Sean Erik Foss
Submitted
in partial fulfilment of the requirements
for the degree of
Doctor Scientarium
Department of Physics
Faculty of Mathematics and Natural Sciences
University of Oslo
Oslo, Norway
September, 2005
Abstract
Combining optics, electronics and mechanics on one miniature platform is
an emerging reality in micro device technology. An important goal of this
is the simplification and enhancement of actions in every-day life, e.g. labon-a-chip for full characterization of blood samples with integrated loading,
transportation, manipulation and analysis on one chip. Many elements are
required for this to work, among them the control and manipulation of light.
This thesis presents a study of the use of porous silicon within this scope.
Optical filters changing the spectral characteristics of light are fabricated
in an electrochemical etching process of silicon in solutions containing hydrofluoric acid. The aim of this investigation is to fabricate high quality optical transmission and reflection filters in the near-infrared making use of the
special properties of porous silicon which are hard to achieve in other optical materials, and at the same time enhance micro-opto-electro-mechanicalsystem technology in silicon by adding the possibilities presented by porous
silicon.
Both discrete layer and graded index optical filters (rugate filters), with
and without laterally dependent filter functions, are fabricated showing the
versatility of this process. However, to obtain high quality optics with
porous silicon, a very good control of the etch process is needed. For this
reason, equipment has been developed for monitoring the most important
etch parameters in situ; depth/time dependent porosity, etch rate, and
interface roughness. The technique is based on interference effects in an
infrared laser beam partly reflected off the different interfaces in a sample
during etching of a porous layer. The information obtained is later used to
control the etching of the designed structures.
Several device designs and ideas incorporating multilayer or graded index
porous silicon are included at the end of the thesis.
Acknowledgements
During the last few years which have brought me to the point where I am
writing these acknowledgements, I have had the great fortune of receiving
help in one form or other from many people. My adviser, Terje G. Finstad,
has guided me patiently through the latest part of my education as a scientist. He has done this by always giving me time and sharing his abundance
of knowledge. I am greatly indebted to him for the opportunity he gave
me! A great thanks is also due my second adviser,
Asmund Sudb. He has
given me an interesting insight into the enlightening world of optics.
Fortunately, I attended a few conferences in connection with my studies. This led me to some interesting discussions with Hans Bohn of
Forschungszentrum J
ulich, Germany, who gave me some suggestions on
rugate filters in porous silicon, and Gilles L`erondel at UTT, France, who
shared some of his indepth understanding of the porous silicon multilayer
etching process.
As ideas evolved and I wanted to test new things in the lab, having the
mechanical workshop and the electronics lab and the people at these places
has been invaluable. Thanks for all the help.
Had I been been alone in the lab or by my computer day after day I had
surely gone mad, so I owe much of my still fairly sound sanity to my colleagues with whom I have discussed everything between science and the
weather. Especially thanks to Ingelin Clausen, Chenglin Heng, Klaus Magnus Johansen, to name a few. Thanks also to H
avard Alnes for being my
bad conscience and keeping me reasonably fit. Erik Marstein deserves a
special thanks for being a good friend and also introducing me into the
secrets of porous silicon. With such an enthusiasm, how can one not think
that whatever he is doing is the most important thing in the world?
My family has always supported me and lent me a helping hand whenever
needed, which I am very grateful for. Thank you mom and mormor. Part
of the reason why I ever thought of doing a PhD is my late uncle Larry. He
has always been an inspiration in both character and career. Thank you for
giving me the opportunity to know you. Last but definity not least I am
forever indebted to my wife, Hilde, for her patience and unfailing confidence
in me. This could not have been done without you my dear.
iii
Contents
Abstract
Acknowledgements
iii
1 Introduction
2.1
2.2
2.2.1
Formation . . . . . . . . . . . . . . . . . . . . . . . .
2.2.1.1
Chemistry . . . . . . . . . . . . . . . . . . .
2.2.1.2
I-V characteristics . . . . . . . . . . . . . .
2.2.1.3
Morphology . . . . . . . . . . . . . . . . . .
2.2.1.4
Formation theories . . . . . . . . . . . . . .
10
2.3.1
Sample . . . . . . . . . . . . . . . . . . . . . . . . . .
10
2.3.1.1
Doping . . . . . . . . . . . . . . . . . . . .
10
2.3.1.2
Preparation . . . . . . . . . . . . . . . . . .
10
2.3.1.3
Resistivity variations . . . . . . . . . . . . .
10
2.3.1.4
Drying . . . . . . . . . . . . . . . . . . . . .
13
Electrolyte properties . . . . . . . . . . . . . . . . . .
14
Etch setup . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
2.4.1
16
17
2.3
2.3.2
2.4
2.5
3 Thin-film calculations
3.1
21
22
vi
3.2
3.3
3.4
Reflectance calculation . . . . . . . . . . . . . . . . . . . . .
25
3.2.1
Characteristic matrix . . . . . . . . . . . . . . . . . .
25
3.2.2
Admittance matrix . . . . . . . . . . . . . . . . . . .
26
Roughness calculation . . . . . . . . . . . . . . . . . . . . .
28
3.3.1
Davies-Bennett theory . . . . . . . . . . . . . . . . .
29
31
3.4.1
31
3.4.2
Inhomogeneous layers . . . . . . . . . . . . . . . . . .
33
4.2
41
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
42
4.1.1
Usage . . . . . . . . . . . . . . . . . . . . . . . . . .
42
4.1.2
Laser . . . . . . . . . . . . . . . . . . . . . . . . . . .
43
4.1.3
Fiber . . . . . . . . . . . . . . . . . . . . . . . . . . .
46
4.1.4
46
4.1.5
Other equipment . . . . . . . . . . . . . . . . . . . .
49
Data analysis . . . . . . . . . . . . . . . . . . . . . . . . . .
49
4.2.1
Chemical etching . . . . . . . . . . . . . . . . . . . .
49
4.2.2
51
4.2.3
Frequency analysis . . . . . . . . . . . . . . . . . . .
51
4.2.4
53
4.2.4.1
53
4.2.4.2
Etch calibration
. . . . . . . . . . . . . . .
55
4.2.4.3
57
Paper I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
61
Paper II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
75
Paper III
83
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Filter fabrication
97
5.1
98
5.2
99
5.2.1
5.2.2
vii
5.2.3
5.3
5.4
5.5
5.4.1
5.4.2
5.4.2.2
5.4.2.3
Paper IV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Paper V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
6 Porous silicon applications for MOEMS and passive optics133
6.1
6.2
6.1.2
6.1.3
6.1.4
6.2.2
6.2.3
6.2.4
7 Conclusion
145
Bibliography
147
Chapter 1
Introduction
Roughly 50 years of development has made silicon the material of choice
in the electronics industry. As so much effort and investment has been
put into silicon technology, an increasing focus is now directed towards silicon photonics. There are other materials available, and with fairly mature
technologies, that are technically superior to silicon for specific uses in photonics. However, to be able to integrate the different elements of a photonic
circuit, which is where a synergy is possible, as well as making the technology commercially viable, only silicon technology has the potential of solving
all the challenges. Why photonic circuits? The basic assumption is that
photons move faster than electrons and can therefore move data quicker,
hence faster data processing. A second assumption is that there is less
of a heating problem with photons, leading to lower power consumption.
Before we see a wholly integrated photonic/optical circuit there are many
problems which need to be addressed. This thesis deals with a technology
which may broaden the usage of silicon for optical elements and thereby
help solve some of the details of these challenges. The last couple of years
have shown great improvement in silicon photonics, with a couple of significant breakthroughs such as an all silicon Raman laser [1, 2] and silicon
optical modulators [3].
By porosifying silicon, several material properties of silicon undergo change.
This may be used to expand the application possibilities of silicon, both by
introducing new concepts to silicon technology and by improving already
applied ideas. A note on terminology is appropriate here. Porosified or
porous silicon has had many abbreviations, in this thesis PS will be used
and where any ambiguity is possible this will be explicitly noted.
Engineered pores in silicon may be in many forms and sizes. This leaves us
with a material in one extreme where the pores do not change any material
properties but act more as a mechanical construction, in the other extreme
one reaches the quantum limit and the material properties are affected by
quantum effects, namely a widening of the band-gap due to quantum confinement in the nm-size silicon crystalline dots and rods. The properties
1
2
affected by porosification are the electrical and thermal resistivity, refractive index, luminescence, stiffness, chemical reactivity and so on. Of these
properties the control of refractive index will be the focus of this thesis.
In principle, by changing the porosity from 0 to 100% (no or all material
removed) the refractive index for a given photon wavelength will change
from that of the silicon bulk material to that of air. This, together with the
fact that the porosity value may be controlled in space and that the size
of the pores is also controllable, most importantly in this case to be much
smaller than the photon wavelength of interest, makes it possible to have
control of the photon trajectory in the material.
The thesis and work done is divided in two parts: the processing/fabrication
of porous silicon and the use of this knowledge for fabrication of devices or
elements of devices. This work has in practice been carried out in parallel with feedback between the two areas indicating where better control
of parameters is needed and which parameters are important for fabrication. Chapter 2 will go into some basics of porous silicon processing and
physics. The next chapter will discuss some of the theory needed for designing optical filters. Chapter 4 presents the work done on etch parameter
monitoring while Chapter 5 discusses some experimental details of the processing of porous silicon. The sixth chapter takes a look at the possibilities
of porous silicon in electronic/photonic devices with an emphasis on microopto-electro-mechanical-systems. A conclusion will be presented in the last
chapter.
Five papers are included in the thesis, presenting the main results:
Paper I:
Paper II:
Paper III:
Paper IV:
Paper V:
3
Mat. Res. Soc. Symp. Proc., 797, W1.6.1 (2004)
Paper I deals with the setup of a fiber optic based in situ infrared reflectance experiment and the theory behind the analysis of the obtained
data. Etch rate, porosity depth profiles and PS-substrate interface roughness are obtained. Some measured data is presented showing the accuracy
of the technique.
Paper II presents experiments on the effect of glycerol and HF concentration in the electrolyte on PS-substrate interface roughness based on in
situ IR reflectance measurements during etching. The roughness showed a
dependence on both glycerol and HF concentration, while porosity showed
only a weak dependence on glycerol concentration.
Paper III discusses in detail the calculation of roughness from in situ infrared reflectance data. Results are presented showing the dependence of
PS-substrate interface roughness on HF and glycerol concentration, temperature and formation current density, and the relative importance of these
parameters.
Paper IV presents calculations of the effect of a laterally graded optical
filter on the filter characteristics. The calculations are based on ray-tracing
through the graded layer stack. It is shown that both a small gradient in
layer thicknesses and a small divergence in the incident beam widens and
reduces the stop band reflectance.
Paper V describes fabrication of laterally graded rugate reflection filters.
Filters with a shift in the reflection band center wavelength of up to 100
nm pr. mm across the filter surface are realized. The shift is close to linear
with position, but a broadening of the reflection band is observed with a
greater gradient.
Contributions to other papers were made during the work on this thesis,
however, these papers are not included as the subject matter is besides the
focus of the thesis. These papers are:
P.Y.Y. Kan, T.G. Finstad, H. Kristiansen and S.E. Foss
Porous silicon for chip cooling applications
Physica Scripta, T114, 2004
P.Y.Y. Kan, S.E. Foss and T.G. Finstad
Thick etch-through macroporous Si membrane from p- & n-Si, and fast
pore etching and tuning the pore size from n-Si
Submitted to the E-MRS 2005 Spring Meeting, Strasbourg
Chapter 2
Porous silicon formation
2.1
6
PS spontaneous emission with the cavity mode of the multilayer structure.
This again led to the incorporation of materials into the porous structure
to take advantage of the increased coupling between field and matter in
the micro-cavity. The doping with erbium ions in a Fabry-Perot structure
resulted in an enhanced infrared (IR) PL at a peak wavelength of 1.536 m
[19].
Due to the flexibility and relative ease of fabricating multilayer structures
in PS, many different structures have been reported. The basic optical multilayer structures such as Bragg mirrors and Fabry-Perot filters have been
mentioned. These may be constructed such that the optical characteristics
change due to an external effect. As the spectral features of these structures
may be very sharp and narrow, slight changes in layer optical thickness will
significantly change the spectral features. Incorporation of liquid crystals
into the pores of a PSM structure lead to the ability of modulating the filter characteristics by controlling a voltage [20, 21, 22], or alternatively, the
filter characteristics could be controlled by a temperature modulation [23].
This may be used for optical switching. Alternatively, one may have polarization dependent filter characteristics as in the case of dichroic filters
etched in 110 Si [24]. These optical elements may also be used as sensors,
either by utilizing the change in spectral characteristics as an indication
of filling ratio, e.g. measure amount of air moisture, amount of liquid in
the pores, or the pore walls may be sensitized, or activated, with different materials which react to more specific molecules [25, 26]. There are
many other structures possible as well; L`erondel et al. [27] have fabricated
a diffraction grating in PS by light assisted etching using two interfering
laser beams incident on the sample surface during normal etching. Volk et
al. [28] have reported lateral PSMs for use as ultraviolet diffraction gratings
using a buried n-doped region to control the current distribution close to
the surface. Further, as will be an important part of this thesis, the possibility of arbitrarily controlling the refractive index with depth has resulted
in the fabrication of rugate filters of different kinds [29, 30].
2.2
2.2.1
Formation
2.2.1.1
Chemistry
(2.1)
(2.2)
Here the SiH2 is bound to the the Si surface. In this reaction hydrogen gas
is formed which may interfere with the etching. In this general reaction
two holes are needed for each Si atom dissolved, hence the valence of the
reaction is two. It may vary, with normal values for PS formation between
2 and 2.8. This reflects both the model used to calculate the valence and
also the complexity of the reaction. The overall valence of the reaction may
be roughly calculated by using the etch rate and porosity:
=
j Ar,Si
.
NA Si e r P
(2.3)
Here is the valence, j the current density (A/m2 ), NA = 6.02 1023 mol1
the Avogadro number, Ar,Si = 28.09 gmol1 the molar mass of silicon,
Si = 2.33 gcm3 the density of silicon, e = 1.60 1019 C the electron
charge, r the etch rate (m/s) and P the porosity (absolute values).
2.2.1.2
I-V characteristics
8
is illuminated during etching there will be photon generated holes which
may react with the F ions. This is often used for n-type PS etching,
while, as can be seen in Fig. 2.1, for anodically biased p-type PS formation
this has only a small, if any, effect as the photon generated holes have
an insignificant concentration. However, to avoid any uncertainty, samples
presented in this thesis have been etched in the dark. The I-V curves have
some similarities with a Schottky diode I-V curve. However, there are a
few important differences, most importantly the two peaks on the curve
under forward bias. The first peak signifies the start of electropolishing,
while the second marks the onset of current oscillation. Electropolishing is
driven by a slightly different reaction than that indicated in the reaction
represented by Eq. 2.1, namely a reaction of valence four. In this case the
dissolution of Si is not direct, but goes through an oxidation step first.
There are some general trends which are seen in I-V curves of this system,
such as Fig. 2.1. By increasing the HF concentration, the first peak shifts to
higher current values (higher electropolishing current), while increasing the
substrate doping concentration shifts the first peak towards lower voltages.
2.2.1.3
Morphology
The result of the electrochemical etching of Si within the limitations discussed above is in general a porous structure. Generally, little will happen
with the pore walls of the already etched structure as the electrochemical
reactions take place at the pore-front. However, depending on the parameters of the etching, the structure will vary greatly. One property which will
be sensitive to several parameters, like current density, sample resistivity,
HF concentration and solvent composition, is the pore size. The classification used is defined by the International Union of Pure and Applied
Chemistry and describes porous materials in general: pore-sizes less than
2 nm are denoted micro, between 2 and 50 nm are meso and above 50 nm
are macro. The PS films fabricated for this thesis are mostly meso-porous
as may be seen in the scanning-electron-microscope (SEM) micrograph in
Fig. 2.2. This picture shows the surface of a typical filter structure with
9
a median pore-size of approximately 15 nm at the surface. The structural
morphology in general is very sensitive to the formation parameters. A
short summary and categorization of different pore morphologies is given
in Ref. [33]. Pores may be sponge-like, straight or branched, random or
aligned along the 100 crystalline axis of the sample. Macro-pores may be
filled by micro-pores or empty, just to mention a few possibilities.
2.2.1.4
Formation theories
How the pores are formed is also a question still under some discussion.
There are three predominant models; the Beale model, the quantum confinement model and the diffusion-limited model. The Beale model [13]
proposes that the pore-walls in meso- and micro-PS are depleted of charge
carriers due to overlapping depletion layers resulting in a concentration of
the electrical field at the pore tips, hence an increased concentration of holes
with a resulting etching at the pore tips. The quantum confinement model
suggested by Lehmann and Gosele [35] is based on the quantum confinement of charge carriers in the nanometer sized Si pore-walls of micro- and
meso-PS. This quantum confinement will lead to an increase in the band
gap compared to bulk Si. This introduces a barrier for the holes going from
the bulk to the porous Si-structure whereby the hole concentration increases
close to the pore tips resulting in a dissolution of Si. The diffusion-limited
model [34] describes the formation of pores as a result of a random walk
process of the holes. In this model the holes moving towards the electrolytesubstrate interface will most likely reach a pore tip first, hence the formation
of PS is limited by the diffusion of the holes. Carstensen et al. [36] have also
introduced a model called the current-burst model to explain pore growth
in Si. Etching in this case occurs in bursts, both temporal and spatial (i.e.
at discrete positions). The passivation of the pore walls is in this model
a result of hydrogen termination. These models may describe micro-, and
10
to some extent, meso-PS, however, macro-PS usually does not have overlapping depletion layers in the pore walls, so the formation mechanism is
somewhat different [37]. Formation of macro-PS will not be discussed here.
2.3
2.3.1
Sample
2.3.1.1
Doping
The doping concentration and type of the sample are crucial parameters
for PS formation. As there is no need for external lighting when etching ptype Si and the fairly low PS-substrate interface roughness obtained, most
PSM structures are etched in p-type samples. The obtained morphology
and porosity ranges are dependent on the resistivity of the sample. Samples of high resistivity tend to give microporous PS which are very brittle
and the controllable porosity range is rather narrow. With lower resistivity
samples, the interface roughness (microscopic) tends to decrease, although
macroscopic roughness, i.e. due to striations, tends to increase. The porosity range of highly doped samples is quite large. The samples presented in
this work are p-type, boron doped with a nominal resistivity of <0.1 cm,
measured to be around 0.018 cm. Nominal sample thickness was 520 m.
2.3.1.2
Preparation
Before etching, an Al-back contact is evaporated on the samples and annealed to give a good ohmic contact. This is crucial for a homogeneous
current density distribution over the etched area, also for highly doped samples. Different back-contact geometries may be used as will be discussed
in Sec. 2.4.1. The samples are ultrasonically cleaned in trichloroethylene,
acetone and DI-water before etching. This process seems crucial, as badly
cleaned samples show inhomogeneities in the spectral characteristics of the
etched optical filters.
2.3.1.3
Resistivity variations
11
the refractive index and rough layer interfaces will be detrimental to the
optical quality.
Both refractive index inhomogeneities and PS-substrate interface roughness
have been observed. In Fig. 2.3 the reflectance spectrum of a rugate optical
reflectance filter is measured. Details of the optical filters are discussed in
Chap. 3. Two measurements are made at different positions, 1 mm apart,
on the filter surface. The reflection bands are shifted relative to each other
which indicates different conditions for interference. This is most likely due
to local differences in etch conditions, e.g. resistivity differences. These
inhomogeneities are often visible on the surface of the optical filters, fabricated for this thesis, as slight deviations in color. These deviations take
the form of concentric circles often roughly coinciding with the center of
the wafer. This may be seen in Paper V where a grayscale optical microscope image clearly shows local differences in color. The difference in color
stems from porosity and layer thickness variations. By selectively removing
the PS by etching in a concentrated alkaline solution, e.g. 40 % NaOH,
the PS-substrate interface is revealed. Figure 2.4 shows a 3D surface plot
of height data obtained by white-light interferometry from a test sample.
One may clearly see ridges caused by spatially inhomogeneous etch rates.
These ridges coincide with the color deviations. This is discussed more in
Paper V. The spatial period with which these ridges occur along the radial
direction is in the 100 m to 1 mm range.
The cause of these inhomogeneities in resistivity is most likely striations,
or fluctuations in dopant concentration caused by the Si-ingot production
process. This is a well known problem in Si-technology and is thoroughly
discussed in technical papers on Si wafer material quality, e.g. Ref. [38]
gives a summary of semiconductor crystal growth specifically discussing
striation formation. The striation induced roughness has rather long spatial
periods and may in p+ samples give quite large surface height fluctuations.
In lower doped p-type Si samples the striation induced roughness is less
pronounced [39]. This may be understood considering that the same dopant
fluctuation relative to the average dopant concentration is likely to occur in
both highly doped and low doped Si, with the result that a different absolute
change in etch rate is observed. However, there will be interface roughness
with smaller spatial periods (micro-roughness) which is more pronounced in
p samples than in p+ samples where this type of roughness is very small.
This results in locally very good optical quality of p+ PS based optics,
and promises good quality optics on larger area when striation effects are
controlled.
The resistivity, , does not only change at a local (m) level, but also on a
wafer level. This has an impact on the reproducibility of filter fabrication.
In Fig. 2.5 the result of a four-point-probe resistivity measurement across
a typical wafer is shown. Standard geometrical correction factors from
Ref. [40] for thin, circular disks are used to calculate the resistivity. Distance
to wafer edge, orientation of the four-point-probe, wafer thickness and probe
12
0.5
Shift: 11 nm
Reflectance (abs)
0.4
0.3
FWHM = 26 nm
0.2
0.1
0.0
550
600
650
700
750
Wavelength (nm)
Figure 2.3: Reflectance spectra of a rugate filter measured at two different positions, 1 mm apart, on the filter surface. There is a small
wavelength shift of the reflection band indicating different
etch conditions due to sample
resistivity inhomogeneities.
Figure 2.4: A surface plot
of white-light interferometry
measurement data from the
PS-substrate interface after removal of the PS by an alkaline etch.
The PS film
was about 117 m thick. The
ridges due to local etch rate
differences are clearly visible.
These are caused by striations,
or dopant inhomogeneities in
the substrate.
U
I
t T2
ln 2
t
s
C0
s
K2
d
d
,
d s
F4 (t, s)
(2.4)
13
different. Different resistivity and bias will likely change the depletion layer
at the pore front which may change the resulting porosity or structure of the
PS. There have been few, if any, systematic investigations of small changes
in wafer resistivity on the morphology of PS with optical applications in
mind. A similar example to that in Fig. 2.3 of the effect of local resistivity
variations on filter characteristics is given by Lerondel et al. in Ref. [41].
2.3.1.4
21
-3
cm)
20
Resistivity (10
19
18
17
16
20
40
60
80
100
Drying
After etching, before the samples are taken out of the etch-bath, the bath
with the sample in it is rinsed out with ethanol. The sample is then taken
out to air dry. Because of the size of the pores the capillary stress within the
pores may be quite high. Depending on the size of the structure (porosity)
and the surface tension of the liquid, cracking of the PS layer may occur.
This limits the maximum porosity obtainable and also suggests a procedure
for drying [42, 43, 44]. When drying in air, a meniscus will always form
in the pores which will result in a stress on the pore walls. This makes it
important to have a very low surface tension liquid in the pores when drying.
As suggested this may be done by rinsing out with pure ethanol which has
a lower surface tension than water (22 mJ/m2 compared to 72 mJ/m2 ), an
alternative is to use pentane (with a surface tension of 14 mJ/m2 ). Pentane,
however, is not water-soluble so the sample is usually rinsed in ethanol first.
The best results, however, have been obtained by supercritical drying [44]
in CO2 (>95 % porosity) where drying is performed above the supercritical
point of a liquid, usually CO2 .
In the electrolyte and immediately after drying, the pore walls are mostly
H-terminated [34]. The hydrogen will be replaced by oxygen to form native
oxide quite rapidly in air. This will change the properties of the PS over
time. Due to the large surface area of the PS the silicon-oxygen ratio may
be quite large resulting in a significant impact on the properties of PS. For
many applications this instability is not acceptable and several methods
14
for surface passivation have been reported in the literature. Among these
are controlled oxidation by anodic or chemical oxidation, rapid thermal
oxidation, capping of the PS layer by a dielectric or metal [45], thermal
nitridation or thermal carbonization [46].
2.3.2
Electrolyte properties
The electrolyte contents used for PS etching may vary substantially, however, the electrolyte is generally based on aqueous HF. For all experiments
reported here, a 40 % aqueous HF has been used as the base. It is quite
possible to etch PS with this base diluted in water, however, to facilitate extraction of hydrogen bubbles formed during etching, ethanol is usually used
as a surfactant. Compared to water, ethanol has better wettability and
lower surface tension which results in better infiltration in the nanometersized pores. Different additions or substitutions may be made to change the
properties of the electrolyte, e.g. to increase viscosity which is thought to influence the PS-substrate interface roughness, glycerol may be added. Other
substitutions include other organic solvents, especially dimethyl formamide
(DMF) and dimethyl sulfoxide (DMSO) which result in p-type macro-PS
for certain parameters. A short overview of the different electrolyte compositions reported in the literature is given in Ref. [32].
In papers II and III the effect of different electrolyte parameters is discussed.
Electrolytes containing different ratios of glycerol are used while measuring
the PS-substrate roughness evolution during etching. A comparison of room
temperature and low temperature etch is also made. Both temperature and
glycerol content seem to affect the interface roughness, however, the degree
depends on other parameters like HF concentration. For some parameter ranges the roughness decreases. It has been suggested [39, 47] that
the reduction in roughness with decreasing temperature, down to -40 C,
and increasing glycerol ratio is due to an increase in viscosity. Data from
Ref. [48] suggest that a mix of water and glycerol (25 %) at 20 C has the
same viscosity as an equivalent mix of water and ethanol at about 12 C.
The exact viscosity values of the electrolytes will differ from these, but the
closeness of the tabulated data indicates that viscosity may be a critical
parameter. However, it is not obvious that an increase in viscosity itself is
the only reason why lower roughness is obtained. Especially in the case of
low temperature etching, the reaction kinetics will most likely be affected.
The tentative explanation why a change in electrolyte viscosity affects the
interface roughness of PS, and also the refractive index inhomogeneity, is
that a situation closer to that of electropolishing is reached. During electropolishing the holes diffuse faster to the interface than do active electrolyte
species (e.g. F ), which results in a guaranteed availability of holes at
the surface. This has the consequence that peaks are etched first, hence
the resulting surface is locally flat, where the extent of the locality depends
15
on a characteristic length (e.g. diffusion length of holes). By reducing the
diffusion of ions in the electrolyte during PS etching, the local differences of
hole availability caused by an inhomogeneous resistivity will be reduced. In
the extreme case of ion diffusion controlled etching, the etch rate and porosity should be independent of resistivity. Some results of etching in glycerol
containing electrolytes and low temperature etching will be presented in
Chapters 4 and 5.
2.4
Etch setup
There are usually two ways of setting up a sample for etching. These are
normally referred to as single etch cell and double etch cell. In the single
etch cell the sample is usually horizontal with a solid back contact, often
a Cu-plate, and the front side is in contact with a reservoir containing the
electrolyte. In the electrolyte there will be a Pt-cathode. In the double cell
the sample is vertically placed between two separate reservoirs containing
the electrolyte, both with Pt-electrodes, one working as cathode and one
as anode. Here the electrolyte reservoir works as a back side contact. All
samples fabricated for this thesis were made with a single cell setup. A
vertical single cell setup was used for some tests, but this resulted in etch
rate and porosity inhomogeneities caused by H2 bubble formation and trapping. Vertical ridges were observed at the PS-substrate interface after PS
stripping by concentrated NaOH which were most likely caused by bubble
induced change in etch rate.
A sketch of a basic etch cell used in this thesis is shown in Fig. 2.6. The
back contact upon which the Si sample is placed is made of copper, on top
of the sample, between the sample and the top part of the cell, is a sealing
ring or a sheet with an opening. The cell is made of Teflon or another
material inert to HF. Several different variations of this basic cell have
been used. The etch current is supplied by a computer-controlled Keithley
2400 sourcemeter. Etching is normally performed under galvanostatic, or
constant current, conditions, where current is the control parameter.
The biasing voltage was monitored during single PS layer fabrication to
detect anomalies, such as significant changes in voltage due to leakage.
In Fig. 2.7 a typical voltage monitored during an etching is shown. It is
included here due to the curious curve shape. The plot contains surprisingly
many features considering the sample was etched with a constant current.
There is a transient region in the beginning which may be ascribed to a build
up of charge before etching begins, e.g. due to an activation energy. The
irregular sawtooth pattern may be due to local oxide build-up and etching,
as in the current-burst model [36] or due to hydrogen bubbles interfering
with electrolyte flow.
16
Figure 2.6: A sketch of the
basic etch cell used. The Si
sample is placed on a Cu-plate
which is positively biased, an
o-ring or silicon rubber sheet
(HF-resistant) is placed on top
of the sample before the top
part of the cell is fastened.
The cell material is either
Teflon or PVC. The electrolyte
is filled into the reservoir and
a Pt-electrode is placed in the
electrolyte. The Pt-electrode
is negatively biased. A computer controlled current source
controls the current/bias.
Voltage (V)
1.50
1.48
1.46
1.44
20
40
60
80
100
120
Time (s)
2.4.1
The geometry of the etch cell will influence the current density distribution
in the sample. In an effort to optimize the etch cell for homogeneous etching, a simplified 2D finite element method (FEM) calculation was done in
Femlab [49] to understand the current density distribution. Three factors
were tested, the geometry of the top part of the cell, i.e. the electrolyte
reservoir, the size of the back-contact and the influence of an opening in the
back-contact. The sketch in Fig. 2.8b) shows the cell geometries tested. The
wide, funnel-shaped cell has a top opening of 2 cm diameter. The sample
opening for both designs is 1 cm diameter, while the height is 2 cm for the
funnel design and 4 cm for the straight reservoir design. For the FEM calculations, an electrostatic model described in Cartesian coordinates was used.
The geometries consist of the sample, with the measured resistivity value,
electrolyte, with a measured resistivity value and an isolating cell material.
Only the potential and current density distributions based on different geometries and voltages were considered in the calculation, disregarding all
17
chemical and electrochemical effects. The results from these calculations
are rough approximations as the potential drop across the Helmholz-layer
close to the Si-surface will normally be quite large. This probably results in
an overestimate of the significance of the placements of the electrodes and
of the cell geometry, however, the trends shown should still be valid.
Figure 2.8a) shows the current density distribution at a depth of 20 or 50 m
in the sample from the center out to the side of the etched area. The top
plot shows a comparison between the two cell geometries. There is a clear
inhomogeneity of the current density when the wide cell is used. In this case
the modeled samples had back-contacts covering the whole back-side and
the current density was calculated for a depth of 20 m. The middle plot
shows a comparison between two samples having back contacts with the
same size as the front opening. One of the samples has a 0.5 mm opening in
the center of the back contact to make possible interference measurements as
will be described in Sec. 4. Current densities were calculated for a depth of
50 m. It is clear that the influence of the back contact opening is significant
at this depth. The selected opening size is the minimum obtainable due to
mechanical constraints in the fabrication of the back plate.
The bottom plot shows a comparison of two samples with wide and narrow
back contacts, i.e. the back contact has the same size as the front opening.
The homogeneity of the current density is significantly better in the case of
the narrow back contact. The current density distribution close to the edge
of the electrolyte in the two cases may also be seen in the cross-sections
shown in Fig. 2.9. In the case of the narrow back contact, the current
density distribution will be more homogeneous with depth also. As can be
seen in Fig. 2.9, the current spreads out more towards the back contact in
the wide contact case. There will be a certain under-etching due to the
spreading of current in the sample. The effect of this is a slight decrease in
current density with depth which may be compensated for by an increase
in the current with time.
2.5
To make optical filters where the filter characteristics are dependent on the
position on the filter, a voltage was set up between two contacts, 1.2 cm
apart, on the back side of the sample. This resulted in a position dependent
current density during etching with in turn gave a change in refractive index
modulation with depth at different positions. Some results on etched filter
structures using this effect are presented in Paper V. The etch cell used
for this is schematically shown in Fig. 2.10 and is basically the same as in
Fig. 2.6 with a difference in the back contact. Two thin Cu-sheets with a
spacing of about 12 mm were used instead of the single Cu-plate. A FEM
simulation of the current density in the sample at a depth of 2 m for a
typical set of parameters is shown in Fig. 2.11. The model used is similar
18
A
24
22
20
C
24
22
20
C
24
22
20
Position (mm)
a)
b)
Figure 2.8: a) A comparison of current density profiles is shown for different etch cell and back contact geometries from FEM calculations of a 2D
electrostatic model in Femlab. The top figure shows profiles calculated at a
substrate depth of 50 m while the two figures below are calculated at 20 m.
The profiles are plotted from the center of the etch area to the edge of the
reservoir. The top figure shows a comparison between a cell with a funnel
shaped reservoir and a back contact wider than the electrolyte contact area
(curve A, the geometry to the right) with a tall and narrow reservoir geometry with a back contact the same size as the front contact area (curve B,
the geometry to the left). The middle figure shows a comparison of two profiles obtained with the narrow geometry with the same narrow back contact
(curve C) save for a 0.5 mm opening for in situ IR reflectance measurements
(curve D). The bottom figure shows a comparison between a narrow (curve
C) vs. wide (curve E) back contact with the tall reservoir geometry. Figure
b) shows the cell geometries.
to that in Sec. 2.4.1 The potential difference across the sample in this case
is 0.2 V while the potential difference between the back contact with the
highest potential and the Pt-cathode is 2 V. The conductivities used for the
materials in the simulation are realistic, although not necessarily measured.
19
2
Reservoir wall
Electrolyte reservoir
Electrolyte reservoir
Reservoir wall
1
1
2
Substrate
3
4
2
Substrate
3
4
5
5.4
0
1
Backcontact
5.2
4.8
4.6
5.4
a)
Backcontact
5.2
4.8
4.6
b)
Figure 2.9: Screen shots from a Femlab calculation showing the current
density distribution and direction for the narrow (a) and wide (b) back contact with a narrow electrolyte reservoir. The effect of the back contact geometry is larger towards the edges of the electrolyte contact area and towards
the back contact.
30
Current density (a.u.)
Figure 2.11:
The current density profile calculated
20 m into the sample from
the electrolyte-sample interface. The two back contacts
are biased to 2 and 1.8 V while
the Pt electrode is grounded.
The calculation is made with
a 2D model in Femlab. In
this calculation the electrolyte
is assumed to be a conducting
material and the electrochemistry is disregarded.
25
20
15
10
-5
-4
-3
-2
-1
Position (mm)
Chapter 3
Thin-film calculations
For many years multilayer dielectric thin film optical interference filters
have been used [50]. Obtaining interference effects by using layers of optical
thickness in the order of the wavelength of the light of interest have been
exploited in many applications. In this thesis multilayer and inhomogeneous
structures in PS are used to some extent to control light. Interference
effects with electromagnetic (EM) fields are obtained only with very specific
structures. Layer optical and physical thicknesses or layer refractive index
modulation periods must be in the order of the wavelength at which one
wants interference. In the visible and near infrared wavelength regions this
translates to structures with elements down to below 100 nm with strict
demands on size accuracy and definition. It is very important to understand
the optical properties of the material, which in this case is PS, to be able
to design such interference systems. In the first section the correspondence
between the porous structure of PS and refractive index will be discussed.
As the correspondence between PS fabrication parameters and obtained
structures is not trivial, a detailed knowledge of the intended structures and
their optical response is needed to be able to understand the actual response
from fabricated structures. The background for calculation of interference
filter responses will be presented in Sec. 3.2. In addition to the most basic multilayer optical system a few physical effects are taken into account.
These include dispersion, absorption and interface roughness. The incorporation of interface roughness into optical response calculations is described
in Sec. 3.3. This will be used both in optical filter calculations as well as
in the analysis of in situ laser reflectance measurements described in detail
in Chap. 4. In the last section, different multilayer and inhomogeneous
refractive index optical filter structures will be presented together with
calculations of optical responses based on the methods described. These
structures are experimentally realized and presented in Chap. 5.
21
22
3.1
A porous medium will exhibit different optical properties than the same
material in bulk. If the typical feature sizes (e.g. pore size) are much smaller
than the wavelengths of the incident electromagnetic field, the field in the
porous medium encounters an effective dielectric function. This effective
dielectric function is dependent on the dielectric functions of both the bulk
material and the filling material (e.g. air) in a ratio controlled by, amongst
other parameters, the porosity.
The theory describing the dielectric function of the mixed media is referred to as effective medium theory. There are several prominent effective
medium formulas, e.g. Bergman [51], Maxwell-Garnett [52], Looyenga [53]
and Bruggeman [54]. The main difference between these formulas lies in
how the microtopology of the pores are taken into account. The optical
response of a porous medium will change with the degree of connectedness (percolation strength) of the network and the sizes of the segments of
material left in the medium. The dependence on microtopology makes the
problem of finding a correspondence between porosity and effective dielectric function non-trivial. Maxwell-Garnett, Looyenga and Bruggeman all
assume certain microtopologies resulting in a more or less limited validity
when considering the effective dielectric function of PS as the microtopology is greatly dependent on formation parameters. An example of this is
reported by Setzu et al. [47] where the same porosity obtained by different
formation parameters gives significantly different refractive indexes.
The Bergman formula is general and takes into account the microtopology
by a spectral density function, however, this function is usually not known
and must be expressed specifically for all the different microtopologies of
different PS films, hence this approach is quite involved. The use of effective medium theory to describe the optical properties of PS is extensively
discussed by Thei in Ref. [55, 56]. Figure 3.1 is taken from [56] and shows
a comparison of several different effective medium formulas.
23
is the Bruggeman formula, often referred to as the Effective Medium Approximation (EMA). The popularity of the EMA is based on a paper by
Aspnes [57] where spectroscopic ellipsometry measurements were performed
on rough amorphous Si. The model describing the roughness which gave
the best fit was the EMA. The EMA is also used in this thesis for all refractive index-porosity calculations. The samples presented in this thesis
should have a narrow range of topologies caused by a narrow range of substrate resistivities and a relatively narrow range of HF concentrations in the
electrolyte. Any error introduced by the choice of the EMA as an effective
medium formula will then be systematic and may be easily adjusted for.
The Bruggeman formula is given by:
p
M
M
eff
eff
+ (1 p)
=0
+ 2 eff
+ 2 eff
(3.1)
where p is the porosity, , M and eff are the dielectric functions of the
embedded material (Si), the host material (air/vacuum) and the effective
medium (PS) respectively.
By using the EMA for all porosities and PS structures, it is clear from the
preceding discussion that for most situations there will be some error in the
calculation due to different microtopologies. From the literature it seems
that this error is small and, in most cases, tolerable for the applications and
measurements in mind. Some papers [58, 59] suggest that the Looyenga formula better describes the dielectric function of highly porous meso-porous
PS. We may calculate the effect of using different effective medium formulas
on the optical response of a simple Bragg reflector. We see from Fig. 3.1
that there is a maximum difference in refractive index of about 0.08 for
80 % porosity, or equivalently a difference in porosity of about 4.5 % for a
refractive index of 1.3. Given a Bragg reflector designed for a reflection
band around a wavelength of 600 nm, the change introduced by going from
the Bruggeman to the Looyenga effective medium formula will result in a
shift of the reflection band (given unchanging layer thicknesses) of about
40 nm. This error may be critical for some types of applications, hence the
choice of model, and possible corrections, must be kept in mind.
If we use a complex, frequency dependent dielectric function (dispersive
media), (), the absorption in the material will be taken into account.
The effective medium models may be used with these complex values. The
complex dielectric function is defined in the following as well as its relation
to the complex refractive index, n
(). We assume for all calculations that
the materials are non-magnetic, i.e. the magnetic permeability, , equals 1
(from [60]):
() = r () + i i ()
n
() = n() + ik()
n
() =
()
(3.2)
24
Solving the equation for n
analytically is not necessary. Note that the frequency dependence will be considered implicit in the following discussion.
The real refractive index, n, and the extinction coefficient, k, are usually
referred to as the optical constants of a material. By solving the Maxwell
equation for a plane wave using the complex dielectric function and considering the field intensity, I, the attenuation of the field as it is transported
in a medium is evident:
4k
0
I = I0 ez
absorption coefficient
(3.3)
7
0.4
6
0.3
0.2
0.1
0.0
3
0.4
0.6
0.8
Wavelength ( m)
10
0.5
Figure 3.2:
This shows
the tabulated data from Palik [61] of the refractive index (blue line - left axis) and
the extinction coefficient (red
line - right axis) used in all
reflectance and transmittance
calculations.
25
3.2
Reflectance calculation
Two equivalent ways of mathematically defining an optical multilayer system will be presented here. Both of these methodologies will be described
and used as they have different advantages. For these calculations we will
assume that the material is homogeneous for each layer, i.e. the refractive
index is constant and identical in all directions: n
(x, y, z) = n
, and that the
planes are parallel.
3.2.1
Characteristic matrix
i
sin (k0 ni di cos i )
cos (k0 ni di cos i )
pi
.
(3.5)
Mi (di ) =
Here pi = (i /i ) cos i and k0 = 2/0 , where 0 is the incident wavelength in vacuum, and di is the layer thickness. To obtain the same characteristic matrix in the case of a transverse magnetic (TM) field (p-polarized),
qi = (i /i ) cos is used instead of pi . With the components of the total
characteristic matrix of the stack denoted by
m11 m12
,
M=
(3.6)
m21 m22
the reflection and transmission coefficients of the system are then given by
r =
,
(3.7)
t =
2p1
(m11 + m12 pl ) p1 + (m21 + m22 pl )
(3.8)
26
The reflectivity and transmissivity are given by
R = |r|2 ,
pl 2
T =
|t|
,
p1
(3.9)
where p1 and pl are for the first and last layers, respectively. A sketch of the
described layered system with notation for both the characteristic matrix
approach and the admittance matrix approach described below is shown in
Fig. 3.3
3.2.2
Admittance matrix
H tR
H tL
=
,
E tR
E tL
(3.10)
27
for the field going left. For each polarization this becomes
Ys = n cos
n
Yp =
.
cos
(3.11)
(3.12)
where the prime denotes that the quantities are on the right side of the
boundary. This may be written in matrix form, relating the field amplitudes
to the right and the left of the interface;
Vi1
E Ri
E
= Vi Ri ,
E Li
E Li
with
Vi =
1
1
,
Yi Yi
Vi1 =
(3.13)
1 Yi1
.
1 Yi1
(3.14)
1
Wi1/i = Vi1
Vi .
(3.15)
The matrix Wij is known as the refractive matrix. In this case we work
from the rightmost layer toward the left, although the incident light will
usually come from the left. However, the expressions are general and light
may come from the left and/or the right. By finding the Fresnel coefficients
from the admittance we get
Wi1/i =
ci1/i 1
rLi
,
r
t
t
tRi
i
Ri Li rRi rLi
(3.16)
where rLi and tLi are the reflection and transmission Fresnel coefficients,
respectively, of the ith interface with the incident field coming in from the
right. ci1/i is chosen for the correct polarization and is given by
ci1/i =
cos i1 / cos i
1
for p-polarization
.
for s-polarization
(3.17)
eii
0
0 eii
E R,i+1
E L,i+1
= Ui
E R,i+1
E L,i+1
(3.18)
28
Ui is called the phase or propagation matrix. The phase-shift, i , is given
by
2
ni di cos i
(3.19)
i =
0
To obtain the total field transformation of a layered system, with layers
from 1 to i, the matrices are multiplied to give, in the general case:
E R1
E L1
E R,i+1
,
E L,i+1
E R,i+1
E L,i+1
(3.20)
s11 s12
.
s21 s22
(3.21)
E Li
,
E Ri
tRi =
E Ri
,
E Ri
(3.22)
rLi
E Ri
,
=
E Li
tLi
E Li
=
.
E Li
s21
,
s11
tR
1
,
s11
(3.23)
rL
s12
= ,
s11
tL
det S
=
.
s11
3.3
Roughness calculation
Roughness in the context of this thesis is based on the height function of the
interface of interest, h(x, y). We assume that the xy plane is parallel to the
29
sample surface. This function describes the deviation of the interface from
a perfectly flat surface. It is more practical to work with a single number
for the roughness than the height function, so we define an interface height
average. It is assumed that the interface height function is isotropic, hence
it is enough to find the average in one direction, which may then be given
by
1 L
h(x)dx.
(3.24)
a=
L 0
The value used for roughness characterization is usually the root mean
square (rms) value of the height function, , given by:
1
L
(3.25)
L is a characteristic distance, e.g. scan length. This length is quite important as average values (large radius bending of surface) and rms values
(different spatial period of roughness/fluctuations) may change considerably with the scale of L. Roughness may be measured by several methods,
i.e. stylus profilometry, white light interferometry or optical scattering (diffuse and specular). Pertaining to this thesis, the consequences of roughness
on the optical field is of most importance, both for determining roughness
and also for evaluating the effect of roughness. A method to measure the
roughness evolution in situ during PS etching, based on the theory described
here, has been developed and will be presented in detail in Chapter 4 and
in Paper I.
3.3.1
Davies-Bennett theory
30
area than covered by the correlation length of an interface, the phase fluctuations will be minimal. On the other hand, if the area measured is larger
than that covered by the interface correlation length, the phase fluctuations
will be stronger. A schematic representation of the principle is shown in
Fig. 3.4
The phase difference between two different parts of the wavefront reflected
from a rough surface where there is a height difference, h, between the
positions of the surface the wave strikes, in a medium with refractive index
n, is given by
2
2nh cos .
(3.26)
r =
0
For the average phase difference over the surface area of interest, h is
replaced by the rms height difference, , as defined in Eq. 3.25. In Eq. 3.26
it is assumed that h < , from this follows that
. For the EM
field will be fully incoherent and the measured intensity follows a different
relation to than for a partially coherent field. In addition, a Gaussian
interface height distribution is assumed and a plane wave EM field.
The phase difference results in a modification of the Fresnel reflection coefficient of
1 2
r s = r 0 e 2 r ,
(3.27)
following [67, 66]. Here r0 is the reflection coefficient of a smooth surface.
In the case of transmission from medium 1 to 2 the phase difference and
modified Fresnel transmission coefficient are
t =
2
h (n2 cos 2 n1 cos 1 ) ,
0
(3.28)
31
1 2
ts = t0 e 2 t .
(3.29)
This results in the following adjusted Fresnel coefficients which will be used
in simulations with the proposed methods in 3.2:
0
0
rRi = rRi
exp 2 (2i ni1 cos i1 /0 )2 = rRi
0
0
rLi = rLi
exp 2 (2i ni cos i /0 )2 = rLi
tRi = t0Ri exp 1/2 (2i /0 )2 (ni cos i ni1 cos i1 )2 = t0Ri
tLi = t0Li exp 1/2 (2i /0 )2 (ni1 cos i1 ni cos i )2 = t0Li
(3.30)
The optical power lost in the specular direction (both reflected and transmitted) because of roughness is regained in the diffuse scattering, i.e. scattering in other directions than specular. For the calculations done in this
thesis, only specular reflection and transmission is considered.
As the roughness, as discussed above, will be incorporated into a calculation of reflectance/transmittance of multilayer structures, the correlation
of interface profiles from one layer to another must be considered. It is
reasonable to assume, due to the nature of PS fabrication, that there will
be a certain degree of correlation of roughness between layers. This would
somewhat decrease the effect of interface roughness compared to fully uncorrelated interface profiles. However, the refractive index contrast between
layers is smaller than between the layer stack and the substrate which results
in a greater effect of roughness at the PS-substrate interface than elsewhere.
This is especially true for rugate filters where the refractive index contrast
between layers is very small. Due to the simplicity of incorporation and the
relative small difference the two options should make on the outcome, only
the effect of fully uncorrelated interface roughness scattering/incoherence
is incorporated into the calculations.
3.4
3.4.1
(3.31)
with one layer having a low refractive index, nL , and the other layer having
a high refractive index, nH . A compact way of denoting the design is
HLHLHL . . . HL = (HL)i , with i being the number of pairs. The thickness
of the layer is d while 0 is the wavelength of the incident EM wave in
32
vacuum for which the maximum reflection will occur (design wavelength).
Plane waves are assumed. This condition results in the reflected EM wave
at the design wavelength constructively interfering in each layer and one
may reflect up to 100 % of the incident energy. The reflectance of a Bragg
reflector designed for a maximum reflectance at a wavelength of 1550 nm
is shown in Fig. 3.5. In this case the stack consists of 10 layer pairs. The
material used is 50 % and 80 % porosity PS. All the calculations in the
following are based on the discussion in Sec. 3.2.
1.0
1.0
0.8
0.8
Reflectance (abs)
Reflectance (abs)
By increasing the contrast, i.e. the difference between the high and low refractive index, one may widen the reflected wavelength band. By increasing
the number of pairs in the stack, the reflection band edges will be sharper,
as shown in Fig. 3.6. The stack is similar to that used in Fig. 3.5 except
a stack of 20 layer pairs was used. In both cases interface roughness, absorption and dispersion were disregarded. In most applications the ideal
reflector would be one where there is no reflection outside the band and
100 % reflection within the band, i.e. a square reflection band.
0.6
0.4
0.2
0.6
0.4
0.2
0.0
0.0
1000
2000
3000
4000
Wavelength (nm)
1000
2000
3000
4000
Wavelength (nm)
Figure 3.6: The design is comparable to Fig. 3.5 except it consists of 20 layer pairs. Note the
sharper band edges due to the increased number of layers. Absorption, dispersion and interface roughness are not taken into account.
The layers may be combined in any way, and many types of filters are
possible for many different uses. Anti-reflection coatings, polarizers, beamsplitters and band-pass filters may all be made by use of multilayer thin
films. A Fabry-Perot band-pass filter is obtained if a spacer layer is introduced between two mirrored Bragg stacks. An example of this is the
structure (HL)i (LH)j where the spacer consists of two L layers. This will
give a sharp resonance peak in the transmission spectra at the design wavelength.
33
3.4.2
Inhomogeneous layers
The types of optical multilayer filters mentioned above are based on discrete layers with homogenous refractive indexes and uniform thicknesses.
It is possible, however, to generalize this to layers of controlled inhomogeneous refractive index (both in depth and laterally). The process of continuously varying the refractive index of a layer with depth has until recently
been quite difficult. There are a few systems in which this is possible,
such as plasma enhanced chemical vapor deposition of silicon oxynitride
(SiOx Ny ) [69] where the refractive index changes with the stoichiometry
which may be controlled with the ratio of formation gases. Other methods include magnetron sputtering deposition with variable control of the
formation gases [70] or deposition of quasi-inhomogeneous layers consisting of many, very thin, homogeneous layers [71], glancing angle deposition
to fabricate porous structures of depth dependent porosity [72, 73], and
codeposition of two oxides by evaporation [74]. However these systems are
usually relatively expensive and complex and have a limited range of refractive indexes available.
With PS, a practically obtainable refractive index contrast in the NIR of
about n = 3.0 1.2 = 1.8 is possible and the method is comparably
straightforward and inexpensive. This flexibility of PS enables both novel
optical elements in intimate connection to Si technology as well as a testing
bed for novel stand-alone optical filters. There are several arguments for
fabricating optical elements with inhomogeneous refractive index. By avoiding internal interfaces, the structure becomes mechanically more stable, e.g.
against scratches, and the probability of delamination is reduced. Fewer interfaces also result in a decrease in the scattering of the EM field within the
structure. One typical use of inhomogeneous refractive index layers is in
anti-reflection coatings. Another possibility is varying the refractive index
sinusoidally with layer depth. This results in a rugate reflection filter. The
refractive index profile is given by
n (z) = na +
np
sin
2
2z
+ ,
na d
(3.32)
with the design wavelength equal to 2na d. Here na is the average of the
maximum (nH ) and minimum (nL ) refractive index used in the layer, np is
the peak-to-peak difference between minimum and maximum values and d
is the physical thickness period of the refractive index sine profile, z is the
depth in the layer and is the phase. The resulting spectral characteristics
of the filter are quite similar to a Bragg reflector. However, in the case of
small np , there are no higher order harmonics (at normal incidence) and by
further exploiting a continuous variation in refractive index one may reduce
sidebands as well. To avoid higher harmonics in the reflectance spectrum for
larger modulations in the refractive index profile, Eq. 3.32 must be modified
so that the exponential of the sinusoidally modulated refractive indexes are
34
obtained [29, 75]:
n (z) = exp
ln nH + ln nL ln nH ln nL
+
sin
2
2
2z
+
na d
(3.33)
This ensures that the optical thickness of the positive sine half is not larger
than the negative sine half, hence there will a be perfect match between
the structure and the incident EM wave. By using Eq. 3.32 with relatively
large np there will be higher order harmonics in the reflectance/transmission
spectrum. Similar to Bragg reflectors the reflection band of a rugate filter
will widen with a greater refractive index contrast (nL to nH ) and the
number of periods gives the quality of the reflection band.
By adding an index matching region between the filter and the main interfaces of air and substrate, the base reflectance is reduced. By apodizing,
i.e. adding a windowing function to the refractive index profile, the sidebands are reduced. The index matching consists of a smooth transition in
the refractive index between two media, i.e. from air to the filter medium,
and from the filter to the substrate, where the filter refractive index is considered to be na . There are many possible gradient functions, but one
which has been shown to give good results is the 5th order polynomial
(quintic) [75, 76]:
n = nL + (nH nL )(10t3 15t4 + 6t5 ),
t [0, 1],
(3.34)
1.0
1.0
0.8
0.8
Reflectance (abs)
Reflectance (abs)
35
0.6
0.4
0.2
0.6
0.4
0.2
0.0
0.0
1000
2000
3000
4000
Wavelength (nm)
1000
2000
3000
4000
Wavelength (nm)
and substrate. This is shown in Fig. 3.10. Narrow filters are obtained by
decreasing the refractive index range. This is shown in Fig. 3.11 which is
based on the same parameters as in Fig. 3.10 but with a refractive index
range corresponding to 50 % to 60 % porosity.
The refractive index profile of rugate filters lends itself to combinations of
different kinds. It is possible to put any number of profiles with different design wavelength in series such that the reflection spectrum will show
corresponding reflection bands [74, 75]. An example of a multiband rugate filter is shown in Fig. 3.12. To obtain this reflectance spectrum three
partly overlapping refractive index profiles are used. Each profile consists
of 40 periods and varies between refractive indexes corresponding to 75 %
and 78 % porosity. The profiles are designed for maximum reflectance at
1000 nm, 2000 nm, and 3000 nm. The calculated spectrum is for the ideal
case with matching outer medium, filter, and substrate and with no absorption, roughness, or dispersion.
The wavelengths in a multiband rugate filter may be so close to each
other that the bands fully or partly overlap resulting in one wide reflection band [77] or a narrow gap transmission band [78], respectively. The
1.0
1.0
0.8
0.8
Reflectance (abs)
Reflectance (abs)
36
0.6
0.4
0.2
0.6
0.4
0.2
0.0
0.0
1000
2000
3000
4000
Wavelength (nm)
Figure 3.9: The calculated reflectance spectrum of a more realistic situation with a Si-substrate and
air as outer medium. The parameters are otherwise similar to those
used in Fig. 3.7. The sideband reflectance is much higher in this case
due to the sharp transitions in refractive index from air to filter and
filter to substrate. Dispersion is
taken into account.
1000
2000
3000
4000
Wavelength (nm)
profiles may also be in parallel or partly shifted relative to each other, i.e.
the sine modulation part of the profiles are added to each other keeping
na as a common base. This results in a thinner filter with the same functionality. However, the total profile is limited by the range of refractive
indexes available. This may be amended by using a partial overlap/shift
of apodized profiles such that the sum of refractive indexes at any position
never goes outside the available range. An example of this is shown in
Fig. 3.13 where two profiles partly overlap. Another possibility of making
thin filters with multiple bands is to add the profiles so that the refractive index range available is exceeded and the resulting profile clipped to
fit within the constraints. How much clipping can be accepted depends on
the tolerance limit of the filter application as this approach will introduce
sidebands and noise in the reflection/transmission spectrum.
With the above in mind, a possible design for a narrow transmission band
rugate filter was designed for use as a graded filter in conjunction with a
strip Schottky detector on the back side of the substrate. This device will be
discussed more in Chaps. 5 and 6. In the calculation of the reflectance spectrum of this filter, shown in Fig. 3.14, both absorption and dispersion were
included. Air and Si-substrate media were used. The filter was designed
with four reflection bands: at 950 nm, 1130 nm, 1450 nm, and 1800 nm,
each refractive index profile consisting of 30 periods with the profiles partly
1.0
1.0
0.8
0.8
Reflectance (abs)
Reflectance (abs)
37
0.6
0.4
0.2
0.6
0.4
0.2
0.0
0.0
1000
2000
3000
4000
Wavelength (nm)
Figure 3.11:
Calculated reflectance spectrum for a narrow rugate reflector similar to the case in
Fig. 3.10 but a variation in refractive index corresponding to porosities between 50 % and 60 %. Due
to absorption and apodization it is
difficult to get unit reflectance with
narrow band rugate filters as in this
case. Dispersion is taken into account.
1000
2000
3000
4000
Wavelength (nm)
Figure 3.12:
Calculated reflectance spectrum of a three band,
narrow band rugate reflector. The
reflector is designed for maximum
reflectance at 1000 nm, 2000 nm and
3000 nm. Each band correspond to
a section in the refractive index profile consisting of 40 periods with
porosities between 75 % and 78 %
nominally.
The total calculated
porosity profile will have a larger
range as the different sections in
the profile will partly or fully overlap. The situation is the same as in
Fig. 3.7 with outer media and substrate being identical, no apodization used, and absorption and dispersion not taken into account.
overlapping. Each profile was apodized with a quintic function and the refractive indexes corresponded to a porosity range of 50 % to 80 %. A quintic
index matching was used with the maximum and minimum refractive index
corresponding to what is obtainable for the etch setup presented earlier in
Sec. 2.4.1. The order of the profiles with respect to the surface is significant
for the reflectance due to the absorption. The physically thinnest profile
should be at the top resulting in more equal significance of each profile compared to the opposite case. In Fig. 3.14 absorption values from Palik [61]
are used, according to Sec. 3.1, for curve C (red) and a 10 % fraction of these
to see the incremental effect are used for curve B (green). The transmitted
spectrum in the latter case is shown in the inset. A narrow band centered
at 1290 nm is clearly visible.
For all the discussed filter types, the reflection/transmission spectrum is
dependent on the incident angle. The spectral features for Bragg reflectors,
Fabry-Perot filters and rugate filters will shift towards shorter wavelengths
38
1.9
Refractive index
1.8
1.7
1.6
1.5
1.4
1.3
1.2
1.1
0
10
15
20
25
Thickness ( m)
with larger incident angle, i.e. blue-shift. This places restrictions on the
usable wavelength range and angular range of a filter. It is possible, however, to device reflection filters having a constant band of reflection for all
angles, as in the PS based omni-directional mirror of Bruyant et al. [79].
39
1.0
0.6
0.8
C
0.6
Transmittance (abs)
Reflectance (abs)
0.4
0.2
B'
0.4
0.0
1000
2000
3000
4000
Wavelength (nm)
0.2
0.0
1000
2000
3000
4000
Wavelength (nm)
Chapter 4
In situ interferometry
experiment
For the efficient and successful fabrication of optical filters it is essential to
have good control of properties of the deposited material. The most important properties are refractive index (also density/porosity, homogeneity
and absorption) and interface roughness when discrete layers are deposited.
When making optical elements in PS, especially filters, this translates to
having a good (instantaneous) control of etch rate, porosity, roughness and,
to some extent, microtopology as discussed in Sec. 3.1.
Etching of PS is a dynamical exercise. The parameters of the system change
during etching which necessitates an in situ monitoring of the most important parameters so the changes may be counteracted, or enable a real-time
feedback to the etching process. The feedback method is often used to obtain optimum results in other conventional optical material systems [69].
Often feedback in this situation is based on interference effects with monoor poly-chromatic coherent light. Interference at single wavelengths and ellipsometry give very good optical density/physical thickness resolution and,
in the case of interference, is easily implemented in a deposition system.
In the PS fabrication setup already described, the sample is in contact
with a liquid electrolyte on one side. In this case it is easier to set up a
system where the optical probing is done from the opposite side than to
have the optics on the electrolyte side which may be closed off or corrosive.
This presupposes a transparent sample which is obtained when using an
IR laser with a wavelength longer than that corresponding to the silicon
bandgap energy (laser
1130 nm). With this setup the IR laser beam
will encounter several interfaces, i.e. back side, PS-substrate, and frontside, with the resulting partially reflected beams phase shifted thus giving
rise to interference in the measured reflected beam. Similar systems are
described by Steinsland et al. [80] for tetramethyl ammonium hydroxide
etching of silicon and by Thonissen et al. [81] and Gaburro et al. [82] for
PS fabrication monitoring.
41
42
The change in interference conditions during PS etching results in a signal
containing information on relative interface movement and the different
refractive indexes. By performing frequency analysis on this signal it is
possible to obtain several key parameters of a PS film during formation.
The parameters depend on a few assumptions about the formation process
and substrate material. Given the correctness of these assumptions, we
obtain the refractive index and porosity profile of the PS film, and the
time dependent etch rate. From this we can calculate the refractive index
profile of the film in air, average porosity and refractive index, PS-substrate
interface roughness, film thickness and instantaneous valence values during
formation.
Most of the work done with the setups presented here is described in Papers I-III, which are included at the end of this chapter. These include
background theory, analysis, experimental setup, and some results. Some
details not discussed in the papers are included in the following sections.
4.1
Setup
4.1.1
Usage
The goal of this work was to design an in situ measurement setup which
would fit in with a standard etch cell as detailed in Sec. 2.4. As the analysis
of the data obtained from the setup gives all the critical parameters without
further processing of the sample, a quick, non-destructive measurement of
the standard calibration parameters for constant current conditions is made
possible. These data, porosity and etch rate versus time and depth at
constant current, may be used for controlling the etch current to obtain
layers of constant porosity. The setups used are schematically shown in
Fig. 4.1. The main descriptions of the setups are given in Paper I and some
of the concepts in the following discussion are defined in this paper.
In addition to the mentioned material properties, the evolution of interface
roughness is also monitored by measuring the introduced partial incoherence
in the reflected laser beam. The incoherence results in a decreasing signal
intensity with increasing roughness. This is needed to better understand
the optical properties of the interfaces encountered in the multilayer PS
optical filters. By changing formation parameters, an optimal parameter
range may be found for filter fabrication with regards to interface roughness.
This understanding is crucial for optimizing the quality of the PS optical
elements.
It is possible to test the calibration data by calculating the needed current profile for a constant porosity layer and measure the actual obtained
porosity with the setup in situ. An example of this will be given later.
One challenge when using calibration data for a multilayer or inhomoge-
43
neous refractive index structure, is that the conditions at the pore front
depend on the history of the etch. The conditions during multilayer
etching is different from those present at the single layer calibration etch.
This effect may be seen by etching a single layer and changing the current
after a certain time. Comparing the porosity obtained after the current
change with porosities obtained by etching a layer at the same current as
this from the start, a shift in porosity is observed.
The measurement of the back side reflected signal will also give information on any change in refractive index or layer thickness of the porous
layer after the etch current is turned off. As the layer thickness is unlikely
to change, one must assume all change comes from the refractive index.
This only applies if all else in the setup is constant, i.e. no thermal expansion/contraction or movement of the sample relative to the optics. A change
in refractive index after the electrochemical etch has stopped may be caused
by a purely chemical etching of the porous structure. This chemical etching
of silicon by HF is very slow for the bulk material, but due to the large
internal surface area of PS the relative time dependent change in Si volume
may be significant, thus will give a measurable change in refractive index
with the in situ reflectance setup. However, there are several possible explanations to the change in refractive index after current is turned off, such
as relaxation of the structure due to decreasing temperature, electrolyte refractive index change due to diffusion, and oxidation of the structure in the
electrolyte. Similar measurements have been performed by Navarro-Urrios
et al. [83].
It is possible to obtain a complete calibration curve, i.e. porosity versus
current density, by performing a current sweep during etching while measuring the interference signal. However, this will only be an approximation
as it does not take into account the time dependent change in etching conditions. For many applications this approach may prove efficient and accurate
enough. This technique is briefly discussed in Ref. [82].
One possible improvement of the current setup is to extend the data analysis
capabilities to real-time operation, hence facilitate the possibility of feedback to the PS formation. With the knowledge of an approximate range
for the starting porosity value, this technique could prove quite accurate.
4.1.2
Laser
Two laser systems were used. For the free space setup a 1310 nm wavelength laser diode (LD), LD1087, with drive electronics, LDP201, from
Power Technolgy was used. The diode was in a housing containing the
collimating optics. The output optical power was 8 mW and the beam diameter was about 2 mm with a divergence of about 10 . By monitoring
the beam the quality of the measurement would increase by increasing the
signal-to-noise ratio. However, this was not necessary for obtaining satis-
44
Figure 4.1: Schematic of the different etch setups. The setup denoted A
is based on free space beam transmission. The laser diode is denoted LD.
The schematics of both setup A and B shows a cross-section of the etch
bath. Setup B shows the wide beam optical fiber variant. MM denotes the
multimode optical fiber and SM denotes single mode optical fiber. A coupler
is shown; this splits the fiber and guides the reflected signal from the sample
to the detector. A graded index lens, as indicated, is used to collimate and
collect the light close to the sample. The third setup, C, shows a fiber variant
where no collimating lens is used, but where the fiber is close enough to the
sample back side (< 500 m) to collect a significant amount of the light
reflected from the sample. This results in a probed area about the size of
the fiber core cross section.
factory results.
For the optical fiber based setup a pigtailed laser system assembled by
Thorlabs was used. The diode used was of type ML976H11F from Mitsubishi, an InGaAsP, multiple quantum well, distributed feedback diode
laser. The output wavelength was 1550 nm with an optical power of 2 mW
when coupled to the single mode fiber.
As the measurement of the interference effect in the reflected signal is dependent on a constant phase of the input beam, and also, for the calculation
of roughness, a constant amplitude, a well regulated system is needed. The
diode lasers are sensitive to temperature fluctuations and temporal effects.
To keep the output power and phase constant the diode needs to be stable
in a specific laser mode. A small temperature change may introduce a mode
jump which may give a change in phase and also a more noisy output as
there will be transient signals, both in the diode itself, and also in the regulation of the diode. Therefor careful regulation of temperature and output
power was needed. This was done using an ILX Lightwave Technologies
LDC 3722 laser diode driver controlling the current by feedback from a
45
monitoring diode in the laser package. The same driver regulated the diode
temperature with a thermo-electric element connected to the laser diode
housing with thermal paste. The diode lasing mode was very sensitive to
changes in temperature which resulted relatively often in unusable periods
in the measured data. This problem was improved by allowing the laser
warm up for a few hours before use.
The lasing in the diode is also very sensitive to back-reflected light from
the pigtailed fiber. To minimize back-reflectance an optical isolator was
connected to the fiber from the laser diode. By adding a second isolator
the stability of the LD seemingly improved further. Because of a small but
significant time delay between an amplitude change in the LD and an amplitude change in the signal reflected off the opposite end of the connected
fiber to the LD, instabilities in the regulation of the output-power may be
introduced. By introducing a light-valve so that the light is not reflected
back to the diode the problem of regulating the power may be reduced.
Even though back-reflectance is very small with two isolators, there may
still be enough to introduce a slight oscillation in the output power which
may explain one persisting artefact in the measurements. A long wavelength oscillation of equal or higher amplitude compared to the measured
interference oscillation was present in all measurements. One possible cause
may be instability in the laser, however, this was most likely ruled out by
monitoring the output signal by using a 2x2 coupler, with a monitoring detector and sample setup opposite the LD, and a reflectance signal detector
on the same side as the LD. This enabled the monitoring of the LD output concurrently with etching to see if the long wavelength oscillation was
present in the output. The conclusion was that the oscillation is introduced
by the etch setup arm of the 2x2 coupler.
46
4.1.3
Fiber
Due to availability, all connectors for the optical fibers were FC/PC. This
ensured a satisfactory coupling between fibers and relatively little interface
back reflection. Possibly, by using FC/APC couplers the back reflectance
seen at the LD might be reduced somewhat, resulting in a more stable LD
output.
Both multimode (MM) and single mode (SM) fiber were used in the setup.
As the fiber pigtailed to the LD was SM, it would be optimal to only
use SM fibers due to back reflectance at connectors. However, it proved
difficult to couple the light reflected from the sample back into the SM
fibers due to the small fiber core diameter and the small acceptance angle
or numerical aperture. Because of the larger numerical aperture and core
diameter a MM fiber was used on the sample side of the setup. The SM fiber
coupled to the LD was a standard SMF-28 type fiber with a core diameter
of 8.3 m, while the multimode fiber used had a 62.5 m diameter core.
Coupling between the two fiber types is not optimal. The core material
may be different leading to Fresnel reflection losses, while the MM fiber
will have many available, unfilled modes which makes the fiber sensitive to
movement/bending etc. as the beam from the SM fiber may change mode
as a response to external influences. To avoid these problems efforts were
made to keep all fibers as rigid and stable as possible by taping fibers and
fastening all measurement equipment to the same metal frame. However, a
short length of the fiber from the 2x1 coupler to the etch setup was free as
the etch setup was inside a flow box with the rest of the equipment outside.
One possible further remedy would be to fill all the modes of the MM fiber
by scrambling the single model from the SM fiber, spreading the beam over
several modes in the MM fiber.
The 2x1 coupler used was connected as shown in Fig. 4.1B. The split ratio
between the two split fibers was 50/50.
4.1.4
How the probe beam interacts with the sample is critical to the quality of
the measurement. The three different setups used, free-space, wide-beam,
and narrow-beam, all have different advantages which mainly depend on
how the beam couples to the sample.
In the case of the free-space setup, as shown in Fig. 4.1A, the probe
beam goes through air from the LD to the sample to the detector. The
beam propagation direction is controlled by mirrors, and to simplify the
setup the beam is at an angle of about 12 to the sample normal. With this
setup there will be no back reflectance to the laser diode, hence, the laser
should be quite stable. However, by traveling through air, the beam may
be distorted by dust particles, changes in air density, moisture content and
47
temperature, thus introducing noise in the measured signal. At the same
time the mechanical positioning of the LD, the detector and the sample
may drift slightly with time introducing amplitude shifts due to misalignment. The beam is easily moved and changed in size. This may be used to
probe different areas of the sample, even scan the whole sample area during
etching by swiping the beam with a high speed beam movement system,
e.g. bar-code scanner setup. By changing the beam size, the same setup
may probe roughness at different spatial scales, although the minimum obtainable beam diameter will be significantly larger than the effective beam
diameter obtained with the narrow beam fiber setup.
It is important for PS layer homogeneity, both in porosity and etch rate,
that the electrical potential in the Si-sample is evenly distributed. This
necessitates a good, and optimally shaped, back contact, even for highly
doped Si-samples. See Sec. 2.4.1 for a discussion on this. However, as the
contact used is non-transparent aluminum a hole must be etched to let
the beam through. Depending on the size of this hole, the homogeneity of
the PS film will be more or less affected. With the beam sizes obtainable
with the free-space setup this will always be a problem. An advantage of
the free-space setup compared with the fiber based setups is that all the
interference comes from interactions in the sample, as opposed to interference between fiber end and sample back side. This assumes that the LD
amplitude and phase is stable or oscillates very slowly compared to the
interference frequencies of interest.
In the wide-beam fiber setup the beam is transported to and from the sample by optical fiber, thus avoiding many of the potential noise sources of the
free-space setup. However, as already noted, movement, bending or temperature gradients may also introduce noise in the fiber, but nonetheless the
beam in the fiber setup is relatively stable. As in the free-space setup, there
is a compromise between beam size and optimal back contact. However in
the fiber case the beam is more rigid. As can be seen in Fig. 4.2, to obtain
a different beam a different graded index (GRIN) lens must be chosen and
fitted.
Once a lens has been fitted, the Cu-plate must be leveled so that the beam
reflected off the sample will be coupled back into the lens. This need only
be done once in a while as there is some reshaping of the materials used
in the sample holder when the sample is clamped down. This setup is
very compact and can be used as a standard addition to the etching of PS.
There are a couple of challenges however. There is always the danger of
leakage of electrolyte, so all parts of the etch setup should be HF resistant.
For the setup used, this means a possible degradation of the the long term
stability of the system as a non-HF-resistant GRIN lens was used as well
as an aluminum lens holder. There is also a possibility of interference
effects introduced in the signal from movement of the sample relative to the
lens, due to either temperature changes or bending of the sample during
etching [84]. This may be one explanation for the long period oscillation
48
observed in the measured reflectance.
By avoiding the GRIN lens setup and only using the cut fiber end, the
problem of PS homogeneity is reduced as only a very narrow hole in the Al
back contact is needed. In this setup a MM fiber is cut straight and placed
as close to the sample as possible. In the used setup a small hole of about
0.5 mm diameter was drilled in a Cu-contact plate in which the bare fiber
end was positioned, see Fig. 4.3.
Figure 4.2: A zoom-in on the coupling between fiber and sample for
the GRIN lens in situ setup. The
Cu-plate with a center hole used as
back contact is shown on top. The
GRIN lens is placed in an aluminum
holder which is shown. There are
screws for adjusting the plane of the
Cu-plate in the top flange of the Al
lens holder so back coupling from
the sample may be maximized. The
fiber is in contact with the end of
the GRIN lens for minimum backreflectance.
The fiber was pushed through as far as practical without leaving the end
free on the front side so it could break when the sample was positioned.
This gave a distance between fiber end and sample of < 0.5 mm. On the
other side of the Cu-plate the fiber was fastened with wax. This was stable
enough and also made it easy to fix and recut the fiber in case of damage or
wear to the end, e.g. by HF etching. This setup is able to measure the most
interesting parameters without interfering with layer homogeneity, and as
the case with the wide beam fiber setup, can be an integral part of the
etch setup. As the probed area of the sample in this setup has a diameter
49
roughly the same as the core diameter, assuming relatively flat interfaces,
the roughness measured may have spatial wavelengths smaller than the
probe diameter and thus will not be able to give a broad characterization
of the interface roughness of the sample. However, to get a good idea of
the roughness, several different beam sizes must be used. The alignment
of this setup is very robust as the fiber has a wide acceptance angle, which
means the fiber end may still capture much of the interference even though
it is somewhat tilted. To improve on back reflectance from the fiber end,
the end may be cut at a small angle. However, the results were satisfactory
with a straight cut.
4.1.5
Other equipment
For both the free-space and fiber based setups, the sensor used was a New
Focus 2011 detector with built in amplification. It was based on a InGaAs
PIN diode. This was connected to a multimeter, either a Keithley 199 DMM
or a HP 34970A. This again was connected to a computer. The electrolyte
temperature was measured with a Pt-based thermocouple for all calibration
measurements. Both temperature and signal was then logged by a LabView
program.
4.2
4.2.1
Data analysis
Chemical etching
In addition to etching at the pore tips there will also be some time dependent
etching of the PS layer as discussed briefly in Sec. 4.1.1. This results in a
gradient in the porosity with depth opposite to the gradient observed by
the in situ measurements. The chemical etching results in an increasing
porosity towards the surface which also will increase with time. The rate of
this etching will change with time as the structure of the pores change due
to etching. A possible result of this is show in Fig. 4.4 where the back side
reflected signal is measured both during etching as well as after etch current
is turned off. The point where the etch current is turned off can be seen
where the short period oscillation stops. A relatively slow oscillation is still
present indicating a continuing change in refractive index of the PS layer.
In the upper right plot the oscillation present in the signal after etch current
is turned off is isolated and the slow variations are filtered out. As can be
seen in the upper left plot of this figure, there is a very significant long
period oscillation present throughout the measurement. This oscillation is
most likely caused by thermal gradients, slow movement of the fiber close
to the sample or bending of the sample as discussed in Sec. 4.1.4.
It is possible that the end-of-etching oscillation is caused by these effects,
50
50
100
120
125
130
2.0
73.2
73.0
72.8
1.8
72.6
87.6s
410.6s
181.2s
Porosity, %
Time, min
1.04
1.02
1.6
1.00
0.98
1.4
2.0
2.0
1.8
1.8
1.6
1.6
100
110
120
130
Reflectance, a.u.
Reflectance, a.u.
39.4s
140
Time, min
however, the period and amplitude of the the most noticeable oscillation
during etching is quite different from the end-of-etch oscillation. There
are other plausible explanations to the latter besides the external effects
and chemical etching. During etching the current through the electrolyte
and the sample will drive up the temperature, possibly changing refractive
indexes of both the electrolyte and PS layer. When the etch current is
turned off this change will relax back to the values at ambient temperature.
The refractive index of electrolyte may also change due to diffusion. Sirich chemical species will diffuse out of the PS layer while HF will diffuse
51
towards the pore front. Assuming that the end-of-etch oscillation is only
due to chemical etching, the average porosity change in the layer is easily
calculated. The data shown in Fig. 4.4 is measured during formation of a
PS layer with an electrolyte containing 26 % HF at 30 mA/cm2 and at an
average temperature of 4.3 C. The final etched thickness is 207.6 m, the
sample was etched for 117 min, while the average porosity of the layer is
72.6 %. The optical thickness difference of the porous layer between two
adjacent oscillation peaks in the upper right plot of Fig. 4.4 corresponds to
half the incident wavelength:
nP S,t1 d nP S,t2 d = 0 /2
(4.1)
with nP S,t being the refractive index of the porous layer in the electrolyte
at a time t and d is the layer thickness. This gives a change in the average
refractive index of the layer of about 0.0037 for each period of the oscillation.
The resulting change in porosity is shown in the upper right plot in Fig. 4.4.
4.2.2
4.2.3
Frequency analysis
Short time Fourier transform (STFT) is used for analysis of the obtained
signal to give both temporal and frequency resolution. The details of the
analysis is discussed in Paper I. When selecting the parameters for the
STFT analysis, it is important to find an optimal balance giving the best
52
result. This is critical for the choice of window, both which window function
and what length it should have. As the partials frequencies are likely to
change with time, increasing a window length will work towards broadening
the observed peak, while at the same time the more periods of the partials
represented within the window, the better frequency resolution one should
get. These effects oppose each other, hence for a partial with only a slight
change in frequency with time, represented in Fig. 4.6 as curve B, the effect
of an increase in the number of oscillation periods overcomes the effect of
the frequency change within the window. On the other hand, for curve A
in Fig. 4.6 the frequency change within the window is too large, resulting in
the broadening of the peak width with an increasing window length. The
data in Fig. 4.6 shows the measured peak widths of the two main partials
of the in situ measured interference signal shown in the spectrogram of Fig.
5 in Paper I varying the window width. Curve A corresponds to the partial
marked 1 while curve B corresponds to the partial marked 2. As can be
seen in Fig. 4.6, for a decreasing window width under a certain threshold
value the peak width increases sharply. This is due to the lack of periods
present within the window, hence the determination of a frequency becomes
more uncertain.
The choice of window function depends on which frequency peak features
are most critical. Some window functions will give very sharp main peaks
53
200
150
100
50
B
20
40
60
80
100
120
but will also introduce a high level of noise, or sidebands, not containing any
information of interest. The shape of three different window functions are
shown in Fig. 4.7a, these are the square window marked A, i.e. the result of
not applying any particular window function, the Blackman window marked
B and the Blackman-Harris window marked C. In the case of STFT analysis,
the selected window is multiplied with the signal of a selected range before
the FFT of the product is computed. By computing a FFT of the window
function itself one can see which effect a particular window function will
have on the result of the signal analysis. The FFT of the window functions
in Fig. 4.7a is shown in Fig. 4.7b. In general a window function giving a
very narrow main peak will have high sidebands, while a window function
with a wide main peak will have more subdued sidebands. For the present
analysis, a balance between the two seems the best. We would like to be
able to determine the frequency-trace of a partial as certain as possibly
requiring sharp peaks, but, due to the tracing procedure, the height of the
peak ridge relative to the noise floor should be as large as possible. As
a compromise the Blackman window function was selected for the STFT
analysis.
4.2.4
4.2.4.1
54
6
1.0
10
10
0.5
-3
10
Power
Amplitude
10
-6
10
C
-9
10
0.0
-12
0.0
0.5
1.0
0.0
Length
a)
0.5
10
1.0
Frequency, a.u.
b)
Figure 4.7: Three normalized window functions are shown in a); a square
function (curve A), a Blackman function (curve B) and a Blackman-Harris
function (curve C). The resulting Fourier transform of these functions are
shown in b) in the corresponding colors.
55
56.5 % to 64.5 %. The porosity changes abruptly as expected at t = 15 min.
The second curve plotted after t = 15 min is the porosity obtained after
15 min for a sample etched under the same conditions but with a constant
etch current of 20 mA from t = 0.
The porosity of the dual current etched sample has shifted toward lower
values compared to the constant current etched sample. This may be understood as a consequence of a change in HF concentration assuming a
simplistic view of the electrolyte chemistry at the pore front. This concentration depends both on diffusion of HF to the pore front as well as on
the usage of HF depending on e.g. current density. The constant current
density etched sample has a certain porosity profile, resulting in a certain
diffusion of HF to the pore front, where the diffusion constant will change
with position through the layer. For the dual current density etched sample
the first half of the etching results in a different porosity profile than that
in the sample of constant current density etched after the first 15 min. As
the current is higher in the first half, the porosity will be higher, hence a
higher diffusion of HF will result. From this one may assume that the HF
concentration at the pore front is higher for the dual etched sample after 15
min of etching than for the constant etched sample after the same time. In
p-type silicon, it is well known [45] that a higher HF concentration, given
the same current density, will result in a lower porosity. This gives a good
understanding of the results in Fig. 4.9.
4.2.4.2
Etch calibration
It is possible to use porosity and etch rate time dependence data to calibrate
etching so that a known porosity with a known etch rate is obtained at any
given time of continuous etching by changing the current density correctly.
A constant time step has been used in the calculation of the needed current
density. The current density needed to obtain the desired porosity is found
by interpolating the porosity data at a given time for all the measured
current densities and extracting the current density for the desired porosity.
This is done for each time step until a given film thickness i reached. The
etched thickness for each time step based on the found current density is
also interpolated from the measured etch rate calibration data.
This procedure is based on the assumption that the etch condition at the
pore front is a function of time and instantaneous etch current density only,
independent of the earlier etch conditions. This is only an approximation,
as exemplified by Fig. 4.9.
The data needed for this procedure are obtained by fabricating several PS
layers at constant current condition for different current densities. This
has been done for two separate electrolyte conditions. A low temperature
etch in a 26 % HF solution was used for the results in Figs. 4.10 and 4.12.
The etching was done in a refrigerator with an average temperature of 5 C,
56
Porosity (%)
65
Current change
40 mA
60
55
20 mA
50
15 min
25
30
40 mA at t =
10
15
20
Time (min)
however, the temperature regulation was slow and the temperature in the
electrolyte varied 3 C. The results shown in Figs. 4.11 and 4.13 were obtained during etching at room temperature with an electrolyte consisting of
15 % HF and 10 % of the ethanol replaced by glycerol. These results show
the generally accepted trends for PS formation in p-type Si; higher current
density leads to higher porosity and etch rate, and for a given current density the porosity increases with a decreasing HF concentration. Also, clearly
the conditions for etching changes with time. This also follows what has
been reported earlier [85, 81, 82] in p-type PS. The trend is toward higher
porosities and lower etch rates with depth/time. Both observations fit with
a picture of decreasing HF concentration with depth/time due to diffusion
limitations. Note that some exceptionally high porosity values are shown
in Figs. 4.10 and 4.12. Porosity values above 90 % are hard to obtain
after drying. Possible reasons for the shown porosity values are the use of
the EMA effective medium formula which may give slightly shifted porosity
values and that highly porous PS may still be mechanically intact during
etching. Highly porous PS is unstable mostly due to the capillary forces
which occur during drying, resulting in breakage of the internal structure.
To test the calibration data, a PS layer was fabricated where the current had
been calculated beforehand to give a layer of uniform porosity throughout
the depth. The back side reflection was measured during this fabrication
57
100
100
40 mA/cm
Porosity (%)
80
15 mA/cm
60
5 mA/cm
40
1 mA/cm
10 mA/cm
60
30 mA/cm2
15 mA/cm2
10 mA/cm2
5 mA/cm2
60
40
20
40
50 mA/cm2
80
20 mA/cm
20
30 mA/cm
70mA/cm
Porosity (%)
55 mA/cm
80
100
120
140
20
20
Time (min)
Figure 4.10: The obtained porosity time profiles for different current
densities. The electrolyte used is
26 % HF without glycerol and etching is done at approximately 5 C.
Porosities increase with current density. Plotted are data for 1, 5, 10,
15, 20, 30, 40, and 55 mA/cm2
40
60
80
100
120
140
Time (min)
Figure 4.11: The obtained porosity time profiles for different current
densities with an electrolyte consisting of 15 % HF and 10 % ethanol
substituted with glycerol. Etching is done at room temperature.
Porosities increase with current density. Plotted are data for 5, 10, 15,
30, 50, and 70 mA/cm2 .
so etch rate and porosity could be calculated. The porosity profile is shown
in Fig. 4.14. Due to the short window used in the STFT analysis there
are some irregularities which would be smoothed out with a longer window. The etch duration was designed to give a layer of 10 m thickness
and the current profile was designed to give a uniform 50 % porosity using
26 % HF at 5 C. The uniformity of the porosity in Fig. 4.14 shows this
is a feasible way of obtaining uniform layer porosity, however the porosity
measured is roughly 3 % (abs.) below the designed value and the thickness
obtained according to the reflectance measurement was 10.69 m. The average porosity was measured by gravimetry which gave a porosity of 52.0 %
and a thickness of 9.56 m. These discrepancies likely show the uncertainty
of the gravimetric method and the error introduced by the choice of an
effective medium approximation as discussed in Sec. 3.1.
4.2.4.3
As seen in Fig. 4.4 and in the Papers I-III, the measured reflection signal
has a good signal-to-noise ratio. Assuming that the problem of the slow
signal oscillation is likely due to mechanical or thermal instabilities of the
setup/fiber, as discussed, may be resolved, the setup may well be used for
real-time monitoring of parameters or feedback control of etching parameters. In the case of real time monitoring, the use of STFT is appropriate,
however, prior knowledge of porosity and etch rate ranges will increase the
58
4
-1
-1
55 mA/cm
40 mA/cm
2
30 mA/cm
20 mA/cm
15 mA/cm
10 mA/cm
5 mA/cm
20
40
60
80
100
70mA/cm
3
50 mA/cm
30 mA/cm
15 mA/cm
10 mA/cm
5 mA/cm
120
140
Time (min)
20
40
60
80
100
120
140
Time (min)
60
Porosity (%)
55
50
45
40
35
30
1.0
1.5
2.0
2.5
Time (min)
3.0
3.5
usability of such a system. This knowledge will ensure that the parameter space is limited such that the two main peaks in the STFT are found
quickly at onset of etching, see paper I for a description of the data analysis.
In addition, optimization of the peak tracing algorithm is needed. In the
case of feedback operation, the use of STFT for frequency determination is
only appropriate if the current density used is constant with time or varies
slowly, i.e. significant variations over much longer time than window function length. In the case of multilayer etching, in which case the feedback
operation would be most useful, the etching time for each layer will often
be too short for STFT based analysis. As seen in Fig. 4.15, which is from
the etching of an infrared Bragg filter with layer thicknesses in the order
of a few 100 nm, each layer etching results in around one period, or less,
of the main interference oscillation (main partial). With prior knowledge
of the system and an appropriate fitting algorithm, the information in this
signal may be used for feedback. Note that in Fig. 4.15 there are three
periods; the high current layer with a high etch rate, the low current layer
59
with a lower etch rate and a break period with no current. This is used to
regenerate the electrolyte at the pore tips instead of adjusting the current
density to give the set porosity with changing electrolyte conditions.
High current layer
0.8
Low current layer
0.6
Break
0.4
0.2
0.0
24.5
25.0
25.5
26.0
Time (min)
26.5
27.0
Paper I
S.E. Foss, P.Y.Y. Kan and T.G. Finstad
Single beam determination of porosity and etch rate
in situ during etching of porous silicon
J. Appl. Phys., 97, 114909 (2005)
Single beam determination of porosity and etch rate in situ during etching
of porous silicon
S. E. Foss, P. Y. Y. Kan, and T. G. Finstada
Department of Physics, University of Oslo, P.O. Box 1048 Blindern, N-0316 Oslo, Norway
Received 14 January 2005; accepted 1 April 2005; published online 31 May 2005
A laser reflection method has been developed and tested for analyzing the etching of porous silicon
PS films. It allows in situ measurement and analysis of the time dependency of the etch rate, the
thickness, the average porosity, the porosity profile, and the interface roughness. The interaction of
an infrared laser beam with a layered system consisting of a PS layer and a substrate during etching
results in interferences in the reflected beam which is analyzed by the short-time Fourier transform.
This method is used for analysis of samples prepared with etching solutions containing different
concentrations of HF and glycerol and at different current densities and temperatures. Variations in
the etch rate and porosity during etching are observed, which are important effects to account for
when optical elements in PS are made. The method enables feedback control of the etching so that
PS films with a well-controlled porosity are obtainable. By using different beam diameters it is
possible to probe interface roughness at different length scales. Obtained porosity, thickness, and
roughness values are in agreement with values measured with standard methods. 2005 American
Institute of Physics. DOI: 10.1063/1.1925762
I. INTRODUCTION
0021-8979/2005/9711/114909/11/$22.50
97, 114909-1
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-2
FIG. 1. Color online Schematic of the different etch setups. The setup
denoted A is based on free-space beam transmission. The laser diode is
denoted LD. The schematics of both setups A and B show a cross section
of the etch bath. Setup B shows the wide beam optical-fiber variant. MM
denotes the multimode optical fiber and SM denotes single mode optical
fiber. A coupler is shown; this splits the fiber and guides the reflected signal
from the sample to the detector. A graded index lens, as indicated, is used to
collimate and collect the light close to the sample. The third setup C shows
a fiber variant where no collimating lens is used, but where the fiber is close
enough to the sample backside 500 m to collect a significant amount
of the light reflected from the sample. This results in a probed area about the
size of the fiber core cross section.
etch rate and porosity. The addition of the interface roughness measurement to the refractive index and etch rate measurements in the presented method results in an extended
characterization ability.
We will describe the experimental setup first in Sec. II.
There we also include the details of the samples where the
method has been used, supplementary characterization techniques, and the determination of supplementary parameters
needed by the method. Then in Sec. III we outline the
method used and the theory behind it. In Sec. IV we present
some examples where the method has been used. These measurements are briefly discussed in Sec. V.
II. EXPERIMENTAL DETAILS
A. Interferometric measurement setup
The preparation of PS films was done in a standard upright etch cell with a solid Cu back contact. Through a hole
cut out in the Cu back contact an IR diode laser was directed
at the sample and the reflected beam was detected with an
InGaAs detector. Three different measurement geometries
were used, with the main difference being the beam diameter. Figure 1 shows a schematic of the different setups. By
using different beam diameters it is possible to probe interface roughness with different spatial wavelengths and also
obtain spatial averaging over probe areas of different sizes.
The latter implies a compromise between a large electrical
contact area on the sample backside, which is needed for
homogenous etching over the etch area, and large beam diameter, needed for obtaining spatially averaged etch rate and
porosity data. Two of the setups were based on fiber optics
while the third was based on free-space optics. By using
optical fibers a compact setup was obtained and beam alignment and positioning were simple.
For both fiber setups, which are shown in Fig. 1 setups
B and C, the monochromatic, coherent light source used
was a diode laser pigtailed to a single mode fiber with an
optical isolator. The wavelength was 1550 nm and the output
Porosities were determined both by analysis of the reflected signal and by gravimetry. The layer thicknesses after
etching were determined by the reflected signal, by crosssection observation in an optical microscope or by stylus
surface profilometry after stripping away the PS film in concentrated NaOH. The interferometrically measured roughness values were compared to values obtained by white-light
interferometry measurements WYKO NT-2000 after stripping away the PS.
Refractive indices of the different electrolytes used were
measured by the amount of parallel shift of a laser beam
transmitted through a holder made of Plexi-glass containing
the electrolyte while it was rotated and using Snells law.
Different concentrations of both HF and glycerol were used.
The inset in Fig. 2 shows a schematic of the experimental
setup. The laser beam used had a wavelength of 1310 nm.
The refractive index was found to change little between 1310
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-3
detector at the same side of the sample as the laser. This total
signal will have contributions from many rays that interfere.
As the porous layer thickness increases as etching
progresses, the optical thickness of each layer changes and
the phase of each ray will vary. Thus the reflection signal
will vary in intensity with etching time. An example of an
experimental reflection intensity signal is shown in Fig. 4. If
we ignore absorption in the layers, the amplitude of each
partially reflected and transmitted ray is given by the product
of Fresnel coefficients from the encountered interfaces. As
will be shown later, the time varying signal can be decomposed into partials of specific frequencies and the frequencies
of these partials are those arising from the interference of ray
pairs. The rate of change of the phase difference between two
detected rays will give a partial with a specific instantaneous
frequency of oscillation. There are many rays contributing to
the detected signal which results in many possible partials of
different frequencies. However, most rays, and thus most
partials, will have very small amplitudes due to multiple reflections and transmissions. By extracting the appropriate
frequencies and their time dependence from the experimental
signal, it is possible to calculate the optical thickness and the
etch rate of each layer at any given time.
The schematic of the system in Fig. 3 shows the principal rays and partials used in the analysis. The phase difference between the ray transmitted at the air-substrate interface, reflected once at the substrate-PS interface, and
transmitted back out ray II in Fig. 3 and the ray reflected at
the substrate ray I is given by
1t =
2 2
n sin2 dsubt + 0 dsub ,
0 sub
1t =
FIG. 3. Color online Ray trace through the sample during etching. Interference between rays I and II is denoted as situation 1 and between I and III
as situation 2. Scattering of light due to a rough PS-substrate interface is
indicated.
1t 2 2
d dsubt
= nsub sin2
.
t
0
dt
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-4
rt =
2t =
1
2ODsubt + 2ODPSt + 0 dsurf ,
0
dPSt
2
nPS
l sin2 dl.
2
nsub
sin2
02t
2rt
+ sin2 . 7
Using the measured refractive indeces for the different electrolytes it is now possible to calculate the porosity profile
and average layer porosity as a function of PS layer depth.
Equations 3 and 7 together give the most important parameters of the PS etching process.
where Asub and APS are the amplitudes and sub and PS are
the phases of the rays reflected from the substrate backside
and the PS-substrate interface, respectively. The cosine component of Iref gives the amplitude and frequency of the interference oscillation. The oscillation amplitude is given by the
cosine prefactor. Asub is constant while APS contains information on the PS-substrate interface scattering, substrate absorption, and PS refractive index in the electrolyte,
APSt = tair-subsubtsR,sub-PStrsub-PStsubttsub-air . 9
Here tair-sub and tsub-air are the transmission amplitude coefficients, rsub-PSt the time-dependent reflection amplitude coefficient at the substrate-PS interface, sR,sub-PSt the timedependent scattering factor as the roughness changes with
time, and subt the absorption factor which changes due to
a decrease in substrate thickness. The time dependence of the
reflection amplitude coefficient is due to the change in the PS
porosity with time.
Transmission and reflection amplitude coefficients, as
well as the absorption factor, are calculated using published
data for the complex refractive index of bulk Si. Normal
Fresnel relations are used for the transmission and reflection
coefficients, while the absorption factor is given by subt
= exp2kzt / , where k is the imaginary part of the complex refractive index and zt is the time-dependent substrate
thickness. The scattering factor value will be extracted from
the interference data and is in the following assumed known.
This extraction is explained in Sec. III C below.
The DaviesBennett theory9,10 attempts to describe the
local phase change in the reflected plane wave front introduced by the height irregularities of the interface. These
phase changes result in a reduced intensity in the specular
direction as conditions for destructive interference will develop between different parts of the wave front resulting in a
loss of coherence. The theory assumes a rms irregularity
height roughness value, , much smaller than the wavelength of the incident light in the medium, , and that the
height function describing the roughness has a Gaussian distribution. In this case the Fresnel reflection coefficient for the
rough PS-substrate interface may be written as
2
= R0 exp
Rsub = R0sR,sub-PS
4sub-PSnsub
0 cos sub
10
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-5
lated based on the obtained sub-PS. The transmission coefficient at the PS-substrate interface has been shown by
Filiski11 to be
2
Tsub = T0sT,sub-PS
= T0exp
11
for the same conditions as for the reflection coefficient in Eq.
10. Here T0 is the transmission coefficient for a perfect
interface, while nPS is the refractive index of the PS and PS
is the angle in the PS layer. Then the reflected amplitude
from the front becomes
2
2
2
tsR,PS-front
Afrontt = tair-subsub
tsT,sub-PS
ttsub-PStPS
trPS-frontttPS-subttsub-air ,
12
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-6
To get the right porosity value of the PS layer, knowledge of the refractive index of the etchant is necessary as the
value obtained from frequency analysis is the refractive index of the PS layer when it is immersed in the etchant. An
effective-medium model must be used to approximate the
porosity, which both accounts for the refractive index of the
silicon substrate and the etchant. The Bruggeman approximation theory was chosen as this is most often used for PS in
the literature.17 The refractive index of glycerol in the visible
is 1.466 while for ethanol it is 1.365 and for water it is 1.350,
hence glycerol will influence the etchant refractive index significantly. Refractive index values for solutions with concentrations different from those measured are extrapolated from
the measurements assuming a linear dependency between refractive index and both HF and glycerol concentration. Some
data with error bars are shown in Fig. 2. For the analysis
presented in Secs. III AIII C the electrolyte refractive index
is assumed constant and independent on time and thickness
of the PS layer. This assumption could be violated if the
electrolyte composition changes close to the pore tips due to,
e.g., diffusion limitations on the HF concentration as well as
buildup of Si-rich chemical species. However, the use of the
assumption seems well justified based on a comparison
between porosities obtained by interferometry and by
gravimetry.
IV. EXPERIMENTAL RESULTS
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-7
fect on average porosity of different HF and glycerol concentrations will be shown next. After this some PS-substrate
interface roughness measurement results will be given as
well as the time dependence of the valence of a few samples.
Some of the measurements used to test this method have
been reported earlier.18 Those measurements were all done
with the free-space optics setup.
The difference in using the wide beam and narrow beam
setups can be seen in Fig. 6. Here two experimental runs are
compared. They use the same parameters; 26% HF,
15 mA/ cm2 at RT, but were performed in different setups.
The traced amplitude of the partial corresponding to curve 1
in Fig. 5a is plotted for both experimental runs. The amplitudes have been scaled so that their starting values are identical. For both cases the general trends in the amplitudes are
the same; there is a decrease in the beginning due to rapidly
increasing roughness. This is overtaken by an increase in
amplitude due to less absorption in the substrate as the PS
film grows. There is a clear difference in the relative importance of these effects, loss due to scattering/coherence loss
and loss due to absorption, between the two experimental
runs using different beam sizes. In the case using the narrow
beam setup the initial decrease is much less, indicating a
significantly smaller measured roughness within the beam
spot than for the wide beam case, hence the relative effect of
the decreased absorption is greater. This is a strong indication that there are different spatial wavelengths of roughness
of importance at different spot sizes. In most of the following
measurements the wide beam free-space or fiber-optic setup
has been used as the signal-to-noise ratio is better.
All the p+ samples etched at constant current conditions
in this study have shown an increase in porosity with longer
etching time. This increase is the result of the increasing
porosity with depth which is exemplified in Fig. 7 where the
porosity profile of one of the samples is calculated and plotted. To be able to compare the porosity measured by gravim-
FIG. 7. An example of a porosity profile solid line in depth upper horizontal axis/time lower horizontal axis and the accompanying etch rate
change. The time development of the average porosity is also indicated
dashed line. Data are from the same sample as in Fig. 5a. The dotted
lines are an indication of uncertainty based on the peak width of the signal
line in the spectrogram.
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-8
TABLE I. Comparison between average gravimetrically measured and interferometrically measured porosities
%. The fit is quite good, values obtained by interferometry even give a more consistent change in porosity
with change in current density. The electrolyte used contains only HF and ethanol. Etching was done at RT.
Average porosity %
after etching in 26% HF
Average porosity %
after etching in 15% HF
Current density
mA/ cm2
Gravimetry
Interferometry
Gravimetry
Interferometry
5
10
15
20
30
36
43
38
48
53
32
40
46
53
69
65
66
80
83
65
66
71
79
method discussed. The plot shown in Fig. 11 is representative of the samples prepared in this experiment. Here the rms
roughnesses of two samples are plotted against layer thickness during etching. The samples are etched with identical
parameters except for a difference in glycerol content. The
sample with 10% glycerol shows a power-law dependence of
roughness on thickness, however, with values substantially
smaller than for the sample etched without glycerol. The
latter sample shows a saturation occurring at around 10 m
after a similar power-law dependence. A saturation of roughness has been reported before in the case of p and p PS.6 In
Table II the interferometrically obtained maximum roughness values of several samples are compared with white-light
interferometry WLI roughness values obtained from the
PS-substrate interface after stripping of the PS layer with
NaOH. The data shown are from samples measured with the
wide beam fiber setup. A WLI spot size similar to the spot
size obtained with the interferometry setup was used. As
there was some curvature over the whole etched interface as
well as fluctuations in thickness with spatial wavelengths
longer than the spot diameter the rms roughnesses when
measured over the whole etched area for these samples were
significantly larger.
From the local porosity data and the etch rate data as
determined by interferometry and the measured anodic current, it is possible to calculate the valence of the reaction,
FIG. 10. The effect of the glycerol content in the electrolyte on average
porosity for different concentrations of HF. The points from samples etched
in 15% HF are marked with and 26% are marked with . The glycerol
ratio is compared to the total ethanol/glycerol content. There is a clear
difference between the 15% and 26% HF cases.
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-9
FIG. 12. Valence calculated based on etch rate and porosity as a function of
time for samples etched with several variations in etching parameters. It is
evident that the different parameters have a large effect.
FIG. 11. Roughness nm rms obtained from calculations using the reflection signal amplitude data for two samples etched with 15% HF at RT with
different glycerol content. 10% glycerol gives the lowest roughness in the
case of 15% HF of all concentrations tested. The roughness saturation of the
0% glycerol data has been reported earlier Ref. 6 for lower-doped p-type
PS.
i.e., the charge required to remove one silicon atom from the
substrate. It is assumed that the etching only occurs at the
pore tips,14,19
j
,
Nerp
13
Etch condition
Reflectance
measurement
rms nm
157
173
168
155
187
157
170
169
153
186
sponds well with gravimetrically obtained porosities. The obtained rms roughness also corresponds well with other measurements. Thus, we consider the obtained results for good.
The present setup is slightly different than setups presented
before1214 and the analysis is more extensive.
In the present setup as well as in Ref. 12, the beam is
transmitted through the sample using an IR laser beam for
which Si is transparent, hence there will be no free-carrier
generation in the sample. When preparing PS it is of great
importance for reproducibility to control the access of charge
carriers, specifically holes, to the etched surface. By using an
IR laser, the beam intensity may be quite high without influencing the etching. This gives a good signal-to-noise ratio.
The use of an IR beam also facilitates the measurement of
thick layers in comparison with a beam in the visible range
which will be absorbed within the first few microns. On the
other hand a shorter wavelength will improve the resolution
of thickness and etch rate. The transmission setup may easily
be adapted to a liquid back contact etch cell. By directing the
beam from the backside, disturbances from hydrogen
bubbles in the beam path through the electrolyte are avoided.
To obtain both etch rate and porosity either one or two beams
may be applied. In the two-beam case, the beams must have
different wavelengths or different incident angles. Only the
top layer needs to be probed in this case, hence a simple
reflection signal is obtained, however, for this to happen
there must be enough absorption or a highly scattering interface to avoid reflection from the back surface.14 In the onebeam case the beam must be able to probe more layers and
this results in a more complex reflection signal also making
it more complicated to analyze. However, having only one
beam simplifies the setup. Further, by using fiber optics, the
footprint of the setup is minimized making it more mobile
and space efficient and making alignment easier.
When preparing multilayer structures for optical filter
applications in PS, thicknesses less than 20 m are normally
used to minimize absorption, for filters used in the visible
range even thinner filters are made. It is clear that the variation in porosity within the first few microns of etching will
be detrimental to the properties of the optical filters if not
taken into account in the design.20 The results shown in Fig.
7 suggest that the greatest change in porosity with depth
occurs within the first 2030 m. By changing current den-
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-10
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
114909-11
This work is supported by the Research Council of Norway. The authors are grateful for help by Chetna Schukla and
Erik S. Marstein during the initial experiments.
G. Vincent, Appl. Phys. Lett. 64, 2367 1994.
M. Berger et al., J. Phys. D 27, 1333 1994.
3
M. Ha, J. Kim, S. Yeo, and Y. Kwon, IEEE Photonics Technol. Lett. 16,
1519 2004.
4
V. Mulloni and L. Pavesi, Appl. Phys. Lett. 76, 2523 2000.
5
W. Zheng, P. Reece, B. Sun, and M. Gal, Appl. Phys. Lett. 84, 3519
2004.
6
G. Lrondel, R. Romestain, and S. Barret, J. Appl. Phys. 81, 6171 1997.
7
S. Setzu, G. Lrondel, and R. Romestain, J. Appl. Phys. 84, 3129 1998.
1
2
Downloaded 01 Jun 2005 to 129.240.153.219. Redistribution subject to AIP license or copyright, see http://jap.aip.org/jap/copyright.jsp
Paper II
P.Y.Y. Kan, S.E. Foss and T.G. Finstad
The effect of etching with glycerol, and the interferometric measurements on the interface roughness
of porous silicon
Phys. Stat. Sol. (a), 202, 8, 1533 (2005)
II
Original
Paper
phys. stat. sol. (a) 202, No. 8, 1533 1538 (2005) / DOI 10.1002/pssa.200461173
Introduction
Porous silicon (PS) can be easily used to fabricate into optical structures such as waveguides [1], optical
filters [2, 3] or luminescent microcavities [4] because its refractive index can be easily modulated.
Chemical and biological sensors based on modulation of the optical properties of PS have also been
made [5]. The interface roughness between PS and silicon (Si) plays a very important role in PS optical
devices. Light transmission can undergo severe losses through scattering if the interfaces are not smooth.
There are many parameters that can affect the roughness of the PS. For example, the type and resistivity
of the Si substrate, the composition of the electrolyte, i.e. the ratio between HF, H2O and ethanol (EtOH),
the current density (J) and the temperature of the electrolyte. The latter two parameters have been studied recently [6, 7] where it was shown that lowering the temperature decreases the interface roughness of
PS which was attributed to the increase of viscosity. Glycerol will also increase the viscosity and was
reported to reduce the roughness of PS-substrate interface [7].
In this study we used glycerol as a partial replacement for ethanol in an electrolyte with ratios from
10 70%. The electrolyte consisted of HF:(EtOH:glycerol). The effect of different current densities and
HF concentrations was also measured. The etching solution was prepared from 40% HF diluted with
EtOH in a ratio depending on the desired percentage of HF. The PS-substrate interface movement and
roughness were measured in-situ during etching by a simple interferometric technique with an infrared
diode laser beam incident from the dry backside of the sample. Experiments were conducted at room
1534
P. Y. Y. Kan et al.: The effect of etching with glycerol, and the interferometric measurements
Pt
HF:EtOH
Si wafer
Copper
Teflon
detector
IR Laser
mirror
temperature with p+-type Si wafers. The average porosity for each experiment was determined by the
gravimetric method. The obtained average thickness was typically around 50 m. The starting material
was p+-type Cz (100) silicon, double-side polished with a thickness of 520 m. The resistivity was 0.01
0.02 cm.
The experimental setup for in-situ interferometric measurement during etching is shown schematically
in Fig. 1. A two-electrode Teflon cell was placed on a stand equipped with two inclined mirrors that
guided the infrared (IR) laser beam ( = 1.31 m, maximum power is 8 mW, spot size is 1 2 mm in
diameter). The copper electrode had the centre opened so that the laser light could reach the Si wafer and
interference could be produced from the bulk Si and the PS layer while etching. A constant current density of 5 to 30 mA/cm2, supplied from a Keithley 2400 current source, was applied for 1 hour while the
interference signal was sampled in each experimental run. An example of the measured interference
signal is shown in Fig. 2. The signal amplitude clearly falls off with time which is an indication of increasing roughness.
For easy parameterization we have defined roughness here as a percentage which is calculated, as
illustrated in Fig. 3a, from the difference between the modulation start amplitude, Amax, and that at time t,
A(t):
Roughness = (Amax A(t))/Amax 100 .
(a)
30% glycerol
Intensity (a.u)
15%HF 10mA
(1)
10
20
30
40
50
60
Time (min)
(b)
Intensity (a.u)
Intensity (a.u)
(c)
54
Time (min)
56
58
Time (min)
60
Original
Paper
1535
PS
Amax
Si
A(t)
laser
Time
a)
b)
Figure 3b illustrates the reflection from various interfaces in a PS sample. Interference between the partially reflected IR beams (1, 2, 3) can easily be measured. The amplitude of the signal will be dependent
on the roughness of the interfaces. Beams (2) and (3) may get weaker as etching proceeds and as the
reflected light undergoes diffuse scattering which gives weaker interference signals. Thus the modulation
amplitude decreases (narrower width) with time (see Fig. 2a). The waveform shape and periodicity can
be analyzed by Fourier decomposition to monitor etching rate and porosity as a function of time, which
will be presented in future works. The roughness parameter as here defined is influenced by uneven etch
rate over the whole area of the laser beam, fluctuations in the effective dielectric constant of the PS layer
due to bubbles and by scattering of the beam. Both interfaces will contribute. A good review of common
ways of quantizing roughness can be found in Ref. [8].
A plot of the roughness against the PS thickness is shown in Fig. 4. There was not much difference
in the roughness for the 26% HF sample (Fig. 4a) as the glycerol percentage was varied. However,
for the sample in Fig. 4b (15% HF, porosity of 70%), there was a large difference in the roughness between samples with 0% and 10% glycerol. The sample with a 10% glycerol replacement of ethanol indicates a relatively smooth interface, whereas the one without glycerol shows a very high roughness percentage.
The 30 and 70% glycerol samples also show a smoother surface than the 0% glycerol sample
(Fig. 4b), but these are comparatively rougher than the 10% case. This implies that glycerol is an effective agent for smoothing the interface roughness. However, the results also indicate that a high glycerol
content (>10%) would not be as effective as the 10% sample. This could be due to the fact that the content of ethanol also plays an important role in the etching solution, which can be speculated is related to
Time (min)
10
100
30
Time (min)
40
50
20
60
40
60
80
100
Roughness (%)
0% glycerol
10%
"
30%
"
50%
"
70%
"
80
Roughness (%)
20
60
40
80
0% glycerol
10%
"
30%
"
70%
"
60
40
porosity ~45%
20
20
26%HF
2
10mA/cm
(a)
0
0
10
20
30
Thickness (um)
40
50
60
0
0
15%HF
2
10mA/cm
porosity ~70%
(b)
10
20
30
40
50
60
Thickness (um)
Fig. 4 Graphs showing the calculated roughness change in samples etched with different electrolytes. (a) An electrolyte with 26% HF is used with changing glycerol content, and (b) an electrolyte with 15% HF is used. In the 15%
case it is clear that 10% glycerol gives the lowest roughness. In the 26% HF case the effect is not so large.
2005 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
P. Y. Y. Kan et al.: The effect of etching with glycerol, and the interferometric measurements
(a)
15%HF
(b)
0% glycerol
10%glycerol
5mA
5mA
Intensity (a.u)
15%HF
Intensity (a.u)
1536
10mA
12mA
10mA
12mA
15mA
15mA
20mA
20mA
10
20
30
40
50
60
10
20
30
40
50
60
Time (min)
Time (min)
Fig. 5 Interferometric results from the15% HF samples at different current densities, 5 20 mA/cm2 for 60 minutes.
Plot (a) with 0% and (b) with 10% glycerol.
the dependency of bubble formation on viscosity and surface energy during the etching, where ethanol
has a favourable effect.
Figure 5 shows the interferometric results from the 15% HF (porosity 70%) samples, one with 10%
glycerol and one without, and the current density varied from 5 to 20 mA/cm2. By observing the modulation amplitude of the interference graphs for the 0% glycerol samples (Fig. 5a), it is evident that
the roughness has an irregular trend. This means that the roughness as measured here varies for nominally identical conditions. However, the trend seems to be steady for the 10% glycerol samples. For all
samples investigated we always produce low roughness with that percentage of glycerol. Detailed analysis indicates that the 10% samples did indeed smooth this irregular trend, as is shown in Fig. 6 (15%
HF).
Other samples with different HF concentration (13 26% HF, porosity 85 50%) are also presented in
Fig. 6 which all indicate a steady trend from the 10% glycerol samples. Largely varying roughness for
nominally identical experimental conditions can only be observed from the samples with the higher porosity and without glycerol replacement (13% and 15% HF samples). This indicates that the roughness of
the high porosity samples, at least in respect to the current measurement method, is critically dependent
13%HF
15%HF
26%HF
20%HF
0% glycerol
10% "
roughness
80
40
0
0
10
20
10
20
30
10
20
30
10
20
30
J (mA/cm )
Fig. 6 (online colour at: www.pss-a.com) A plot of roughness against current density at 13% to 26% HF which
correspond to the porosity of 85 to 50%, respectively. The 15% HF sample with 0% glycerol has two attempts of
measurements.
2005 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
Original
Paper
1537
100
100
(a)
13% HF
15% "
26% "
80
Porosity (%)
Porosity (%)
80
60
40
13% HF
15% "
26% "
20
60
40
0.01 ohmcm
2
10mA/cm
20
0.01 ohm-cm
(b)
0
0
10
15
20
2
J (mA/cm )
25
30
10
20
30
40
50
60
70
% glycerol
Fig. 7 A plot of porosity against current density (a), and against glycerol percentage (b), at 13% to 26% HF. Differences between 0% and 10% glycerol are also shown in (a).
on the etching parameters. The addition of glycerol seems to yield conditions which are in a part of the
parameter space which is much less sensitive to either the etching conditions or parameters of the roughness measurements. It has been suggested [7] that the smoothness of interfaces of PS and Si can be understood in terms of diffusion limited asperity smoothing. The nature of the roughness on several length
scales should be investigated further. Lrondel et al. [8] has made measurements of the evolution of
interfaces which suggest a saturation. The present experiments were not designed to test this.
From an optical application point of view a key issue is whether the same porosity and the same optical properties can be achieved with and without glycerol. To ensure that the 0% and 10% glycerol replacement is comparable, their porosities should be in a narrow range. We thus carried out a porosity
analysis on each sample and the results are shown in Fig. 7a. Interestingly, they only show a little difference, the porosities between samples with 0% and 10% glycerol varied only slightly, indicating that the
results were comparable.
Figure 7b shows the porosity of the samples made with different percentage of glycerol replaced with
ethanol, measured at three different concentrations of HF. The lower percentage of HF (13% HF) has a
higher porosity (85%), whereas the higher percentage of HF (26% HF) has a lower porosity (45%). The
porosity change with varying glycerol percentage at a particular HF concentration does not vary too
much, but seems to be in an acceptable range. For the 15% HF sample, the porosity was only varied from
60 to 70% as the glycerol content changed from 0 to 50%.
Conclusion
The simplicity and the usefulness of the IR laser in-situ interferometric measurement of etch rate and
roughness have been demonstrated, and the effect of glycerol on PS interface roughness have been
measured. These measurements show that glycerol is effective for smoothing the interface roughness of
the PS for a small range of glycerol concentrations and for certain HF concentrations, while glycerol
replacement has little or no effect when higher concentrations of glycerol, up to 70% replacement of
ethanol, are used. The 10% glycerol replacement in 15% HF solution seems to be the most effective in
smoothing the interface compared to no glycerol.
Acknowledgements We thank Chetna Shukla and Erik Marstein for help during the initial stages. This work is
supported by the Norwegian Research Council.
2005 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
1538
P. Y. Y. Kan et al.: The effect of etching with glycerol, and the interferometric measurements
References
[1]
[2]
[3]
[4]
[5]
[6]
[7]
[8]
A. Loni, L. T. Canham, M. G. Berger, R. Arens-Fischer, H. Munder et al., Thin Solid Films 276, 143 (1996).
G. Lammel, S. Schweizer, and Ph. Renaud, Sens. Actuators A 92, 52 (2001).
L. Pavesi, C. Mazzoleni, A. Tredicucci, and V. Pellegrini, Appl. Phys. Lett. 67, 3280 (1995).
M. Cazzanelli, C. Vinegoni, and L. Pavesi J. Appl. Phys. 85, 1760 (1999).
V. S.-Y. Lin, K. Motesharei, K.-P. S. Dancil, M. J. Sailor, and M. R. Ghadiri, Science 278, 840 (1997).
S. Setzu, G. Lrondel, and R. Romestain, J. Appl. Phys. 84, 3129 (1998).
M. Servidori, C. Ferrero, S. Lequien, S. Milita, A. Parisini et al., Solid State Commun. 118, 85 (2001).
G. Lrondel, R. Romestain, and S. Barret, J. Appl. Phys. 81, 6171 (1997).
Paper III
S.E. Foss, P.Y.Y. Kan and T.G. Finstad
In situ porous silicon interface roughness characterization by laser interferometry
Accepted for publication in the Proceedings of the
3rd Pits and Pores symposium, 206th Meeting,
ECS, Hawaii, 2004
III
The beam was oriented normal to the sample and the lens both collimated the incident
beam and collected the reflected beam. A sketch of the setup is shown in Fig. 1. A series
of experiments used a setup where the beam directed at the sample was not coupled
through fiber. This free space laser interferometer has been described before (16).
Figure 1 Sketch of the measurement setup. A standard etch cell is modified so that an IR diode laser
(LD) beam may be directed from a multimode (MM) optical fiber towards the back side of the sample
and the reflected beam collected back into the fiber, either via a collimating GRIN lens (left image) or
by direct coupling (right). The return signal is returned by a 2x1coupler to a GaAs PIN detector. The
LD is first connected to a single-mode fiber with an optical isolator, then it is connected with one input
of the MM 2x1 coupler. While etching, the PS/substrate interface moves giving rise to interference in
the reflected beam.
[1]
where Asub and APS are the amplitudes and sub and PS are the phases of the beams
reflected from the substrate back side and the PS/substrate interface respectively. As the
optical path length of the beam within the substrate changes during etching, PS changes.
This will lead to an oscillation in Iref with a frequency given by d ( sub PS ) dt . This
frequency may not be constant as the etch rate generally will change with time or depth.
The cosine component of Iref gives the amplitude and frequency of this oscillation and is
traced in the spectrogram. The oscillation amplitude is given by the cosine prefactor. Asub
is constant while APS contains information on the PS/substrate interface scattering,
substrate absorption and PS refractive index (in the electrolyte):
APS (t ) = t air sub sub (t ) s R , sub PS (t ) rsub PS (t ) sub (t )t sub air .
[2]
Here tair-sub and tsub-air are the transmission amplitude coefficients, rsub-PS(t) the time
dependent reflection amplitude coefficient as the PS refractive index changes during
ethcing, sR,sub-PS(t) the time dependent scattering factor as the roughness changes with
time and sub(t) the absorption factor which changes due to a decrease in substrate
thickness.
The scattering factor is calculated by fitting the extracted amplitude to a double
exponential as in Fig. 4, where a reflection amplitude from a wide beam measurement has
been extracted from the spectrogram and fitted. Fitting is done to avoid fluctuations
caused by crossing, interfering signals in the STFT as can be seen in Fig. 3. The fitted
amplitude is then scaled so that the extrapolated value at t=0 corresponds to the
theoretical amplitude with zero scattering. The scaling is necessary because the measured
data are not normalized to unit reflectance. The STFT uses a movable windowing
function which results in average amplitude values over the time range the window
covers, hence the STFT amplitude values for t=0 are correct for the signal at a time in the
middle of the first window. This necessitates an extrapolation of amplitude to t=0. The
fitted and scaled amplitude function corresponds to APS(t). Transmission and reflection
Amplitude, a.u.
0.10
0.08
Rsub = R s
2
0 R , sub PS
4
sub PS n sub
= R0 exp
0 cos sub
[3]
where nsub is the refractive index of the incident medium, 0 the wavelength in vacuum
and sub-PS the PS/substrate interface roughness. Note that the spatial frequency of the
roughness does not enter into this equation, so both long period (striations) and short
period (small scale) roughness will have an equal effect, depending only on sub-PS.
To calculate the rms roughness of the front surface, a scattering factor for
transmission, sT,sub-PS, must be calculated based on the obtained sub-PS. The transmission
coefficient at the PS/substrate interface has been shown by Filiski (12) to be
2
n sub
n PS
2
cos sub
cos PS
,
= T0 exp
0
[4]
for the same conditions as for the reflection coefficient in Eq. 3. Here T0 is the
transmission coefficient for a perfect interface, while nPS is the refractive index of the PS
and PS is the angle in the PS layer. Then the reflected amplitude from the front becomes
2
2
A front (t ) = t air sub sub
(t )sT2,sub PS (t )t sub PS (t ) PS
(t )s R, PS front (t )rPS front (t )t PS sub (t )t subair
[5]
where most of the parameters are the same as in Eq. 2 with the addition of a time
dependent transmission amplitude coefficient for the PS/substrate interface, tsub-PS(t), a
time dependent reflection amplitude coefficient for the PS/front interface, rPS-front(t), and
an absorption factor in the PS layer, PS(t), defined as for sub(t) with kPS calculated by
effective medium theory using the complex refractive index of bulk Si. The reflected
amplitude is given by the same calculations on the extracted amplitude of the signal in
case 1 in Figs. 2 and 3 as for APS. The calculation of the scattering factor, sR,PS-front, is the
same as for sR,sub-PS and gives the rms roughness value at the front interface.
RESULTS
Temperature effect
Assuming that the reflectance measurements give a good estimate of the roughness at
the beam spot, and that the roughness in this limited area is an indication of both
roughness time evolution and the roughness of the whole sample, the data plotted in Fig.
5 shows that roughness of samples etched with 15 % HF are influenced by etchant
temperature while those samples etched with 26 % HF are not significantly influenced.
The porosity and etchrate data for the 15 % samples are calculated and shown in Fig. 6,
as well as data for low temperature etching with 15 % HF and 1:9 glycerol:ethanol. There
is an increase in porosity and decrease of etch rate with decreased temperature. The
porosity data describe the instantaneous porosity closest to the interface, and therefore
also gives a porosity profile of the PS layer with depth. When calculating these data the
dissolution of PS closer to the surface, i.e. chemical dissolution, has been disregarded as
this is very small for p+ PS. Note the sharper increase in the RT porosity profile
compared to the low temperature profile. A similar effect has been reported by Servidori
in (8) for PS in p-type Si.
15%,RT
26%,RT
26%,LT
1.4
0.80
15%,LT
Porosity, abs.
Roughness, rms nm
120
100
0.85
80
60
40
1.2
0.75
0.70
1.0
15 % HF, RT, 0 % Gly
0.65
15 % HF, LT, 10 % Gly
20
0.8
0.60
140
0.55
0
20
40
60
80
100
Thickness, m
Figure 5 Roughness dependence on temperature
for different HF concentrations. In the 15 % HF
case there is a decrease in roughness with
temperature while the 26 % HF case show no
such dependence.
20
40
60
80
100
120
0.6
Time, min
Figure 6 Comparison of porosity and etch rate
between low temperature (4 C) and RT etched
samples, with electrolyte consisting of 15 % HF and
0 % or 10 % glycerol. Current density used was 15
mA/cm2 for all three.
Glycerol effect
The effect of replacing 10 % of the ethanol volume with glycerol on roughness at low
temperature is shown in Fig. 7 where a 15 % HF solution is used. In this case roughness
increases with the added glycerol compared to roughness measured without. A
comparison with results obtained at RT, shown in Fig. 9, may indicate an explanation.
These samples are etched with varying glycerol percentage from 0 to 70 % and HF
concentrations of 15 and 26 %. It is evident that a significant decrease in roughness is
obtained when replacing 10 % ethanol with glycerol at RT in the 15 % HF case. At low
temperature the viscosity of the electrolyte increases compared to at RT. By adding 10 %
glycerol to the electrolyte at low temperature, the viscosity increases further and the
roughness evolution is comparable to the evolution with a glycerol content higher than 10
% at RT. In Fig. 9 it is evident that roughness increases with glycerol content above 10
%. As seen for the low temperature porosity and etch rate data in Fig. 6, there is very
little difference between the samples with and without glycerol. This differs from the
same situation at RT in Figs. 8 and 9, where roughness decreases, porosity increases and
etch rate decreases by adding 10 % glycerol.
For other glycerol concentrations than 10 % in the 15 % HF case, the effect of adding
glycerol is minimal. Below 25 m the roughness is greatest with no glycerol and
1.4
0.80
0.75
15%, 10 % Gly.
120
100
15%, 0 % Gly.
80
60
Porosity, abs.
Roughness, rms nm
140
1.3
10 % Gly
0.70
1.2
0 % Gly
0.65
1.1
10 % Gly
0.60
1.0
40
0 % Gly
0.55
20
160
0.9
0.50
20
40
60
80
100
Thickness, m
Figure 7 Rms roughness plotted vs. PS layer
thickness. There is a change in the roughness
evolution when adding glycerol to a 15 % HF
electrolyte at low temperature. Samples are
etched at 15 mA/cm2.
10
20
30
40
50
Time, min
Figure 8 Porosity and etch rate obtained by reflection
measurements showing the dependence on
temperature and time. Samples were etched at RT
with a 15 % HF solution.
decreases with decreasing glycerol concentration from 70 % and down, however, for
higher concentrations of glycerol or no glycerol at all the roughness seems to stabilize,
while for lower concentrations of glycerol the roughness steadily increases with PS
thickness. A similar comparison is done in the 26 % HF case. Here the final roughness
increases with increasing glycerol concentration, with the sample etched with no glycerol
having the lowest roughness, although for thicknesses up to about 20 m the electrolyte
with 20 % glycerol gives the lowest roughness.
HF concentration, current density effect and roughness saturation
Comparing the data for 15 and 26 % HF in Fig. 9 a general tendency is apparent, most
of the data for 26 % HF shows a lower rms roughness than for 15 % HF, with the
exception of the 15 % HF + 10 % glycerol plot. However, HF concentration is reported to
have no effect on roughness in p and p- PS (5). When etching with 15 % HF and 10 %
glycerol and changing current density from 5 to 30 mA/cm2 a smaller spread in roughness
values is obtained compared to the same experiment done with no glycerol (16). In fact,
at a 20 m PS layer thickness the rms roughness varies from 30 to 70 nm with 10 %
glycerol and from 40 to 110 nm without. However, there is no apparent systematic effect
of current density on roughness in these data. This differs with the case of etching whit a
26 % HF solution. Here the roughness seems to increase with current density. At 60 m
thickness rms values are 66 nm, 75 nm and 92 nm for 10, 20 and 30 mA/cm2
respectively. According to (5), roughness decreases with an increase in current density in
p and p- PS.
Roughness, rms nm
15 % HF
26 % HF
120
100
80
60
0 % Gly
10 % Gly
20 % Gly
30% Gly
70 % Gly
0 % Gly
20 % Gly
30 % Gly
40 % Gly
40
20
0
0
20
40
60
80
20
40
60
Figure 9 Comparison of
rms roughness data for
different combinations of
HF concentrations and
glycerol amounts. All
samples are etched at RT
and with 10 mA/cm2.
80
Thickness, m
In (5) two distinct regimes are shown to exist in the roughness evolution for p and ptype PS. With roughness plotted against layer thickness in a log-log plot, it is clear that
roughness enters a saturation regime after a linearly increasing regime. In the data
discussed here, roughness clearly increase faster in the beginning, however, in a log-log
plot some show a clear correspondence with the two-regime picture, while some are
closer to linear. In Fig. 10 a sample etched at RT in 15 % HF with 10 mA/cm2 shows a
saturation regime while a sample etched with 20 mA/cm2 does not show any such
saturation. The saturation regime is explained as a result of a transition from hole limited
to a diffusion limited etching. A diffusion limited etching will be dominant when
diffusion of reaction species is decreased due to increased thickness of the porous layer,
hence fluctuations in resistivity at the PS/substrate interface have no influence on hole
availability and an asperity smoothening effect occurs. Another indication of the change
in diffusion is the porosity depth profile and etch rate change with time obtained for all
the samples investigated in this study, as in Figs. 6 and 8.
2
Roughness, rms nm
10
1
10
100
Thickness, m
Figure 10 Plot of roughness showing clearly the
increase and the saturation regime in the low current
case, and the lack of saturation in the high current
case. Both samples etched in 15 % HF at RT without
glycerol. The line is only a guide for the eye.
was not possible to obtain a value for the rms roughness, indicating that short period
roughness was below the detection limit of about 10 nm (5) roughness determination by
specular reflection, hence practically all of the surface irregularity is caused by striations.
Roughness of the PS/electrolyte interface
Analysis of the PS/electrolyte interface roughness was attempted. As explained in the
theory and method part, both the signal from case 1 and case 2 in Figs. 2 and 3 are used
to obtain this roughness. Due to the small roughness (<10 nm) at this interface as
reported by Servidori et al. (8), no reliable data were obtained. The absorption in PS
during etching will likely be large due to H2 bubble formation during etching of PS, this
is not accounted for in the calculation of the absorption factor.
Comparison between reflection and WLI
measurement
To verify that the method of roughness
calculation by specular reflection gives a good
indication of roughness, WLI measurements
were performed on a few samples that were
stripped of PS. A 2 mm diameter circular area
in the center of the samples where the laser
beam was thought to have been, was extracted
Figure 11 Surface plot of PS/substrate
from the WLI data. The rms roughness values
interface height data obtained by WLI
from these areas are compared to the
measurement. Parallel ridges with at least two
maximum roughness values obtained by
distinct spatial repeatability wavelengths are
reflection in Table 1 and show an excellent fit. clearly visible. These are caused by slightly
The true position of the beam is not known, so different etch rates due to inhomogeneities in
resistivity, i.e. striations. Rms irregularity
these values are only indicative of what is
height is in this case about 180 nm.
likely. These samples were all etched for long
times, up to 130 minutes, and were therefore quite deep, 2 00 m. Figure 11 shows a
surface plot of one such measurement. The parallel ridges caused by striations are clearly
visible. The rms roughness values obtained by this method show a very strong positional
dependence with a total rms value over the whole etched area up to a factor 2 larger than
that measured by reflection.
Figure 11 indicates that the roughness has several spatial wavelengths. A comparably
small period roughness, with a wavelength of about 200 m in the direction normal to the
ridges, is evident in the smaller ridges in the surface plot, as well as a much larger period
roughness with a wavelength in the order of mm. This large period roughness is very
irregular compared with the smaller and gives rise to the positional dependence of the
discussed WLI measurements, but both are most likely caused by an irregular dopant
distribution. The rms roughness values for the whole etched area are also tabulated in
Table 1. Note that this value is obtained after correcting for interface curvature, which is
caused by an inhomogeneous current distribution due to the opening in the Al back
contact and the design of the etch cell. This was done by an algorithm in the WYKO
analysis program. The large period roughness was not observed in WLI measurements
done on similar samples etched much shorter, to depths of about 10 m. At this depth,
only striations similar to the short period ones in Fig. 11 were present, and the rms
roughness was a few tenths of nm. This may indicate a slower development of the long
wavelength roughness.
Table 1 Comparison between measured roughness rms values by reflectance, data from a 2 mm diameter
circular area at the center of each sample with white light interferometry (WLI) and data from the whole
sample by WLI. LT is low temperature (4 C) and RT is room temperature. Note that thicknesses will not
be the same for different samples.
Sample
number
46
47
48
52
53
Etch condition
Small area
WLI (nm)
26%HF,15mA/cm2,RT,100min 157
15%HF,15mA/cm2,RT,120min 173
26%HF,15mA/cm2,RT,120min 168
26%HF,15mA/cm2,LT,120min 155
26%HF,30mA/cm2,LT,120min 187
Reflectance
measurement(nm)
157
170
169
153
186
Full area
WLI (nm)
272
336
273
429
339
Roughness, rms nm
Plots of calculated rms roughness values vs. depth from the reflectance measurements
for the samples that were analyzed with
WLI are shown in Fig. 12. The data
obtained for 26 % HF show a remarkable
consistency. It seems the roughness
15%RT 15 mA
160
evolution does not depend on current
density or temperature. The full area WLI
26%RT/5 deg/15 and 30 mA
data do not follow the same trends,
80
however, indication that the larger period
roughness is affected by the parameters in
a different way. Perhaps there is a
maximum correlation length over which
0
saturation of roughness is achieved, while
for distances larger than this there is little
0
80
160
correlation between diffusion constants in
PS layer thickness, m
the electrolyte or local hole concentration
at the interface.
DISCUSSION AND CONCLUSION
There seems, as shown, as if there is only a very narrow region of parameter space
where an increase in electrolyte viscosity by addition of glycerol is beneficial for
PS/substrate interface roughness in p+ PS, and even this is different for room temperature
and low temperature. All the parameters studied may affect the viscosity of the
electrolyte, or rather reaction species diffusion, as has been suggested (7,8). However, it
seems a too high diffusion constant increases roughness and the same for a too low
constant, at least in the case of p+ PS. The reduction of roughness seems to be limited
to a factor of maximum two for fairly thick layers. Substrate quality is of importance, as
the roughness discussed here is more or less completely dependent on dopant
distribution. It seems more difficult to control roughness of p+ PS compared to what has
been published on p and p- PS.
The laser reflection interference roughness measurement method presented in this
paper gives a good estimate of the roughness rms value within the laser beam spot,
whether this value is representative for the whole sample depends on the spatial
wavelength of the roughness. In the case of some of the p+ Si wafers used in this study,
dopant variations with several wavelengths are present in WLI measurements, at least
with wavelengths of about 200 m and 1 mm. This may explain some discrepancies or
unexpected results within the presented data. PS/substrate roughness on a smaller scale
(micro-roughness) and low rms value striation roughness are undetectable with the
presented method. This is in agreement with other reports stating that microscale
roughness is very small at a p+ PS/substrate interface and that specular reflection
methods are reliably for rms roughness larger than 10 nm. There are both similarities and
differences between roughness progression in p+ PS and other p-type PS. There are
indications of a saturation regime for some of the samples, where the rms roughness
value increases little more with PS layer depth, while for others a saturation regime was
not reached within the measurement time/depth.
REFERENCES
1. H. Man-Lyun, K. Jae-Ho, Y. Sung-Ku and K. Young-Se, IEEE Phot. Tech. Lett., 16, 1519
(2004).
2. V. Mulloni, L. Pavesi, Appl. Phys. Lett., 76, 2523 (2000).
3. G. Vincent, Appl. Phys. Lett., 64, 2367 (1994).
4. M. G. Berger, C. Dieker, M. Thnissen, L. Vescan, H. Lth, H. Mnder, W. Thei, M. Wernke
and P. Grosse, J. Phys. D, 27, 1333 (1994).
5. G. Lrondel, R. Romestain and S. Barret, J. Appl. Phys., 81, 6171 (1997).
6. G. Lrondel, P. Reece, A. Bryant and M. Gal, Mat. Res. Soc. Symp. Proc., 797, 15 (2004).
7. S. Setzu, G. Lrondel and R. Romestain, J. Appl. Phys., 84, 3129 (1998).
8. M. Servidori, C. Ferrero, S. Lequien, S. Milita, A. Parisini, R. Romestain, S. Sama, S. Setzu
and D. Thiaudire, Solid State Comm., 118, 85 (2001).
9. X.G. Zhang, J. Electrochem. Soc., 151, C69 (2004).
10. H. Davies, Proc. Inst. Electr. Eng., 101, 209 (1954).
11. H.E. Bennett and J.O. Porteus, J. Opt. Soc. Amer., 51, 123 (1961).
12. I. Filiski, Phys. Stat. Sol. (b), 49, 577 (1972).
13. E. Steinsland, T. Finstad and A. Hanneborg, J. Electrochem. Soc., 146, 3890 (1999).
14. Z. Gaburro, C.J. Oton, P. Bettotti, L. Dal Negro, G. Vijaya Prakesh, M. Cazzanelli and L.
Pavesi, J. Electrochem. Soc., 150, C281 (2003).
15. M. Thnissen, M.G. Berger, S. Billat, R. Arens-Fischer, M. Krger, H. Lth, W. Thei, S.
Hillbrich, P. Grosse, G. Lerondel and U. Frotscher, Thin Solid Films, 297, 92 (1997).
16. P.Y.Y. Kan, S.E. Foss, T.G. Finstad, Phys. Stat. Sol. a/c (in press).
Chapter 5
Filter fabrication
The goal of producing multilayer structures, or refractive index modulated
structures, by porosification of silicon, is to add the possibility of controlling
light, its spectral composition, phase, and movement, through a device.
In this thesis the focus of the device fabrication is on the control of
the spectral component of incident light, while many of the findings and
techniques may be used for other elements within silicon photonics.
The principles of optical filter design has been described earlier in Chapter 3.
From these principles we may draw the conclusion that a good refractive
index control, layer thickness control, and maximum interface smoothness
is imperative to obtain filters of good quality. In PSM fabrication this
translates to control of porosity with depth and time, control of etch rate
with depth and time and minimization of interface roughness. To a certain
degree the control, or at least a knowledge of, the microtopology of the
porous structure and the relation to refractive index is also necessary as
discussed in Sec. 3.1.
A specific goal of this thesis has been to obtain good quality IR optical
filters of different kinds in PS. A parallel goal has been to use the flexibility
of PSM fabrication to extend the usability of the standard filter. In the
case of graded filter structures the obtained quality of the filters depends
strongly on the level of understanding of flat filter fabrication in PS as some
quality will be lost to the extra design freedom obtained. Ideally, a filters
spectral features will become sharper with an increasing number of layers.
This assumes no or very weak absorption. With PS, this is possible in the
IR, especially for wavelengths around 1500 nm where the absorption is at
a minimum.
To list some of the points we must take care of to be able to fabricate good
quality PSMs:
Porosity must be known.
Etch rate must be known.
97
98
PS morphology should be known.
Change in parameters/factors during etching (as a function of time,
depth, etch area geometry, structural history) should be known.
A model for calculating the effective refractive index as a function of
porosity (and morphology) must be known.
The transition of effective refractive index across an interface between
two layers should be understood and controlled.
The roughness of an interface should be understood and controlled.
The effect of chemical etching of the porous structure should be taken
account of.
A good understanding of the etching process is helpful; where and how
(only pore tip or partly up the walls of the pore, oxidation, passivation by hydrogenation of walls etc., the effect of incident light during
etching and bubble formation).
Most of these points have been or will be addressed in this thesis. Those
points not addressed are assumed to play relatively small roles in the overall
results, however, they are still likely to be measurable.
5.1
where P (j) and P0 are the current density dependent and constant contribution porosities respectively, j is the current density and and are
fitting parameters. Porosity vs. current density data with a fit using Eq. 5.1
is shown in Fig. 5.1. The data was obtained from etching p+ samples in
13.3 % HF for 60 seconds at room temperature. The fit parameters in this
case are P0 =15, =6.5 and =1.6. A similar approximation to the etch
99
rate versus current density may be obtained by using the simple model of
the PS formation described by Eq. 2.3 with porosity as a parameter. By
assuming the valence to be constant with current density and using it as a
fitting parameter, a relatively good fit of the etch rate may be obtained.
Porosity (%)
40
30
Equation: P1+P2*ln(x/P3+1)
Chi^2/DoF
R^2
20
P1
10
20
40
= 0.29219
= 0.99545
15
P2
6.49619
0.38145
P3
1.60483
0.33501
60
80
100
120
5.2
A good quality IR filter designed with sharp spectral features may reach
a thickness of 50 m or more. With an average etch rate, from Fig. 4.13,
100
1.0
Reflectance (abs.)
0.8
0.6
0.4
0.2
0.0
1000
1200
1400
1600
W avelength (nm)
5.2.1
Effect of HF diffusion
101
and the concentration gradient. The valence is sensitive to current density,
temperature and HF concentration. The diffusion coefficient function is
likely to be dependent on the electrolyte properties, e.g. viscosity, as well
as on the porous structure.
In this discussion a couple of factors are not taken into account, namely the
effect of a finite electrolyte reservoir and the behavior of other electrolyte
species, especially those resulting from the reactions at the etch front. A
simplification in a discussion like this would be to assume that the HF
concentration at the surface of the PS is constant, however, with a finite
reservoir without stirring this is not likely to be the case. A constant surface
concentration of HF could probably be used as a first approximation in a
model, however. The behavior of other electrolyte species will most likely
be linked to the HF concentration and not be rate limiting in themselves. In
an empirical model it should be possible to only use the HF concentration
and diffusion coefficient as the electrolyte parameters.
We may assume that the changes in etch rate and porosity with time already
shown in the constant current plots are only due to changes in the electrolyte
composition at the etch front. Thus, according to the discussion above, it
should be possible to correlate each porosity value at any given time to a
starting porosity for a certain HF concentration. However, to project the
time evolution of etch rate and porosity, the microtopology must also be
known, as this will influence the HF diffusion.
In Fig. 4.9 the complicating effect of different microtopologies on the porosity evolution for otherwise equivalent parameters is shown. This has a significant effect on the etching of multilayer structures. The effect of changing
the current density from high to low will be twofold: the concentration at
the etch front will most likely be lower than if the layer had been etched
at the same low current density for the same amount of time, this is due
to the higher usage of HF. However, the concentration also depends on the
diffusion which will be higher because of a steeper gradient and because the
pores are larger. The diffusion to the etch front will be higher compared to
a constant low current density etch which means that the evolution of the
following etch will be different.
To get an idea of the effect this drift in porosity and etch rate will have on
PS optical filters we may use the thick Bragg mirror mentioned above as
an example. Say that we make one filter where the current densities used
for both high and low refractive index are constant throughout the etching
assuming that there is no drift in etch parameters. For a layer etched at
38 mA/cm2 to get a porosity of 80 % (n=1.32) for 6 s to get a thickness of
200 nm, at the top of the filter, a layer etched with the same current density
and etch time after 30 min of etching will give a layer of about 90 % porosity
(n=1.14) and a thickness of 190 nm, shifting the wavelength fulfilling the
Bragg condition Bragg = 4nd (Bragg wavelength) by an amount given by
102
(assuming small changes):
center = 4 (n d + n d)
(5.2)
Inserting the above values gives a shift of about 200 nm. When the Bragg
wavelength shifts from front to back in a Bragg mirror the resulting reflectance band broadens and chirps.
5.2.2
Effect of temperature
103
etching of PSMs is important. These calculations only hold for the case of
15 mA/cm2 current density with the described electrolyte. The temperature dependence of PS etching will most likely depend on the electrolyte
composition and the current density.
70
Porosity (%)
-1
2.0
1.5
23 C
60
4 C
23 C
50
4 C
40
1.0
20
40
60
80
100
120
Time (min)
20
40
60
80
100
120
Time (min)
As current passes through the electrolyte and the sample, resistive heating
will occur. This will in the same way as above introduce changes in etch
rate and porosity, but in this case the changes occur during etching. The
temperature was measured together with the data shown in Figs. 4.104.13
showing the temperature change due to resistive heating and ambient temperature variations. In Fig. 5.5 the temperature profiles measured during
constant current etching with 70, 30, and 5 mA/cm2 in 15 % HF with 10 %
glycerol, at room temperature are shown. One would expect that the temperature change would be dependent on the current density with resistive
heating. This is observed in Fig. 5.5 with the 70 mA/cm2 etching increasing
in temperature from 18 C to 23 C, while the 30 mA/cm2 etching has a
fairly stable temperature. The decrease in temperature for the 5 mA/cm2
etching is most likely due to a decrease in ambient temperature. This etch
current dependent temperature change is likely to have a significant impact
on optical filter characteristics. The obtained temperature evolution will
depend on the current profile needed for each specific PSM. Following the
discussion relating to temperature differences between etches, a change in
temperature from start to finish of 5 C during etching of a Bragg mirror
could result in a substantial chirping or broadening of the filter optical
104
response due to a shift of 140 nm in the Bragg wavelength between front
and back.
24
Temperature ( C)
22
20
18
20
40
60
80
100
Time (min)
5.2.3
120
140
Chemical etching
Chemical etching has been briefly discussed in Sec. 4.2.1. The change in
refractive index in p+ Si due to this effect should be small, but for very
good quality PS filters in p+ Si it may still be significant. To take chemical
etching into account, the porosity profiles obtained by the in situ reflectance
method must be calculated with this in mind. When calculating a current
profile for filter etching, an iterative process may be adopted where the
a profile is calculated disregarding chemical etching, the total time each
layer is in the etchant is calculated and subsequently the change in porosity
due to chemical etching. Next, a new profile is calculated attempting to
counteract the porosity change due to chemical etching.
Assuming that the oscillation in the signal shown in Fig. 4.4 is a result of
chemical etching, we have in this case (4.3 C, 30 mA/cm2 and 26 % HF
) that the average porosity of the film changes from 72.6 to 73.2 % over
12.5 min. For simplicity we assume that the corresponding refractive index
change (in air) is independent of porosity and constant, hence we have a
constant refractive index change due to chemical etching of 2 105 s1 .
The topmost layer in a Bragg mirror, similar to the one discussed above,
will then have a change in refractive index due to long exposure (30 min)
to the electrolyte. This change would be about 0.036 in this case. This
would shift the reflection band center wavelength about 7 nm. Chemical
etching has been reported to depend strongly on the substrate used as the
obtained microtopology will change with doping. The dependence on HF
concentration is also significant [86] with higher concentrations resulting in
less etching. There will most likely be a temperature dependence as well.
105
5.3
Etch calibration
To improve the control of porosity and layer thickness during filter etching
compared to the basic filter etching discussed above in Sec. 5.1, the constant
current porosity and etch rate curves will be used to take into account the
drift of these parameters with time.
A first approximation to a complete calibration of a current profile may be to
assume that the observed drift in the constant current plots is independent
of etch history or microtopology and dependent only on time and current
density. Thus the needed current density at a given time for a given porosity
is obtained by taking a slice from the porosity constant current plot at the
given time and interpolating the porosities to get a porosity vs. current
density plot from which the current density is obtained. To obtain a current
density profile for filter etching, this is done for any number of discrete time
steps depending on the wanted time resolution. The necessary duration at
each porosity depends on the corresponding etch rate which is found in a
similar way. When the current density is found, this is used to find the
etch rate in an interpolated etch rate versus current density plot. For each
time step the total etched thickness and layer thickness is calculated. When
the designed layer thickness is reached, the porosity is changed according
to the design. By using small enough time steps a good approximation to
the designed refractive index profile is reached. However, there will to some
degree always be discretization errors with this method.
A porosity versus current density versus time plot for the 15 % HF, 10 %
glycerol electrolyte at room temperature is shown in Fig. 5.6. The data are
the same as in Fig. 4.11 but interpolated and smoothed. Comparing with
Fig. 5.1, it is clear that the porosity versus current plots for given times are
not as smooth. This is most likely due to the observed variation in ambient
temperature as well as resistive heating during etching.
One way to take into account the etch history when calculating the current
profile for a filter, is to shift the time axis in the constant current data
106
in Figs. 4.104.13 with an amount depending on the current density. We
may assume that the high current density etch will be the controlling factor
for HF concentration at the etch front. If the etching starts with a high
current density etched layer, the etch front HF concentration at the end of
this etch will correspond to a concentration at a much later time for a lower
current density etch due to greater HF usage, hence the time axis is shifted
compared to the constant current plot. We may assume that the change
in etch front concentration is negligible for the low current etch, hence the
etching at the next high current density layer picks up at the same time
the last high current density etch stopped. A change in concentration due
to the low current etch may also be taken into account by shifting the
time axis of the high current density etch correspondingly. The necessary
added shift may be found by fitting the reflectance data for a PSM structure
etched assuming no drift in etch parameters with a model describing these
effects, where the time axis shift is a fitting parameter. In this conceptual
model it is assumed that etch front HF concentration is always decreasing,
however this may not be correct in all situations. Changing from high to
low current density may result in an increase in HF concentration due to
higher diffusion than usage. This calculation has not been implemented for
current profile calculations, but, as will be shown below, the idea has been
used to understand some measured reflectance spectra.
In all the examples and discussions above the emphasis has been on discrete
filters. This is only due to the ease of illustration, the same arguments apply
to the case of an inhomogeneous, e.g. rugate, filter, especially when this is
approximated by a number of discrete layers.
5.4
5.4.1
Prepared filters
Reflectance measurement setup
107
To obtain an optimal wavelength resolution, a narrow slit was used at the
output of the monochromator. Also, to reduce the contribution of light
at different angles due to a diverging reflected beam and to reduce the
measured spot size as the filters at times were somewhat inhomogeneous, a
narrow slit at the monochromator input was used. The narrow slits together
with the beam narrowing by the iris resulted in a fairly low light intensity
reaching the detector. For some measurements this is observed as a nonlinear intensity response of the detector with higher normalized reflectance
than expected for wavelengths outside the optimal wavelength range of the
detector.
5.4.2
Reflectance analysis
5.4.2.1
Discrete filters
In Fig. 5.8 the calculated current profile of a Bragg reflector using the
discussed calibration is shown. The reflector is designed for a wide reflection
band at a center wavelength of 1300 nm at normal incidence with 10 layer
pairs where a pair consists of one 60 %, 175 nm layer and one 83.75 %,
260 nm layer, nominally. The resulting reflected spectral characteristics of
the etched reflector measured at 45 is shown in Fig. 5.9. In the same figure
the calculated spectrum based on the design is also plotted. This calculation
is based on the assumption of equal amounts of s- and p-polarized light.
Interface roughness is not taken into account.
The fit between the calculated spectrum and the measured spectrum in
Fig. 5.9 is quite good. As indicated in the plot, the measured reflectance
band is about 80 nm shifted towards the red compared with the calculated
(designed) peak. The most likely causes of this are the effects discussed
above, including the potential error in the effective medium function used
(see Sec. 3.1), a possible difference in ambient temperature at the time of
108
Current (mA)
60
40
20
0
50
100
150
Time (s)
Reflectance (abs.)
1.0
0.8
0.6
0.4
0.2
0.0
1000
1200
1400
W avelength (nm)
1600
etching compared to the calibration data and the use of a simplified model
for calibration. A red shift indicates optically thicker layers; physically
thicker and/or lower porosity. The observed shift of about 80 nm could be
accounted for by a difference in porosity of about 5 % (absolute difference)
compared to the designed value. This number compares well with the discussion in Sec. 3.1 and above. There are other features in the spectrum
indicating that the current profile is not perfectly adjusted for the effect
of electrolyte species diffusion. A distortion in the spectrum may also be
caused by the striation induced lateral inhomogeneity of the filter if the
beam spot is large enough, which is the case with the spot size of about
1 mm used for most of the reflection measurements.
Figures 5.2, 5.10 and 5.11 compare reflection spectra from Bragg reflectors
designed to reflect at 1500 nm. The spectrum in Fig. 5.2 is from a structure
consisting of 10 pairs of layers with nominally 53 % and 83 % porosities and
thicknesses of 183 and 301 nm, respectively, etched without time calibration.
The peak is shifted about 80 nm towards the red from the designed peak
109
which is similar to Fig. 5.9. The fit between the calculated and the measured
spectrum is quite good. This is expected as the filter is rather thin, about
5 m, so the upper layer pair is not much different from the lowest layer pair.
In Figs. 5.10 and 5.11 the reflectors measured contain 40 layer pairs and one
may clearly see tendencies of drift in layer optical thickness within the layer
stack. In Fig. 5.10 the spectrum from a filter etched without calibration is
shown. In Fig. 5.11 a filter etched with current density adjusted for porosity
and etch rate drift is shown. Note that the calculated spectra in these two
figures take into account a 70 nm PS-substrate interface rms roughness and
an interlayer interface rms roughness of 15 nm. These numbers are used
based on measurements presented in Paper III. The sample etched with
adjustments for drift show a pronounced broadening indicating a possible
over-adjustment. These three spectra are obtained with unpolarized light.
1.0
Reflectance (abs.)
0.8
0.6
0.4
0.2
0.0
1000
1200
1400
1600
W avelength (nm)
To evaluate the structure of the etched filters and try and quantize the
amount of drift and over-compensation, a fitting procedure was used. The
measured reflectance spectra were fitted to the calculated spectra of a simple model of the structure incorporating a linear drift effect. As most of the
layer information is in the optical thickness, the variation in layer optical
thickness was parameterized by a variation in physical layer thickness. By
doing this, some information is lost because individual layer reflectances
are not changed due to constant refractive indexes. The resulting fit, using
the Levenberg-Marquardt algorithm as implemented in the IMD software
for thin film calculations by David L. Windt, for the structure measured in
Fig. 5.10 is shown in the same figure with open circles. The resulting optical
110
Reflectance (abs.)
1.0
0.8
0.6
0.4
0.2
0.0
1000
1200
1400
1600
W avelength (nm)
53 % layers
1600
1500
1400
83.75 % layers
1300
20
40
Layer index
60
80
Figure 5.12: The resulting bragg wavelength (optical thickness4) of the layers
of the filter corresponding to
Fig. 5.10 after fitting a simplified model to the measured
spectrum. The drift in porosity and etch rate with time
is assumed to result in a linear change in optical thickness
with depth in this model.
thickness for each layer, transformed to Bragg wavelengths (4optical thickness), is shown in Fig. 5.12. The resulting spectrum fits the measurement
quite well. Note that adjacent averaging is used with the fitted spectrum
to simulate the effect of wavelength resolution limitations in the monochromator. The layer optical thicknesses shown in Fig. 5.12 show that there is
a marked drift in optical thickness which may be caused by one or both of
etch rate and porosity drift. The optical thickness of all the layers should
have been the same according to the design, in this case corresponding to
a Bragg wavelength of 1500 nm, but the resulting optical thicknesses differ
quite markedly between high and low porosity layers. This may be that
the starting point of the etch was incorrect, i.e. that either the calibration
curves were slightly off true value because of e.g. temperature differences, or
that the effective medium function introduced an error as discussed earlier.
The same procedure was performed for the filter measured in Fig. 5.11. The
111
1800
Figure 5.13:
Same as
Fig. 5.12 for the case where the
current profile was calibrated
(Fig. 5.11).
83.75 % layers
1600
1400
60 % layers
1200
1000
20
40
60
80
Layer index
resulting fitted spectrum is shown in the same figure with open circles. Not
all features seem explained, but the widening and the general band edge
gradients seem explained quite well. Figure 5.13 shows the layer Bragg
wavelengths of the fitted spectrum. It is clear that the calibration procedure
has had a significant effect on the structure. The low porosity layers seem
most affected by the calibration, while the high porosity/high current layers
seems well adjusted and nearly constant throughout the structure. This
may indicate that a time shift of the calibration curves when designing a
filters is appropriate. The etch rate decrease or porosity increase with time
for the low porosity layers seem much stronger than that taken into account.
5.4.2.2
Rugate filters
Rugate filters were fabricated in the same manner with a time calibration
done on the current profile. In Figs. 5.14 and 5.15 the measured and calculated reflection spectra are shown for two fabricated rugate reflectors. The
sample used for Fig. 5.14 was designed for a narrow reflection band at 600
nm. Index matching was used while no apodization was used. The porosity
varied from 49 % to 53 % nominally between the index matching regions.
The filter contained 65 periods in the refractive index profile resulting in a
thickness of about 10 m. A low temperature of about 6 C was used along
with an aqueous electrolyte consisting of 26 % HF and ethanol. The low
temperature was used in an attempt to minimize roughness as discussed in
Paper III. The values of the measured data are only indicative as a baseline was subtracted from the data as very a low light intensity resulted in
a non-linear detector response. The same trend as discussed for the other
measured spectra is present in Fig. 5.14 with the measured spectrum redshifted compared to the calculated spectrum, however the band shapes are
quite similar, with a measured full width at half maximum of 10 nm, indicating little drift of porosity and etch rate. The effect of interface roughness is
clearly seen comparing the two calculated spectra; one taking into account
interface roughness, one without roughness. In Fig. 5.15 the reflectance of a
thick IR rugate filter is measured. This filter was designed with a medium
112
width reflection band at 1550 nm with 75 periods in the refractive index,
both index matching and quintic apodization were used. The total thickness was about 30 m. A time calibration of the current profile was done in
this case also. This filter was etched at low temperature (4 C) with a 26 %
HF electrolyte. A calculated spectrum based on the design is shown. It is
evident that the refractive index profile is not optimal as the peak is much
too wide and non-uniform. To test the suggestion that this widening is due
to a drift in the layer optical thickness as discussed, a calculation of the
reflection spectrum with a linear increase in the period was done. This is
shown as a dashed line in Fig. 5.15. The correspondence is good indicating
an over-adjustment due to a non-optimal calibration. In both Figs. 5.14
and 5.15 the calculated spectra shown in whole line are without roughness
taken into account, while the calculations shown in dashed lines are with a
rms roughness at the interface of 70 nm and an interlayer rms roughness of
15 nm.
Reflectance (abs.)
0.4
0.3
0.2
0.1
0.0
500
600
700
800
W avelength (nm)
900
1000
In Fig. 5.16 the measured reflectance spectrum of a three band rugate spectrum is shown. The filter was designed as a narrow band pass filter with a
wide blocking band on both sides of the pass band. This was done similar
to the design calculated in Fig. 3.14 with three reflectance bands very close
to each other. The calculated current profile, adjusted for drift in porosity
and etch rate, is shown in Fig. 5.17. The regions in the current profile corresponding to the different bands overlap in such a way as to minimize the
thickness without allowing the total refractive index to go over a certain
threshold value. The lower current values reach a limiting threshold several
places whereby the current is clipped. This procedure introduces some features in the spectral characteristics, however, with little clipping the result
is acceptable. As can be seen in Fig. 5.16 the main spectral characteristics
in the designed filter are recognizable in the measured spectrum. Again,
due to roughness and overcompensation for drift, the result is not optimal.
Note that the calculated spectrum in this case is only for s-polarized light.
113
Figure 5.16:
Reflectance
measurement of a three peak
rugate filter (filled squares).
The filter was designed to have
a narrow transmittance band
as can be seen in the calculated spectrum (open circles). The calculated spectrum is for s-polarized light
while the measurement is unpolarized. The pass band is
present in the measured spectrum, but much wider than designed due to drift in porosity
and etch rate with depth and
interface roughness.
1.0
Reflectance (abs.)
0.8
0.6
0.4
0.2
0.0
1000
1200
1400
1600
W avelength (nm)
1.0
0.8
Reflectance (abs.)
Figure 5.15: A thick rugate reflection filter is compared with calculated spectra.
The design was for a peak
at 1550 nm, using 75 refractive index periods with porosities varying between 30 % and
53 %. The etch current profile was adjusted for drift resulting seemingly in an overcompensation. A calculated
spectrum incorporating a shift
in design peak wavelength is
shown as the dashed red curve.
This compares well with the
measured spectrum.
0.6
0.4
0.2
0.0
1000
1200
1400
1600
W avelength (nm)
Etch brakes have been used quite successfully by Billat et al. [87] and Reece
et al. [59]. The etch breaks allows the electrolyte at the etch front to
regenerate, avoiding problems caused by diffusion of the electrolyte species.
However, the etch time of a filter increases substantially compared with a
continuous etch as typically break times are in the order of 10 to 20 times
the time it takes to etch the thickest of the layers in the layer pair [88]. This
results in an etch time of about one hour for a 20 layer pair Bragg stack
with peak wavelength at 1300 nm. By increasing the etch time this much
the effect of chemical etching must be considered, even for p+ samples.
114
50
Current (mA)
40
30
20
10
0
500
5.4.2.3
Graded filters
The setup for etching graded filters is explained in Sec. 2.5. A detailed
analysis of a series of graded rugate reflection filters is presented in Paper
V, while some possibilities of this technique will be discussed in Chap. 6.
The measured spectra from a near-infrared graded rugate reflection filter
is shown in Fig. 5.18. These data are from measurements at an incident
angle of 22 . The current profile used was based on a filter designed for a
reflection peak at 1000 nm, porosities varying from 75 % to 85 %, and a
total thickness of about 10 m. A calculated reflection spectrum based on
this design is shown in the figure. The lateral voltage used during etching
was 1 V, resulting in a shift of the peak wavelength of about 35 nm per mm
across the filter.
1.0
Pos.3
Pos.4
Pos.5
2mm
5mm
7mm
Reflectance (abs)
Pos.2
0.6
Designed
1mm
0.8
Edge
0.4
0.2
0.0
500
600
700
800
900
Wavelength (nm)
1000
1100
Paper IV discuss some of the effects the grading of the filter have on the
optical response using a simple single ray model. Particularly if the graded
layers introduce an angular dispersion in the reflected beam and how the
spectral response change with graded layers. The conclusion seems to be
115
the errors introduced by an angular dispersion in the incident beam overshadows any effect of the graded layers.
An indication of the regularity of the periods of the fabricated filters may
be obtained by high resolution field emission SEM (FESEM) images. Figure 5.19 shows a cross-section FESEM image of the reference (non-graded)
rugate reflection filter presented in Paper V. The average image intensity
profile taken across the image is overlayed in yellow. A sine like profile is
clearly recognizable. The obtained sine profile is not accurate enough to
show a difference in period from top to bottom which seems present when
considering the chirping of the spectral responses of the filters in the
paper.
5.5
Several potential and known causes of errors in the PSM structures have
been discussed. Some possible remedies for these problems are suggested
in the following:
stirring to keep sample surface electrolyte composition constant as
well as reduce temperature gradients/changes, pump induced flow for
the same reasons or bubling of e.g. nitrogen gas.
temperature control of electrolyte (main volume) and of sample.
introduce breaks in etching to regain original electrolyte composition
in active region as briefly discussed.
the effect of bubble formation in pores during etching may be reduced
by using a different wetting agent (may be improved further compared
with ethanol).
116
(micro-) roughness may be reduced by a limited viscosity increase
(lower temp or addition of glycerol) as indicated in Papers II and III.
use a cathode integrated in the etch cell to keep the position constant
relative to the sample.
Paper IV
S.E. Foss and T.G. Finstad
Multilayer interference filters with non-parallel interfaces
Proceedings of the Nordic Matlab Conference,
Copenhagen, Denmark, 2003
IV
Incident
beam
Graded PS film
2
3
4
5
6
7
Si-substrate
Paper V
S.E. Foss and T.G. Finstad
Laterally graded rugate filters in porous silicon
Mat. Res. Soc. Symp. Proc., 797, W1.6.1 2004
Mat. Res. Soc. Symp. Proc. Vol. 797 2004 Materials Research Society
W1.6.1
INTRODUCTION
The process of discretely varying the current density during etching of porous silicon (PS) to
obtain thin layers with different refractive indexes was first reported by Vincent [1] and Berger
et al. [2] in 1994. With this method Bragg reflectors and Fabry-Perot filters are now routinely
made. By applying an in-plane voltage across the sample during etching, Hunkel et al. have
shown that it is possible to produce PS discrete filters with laterally dependent filter
characteristics [3,4]. In addition to the discrete layer filters (i.e. Bragg and Fabry-Perot) one may
also realize structures in PS that are uniquely simple to this system. By varying the current
density continuously and periodically, the refractive index into the PS layer will vary accordingly
[5]. The refractive index may be calculated from the porosity by the effective medium
approximation. With this approach one obtains reflecting rugate filters which may have narrow
reflection peaks, no higher order harmonics and suppressed sidebands. For an overview of rugate
filter theory, see Bovard [6]. Here we report on rugate filters with laterally varying
characteristics which have been made for the visible to near infrared optical spectrum.
EXPERIMENTAL DETAILS
The Si substrate used was 0.018 cm B-doped, p-type Czochralski-grown single crystal
with <100> orientation polished on both sides. The HF-based solution used for etching consisted
of 1:2 HF(40 %):ethanol. The etching current was controlled by a computer, in this case with a
time step of 1 second which sufficiently reproduced the refractive index sinusoid. The used
current profile is shown in Fig. 1. The etching occurs mostly at the pore tips which is why
porosity may be modulated by the current. There will also be a small chemical etching which is
W1.6.2
dependent on time. This is corrected for by etching the deeper part of the layer with a slightly
higher current.
The starting point for the design was a 1 cm diameter, circular, non-graded rugate filter with
a reflectance peak-wavelength at 738 nm measured with light incident at 24 deg. The filter was
made relatively thin with 23 periods of the porosity sinusoid over 5 m. Etching was done with
current density varying between 9.4 mAcm-2 and 19.7 mAcm-2, which corresponds to refractive
indexes at the peak wavelength of 1.78 and 1.48 respectively. Index matching for the air-PS
interface and the PS-substrate interface was employed to reduce sidebands. This may be
observed in the beginning and end of the current profile of Fig. 1 where there are large slopes.
A schematic of the etching setup is shown in Fig. 2. A lateral gradient in porosities and etch
rates was obtained by applying a constant in-plane voltage up to 1.5 V between two contacts on
the sample back side while etching with the same current profile as for the starting point design.
By doing this the local current density varies. Ohmic contacts were made from evaporating Al on
the samples followed by a short heat treatment. The resistance between the contacts varied
between 0.6 and 0.7 for the different samples. Contact resistance is a substantial fraction of
this, so good, reproducible contacts were important to achieve control of the potential drop
within the sample.
Reflectance measurements were conducted with a 0.275 m focal-length monochromator with
a Si-detector. The focused probe beam had a diameter of less than 1 mm and was directed at the
filter at an angle of about 24 deg. Both the size of the beam and the angle contribute to a
widening of the reflectance peak to some extent compared to the peak from the non-graded filter.
The effect depends on the gradient of the filter. Reflectance spectra are only plotted from 600 nm
to 1100 nm because of the combination of the detector and diffraction grating used in the
monochromator.
For the optical images a microscope with a mounted digital camera was used. Several images
were stitched together to get a better overview. To show the striation effects on the filter
W1.6.3
reflectance clearly in these images, the RGB colors were split into separate images and only the
green was used for analysis as this had the largest contrast between striation minimum and
maximum.
A white light interferometer, WYKO NT-2000 by Veeco, was used to measure surface
topography profiles of the samples after the PS was stripped away with a concentrated NaOH
etch. This instrument has a large dynamical range and can measure m and nm height
differences in the same measurement. Even with the gradient in the samples present,
measurements showing the ridges caused by striations, in the order of 100 nm high, were
possible.
W1.6.4
dependence on current density [3]. Therefore the amplitude and period in the refractive index
sinusoid will only be optimal at one point along the filter surface. All the graded filters show
similar behavior. A decrease of amplitude between samples with increasing grading is also
observed. This is most likely caused by the finite spot size of the probe beam. When this covers a
too large area compared to the gradient at that point, the resulting measurement will be an
average of spectra and hence the filter window will be smeared out.
Figure 5 shows the peak wavelengths at different positions for a series of filters with
changing in-plane voltage. As expected, the shift increases with the voltage caused by the
increase in the porosity and etch rate gradient. The horizontal line across the plot shows the peak
wavelength of the non-graded filter. The reflectance peak is clearly not at 738 nm for the mid
position for any of the graded filters. Thus the current density does not vary linearly with
position. This is as expected considering the geometry of the etch setup.
Although the shift in peak wavelength seems close to linear, depth profiles of the PS layer
made by white light interferometry after stripping, shown in Fig. 6, show tendencies towards
non-linearity. This is also supported by [7] where Bohn and Marso describe an equivalent circuit
model for the etch situation. A non-linear current density distribution will result in a varying
position of the crossing point between the peak shift curves and the non-graded peak wavelength.
One would, however, expect that at some point the filter would have the characteristics of the
non-graded filter as the total current is the integral of the local current density at all positions on
the filter area. This is true for the filters of Fig. 5 except for the 0.1 V case. A slight difference in
process parameters is most likely the cause. In Fig. 7 the change in peak shift with applied inplane voltage is shown. The line corresponds to a linear fit which seems appropriate for this
range of voltages. A non-linearity might be expected for higher voltages as the necessary current
0
1.5 V
Depth, m
-2
0.5 V
-4
0.1 V
-6
-8
-10
0
10
12
W1.6.5
reaches amperes, hence heating will probably affect the process. The geometry of the filter has
an effect on the local current density, as may be observed visually. Perpendicular to the gradient
direction reflection colors shift somewhat away from the center line, indicating a non-constant
current density towards the edges.
Visible stripes of slightly different colors are present in all the filters made. These seem
circular and centered around the center of the 4 wafers used for samples. After stripping away
the PS with an NaOH etch these stripes are still present as ridges. Figure 8 shows a compilation
of a light microscope image representing the intensity of the green component from a color
image (to increase contrast) and the subsequent intensity plot of the 2-dimensional height profile
from white light interferometry measurements at the same position after stripping of PS. Peak to
peak height of the ridges are about 100 nm. These ridges indicate locally different etch rates.
Similar results have been reported earlier by Lrondel et al. [8] as likely being caused by
striations, i.e. radially symmetric resistivity inhomogeneities in the substrate due to an
inhomogeneous dopant distribution. One way of smoothing the PS-substrate interface is
suggested by Setzu et al. [9] where etching is done at low temperatures down to -35 C.
CONCLUSION
We have shown it is possible to make a laterally graded rugate filter with good reflectance.
The filter may be improved by optimizing the parameters used, especially by using a smaller
difference between minimum and maximum refractive index and more periods, hence a thicker
PS film. Apodization may also be used to optimize the reflectance characteristics. Striations
causing locally differing etch rates have been shown.
W1.6.6
ACKNOWLEDGMENTS
This work has been carried out under the MOEMS and MEMS research program of the
Research Council of Norway. The authors would like to thank Maaike Taklo Wisser at SINTEF
for help with the WYKO measurements, and H. G. Bohn at Forschungszentrum Jlich, Germany,
for help with the program for designing the rugate filter.
REFERENCES
1. G. Vincent, Appl. Phys. Lett. 64, 2367 (1994).
2. M. G. Berger, C. Dieker, M. Thnissen, L. Vescan, H. Lth, H. Mnder, W. Theiss, M.
Wernke and P. Grosse, J. Phys. D: Appl. Phys. 27, 1333 (1994).
3. D. Hunkel, R. Butz, R. Arens-Fischer, M. Marso and H. Lth, J. Lumin. 80, 133 (1999).
4. D. Hunkel, M. Marso, R. Butz, R. Arens-Fischer and H. Lth, Mater. Sci. Eng. B 69-70, 100
(2000).
5. M. G. Berger, R. Arens-Fischer, M. Thnissen, M. Krger, S. Billat, H. Lth, S. Hilbrich, W.
Theiss and P. Grosse, Thin Solid Films 297, 237 (1997).
6. B. G. Bovard, Applied Optics 32, 5427 (1993).
7. H. G. Bohn and M. Marso, (unpublished report).
8. G. Lrondel, R. Romestain and S. Barret, J. Appl. Phys. 81, 6171 (1997).
9. S. Setzu, G. Lrondel and R. Romestain, J. Appl. Phys. 84, 3129 (1998).
Chapter 6
Porous silicon applications for
MOEMS and passive optics
The techniques and results presented in this thesis may be considered as
part of a toolbox to build novel devices in silicon microtechnology. Some
possibly new and untested ideas will be presented in the following showing some of the many possibilities of porous silicon as an optical material.
The two main areas these ideas describe are passive optical elements and
micro-opto-electro-mechanical-systems (MOEMS). Several different passive
optical elements and applications have been presented in the literature,
some of which have been mentioned in Sec. 2.1. In the area of MOEMS
there are very few reported devices employing PS as a critical material.
The most striking device is the spectrometer by Lammel [89].
6.1
6.1.1
One aim of the presented work has been to prepare graded optical band-pass
filters in PS, which for example could be used in a monolithically integrated
sensor-array system. A fairly simple sensor design could be based on several
separate Schottky barrier sensors on the back side with the graded band
pass filter, either Fabry-Perot or rugate, on the front side. This detector
would be for use in the near- to mid-IR as photons with wavelengths below
the absorption edge of the silicon substrate would be absorbed and would
133
134
not reach the Schottky barrier sensors. The sensors would be in the form of
parallel strips with the length of each strip perpendicular to the filter gradient direction. In this way each strip would respond to a specific transmitted
band of photon wavelengths, with the position of each strip relative to the
filter deciding which band will be detected. The total of the strips would
than give a spectroscopic detector. A schematic drawing of the design is
shown in Fig. 6.1.
To obtain sharp features in the filter spectral characteristics, many layers or
periods are needed resulting in fairly thick filters. This necessitates a very
good control of etch rate, porosity and interface roughness. A standard
Bragg reflector of 50 layer pairs with a peak reflecting wavelength of 1.5 m
would have a total thickness of roughly 20 m. Reasonable filter thicknesses
for more complex rugate filters may reach 50 m. The good control of the
parameters is also necessary to minimize some of the problems introduced
by grading the filter.
The Schottky barrier sensor strips may be fabricated by different metals
depending on the wavelength range of interest. Deposition of Ti, Ir and
Pt with a subsequent annealing produces silicides with low enough work
functions so that photons not absorbed by the substrate may induce a
current. The possible wavelength range of the detector is determined on
the high energy side by the absorbtion edge of the Si substrate, about
1.1 eV, and on the low energy side by the barrier height, which for Al on
p-type Si is about 0.55 eV and for PtSi on p-Si is normally 0.3 eV, but has
been reported for special cases to be as low as 0.13 eV [90]. This results
in a detectable wavelength range from 1.127 m to 2.254 m for Al and
from 1.127 m to 4.133 m or 9.5 m for PtSi. The performance of such a
spectroscopic detector would depend on the filter gradient and number and
width of the sensor strips for wavelength resolution and on the filter size
and Schottky barrier material for wavelength range.
By miniaturizing each element it may be possible to fabricate arrays of
such detectors, with each identical detector detecting a range of discrete
wavelength bands. One would then have an imaging IR spectroscopic device. Instead of basing the sensing on the Schottky effect, one may use
pyroelectric materials, such as BaTi, for heat sensitive detector arrays. It
is possible to use PS on the contact side also. Raissi and Far reported in
Ref. [91] that electroplating of Pt within the pores of PS with a subsequent
anneal produces Schottky barrier diodes with low barrier height and high
efficiency.
6.1.2
2D photonic crystal
A subject of much research recently has been photonic crystals. A multilayer thin film optical filter can be said to be a 1D photonic crystal with a
photonic band gap, or forbidden band, i.e. reflection band. 2D and even
135
3D photonic research has been proposed and tested for applications such
as waveguides in photonic circuits and resonators for enhancing LED emission [92]. Macro PS has been used for 2D photonic crystals [93] due to
the well ordered, high aspect ratio pores. Most 2D photonic crystal structures are based on two materials, air and a dielectric. This is mostly due
to fabrication limitations, but also the high contrast in refractive index obtainable. However, it could be interesting to have a controllable refractive
index contrast, as shown by Weiss in Ref. [22] in the case of 1D photonic
crystals. By fabricating macro-PS with pores filled with micro-PS instead
of air, and filling this again with liquid crystals, a controllable band gap
may be realized for 2D photonic crystals as well. The fabrication of this
structure may be done by micro-PS etching on a masked Si-substrate or by
etching macro-PS under certain conditions [94] obtaining filled macropores.
The obtained photonic crystals could be used as reconfigurable waveguides
where, in principle, each column of liquid crystal filled micropores could be
addressed individually. This would enable switching, modulation, and beam
shaping of the light in the crystal. Potential applications could be within
lab-on-a-chip technology (beam steering) or light sources (beam shaping
and directing). A conceptual drawing of the discussed design is shown in
Fig. 6.2.
136
6.1.3
GRIN optics
By controlling the potential distribution through the sample both temporally and spatially during etching, it is possible to form any conceivable
refractive index geometry within the limits of the etch parameters. One
possibility is to form a graded index (GRIN) lens, a plane parallel centrosymmetric structure with a given radial function describing the refractive
index. By choosing e.g. a quadratic function with the largest refractive
index at the central axis, and etching such that the refractive indexes are
constant through the structure, possibly through the sample, a collecting
lens is fabricated.
As the practical refractive index range is limited to roughly 1.15 to 2.7 (90
- 30 % porosity at 1500 nm wavelength), and the thickness of standard Siwafers is around 500 m, the obtainable focal length will be fairly large.
The brand of SELFOC GRIN lenses uses a refractive index profile of
n2 (y) = n20 (1 y 2 ),
where y is the radial distance and a constant. With 2 y 2
interest the focal length may be given as
f=
1
,
n0 sin (d)
(6.1)
1 for all y of
(6.2)
where d is the thickness of the sample [95]. With = 0.385 and n0 = 2.7 the
focal length is 5 m giving a reasonable focal length for compact detector
137
applications. The circular lens would then be etched through the sample
and have a refractive index along the central axis of 2.7 decreasing out to
the rim to 1.15 at a radius of about 2.5 mm.
The GRIN design may also be used for waveguiding, as is the case for a
certain type of optical fiber. One challenging element of combining microtechnology and optics, usually referred to as micro photonics, is the coupling of light from the transportation level (fiber) to the manipulation level
(chip/device). Aligning fibers to waveguides on an optical integrated circuit (OIC) is difficult as the areas to be aligned have a typical size in the
m range. One way of doing this which has proven quite efficient has been
to etch grooves adjacent to the waveguide-entrance during processing of
the OIC. The fiber will then be centered upon placement and fastening.
However, there will be a substantial loss of signal as there will most likely
be an air gap between the fiber-end and the waveguide-entrance. This
method is also permanent. One potential application where the coupling
of light and OICs is non-permanent is in the processing of lab-on-a-chips.
In this case a specially designed chip may manipulate a physical sample,
e.g. liquid containing DNA, through heat treatment and transport through
micro-conduits. The measurement of parameters of interest may be done by
luminescense measurements at some position on the lab-on-a-chip with an
external excitation source and detector. By integrating some of the optics,
e.g. waveguides, it may be possible to measure more parameters and make
the measurement more selective or sensitive. To make a non-permanent,
robust coupling between a fiber and an optical circuit the GRIN properties
of PS may be utilized. It is also possible to integrate the light sources with
the lab-on-a-chip also, possibly PS LEDs or nanodot LEDs/lasers. In this
case the system will be even more compact and coupling to and from the
chip is avoided.
Waveguides fabricated with PS have been reported by Loni [96] and others [97, 98, 99]. By combining a standard waveguide with a specially designed coupling area, a good coupling with little loss may be possible. A
schematic of the idea is shown in Fig. 6.3. Guiding of light is most often
based on total internal reflection which is possible when light in a medium
is reflected off an interface to another medium with a lower refractive index
at a minimum angle. This is shown in Fig. 6.3 as a darkly colored guiding core (high index) surrounded by a lighter colored cladding layer (low
index). However, to minimize back-reflectance from the air-guide interface
the guide refractive index should match as well as possible the air refractive index. These conflicting requirements may be resolved by gradually
increasing the core index from the coupling area at the surface to the guide.
The curvature of the core in the coupling area should be small as this will
reduce the loss due to non-total internal reflection conditions close to the
surface - the core has a lower index than the cladding here. This design
seems quite robust in that there is a range of acceptable angles and there
are no movable parts.
138
6.1.4
139
oxidized.
6.2
MOEMS devices
Enabling optical elements, like filters, to move on a chip level opens up vast
possibilities within practical applications.
6.2.1
140
of the cavity may be etched to form a PS reflectance filter. With a higher
reflection from the mirrors on both sides of the cavity, the cavity mode
will be sharper, hence smaller shifts in the cavity mode due to membrane
movement may be detected. A sketch showing the principle of the device
is shown in Fig. 6.4b. This type of pressure sensor is being used today as
microphones because of the quick response and high sensitivity.
a)
b)
6.2.2
141
of the array such that the length is perpendicular to scan direction and the
length of the array similar to the width of the filter, hence quite a wide filter
would be necessary. This would increase the inertial mass and reaction time
suggesting that perhaps several filters in parallel could be used. This setup
would enable a multispectral line array detector which could be used for,
e.g., spatially resolved gas detection or environmental monitoring.
Instead of a graded filter it is possible to make binary filters such as reported
by Arens-Fischer et al. [103]. By fabricating an ARC on one end of the filter
and a reflection filter on the other end a switch may be realized. The input
beam to be switched on or off may be provided by fiber, and the output
beam may be coupled to a fiber. This setup should give a very high signal
on/off ratio. The switching rate of such a device should be in the single
digit to double digit kHz following the resonance frequency of reported
comb drive devices [104]. This is not high enough for data package routing
in telecom networks, but should be enough for network reconfiguration and
perhaps other applications.
Figure 6.5: A MOEMS device making use of a graded PS filter. This basic
element may be used in several different applications. The filter element is
coupled to a system of comb-drives which is able to horizontally translate
the filter. One application as suggested in the figure is an imaging spectral
scanner.
6.2.3
142
factor quite low as it would be necessary to have quite long filters compared
to the size of the detector for a practical measurable spectral range. The
benefits would be a very fast, compact and monolithic imaging multispectral detector.
6.2.4
Holographic scanner
(6.3)
143
with the incident angle, , being 75 . Assuming the beam to be scanned
is a monochromatic laser beam of diameter 50 m and that the change of
grating period within one diameter is 20 nm (roughly 2 output difference
between maximum and minimum diffraction period within the beam) the
needed scan length would be about 250 m. This may result in a fairly
quick and compact laser scanner. A schematic drawing showing the graded
grating is shown in Fig. 6.6.
Chapter 7
Conclusion
In this thesis a method for the in situ monitoring of parameters critical
for optical applications during etching of porous silicon has been described.
Data obtained by this method has been used to fabricate different types of
interference based optical filters in porous silicon.
The development of the method for in situ monitoring includes the development of a optical fiber based measurement system composed of an infrared
laser coupled to the dry side of the Si-sample in the etch cell during etching and a detector to measure the intensity of the reflected beam. As the
system is fiber based, it is compact and the system hardware may be at
a distance from any harmful chemicals. The reflected beam contains an
oscillating signal due to interference between the beams partially reflected
off the different interfaces in the porous silicon sample; front side, porous
siliconsubstrate interface and back side. By analyzing the reflected interference signal with a short-time Fourier transform, the instantaneous or
depth dependent porosity is obtained along with the instantaneous or depth
dependent etch rate and porous siliconsubstrate interface roughness.
Information on porosity and etch rate from both gravimetrical measurements as well as the in situ reflectance method has been used to etch both
discrete and inhomogeneous (rugate) optical interference filters in the visible and near-infrared spectral range. By applying a voltage laterally across
the Si-samples during porous silicon filter fabrication, the resulting filters
had a gradient in the filter response in the direction of the voltage drop.
This could be developed into a near-infrared spectrometer.
The porosities and etch rates obtained by the in situ reflectance method
show a very strong dependence on etch time. This affects the filter etching
such that the resulting filter response is non-optimal. Attempts at counteracting this time variation is shown. A discussion of the causes of the
non-optimality of the filter responses is given as well as possible ways of
avoiding the detrimental effects of this time variation which is important
for fabricating infrared optical filters of very good quality.
145
146
Possible uses of porous silicon in other novel applications, both passive and
in conjunction with micro-opto-electro-mechanical-systems are discussed in
the last chapter.
Bibliography
[1] O. Boyraz and B. Jalali. Demonstration of a silicon raman laser. Opt.
Express, 12(21):5269, 2004.
[2] H. S. Rong, A. S. Liu, R. Jones, O. Cohen, D. Hak, R. Nicolaescu,
A. Fang, and M. Paniccia. An all-silicon raman laser. Nature,
433(7023):292, 2005.
[3] A. Irace, G. Breglio, M. Iodice, and A. Cutolo. Light modulation with
silicon devices. In L Pavesi and Dj Lockwood, editors, Silicon Photonics, volume 94 of Topics in Applied Physics, page 361. Springer-Verlag
GmbH, 2004.
[4] A. Uhlir. Electrolytic shaping of germanium and silicon. Bell Syst.
Tech. J., 35(2):333, 1956.
[5] D. R. Turner. Electropolishing silicon in hydrofluoric acid solutions.
J. Electrochem. Soc., 105(7):402, 1958.
[6] Y. Watanabe and T. Sakai. Application of a thick anode film to
semiconductor devices. Rev. Elec. Commun. Lab., 19(7-8):899, 1971.
[7] M. J. J. Theunissen. Etch channel formation during anodic dissolution
of n-type silicon in aqueous hydrofluoric acid. J. Electrochem. Soc.,
119(3):351, 1972.
[8] Y. Watanabe, Y. Arita, T. Yokoyama, and Y. Igarashi. Formation
and properties of porous silicon and its application. J. Electrochem.
Soc., 122(10):1351, 1975.
[9] T. Unagami. Oxidation of porous silicon and properties of its oxide
film. Jpn. J. Appl. Phys., 19(2):231, 1980.
[10] L. T. Canham. Silicon quantum wire array fabrication by electrochemical and chemical dissolution of wafers. Appl. Phys. Lett., 57(10):1046,
1990.
[11] N. Koshida and H. Koyama. Visible electroluminescence from porous
silicon. Appl. Phys. Lett., 60(3):347, 1992.
147
148
[12] O. Bisi, S. Ossicini, and L. Pavesi. Porous silicon: A quantum sponge
structure for silicon based optoelectronics. Surf. Sci. Rep., 38(1-3):5,
2000.
[13] M. I. J. Beale, J. D. Benjamin, M. J. Uren, N. G. Chew, and A. G.
Cullis. An experimental and theoretical-study of the formation and
microstructure of porous silicon. J. Cryst. Growth, 73(3):622, 1985.
[14] C. Pickering, M. I. J. Beale, D. J. Robbins, P. J. Pearson, and
R. Greef. Optical-properties of porous silicon films. Thin Solid Films,
125(1-2):157, 1985.
[15] G. Vincent. Optical-properties of porous silicon superlattices. Appl.
Phys. Lett., 64(18):2367, 1994.
[16] M. G. Berger, C. Dieker, M. Thonissen, L. Vescan, H. L
uth,
H. M
under, W. Thei, M. Wernke, and P. Grosse. Porosity superlattices - a new class of si heterostructures. J. Phys. D Appl. Phys.,
27(6):1333, 1994.
[17] C. Mazzoleni and L. Pavesi. Application to optical-components of
dielectric porous silicon multilayers. Appl. Phys. Lett., 67(20):2983,
1995.
[18] L. Pavesi, C. Mazzoleni, A. Tredicucci, and V. Pellegrini. Controlled
photon-emission in porous silicon microcavities. Appl. Phys. Lett.,
67(22):3280, 1995.
[19] Y. Zhou, P. A. Snow, and P. S. J. Russell. Strong modification of photoluminescence in erbium-doped porous silicon microcavities. Appl.
Phys. Lett., 77(16):2440, 2000.
[20] M. Thonissen, M. Marso, R. Arens-Fischer, D. Hunkel, M. Kr
uger,
V. Ganse, H. L
uth, and W. Thei. Electrical control of the reflectance
of porous silicon layers. J. Porous Mat., 7(1-3):205, 2000.
[21] M. V. Wolkin, S. Chan, and P. M. Fauchet. Porous silicon encapsulated nematic liquid crystals for electro-optic applications. Phys.
Stat. Sol. a, 182(1):573, 2000.
[22] S. M. Weiss and P. M. Fauchet. Electrically tunable porous silicon
active mirrors. Phys. Stat. Sol. a, 197(2):556, 2003.
[23] S. Weiss, M. Molinari, and P. Fauchet. Temperature stability
for silicon-based photonic band-gap structures. Appl. Phys. Lett.,
83(10):1980, 2003.
[24] J. Diener, N. Kunzer, E. Gross, D. Kovalev, and M. Fujii. Planar
silicon-based light polarizers. Opt. Lett., 29(2):195, 2004.
149
[25] M. Thust, M. J. Schoning, S. Frohnhoff, R. Arens-Fischer, P. Kordos,
and H. L
uth. Porous silicon as a substrate material for potentiometric
biosensors. Meas. Sci Technol., 7(1):26, 1996.
[26] V. S. Y. Lin, K. Motesharei, K. P. S. Dancil, M. J. Sailor, and M. R.
Ghadiri. A porous silicon-based optical interferometric biosensor. Science, 278(5339):840, 1997.
[27] G. Lerondel, R. Romestain, J. C. Vial, and M. Thonissen. Porous
silicon lateral superlattices. Appl. Phys. Lett., 71(2):196, 1997.
[28] J. Volk, N. Norbert, and I. Barsony. Laterally stacked porous silicon
multilayers for subquart micron period uv gratings. Phys. Stat. Sol.
a, 202(8):1707, 2005.
[29] E. Lorenzo-Cabrera, C. J. Oton, N. E. Capuj, M. Ghulinyan,
D. Navarro-Urrios, Z. Gaburro, and L. Pavesi. Porous silicon-based
rugate filters. Appl. Opt., 2005.
[30] M. G. Berger, R. Arens-Fischer, M. Thonissen, M. Kr
uger, S. Billat,
H. L
uth, S. Hilbrich, W. Thei, and P. Grosse. Dielectric filters made
of ps: Advanced performance by oxidation and new layer structures.
Thin Solid Films, 297(1-2):237, 1997.
[31] J. L. Coffer. Porous silicon formation by stain etching. In L Canham,
editor, Properties of Porous Silicon, volume 18 of Emis Datareviews
Series, page 223. INSPEC, IEE, London, United Kingdom, 1997.
[32] H. Foll, M. Christophersen, J. Carstensen, and G. Hasse. Formation
and application of porous silicon. Mat. Sci. Eng. R, 39(4):93, 2002.
[33] X. G. Zhang. Morphology and formation mechanisms of porous silicon. J. Electrochem. Soc., 151(1):C69, 2004.
[34] R. L. Smith and S. D. Collins. Porous silicon formation mechanisms.
J. Appl. Phys., 71(8):R1, 1992.
[35] V. Lehmann and U. Gosele. Porous silicon formation - a quantum
wire effect. Appl. Phys. Lett., 58(8):856, 1991.
[36] J. Carstensen, M. Christophersen, and H. Foll. Pore formation mechanisms for the si-hf system. Mat. Sci. Eng. B-Solid, 69:23, 2000.
[37] V. Lehmann and S. Ronnebeck. The physics of macropore formation
in low-doped p-type silicon. J. Electrochem. Soc., 146(8):2968, 1999.
[38] D. T. J. Hurle and P. Rudolph. A brief history of defect formation,
segregation, faceting, and twinning in melt-grown semiconductors. J.
Cryst. Growth, 264(4):550, 2004.
150
[39] G. Lerondel, R. Romestain, and S. Barret. Roughness of the porous
silicon dissolution interface. J. Appl. Phys., 81(9):6171, 1997.
[40] H. Topse. Geometric Factors in Four Point Resistivity Measurement.
http://www.four-point-probes.com/haldor.html, 1968.
[41] G. Lerondel, P. Reece, A. Bruyant, and M. Gal. Strong light confinement in microporous photonic silicon structures. Materials Research
Society Symposium Proceeding, 797:W1.7.1, 2004.
[42] G. Lerondel, G. Amato, A. Parisini, and L. Boarino. Porous silicon
nanocracking. Mat. Sci. Eng. B-Solid, 69:161, 2000.
[43] D. Bellet and L. T. Canham. Controlled drying: The key to better
quality porous semiconductors. Adv. Mater., 10(6):487, 1998.
[44] L. T. Canham, A. G. Cullis, C. Pickering, O. D. Dosser, T. I. Cox,
and T. P. Lynch. Luminescent anodized silicon aerocrystal networks
prepared by supercritical drying. Nature, 368(6467):133, 1994.
[45] L. Canham, editor. Properties of Porous Silicon, volume 18 of Emis
Datareviews Series. INSPEC, IEE, London, United Kingdom, 1997.
[46] J. Salonen, E. Laine, and L. Niinisto. Thermal carbonization of porous
silicon surface by acetylene. J. Appl. Phys., 91(1):456, 2002.
[47] S. Setzu, G. Lerondel, and R. Romestain. Temperature effect on the
roughness of the formation interface of p-type porous silicon. J. Appl.
Phys., 84(6):3129, 1998.
[48] Crc Handbook of Chemistry and Physics. CRC Press, 78th edition,
1998.
[49] Femlab, comsol, inc., burlington, massachusetts, usa, 2004.
[50] O. S. Heavens. Optical Properties of Thin Solid Films. Dover Classics of Science and Mathematics. Dover Publications, Inc., New York,
1991.
[51] D. J. Bergman. Dielectric-constant of a composite-material - problem
in classical physics. Phys. Rep., 43(9):378, 1978.
[52] J. C. Maxwell Garnett. Philos. Trans. R. Soc. London, 203:385, 1904.
[53] H. Looyenga. Dielectric constants of heterogeneous mixtures. Physica,
31(3):401, 1965.
[54] D. A. G. Bruggeman. Ann. Phys., 24:636, 1935.
[55] W. Thei. Optical properties of porous silicon. Surf. Sci. Rep., 29(34):95, 1997.
151
[56] W. Thei and S. Hilbrich. Refractive index of porous silicon. In
L Canham, editor, Properties of Porous Silicon, volume 18 of Emis
Datareviews Series, page 223. INSPEC, IEE, London, United Kingdom, 1997.
[57] D. E. Aspnes and J. B. Theeten. Investigation of effective-medium
models of microscopic surface-roughness by spectroscopic ellipsometry. Phys. Rev. B, 20(8):3292, 1979.
[58] E. K. Squire, P. A. Snow, P. S. Russell, L. T. Canham, A. J. Simons, and C. L. Reeves. Light emission from porous silicon single
and multiple cavities. J. Lumin., 80(1-4):125, 1998.
[59] P. J. Reece, G. Lerondel, W. H. Zheng, and M. Gal. Optical microcavities with subnanometer linewidths based on porous silicon. Appl.
Phys. Lett., 81(26):4895, 2002.
[60] J. R. Reitz, F. J. Milford, and R. W. Christy. Foundations of Electromagnetic Theory. Addison-Wesley Publishing Company, Inc., Reading, Massachusetts, USA, 4th edition, 1993.
[61] E. D. Palik. Handbook of Optical Constants of Solids. Academic,
Orlando, FL, USA, 1985.
[62] P. A. Schumann, W. A. Keenan, A. H. Tong, G. H. H, and S. C. P.
Silicon optical constants in infrared. J. Electrochem. Soc., 118(1):145,
1971.
[63] M. Born and E. Wolf. Principles of Optics. University Press, Cambridge, UK, 7th edition, 1999.
[64] Z. Knittl. Optics of Thin Films. Wiley Series in Pure and Applied
Optics. Wiley, London, 1976.
[65] C. L. Mitsas and D. I. Siapkas. Generalized matrix-method for analysis of coherent and incoherent reflectance and transmittance of multilayer structures with rough surfaces, interfaces, and finite substrates.
Appl. Opt., 34(10):1678, 1995.
[66] H. E. Bennett and J. O. Porteus. Relation between surface roughness and specular reflectance at normal incidence. J. Opt. Soc. Am.,
51(2):123, 1961.
[67] H. Davies. The reflection of electromagnetic waves from a rough surface. P. I. Electr. Eng., 101(7):209, 1954.
[68] I. Fili
nski. Effects of sample imperfections on optical-spectra. Phys.
Status Solidi B, 49(2):577, 1972.
152
[69] M. Kildemo. Real-time monitoring and growth control of sigradient-index structures by multiwavelength ellipsometry. Appl.
Opt., 37(1):113, 1998.
[70] H. Bartzsch, S. Lange, R. Frach, and K. Goedicke. Graded refractive index layer systems for antireflective coatings and rugate filters
deposited by reactive pulse magnetron sputtering. Surf. Coat. Tech.,
180-81:616, 2004.
[71] A. V. Tikhonravov, M. K. Trubetskov, J. Hrdina, and J. Sobota.
Characterization of quasi-rugate filters using ellipsometric measurements. Thin Solid Films, 277(1-2):83, 1996.
[72] K. Robbie, A. J. P. Hnatiw, M. J. Brett, R. I. Macdonald, and J. N.
Mcmullin. Inhomogeneous thin film optical filters fabricated using
glancing angle deposition. Electron. Lett., 33(14):1213, 1997.
[73] K. Kaminska, T. Brown, G. Beydaghyan, and K. Robbie. Vacuum
evaporated porous silicon photonic interference filters. Appl. Opt.,
42(20):4212, 2003.
[74] W. J. Gunning, R. L. Hall, F. J. Woodberry, W. H. Southwell, and
N. S. Gluck. Codeposition of continuous composition rugate filters.
Appl. Opt., 28(14):2945, 1989.
[75] B. G. Bovard. Rugate filter theory - an overview.
32(28):5427, 1993.
Appl. Opt.,
[76] W. H. Southwell and R. L. Hall. Rugate filter sidelobe suppression using quintic and rugated quintic matching layers. Appl. Opt.,
28(14):2949, 1989.
[77] W. H. Southwell. Extended-bandwidth reflector designs by using
wavelets. Appl. Opt., 36(1):314, 1997.
[78] A. C. Van Popta, M. M. Hawkeye, J. C. Sit, and M. J. Brett. Gradientindex narrow-bandpass filter fabricated with glancing-angle deposition. Opt. Lett., 29(21):2545, 2004.
[79] A. Bruyant, G. Lerondel, P. J. Reece, and M. Gal. All-silicon omnidirectional mirrors based on one-dimensional photonic crystals. Appl.
Phys. Lett., 82(19):3227, 2003.
[80] E. Steinsland, T. Finstad, and A. Hanneborg. Laser reflectance interferometry for in situ determination of silicon etch rate in various
solutions. J. Electrochem. Soc., 146(10):3890, 1999.
[81] M. Thonissen, M. G. Berger, S. Billat, R. Arens-Fischer, M. Kr
uger,
H. L
uth, W. Thei, S. Hillbrich, P. Grosse, G. Lerondel, and
U. Frotscher. Analysis of the depth homogeneity of p-ps by reflectance
measurements. Thin Solid Films, 297(1-2):92, 1997.
153
[82] Z. Gaburro, C. J. Oton, P. Bettotti, L. Dal Negro, G. V. Prakash,
M. Cazzanelli, and L. Pavesi. Interferometric method for monitoring electrochemical etching of thin films. J. Electrochem. Soc.,
150(6):C381, 2003.
[83] D. Navarro-Urrios, C. Perez-Padrn, E. Lorenzo, N. E. Capuj,
Z. Gaburro, C. J. Oton, and L. Pavesi. Chemical etching effects in
porous silicon layers. Proc. SPIE, 5118:109, 2003.
[84] T. Unagami. Intrinsic stress in porous silicon layers formed by anodization in hf solution. J. Electrochem. Soc., 144(5):1835, 1997.
[85] M. Thonissen, S. Billat, M. Kr
uger, H. L
uth, M. G. Berger,
U. Frotscher, and U. Rossow. Depth inhomogeneity of porous silicon layers. J. Appl. Phys., 80(5):2990, 1996.
[86] M. Thonissen and M. G. Berger. Multilayer structures of porous
silicon. In L Canham, editor, Properties of Porous Silicon, volume 18
of Emis Datareviews Series, page 30. INSPEC, IEE, London, United
Kingdom, 1997.
[87] S. Billat, M. Thonissen, R. Arens-Fischer, M. G. Berger, M. Kr
uger,
and H. L
uth. Influence of etch stops on the microstructure of porous
silicon layers. Thin Solid Films, 297(1-2):22, 1997.
[88] M. Thonissen. Spektoskopische Charakterisierung Von Schichten Und
Schichtsystemen Aus Por
osem Silicium Im Hinblick Auf Optische
Und Optoelektronische Anwendungen. PhD thesis, Forschungszentrum J
ulich, 1999.
[89] G. Lammel. New Micromachining Technologies Using Porous Silicon.
Phd thesis, Swiss Federal Institute of Technology, 2001.
[90] M. C. Li, L. C. Zhao, and X. K. Chen. Reducing the effective barrier
height of a ptsi schottky diode by a p+ doping spike using pulsed laser
doping. J. Phys. D Appl. Phys., 36:2347, 2003.
[91] F. Raissi and M. M. Far. Highly sensitive ptsi/porous si schottky
detectors. IEEE Sens. J., 2(3):476, 2002.
[92] G. Parker and M. Charlton. Photonic crystals. Phys. World, 13(8):29,
2000.
[93] F. M
uller, A. Birner, U. Gosele, V. Lehmann, S. Ottow, and H. Foll.
Structuring of macroporous silicon for applications as photonic crystals. J. Porous Mat., 7(1-3):201, 2000.
[94] D. Hamm, T. Sakka, and Y. Ogata. Porous silicon formation under
constant anodization conditions: Homogeneous regime or transition?
J. Electrochem. Soc., 151(1):C32, 2004.
154
[95] B. E. A. Saleh and M. C. Teich. Fundamentals of Photonics. Wiley
Series in Pure and Applied Optics. John Wiley & Sons, Inc., New
York, NY, USA, 1991.
[96] A. Loni, L. T. Canham, M. G. Berger, R. Arens-Fischer, H. M
under,
H. L
uth, H. F. Arrand, and T. M. Benson. Porous silicon multilayer
optical waveguides. Thin Solid Films, 276(1-2):143, 1996.
[97] M. Araki, H. Koyama, and N. Koshida. Fabrication and fundamental
properties of an edge-emitting device with step-index porous silicon
waveguide. Appl. Phys. Lett., 68(21):2999, 1996.
[98] T. M. Benson, H. F. Arrand, P. Sewell, D. Niemeyer, A. Loni,
R. J. Bozeat, M. Kr
uger, R. Arens-Fischer, M. Thonissen, and
H. L
uth. Progress towards achieving integrated circuit functionality using porous silicon optoelectronic components. Mat. Sci. Eng.
B-Solid, 69:92, 2000.
[99] P. Ferrand and R. Romestain. Optical losses in porous silicon waveguides in the near-infrared: Effects of scattering. Appl. Phys. Lett.,
77(22):3535, 2000.
[100] C. C. Striemer and P. M. Fauchet. Dynamic etching of silicon for
broadband antireflection applications. Appl. Phys. Lett., 81(16):2980,
2002.
[101] J. A. Dobrowolski, D. Poitras, P. Ma, H. Vakil, and M. Acree. Toward
perfect antireflection coatings: Numerical investigation. Appl. Opt.,
41(16):3075, 2002.
[102] G. Lammel, S. Schweizer, S. Schiesser, and P. Renaud. Tunable optical filter of porous silicon as key component for a mems spectrometer.
J. Microelectromech. S., 11(6):815, 2002.
[103] R. Arens-Fischer, M. Kr
uger, M. Thonissen, V. Ganse, D. Hunkel,
M. Marso, and H. L
uth. Formation of porous silicon filter structures
with different properties on small areas. J. Porous Mat., 7(1-3):223,
2000.
[104] S. D. Senturia. Microsystem Design. Kluwer Academic Publishers,
Norwell, MA, USA, 2001.