = mm
Theoretical value = .mm
Actual value = mm



Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 32






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 33






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 34






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 35
Experiment No: 5 Date: / / 20
ANGLE AND TAPER MEASUREMENTS BY BEVEL PROTRACTOR, SINE BARS
Aim: - To measure the taper angle of the given specimen using bevel protractor and sine bar
method.
APPARATUS REQUIRED:
1. Sine bar 2. Micrometer3. Slip gauge set 4. Surface plate5. Dial gauge withstand
6. Vernier caliper7.Combination Sets 8. Bevel Protector
Theory:
BEVEL PROTECTOR
A universal bevel protractor is used to measure angles between two planes. This consists of
stem, which is rigidly attached to main scale and a blade, which is attached to the Vernier
scale and can be rotated to read angles. To improve the accessibility, the blade can also slide.
The least count is calculated by knowing the value of the smallest division on the main scale
and number of division on the Vernier scale. It should be noted that the divisions on the main
scale is in degrees and that the fractional divisions of degrees are minutes (i.e. with 60
minutes/degree, denoted). To measure angle between two planes, rest the stem on one of the
planes (reference plane). Rotate the blade such that blade is flush with second plane.
Readings are taken after ensuring that the stem and blade are in flush with the two planes.
Lock the protractor at this point and note sown the readings.

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 36

OBSERVATIONS:
S.NO. ANGLE MEASURED





PRECAUTIONS:
1. The sine bar should not be used for angle greater than 600 because any possible error in
construction is accentuated at this limit.
2. A compound angle should not be formed by miss-aligning of work piece with the sine bar.
This can be avoided by attaching the sine bar and work against an angle plate.
3. As far as possible longer sine bar should be used since using longer sine bars reduces many
errors.
RESULT:
The angle of the given specimen measured with the Bevel Protractor is..
SINE BAR
The sine principle uses the ratio of the length of two sides of a right triangle in deriving a
given angle. It may be noted that devices operating on sine principal are capable of self-
generation. The measurement is usually limited to 45 degree from loss of accuracy point of
view. The accuracy with which the sine principle can be put to use is dependent in practice,
on some from linear measurement. The sine bar itself is not complete measuring instrument.
Another datum such as surface plate is needed, as well as other auxiliary instrument, notably
slip gauge, and indicating device to make measurements.
Sine bar:

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 37
A sine bar is a tool used to measure angles in metalworking.

FIG: SINE BAR
It consists of a hardened, precision ground body with two precision ground cylinders fixed at
each end. The distance between the centers of the cylinders is precisely controlled, and the
top of the bar is parallel to a line through the centers of the two rollers. The dimension
between the two rollers is chosen to be a whole number (for ease of later calculations) and
forms the hypotenuse of a triangle when in use. The image shows a 10 inch and a 100 mm
sine bar.
When a sine bar is placed on a level surface the top edge will be parallel to that surface. If
one roller is raised by a known distance then the top edge of the bar will be tilted by the same
amount forming an angle that may be calculated by the application of the sine rule.
The hypotenuse is a constant dimension (100 mm or 10 inches in the examples
shown).
The height is obtained from the dimension between the bottom of one roller and the
table's surface.
The angle is calculated by using the sine rule.

Angles may be measured or set with this tool. For precision measurements where the bar
must be set at an angle, gauge blocks are traditionally used.
The sine bar is set up on a surface plate to the nominal angle of the taper plug, which is then
placed in position on the bar, being prevented from sliding down by the stop plate at the end.
Care must be taken to ensure that the axis of the plug gauge is aligned with the sine bar.
Pieces of plasticine will be found to be useful for preventing sideways movement. The dial
gauge, supported in a stand on the surface plate, is then passed over the plug gauge near each
end and also at one or two positions between the ends. If there is any variation in the
readings, two alternatives are available for finding the true angle of the cone. Either the
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 38
variation over a measured distance along the surface of the plug gauge can be used to obtain
the difference between the true angles or the angle set up, as the height of the slip gauge pile
can be adjusted until no variation occurs in the reading of the dial gauge.
Checking of Unknown Angles: - Many a times, angle of component to be checked is
unknown. In such a case it is necessary to first find the angle approximately with the help of a
bevel protractor. Let the angle. Then the sine bar is set at an angle () and clamped to an angle
plate. Next the work is placed on sine bar and clamped to Angle plate as shown in figure. Slip
gauges are so arranged (according to deviation) that the sprit level is at center (the air
bubble)
If the deviation is noted down by the spirit level is h over a length l of work ,then height of
slip gauges by which it should be adjusted is equal to = h 1
FORMULA:
Sin = h / L
Where,
H - Height of the slip gauge
L - Distance between the centres
- Inclined angle of the specimen
PROCEDURE:
The given component is placed on the surface plate.
One roller of sine bar is placed on surface plate and bottom surface of sine bar is seated on
the taper surface of the component.
The combination of slip gauges is inserted between the second rollers of sine bar and the
surface plate.
The angle of the component is then calculated by the formula given above.
S.No Length of the sine
bar (L) mm
Height
(h)mm
Taper angle ()






Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 39
CALCULATION:
Sin = h / L
Precaution in Sine Bars:-
(a) A Compound angle should not be formed by miss dignity of w/p with the sine bar. This
can be avoided by attaching the sine brand work against an angle plate.
(b) Accuracy of sine bar should be ensured.
(c) As far as possible longer sine bar should be used since4 many errors are reduced by using
longer sine bar.
Precautions:-
1. Angle of instrument must coincide with the angular scale
2. Gripped the instrument to the measuring face exactly
Result:-
Thus the taper angle of the given specimen is measured using sine bar.
The external taper angle is..
VIVA QUESTIONS
1. What is the use of angle plates?
2. Name some angle measuring devices?
3. What is the least count of mechanical Bevel Protractor?
4. What is the least count of optical Bevel Protractor?
5. What is a sine bar?
6. What are the limitations of Sine bar?
7. What is the difference between the sine bar and sine center?
8. What is the use of V-block?
9. What is the purpose of adjusting nuts in a micro meter?
10. What is the least count of dial indicator?
11. How do you specify sine bar?
12. How to maintain constant pressure in micro meter?
13. What are the applications of Gear toothvernier caliper?
14. How do we check the profile of a Gear tooth?
15. Name some angle measuring devices?
16. Why do we use Feeler gauges?
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 40
17. What are slip gauges and why do we use them?
18. What are slip gauges and why do we use them?
19. Explain zero error and zero correction in case of micrometers?
20. What is the principle involved in sprit levels?
21. What is the least count of digital vernier caliper?
22. What is the difference between vernier height gauge, vernier depth gauge, and vernier
caliper?
23. Explain briefly about the different types of micrometers?























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 41






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 42






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 43
Experiment No: 6 Date: / / 20
MACHINE TOOL ALIGNMENT TEST ON THE LATHE.
Aim:-
1. Test for level of installation:
(a) In a longitudinal b) In transverse direction
Measuring instruments: Spirit level, gauge block to suit the guide ways of the lathe bed.
Theory:-
The following are the alignment tests on lathe.
Levelling of machine:-
It is essential that a machine tool should be installed truly horizontal and vertical plane and this
accuracy must be maintained. The level of machine base in longitudinal and transverse
direction is tested by spirit level or precision level. The spirit level is placed at to measure the
level.
True running of main spindle:-
The true mandrel is placed in the main spindle and test is conducted on the surface of material
if dial gauge shows any deviation in the reading then it is said that the main spindle is running
in the proper way.
Parallelism of main spindle to saddle movement:-
If the axis of the spindle is not parallel to the saddle movement then it is not possible to get
required dimension of work piece while doing the operation on lathe. The spindle is moved
and the deviation in the reading of dial gauge are noted.
Parallelism of Tailstock guide ways to saddle movement:-
To check the parallelism of guide ways with the saddle movement in the both vertical and
horizontal directions. The dial indicator is held on the spindle and block is moved
simultaneously any deviation in reading of dial gauge is noted if no deviation in the reading
then tail stock guide ways is parallel to saddle movement otherwise it is not parallel to saddle
movement.
Parallelism of tail stock guide ways to carriage movement:-
To check the parallelism of guide ways with the carriage in both vertical and horizontal
objections. A block is placed on the guide ways of tail stock. The dial indicator is held on the
carriage and block is moved simultaneously any deviation in reading of dial gauge is noted
Parallelism of main spindle to carriage movement:-
To check the parallelism of main spindle to carriage in both vertical and horizontal. The
deviation is observed the spindle is not parallel to the carriage.

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 44
True Running of head stock centre:-
The test mandrel is placed in the head stock and test is conducted on the surface of carriage.
The dial gauge shows any deviations in the reading then the head stock is not running in proper.
Procedure: - The gauge block with the spirit level is placed on the bed ways on the front
position, back position and in the cross wise direction. The position of the bubble in the spirit
level is checked and the readings are taken.
1. Permissible error: Front guide ways. 0.02 mm/meter convex only. Rear guide ways, 0.01
to0.02 convexity. Bed level in cross-wise direction 0.02/meters. Straightness of slide ways(for
machines more than 3 mm turning length only, measurement taken by measuring tight wire
and microscope or long straight edge). Tailstock guide ways parallel with movement of
carriage 0.02 mm/m. No twist is permitted.

The error in level may be corrected by setting wedges at suitable points under the support feel
or pads of the machine.
2. Straightness of saddle in horizontal plane:-
Measuring instruments: Cylindrical test mandrel (600mm long), dial indicator.
Procedure: - The mandrel is held between centres. The dial indicator is mounted on the saddle.
The spindle of the dial indicator is allowed to touch the mandrel. The saddle is then moved
longitudinally along the length of the mandrel. Readings are taken at different places.
Permissible error: 0.02 mm over length of mandrel.

3. Alignment of both the centres in the vertical plane:
Measuring instruments: Cylindrical mandrel 600 mm long, dial gauge.
Procedure: The test mandrel is held between centres. The dial indicator is mounted on the
saddle in vertical plane as shown in figure. Then the saddle along with the dial gauge is
travelled longitudinally along the bed ways, over the entire length of the mandrel and the
readings are taken at different places.
Permissible error: 0.02 mm over 600 mm length of mandrel (Tail stock centre is to lie higher
only).
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 45

4. True running of taper socket in main spindle
Instruments required: Test mandrel with taper shank and 300 mm long cylindrical measuring
part, dial gauge.
Procedure: The test mandrel is held with its taper shank in a head stock spindle socket. The
dial gauge is mounted on the saddle. The dial gauge spindle is made to touch with the mandrel.
The saddle is then travelled longitudinally along the bed ways and readings are taken at the
points A and B as shown in figure.
Permissible error: Position A 0.01 mm, position B 0.02 mm.

5. Parallelism of main spindle to movement:
(a) In a vertical plane (b) In a horizontal plane
Measuring instruments: Test mandrel with taper shank and 300 mm long cylindrical
measuring part, dial gauge.
Procedure: The dial gauge is mounted on the saddle. The dial gauge spindle is made to touch
the mandrel and the saddle is moved to and fro. It is checked in vertical as well as in horizontal
plane.
Permissible error: (a) 0.02/300 mm mandrel rising towards free end only.
(b) 0.02/300 mm mandrel inclined at free end towards tool pressure only.

6. Movement of upper slide parallel with main spindle in vertical plane:
Measuring instruments: Test mandrel with taper shank and 300 mm long cylindrical
measuring part, dial gauge.
Procedure: The test mandrel is fitted into the spindle and a dial gauge clamped to the upper
slide. The slide is transversed along with the dial gauge plunger on the top of the stationary
mandrel.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 46
Permissible error: 0.02 mm over the total movement of the slide.

7. True running of locating cylinder of main spindle:
Measuring instrument: Dial gauge.
Procedure: The dial gauge is mounted on the bed, touching at a point on main spindle. The
main spindle is rotated by hand and readings of dial gauge are taken.
Permissible error: 0.01 mm.

8. True running of head stock centre:
Measuring instruments: Dial gauge.
Procedure: The live centre is held in the tail stock spindle and it is rotated. Its trueness is
checked by means of a dial gauge.
Permissible error: 0.01 mm.

9. Parallelism of tailstock sleeve to saddle movement:
Measuring instruments: Dial indicator
Procedure: Tailstock sleeve is fed towards. The dial gauge is mounted on the saddle. Its
spindle is touched to the sleeve at one end and the saddle is moved to and fro, it is checked in
H.P. and V.P. also.

Permissible error: (a) 0.0 1/100 mm (Tailstock sleeve inclined towards tool pressure only).
(b) 0.0 1/100 mm (Tailstock sleeve rising towards free end only).
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 47

10. Paralle1ism of tail stock sleeve taper socket to saddle movement
(a) In V.P.
(b) In H.P.
Measuring instruments: - The mandrel with taper shank and a cylindrical measuring part of
300mm length, dial gauge.
Procedure: - Test mandrel is held with its taper shank in tail-stock sleeve taper socket. The
dial gauge is mounted on spindle. The dial gauge spindle is made to touch with the mandrel.
The saddle is then transverse longitudinally along the bed way and readings are taken.
Permissible error:-
(a) 0.03/300 mm (mandrel rising towards free end only).
(b) 0.03/300 mm (Mandrel inclined towards tool pressure only).
PRECAUTIONS:
i) The mandrel must be so proportioned that its overhang does not produce appreciable sag,
else the sag must be calculated and accounted for.
ii) The indicator set up must be rigid, otherwise variations in readings as recorded by point may
be solely due to deflection of the indicator.
REVIEW QUESTIONS
a) What is the necessity of conducting various alignment tests on lathe?
b) What are the various alignment tests to be conducted on the lathe?
c) What is straightness?
d) What is flatness?
e) What is square ness?
f) What is parallelism?
g) What do you mean by axial slip of main spindle?
h) It is necessary to conduct alignment tests on other machine tools? If so why? Not, why not?




Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 48






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 49



























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 50







CYCLE-2 EXPERIMENTS
















Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 51
Experiment No: 7 Date: / / 20
MACHINE TOOL ALIGNMENT TEST ON MILLING MACHINE.
Aim:-
To perform the alignment test on milling.
Apparatus:-
Spirit level, gauge blocks, dial gauge
Theory:-
Following are the tests on milling machine
Test for levelling of milling machine:-
It is essential that a machine tool should be installed truly horizontal and vertical plane and this
accuracy must be maintained. If milling base is not installed truly horizontal then bed will
undergo a deflection and produce a simple bend.
True Running of spindle:-
A mandrel placed in the spindle and test is conducted on the surface of mandrel. A dial gauge
is fixed on the machine table and feeler of the dial gauge is made to touch the lower surface of
it clearance is noted then it is said that the table is not flat otherwise it is flat.
True Running of spindle:-
For this test the mandrel is placed in the spindle and dial indicator is fixed on the table. The
feeler of dial gauge is made to touch the surface of manderal.
Parallelism of spindle Axis with its vertical moment:-
For this test the manderal is placed in the spindle and dial indicator is fixed on the table. The
feeler of dial gauge is made to touch the surface of mandrel also moved up and down, the
mandrel also moved up and down observe any direction in the reading of dial gauge is noted
then that is said that it is not running in proper way mandrel. Axis slip of main spindle is
developed due to the error in bearing support for this test feeler of the dial gauge is placed on
the face of main spindle and the dial gauge.
Parallelism (or) Table Surface with longitudinal surface:-
A machine is placed in the spindle and test is conducted on the surface of mandrel. If any
degration is noted then it is noted then it is said that spindle is not parallel to the table.
Parallelism of Table Surface with main spindle:-
A mandrel is placed in the spindle and test is conducted on the surface of mandrel. A dial gauge
is fixed on the table and feeler is touched to the spindle. If any deviation takes place the spindle
is not machine to the table.


Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 52
Parallelism of Table Surface with Arbor:-
Arbor is placed in the spindle and test is conducted on the surface of order. If any degration is
noted than it is said that arbor is not parallel to the table.
Procedure:
(1) Flatness of work table
(a) In longitudinal direction.
(b) In transverse direction.
Measuring instruments: - Spirit level.
Procedure: - A spirit level is placed directly on the table at points about 25 to 30 cm apart, at
A, B, C for longitudinal tests and D, E and F for the transverse test. The readings are noted.
Permissible error:
Direction A-B-C, 0.04 mm
Direction D-E-F, 0.04 mm

(2) Parallelism of the work table surface to the main spindle
Measuring instruments: Dial indicator, test mandrel 300 mm long, spirit level.
Procedure: The table is adjusted in the horizontal plane by spirit level and is then set in its
mean position longitudinally. The mandrel is fixed in the spindle taper. The dial gauge is set
on the machine table, and the feeder adjusted to touch the lower surface of the mandrel. The
dial gauge readings at (A) and (B) are observed, the stand of the dial gauge being moved while
the machine table remains stationary.
Permissible error: 0.02/3 00 mm.

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 53
(3) Parallelism of the clamping surface of the work table in its longitudinal motion:
Instruments: Dial gauge, straight edge.
Procedure: A dial gauge is fixed to the spindle. The dial gauge spindle is adjusted to touch the
table surface. The table is then moved in longitudinal direction and readings are noted. If the
table surface is uneven it is necessary to place a straight edge on its surface and the dial gauge
feeder is made to rest on the top surface of the straight edge.
Permissible error: 0.02 up to 500 mm length of transverse, 0.03 up to 1000 mm and 0.04
above1000 mm length of transverse.

(4) Parallelism of the cross (transverse) movement of the worktable to the main spindle
:( a) In vertical plane
(b) In horizontal plane
Instruments: Dial gauge, test mandrel with taper shank.
Procedure: The work table is set in its mean position. The mandrel is held in the spindle. A
dial gauge fixed to the table is adjusted so that its spindle touches the surface of the mandrel.
The table is moved cross-wise and the error is measured in the vertical plane and also in the
horizontal plane.
Permissible error: 0.02 for the overall transverse movement of the work table.


(5) True running of internal taper of the spindle:
Instruments: 300 mm long test mandrel, dial gauge.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 54
Procedure: The test mandrel with its taper shank is held in the main spindle. Dial gauge is
kept scanning the periphery of the mandrel. Spindle is rotated and dial gauge readings are noted
at different points say A and B as shown.
Permissible error: Position A: 0.01 mm, Position B: 0.02 mm.

(6) Square nests of the centre T-slot of worktable with main spindle
Instruments: Dial gauge, special bracket.
Procedure: To check the perpendicularity of the locating slot and the axis of the main spindle.
The table should be arranged in the middle position of its longitudinal movement, and a bracket
with a tenon at least 150 mm long inserted in the locating slot as shown in figure. A dial gauge
should be fixed in the taper, the feeder being adjusted to touch the vertical face of the bracket.
Observe the reading on the dial gauge when the bracket is near one end of the table, the swing
over the dial gauge and move the bracket so that the corresponding readings can be taken near
the other end of the table.
Permissible error: 0.025 mm in 300 mm.

(7) Parallelism of the T-slot with the longitudinal movement of the table:
Instruments: Dial gauge, special bracket.
Procedure: The general parallelism of the T-slot with the longitudinal movement of the table
is checked by using 150 mm long braked having a tenon which enters the slot, the dial gauge
is fixed to the spindle taper and adjusted so that its feeder touches the upper surface of the
bracket. The table is then moved longitudinally while the bracket is held stationary by the hand
of the operator and dial gauge deviations from parallelism are note down.
Permissible error: 0.0 125 mm in 300 mm.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 55

(8) Parallelism between the main spindle and guiding surface of the overhanging arm
Instruments: Dial gauge, mandrel.
Procedure: The overhanging arm is clamped in its extreme extended position. The dial
gauge is fixed to the arbor support. The feeder of the dial gauge is adjusted to touch the
top or ride of the test mandrel. The arbor can then be moved along the overhanging arm
and the deviations from parallelism observed on the dial gauge.



PRECAUTIONS:
i) All moving parts of the machine must be locked while reading the dial gauge,
ii) If the table surface is uneven, straight edges must be used.
REVIEW QUESTIONS:
i) Distinguish between geometric tests and practical tests.
ii) How will you measure the flatness of the table surface?
iii) What are the various alignment tests conducted on vertical milling machine?
iv) What are the various measuring instruments used in alignment test of a milling
machine
v) What are the dimensions of a test piece used in practical test?





Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 56






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 57






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 58






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 59
Experiment No: 8 Date: / / 20
TOOL MAKERS MICROSCOPE AND ITS APPLICATION.
Aim:-
Study of Tool Makers Microsocope.
Objectives:
After performing this experiment, you should be able to
appreciate the importance of precision measurement,
know how precise measurements can be taken with this instrument,
explain the field of application/working of this instrument, and
understand the principle of working of tool room microscope.
Introduction:
Engineering microscopes designed to satisfy various measuring needs of toolmakers are known
as toolmakers microscopes. A plain toolmakers microscope is primarily intended for a
particular application. On the other hand, universal toolmakers microscope is adaptable to an
uncommonly wide range of measuring tasks. A toolmakers microscope is designed for
measurements of parts of complex forms, e.g. profile of external threads, tools, templates and
gauges. It can also be used for measuring centretocentre distance of holes in any planes, as
well as the coordinate of the outline of a complex template gauges.
Apparatus:-
Tool makers microscope, specimen

FIG: TOOL MAKERS MICROSCOPE
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 60
BRIEF DESCRIPTION OF INSTRUMENT:
It consists of optical head, which can be adjusted vertically along the ways of the vertical
column and can be clamped in any position. The working table is secured on a heavy hollow
base. The table has a compound slide to give longitudinal and lateral movements actuated by
accurate micrometre screws having thimble scales and vernier. At the back of the base is a light
source, which provides a horizontal beam of light reflected upwards by 90
o
towards the table.
This beam of light passes through a transparent glass plate on which flat parts to be checked
are placed. A shadow image of the outline of the contour passes the objective of the optical
head and is projected by a combination of three prisms to a ground glass screen. Observations
are made through the eyepiece of the optical head. Figure gives the views of a tool room
microscope. Cross lines are engraved on the glass screen, which can be rotated through 360
o
,
and these lines make the measurements. The angle of rotation of screen can be read on the
optical head. The eyepiece field of view contains an illuminated circular scale with a division
value of one minute. Adjusting optical head tube performs focussing.
Theory:-
The tool maker microscope is designed for measurement of components of difficult forms.
Ex: - profile of external threads, tools, gauge. It can be used for measuring center to center
distance of holes in any plane it consists of optical head which can be adjusted vertically along
inspection the table can be moved in longitudinal direction and lateral direction by micro meter
screws, which are having barrel and thimble at back of base light is arranged which provides
on the optical head. The image of component passes through optical head and observations.
The reading of longitudinal micro meter is noted. The difference gives the pitch of the thread.


FIG: TOOL MAKERS MICROSCOPE
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 61
Procedure:-
PROCEDURE:
1. Switch on the main.
2. Switch on the micros scope lights.
3. Select the capacity of the lens for precision operation.
4. Place the object on the class table to get the clear image rotate the wheel provided at the light
side.
5. After getting the clear image, locate the crosswire at the initial point on the image. Now note
down the micro meter reading.
6. Move the cross wire from initial point to the finial point on the image, which is to be
measured. Note down the micro meter reading, this operation is done by usingmicrometer.
7. Now the different but when the initial and the finial reading i.e. distance travelled gives the
size of the object.
8. Graph can be plotted actual micro meter reading vs. % of error.
Precautions:-
1) Obtain clean picture of cross line and the cross thread seen through the eyepiece.
2) For angular measurements lines must remain parallel to flank edge to the tooth.
RESULT:
Thus the all dimensions of the given particular screw were measured by using toolmakers
microscope.












Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 62






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 63






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 64
Experiment No: 9 Date: / / 20
USE OF SPIRIT LEVEL IN FINDING THE FLATNESS OF SURFACE
PLATE.
Aim:-
To check the flatness of given surface plate
Apparatus:-
Spirit level, surface plate
Theory:-
Generally spirit level is used for levelling the machinery and other instruments. But spirit levels
are also used to measure the angles. It is also called precision level. It consists of glass tube
and of the tube. If the tube is fitted through a small angle if R- radius of tube L distance of
bubble moved when spirit level is fitted to same angle
The simplest form of flatness testing is possible by comparing the surface with an accurate
surface. Spirit level is used in special cases and called Clinometers, precision micro-optic
clinometers utilizes bubble unit with a prismatic coincidence reader which presents both ends
of the bubble an adjacent images in a spirit field.Leveling helps in the coincidence of the 2
images, making it very easy to sec when the bubble is exactly centered without reference to
any graduations. The special features to precision micro-optic clinometers arc direct reading
over range 0-360, optically reading system, main coarse setting, slow motion screw to fine
setting. The least count of precision spirit level is 0.01 mm.
The spirit of level consists of a sealed glass tube mounted on a base. The inside surface of the
tube is ground to a convex barrel shape having large radius. The precision of the level depends
on the accuracy of this radius of the tube. A scale is engraved on the top of the glass tube. The
tube is nearly filled with either ether or alcohol, except a small air or vapour in the form of a
bubble.
The bubble always tries to remain at the highest point of the tube. If the base of the spirit level
is horizontal, the centre point is the highest point of the tube. So, that when the level is placed
on a horizontal surface, the bubble rests at the centre of the scale. If the base of the level is
fitted through a small angle, the bubble will more relative to the tube a distance along its radius
corresponding to the angle.


Fig: surface plate Fig: Spirit level
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 65

The figure shows two positions of the base of the level (OA1 and OA2) and corresponding
positions of the bubble (Bl, B2). When the base OA1 is horizontal, the bubble occupies
positionB1. Let be the small angle through which the base is fitted. The bubble now
occupies the position B2.Let L be the distance travelled by bubble along the tube and h the
difference in heights between the ends of the base. Then L= R and h =. L
Therefore =


Therefore =


Where R = radius of curvature of the tube
L = length of base
Finally =


Procedure:
1 Place the spirit level on the surface plate for which we have to find out the flatness
2 Find the base length of the spirit level
3 Note the radius of curvature of the spirit level tube
4 Find the tilt in the bubble
5 Finally find out the difference in heights between the ends of the base.
Flatness of the specimen:
S.NO Distance travelled
by the bubble
Difference in height
between
ends
Angle

1
2
3
4
5

Precautions:
1 .Clean the surface plate and ensure there is no dust particles
2. Take the bubble reading without any parallax error.
Result:-The experiment has been conducted on spirit level to check the flatness of given
surface plate. The given surface plate is flat/not flat---------------------


Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 66






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 67






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 68
Experiment No: 10 Date: / / 20
THREAD MEASUREMENT BY TWO/THREE WIRE METHOD OR
TOOL MAKERS MICROSCOPE.
Aim:-
To measure the effective diameter of the screw thread by using two/three wire method or Tool
makers microscope
Apparatus:-Tool Makers Microscope
Theory:-


1. Screw thread. A screw thread is the helical ridge produced by forming a continuous helical
groove of uniform section on the external or internal surface of a cylinder or cone. A screw
thread formed on a cylinder is known as straight or parallel screw thread, while the one formed
on a cone or frustum of a cone is known as tapered screw thread.
2. External thread. A thread formed on the outside of a work piece is called external thread
e.g., on bolts or studs etc.
3. Internal thread. A thread formed on the inside of a work piece is called internal thread e.g.
on a nut or female screw gauge.
4. Multiple-start screw thread. This is produced by forming two or more helicalgrooves,
equally spaced and similarly formed in an axial section on a cylinder. This gives a quick
traverse without sacrificing core strength.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 69
5. Axis of a thread. This is imaginary line running longitudinally through the centre of the
screw.
6. Hand (Right or left hand threads). Suppose a screw is held such that the observer is
looking along the axis. If a point moves along the thread in clockwise direction and thus moves
away from the observer, the thread is right hand; and if it moves towards the observer, the
thread is left hand.
7. Form, of thread. This is the shape of the contour of one- complete thread as.seen in axial
section.
8. Crest of thread. This is defined as the prominent part of thread, whether it be external or
internal.
9. Root of thread. This is defined as the bottom of the groove between the two flanks of the
thread, whether it be external or internal.
10. Flanks of thread. These are straight edges which connect the crest with the root.
11. Angle of thread {Included angle). This is the angle between the flanks or slope of the
thread measured in an axial plane.
12. Flank angle. The flank angles are the angles between individual flanks and the
perpendicular to the axis of the thread which passes through the vertex of the fundamental
triangle. The flank angle of a symmetrical thread is commonly termed as the half- angle of
thread.
13. Pitch. The pitch of a thread is the distance, measured parallel to the axis of the thread,
between corresponding points on adjacent thread forms in the same axial plane and on the same
side of axis. The basic pitch is equal to the lead divided by the number of thread starts. On
drawings of thread sections, the pitch is shown as the distance from the centre of one thread
crest to the centre of the next, and this representation is correct for single start as well as multi-
start threads.
14. Lead. Lead is the axial distance moved by the threaded part, when it is given one complete
revolution about its axis with respect to a fixed mating thread. It is necessary to distinguish
between measurements of lead from measurement of pitch, as uniformity of pitch measurement
does not assure uniformity of lead. Variations in either lead or pitch cause the functional or
virtual diameter of thread to differ from the pitch diameter.
15. Thread per inch. This is the reciprocal of the pitch in inches.
16. Lead angle. On a straight thread, lead angle is the angle made by the helix of the thread at
the pitch line with plane perpendicular to the axis. The angle is measured in an axial plane.
17. Helix angle. On straight thread, the helix angle is the angle made by the helix of the thread
at the pitch line with the axis. The angle is measured in an axial plane.
18. Depth of thread. This is the distance from the crest or tip of the thread to the root of the
thread measured perpendicular to the longitudinal axis or this could be defined as the distance
measured radially between the major and minor cylinders.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 70
19. Axial thickness. This is the distance between the opposite faces of the same thread
measured on the pitch cylinder in a direction parallel to the axis of thread.
20. Fundamental triangle. This is found by extending the flanks and joining the points and C.
Thus in Fig. 13.2, triangle ABC is referred to as fundamental triangle.
Here BC=pitch and the vertical height of the triangle is called the angular or theoretical depth.
The point A is the apex of the triangle ABC.
21. Truncation. A thread is sometimes truncated at the crest or at the root or at both crest and
root. The truncation at the crest is the radial distance from the crest to the nearest apex of the
fundamental triangle. Similarly the truncation at the root is the radial distance from the root to
the nearest apex.
22. Addendum. For an external thread, this is defined as the radial distance between the major
and pitch cylinders. For an internal thread this is the radial distance between the minor and
pitch cylinders.
23. Dedendum. This is the radial distance between the pitch and minor cylinder for external
thread, and for internal thread, this is the radial distance between the major and pitch cylinders.
24. Major diameter. In case of a straight thread, this is the diameter of the major cylinder
(imaginary cylinder, co-axial with the screw, which just touches the crests of an external thread
or the root of an internal thread). It is often referred to as the outside diameter, crest diameter
or full diameter of external threads.
25. Minor diameter. In case of straight thread, this is the diameter of the minor cylinder (an
imaginary cylinder, co-axial with the screw which just touches the roots of an external thread
or the crest of an internal thread). It is often referred to as the root diameter or cone diameter
of external threads.
26. Effective diameter or pitch diameter. In case of straight thread, this is the diameter of
the pitch cylinder (the imaginary cylinder which is co-axial with the axis of the screw, and
intersects the flank of the threads in such a way as to make the width of threads and width of
the spaces between the threads equal). If the pitch cylinder be imagined as generated by a
straight line parallel to the axis of screw that straight line is then referred to as the pitch line.
Along the pitch line, the widths of the threads and the widths of the spaces are equal on a
perfect thread. This is the most important dimension at it decides the quality of the fit between
the screw and the nut.
27. Functional (virtual) diameter. For an external or internal thread, this is the pitch diameter
of the enveloping thread of perfect pitch, lead and flank angles having full depth of engagement
but clear at crests and roots. This is defined over a specified length of thread. This may be
greater than the simple effective diameter by an amount due to errors in pitch and angle of
thread. The virtual diameter being the modified effective diameter by pitch and angle errors, is
the most important single dimension of screw thread gauge.
(i) Measurement of Major Diameter.
For the measurement of major diameter of external threads, a good quality hand micrometers
is quite suitable. In taking readings, a light pressure must be used as the anvils make contact
with the gauge at points only and otherwise the errors due to compression can be introduced.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 71
It is, however, also desirable to check the micrometers reading on a cylindrical standard of
approximately the same size, so that the zero error etc., might not come into picture.
For greater accuracy and convenience, the major diameter is measured by bench micrometers.
This instrument was designed by N.P.L. to estimate some deficiencies inherent in the normal
hand micrometers. It uses constant measuring pressure and with this machine terror due to pitch
error in the micrometers thread is avoided. In order that all measurements be made at the same
pressure, a fiducially indicator is used in place of the fixed anvil. In this machine there is no
provision for mounting the work piece between the centres and it is to be held in hand. This is
so, because, generally the centres of the work piece are not true with its diameter. This machine
is used as a comparator in order to avoid any pitch errors micrometers, zero error setting etc.
A calibrated setting cylinder is used as the setting standard.
The advantage of using cylinder as setting standard and not slip gauges etc., is that it gives
greater similarity of contact at the anvils. The diameter of the setting cylinder must be nearly
same as the major diameter. The cylinder is held and the reading of the micrometers is noted
down. This is then replaced by threaded work piece and again micrometers reading is noted for
the same reading of fiducially indicator. Thus, if the size of cylinder is approaching, that of
major diameter, then for a given reading the micrometers thread is used over a short length of
travel and any pitch errors it contains are virtually eliminated.

Then major diameter=D1+ (R2R1).
In order- to determine the amount of taper, the readings should be taken at various positions
along the thread and to detect the ovality, two or three readings must be taken at one plane in
angular positions.
(ii) Measurement of Minor Diameter
This is also measured by a comparative process using small Vee-pieces which make contact
with a root of the thread. The Vee-pieces are available in several sizes having suitable radii at
the edges. The included angle of Vee-pieces is less than the angle of the thread to be checked
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 72
so that it can easily probe to the root of the thread. To measure the minor diameter by Vee-
pieces is suitable for only Whitworth and B.A. threads which have a definite radius at the root
of the thread. For other threads, the minor diameter is measured by the projector or microscope.


The measurement is carried out on a floating carriage diameter measuring machine in which
the threaded work-piece is mounted between centres and a bench micrometers is constrained
to move at right angles to the axis of the centre by a Vee-ball slide. The method of the
application of Vee-pieces in the machine is shown diagrammatically in Fig... The dimensions
of Vee-pieces play no important function as they are interposed between the micrometers faces
and the cylindrical standard when standard reading is taken.
It is important while taking readings, to ensure that the micrometers be located at right angles
to the axis of the screw being measured. The selected Vees are placed on each side of the screw
with their bases against the micrometers faces. The micrometers head is then advanced until
the pointer of the indicator is opposite the zero mark, and note being made of the reading. The
screw is then replaced by standard reference disc or a plain cylindrical standard plug gauge of
approximately the core diameter of the screw to be measured and second reading of the
micrometers is taken.
If reading on setting cylinder with Vee-pieces in position=R1
And reading on thread =R2
And diameter of setting cylinder=D1
Then minor diameter =D1+ (R2R1)
Readings may be taken at various positions in order to determine the taper and joviality.
(iii) Effective Diameter Measurements.
The effective diameter or the pitch diameter can be measured by any one of the following
methods:
(i) The micrometre method
(ii) The one wire, two wire, or three wire or rod method.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 73
Procedure:
1. Two Wire Method.
The effective diameter of a screw thread may be ascertained by placing two wires or rods of
identical diameter between the flanks of the thread, as shown in Fig. 13.15, and measuring the
distance over the outside of these wires. The effective diameter E I s then calculated as
E=T+P
Where T= Dimension under the wires
=M2d
M=dimension over the wires, d= diameter of each wire

Fig (a) Fig (b)
The wires used are made of hardened steel to sustain the wear and tear in use. These are given
a high degree of accuracy and finish by lapping to suit different pitches. Dimension T can also
be determined by placing wires over a standard cylinder of diameter greater than the diameter
under the wires and noting the reading R1 and then taking reading with over the gauge, say R2.
Then T=S(R1R2).
P=It is a value which depends upon the dia of wire and pitch of the thread.
If P= pitch of the thread, then
P= 0.9605p1.1657d (for Whitworth thread).
P= 0.866pd (for metric thread).
Give the effective diameter. The expression for the value of P in terms of p (pitch), d
(Diameter of wire) and x (thread angle) can be derived as follows:
In Fig.13.15 (b), since BC lies on the effective diameter line
BC= pitch= p
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 74
OP=d cosec x/22
PA=d (cosecx21) 2
PQ=QC cot x2=p4 cot x2
AQ=PQAP=p cot x24 d (cosec x2 1) 2
AQ is half the value of P
.. P value=2AQ
=p2 cot x2 d (cosecx21)
Two wire method can be carried out only on the diameter measuring machine described for
measuring the minor diameter, because alignment is not possible by 2 wires and can be
provided only by the floating carriage machine. In the case of three wire method, 2wire, on one
side help in aligning the micrometre square to the thread while the third placed on the other
side permits taking of readings.
2. Three Wire Method.
This method of measuring the effective diameter is an accurate method. In this three wires or
rods of known diameter are used; one on one side and two on the other side {Fig.13.17 (a) and
(&)]. This method ensures the alignment of micrometre anvil faces parallel to the thread axis.
The wires may be either held in hand or hung from a stand so as to ensure freedom to the wires
to adjust themselves under micrometre pressure.
M=distance over wires E=effective diameter
r=radius of the wires d=diameter of wires
h =height of the centre or the wire or rod from the effective
x=angle of thread.

Fig: a fig: b
From fig. (b),
AD = AB cosec x2 = r cosec x2
H = DE cot x2 = p2 cot x2
CD = H = p4 cot x2
H = ADCD
r = cosec x2 p4 cot x2
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 75
Distance over wires=M = E+2h+2r
= E+2(r cosec x2 p4 cot x2) +2r
= E+2r (louse x2) p2 cot x2
Or M = E+d (1+cosec x2) p2 cot x2
(Since 2r = 0)
(i) In case of Whitworth thread:
X = 55, depth of thread = 0.64 p, so that
E= D0.64 p and cosec x2 = 2.1657
Cot x2 = 1.921
M = E+d (1+cosec x2) p2 cot x2
= D0.64p+d (1+2.1657) p2 (1.921)
= D+3.1657d1.6005p
M = D+3.1657d1.6p
Where D=outside dia.
(ii) In case of metric threads:
Depth of thread=0.6495p
So, E = D-0.6495p.
x = 60, cosec x2 = 2; cot x2 = 1.732
M = D0.6495 p+d (l+2)p2 (1.732)
= D+3d(0.6495+0.866)p
= D+3d1.5155p.
Result:

B) TOOL MAKERS MICROSCOPE:
Procedure:
1. Switch on the main.
2. Switch on the micros scope lights.
3. Select the capacity of the lens for precision operation.
4. Place the object on the class table to get the clear image rotate the wheel provided at the light
side.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 76
5. After getting the clear image, locate the crosswire at the initial point on the image. Now note
down the micrometre reading.
6. Move the cross wire from initial point to the finial point on the image, which is to be
measured. Note down the micrometre reading, this operation is done by using micrometre.
7. Now the different but when the initial and the finial reading i.e. distance travelled gives the
size of the object.
8. Graph can be plotted actual micrometre reading vs. % of error.
TABULATION:


Sl.
No



Actual
micrometre
reading in mm
(A)

Profile projector micrometre reading




Error
(A-D)
In
mm



% error
Initial(B)
mm

Final (c)
mm
Difference
b/w B& C (D)






CALCULATION:
% error = (error/actual micrometre reading) x100
FIND THE FLANK ANGLE:
Sl.
No

Initial angle in degree
(A)
Final angle in
degree (B)
Difference b/w
(A&B)
Average






RESULT:
Thus the all dimensions of the given particular screw were measured by using tool makers
microscope.




Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 77






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 78






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 79
Experiment No: 11
SURFACE ROUGHNESS MEASUREMENT BY TALY SURF
Aim: To measure the surface roughness using Taly surf instrument.
Apparatus: - Taly surf, work piece, surface plate.
Theory:
On any finished surface, imperfections are bound to be there and these take the form of a
succession of hills and valleys which vary both in height and in spacing and result in a kind of
texture which in appearance or feel is often characteristic of the machining process and
accompanying defects. The several kinds of departures are there on the surface and these are
due to various causes.
Methods of Measuring Surface Roughness:
Surface inspection of comparison methods
Direct instrument measurements
In comparative methods the surface texture is assessed by observation of the surface. But these
methods are not reliable as they can be misleading, if comparison is not made with surfaces
produced by same techniques. The various methods available under comparison method are:
(i) Touch Inspection (ii)Scratch Inspection (iii) Microscopic Inspection (iv) Visual Inspection
(v) Surface Photographs (vi) Reflected Light Intensity Direct Instrument Measurements enable
to determine a numerical value of the surface finish of any surface. Nearly all instruments used
are stylus probe type of instruments. These operate on electrical principles
Taylor Hobson Talysurf:
Talysurf is an electronic instrument working on carrier modulating principle. The measuring
head of this instrument consists of a diamond stylus of about 0.002mm tip radius and skid or
shoe which is drawn across the surface by means of a motorized driving unit(gear box), which
provides three motorized speeds giving respectively X20 and X100 horizontal magnification
and a speed suitable for average reading.


Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 80


FIG: NAME OF THE EACH PART ON THE DISPLAY BOARD
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 81


A neutral position in which the pick-up can be traversed manually is also provided. In this case
the arm carrying the stylus forms an armature which pivots about the centre piece of E-shaped
stamping as shown in fig. On two legs of the-shaped stamping there are coils carrying an a.c.
current. These two coils with two other resistances form an oscillator. As the armature is
pivoted about the central leg, any movement of the stylus causes the air gap to vary and thus
the amplitude of the original a.c. current flowing in the coils is modulated. The output of the
bridge thus consists of modulation only as shown in fig. this is further demodulated so that the
current now is directly proportional to the vertical displacement of the stylus only.
The demodulated output is caused to operate a pen recorder to produce a permanent record and
a meter to give a numerical assessment directly. In record of this statement the marking medium
is an electric discharge through a specially treated paper which blackens at the point of stylus
so this has no distortion due to drag and the record strictly rectilinear one.
Analysis of Surface Traces: A numerical assessment is assigned to indicate the degree of
smoothness (roughness) in a number of ways. In practice three roughness measures have shown
themselves to be particularly useful.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 82
1. Ra Centre Line Average (CLA) or Arithmetic Average (AA):
This is most widely used parameter for specifying surface roughness. It is the arithmetic mean
of the departures Y of the points on the profile from the mean line shown in fig.



The mean line is first determined and the ordinates of the points from the mean line are added
without considering the sign (i.e. irrespective of whether the points are above or below the
mean line)

Where n is the number of ordinates in the sampling length L and Y is the ordinate height.
Ordinates are taken at equal intervals. The CLA can also be calculated from the areas between
the profile and the mean line (shown in fig.)


Where P1, P2, P3 and Q1, Q2, Q3 are areas in mm2
L is the sampling length in mm Mis the vertical magnification of the recorded profile.
Talysurf has got built in arrangement from integrating the areas and the average value is
directly given.
2. Roughness Average or Root Mean Square Average Height (RMS) R q:
It is defined as the average root mean square deviation Y of the profile from its mean length
within the sampling length L.

Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 83
Where n is the number of ordinates in the sampling length L.
3. Rz Ten Point Peak to Valley Average Height: Rz
Is the average difference between the five highest peaks and five deepest valleys within the
sampling length, the heights being measured from a line parallel to the mean line and not
crossing the profile (shown in fig.)



PROCEDURE:
The finished component is placed on the surface plate.
Talysurf tester is fixed to the vernier height gauge using adopter at a convenient height.
Make sure that the stylus probe touches the work piece.
Fix the sampling length in the tester.
Then the power button is pressed so that the probe moves on the surface to and fro.
Take the readings of the surface roughness directly from the instrument.
Repeat the above process for the remaining specimen and tabulate the readings
PRECAUTIONS:
The surface to be tested should be cleaned properly.
The tester should be fixed to the height gauge properly so that the movement of the probe is
exactly parallel to the surface of work.
Make sure that the probe gently touches the work






Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 84
Observations and Tabulation:
S.No

Measurement roughness value
m
Sample, direction Ra, Rz
Average
Ra

Average
Rz

Grade
1


2


3


4


5



Result:



















Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 85






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 86
Experiment No. 12 Date: / / 20
PROFIILE PROJECTOR
AIM:
To calibrate the profile projector using given samples which dimensions is measured by
micrometre as standard.
CONSTRUCTIONAL DETAILS:
Profile projections are highly sophisticated and versatile designed as per international
standards. This comprehensive range covers all conceivable application its ideal for the rapid
inspection and measurement (linear and angular) of small to medium size components such as
watch parts, tools, rubber components, miniature electronic assemblies and so on.
Its best quality high resolution optics provides accurate, bright, clear and sharp images. The
special front and back surface mirror are highly polished and lobbied distortion and
reproduction. Three element condenser system and high intensity halogen lamps provides
brilliant images even in day light condition commitment to quality insure that offer the highest
level of precision, quality, reliability and performance.
Two types profile projectors are,
Vertical floor model, ideal for the rapid inspection
Horizontal floor model, ideal for the tracing for projected images
PROJECTION CAPACITY:
Projection lens Dial of view field Free working distance
between object and lens
10:1 32mm 23mm
20:1 12.5mm 38mm


FIG: PROFILE PROJECTOR
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 87
TABULATION:


Sl.
No



Actual
micrometre
reading in mm
(A)

Profile projector micrometre reading




Error
(A-D)
In
mm



% error
Initial(B)
mm

Final (c)
mm
Difference
b/w B& C (D)








CALCULATION:
% error = (error/actual micrometre reading) x100
FIND THE FLANK ANGLE:
Sl.
No

Initial angle in degree
(A)
Final angle in
degree (B)
Difference b/w
(A&B)
Average






OPERATING PROCEDURE:
1. Switch ON the main. The induction glows, then the instrument is ON.
2. Switch ON the toggle switch. For cooling fan below the light house.
Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 88
3. Toggle switch 2 for light source ON.
4. Select the capacity of the length for precision operation measured and fix that flow the
projection head.
5. Place the object (balls) on the glass table. TO get the clear image rotate the arm wheel
provided at the right side.
6. After getting the clear image locate the cross wire of the initial point on the image which to
be measured, and then the cross wire moved to the final point. Note down the micro meter
reading. This is done by using micro meter.
7. Now find the difference between initial and final readings.
8. Graph can be plotted between actual micro meter readings Vs. % Error.
RESULT:
Thus the dimension of given samples are measured by using profile projector.




















Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 89






























Prepared by SUDARSHAN BOLLAPU., M.TECH (Ph.D.) DEPARTMENT OF MECHANICAL ENGINEERING, CUTM 90
STUDY OF MACHINIG CNC:
1. Machining of simple components on CNC lathe
2. Machining of simple components on CNC Milling
3. Inspection of quality and dimensional practice using Coordinate Measuring Machine