Академический Документы
Профессиональный Документы
Культура Документы
Operators Manual
Contents-1
Contents
Page
Preface
Preface-1
1-1
1-3
1-3
1-4
1-5
1-6
1-7
1-8
2-1
2-3
Intended Use
2-3
System Features
2-4
2-4
2-5
2-5
2-6
2-8
2-9
Fluid Subsystem
2-12
Cuvette Subsystem
2-17
Photometer Subsystem
2-20
Communications Subsystem
2-22
Contents-2
Contents (Continued)
Page
Express Plus Subsystems: Principles of Operation (Continued)
User Interface Subsystem
Moving Through the Software
2-24
2-32
Screen Zones
2-34
2-35
Operational Sequence
2-37
3-1
3-5
Calculating Absorbance
3-6
3-8
Polychromatics
3-8
Monochromatics
3-9
Common Bichromatics
3-9
Sample Blanking
3-9
3-10
Data Selection
3-11
Substrate Depletion
3-11
Outlier Rejection
3-11
3-13
3-13
3-15
3-15
Contents-3
Contents (Continued)
Page
3-17
3-17
3-22
Two-Point Normalization
3-26
3-27
3-28
3-29
Calibration Failures
3-30
3-30
3-31
3-33
3-33
1-2s Rule
3-34
1-3s Rule
3-34
2-2s Rule
3-34
4-1s Rule
3-34
10x Rule
3-35
3-35
Sample Interpretation
3-37
3-40
Contents-4
Contents (Continued)
Page
4-1
4-3
4-3
Worklist Program
4-7
4-11
4-14
4-21
Definitions Program
4-32
Options Program
4-49
Diagnostics/Maintenance Program
4-57
4-64
5-1
5-3
5-5
5-11
5-12
5-14
Defining Profiles
5-24
5-26
Defining Controls
5-27
Defining Calibrators
5-30
Defining Blanks
5-33
Contents-5
Contents (Continued)
Page
5-37
5-41
6-1
6-5
Operating Summary
6-5
6-7
Starting Up
6-11
6-13
6-14
6-14
6-15
6-16
6-20
6-21
6-22
6-23
6-24
6-25
6-26
6-29
6-30
6-31
6-33
Contents-6
Contents (Continued)
Page
6-34
6-36
6-37
6-39
6-43
6-43
6-45
6-47
Shutting Down
6-51
7-5
7-7
7-8
7-10
7-13
7-14
7-16
7-17
7-21
7-24
7-26
Contents-7
Contents (Continued)
Page
7-28
7-30
7-33
7-35
7-39
7-41
7-44
8-1
8-3
8-3
8-5
8-8
8-11
8-14
8-17
8-20
9-1
9-3
9-4
9-5
9-6
9-7
9-8
Contents-8
Contents (Continued)
Page
Performing Preventive Maintenance Tasks (Continued)
Performing a Precision Check
9-9
9-12
9-14
9-16
9-19
9-21
9-26
9-27
9-31
9-32
9-37
9-43
9-45
9-46
9-50
9-52
10-1
10-3
10-4
10-5
10-11
10-13
Contents-9
Contents (Continued)
Page
10-44
10-47
10-47
10-48
10-52
10-53
10-60
10-64
10-68
10-70
Appendixes
A Glossary
A-1
B Safety Summary
B-1
C-1
D-1
E Specifications
E-1
F LIS Communications
F-1
G Relocation Instructions
G-1
Index
Index-1
Contents-10
Figures
Page
2-1
Express Plus
2-8
2-2
2-10
2-3
2-11
2-4
2-11
2-5
2-12
2-6
2-13
2-7
2-14
2-8
Probe Assembly
2-16
2-9
2-18
2-19
2-21
2-22
2-23
2-24
2-26
2-26
2-27
2-28
2-34
2-35
2-37
2-39
Contents-11
Figures (Continued)
Page
3-1
Outlier Rejection
3-12
3-2
3-36
3-3
Control 1 Results
3-39
3-4
Control 2 Results
3-39
4-1
4-3
4-2
4-5
4-3
4-7
4-4
4-11
4-5
4-14
4-6
4-22
4-7
4-33
4-8
4-35
4-9
4-36
4-37
4-38
4-39
4-49
4-57
4-59
4-60
4-61
4-62
4-64
4-64
Contents-12
Figures (Continued)
Page
5-1
5-5
5-2
5-7
5-3
5-9
5-4
Definitions/Tests Screen
5-12
5-5
5-14
5-6
5-15
5-7
5-15
5-8
5-17
5-9
5-18
5-19
5-20
5-22
5-24
5-25
5-27
5-29
5-30
5-31
5-33
5-34
5-39
6-1
6-8
6-2
6-8
6-3
Startup Screen
6-11
6-4
6-12
Contents-13
Figures (Continued)
Page
6-5
6-14
6-6
6-22
6-7
6-24
6-8
6-25
6-9
6-26
6-31
6-33
6-37
6-38
6-39
6-43
6-49
6-52
7-1
7-8
7-2
7-9
7-3
7-10
7-4
7-11
7-5
7-14
7-6
7-16
7-7
7-17
7-8
7-22
Contents-14
Figures (Continued)
Page
7-9
Definitions/Tests/QCSystems Screen
7-24
7-26
7-28
7-30
7-33
7-35
7-37
7-38
7-39
7-40
7-41
8-1
8-4
8-2
8-5
8-3
8-6
8-4
8-8
8-5
8-8
8-6
8-11
8-7
8-11
8-8
8-14
8-9
8-15
8-17
8-18
8-20
8-21
9-1
9-8
Contents-15
Figures (Continued)
Page
9-2
9-9
9-3
9-11
9-4
Front Panel
9-12
9-5
9-14
9-6
9-15
9-7
9-16
9-8
9-17
9-9
9-18
9-19
9-20
9-21
9-22
9-23
9-25
9-25
9-27
9-28
9-30
9-31
9-32
9-33
9-34
9-35
9-36
9-36
Contents-16
Figures (Continued)
Page
9-27 Removing the Probe Cover
9-37
9-38
9-39
9-40
9-41
9-41
9-44
9-45
10-11
10-13
10-22
10-44
10-48
10-52
10-53
G-1
G-7
G-2
G-7
G-3
G-8
G-4
G-9
G-5
G-11
G-6
G-12
G-7
G-13
Contents-17
Tables
Page
2-1
2-9
2-2
2-12
2-3
2-17
2-4
2-20
2-5
2-23
2-6
2-25
2-7
2-29
2-8
2-32
2-9
2-33
2-34
3-1
3-28
3-2
3-29
3-3
3-30
3-4
Automatic Reruns
3-31
3-5
3-32
6-1
6-7
6-2
6-9
6-3
Automatic Reruns
6-34
6-4
6-35
6-5
6-40
7-1
7-18
7-2
7-19
7-3
7-20
7-4
7-42
Contents-18
Tables (Continued)
Page
9-1
9-2
Fuse Ratings
9-4
9-30
10-5
10-14
10-19
10-45
10-49
10-54
10-60
10-65
10-66
C-1
Replacement Parts
C-5
C-2
C-6
C-3
Ordering Supplies
C-7
C-4
C-8
F-1
Communication Protocols
F-4
F-2
F-5
F-3
F-6
G-1
Fuse Ratings
G-5
Preface-1
Preface
This manual accompanies your Express Plus cinical chemistry
system and serves two basic functions:
The first function of this manual is to teach new operators the
Section 2
Section 3
Section 4
Section 5
Section 6
Section 7
Section 8
Section 9
Section 10
Appendix A
Glossary
Appendix B
Safety Summary
Appendix C
Appendix D
Appendix E
Specifications
Appendix F
LIS Communications
Appendix G
Relocation Instructions
Preface
1-1
Page
1-3
1-3
1-4
1-5
1-6
1-7
1-8
1-2
1-3
1-4
Refer to section . . .
10 Troubleshooting the
Express Plus
10 Troubleshooting the
Express Plus
Appendix A
Glossary
Appendix C
Service and
Supplies
Appendix F
LIS
Communications
Appendix G
Relocation
Instructions
1-5
Then . . .
WARNING
BIOHAZARD
CAUTION
NOTE
1-6
It means . . .
type
scan
select
selected
highlighted
(reverse video)
removed
1-7
It means . . .
alternating current (AC)
ON (Mains supply)
OFF (Mains supply)
Fuse
Keyboard
Bar Code Wand (Hand-held)
Printer
Water
This symbol indicates the system meets the
EMC Directive of the European Union.
1-8
Then refer to . . .
1-9
Then refer to . . .
2-1
Page
2-3
Intended Use
2-3
System Features
2-4
2-4
2-5
2-5
2-6
2-8
2-9
Fluid Subsystem
2-12
Cuvette Subsystem
2-17
Photometer Subsystem
2-20
Communications Subsystem
2-22
2-24
2-30
2-32
Screen Zones
2-34
2-35
Operational Sequence
2-37
2-2
Refer to . . .
Intended Use
System Features
Fluid Subsystem
Cuvette Subsystem
Photometer Subsystem
Communications Subsystem
Operational Sequence
2-3
Intended Use
The Express Plus is a clinical chemistry system that measures and
quantitates concentrations of chemistry analytes in serum or
urine. The system is intended for use by clinical laboratory
professionals.
The Express Plus has the following system capabilities:
assays up to 180 tests per hour for single-reagent applications
accepts up to 26 reagents on-board at one time
allows you to load and assay stat patient samples and controls
allows you to enter requests for up to 360 samples and 48 tests
allows random-access and batch testing
automates pipetting and diluting samples and reagents, and
reagent tray
allows you to enter test requests from the keyboard, by using
program
provides current information about the operating status of the
by the system
provides user interface software languages as a user-selectable
option
allows communication from three RS-232 ports
2-4
System Features
The Express Plus uses extensive software functions for system
setup and verification, automated dilution and rerun of samples,
and data management. These major system features contribute to
the efficiency and ease of use of the Express Plus. The following
paragraphs explain these features of your system.
The Express Plus offers two bar code features. These features
reduce the sample and reagent identification errors by
eliminating or minimizing keyboard entry of information about
samples, reagents, and tests.
The bar code features include the reagent bar code scanner and
the optional, hand-held bar code wand and Test Scheduler.
Reagent Bar Code Scanner
The reagent bar code scanner is provided as standard hardware
on the Express Plus. The reagent bar code scanner allows the
system to verify reagent loading, reagent positions on the reagent
tray, and reagent bottle types before the start of each run.
Optional Bar Code Accessories
The optional, hand-held bar code wand and Test Scheduler are
provided as part of the Positive Sample Identification software
option. This option allows you to schedule samples and tests
with minimal keyboard entry, reducing typing errors and the
possibility of incorrect patient identification.
The bar code wand and Test Scheduler allow:
entry of requests into the worklist without typing at the
keyboard
control of sample placement on the sample tray
aspiration of samples directly from the patients primary
sample tubes
association of a unique bar-coded request number with the
Reagent Carryover
Reduction Feature
2-5
2-6
Data Management
Features
the printed results reports and error flags that print on those
reports
the ability to back up and restore all data files on the system
Results Reports
The Express Plus offers several options for printing patient
results and other results. Two kinds of result reports are
available: run-time reports or collated patient reports. Refer to
Section 5, Setting Up the Express Plus, for more information
about setting the options for your run-time and collated reports.
A run-time report contains the results for an individual test run
for a sample (for example, a Glucose HK result for sample
number 20). This report prints on the system printer or optional
external printer as the system calculates the result at the time of
the run. You can set the printing options for the run-time
reports and modify the option settings as required. Run-time
reports can include any combination of the following data:
patient results, calibration and control results, and/or result
(error) flags that print with the result and provide added
information about the result.
A collated patient report contains the results for all tests run for
a sample. This report can also include demographic information
about the patient. After the tests are completed for a sample,
you can instruct the Express Plus to print the collated report for
that sample. You can set the printing options for the collated
report to include routine samples and/or Stat samples, to contain
a customized report heading, and to print on the system printer
or optional external printer.
Data Management
Features
(Continued)
2-7
plot
the Summary Statistics screen shows the expected, actual, and
2-9
Sample Transport
NOTE
Fluid Subsystem
2-15
(Continued)
Component
Probe
WARNING
BIOHAZARD
Cuvette Subsystem
2-17
Cuvette Strips
Cuvette Magazine
Cuvette Feeder
Cuvette Preheater
Cuvette Track
2-20
Photometer Subsystem
Photometer Board
2-22
Communications
Subsystem
Screen
Probe Board
Track Board
Keyboard
Transport Boards
System Keys
RS-232 Ports
Central
Processing
Unit
(CPU)
Photometer Board
External Printer
Parallel Port
Rev. B 1/2000
User Interface
Subsystem
2-25
(Continued)
Table 2-6. User Interface Subsystem Components
Component
Screen
Keyboard
Paper Advance
Key
System Printer
NOTE
User Interface
Subsystem
(Continued)
2-29
To . . .
start the testing sequence from the beginning, or
resume testing that you previously stopped or paused.
When you press START, the Express Plus:
primes the lines if you filled the distilled water
bottle
verifies the system operating temperatures
goes to the Warming Up or Running state
identifies the reagents on the reagent tray
waits for you to verify reagent loading
p use
in progress, dispenses the reagents for multireagent tests, then stops all dispensing operations
continues measuring reaction mixtures that contain
samples
waits for you to press START to resume assaying
2-30
Hardware Accessories
and Software Options
controls, and blanks defined for the Express Plus, which allows
for faster, more efficient worklist scheduling
enter data and select options without using the keyboard,
Hardware Accessories
and Software Options
(Continued)
2-31
System (LIS)
For information about cabling, configuring the ports, enabling the
software option, and programming for communications, refer to
the following documents that are shipped with the
communications software option:
Express Plus Communications Enabling Disk Instructions
Express Communications Manual
Express Interface Specifications Manual
To order accessories or software options, call Bayer Diagnostics.
Refer to Appendix C, Service and Supplies, for
Bayer Diagnostics addresses and telephone numbers for
placing your orders.
2-32
Then . . .
a menu shows on
the screen
a screen shows
without a blinking
cursor
2-33
Reverse Video
Cursor
Cursor Keys
SELECT
TAB
ENTER
EXIT
CANCEL
MAIN MENU
2-34
Screen Zones
STATUS ZONE
shows the systems operating status, date, time,
and the number of used cuvettes. Refer to
Table 2-10 for more information.
31-MAR-1992
15:11
Definitions
PROGRAM ZONE
shows the information or program that you can
select from the current screen.
Error detected
Press Status
Test: Gluc_HK
Tests
PLACEKEEPING ZONE
shows where you have moved in the software, from
the Main Menu selection to where you are now.
PROMPT ZONE
shows the prompts and messages that explain what
you should do or what the system requires.
Status:
Warming Up
Parameters
Press <EXIT> to
store definition
or review another
test.
F1
F2 Reagents
F7
F8
And it means . . .
31-MAR-1992
15:11
Status:
Error detected
Press Status
2-35
The Main Menu contains eight programs (F1 through F8). Each
program has various functions and/or actions that you can select.
Figure 2-20 is a map of the Express Plus programs.
Figure 2-20. Software Map for the Express Plus
Main Menu
F1 Worklist
F2 Patient Results
F3 Calibration Data
Worklist/Tests:
(Patient) Screen
Patient Results
Screen
Calibration Data
Screen
F1
Next Request
F1
Next Request
Previous Request
Previous Request
F2
F3
Find Request
F3
Find Request
F4
Insert Request
F4
Edit Request
F2
F5
F6
F7
F8
Erase Request
F7
Load Sample
F8
Worklist
Calibration Data/
Parameters
Screen
F4 Q.C. Statistics
Q.C. Statistics
Screen
F1
F2
F8
Data
Data
Levey-Jennings
F3
Summary Stats
F8
QCSystem
Calibration Data/
Data Screen
More
F4
Edit
Results
Demographics
F1
F2
F3
F6
F8
Recalculate
Rerun
Accept
Erase
Parameters
2-36
(Continued)
Main Menu
(Continued)
F5 Definitions
Definitions
Menu
F1
F2
Test
Profile
F6 Options
Options Menu
F1
F2
System Options
Printing Options
F3
Control
F3
Test Order
F4
Calibrator
F4
Communications
F5
Blank
F5
F6
Disk Options
Diagnostics/
F7 Maintenance
Diagnostics/
Maintenance
Menu Screen
Alternate
F8 Load
Program
Load Alternate
Program Screen
Precision Check
F1
F2
Manual Operations
F8
F3
Lamp Alignment
F1
F4
F5
Boot Floppy
More
System Version
Probe Tip Calibration
Reset System
The Express Plus also uses an online Help program that provides
instructions and explanations about screens, screen fields,
function keys, fixed-function keys (such as STATUS or PRINT),
and system keys (such as START or PAUSE).
There is also a Status program, which provides online
information about the operating status of the system, reagents,
samples, and worklist. The Status program is very useful in
troubleshooting and correcting system status errors.
For more information about the Express Plus software, refer to
Section 4, Understanding the Express Plus Programs, and to
Section 8, Using the HELP and STATUS Programs.
2-38
2-40
3-1
Page
3-5
Calculating Absorbance
3-6
3-8
Polychromatics
3-8
Monochromatics
3-9
Common Bichromatics
3-9
Sample Blanking
3-9
3-10
Data Selection
3-11
Substrate Depletion
3-11
Outlier Rejection
3-11
3-13
3-13
3-15
3-15
Calibration Methods
3-17
3-17
3-20
3-21
3-21
3-2
Page
3-22
3-24
3-24
3-25
3-25
Two-Point Normalization
3-26
3-27
3-28
3-29
Calibration Failures
3-30
3-30
3-31
3-33
3-33
1-2s Rule
3-34
1-3s Rule
3-34
2-2s Rule
3-34
4-1s Rule
3-34
10x Rule
3-35
3-35
Sample Interpretation
3-37
3-40
3-3
Refer to . . .
Calculating Absorbance
Polychromatics
Monochromatics
Common Bichromatics
Sample Blanking
Monochromatic Absorbance
Check
Data Selection
Substrate Depletion
Outlier Rejection
Calibration Methods
3-4
Refer to . . .
Using a Combination of
Control Rules
Sample Interpretation
3-5
3-6
Calculating Absorbance
Absorbance readings are taken on samples every ten seconds as
the samples move in front of the photometer. As light passes
through the cuvette and sample onto the photodiode, the
photometer circuit generates a current for each wavelength. This
current is inversely proportional to the amount of light absorbed
at each wavelength and is called the sample current.
The sample current readings are taken until the read interval
entered at the Read Time/Interval field on the Definitions/Tests/
Parameters screen, elapses. The readings are then calculated into
a result.
Zero currents are the readings taken when there is nothing
blocking the light source and the photodiode. The zero current
is a reading taken on air. This reading represents zero
absorbance.
After each retrace of the cuvette track, a new zero current for
each wavelength is taken at least once every twenty seconds. If
zero currents cannot be taken, the last set of zero currents are
used.
Absorbance is calculated as:
Abs(x) =
2-Log (%T)
where:
Abs (x) =
Log
%T
%T
100
3-7
where:
Sample Current
Zero Current
Since the ideal case is not usually observed, the equation for %T
requires another term. When the shutter is closed and no light
enters the photometer, there is still a current generated by the
circuit. This current is called the dark current.
During each retrace of the cuvette track, the system takes several
dark current readings at each wavelength, then averages the
readings for each wavelength. The system then subtracts the
average from the sample current and zero current for each
wavelength before it calculates %T as:
%T=
where:
Dark Current
3-8
NOTE
Absorbance (x)=
Sample Current (x) -Average Dark Current (x)
-Ln
where:
-Ln
Average Dark
Current (x)
Polychromatics
The system takes absorbance readings every ten seconds at each
wavelength defined at the Primary Wavelength and Secondary
Wavelength fields on the Definitions/Tests/Parameters screen for
a test. If one or two wavelengths are defined and the Sample
Blank? field is set to Yes, the system adjusts the absorbance
readings to yield a single corrected absorbance reading for every
ten seconds in the read interval. The system uses these corrected
absorbance readings for data selection and the remainder of the
result calculations.
3-9
Monochromatics
Common Bichromatics
Sample Blanking
Abs1 + Abs2
2
VB
VA
3-10
Sample Blanking
(Continued)
where:
Abs1 =
Abs2 =
VB
VA
3-11
Data Selection
The Express Plus uses the data selection algorithms to select the
subset of readings taken on each cuvette. The system then uses
these readings to calculate results. Unless the test is
monochromatic, the system corrects all readings by bichromatics
and sample blanking. All testing uses only the readings taken
during the read time/interval defined at the Definitions/Tests/
Parameters screen for the test.
Substrate Depletion
Outlier Rejection
seconds)
an endpoint test type has at least three readings (read time is
20 seconds)
If the minimum number of readings is not met for the test type,
the system does not perform outlier testing. The minimum
number of readings is required since the system uses linear and
quadratic regression to detect the outliers.
3-12
Outlier Rejection
(Continued)
The system does not flag when outlier rejection testing stops
because there are not enough readings. Outlier rejection testing
does stop when the following occurs:
three readings are left for a kinetic test type
two readings are left for an endpoint test type
590
600
0.4
Absorbance
0.3
0.2
530
540
560
560
570
Time
580
3-13
N XY - X Y
N X2 - ( X)2
where:
N =
X =
time value
Y =
bichromatically-corrected absorbance
3-14
(Continued)
Nonlinearity
a2T2
(RT)2+(NMultNoiseSys)2
Noise =
N-3
where:
R
NMult
NoiseSys
NoiseSys=
(SDp2+SDs2)
where:
SDp = expected S.D. of the primary wavelength readings at
their average absorbance level
SDs = expected S.D. of the secondary wavelength readings at
their average absorbance level
The flagging limit for noise is two times the system noise. The
flagging limit for nonlinearity is 0.4
NOTE
3-15
Y
N
Noise =
N-3
where:
Slope
For a two-point test type, there is one reading taken at each read
time/interval defined for the test at the Definitions/Tests/
Parameters screen. The response obtained for a two-point test
type is the difference between the second interval absorbance
and the first interval absorbance. The system calculates the
response as:
Response = Abs Interval2 - Abs Interval1
where:
Abs Interval2
Abs Interval1
3-16
Calibration Methods
The Express Plus can use any of eight curve types to calculate a
calibration curve. The linear curve type calculations determine
the response versus concentration calibration curve from the
calibrators. The nonlinear curve type calculations use the curve
to quantitate sample concentrations from the responses.
If the acceptance criteria for a linear or nonlinear calibration is
not met, the calibration or normalization is invalid and the curve
is not automatically accepted.
This section explains the following:
linear curve types
reblanking of stored curves
nonlinear curve types
two-point normalizations
qualitative curve types
summary of curve types
3-17
3-18
ConcStd=
where:
ConcStd
RStd
RBlk
When using more than one blank and calibrator, the factor is
determined as:
X(Xi - X0)2
X(Xi - X0)(yi -y0)
Response=
where:
Xi =
X0 =
yi
y0
If there are at least three data points when using two or more
calibrator concentrations or at least two data points when using
one calibrator concentration, the system calculates a standard
deviation of the regression line (residual error). The standard
deviation of the regression line using N-2 is calculated as:
S.D. =
Factor
[ (yi - yi)2]
(N-2)
where:
yi
(yi - yi)2=
N
3-19
If there are at least three data points when using two or more
calibrator concentrations or at least two data points when using
one calibrator concentration, the system calculates a standard
deviation of the regression line (residual error). The standard
deviation of the regression line using N-2 is calculated as:
S.D. =
Factor
[ (yi - yi)2]
(N-2)
where:
yi
(yi - yi)2
3-20
The result for a blanked linear curve type is the average of the
reagent blank responses subtracted from the sample response
before multiplying by the factor to give the concentration.
The blanked linear result is calculated as follows:
Result = (Sample Response-Average Reagent Blank)Factor
Reblanking Blanked Linear Curves
The Express Plus can reblank the curve of a test with a blankedlinear curve type. This resets only the intercept of the calibration
curve for the test. It does not affect the slope of the calibration
curve.
A current, valid calibration curve for the test must exist to
initially reblank a test. Once a test is reblanked, it can be
successively reblanked until a full calibration is required for the
test.
The result for a blanked-linear test is the average of the reagent
blank responses subtracted from the sample response before
multiplying by the factor to give the concentration. After the
system performs a full calibration, it calculates the blanked linear
result as follows:
Result = (Sample Response-Average reagent blank)Factor
CAUTION
3-21
The result for a limited linear curve type is the average of the
low calibrator replicates subtracted from the sample response
before multiplying by the factor and adding the low calibrator
concentration to the given concentration.
The limited linear result is calculated as:
Result=
The result for an enzyme linear curve type is the sample rate
(response) in delta absorbance per minute multiplied by the
factor. For an enzyme linear curve type, the factor is always
user-programmed. The blank is used only as a check of reagent
quality (refer to the Blanks and Zero Calibrators section.) The
enzyme linear result is calculated as follows:
Result = Sample ResponseFactor
NOTE
3-22
rate
= concentration of calibrators
a'
= ea
= a'Kc
3-23
After fitting the curve to the data, the system accepts the curve if
the curve standard deviation is less than any user-defined limit
for that test. If no limit is specified, no check on the S.D. is
performed. The equation for standard deviation is:
NConc Reps(i)
S.D. =
i=1
j=1
[(yij-y
i)2 + (y
i-yi)2]
(NConc-x)
where:
NConc =
Reps(i) =
yij
yi
yi
calibrator concentration.
Samples with rates outside the range defined by the highest
calibrator print with the flag Over, followed by the highest
calibrator concentration.
3-24
Z(Cb)
1 + a(Cb)
C = ex
where:
x=
-Ln
-a
(R-R0)
b
For a 2-Logit5 curve type, you must run at least one zero
calibrator to ensure convergence of the curve fit iterations. This
convergence of the curve fit iterations is accepted for a 2-Logit5
curve type and 3-Exponential5 curve type only if the mean
square error decreases.
For a 2-Logit5 curve type, the parameters a, b, and c are all
constrained at each iteration to prevent them from going
negative. If parameters b and c are zero after the iterations
converge, the flat curve is useless and the system does not accept
the curve. Sample quantitation uses the Gauss-Newton method
of iteration since the curve equation does not have a solution for
C in closed form.
The calculation for a 2-Logit5 curve type is:
R = R0 +
Z(Cb)(e(cC))
1 + a(Cb)(e(cC))
3-25
4-Polynomial5 Curve
Type
CAUTION
For a 4-Polynomial5 curve type, you must run at least one zero
calibrator, since its average rate is used as curve parameter R0.
This calculation is linear in the other four parameters, allowing
linear regression to be used instead of Gauss-Newton iterations.
The linear regression curve fit minimizes the mean square error
in Ln(C) rather than in rate.
Since the curve predicts a rate of R0 at concentration C0=ea
instead of at zero, Under prints when the samples rate is below
that of the lowest nonzero calibrator. If the curve equation is
satisfied by more than one value of R for any value of C between
the lowest nonzero and highest calibrator concentrations, the
system prints Curve is nonmonotonic and does not accept the
calibration. The calculation for a 4-Polynomial5 curve type is:
Ln(C)=a+b[(R-R0)/100]+c[(R-R0)/100]2+d[(R-R0)/100]3
Sample quantitation consists of raising e to the power given by
the right-hand side of the curve equation.
3-26
Two-Point
Normalization
S(NH) - S(NL)
RH - RL
where:
S(NH) =
S(NL) =
RH
RL
3-27
S.D. =
(y
i - y0)2
N-1
where:
N
yi
y0
The S.D. is compared to the limit defined at the Curve S.D. Limit
field on the Definitions/Tests/Parameters screen. If the S.D.
exceeds the limit, the calibration is flagged as invalid. If no curve
S.D. limit is defined, the system does not check the standard
deviation.
3-28
Curve Type
Factor
Blank or
Zero Cal.
Used?
Blanked Linea
User-defined
Yes
Blanked Linear
Calculated
Yes
2 through 6
Enzyme Linear
User-defined
Yes
Limited Linear
User-defined
No
Limited Linear
Calculated
No
2 through 6
Qualitative
User-defined
No
Logit 4
User-defined
No
4 through 6
Logit 5
User-defined
Yes
5 and 6
Exponential 5
User-defined
No
5 and 6
Polynomial 5
User-defined
Yes
5 and 6
NOTE
Calibrator Concs.
(including blanks)
3-29
Refer to Tables 3-2 and 3-3 for a summary of the flags that can
show on the Patient Results screen while the system calculates a
result or calibrates a test:
Table 3-2. Error Flags for a Rate and/or Response
This flag . . .
Means . . .
High abs
High pr abs
Low abs
Low pr abs
Low sc abs
Nonlinear
Outlier
3-30
Means . . .
Control out
High conc.
Calibration Failures
Over
Under
system screen.
Normalization or
Reblanking Failures
3-31
Is rerun if . . .
And is rerun . . .
patients and
controls
once
>1
until twice the
number of scheduled
replicates are run
3-32
Dilution?
Control out
No
No
Yes
No
High Concentration.
Yes
Yes
No
No
No
No
Outlier
No
Over
Yes
No
No
No
No
Under
No
3-33
3-34
1-2s Rule
1-3s Rule
2-2s Rule
4-1s Rule
materials that do not violate the 4-1s rule, then it evaluates the
4-1s rule within a single control material.
If the system cannot find four control results across control
materials, it cannot evaluate the 4-1s rule, so it prints 1-2s next
to the result.
If the Express Plus finds four control results within a control
material that do not violate the 4-1s rule, then it evaluates the
10x
rule.
If the system cannot find four control results within the control
10x
Rule
3-35
The 10x
rule detects small shifts in control results. If the current
control result and the nine preceding control results for the
selected test fall on the same side of the mean, the system prints
10x
next to the result.
The Express Plus first applies this rule across control materials
by using the most recent control result and the nine preceding
control results, regardless of control materials. Since this rule
uses the most recent control results to evaluate across control
materials, this evaluation might not always include each different
control material defined for the test.
If the Express Plus finds ten control results across control
Using a Combination
of Control Rules
3-36
Using a Combination
of Control Rules
Figure 3-2. Flowchart of Control Rule Evaluations
(Continued)
Start
Evaluating
Rules
Does
result violate
1-2s rule?
No
System reports
the result
No
Stop
Evaluating
Rules
Yes
System reports
the result with
the 1-3s flag
Yes
System reports
the result with
the 2-2s flag
Yes
System reports
the result with
the 4-1s flag
Yes
Is 1-3s rule
defined for
the test?
Yes
No
Is 2-2s rule
defined for
the test?
No
Yes
No
Is 4-1s rule
defined for
the test?
No
System reports
the result with
the 1-2s flag
Does
result violate
2-2s rule?
No
Yes
No
Is 10x rule
defined for
the test?
Does
result violate
1-3s rule?
Does
result violate
4-1s rule?
System
automatically
reruns sample
No
Yes
Does
result violate
10x rule?
Yes
System reports
the result with
the 10x flag
Yes
No
Is
Automatic Test
Shutdown
ON?
Yes
System shuts
down the
test
No
Is
this an initial
result (not a
rerun)?
No
Stop
Evaluating
Rules
Sample
Interpretation
3-37
Expected Mean
Expected 1 S.D.
Control 1
2.5
0.5
Control 2
6.0
1.0
Result
Date
Interpretation
Control 1
2.9
1-1-92
Control 2
6.6
1-1-92
Control 1
2.9
1-2-92
Control 2
7.1
1-2-92
Control 1
3.3
1-3-92
Control 2
7.2
1-3-92
Control 1
3.2
1-4-92
Control 1
3.0
1-5-92
3-38
Sample Interpretation
(Continued)
Control
Result
Date
Interpretation
Control 1
3.6
1-6-92
control rule(s)
preceding Control 1 value (3.0)
Control 2
8.2
1-7-92
control rule(s)
preceding Control 2 value (7.2)
rule is violated)
result is flagged 10x
and further
3.9
1-8-92
Sample Interpretation
(Continued)
3-39
3.5
3.0
2.5
2.0
1.5
1.0
01-MAR-1992
08-MAR-1992
8.0
7.0
6.0
5.0
4.0
3.0
01-MAR-1992
08-MAR-1992
3-40
Statistics.
The number of possible points stored per test is 2047. The
system uses all the data points when calculating the lifetime
mean, and stores the data points in reverse chronological
order, such that 2048 overwrites 1, and 2049 overwrites 2, etc.,
when calculating the actual mean.
For QCSystems Summary Reports, if the number of points for
is not printed).
Only control results for a defined control go into the Q.C.
database.
If you select Expected Mean at the Control Rule Evaluation field
control rules.
4-1
Page
4-3
4-3
Worklist Program
4-7
4-11
4-14
4-21
Definitions Program
4-32
Options Program
4-49
Diagnostics/Maintenance Program
4-57
4-64
4-2
Refer to . . .
Worklist Program
Definitions Program
Options Program
Diagnostics/Maintenance
Program
4-3
Main Menu
F1 Worklist
F2 Patient Results
F3 Calibration Data
Press the required
function key.
F4 Q.C. Statistics
F5 Definitions
F6 Options
F7 Diagnostics/Maintenance
F8 Load Alternate Program
4-4
Press . . .
F1 Worklist
F2 Patient Results
F3 Calibration Data
F4 Q.C. Statistics
F5 Definitions
F6 Options
F7 Diagnostics/
F8 Load Alternate
Maintenance
Program
The rest of this section explains each Main Menu program and
provides:
a software map of each program
a chart that describes the functions of each program
a chart that briefly describes the screen fields of each program
4-5
Patient Results
F1 Worklist
F2
Worklist/Tests:
(Patient) Screen
Patient Results
Screen
Next Request
Next Request
F1
Previous Request
F2
F3
Find Request
F3
Find Request
F4
Insert Request
F4
Edit Request
Erase Request
F7
Load Sample
F8
F1
F2
F5
F6
Previous Request
Worklist
F3
Calibration Data
Screen
Calibration Data/
Parameters
Screen
F8
F4
Q.C. Statistics
Q.C. Statistics
Screen
F1
F2
F8
Data
F3
F8
Data
Levey-Jennings
Summary Stats
QCSystem
Calibration Data/
Data Screen
More
F4
F7
Calibration Data
Edit
Results
Demographics
F1
F2
Recalculate
Rerun
F3
Accept
F6
Erase
F8
Parameters
4-6
F5 Definitions
Definitions
Menu
F1
F2
Test
Profile
F6 Options
Options Menu
F1
F2
System Options
Printing Options
F3
Control
F3
Test Order
F4
Calibrator
F4
Communications
F5
Blank
F5
F6
Disk Options
Reset System
Diagnostics/
F7 Maintenance
Diagnostics/
Maintenance
Menu
F1
Precision Check
F2
Manual Operations
F3
Lamp Alignment
F4
F5
System Version
Probe Tip Calibration
Alternate
F8 Load
Program
Load Alternate
Program Screen
F1
Boot Floppy
4-7
Worklist Program
The Worklist program contains the options that allow you to
create and manage a worklist of your patient samples, controls,
calibrators, and blanks. You can also move to the Patient Results
program from the Worklist program. Refer to Section 6,
Operating the Express Plus, for more information about using the
Worklist program to schedule your daily worklists.
Use the Worklist program to:
schedule your daily worklist of tests for patient samples,
Worklist/Tests screen)
Figure 4-3 shows the Worklist program and its options:
Figure 4-3. Worklist Program Map
F1 Worklist
Worklist/Tests
Screen
F1
Next
Request
F2
Previous
Request
F3
Find
Request
F1
F1
Request
Number
F2
Cup
Number
F3
F4
Erase
Worklist
Patient
Name
F4
Insert
Request
F2
F5
Erase
Request
Patient
ID
F5
Erase
Request
F3
F6
Patient
Name
First
Request
F6
Load
Sample
F4
F7
Patient
ID
Last
Request
F7
Repeat
F8
4-8
Press . . .
F1 Next Request
F2 Previous
F3 Find Request
Request
F4 Insert Request
F5 Erase Request
F6 Load Sample
(Continued on the next page)
4-9
Press . . .
F7 Schedule
Calibs.
F7 Results
F8 Demo-
F8 Concen-
F8 Tests
graphics
trations
Shows . . .
At the Worklist/
Age
Demographics screen
Blanks
Blanks screen
Calibrator
Name
Concentrations screen
Calibrators
Calibrators screen
Comments
Demographics screen
(Continued on the next page)
4-10
At the Worklist/
Conc.
Concentrations screen
Controls
Controls screen
Patient ID
Demographics screen
Patient
Name
Demographics screen
Tests (Patient) screen
Profiles
Profiles screen
Request
Number
Sample
Type
Sex
Demographics screen
Stat?
Tests
Tray/Cup
Number
4-11
Figure 4-4 shows the Patient Results program and its options:
Figure 4-4. Patient Results Program Map
F2
Patient Results
Patient Results
Screen
F1 Next
Request
F1
Request
Number
F2
F2 Previous
Request
Cup
Number
F3
Patient
Name
F3 Find
Request
F4
Patient
ID
F4 Edit
Request
F5
First
Request
F7 Worklist
F6
Last
Request
F8 More
F7
Repeat
4-12
Press . . .
F1 Next
F2 Previous
Request
Request
F3 Find
Request
F4 Edit
F7 Worklist
F8 More
F3 Patient ID
F4 Cup Range
F8 All Requests
Request
4-13
Shows . . .
Control Name
Flag
Patient Name
Request
Number
Result
Sample Type
Test
Tray/Cup
Number
4-14
Figure 4-5 shows the Calibration Data program and its options.
Figure 4-5. Calibration Data Program Map
F3
Calibration Data
Calibration Data
Screen
Calibration Data/
Parameters
Screen
F8
Data
Calibration Data/
Data Screen
F4
F1
Edit
Recalculate
F2
Rerun
F3
Accept
F6
Erase
F8
Parameters
Calibration Data/
Parameters
Screen
F1
Recalculate
F2
Rerun
F3
Accept
F6
Erase
F8
Data
4-15
To . . .
Press . . .
F1 Recalculate
F2 Rerun
F3 Accept
F6 Erase
F6 Erase
F8 Parameters
4-16
At the Calibration
Calibration
Status
Data and
Data/Data screens
Conc.
Data/Data screen
Curve
Parameters
Data/
Parameters screen
Where:
Low Cal. is the mean response of
the lowest calibrator
Factor is the factor calculated from
the calibrator values
K, R0, and Z are nonlinear
parameters used to scale the curve:
K and Z are slopes
R0 is the predicted rate of a
concentration of zero (intercept)
a, a, b, c, and d are nonlinear curve
parameters
Curve S.D.
Data/
Parameters screen
4-17
Shows . . .
At the Calibration
Data (1, 2, 3)
Data/Data screen
Error (1, 2, 3)
Data/Data screen
Last
Calibration
Data/Parameters
screen
Last
Normalization
Data/Parameters
screen
Last
Reblanking
Data/Parameters
screen
Maximum
Data/Parameters
screen (for
normalized curves)
Mean
Response
Data/Data screen
Minimum
Data/Parameters
screen (for
normalized curves)
Normalization
Parameters
Data/Parameters
screen (for
normalized curves)
Normalization
Status
Data/Parameters
screen (for
normalized curves)
(Continued on the next page)
4-18
Shows . . .
At the Calibration
Reason for
rejection
Data/Parameters
screen
Data/Parameters
screen (for
reblanked curves)
Reblanking
Required
Data/Parameters
screen
Recalibration
Required
Data/Parameters
screen
Renormalization Required
Data/Parameters
screen
Reblanking
Status
Data/Parameters
screen (for
reblanked curves)
(Continued on the next page)
4-19
Shows . . .
At the Calibration
Rep. S.D.
Data/Data screen
Test
Data/Parameters
and Data/Data
screens
Test Name
Data/Parameters
and Data/Data
screens
4-20
Means . . .
CURRENT
EXPIRED
INCOMPLETE
the test is scheduled, but not all the data points are
collected, or you edited data point(s) but did not
recalculate the curve.
IN PROCESS
INVALID
NONE
NORMALIZED
REBLANKED
SCHEDULED
Figure 4-6 shows the Q.C. Statistics program and its options.
4-21
4-22
Q.C. Statistics
Q.C. Statistics
Screen
F1 Data
LeveyF2 Jennings
Summary
F3 Stats
F8
Q.C. Statistics/
Summary
Statistics Screen
Q.C. Statistics/
Levey-Jennings
Screen
F1 Data
F2
F1
Data
Details
F3
Q.C. Statistics/
QCSystems
Screen
LeveyF2 Jennings
Summary
Stats
F4
F4
Q.C. Statistics/
Levey-Jennings/
Details Screen
LeveyF2 Jennings
QCSystem
F4
Edit
Update
F4 Stats
Edit
F5
F8
Next
Control
More
F6
Previous
Control
F7
Move
Left
F8
Update
Plot
Next
F7 Entry
Previous
F8 Entry
Q.C. Statistics/
Data Screen
F1 Details
LeveyF2 Jennings
Summary
F3 Stats
F4
Update
Data
Q.C. Statistics/
Data/Details
Screen
F1 Data
F4
Edit
Next
F7 Entry
Previous
F8 Entry
Next
F5 Control
Previous
F6 Control
Next
F7 Page
Previous
F8 Page
Move
Right
4-23
Press . . .
F1 Data
F2 Levey-Jennings
F3 Summary Stats
F8 QCSystem
4-24
First . . .
Then press . . .
F1 Details
F2 Levey-Jennings
F3 Summary Stats
F4 Update Data
F5 Next Control
F6 Previous Control
F7 Next Page
F8 Previous Page
press PRINT
4-25
First . . .
F1 Data
F2 Details
F3 Summary Stats
Then press . . .
F4 Update Plot
2. Plotted Against,
then press SELECT
to select the
vertical axis
F5 Next Control
F6 Previous
F7 Move Left
F8 Move Right
Control
press PRINT
F1 Test
F2 Control
4-26
First . . .
Then press . . .
F1 Data
F2 Levey-Jennings
F4 Update Stats
F8 More
4-27
Press . . .
F1 Data
Data/Details screen
F2 Levey-
Levey-Jennings/Details
screen
F4 Edit
Data/Details or
Levey-Jennings/Details
screens
F4 Edit
Data/Details or
Levey-Jennings/Details
screens
F7 Next
Data/Details or
Levey-Jennings/Details
screens
F8 Previous
Jennings
Entry
Entry
Data/Details or
Levey-Jennings/Details
screens
4-28
Actual
Data;
Data/Details;
Levey-Jennings/
Details;
Summary Statistics
screens
Comments
modifiable, operator-entered
comments (up to 30
alphanumeric characters)
Control
Control
Rules
Date
(Time)
Data;
Data/Details; LeveyJennings; LeveyJennings/Details
screens
End Date
Data;
Levey-Jennings;
Summary Statistics;
QCSystems screens
Excluded
Data; Data/Details;
Levey-Jennings/Details
screens
4-29
Mean
Expiration
Date
Data/Details;
Levey-Jennings/Details;
Summary Statistics
screens
Lifetime
Summary Statistics
screen
1 SD
CV%
coefficient of variation
calculated since the last
time the Q.C. database
was erased
Mean
Lot
Number
Data/Details;
Levey-Jennings/Details;
Summary Statistics
screens
Operator
Data Details;
Levey-Jennings/Details
screens
Plot
Statistics
Levey-Jennings screen
4-30
ance is plotted
against the expected
mean and one S.D.
values defined for
the selected control
Lifetime
Result
Data;
Data/Details;
Levey-Jennings;
Levey-Jennings/Details
screens
Rule
Data;
Data/Details;
Levey-Jennings;
Levey-Jennings/Details
screens
Start Date
Data;
Levey-Jennings;
Levey-Jennings/Details;
Summary Statistics
screens
4-31
Status*
Data;
Data/Details;
Levey-Jennings/Details
screens
Test
Data;
Levey-Jennings;
Levey-Jennings/Details;
Summary Statistics
screens
Time
Data;
Data/Details;
Levey-Jennings;
Levey-Jennings/Details
screens
4-32
Definitions Program
The Definitions program enables you to define, modify, or erase
parameter values for tests, profiles, controls, calibrators, and
blanks, and associated parameter values for assay order, reagents,
calibrator concentrations, and control rules. For more
information about using the Definitions program to set up your
system, refer to Section 5, Setting Up the Express Plus.
Use the Definitions program to:
define, modify, or erase customized tests, profiles, controls,
4-33
Definitions
Definitions Menu
Screen
F1 Test
Profile
F2
F3
Definitions/
Profiles Screen
F1 Insert
F4
Definitions/
Controls Screen
Erase
F2
Control
F1 Insert
Definitions/
Profiles/Selection
Screen
F2
Calibrator
Definitions/
Calibrators Screen
Erase
F1 Insert
F2
Definitions/
Blanks Screen
Erase
F1 Insert
Definitions/
Calibrator/
Concentrations
Screen
Definitions/
Control/Limits
Screen
Blank
F5
F2
Erase
Definitions/
Blanks/Selection
Screen
Definitions/Tests
Screen
F1
Insert
F2
Erase
F3
Restore
Definitions/Tests/
Parameters
Screen
F2
Reagents
F3
Assay Order
F4
Calib. Concs.
F5
Controls
F6
Definitions/Tests/
Assay Order
Screen
Definitions/Tests/
Reagents
Screen
ParamF1 eters
ParamAssay
F1 eters
F3 Order
F2
ParamReagents
F1 eters
F2
Calib.
F4 Concs.
Calib.
Controls
F4 Concs. F5
F6
More
ParamReagents
F1 eters
F2
ParamReagents
F1 eters
F2
Reagents
F8
Definitions/Tests/
QCSystems
Screen
Definitions/Tests/
Controls
Screen
Definitions/Tests/
Calibrator
Concentrations
Screen
QCSystem
F5
Assay
F3 Order
Controls
QCSystem
F6
F5
Assay
F3 Order
Controls
QCSystem
F6
Assay
F3 Order
Calib.
F4 Concs.
F5
Calib.
QCSystem
F4 Concs. F6
QCSystem
Controls
4-34
Press . . .
F1 Test
F2 Profile
F3 Control
F4 Calibrator
F5 Blank
Definitions/Tests Screens
When you press F1 Test at the Definitions menu screen, the
Definitions/Tests screen shows the tests currently defined for the
Express Plus and has these functions available:
To . . .
First . . .
Then press . . .
SELECT
F1 Insert
F2 Erase
F3 Restore
NOTE
4-35
Press . . .
F2 Reagents
F3 Assay Order
F4 Calib. Concs.
F5 Controls
F6 QCSystem
Definitions
Test: Alb
Tests
Parameters
F1
F2 Reagents F3 Assay
Order
F4 Calib.
Concs.
F5 Controls F6 QCSystem F7
F8
4-36
First . . .
Then press . . .
SELECT
F1 Insert
F2 Erase
Definitions
Profiles
Selection
Press <SELECT>
to select the test
indicated by the
cursor.
Press <EXIT> to
store definition
or review another
profile.
ALP
BUN
GGT
UA
ACP_P
ALT
Ca
Gluc_HK
HBD
Mg
Amy
Chlor
IPhos
HDL
Fe
AST
Chol
LDH_L
CO2
ExFe
DBili
CK
TP
ACP_NP
UIBC
4-37
First . . .
Then press . . .
SELECT
F1 Insert
F2 Erase
Definitions
Control
Limits
F1
F2
Expected
Test
Mean
Alb
3.3
ALT
AST
TBili
0.7
Ca
Chol
141.
Creat
Gluc_HK 88.
LDH_L
Trig
74.
HDL
CO2
F3
F4
Expected
Expected Expected
1 SD
Test
Mean
1 SD
0.30
ALP
Amy
DBili
0.1
0.10
0.20
BUN
16.0
3.00
Chlor
98.
4.0
21.0
CK
GGT
9.0
IPhos
TP
5.5
0.30
11.0
UA
5.1
0.50
HBD
ACP_NP
F5
F6
F7
F8 More
4-38
First . . .
Then press . . .
SELECT
F1 Insert
F2 Erase
Definitions
Calibrator
Concentrations
Test
Conc.
ALT
DBili
0.4
Ca
9.9
CK
Gluc_HK 114.
TP
7.6
HDL
ACP_NP
Mg
2.4
UIBC RATIO
4-39
First . . .
Then press . . .
SELECT
F1 Insert
F2 Erase
Definitions
Blanks
Selection
Press <SELECT> to
select the test
indicated by the
cursor.
Press <EXIT> to
store the definition
or review another
blank.
ALP
BUN
GGT
UA
ACP_P
ALT
Ca
Gluc_HK
HBD
Mg
Amy
Chlor
IPhos
HDL
Fe
AST
Chol
LDH_L
CO2
ExFe
DBili
CK
TP
ACP_NP
UIBC
4-40
Shows . . .
At the Definitions/
Bar Code
(Rgnt 1, 2,
3, 4)
Tests/Reagents
screen
Blank Bar
Code
Blanks screen
Blank Name
Blanks screen
Bottle Type
Tests/Reagents
screen
Calibration
Interval
Tests/Parameters
screen
Calibrator
Bar Code
Calibrator/
Concentrations
screen
Calibrator
Name
Calibrator/
Concentrations
screen
Conc.
Calibrator/
Concentrations
screen
Conc. n
Tests/Calibrator
Concentrations
screen
Control Bar
Code
Tests/Controls and
Control/Limits
screens
(Continued on the next page)
4-41
Shows . . .
At the Definitions/
Control
Name
Control/Limits
screen
Control Rule
Evaluation
Tests/Controls
screen
Lifetime
Mean
Control
Rules
Tests/Controls
screen
Controls
Tests/Controls
screen
Curve S.D.
Limit
Tests/Parameters
screen
Curve
Type
Tests/Parameters
screen
4-42
Shows . . .
At the Definitions/
Diluent
Volume
Tests/Reagents
screen
Expected 1
S.D.
Control/Limits
screen
Expected
Mean
Control/Limits
screen
Expiration
Date
Control/Limits
screen
Factor
Tests/Parameters
screen
Fluid 1 Bar
Code
Tests/Assay Order
screen
Fluid 2 Bar
Code
Tests/Assay Order
screen
High A Limit
Tests/Parameters
screen
Tests/Parameters
screen
4-43
Shows . . .
At the Definitions/
High Normal
Tests/Parameters
screen
Lag Time
Tests/Reagents
screen
Linearity
Limit
Tests/Parameters
screen
Lot Number
Calibrator/
Concentrations
screen
Lot Number
Control/Limits
screen
(Continued on the next page)
4-44
Shows . . .
At the Definitions/
Low A Limit
lowest bichromatic
Tests/Parameters
screen
Tests/Parameters
screen
Low Normal
Tests/Parameters
screen
Method Code
Tests/QCSystems
screen
No. of
Calibrators
number of calibrators
(including blanks) the system
assays and uses to calibrate the
curve for the test (range: 0 to 6;
default: 1)
Tests/Parameters
screen
No. of Decimal
Places
Tests/Parameters
screen
No. of
Replicates
Tests/Parameters
screen
Normalization/
Reblanking
Interval
Tests/Parameters
screen
Pool Number
Control/Limits
screen
4-45
Shows . . .
At the Definitions/
Predilution
Ratio
Tests/Reagents
screen
Primary
Wavelength
Tests/Parameters
screen
Profile Bar
Code
Profiles/Selection
screen
Profile Name
Profiles/Selection
screen
Ratio
Definition
mathematical expression of up
to 58 characters used to
describe how to calculate a ratio
test (allowable math operators
are: +, -, , (, ), and /)
Tests/Parameters
screen
Read
Time/Interval
Tests/Parameters
screen
Reagent Code
Tests/QCSystems
screen
Reagent
Diluent Bar
Code
Tests/Reagents
screen
Reagent
Diluent Bottle
Type
Tests/Reagents
screen
4-46
Shows . . .
At the Definitions/
Reagent
Volume
Tests/Reagents
screen
Rerun Dilution
Ratio
Tests/Reagents
screen
Rgnt 1, 2, 3, 4
Tests/Reagents
screen
Sample Blank?
Tests/Parameters
screen
4-47
Shows . . .
At the Definitions/
Sample
Diluent Bar
Code
Tests/Reagents
screen
Sample
Diluent Bottle
Type
Tests/Reagents
screen
Sample
Volume
Tests/Reagents
screen
Secondary
Wavelength
Tests/Parameters
screen
Test
Test Code
Tests/QCSystems
screen
Test Name
Tests
Profiles, Controls,
Calibrators, and
Blanks screens
(Continued on the next page)
4-48
Shows . . .
At the Definitions/
Test Type
Tests/Parameters
screen
Units
Tests/Parameters,
Tests/QCSystems
screens
Units Code
Tests/QCSystems
screen
Wash Count
Tests/Assay Order
screen
Wash Fluid
Tests/Assay Order
screen
4-49
Options Program
The Options program contains the system, printing,
communications, and disk options that allow you to manage the
set up and customizing of your Express Plus.
Refer to Section 5, Setting Up the Express Plus, for more
information about using the Options program to set up and
customize the available options.
Refer to Section 9, Maintaining the Express Plus for procedures
on backing up and restoring data and system calibration files, and
resetting the system.
If you have communications enabled for your Express Plus, refer
to the documents that accompany your Express Plus
Communications software option for details on setting the
communications configuration and protocols.
Use the Options program to:
customize the installation option settings
customize the selection of the default printer and the printing
of collated reports
customize the order in which the system shows the tests on the
programmed values
Figure 4-13 shows the Options program and its functions:
Figure 4-13. Options Program Map
F6 Options
Options Menu
Screen
System
F1 Options
Printing
F2 Options
F3
Backup
Data
Files
Restore
F2 Data
Files
F3 Backup
F4 Restore
F5
System
System
Calibrations
Calibrations
F1
Test
Order
CommuniF4 cations
F5
Disk
Options
Backup
Program
Files
F6
F6
Reset
System
Format
Floppy
Disk
4-50
Press . . .
F1 System
F2 Printing
F3 Test
F4 Communi-
Options
Options
Order
cations
F3 Reset
Channel
CAUTION
4-51
Press . . .
F5 Disk
Options
F2 Restore
Data Files
F3 Backup
System
Calibrations
F4 Restore
System
Calibrations
F5 Backup
Program
Files
F6 Format
Floppy Disk
NOTE
F6 Reset
System
4-52
At the Options/
Automatic
Test
Shutdown
System Options
screen
Baud Rate
Communications
screen
Collate By
Communications
screen
Collated
Reports
Printing Options
screen
Communications
screen
4-53
At the Options/
Default
Printer
Printing Options
screen
Device
Communications
screen
Device
Number
Communications
screen
Enable
Polling
Communications
screen
Enable
Screen
Saver
System Options
screen
Host Device/
Remote
Device
Communications
screen
Operating
Temperature
System Options
screen
Options
System Options
screen
4-54
Shows . . .
At the Options/
Parity
Communications
screen
QCSystems
Instrument
Code
System Options
screen
Receiving
Tray/Cup
Communications
screen
Report
Heading
System Options
screen
Request
Number
Communications
screen
RS-232 Ports:
Port Number 1
Port Number 2
Port Number 3
Communications
screen
4-55
Shows . . .
At the Options/
Run-Time
Reports
Printing Options
screen
Sample Tray
Format
System Options
screen
System
Language
System Options
screen
4-56
Shows . . .
At the Options/
Stop Bits
Communications
screen
System Clock
System Options
screen
Transmit
Results
Incrementally?
Communications
screen
XON/XOFF
Communications
screen
NOTE
4-57
Diagnostics/Maintenance Program
The Diagnostics/Maintenance program enables you to perform
several maintenance, calibration, and troubleshooting procedures
for your Express Plus. Refer to Section 9, Maintaining the
Express Plus, for complete maintenance information. Refer also
to Section 10, Troubleshooting the Express Plus, for complete
troubleshooting information.
Use the Diagnostics/Maintenance program to:
perform startup and end-of-day procedures
perform a precision check of the system
align the lamp
calibrate the probe tip
perform various manual operations of the system components
view the current system software, hardware, and firmware
Diagnostics/
Maintenance
Menu Screen
Precision
F1 Check
F1
Start-ofDay
F2
End-ofDay
Manual
F2 Operations
Prime
F3 Digital
Diluter
F4
Lamp
F3 Alignment
Select
Reagent
F5
Select
Sample
System
F4 Version
F6
Wash
Position
Probe Tip
F5 Calibration
F7
Load
Cuvette
Strip
F8
Purge
Cuvette
Track
4-58
Press . . .
F1 Precision
F2 Manual
Check
Operations
F2 End-of-Day
F3 Prime Digital
Diluter
F4 Select
Reagent
F5 Select
Sample
F6 Wash
Position
F7 Load Cuvette
Strip
F8 Purge Cuvette
Track
4-59
Press . . .
F3 Lamp
F4 System
F5 Probe Tip
Alignment
Version
Calibration
Diagnostics/
Maintenance
Precision
Check
Precision check
complete. Press
<EXIT> to select
another test, or
press <START> to
rerun.
Flag
Replicate
Number
2
4
6
8
10
12
14
16
18
20
Result
265.908
259.001
260.783
259.228
264.489
262.696
265.003
263.995
261.489
263.300
Flag
4-60
Diagnostics/
Maintenance
Lamp
Alignment
4-61
Diagnostics/
Maintenance
System Version
Main CPU
Software Version: MAIN V1.00
Operating System: MTSAPO V1.00
Sequencer
Software Version: SEQ V1.00
Operating System: MSQAPO V1.00
I/O Processor Version: IO V1.00
Machine Controllers
Track:
Probe:
Diluter:
Transports:
Hardware
3
3
3
3
Firmware
TRACK
V2.0
PROBE
V2.01
DILUTER
V2.0
TRANSPORT V2.08
4-62
Diagnostics/
Maintenance
Probe Tip
Calibration
4-63
Shows . . .
At the Diagnostics/
Maintenance
C.V.
Precision Check
screen
Firmware
System Version
screen
Flag
Precision Check
screen
Hardware
System Version
screen
I/O
Processor
Version
System Version
screen
Machine
Controllers
System Version
screen
Main CPU
System Version
screen
Mean
Precision Check
screen
Photometer
Reading
Lamp Alignment
screen
Replicate
Number
Precision Check
screen
Result
Precision Check
screen
S.D.
Precision Check
screen
Sequencer
System Version
screen
Test Name
Precision Check
screen
4-64
Load Alternate
Program
Screen
F1
Boot
Floppy
CAUTION
When you press F8 Load Alternate Program at the Main Menu, the
Load Alternate Program screen shows with the cursor at Express
Plus User Programs. Use this program only if you need to reload
your system software. Refer to Figure 4-20.
Figure 4-20. Load Alternate Program Screen
Load Alternate
Program
Alternate Programs
Express Plus User Programs
Calibrations and Manual Tests
Press <SELECT>
to select the
program indicated
by the cursor.
F1 Boot
Floppy
F2
F3
F4
F5
F6
F7
F8
5-1
Page
5-3
5-5
5-11
5-12
5-14
Defining Profiles
5-24
5-26
Defining Controls
5-27
Defining Calibrators
5-30
Defining Blanks
5-33
5-37
5-41
5-2
Refer to . . .
Tab:
Setting Installation Options
Tab:
Setting Installation Options
Tab:
Defining and Customizing
Parameters
Tab:
Backing Up the Settings
Tab:
Backing Up the Settings
5-3
5-5
Options
System Options
Press <EXIT> to
store what you
modified.
Options:
Positive Sample Identification:
DISABLED
Express Communications Package: DISABLED
5-6
(hours:minutes)
NOTE
5-7
Options
Printing
Options
Run-Time Reports
Calibrator/Control Log: ON
Error Log: ON
Result Log: ON
Send Run-Time Reports to default printer: OFF
Default Printer: System
Collated Reports
Automatic print for Stats: ON
Automatic print for routine samples:OFF
Press <EXIT> to
store what you
modified.
Report Heading:
General Laboratories
123 Main Avenue
Oberlin, Ohio 44012
5-8
5-9
5-10
Options
Test Order
Alb
TBili
Creat
Trig
ACP_T
TIBC
ALP
BUN
GGT
UA
ACP_P
ALT
Ca
Gluc_HK
HDL
Mg
Amy
Chlor
IPhos
HBD
Fe
AST
Chol
LDH_L
CO2
ExFe
DBili
CK
TP
ACP_NP
UIBC
Press <SELECT> to
select the test
that you want to
move.
Press <EXIT> to
store what you
modified.
12. Move the cursor to the test that you want to show after the
selected test in the test order, then press SELECT again. The
test that you selected moves to the position you indicate.
NOTE
5-11
Press . . .
Then . . .
F1 Test
go to step 3 of this
procedure
profiles
F2 Profile
go to Defining
Profiles
controls
F3 Control
go to Defining
Controls
calibrators
F4 Calibrator
go to Defining
Calibrators
blanks
F5 Blank
go to Defining
Blanks
tests
reagents
assay order
calibrator concentrations
controls
QCSystems
5-12
Definitions
Tests
Alb
TBili
Creat
Trig
ACP_T
TIBC
ALP
BUN
GGT
UA
ACP_P
ALT
Ca
Gluc_HK
HDL
Mg
Amy
Chlor
IPhos
HBD
Fe
AST
Chol
LDH_L
CO2
ExFe
DBili
CK
TP
ACP_NP
UIBC
Press <SELECT> to
select the test
indicated by the
cursor.
F1 Insert
F2 Erase
F3 Restore F4
F5
F6
F7
F8
5-13
Then . . .
NOTE
No:
5-14
Defining Test
Parameters
Definitions
Test: Alb
Tests
Parameters
F1
F2 Reagents F3 Assay
Order
F4 Calib.
Concs.
F5 Controls F6 QCSystem F7
F8
Defining Test
Parameters
5-15
(Continued)
Definitions
Test: ACP_P
Tests
Parameters
Ratio Definition:
ACP_T - ACP_NP
Review data then
modify if
required.
Press <EXIT> to
store the
definition or
review another
test.
Definitions
Test: Na
Tests
Parameters
5-16
Defining Test
Parameters
(Continued)
Press . . .
Then go to . . .
F2 Reagents
F3 Assay
F4 Calib.
F5 Controls
F6 QCSystem
NOTE
Order
Concs.
Defining Test
Parameters
(Continued)
5-17
Definitions
Test: Alb
Tests
Reagents
Sample Volume: 3 L
Rerun Dilution Ratio: 1
Predilution Ratio: 1
F1 Param- F2
eters
7.
Rgnt 1:
Rgnt 2:
Rgnt 3:
Rgnt 4:
Reagent Bar
Volume Code
300 uL AB1A
uL
uL
uL
F3 Assay
F4 Calib.
Order
Diluent
Volume
uL
uL
uL
uL
Lag
Time
60 secs
secs
secs
secs
F5 Controls F6 QCSystem F7
Bottle
Type
Plastic
Plastic
Plastic
Plastic
F8
Concs.
5-18
Defining Test
Parameters
NOTE
(Continued)
Definitions
Tests
Assay Order
Select a wash
fluid and the
number of washes
for the test(s)
that interfere
with the test
being defined.
Test
Alb
ALT
AST
TBili
Ca
Chol
Creat
Gluc_HK
LDH_L
Trig
HBD
CO2
Press <EXIT> to
store definition
or review another
test.
F1 Param- F2 Reagents F3
eters
Wash
Fluid
NONE
NONE
NONE
NONE
Fluid 1
NONE
NONE
NONE
NONE
NONE
NONE
NONE
F4 Calib.
Concs.
Wash
Count
1
1
1
1
2
1
1
1
1
1
1
1
Test: Alb
Fluid 2 Bar Code:
Test
ALP
Amy
DBili
BUN
Chlor
CK
GGT
IPhos
TP
UA
HDL
ACP_NP
F5 Controls F6 QCSystem F7
Wash
Fluid
NONE
NONE
NONE
NONE
NONE
NONE
NONE
NONE
NONE
NONE
NONE
NONE
Wash
Count
1
1
1
1
1
1
1
1
1
1
1
1
F8 More
Defining Test
Parameters
(Continued)
5-19
Definitions
Tests
Calibrator
Concentrations
Conc. 2
1.0
Conc. 3
2.0
Test: NAPA
Conc. 4
4.0
Conc. 5
8.0
Conc. 6
16.0
Order
F4
F5 Controls F6 QCSystem F7
F8
5-20
Defining Test
Parameters
(Continued)
Definitions
Tests
Controls
Control Rules:
1-3s
2-2s
Order
NOTE
4-1s
QCSAbn
F4 Calib.
F5
Test: Alb
Uniqual
10x
Expected Mean
F6 QCSystem F7
F8
Concs.
a. Move the cursor to each control to use for the test, then
press SELECT. You can select up to six controls.
b. Move the cursor to each control rule to use for the
controls you selected, then press SELECT.
The system automatically uses the 1-2s control rule for all
tests; that rule does not show on this screen.
For information about using the Automatic Test
Shutdown option, refer to the Section 5 tab, Setting
Installation Options, and to Section 3, Understanding the
System Calculations.
Defining Test
Parameters
(Continued)
5-21
5-22
Defining Test
Parameters
(Continued)
Then . . .
Test Code
Method Code
Units Code
Reagent Code
Definitions
Tests
QCSystems
Test: Alb
Test Code: 6
Method Code: 2
Units: g/dL
Units Code: 10
Reagent Code: 99
Press <EXIT> to
store definition
or review another
test.
Order
F4 Calib.
Concs.
F5 Controls F6
F7
F8
Defining Test
Parameters
5-23
NOTE
(Continued)
11. After you define or modify all the required parameters for
the test definition, press EXIT to store the definition in system
memory and return to the Definitions/Tests screen.
12. At the Definitions/Tests screen:
To . . .
Then . . .
NOTE
5-24
Defining Profiles
At the Definitions/Profiles screen, shown in Figure 5-13, you can:
select the profile that you want to modify
define a new profile
erase a defined profile
Figure 5-13. Definitions/Profiles Screen (Example)
Definitions
Chem
Liver
Cardiac
Renal
Lytes
Profiles
Press <SELECT>
to select the
profile indicated
by the cursor.
F1 Insert
F2 Erase
F3
F4
F5
F6
F7
F8
Then . . .
NOTE
5-25
Definitions
Profiles
Selection
Press <SELECT>
to select the test
indicated by the
cursor.
ALP
BUN
GGT
UA
ACP_P
ALT
Ca
Gluc_HK
HBD
Mg
Amy
Chlor
IPhos
HDL
Fe
AST
Chol
LDH_L
CO2
ExFe
DBili
CK
TP
ACP_NP
UIBC
Press <EXIT> to
store definition
or review another
profile.
5-26
Then . . .
erase a profile
5-27
Defining Controls
At the Definitions/Controls screen, shown in Figure 5-15, you
can:
select the control that you want to modify
define a new control
erase a defined control
Figure 5-15. Definitions/Controls Screen (Example)
Definitions
Normal
QCSNorm
QCSAbn
Lipid
Bili
Controls
Press <SELECT> to
select the control
indicated by the
cursor.
F1 Insert
F2 Erase
F3
F4
F5
F6
F7
F8
Then . . .
erase a currently-defined
control
NOTE
5-28
5-29
Then . . .
erase a control
Definitions
Control
Limits
F1
F2
Expected
Test
Mean
Alb
3.3
ALT
AST
TBili
0.7
Ca
Chol
141.
Creat
Gluc_HK 88.
LDH_L
Trig
74.
HDL
CO2
F3
F4
Expected
Expected Expected
1 SD
Test
Mean
1 SD
0.30
ALP
Amy
DBili
0.1
0.10
0.20
BUN
16.0
3.00
Chlor
98.
4.0
21.0
CK
GGT
9.0
IPhos
TP
5.5
0.30
11.0
UA
5.1
0.50
HBD
ACP_NP
F5
F6
F7
F8 More
5-30
Defining Calibrators
At the Definitions/Calibrators screen, shown in Figure 5-17, you
can:
select the calibrator that you want to modify
define a new calibrator
erase a defined calibrator
Figure 5-17. Definitions/Calibrators Screen (Example)
Definitions
Multi1
Multi2
Calibrators
Press <SELECT> to
select the
calibrator
indicated by the
cursor.
F1 Insert
F2 Erase
F3
F4
F5
F6
F7
F8
Then . . .
erase a currently-defined
calibrator
modify a defined
calibrator
NOTE
5-31
Definitions
Calibrator
Concentrations
5-32
Then . . .
erase a calibrator
5-33
Defining Blanks
At the Definitions/Blanks screen, shown in Figure 5-19, you can:
select the blank that you want to modify
define a new blank
erase a defined blank
Figure 5-19. Definitions/Blanks Screen (Example)
Definitions
H2O1
H2O2
H2O3
Blanks
Press <SELECT>
to select the
blank indicated
by the cursor.
F1 Insert
F2Erase
F3
F4
F5
F6
F7
F8
2.
3.
Then . . .
erase a currently-defined
blank
NOTE
5-34
Go to step 5.
No:
Definitions
Blanks
Selection
Press <SELECT> to
select the test
indicated by the
cursor.
Press <EXIT> to
store the definition
or review another
blank.
ALP
BUN
GGT
UA
ACP_P
ALT
Ca
Gluc_HK
HBD
Mg
Amy
Chlor
IPhos
HDL
Fe
AST
Chol
LDH_L
CO2
ExFe
DBili
CK
TP
ACP_NP
UIBC
5-35
Then . . .
erase a blank
5-37
NOTE
5-38
NOTE
4.
If you are
backing up . . .
5-39
Options
Disk Options
Press the
required function
key.
5-40
Place the write-protect tab on each backup disk in the readonly position.
5-41
NOTE
5-42
2.
3.
4.
verifies that the backup disks you insert are from a set of
backup disks that were made at the same time.
verifies that you insert each backup disk from the set of
disks in the correct sequence.
5.
6.
Remove the backup disk from the disk drive and store the
backup disk(s) in a safe place.
NOTE
5-43
6-1
Page
6-5
Operating Summary
6-5
6-7
6-11
6-13
6-14
6-14
6-15
6-16
6-20
6-21
6-22
6-23
6-24
6-25
6-26
6-2
Page
6-29
6-30
6-31
6-33
6-34
6-36
6-37
6-39
6-43
6-43
6-45
6-47
6-51
6-3
Refer to . . .
Tab:
Preparing the Express Plus
Starting Up
Tab:
Preparing the Express Plus
Tab:
Preparing the Express Plus
Tab:
Assaying the Scheduled
Samples
Starting the Run
Loading the Next Sample
Tray(s)
Tab:
Assaying the Scheduled
Samples
Loading Stats or Single
Samples During a Run
Tab:
Assaying the Scheduled
Samples
Manually Rerunning and
Diluting Samples
Tab:
Assaying the Scheduled
Samples
Reviewing Run-time Results
(Continued on the next page)
6-4
Refer to . . .
Tab:
Managing the Patient Results
Reviewing Sample Results
Editing Sample Results
Tab:
Managing the Patient Results
Printing Sample Results
Tab:
Performing the End-of-Day
Tasks
Shutting Down
6-5
Operating Summary
The following steps briefly describe the daily operating
procedures for your Express Plus. The procedures in this section
provide the details.
To . . .
Then . . .
6-6
Then . . .
6-7
To . . .
HELP
STATUS
MAIN MENU
F4 Sample Status
F5 Pending Status
F6 Communications Status
Worklist
Patient Results
Calibration Data
Q.C. Statistics
F5
F6
F7
F8
Definitions
Options
Diagnostics/Maintenance
Load Alternate Program
PRINT SCRN
EXIT
CANCEL
SELECT
ENTER
DELETE
CURSOR Keys
6-9
To . . .
start the testing sequence from the beginning, or
resume testing that you stopped or paused. When
you press START, the Express Plus:
primes the lines if you filled the distilled water
bottle
verifies the system operating temperatures
goes to the Warming Up or Running state
identifies the reagents on the reagent tray
waits for you to verify Reagent Loading
p use
6-11
Starting Up
To start up your Express Plus at the beginning of the day or shift:
1. Power up the system:
If the Express Plus is . . .
Then . . .
powered ON (I)
go to step 2.
Startup
If you removed
the waste water
bottle, install the
bottle, then press
<ENTER>.
NOTE
12:31
6-13
Starting Up (Continued)
2. c. Verify that the waste bottle is empty, then install the
empty waste bottle in the system (refer to Figure 6-4).
d. Verify that there are no used cuvettes are in the cuvette
waste drawer. If there are used cuvettes in the waste
drawer, remove the biohazard liner and safely discard it,
then install a clean biohazard liner in the cuvette waste
drawer and install the drawer in the system (refer to
Figure 6-4).
e. Close the front panel.
f.
6-14
Erasing a Stored
Worklist
CAUTION
Worklist
Tray/Cup Number: A2
Tests
Press <SELECT>
to select the blank
indicated by
the cursor.
F1 Next
Request
Request Number: 2
Blanks:
Blank 1
Blank 6
Tests:
Alb
TBili
Creat
TRIG
ACP_T
TIBC
F2 Previous F3 Find
Request
Request
Blank 2
Blank 7
ALP
BUN
GGT
UA
ACP_P
Blank 3
Blank 8
ALT
Ca
Gluc_HK
HBD
Mg
F4 Insert
Request
Blank 4
Blank 9
Amy
Chlor
IPhos
HDL
Fe
F5 Erase
Request
Blank 5
Blank A
AST
Chol
LDH_L
CO2
ExFe
F6 Load
Sample
F7
DBili
CK
TP
ACP_NP
UIBC
F8
Erasing a Stored
Worklist
(Continued)
6-15
Reviewing Calibration
Status
Before you schedule tests and samples in the worklist, you might
want to review the calibration status of the tests you require:
1. At the Main Menu, press F3 Calibration Data.
2. Review the calibration status of the tests you need to
schedule in the worklist:
If the status is . . .
Then . . .
CURRENT,
NORMALIZED, or
REBLANKED
6-16
Scheduling Samples
in the Worklist
on tray A
A1/90 the system schedules the same tests for cups 1 through 40
press DELETE
to delete the
current number
And . . .
Random access testing:
type the sample tray identifier
(A through I), then type the cup
number (1 through 40).
Batch testing:
type the sample tray identifier
(A through I), the starting cup
number, a slash (/), then the total
number of samples for the batch.
Request
Number
press DELETE
to delete the
request number
Scheduling Samples
in the Worklist
(Continued)
6-17
3. Move the cursor to the Sample Type field, then press SELECT
to select the sample type for the current request.
If you select Blank, Control, or Patient, go to step 5.
If you select Calibrator, take the required actions:
If you want to . . .
And . . .
Then . . .
go to step 4.
schedule a multicalibrator
calibrate the
curve
go to step 5 and
schedule the multicalibrator.
reblank a
go to step 5 to schedule
the blank (zero
calibrator) .
stored, blanked
linear curve
NOTE
And . . .
Then . . .
the calibrator
concentrations are
defined for the test
the calibrator
concentrations are
defined for the
test, and you
selected the
normalizers to use
to normalize the
test
Go to step 6.
the Express Plus waits
for you to select the
calibration you need:
1. Press F1 Calibrate to
schedule a full
calibration or,
press F2 Normalize
to normalize the
stored curve.
2. Go to step 6.
the calibrator
concentrations are
defined for the
test, but you did
not select
normalizers to use
for the test
6-18
Scheduling Samples
in the Worklist
4.
(Continued)
If . . .
And . . .
Then . . .
the calibrator
concentrations are
defined for the test
1. Press F1 Calibrate to
schedule a full
calibration, or press
F2 Reblank to reblank
the stored curve.
(Continued)
2. Go to step 6.
calibrator
concentrations
are not defined
for the test
1. Press
F8 Concentrations.
5. Move the cursor off the Sample Type field to show the
blanks, calibrators, controls, or profiles and tests, then take
the required actions to schedule the worklist request for the
selected sample type:
If the Sample
Type is . . .
Move the
cursor to . . .
Blank
Calibrator
the defined
calibrator to
schedule for the
request
Control
Patient
1. the Patient
Name field
2. the defined
profiles and/or
tests to schedule
for the patient
request
(multi-calibrator)
And . . .
Scheduling Samples
in the Worklist
(Continued)
6-19
6-20
Entering Patient
Demographics
You can enter patient demographics while the system assays the
samples. To type the demographics for each patient request in
the worklist:
1. At the Worklist/Tests screen, press F8 Demographics.
2. At the Worklist/Demographics screen, type the patient
demographic information:
Move the
cursor to . . .
Then type . . .
Patient Name
Patient ID
Age
Sex
Comments
6-21
Press . . .
F1 Next
F2 Previous
F3 Find
F4 Insert
erase a request;
these Erase function keys show on the screen:
F1 Erase Worklist
to erase the stored worklist
F2 Erase Request
to erase selected request(s)
F3 Patient Name
to erase by name
F4 Patient ID
to erase by ID
F5 Erase
F6 Load
Request
Request
Request
Request
Request
Sample
6-22
13:28
Cup
Request
Sample Type
Patient Name
Patient ID
B14
B15
B18
B38
C1
C2
C5
C6
C7
112233
112256
968245
747220
222994
223344
234567
234568
234569
Patient
Patient
Control
Calibrator
Patient
Blank
Patient
Patient
Patient
Donna J. Burrer
Sally Brown
Control 3
Calib 1
Jeri Scheer
Blank 1
Nancy Swiz
Rose Freeman
Scott Mathis
123-45-6789
736-45-9876
250-91-6449
248-50-4199
249-51-2121
234-13-3939
6-24
NOTE
Load the sample cups into the scheduled sample tray positions:
1. Remove the 40-position sample tray cover.
2. Dispense the required volume of samples into sample cups.
NOTE
3. Use the sample loadlist to load each sample into the correct
cup position on the required sample tray.
4. Install the tray of samples on the transport.
5. Remove the humidifier ring from the tray cover, as shown in
Figure 6-8.
6. Moisten the blue felt with a laboratory wash bottle containing
deionized water.
7.
8.
9.
6-26
rotates the reagent tray and reads the bar codes on the
reagent bottle labels to identify the reagents that are
loaded and the positions on the reagent tray.
Reagent Reagent
Position Bar Code Test
reagent use.
Press <ENTER> to
continue.
F1
F2
1
2
3
4
5
6
7
8
9
10
11
12
13
F3 Waive
AB1A
ABC1
AP1B
CH1B
CR1A
TG1B
CA1A
GL1A
CB1A
AC2B
BU1A
NA1B
WS1A
F4
Reagent
Number Calibration
Bottle
Type
Plastic
Plastic
Diluent
CURRENT
Glass
ALP
Chol
1
CURRENT
Plastic
IN PROCESS Plastic
Creat
1
Trig
1
EXPIRED
Plastic
Plastic
*
*
Gluc_HK 1
REBLANKED Glass
INVALID
Special
Carb
2
Plastic
*
*
SCHEDULED Plastic
BUN
1
NAPA
2
NORMALIZED Glass
Wash Fluid
Plastic
*
Alb
F5
F6
EXPIRED
F7
F8 More
6-27
Then . . .
go to Definitions/Tests/Reagents
screen to modify the bottle type.
go to Definitions/Tests/Reagents
screen to modify dispensing
sequence, or to Definitions/Tests/
Assay Order screen to modify
reagents used as wash fluids.
go to the Definitions/Tests/
Reagents screen to modify
test/reagent combinations.
press F8 More.
6-29
After you load your sample trays and reagent trays, use this
procedure to start the run of samples scheduled in the worklist:
1. At the Reagent Loading screen, press ENTER to start the run
for the first sample tray.
2. These tasks can take place while the system assays samples:
To . . .
correct an operating or
Status error
6-30
3. Press START.
If you removed the reagent tray cover in step 2, the system
scans the reagent bar code labels, then shows the next
Reagent Loading screen.
Verify the reagent loading information, if required, then press
ENTER (refer to the Verifying Reagent Loading procedure).
4. Type the tray identifier for the sample tray that you loaded
on the system in step 1.
5. Repeat steps 1 through 4 for each sample tray of scheduled
samples that you need to run to complete the worklist.
6-31
4.
5.
6.
Worklist
Tests
Request Number: 30
Stat? Yes
Press <SELECT>
to select the blank
indicated by
the cursor.
F1 Next
F2 Previous F3 Find
Request
7.
Tests:
Alb
TBili
Creat
TRIG
ACP_T
TIBC
Request
Request
Liver
ALP
BUN
GGT
UA
ACP_P
Lytes
ALT
Ca
Gluc_HK
HBD
Mg
F4 Insert
Request
Routine
Amy
Chlor
IPhos
HDL
Fe
F5 Erase
Request
AST
Chol
LDH_L
CO2
ExFe
F6 Load
DBili
CK
TP
ACP_NP
UIBC
F7
Sample
Open the Stat cover to load the sample; do not remove the
sample tray or tray cover from the system to load the sample.
NOTE
F8
6-32
6-33
System Status
Pending Status
Press required
function key for
further status
information, or
press <EXIT> to
exit Status.
Currently Dispensing:
Tray/Cup Number: A10
Operation: Initial dispense
Pending Tests:
Creat
Alb
DBili
TBili
CK
GGT
Trig
UA
ACP_T
ACP_P
TIBC
Details
ALP
BUN
Gluc_HK
HDL
Mg
Status
Status
F5
ALT
Ca
IPhos
HBD
Fe
Amy
Chlor
LDH_L
CO2
ExFe
F6 Communi-
AST
Chol
TP
ACP_NP
UIBC
F7
cations
3. Press EXIT to return to the screen where you started (step 1).
F8
6-34
Is rerun if . . .
And is rerun . . .
patients and
controls
once
6-35
Dilution?
Control out
No
No
Yes
No
High Concentration.
Yes
Yes
No
No
No
No
Outlier
No
Over
Yes
No
No
No
No
Under
No
6-36
3.
4.
5.
6.
7.
Open the Stat cover on the tray cover to load the sample; do
not remove the cover or tray from the system.
NOTE
8.
9.
6-37
NOTE
Status:
Cuvettes Used: 20
Standby
Maximum: 195
Currently Dispensing:
Tray/Cup Number: A10
Operation: Initial dispense
Error detected
Press Status
Test Name: Glucose
Status
Status
Status
cations
F7
F8
6-38
Press required
function key for
further status
information, or
press <EXIT> to
exit Status.
Errors shown are
the most recently
detected errors.
F1 System F2
Status
Status:
Cuvettes Used: 20
Error detected
Standby
Maximum: 195
Press Status
System Status: Distilled water empty
Cuvette Waste Drawer: Closed
Water Bath: OK
Cuvettes in Feeder: 170
Waste Water: OK
Reagent Cover: Installed
Wash Water: LOW
Sample Cover: Sample cup cover installed
C
Subsystems
Track: OK
Probe: OK
Valve: OK
Diluter: OK
Wash Pump: OK
Waste Pump: OK
Sample Transport: OK
Reagent Transport: OK
Photometer: OK
Lamp: Standby
Track: Ready
Water Bath: Ready
Cuvette Shuttle: Standby
F3 Reagent F4 Sample
Status
Status
F5 Pending F6 CommuniStatus
F7
cations
F8
6-39
Table 6-5 explains the flags that can show on the run-time reports
for patient results (some of the same flags can also print for
calibrators, controls, and blanks).
6-40
Result
Cup Number:
Patient Name:
Gluc_HK
Result:
25-Mar-1992
10:30
A3
Request Number 23
Maggie Brown
A, dA, dA/T:
0.8326
87.1
Cup Number:
Patient Name:
Gluc_HK
Result:
25-Mar-1992
10:31
A4
Request Number 24
Norma Schmidt
A, dA, dA/T:
0.3393
120.4 HIGH
Flag
Response
6-41
Means . . .
Diluted Rerun
HIGH
High absorbance
limit exceeded.
High
Concentration.
Invalid
LOW
Low absorbance
limit exceeded.
Noise limit
exceeded.
Nonlinearity limit
exceeded.
Outlier
Over
Primary
wavelength
abs. is high.
6-42
Means . . .
Primary
wavelength
abs. is low.
Rerun
Secondary
wavelength
abs. is high.
Secondary
wavelength
abs. is low.
Under
6-43
off-system test
print the collated patient reports
Reviewing Sample
Results
Press . . .
Then . . .
Main Menu
F2 Patient
Results
Worklist/Tests screen
F7 Results
Patient Results
Request Number: 27
F1 Next
Request
Result
Test
Alb
ALT
35.9
AST
TBili
1.2
Ca
Chol
115.2
Creat
3.5
Gluc_HK 85.25
LDH_L
Trig
HBD
CO2
F2 Previous F3 Find
Request
Request
Flag
Test
Result
ALP
Amy
DBili
BUN
8.7
Chlor
CK
55.9
GGT
IPhos
TP
UA
HDL
ACP_NP
High
Low
Dil. rerun
F4 Edit
Request
F5
F6
Flag
Nonlinear
F7 Worklist F8 More
6-44
Reviewing Sample
Results
(Continued)
To . . .
Press . . .
F1 Next
F2 Previous
F3 Find
Request
Request
Request
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
F7 Worklist
F8 More
Then . . .
6-45
Press . . .
Then . . .
Main Menu
F2 Patient
Results
Worklist/Tests screen
F7 Results
Press . . .
F1 Next
F2 Previous
F3 Find
Request
Request
Request
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
F7 Worklist
F8 More
6-46
Then . . .
press EXIT
6-47
Press . . .
Then . . .
Main Menu
F2 Patient
Results
Worklist/Tests screen
CAUTION
F7 Results
NOTE
6-48
(Continued)
To . . .
Press . . .
Then . . .
F1 Current
F2 Patient
F3 Patient
F4 Cup
F8 All
Request
Name
ID
Range
Requests
Press . . .
F1 Next
F2 Previous
F3 Find
Request
Request
Request
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
F7 Worklist
6-49
ID: 123-27-4567
Age: 10/26/68
Sex: F
Fasting sample
Time of draw: 7:15AM
Room 221A
Pre-op patient for 16-JAN
Physician: Dr. Seth Daniels
Insurance: Health Insurance Inc.
Test
Result
Flags
Low
High
BELOW NORM
3.5
7.0
0.7
65.
6.0
5.3
18.0
1.5
110.
8.3
ABOVE NORM
37.
11.8
13.2
147.
29.4
33.0
8.4
140.
2.7
89.
10.2
220.
4.5
221.
Profile: profile1
Albumin
BUN
Creatinine
Glucose
Total Protein
0.1
Pending
Pending
Pending
6.5
mg/dL
Profile: profile2
Alk. Phos
ALT
AST
160.2
25.5
33.0
U/L
U/L
U/L
Tests:
Calcium
Cholesterol
Phosphorus
LDH
Pending
Pending
3.7
200
g/dL
mg/dL
U/L
ABOVE NORM
6-50
6-51
Shutting Down
At the end of the day or the end of each work shift, perform the
following procedure to clean the Express Plus:
1. At the Main Menu, press F7 Diagnostics/Maintenance.
The Diagnostics/Maintenance screen shows.
2. At the Diagnostics/Maintenance menu screen, press
F2 Manual Operations.
The Manual Operations screen shows.
3. To begin the manual operations, press ENTER.
The system homes the probe and transports.
4. Load a bottle of deproteinizing solution in position 1 on the
reagent tray, then install and secure the reagent tray cover.
5. At the Manual Operations screen, press F2 End-of-Day, then
follow the instructions shown in the prompt area of the
screen.
The system begins the end-of-day tasks by cleaning the probe
as follows:
a. The system moves the probe to the wash station, then
pumps approximately 5.0 mL of distilled water into the
wash station.
b. The system rotates the deproteinizing solution to the
aspiration position, aspirates 500.0 L of solution, then
dispenses the deproteinizing solution into the wash
station.
c. The system repeats steps 5a and 5b five times.
d. The system pumps approximately 5.0 mL of distilled
water into the wash station. The probe remains at the
wash station after this cleaning procedure is completed.
e. The system then sets the sample tray to A and purges all
the cuvette strips (except one) from the cuvette track into
the cuvette waste drawer.
6-53
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Rev. B 1/2000
Glossary
A-1
Glossary
Absorbance (A)
Accessory
Alphanumeric Keys
Automatic Test
Shutdown
Bar Code
Bichromatic
Absorbance
BIOHAZARD
Calibration Data
Program
Carryover
CAUTION
Collated Report
A-2
Glossary
Glossary (Continued)
Communication Ports
Communications
Subsystem
Concave Holographic
Grating
CPU (Central
Processing Unit)
Cursor
Cursor Keys
Cuvette Magazine
Cuvette Strip
Cuvette Subsystem
Cuvette Track
Glossary
A-3
Glossary (Continued)
Default
Definitions Program
Demographics
Diagnostics/
Maintenance Program
Digital Diluter
External Printer
Factory-Programmed
Fluid Subsystem
Function
Function Keys
A-4
Glossary
Glossary (Continued)
Hard Disk Drive
Help Program
Highlighted
Humidifier Ring
Keyboard
Lag Time
LIS 2 Protocol
LIS 3
LIS Device
Main Menu
Menu Screen
Glossary
A-5
Glossary (Continued)
Normalize
NOTE
Numeric Keypad
Options Program
Outlier Rejection
Patient Results
Program
Photodiode
Photometer Subsystem
Placekeeping Zone
Polychromator
Positive Sample
Identification
A-6
Glossary
Glossary (Continued)
Power Switch
Printing Options
Probe
Profile
Program
Program Zone
Prompt
Prompt Zone
Purge
Q.C. Statistics
Glossary
A-7
Glossary (Continued)
QCSystems
Ratio Test
Reaction Mixture
Reagent Carryover
Reduction
Reagent Transport
Reagent Tray
Reblanking
Remove
Replicate
Reverse Video
Rule Evaluation
Run-time Report
A-8
Glossary
Glossary (Continued)
Sample Cup
Sample Loadlist
Sample Transport
Sample Type
Scan
Screen
Screen Field
Screen Zones
Select
Startup
Glossary
A-9
Glossary (Continued)
Status Program
Status Zone
System Clock
System Keys
System Printer
Test Scheduler
WARNING
Waste Bottle
Water Bath
Worklist
Worklist Program
Glossary
Safety Summary
B-1
Safety Summary
The Express Plus is designed to operate safely. When you
operate the system in a safe environment, there are no known
operating hazards.
Warnings are used in this manual to alert you to electrical
hazards and other conditions that can lead to personal injury.
Ensure that you read and follow the instructions contained in
these warnings.
Warnings are shown on the following pages of this manual:
WARNING
2-15
9-31
9-5
9-32
9-6
E-3
9-8
G-4
9-26
G-5
9-27
G-9
B-2
Safety Summary
2-15
9-8
9-27
10-68
2-18
9-12
9-32
C-8
6-12
9-16
9-33
G-7
6-24
9-19
9-37
G-12
6-52
9-21
9-43
9-5
9-26
10-18
Safety Summary
B-3
B-4
Safety Summary
3-20
5-39
6-47
9-7
9-27
9-45
10-18
C-8
3-25
5-41
7-10
9-9
9-31
9-46
10-29
G-3
4-50
5-42
7-44
9-18
9-34
9-48
10-37
G-4
4-64
6-7
9-3
9-22
9-38
9-50
10-64
G-5
5-5
6-12
9-5
9-23
9-40
9-51
10-65
G-6
5-37
6-14
9-6
9-26
9-43
10-3
10-68
G-9
Safety Summary
B-5
It means . . .
alternating current (AC)
ON (Mains supply)
OFF (Mains supply)
Fuse
Keyboard
Water
This symbol indicates the system meets the
EMC Directive of the European Union.
This symbol indicates the system meets the
specifications of the Underwriters
Laboratories, Inc.
Safety Summary
C-1
Page
C-3
C-3
C-5
C-6
Ordering Supplies
C-7
C-8
C-3
Bayer Inc.
Bayer Diagnostics
Healthcare Division
77 Belfield Road
Toronto, Ontario
Canada M9W 1G6
416 248-0771
Bayer S.A.
Divisin Diagnstica
A.A. 80387
Av. De las Amricas No57-52
Santaf de Bogot, D.C. - Colombia
(571) 4234199-4234352
Bayer A/S
Norgaardsvej 32
DK-2800 Lyngby, Denmark
+ 45 45 23 50 00
Bayer OY
Suomalaistentie 7
FIN 02270 Espoo, Finland
+ 35 89 88 78 87
Bayer Diagnostics
Tour Horizion
52, quai de Dion Bouton
92807 Puteaux Cedex, France
01 49 06 56 00
Bayer Vital GmbH & Co. KG
Geschftsbereich Diagnostics
Siemensstrae 3
D-35463 Fernwald, Germany
0049-(0)641-4003-0
C-4
Bayer AS
Brennaveien 18
N-1483 Skytta, Norway
+ 47 67 06 86 00
C-5
Part
Number
Quantity
23517
23174
23378
107099
107100
107098
107101
107125
107122
107123
24067
24167
23481
23482
Paper Pin
23490
23542
23513
Probe Tip
1453x185
23334
Reagent Cover
24078
24076
Sample Cover
1453x261
1453x249
24162
1453x225
23361
Waste Bottle
23379
C-6
Part
Number
Quantity
1453x7
1453x125
23626
60048x57
60044x12
Lamp Assembly
1453x135
1453x156
Probe Tip
1453x185
19819
1 piece
E34031
* You can order the Operators Service Kit as a kit, or order individual
parts contained in the Operators Service Kit.
C-7
Ordering Supplies
This list shows the Express Plus supplies that you can order:
Table C-3. Supplies
Description
Part
Number
Quantity
23531
751060
20764
1 box
24068
60048x60
60048x57
60044x13
60044x12
1453x135
751063
1 box
23520x13
23520x16
751058
1 box
Consumables Kit
1453x506
751059
1453x108
Lamp Assembly
1453x135
23490
1453x156
Probe Tip
1453x185
751058
1 box
23372
1 package
BIOHAZARD
CAUTION
C-8
Part
Number
Quantity
23678
1453x510
107112
24067
107113
24068
1453x507
23531
1 set
23409
23363
107087
1453x226
23520x13
20
23520x16
20
23513
23542
Test Scheduler
23541B
23511
D-1
Page
D-3
D-4
D-5
D-3
D-4
Part Number
D-5
and controls defined for the Express Plus, which allows for
faster, more efficient worklist scheduling
enter data and select options without using the keyboard,
D-6
System (LIS)
For complete information about cabling, configuring the ports,
enabling the software option, and programming for
communications, refer to the following documents that are
shipped with the communications software option:
Express Plus Communications Enabling Disk Instructions
Express Communications Manual
Express Interface Specifications Manual
Appendix F, LIS Communications, provides a brief overview of
the Express Plus communications options.
To order accessories or software options, call Bayer Diagnostics.
Refer to Appendix C, Service and Supplies, for Bayer
Diagnostics addresses and telephone numbers, and part numbers
for placing your orders.
Specifications
E-1
Appendix E Specifications
Page
E-3
System Specifications
E-3
E-7
Environmental Conditions
E-7
Microprocessor (CPU)
E-8
Outputs
E-8
Polychromator
E-8
E-9
System Screen
E-9
Specifications
Specifications
E-3
Agency Approvals
European Union: CE Marking
E-4
Specifications
Agency Approvals
International Electrotechnical Commission (IEC)
System Specifications
These specifications apply to the Express Plus, in general, and are
shown in alphabetical order:
Application
Controls
Cuvette Feeder
Cuvette Magazines
Cuvette Preheater
Cuvette Strip
Cuvette Track
Specifications
E-5
Drawer
Data Storage
Digital Diluter
Dispense Volume
Distilled Water
Bottle
External Printer
Imaging Optics
Keyboard
Language
user-selectable
Light Source
Machine Controller
Pediatric Cup
Polychromator
Primary Sample
Tubes
Primary Sample
Tube Tray
E-6
Specifications
temperature-controlled, water-jacketed
probe for heating reaction mixtures
before dispensing into cuvette wells;
probe tip uses a replaceable stainless steel
aspiration tip and fluid detector
Productivity
Reaction Temp.
Reaction Time
Reagent Bar
Code Scanner
Reagent Bottles
Reagent Capacity
Reagent Transport
Reagent Tray
Reagent Volume
Sample Cup
Sample Transport
Sample Integrity
Tray
Sample Volume
Specifications
E-7
Startup
Stat Assays
System Control
System Printer
System Type
Test Calibration
Test Selection
Thermal Electric
Devices (TEDs)
Water Bath
Wash Station
Waste Bottle
Wavelengths
E-8
Specifications
Nominal Voltages
Fuse Ratings
100 V
120 V
200 V
240 V
7 amp, T (SB)
6.25 amp, T (SB)
4 amp, T (SB)
3.15 amp, T (SB)
Power Consumption
Electrical Requirements
Environmental Conditions
NOTE
Intended Environment
Ambient Operating
Temperatures
40 to 80% (maximum),
noncondensing
Storage Temperature
0.0 to 50.0 C
Installation Category
Specifications
E-9
Microprocessor (CPU)
Main CPU
Mass Storage
Parallel
RS-232C
Screen
System Printer
Accuracy
2.0 nm
Bandwidth
9 nm
Detector
Light Source
Linearity
1%
Measurement
bichromatic measurement
Noise
S.D. of 8 x 10-5
Range
Reference Band
Stray Light
Warmup Time
Wavelengths
Outputs
Polychromator
E-10
Specifications
Height
Weight
Width
Screen
Characters
Display Lines
24 Lines
Equipment
System Screen
Classification
LIS Communications
F-1
Page
F-3
F-4
F-5
Physical Interface
F-5
Software Interface
F-6
F-7
F-7
F-8
F-9
F-9
LIS Communications
LIS Communications
F-3
F-4
LIS Communications
LIS 2
LIS 2 Plus
LIS 3
Application
appropriate for
applications
that do not
require LIS
device
interaction with
the Express
Plus
appropriate for
applications
that do not
require LIS
device
interaction with
the Express
Plus
appropriate for
applications
that requires
LIS device
interaction with
the Express
Plus
use in an
environment
with low
transmission
errors
use in an
environment
with low
transmission
errors
Application Layer
unidirectional
unidirectional
bidirectional
Data Format
outputs data
with parameter
delimiters and
field separators
outputs data
with parameter
delimiters and
field separators
outputs data
with parameter
delimiters and
field separators
unidirectional
bidirectional
bidirectional
Error Detection
and Error
Recovery
no error
detection or
recovery
error detection
and no error
recovery
error detection
and error
recovery
Message
Acknowledgment
no
yes
yes
error detection
and recovery
allows you to
use this
protocol in an
environment
with a high
noise level
LIS Communications
F-5
Physical Interface
The LIS device connects to the Express Plus through one of the
three serial RS-232 ports on the Express Plus.
NOTE
RS-232 Signal
Protective Ground
Transmit Data
Receive Data
Signal Ground
The remaining pins are not enabled for the cable sold by
Bayer Diagnostics.
F-6
Software Interface
LIS Communications
Description
Selections
Port No.
1, 2, or 3
Device
LIS 2
LIS 2 Plus
LIS 3
Baud Rate
9600
4800
2400
1200
600
300
110
Stop Bits
1
1
2
Parity
ODD
EVEN
MARK
NONE
SPACE
Data Bits
7 or 8
Xon/Xoff
No
Enable Polling?
Yes or No
Transmit Results
Incrementally?
Yes or No
LIS Communications
F-7
F-8
Downloading a Worklist
to the Express Plus
LIS Communications
LIS Communications
Transferring Results
Incrementally
F-9
Transferring Worklists
Incrementally
the remote Express Plus describes the set of samples that are
loaded, and the set of tests that it can run (determined from
the reagents that are loaded on the system)
LIS Communications
Relocation Instructions
G-1
Page
G-3
G-4
G-4
G-5
G-6
G-6
G-9
G-10
G-11
G-14
G-2
Relocation Instructions
Relocation Instructions
G-3
G-4
Relocation Instructions
WARNING
CAUTION
Depth:
Height:
Relocation Instructions
Understanding Power
Requirements
G-5
CAUTION
Fuse Rating
7.00 A, T (SB)
6.25 A, T (SB)
4.00 A, T (SB)
3.15 A, T (SB)
NOTE
G-6
Relocation Instructions
Preparing to Relocate
the Express Plus
Relocation Instructions
Preparing to Relocate
the Express Plus
(Continued)
G-7
all cables
keyboard
8. Open the front panel, then remove the distilled water bottle,
the waste bottle, and the cuvette waste drawer (refer to
Figure G-2).
Figure G-2. Removing the Bottles and Cuvette Waste Drawer
G-8
Preparing to Relocate
the Express Plus
(Continued)
Relocation Instructions
9. Insert the caps into both vent holes for the water bath.
(Vent holes are located to the right of the digital diluter,
and one vent hole is open (uncapped) during normal
operation.) Refer to Figure G-3.
Figure G-3. Installing Caps in the Water Bath Vent Holes
Relocation Instructions
G-9
CAUTION
G-10
Relocation Instructions
Relocation Instructions
G-11
G-12
Relocation Instructions
7. Open the front panel, then remove the distilled water bottle,
waste bottle, and cuvette waste drawer (refer to Figure G-6).
8. Fill the distilled water bottle with fresh, distilled water.
9. Safely discard any waste water from the waste bottle.
Figure G-6. Replenishing the Supplies
Relocation Instructions
G-13
G-14
Relocation Instructions
The water bath vent holes are capped correctly (one vent
open, one vent capped).
Index-1
Index
A
Absorbance (see also Response),
A-1
bichromatic, 3-9, A-1
blank, 7-13
calculating, 2-20, 3-63-8,
3-133-16
checks, 3-8, 3-10, 3-11
corrected, 3-8, 3-9
differences, 3-9
flags (errors), 3-10, 3-15, 3-29,
3-30, 6-396-41, 7-19, 7-20,
7-417-43, 8-15, 10-45, 10-47,
10-49, 10-50, 10-51,
10-6110-63
high (see also Absorbance
Limits), 6-35
blank, 3-8
primary, 3-10
secondary, 3-10
limits, 3-10, 3-11, 3-15, 4-42, 4-44,
5-14, 10-45, 10-49, 10-50,
10-51, 10-61, 10-62
low (see also Absorbance
Limits), 6-35
blank, 3-8
primary, 3-10
secondary, 3-10
monochromatic, 3-9, 3-10
range, E-8
reported, 7-41
readings, 3-8, 3-9
sample blank, 3-9
zero, 3-6, 3-7
Accessories (see also Software
Options), A-1, D-3D-6
bar code, 2-4, 2-30, 6-14, 6-16, C-8,
D-5
documents for, 1-9, 5-6, C-8
external printer, 2-30, 6-47, C-8,
D-5
ordering, C-8
Test Scheduler (bar code chart),
2-4, 6-14, 6-16, C-8, D-5
Actual,
C.V., 4-28, 7-33, 7-35, 7-37, 7-38,
9-9
Mean, 4-28, 4-30, 7-33, 7-35, 7-37,
7-38
S.D., 4-28, 4-30, 7-33, 7-35, 7-37,
7-38
Age, Patient, 4-9, 6-20
Aligning,
lamp, 9-28
probe tip, 9-31
Alphanumeric Keys, 2-25, A-1
Alternating Current, 1-7
Alphanumeric Keys, 2-25, A-1
B
Backing Up, 9-4
calibration files, 4-494-51,
9-459-48
data files, 4-494-51, 5-375-40,
9-459-48
definitions, 9-45
patient results, 9-45
program files, 4-51, 9-459-48
Q.C. databases, 4-51, 5-38, 5-39,
9-459-48
system calibrations, 4-51,
9-459-48
worklist data, 9-45
BACKSPACE (DELETE) Key,
2-27, 6-7, 6-46
Bar Code, 2-3, A-1
accessories, 2-4
blank, 4-39, 4-40, 5-34
calibrator, 4-38, 4-40, 5-19, 5-31
control, 4-37, 4-40, 5-20, 5-29
features, 2-4
fluid (1 or 2), 4-42, 5-17, 5-18
labels, 6-23, 6-26, C-7, C-8, D-5
profile, 4-36, 4-45, 5-25
reader failure, 10-19
reagent, 4-40, 5-17, 6-23, 6-26, 6-27
diluent, 4-45
scanner, 2-4, 2-10, 2-22, 2-23,
2-37, 6-30, A-1, E-5
test, 4-47, 5-14, 5-15, 5-18, 5-19
wand (hand-held), 1-7, 2-4, 2-22,
2-24, 2-30, 6-14, 6-16, A-1,
C-8, D-5, E-3, G-7, G-9
cables, G-7, G-9
Batch Testing, 6-16
Bayer Diagnostics, Contacting, C-3
C-4
Bichromatic Absorbance, 3-9, 3-13,
A-1
Biconvex Lens, 2-20, 2-21, 2-40
Bin, Cuvette (see Cuvette
Waste Drawer )
Index-2
Index (Continued)
Biohazard,
liner, 2-18, 2-19, 6-53
warning, 1-5, 1-7, A-1, B-2, B-3,
B-5
Blanked Linear Curve Type, 3-17,
3-20, 3-28
reblanking, 3-20, 3-30, 4-9, 4-15,
4-17, 4-18, 4-20, 6-15, 6-17,
6-18, 7-7, 7-8, 7-107-13
Blanks (see also Zero Calibrator), 3-8,
3-17, 3-28, 4-9, 6-29
absorbance, 7-13
bar codes, 4-39, 4-40, 5-34
defining, 4-34, 4-39, 4-40, 5-11,
5-335-35
erasing, 4-34, 5-33, 5-35
limits (see Absorbance, Limit)
modifying, 4-34, 5-11, 5-335-35
reruns, 3-31, 3-32, 6-34
sample, 3-9, 6-18
scheduling in worklist, 2-30, 6-18
storing, 5-34, 5-35
Bottle,
distilled water, 2-12, 2-13, 2-37,
2-38, 6-5, 6-12, 6-53, 10-29,
A-3, C-5, E-4
reagent, 2-10, 2-38, 4-45, 6-23,
6-27, 6-51, E-5
special, 4-40, 4-45, 4-47
types, 2-37, 4-40, 4-47, 5-17, 6-27
waste, 2-13, 2-37, 2-38, 6-5, 6-13,
6-38, 6-53, 10-43, A-9, C-5,
E-6
C
Cables, 10-68, B-1, G-7, G-9
bar code wand, G-7, G-9
communication, 2-31,
10-1910-21, 10-27, 10-41,
10-5410-57, 10-59, 10-68, F-5
external printer, 10-30, C-5
keyboard, 10-31, 10-67, G-9
power (line), 9-5, 9-25, 9-30,
10-66, C-5, G-5, G-7, G-9
Calculations,
absorbance, 2-20, 3-63-8,
3-133-16
calibration curve, 3-173-27, 7-10,
7-12, 7-13, 7-15, 7-18
factors, 3-28, 7-17
nonlinearity, 3-13, 3-14
noise, 3-133-15
normalizing slope, 3-26
polychromatic, 3-8
Index-3
Index (Continued)
normalizing, 3-26, 4-15, 4-17, 4-20,
6-15, 7-11, 7-12, 7-13
parameters (see also Calibration
Data Program), 4-16, 4-17,
4-18, 7-7, 7-11
qualitative, 3-27, 3-28
quantitative, 3-223-25
reblanking, 3-20, 3-30, 4-9, 4-15,
4-17, 4-18, 4-20, 6-15, 7-1, 7-2,
7-5, 7-7, 7-8, 7-10, 7-11, 7-12,
7-13
recalculating, 4-15, 7-14, 7-15
rejection, 4-18
reviewing, 7-10, 7-11
rerunning, 4-15, 7-14, 7-15
S.D., 3-19, 3-23, 4-16, 7-17, 10-61
status (see Calibration Status)
stored, 7-13
type, 3-28, 4-16
linear, 3-173-21, 3-28, 6-17,
6-18
nonlinear, 3-223-25, 6-17
qualitative, 3-27, 3-28
quantitative, 3-223-25, 3-28
waiving an expired, 7-16, 10-47
Calibration Data Program, 2-35,
4-144-20, 7-1, A-1
data, 4-17, 7-14
Data screen, 4-15
concentrations, 4-16
data, 4-17
error, 4-17
functions, 4-15, 7-15
replicate S.D., 4-19
responses, 4-17
map, 4-14
features, 7-5
fields, 4-164-19
functions, 4-15
Parameters screen, 4-154-19
curve parameters, 4-164-18
curve S.D., 4-16
functions, 4-15
status, 4-164-18, 4-20
status (see also Calibration Status),
reviewing, 6-15, 7-107-12
Calibration Status, 4-16, 4-20, 6-15,
6-27, 7-107-12, 7-17, 10-44
current, 4-20, 6-15, 7-15, 10-45
expired, 4-15, 4-20, 6-28, 7-15,
7-16, 10-45
incomplete, 4-15, 4-20, 7-12, 7-14,
7-15, 10-45
in process, 4-20, 10-46
invalid, 3-19, 4-15, 4-20, 6-40,
7-13, 7-14, 7-15, 10-46
locked, 10-46
none, 4-20, 7-12, 10-46
Index-4
Index (Continued)
documents, 1-9, 2-31
errors, 8-218-23, 10-5310-59
fields, 4-524-56
options, 4-49, 4-50
ports, 2-23, 4-54, 10-53, 10-56,
10-58, 10-68, A-2, F-5
protocols, F-4
software option, 2-31, 4-49, 4-56,
5-6, C-8, D-6, F-1
Status screen (see also Status
Program), 8-5, 8-20
fields, 8-218-23
messages and errors, 8-21, 8-23,
10-5310-59
subsystem, 2-22, 2-23, A-2
components, 2-23
Components,
communication subsystem,
central processing unit (CPU),
2-23, 2-40, A-2, E-8
hard disk drive, 2-23, A-4
ports, 2-23, A-2
reagent bar code scanner, 2-4,
2-10, 2-22, 2-23, 2-37, 6-30,
A-1, E-5
cuvette subsystem,
biohazard liner, 2-18, 6-13, 6-53
edge detector, 2-17, 2-39
feeder, 2-17, 2-39, 6-13, E-3
magazine, 2-17, 6-13, A-2, E-3
preheater, 2-17, E-3
strips, 2-17, 2-39, 2-40, A-2
track, 2-17, 2-39, 2-40, A-2, E-4
waste drawer, 2-18, 2-40, 6-13,
6-53, A-2, E-4
fluid subsystem,
digital diluter, 2-13, 2-14, 2-37,
2-38, 2-39, A-3, E-4
distilled water bottle, 2-13, 2-37,
2-38, 6-5, 6-12, 6-53, 10-29,
A-3, C-5, E-4
in-line filter, 2-12, 2-37, 2-38,
9-18, 9-19
peristaltic pumps, 2-12, 2-37,
2-38, 9-15
probe, 2-15, 2-16, 2-37, 2-38,
2-39, A-6
solenoid valve, 2-13, 2-14, 2-37,
2-38
syringe, 2-13, 2-14
wash station, 2-12, 2-37, 2-38,
E-6
waste bottle, 2-13, 2-37, 2-38,
6-15, 6-13, 6-38, 6-53, 10-43,
A-9, C-5, E-6
water bath, 2-13, 2-38, A-9, E-6
operation, 2-8, 2-372-40
photometer subsystem,
photometer board, 2-20, 2-21,
2-40
Index-5
Index (Continued)
Cord, Power (see Cables)
Correcting Absorbances, 3-8, 3-9
CPU (Central Processing Unit),
2-23, 2-40, A-2, E-6
CTRL Key, 2-27
Cumulative Statistics (see
Q.C. Statistics)
Cup, Sample, 2-38, 6-16, A-8, E-4,
E-5
Current,
alternating, 1-7
dark, 3-7, 3-8
sample, 3-7, 3-8
zero, 3-6, 3-7, 3-8
Current (Curve Status), 4-20, 6-15,
7-12, 7-15, 10-45
Cursor, 2-32, 2-33, A-2
Cursor Keys, 2-26, 2-33, 6-7, A-2
Curve (see Calibration Curve),
data (see Calibration Curve;
Calibration Data Program)
status (see Calibration Status)
Cuvette,
biohazard liner, 2-18, 6-13, 6-53
edge detector, 2-17, 2-39
feeder, 2-17, 2-39, 6-13, E-3
handling system, 2-18
magazine, 2-17, 6-13, A-2, E-3
preheater, 2-17, E-3
strips, 2-17, 2-39, 6-13, A-2, E-3
subsystem components, 2-17, 2-39
track, 2-17, 2-39, 2-40, A-2, E-4
waste drawer, 2-18, 2-40, 6-13,
6-53, A-2, E-4
C.V. (Coefficient of Variation), 7-33
7-35, 7-37, 7-38, 9-9, 9-10
D
Dark Current, 3-7, 3-8
Data (see also Patient Results
Program; Results),
calibration (see also Calibration
Data Program), 3-173-27,
4-17, 7-14, 7-15, 7-17
DAU Applications Guide,
1-8
files, 4-494-51, 5-375-40
backing up, 4-51, 5-375-40,
9-459-48
restoring, 4-51, 9-499-51
management, 2-6, 2-7
quality control (Q.C.) (see also
Q.C. Statistics Program),
2-7
selection, 3-11
outlier rejection, 3-11, 3-12
substrate depletion, 3-9, 3-11
storage specification, E-4
Index-6
Index (Continued)
defining, 4-34, 5-115-13,
7-7
erasing, 4-34, 5-115-13,
5-23
features, 5-12
functions, 4-34
modifying, 4-34, 5-115-13
parameters, 4-34, 4-35,
5-11, 5-145-23, 7-7, 7-8, 7-22
QCSystems, 4-35, 5-11, 5-16,
5-22, 5-23, 7-24
reagents, 5-11, 5-16, 5-17
selecting, 5-12, 5-23
storing, 5-23
type, 5-16
DELETE Key (BACKSPACE), 2-27,
6-7, 6-46
Deleting (see Erasing)
Demographics (see Patient,
Demographics)
Depletion, Substrate, 3-9, 3-11
Diagnostics, 2-34
Diagnostics/Maintenance Program,
4-4, 4-574-63, A-3
features, 4-57
functions, 4-58, 4-59
start-of-day, 4-58, 6-116-13
map, 4-57
screens,
Lamp Alignment, 4-60
Precision Check, 4-59, 9-89-10
Probe Tip Calibration, 4-62
System Version, 4-61
Digital Diluter, 2-13, 2-37, 2-38, 2-39,
A-3, E-4
adjusting, 9-22, 9-40
cleaning, 9-4, 9-209-24, 10-28
priming, 9-19, 9-41
replacing,
syringe, 9-369-41
tubing, 9-379-41
removing air from tubing, 9-41
Diluent,
reagent, 4-40, 4-42, 4-45, 4-46, 5-17
sample, 4-47, 5-17
bar code, 5-17
bottle type, 4-40, 4-45, 4-47, 5-17
volume, 4-42, 4-47, 5-17, E-4
Diluter (see Digital Diluter)
Dilution, Sample,
automatic, 2-5, 3-31, 3-32, 3-36,
6-34, 6-35, 6-40
manual, 2-6, 6-36
parameters, 2-5
ratios, 2-5, 5-17, 6-34
Diode Array, 2-20, 2-40
Disk Drive, 2-22, 2-23, 2-25, A-4
Disk, Floppy, 2-25, 2-26, A-3
format, 4-51, 5-38
E
Edge Detector, Cuvette, 2-17, 2-39
Editing,
blanks, 4-34, 5-11, 5-33, 5-35
calibration curve, 4-15, 7-14, 7-15
calibrators, 4-34, 4-35, 5-11,
5-305-32
controls, 4-34, 5-27, 5-29
profiles, 4-34, 5-11, 5-245-26
requests, 6-46
sample results, 6-46
tests, 4-34, 5-115-13, 5-23
Electrical,
ratings, E-7
requirements, E-7
End-of-day,
cleaning, 6-6, 6-52, 6-53
procedures, 6-516-53
Endpoint Test, 3-11, 3-12, 3-15, 5-16
response, 3-21
ENTER Key, 2-33, 6-7
Environmental Conditions, E-7
Erasing,
calibration curve, 4-15, 7-14, 7-15
definitions,
blanks, 4-34, 5-33, 5-35
calibrators, 4-34, 5-30, 5-32
controls, 4-34, 5-27, 5-29
profiles, 4-34, 5-245-26
tests, 4-34, 5-115-13, 5-23
normalized curve, 3-24
stored worklist, 6-14, 6-15
Error Conditions, 10-6410-67
Errors (see also Messages; Flags),
communications, 8-21, 8-23,
10-5310-59
correcting during a run, 6-37,
6-38, 10-65
fatal disk, 10-70
flags (see Flags)
reagent status, 10-4410-47
Index-7
Index (Continued)
run-time, 10-510-10, 10-6010-63
sample status, 10-4810-51
status, 10-510-10
system details status, 10-1910-43
EXIT Key, 2-33, 6-7
Expected,
C.V., 7-33, 7-35, 7-37, 7-38, 9-9
Mean, 4-29, 4-30, 5-20, 5-21, 7-22,
7-23, 7-33, 7-35, 7-37, 7-38
S.D., 4-29, 4-30, 7-33, 7-35, 7-37,
7-38
Expired (Curve Status) (see also
Calibration Status), 6-27,
7-16
External Printer, 2-30, A-3, C-5,
D-5, E-4
cables, 10-30, C-5
documents for, 1-9, 2-30
F
Factors, Calculation, 3-28, 7-17
Failure,
bar code reader, 10-19
calibration curve, 3-30, 10-45,
10-47
normalization/reblanking, 3-30
power, 10-68, 10-69
waste pump, 6-12
Fan Filters,
cleaning, 9-4, 9-13, 9-14
Fatal Disk Error, 10-70
Files, Calibration (see Calibration,
Files)
Filters,
fan, 9-4, 9-13, 9-14
in-line, 2-12, 2-37, 2-38, 9-18, 9-19
Fixed Function Keys, 2-33, 6-7
Flags (errors),
absorbance, 3-10, 3-15, 3-29, 3-30,
6-396-41, 7-19, 7-20,
7-417-43, 8-15, 10-45, 10-47,
10-49, 10-50, 10-6110-63
automatic rerun and dilution,
6-34, 6-35
calibration, 3-29, 7-177-20
calibration curve, 3-27, 3-29, 3-30,
3-32, 7-177-19, 10-46,
10-6110-63
data selection, 3-11, 3-12
run-time report, 6-40, 6-41,
7-417-43, 10-6010-63
Floppy Disk Drive, 2-25, 2-26, A-3
Fluid,
bar code, 4-42, 5-18
subsystem components, 2-122-15,
2-372-39
wash, 5-17
Fluid-level Detector Wire, 2-15,
2-37, 2-38
H
Hard-disk Drive, 2-23, A-4
Heading, Report, 5-8
HELP Key, 6-7, 8-3
Help Program, 2-36, 4-3, 8-3, 8-4,
A-4
features, 8-3
operating, 8-3
screen example, 8-4
selecting, 8-3
High Absorbance (see Absorbance,
High; Absorbance, Limits)
I
Incomplete (Curve Status), 4-15,
4-20, 7-12, 7-14, 7-15, 10-46
In-line Filter, 2-12, 2-37, 2-38, 9-18,
9-19
In Process (Curve Status), 4-20,
7-12, 10-46
Interface Protocols,
documents for, 1-9
Interval,
calibration, 4-35, 4-40, 4-44, 7-7
normalization, 7-7
reblanking, 7-7
Invalid (Curve Status), 3-19, 4-15,
4-20, 6-40, 7-13, 7-14, 7-15,
10-46
K
Keyboard, 1-7, 2-252-29, A-4, E-4
cables, 10-31, 10-67, G-9
Keys,
alphanumeric, 2-25, A-1
ALT, 2-27
BACKSPACE (DELETE), 2-27,
6-7, 6-46
CAPS LOCK, 2-27
COMPOSE, 2-27
CTRL, 2-27
cursor, 2-26, 2-33, 6-7, A-2
DELETE (BACKSPACE), 2-27,
6-7, 6-46
fixed-function, 2-33, 6-7
SELECT, 1-6, 2-33, 6-7
ENTER, 2-33, 6-7
EXIT, 2-33, 6-7
Index-8
Index (Continued)
CANCEL, 2-27, 2-33, 5-7, 5-16,
6-7
MAIN MENU, 2-33, 4-3, 6-7
TAB, 2-27, 2-33
function (F1F8), 2-26, 2-27, A-4
HELP, 6-7
paper advance, 2-25, 2-28, 6-8, 6-9,
A-5
PAUSE (see System Keys)
PRINT, 6-7, 6-22, 7-36
PRINT SCRN, 6-7
SHIFT, 2-27
START (see System Keys)
STATUS, 6-7, 6-37
STOP (see System Keys)
system, 2-25, 2-28, 2-29, 6-8, 6-9,
A-9
PAUSE, 2-29, 6-8, 6-9
START, 2-29, 6-8, 6-9, 6-30,
6-38
STOP, 2-29, 6-8, 6-9
Kinetic Tests, 3-11, 3-13, 3-14, 5-16
L
Labels, Bar Code, 6-23, 6-26, C-7,
C-8, D-5,
Laboratory Information System (LIS),
F1F9
communications protocols, F-4
connecting, F-5, F-6
physical interface, F-5
software interface, F-6
data,
receiving, F-7
transferring,
incremental results, F-9
worklists, F-9
transmitting, F-7, F-8
documents for, F-3, F-4
Lag Time, 5-17, A-4
Lamp,
adjusting, 9-28, 9-29
aligning, 4-60, 9-28, 9-29
assembly, 9-27
quartz-halogen, 2-20, 2-40
replacing, 9-269-29
Lens,
biconvex, 2-20, 2-21, 2-40
plano convex, 2-20, 2-21, 2-40
Levey-Jennings (see Q.C. Statistics
Program)
Lifetime,
C.V., 4-29, 7-33, 7-37, 7-38
Mean, 4-29, 5-21, 7-23, 7-33, 7-37,
7-38
S.D., 4-29, 4-30, 7-33, 7-37, 7-38
Light Source (see Lamp)
Limits, Absorbance (see
Absorbance, Limits)
M
Main Menu, A-4
key, 2-33, 4-3, 6-7
map, 4-5, 4-6
Maintenance Tasks, Corrective,
9-259-51
backing up files, 9-459-48
calibrating the probe tip,
9-319-35
filling the water bath, 9-42, 9-43
hazards, 9-25
replacing,
digital diluter syringe, 9-369-41
fuses, 9-30, 9-31
lamp, 9-269-29
printer paper, 9-44, 9-45
probe tip, 9-319-35
resetting the system, 9-51
restoring files, 9-499-51
Maintenance Tasks, Preventive,
9-49-23
cleaning,
digital diluter syringe, 9-4,
9-209-24, 10-28
fan filters, 9-4, 9-13, 9-14
probe tip, 9-4, 9-7, 9-9, 10-43
screen, 9-4, 9-6
system cabinet, 9-4, 9-6
Index-9
Index (Continued)
system tubing, 9-4, 9-9, 9-11,
9-12
hazards, 9-5
precision check, 4-58, 4-59,
9-89-10
replacing,
in-line filter, 9-18
wash pump tubing, 9-159-17
waste pump tubing, 9-159-17
Manual,
dilutions, 2-6, 6-36
reruns, 2-6, 6-36
Map (see Software Map)
Mean,
actual, 4-28, 4-30, 7-33, 7-35, 7-37,
7-38
establishing, 5-26
expected, 4-29, 4-30, 5-20, 5-21,
7-22, 7-23, 7-33, 7-35, 7-37,
7-38
lifetime, 4-29, 5-21, 7-23, 7-33,
7-37, 7-38
Menu, 2-32
Messages,
calibration curve, 3-23, 3-25, 3-26,
4-18, 10-61, 10-63
communications, 8-218-23
error, 10-510-10
run-time report flags, 6-40, 6-41,
7-417-43, 10-6010-63
status, 10-510-10
system detail, 10-1310-16
Methods, Calibration,
3-173-27
Microprocessor, E-8
Modifying,
blanks, 4-34, 5-11, 5-335-35
calibration curve, 4-15, 7-14, 7-15
calibrators, 4-34, 5-11, 5-305-32
controls, 4-34, 5-11, 5-275-29
profiles, 4-34, 5-11, 5-245-26
reagent loading, 6-27
requests, 6-46
sample,
loading, 6-27
results, 6-46
tests, 4-34, 5-115-23
Monochromatic Absorbance, 3-9,
3-10
Multi-test Calibrators, 6-17, 6-18
N
Noise,
calculations, 3-133-15
flags, 6-35, 6-40, 7-19
None (Curve Status), 4-20, 7-12,
10-46
Nonlinear,
calculations, 3-13, 3-14
calibration curve, 3-223-25, 6-17
parameters, 3-22, 3-28
curve types,
1-Logit 4, 3-24, 3-28
2-Logit 5, 3-24, 3-28
3-Exponential 5, 3-25, 3-28
4-Polynomial 5, 3-25, 3-28
flags, 6-35, 6-40
Normalized (Curve Status), 3-26,
4-15, 4-17, 4-20, 6-15, 7-11,
7-12, 7-13, 10-46
Normalizers, 7-8, 7-13
Normalizing, A-5
calibration curve, 3-26, 4-15, 4-17,
4-20, 6-15, 7-11, 7-12, 7-13,
10-46
failures, 3-30
intervals, 7-7
slope, 3-26
two-point, 3-26
O
Off-system Test, 5-14
Operating Temperature, 5-6
Operational Sequence, 2-372-40
Optics, Imaging, 2-20, 2-21, 2-40
Options Program, 4-4,
4-494-56, A-5
communications, 4-49, 4-50
data fields, 4-524-56
disk options, 4-51
features, 4-49
functions, 4-50, 4-51
map, 4-49
printing options, A-6
data fields, 5-8
defining, 5-7, 5-8
modifying, 5-7, 5-8
screen, 5-7
selecting, 5-8
reset system, 4-51, 9-51
system options,
data fields, 5-6
defining, 5-55-7
modifying, 5-55-7, 7-25
screen, 5-5
selecting, 5-6
test order, 4-50
Ordering,
accessories, C-8
replacement parts, C-5
service kit parts, C-6
supplies, C-7
Outlier Rejection, 3-11, 3-12, A-5
flags, 6-35, 6-40
Outputs, E-8
Over Flags, 6-35, 6-40
Index-10
Index (Continued)
P
Paper Advance Key, 2-25, 2-28, 6-8,
6-9, A-5
Parameters,
calibration (see Calibration Curve)
calibration curve (see also
Calibration Data Program),
4-164-18, 7-7
nonlinear curve, 3-22
reagent, 5-16
test, 5-145-23
Parts,
replacement, C-5
service kit, C-6
Patient,
age, 4-9, 6-20
comments, 6-20
demographics, 6-20, A-3
ID, 6-20
name, 6-20
results,
backing up, 9-45
sex, 6-20
Patient Results Program, 4-4,
4-114-13, A-5
data fields, 4-12
editing, 6-45, 6-46
features, 6-43
finding requests, 6-48
functions, 4-12, 6-436-48
map, 4-11
printing results, 6-476-49
reviewing results, 6-43, 6-44
screen, 4-12, 4-13
PAUSE Key, 2-29, 6-8, 6-9
Pending Status (see also Status
Program), 6-33
Peristaltic Pumps (see Pumps)
Photodiode Array, 2-20, 2-40, A-5
Photometer Subsystem Components,
2-20, 2-21, A-5
Placekeeping Zone, 2-34, A-5
Plano Convex Lens, 2-20, 2-21, 2-40
Plots (see Q.C. Statistics Program)
Polychromator, 2-20, 2-21, 2-40, A-5,
E-4, E-8
calculations, 3-8
Port,
communications, 2-23, 4-54, 10-53,
10-56, 10-58, 10-68, A-2, F-5
parallel, 2-23, E-8
probe inlet, 2-15
probe inlet/outlet, 2-15
RS-232, 2-23, E-8
Positive Sample Identification
(PSID) Option, 2-30, 6-14,
6-16, 6-24, A-6, C-8, D-5
documents for, 1-9, 2-30, C-8
features, 2-30
loading sample tube tray, 6-24
scheduling samples, 6-16
Power,
cable, 9-5, 9-25, 9-30, 10-66, C-5,
G-5, G-7, G-9
failure, 10-68, 10-69
switch, 6-11, 6-53, 9-11, A-6
Precision Check, 4-58, 4-59,
9-89-10
Predilution Ratio, 5-17
Preparing Worklist, 6-146-22
Preventive Maintenance,
9-49-23
Primary,
sample tube, 2-30, A-6
sample tube tray, 2-9, A-6, E-5
wavelength, 3-8, 3-9, 7-19
Printer,
default, 5-8
external, 1-9, 2-30, 5-8, 10-30, A-3,
C-5, D-5, E-4, E-8
paper, 9-449-45
system, 2-25, 2-28, 5-8, A-9, E-6
Printing Options, 4-50, A-6
data fields, 5-8
defining, 5-7, 5-8
modifying, 5-7, 5-8
screen, 5-7
selecting, 5-8
sample results, 6-476-49
PRINT Key, 6-7, 6-22, 7-36
PRINT SCRN Key, 6-7
Probe, 2-15, 2-16, 2-37, 2-38, 2-39,
A-6
assembly, 2-16, E-5
biohazard, 2-15
calibration constants report, 9-35
ports, 2-15
safety guard, 2-15
thermistor, 2-15
tip (aspiration), 2-15, 2-16, 2-37,
2-38, 2-39
aligning, 9-31
calibrating, 9-319-35, 10-42,
10-43
cleaning, 9-4, 9-7, 9-9, 10-43
replacing, 9-319-35
Profile(s), A-6
bar code, 4-36, 4-45, 5-25
defining, 4-34, 5-245-26
erasing, 4-34, 5-245-26
modifying, 4-34, 5-245-26
selecting, 5-24, 5-26
Selection screen, 4-36, 5-25
storing, 5-26
Program, 2-32, 2-35, A-6
Calibration Data (see also
Calibration Data Program),
4-4, A-1
Index-11
Index (Continued)
Definitions (see also Definitions
Program), 4-4, A-3
Diagnostics/Maintenance
(see also Diagnostics/
Maintenance Program), 4-4,
A-3
files,
backing up, 4-51, 9-459-48
restoring, 9-499-51
Help (see Help Program)
Load Alternate (see also Load
Alternate Program), 4-4, A-4
Options (see also Options
Program), 4-4, A-5
Patient Results (see also Patient
Results Program), 4-4, A-5
Q.C. Statistics (see also
Q.C. Statistics Program),
4-4, A-7
software map, 2-35, 2-36
Status (see also Status Program),
A-9
Worklist (see also Worklist
Program), A-10
zone, 2-34, A-6
Prompts, 2-32, A-6
Prompt Zone, 2-34, A-6
Protocols, Communications, F-4
PSID (Positive Sample Identification)
Option, 2-30, 6-14, 6-24, A-6,
C-8, D-5
documents for, 1-9, 2-30
features, 2-30
loading sample tube tray, 6-24
scheduling samples, 6-16
Pumps, Peristaltic, 2-12, 2-37, 2-38,
9-15
Q
Q.C. Databases,
backing up, 4-51, 5-38, 5-39,
9-459-48
deleting data, 7-44
restoring, 5-41, 9-49, 9-51
Q.C. Statistics Program, 4-4
4-214-31, 7-217-44, A-7
calculating, 3-40, 4-284-30, 7-33
control rules, 3-333-39, A-2
cumulative status, 4-27
data,
deleting, 7-44
modifying, 7-27, 7-307-32
printing, 7-27
reports, 7-3, 7-357-38
selecting, 7-27, 7-30, 7-32
viewing, 7-3, 7-26, 7-27
Data screen, 2-7, 4-24
data fields, 4-284-31
Details screen, 4-274-30
functions, 4-24
print functions, 7-35
features, 2-7, 4-21, 7-6, 7-26
flags, 7-417-43
functions, 4-23
Levey-Jennings plots,
data fields, 4-284-31
Details screen, 7-287-32
functions, 4-25
modifying, 7-287-32
printing, 7-29
sample interpretation, 3-373-39
screen, 2-7, 4-25, 7-28
selecting, 7-29, 7-30
viewing, 7-3, 7-27, 7-28, 7-29,
7-31
map, 4-22
QCSystems screen, 4-26, A-7
data fields, 4-28
screen, 4-23
special considerations, 3-40
Summary Statistics screen, 2-7,
4-26
data fields, 4-284-31
functions, 4-26
printing, 7-34
selecting, 7-34
QCSystems Codes, 5-11, 5-16,
5-22, 5-23, 7-3, A-7
defining, 5-16, 5-22, 5-23
method code, 5-22
modifying, 5-16, 5-22, 5-23
printing, 7-39, 7-40
reagent code, 5-22
selecting, 5-16, 5-22, 5-23
test code, 5-22
units code, 5-22
Qualitative Curve Type, 3-27, 3-28
Quality Control (see Q.C. Statistics)
Quantitative Curve Type,
3-223-25, 3-28
Quartz-Halogen Lamp, 2-20, 2-40
R
Random Access Testing, 6-16
Range, Absorbance, E-8
Rates, Calculation, 3-22
Ratio Test, 5-14, A-7
Readings, Absorbance, 3-8, 3-9
Reaction,
temperature, E-5
time, E-5
Reading Status, 2-34
Ready Status, 2-34
Reagent,
Applications Guides, 1-8,
bar code, 4-40, 5-17, 6-23, 6-26,
6-27
Index-12
Index (Continued)
bar code scanner, 2-4, 2-10, 2-22,
2-23, 2-37, 6-30, A-1, E-5
bottles, 2-10, 2-38, 4-45, 6-23, 6-27,
6-51, E-5
capacity, E-5
carryover, 2-5, A-1
defining, 5-11, 5-16, 5-17
diluent, 4-40, 4-42, 4-45, 4-46, 5-17
loading, 6-23, 6-25, 6-26, 6-27, 6-30
screen, 7-16
modifying, 5-11, 5-16, 5-17
selecting, 5-11, 5-16, 5-17
status (see Status Program)
errors, 10-4410-47
subsystem components (see also
Components), 2-10
transport, 2-10, 2-37, A-7, E-5
tray, 2-10, 2-37, 2-38, 6-23, 6-25,
6-27, A-7, E-5
tray cover, 2-10, 6-256-27
volume, 5-17, E-5
Reblanked (Curve Status), 4-18,
4-20, 6-15, 7-12, 7-13, 10-47
Reblanking, A-7
blanked linear curves, 3-20
calculations, 3-20, 7-10
calibration curve, 3-20, 3-30, 4-9,
4-15, 4-17, 4-18, 4-20, 6-15,
7-1, 7-2, 7-7, 7-8, 7-10, 7-11,
7-12, 7-13
failures, 3-30
intervals, 7-7
parameters, 4-164-18
status, 4-17, 4-18, 4-20
stored curves, 6-17, 7-2
Recalculating Curves, 4-15, 7-14, 7-15
Receiving Data (LIS), F-7
Rejection, Calibration Curve, 4-18
Relocation, System, G-1G-14
Replacement Parts, C-5
Replacing,
digital diluter, 9-369-41
fuses, 9-30, 9-31
in-line filter, 9-18, 9-19
lamp, 9-269-29
printer paper, 9-44, 9-45
probe tip, 9-319-35
tubing, 9-159-17,
Replicates, Calibrator, 4-44, 7-7,
10-63
Report,
collated patient, 2-6, 6-476-49,
A-1
headings, 5-8
run-time, 2-6, 6-396-41, A-8
Request Number, 6-16
Rerunning,
blanks, 3-31, 3-32, 6-34
calibration curves, 4-15, 7-14, 7-15
calibrators, 6-34, 7-12, 7-19, 7-20
samples, patient, 6-346-36
Rerun(s),
automatic, 2-5, 3-19, 3-31, 3-32,
3-36, 4-46, 6-34, 6-35, 6-40
diluted, 2-5, 3-31, 3-32, 3-36, 4-46,
5-17, 6-34, 6-35, 6-40
dilution ratio, 4-46 5-17
flags, 3-32, 6-35, 6-40, 6-41, 7-20
manual, 2-6, 6-36
Reset System, 4-51, 9-51
Response,
calibrator, 7-17
values (delta absorbance),
3-133-16, 7-17, 7-41
calculating, 3-133-16
printing, 7-17, 7-41
Restoring,
data files, 4-51, 5-415-43,
9-499-51
Q.C. data, 5-41, 9-499-51
system calibrations, 4-51,
9-499-51
Results (see also Data),
blanked linear curve, 3-20
calculations, 2-40, 3-11, 4-48
enzyme linear curve, 3-21
limited linear curve, 3-21
patient (see also Patient Results
Program), 4-11
backing up, 9-45
precision check, 9-10
reports,
collated patient, 2-6, 6-476-49
run-time, 2-6, 6-396-41, A-8
Reverse Video, 2-33, A-7
Reviewing,
calibration,
curve, 7-10, 7-11
status, 6-15, 7-107-12
pending samples, 6-33
Q.C. data, 7-417-43
run-time,
error flags, 7-177-20, 7-417-43
reports, 6-39
results, 6-396-41
scheduled samples, 6-33
Rule(s), Control (see Q.C. Statistics
Program)
Running Status, 2-34
Run-time,
error log, 5-8, 7-17, 10-60
reports, 2-6, 6-396-41, A-8
calibrator, 5-8, 7-177-20
errors, 10-510-10, 10-6010-63
S
Safety Summary, B-1B-5
Sample(s),
absorbance, 3-9
blanking, 3-9
Index-13
Index (Continued)
cup, 2-38, 6-16, A-8, E-4, E-5
diluent, 4-47, 5-17
diluting manually, 6-36
loadlist, 6-22, A-8
reviewing scheduled, 6-33
scheduling, 6-166-19
single, 6-31, 6-32
status, (see Status Program)
transport, 2-9, 2-37, A-8, E-5
tray, 2-9, 2-10, 2-11, 6-16, 6-24,
6-256-27, A-6, A-8, E-6
capacity, E-6
cover, 2-9, 2-10, 2-11, 6-16, 6-24
format, 5-6, 5-7
loading, 6-24, 6-25
tube, primary, 2-30, A-6
types, A-8
blank, 2-5, 3-7, 3-8, 3-9, 3-10,
6-18, 6-19
calibrator, 6-18, 6-19
control, 6-18, 6-19
patient, 6-18, 6-19
volume, 5-17, 6-34, E-6
Sample Subsystem Components
(see also Components), 2-9
Scheduled (Curve Status), 4-20,
7-12, 10-47
Scheduling,
bar-coded patient samples, 2-30
blanks, 2-30, 6-18
calibrators, 2-30, 6-17, 6-18, 7-13,
7-15, 7-18, 10-46
multi-test, 6-17, 6-18
test-specific, 6-17, 6-18
controls, 2-30
manual reruns, 6-36
profiles, 2-30
samples, 6-166-19
tests, 2-30
Screen, Cleaning, (see System,
Cleaning)
Screens (see also Screen Zones), 2-25,
2-32, A-8
Screen Saver Option, 5-6
Screen Zones, A-8
function key, 2-34
placekeeping, 2-34
program, 2-34
prompt, 2-34
status, 2-34
S.D.,
calibration curve, 3-19, 3-23, 4-16,
7-17, 10-61
establishing control values, 5-26
Secondary Wavelength, 3-8, 3-9
Selecting,
calibrators, 5-305-32
controls, 5-20, 5-21, 5-27, 5-29,
7-217-23
test parameters, 5-145-23
SELECT Key, 1-6, 2-33, 6-7
Index-14
Index (Continued)
reagent status,
data fields, 8-12
features, 8-11
functions, 8-13
map, 8-11
screen, 8-11
sample status,
data fields, 8-15,
errors, 10-4810-51
features, 8-14
functions, 8-16
map, 8-15
screen, 8-14
screen, 8-6
system details,
data fields, 8-9
features, 8-8
functions, 8-10
map, 8-8
screen, 8-8
troubleshooting,
10-11, 10-12
Status Zone, 2-34
STOP Key, 2-29, 6-8, 6-9
Storing,
blanks, 5-34, 5-35
calibrators, 5-32
controls, 5-28
profiles, 5-26
tests, 5-23
Substrate Depletion, 3-9, 3-11
Subsystem,
communications, 2-22, 2-23, A-2
cuvette, 2-172-19, A-2
fluid, 2-122-16, A-3
operation, 2-8, 2-372-40
photometer, 2-20, 2-21, A-5
reagent, 2-92-11
sample, 2-92-11
user interface, 2-242-31
Symbols, 1-7
Syringe, Diluter, 2-13, 9-21
System,
application, E-1
calculations, 3-6
calibrations,
backing up, 4-51, 9-459-48
restoring, 4-51, 9-459-51
capabilities, 2-3
cleaning, 6-53, 9-6
cabinet, 9-4, 9-6
screen, 9-4, 9-6
tubing, 9-9, 9-11, 9-12
clock, 5-5, 5-6, 6-11, A-9
conditions, E-7
controls, E-1, E-6
dimensions, E-7, G-4
features, 2-42-7
bar code, 2-4
data management, 2-6, 2-7
T
TAB Key, 2-27, 2-33
TDM Applications Guide, 1-8
Temperature,
operating, 5-6
reaction, E-5
Index-15
Index (Continued)
Test,
assay order screen, 4-34, 4-35,
4-48, 5-11, 5-165-18
automatic shutdown, 3-36, 4-52,
5-6, A-1
bar code, 4-47, 5-14, 5-15, 5-18,
5-19
calibrating (see Calibration Curve)
name, 5-18
order options, 4-50, 5-9
screen, 5-9
selecting, 5-9
parameters (see Definitions
Program)
types,
DAU, 3-27
endpoint, 3-11, 3-12, 3-15, 5-16
kinetic, 3-11, 3-13, 3-14, 5-16
off-system, 5-14
ratio, 5-14, A-7
two-point, 3-15, 3-16, 3-26, 5-16
Testing,
batch, 6-16
random access, 6-16
Tests (see also Definitions
Program; Worklist Program),
defining, 4-34, 5-115-13, 7-7
erasing, 4-34, 5-115-13, 5-23
modifying, 4-34, 5-115-13
pending status, 6-33
reviewing status, 6-33
selecting, 4-34
Test Scheduler (bar code chart),
2-4, 6-14, 6-16, A-9, C-8, D-5
Test-specific Calibrators, 6-17, 6-18
Thermal Electric Device (TED), E-6
Thermistor, Probe, 2-15
Transferring (LIS),
incremental results, F-9
incremental worklists, F-9
Transmitting Data (LIS), F-7, F-8
Transport,
reagent, 2-10, 2-37, A-7, E-5
sample, 2-9, 2-37, A-8, E-5
Tray,
reagent, 2-10, 2-37, 2-38, 6-23,
6-256-27, A-7, E-5
sample, 2-9, 2-10, 6-16, 6-24,
6-256-27, A-8, E-6
Tray Cover,
reagent, 2-10, 6-256-27
sample, 2-9, 2-10, 2-11, 6-16, 6-24
Troubleshooting, 10-110-70
correcting,
communications status errors,
10-5310-59
error conditions, 10-6410-67
U
Under Flag, 6-35, 6-41
User Interface Subsystem
Components (see also
Components), 2-252-31
V
Value, Response (see Response
Value)
Valve, Solenoid, 2-13, 2-14, 2-37,
2-38
Verifying,
reagent loading, 6-26, 6-27, 6-30
system conditions, 6-25
Volume,
dispense, E-4
reagent, E-5
sample, E-6
W
Waived (Curve Status), 7-16, 10-47
Waiving Expired Curves, 7-16
10-47
Wand, Bar Code (see Bar Code)
Warming Up Status, 2-34
Index-16
Index (Continued)
Warning, 1-5, A-9, B-1, B-2
Wash,
count, 5-17, 5-18
fluid, 5-17, 5-18
pump, 2-12, 2-37, 2-38
pump tubing, replacing, 9-159-17
station, 2-12, 2-37, 2-38, E-6
Waste,
bottle, 2-13, 2-37, 2-38, 6-5, 6-13,
6-38, 6-53, 10-43, A-9, C-5,
E-6
drawer, cuvette, 2-18, 2-40, 6-13,
6-53, E-4
pump, 2-12, 2-38
pump tubing, replacing, 9-159-17
water, 2-38
Water,
bath, 2-13, 2-38, A-9, E-6
filling, 9-42, 9-43
bottle, distilled, 2-12, 2-13, 2-37,
2-38, 6-12, 6-53, 10-29, A-3,
C-5, E-4
Wavelength Flags, 6-35, 6-40, 6-41
Worklist, A-9
backing up, 9-45
entering patient demographics,
6-20
erasing,
requests, 6-21, 6-52
stored worklists, 6-14, 6-15
modifying, 6-20, 6-21
preparing, 6-146-22
printing, 6-20, 6-22
reviewing calibration status, 6-15
scheduling,
blank, 2-30, 4-9, 6-18
calibrators, 2-30, 4-9, 6-17, 6-18,
7-13, 7-15, 7-18, 10-46
controls, 2-30, 4-10, 6-18
profiles, 2-30, 4-10, 6-18
samples, 4-10, 6-166-19
tests, 2-30, 4-9, 4-10, 6-146-22
Z
Zero,
absorbance, 3-6, 3-7
calibrator (see also Blanks), 3-8,
3-28, 7-13, 7-20
currents, 3-6, 3-7, 3-8
Zones, Screen (see Screen Zones)