0 оценок0% нашли этот документ полезным (0 голосов)
57 просмотров7 страниц
Digital-multimeters are being used in multichannel automatic data-acquisition systems to determina strain values from strain-gage resistance evaluation. 8oth the direct-resistance method and the reverse current one are examined in two tests. On the first system the intrinsic zeroing of the instrumentation at the gage location can be removed by shortening, with a resistance very small compared to circuit resistance.
Digital-multimeters are being used in multichannel automatic data-acquisition systems to determina strain values from strain-gage resistance evaluation. 8oth the direct-resistance method and the reverse current one are examined in two tests. On the first system the intrinsic zeroing of the instrumentation at the gage location can be removed by shortening, with a resistance very small compared to circuit resistance.
Digital-multimeters are being used in multichannel automatic data-acquisition systems to determina strain values from strain-gage resistance evaluation. 8oth the direct-resistance method and the reverse current one are examined in two tests. On the first system the intrinsic zeroing of the instrumentation at the gage location can be removed by shortening, with a resistance very small compared to circuit resistance.
from the SAGE Social Science Collections. All Rights Reserved.
Zero-shift Evaluation of Automatic Strain-gage
Systems Based on Direct and Reverse Current Method by P. Cappa, Z. Del Prete, K.G. McConnell and L. Zachary ABSTRACT -Sol id-state digital-multimeters are being used in multichannel automatic data-acquisition systems to determina strain values from strain-gage resistance evaluation rather than using conventional Wheatstone bridge. 8oth the direct-resistance method and the reverse current one are examined in two tests. Each one lasted far two weeks. lrrelevant differences were observed between the two methods in spite of the adopted general-purpose low-cost switch-control unit. lntroduction The effects of advances in technology on digitai mul- timeter caused, as is well known, a sensible increase of metrological performances as accuracy and precision in Iong-term measurements. The previous observation deter- mined in 1983 1 the proposal of a nontraditional system to measure electrical-strain-gage outputs collected with a digitai ohmmeter; the method is simply based on the direct measurements, carried out by a four-wire connection, of the transducer resistance values. Moreover, the Iead-wire length is theoretically unimportant. The intrinsic Iimitation that is the zeroing of the instrumentation at the gage location can be removed by shortening, with a resistance very small compared to circuit resistance, at the instru- ment.Z The strain system is then significantly simplified; in fact, from a comparative examination of the direct -resistance sys- tem with the traditional one based on the Wheatstone-bridge configuration, evidence emerges that on the first system a P. Cappa (SEM Member) is Associate Professar, and Z. Del Prete is Researcher, University of Rome, "La Sapienza," Via Eudossiana, 18, 00184 Rome, ltaly. K.G. McConnell (SEM Member) and L Zachary (SEM Member) are Professors, Department of Aerospace and Engineer- i ~ Mechanics, University of 1owa, Ames, lA 50011. Paper was presented at the SEM Vll1nternational Congress on Experi- mental Mechanics held in Las Vegas, NV on June 8-11. Originai manuscript submitted: September 27, 1992. Fina[ manuscript received: February 25, 1993. constant voltage supply unit, precision resistors to com- plete the bridge, to balance and to calibrate the bridge output are not utilized. However, the intrinsic simplicity of the direct-resistance method implies the utilization of a constant-current supply unit and the impossibility to com- pensate the effects caused by test-area temperature vari- ations as, on the contrary, it is possible, as is well known, utilizing half-bridge and full-bridge configurations. Moreover, the development of intelligent instrumenta- tions makes possible sequential multichannel automatic strain-gage systems simply based on a computer, a switch unit and a digitai ohmmeter. In previous papers 3 ' 4 the Iimits of stability of automatic sequential systems based on the four-wire direct-resistance method were experimentally examined. It was decided to test the effectiveness of the method w ben general-purpose low-cost mechanical switches are utilized. The zero-drift values observed in six-day static tests were always in ranges of + 10 to -30 Jlmlm in test-area temperature vari- ations of 24 to 50C. Successive experimental tests 5 on the utilization of a high-speed digitai multimeter (100,000 readings/s) showed significant inaccuracy (200 )lm/m) that can be reduced adopting specific averaging scheme, to a value of 50 )lm/m. Therefore, the direct-resistance method seems to bave some effectiveness only in quasistatic measure- ments. However, the direct-resistance method does not cancel factors such as thermoelectric and electrochemical effects that can change measurement accuracy. To overcome these limitations, the reverse-current method was pro- posed and experimentally validated in preliminary tests 6 conducted on a single-channel configuration. The results showed that the reverse-current method is superior in holding the zero-gage resistance reading at the cost of using severa! meters (reversible-current source, digitai ammeter, digitai voltmeter). This paper will comparatively examine the limits of stability of two sequential multichannel automated sys- tems that use a four-wire connection wiring scheme, one based on the direct-resistance method and the other based on the reverse-current method. Experimental Mechanics 293 COINTROLn..IEIR CONTROIL.IL.IE .. CONISTANT enal I'IEVEIRS CUJIRI'IENT SOUJRCE DOGO TAL AM IMl ElJlEI'I DIOUTAIL. VOIL.TMETIE .. MIIJLTOPLEXEI'I Thermocouple reference CONSTANT enal I'IEVERS CURRENT SOURCE DOGOTAL AMIMJIETEI'I DUOUTAIL. VOIL.TIMIETE .. MIIJLTOPLEXEI'I Thermocouple reference l .... l +or- Fig. 1-Scheme of the experimental arrangement; (a) test A, (b) test B 294 December 1993 CONTROl UNI T l ..... l +or- l -o.-+ o v lE INl c Hl A M IB lE R lC LAB SWITCH CONTIROl UN!T -- :SG #2 re l ...... Experimental Test Setup and Procedure T o evaiuate the metrological performances of automatic systems in long-term data-acquisitions based on both the direct-resistance method (DRM) and on the reverse-cur- rent method (RCM), two electrical resistance strain gages were applied on a cantilever. 1t was decided to utilize general-purpose strain gages (constantan foil in combination with polymide hacking). A two-element rosette was chosen for this work (gage length: 1.5 mm; grid width: 1.6 mm; nominai resistance: 120 Q). The rosette was applied to the cantilever, in accordance with the manufacturer' s recommendations, by means of a methyl-2-cyanoacrylate adhesive that is gener- ally suggested for routine experimentai anaiysis applica- tions (long-term operating temperature range: -32 to 65C). The gage installations were checked before and after the tests by measuring the resistance between the gages and the ground. The vaiues observed, aiways higher than 20 GQ, assured good mechanical properties of the adhesive layers. The observed behavior of the two-element rosette installation is judged acceptable. Two different lead wires were chosen in the two tests that w ere conducted: a four-conductor flat cab le ( diameter 0.4 mm) and a four-conductor twisted cable brieded shielded (diameter 0.25 mm). The adopted experimental test setups are schematized in Fig.l(a) and l(b). Alow-costgenerai-purpose switch-con- trol unit was utilized. The switches had low thermai offset ( < 3 j..LV at the end of their life); but this source of error was not compensated. T o measure the current injected and the voltage drop developed over the strain gages, a digitai ammeter and a digitai voltmeter were adopted; the metrologicai performances of the chosen devices are re- ported in Tables lA-lB. The reversible dc-current source, used to drive current through the switches, the lead wires and the gage resistances is characterized, for the selected current values, by the metrologicai characteristics reported in Table IC. To monitor the environmentai temperature variations, chromel aiumel thermocouples were utilized. The thermo- couple outputs were measured by a digitai voltmeter via a multiplexer capable to complete the thermocouple circuits with a cold junction and to automaticaily compensate the cold-junction temperature differences from ooc (reference junction compensation accuracy of O. l 0 C). The devices are interfaced by means of the IEEE-488 Bus. The experimentai data, are gathered reduced and stored by a microcomputer. In the DRM anaiysis, the current source is set to drive, through the gages, four current vaiues equai to 500 j..LA, l mA, 5 mA and l O mA; the developed voltage drops are measured by the digitai voltmeter. By means of the adopted procedure, the four-wire resistance measurement available in the digitai multimeter were simulated. How- ever, it was observed that in the commerciaily available digitai ohmmeter, it was possible to drive only one current vaiue which was generaily ::::1 mA, fora load resistance of :::100 Q. The DRM tests were conducted gathering the voltage drops, and calculating the resistance values by means of the selected nominai values. Hence, the resis- tance vaiue Rn,i at the generic nth acquisition and for the ith current vaiue selected, was obtained from R =V.,; n,l /; (l) where V.,; = the generic voltage drop measured for the ith current vaiue selected l; = the nominai current vaiue injected by the current source The frrst resistance values, relative to the two examined gages, were used as the reference readings. The zero shift vaiues Ezs,n,i for the nth reading and for the ith current value was then calculated from the relationship w bere R -RI Ezs.n,i = ; R ' . l,i R . ; = the generic resistance vaiue relative to one of the four current vaiues selected Ru = is the frrst resistance vaiue relative to one of the four current vaiues selected F = the gage factor (2) The RCM has the metrologicai advantage of canceling the effects of any dc-offset voltage that occurs because of wire-connection thermocouple effects. In the chosen ex- perimental procedure the current source is set to drive a selected current vaiue through the switches, the lead wires and the gages in both directions. In the RCM anaiysis, both the current injected and the voltage drops are measured by the digitai ammeter and the digitai voltmeter. The voltage- current relationships for the strain-gage resistance R 8 are given by where Vp = the positive voltage reading lp = the positive current reading Ll VP = the positive current error voltage v. = the negative voltage reading I. = the negative current reading Ll V. = the negative current error voltage Subtracting eq (4) from eq (3) gives where M is the error induced by Ll Vp-Ll v . (3) (4) (5) If the thermoelectric effects can be assumed to be the same for both positive and negative currents, then M should be nearly zero. Thus, the RCM appears to be able to reduce this source of error. In RCM anaiysis, a complete set of data for strain-gage resistance evaluation consists of Experimental Mechanics 295 TABLE 1-METROLOGICAL PERFORMANCES OF CHOSEN DEVICES a) Current Source Accuracy: (percent of programmed output + amps) Range 1 mA 10mA Maxlmum 24 Hour Resolutlon (T s ~ 1 C) 1 nA 0.001 percent + 9.5 nA 1 O nA 0.0031 percent + 98 nA Temperature Coefficient: (percent of programmed output + amps)fOC 90Day (TcatSoC) 0.0052 percent + 20 nA 0.0074 percent + 220 nA 1 mA 0.00036 percent + 1.2 nA 10 mA 0.00036 percent + 14 nA Noise and spurious responses (to 250kHZ) peak-to-peak rms 0.02 percent of range + 0.7 !JA 0.003 percent of range + 70 nA b) Digitai Ammeter Accuracy: (ppm of readings + ppm of range) Maxlmum Range Resolution 24 Hour 1 mA 10mA 100 pA 1 nA Temperature Coefficient: (ppm pf readings + ppm of rangefOC) 1mA 2+1 10mA 2+1 c) Digitai Integrating Voltmeter + Relay Multiplexer Range 300mV 3V Maximum Resolutlon 100nV 1 flV NPLC = Number of Power Line Cycles (integration time) 10 +3 10 + 3 90Day 15 + 5 15 + 5 90Day 18to 28C 0.008 percent + 8flV 0.008 percent + 10 flV Temperature Coefficient =no additional accuracy error occurs when operating inside 18 to 28C. TABLE 2-TEST EXPERIMENTAL PROCEDURES DRM RCM Remarks CH.l CH.2 CH.l CH.2 Data row collected: every 20 min. for 2 weeks +500 flA +500 flA 500 !lA 500 !lA TestA +l mA +l mA l mA l mA l O m shielded twisted wires from devices to strain gages into +5mA +5mA 5mA 5mA the oven; two thermocouple measurements T 1ab and T oven +lO mA +lO mA lO mA lO mA +500 !lA +500 flA 500 flA 500 IlA TestB +l mA +l mA l mA l mA l m unshielded wires from devices to strain gages in +5mA +5mA 5mA 5mA laboratory environment; one thermocouple measurements +lO mA +lO mA lO mA lO mA T,ab 296 December1993 positive and negative voltage and current values. The RCM metrological performances were tested with the same current values chosen in the DIRM analysis, i.e., 500 J.lA, l mA, 5 mA, 10 mA. Finally, the Ez.r.n.i values are calculated utilizing eq (2). Both the DRM and the RCM were examined by conduct- ing two tests; in the frrst test, test A, the two-element rosette is connected to the switch-control unit by shielded wire leads approximately 10-m long. T o smooth the gage- installation temperature variations, the two-element ro- sette installation was placed in an oven. One of the two thermocouples was utilized to measure Toven inside the oven near the gages. The latter T /ab was utilized to monitor the temperature outside the oven. In the second test, test B, the gages were connected to the switch-control unit by unshielded wires approximately l-m long. The gage instal- lations followed the laboratory temperature T tab variations and only one lab thermocouple was used and placed at the strain-gage locations. Both tests A and B lasted for two weeks and each data row, summarized in Table 2 was collected every 20 rninutes and the current values were forced only for the interval of time necessary for the controller to read the current and the voltage drop devel- oped (=3 ms). The sequence followed in the data-acquisi- tion process is reported in Fig. 2. Test Results The experimental results relative to DRM analysis are reportedinFig. 3(a) andFig. 3(b) fortestAandinFig.4(a) and Fig. 4(b) for test B. From an examination ofFig. 3(a) and Fig. 3(b ), a sirnilar behavior of the two examined Genere.toreuppliescurrenti. DAM reads current.I. MPX and SCU cloe CH.l l :; l DVM reada v. o:r CH.l l MPX and SCU close CH.l DVM reeds V_ of CH. 1 YES YES Fig. 2-Fiowchart of the data-acquisition process w ' o E " ..s DIRECT RESISTANCE METHOD 30 10 ;:::
., l
"' N Test A - Stroin goge #1 .. w ' o - :C-1 ., l
"' N Time (hour) (a) DIRECT RESISTANCE METHQD Test A - Stroin goge #2 (b) Fig. 3--Direct-resistance method, test A: zero-shift ver- sus time; a) strain gage #1, b) strain gage #2 DIRECT RESISTANCE METHOD Test B - Strcin goge fl (a) DIRECT RESISTANCE METHOD Test 8 - Strcin goge #2 (b) Fig. 4--Direct-resistance method, test 8: zero-shift ver- sus time; a) strain gage #1, b) strain gage #2 Experimental Mechanics 297 channels emerges. The same observation is not confirmed for test B, see Fig. 4(a) and Fig. 4(b). In fact, test B seems to confirm the high dependence of zero shift values, Ezs.n,;, on the first resistance readings, Rt,;, accordingly with eq (2). If Rl.i indeed is significantly different from the mean value of the all successive Rn,; values, a relevant zero-shift obviously appears; this behavior is more evident for low- current supply values, i.e., 500 J.I.A. The current values chosen and the adopted experimental procedure do not cause appreciable heating effects of the two gag es. The environmental temperature variations, showed in Fig. 5 for test A and in Fig. 6 for test B, seems to cause relevant effects on Ezs.n.i values. From a comparative exami- nation ofFig. 3(a) and Fig. 3(b) with Fig. 5, the dependence of Ezs,n,i on laboratory temperature variations emerges; in fact the five temperature 'spikes' of =.soc cause a Ezs,n,i variation of :::15 J.l.ffi/m. This observed 'temperature in- duced apparent strain' seems to be mainly caused by the temperature coefficients of the devices, see T ab le l. The dependence of Ezs.n.i on T IDb is not denied by a comparative examination ofFig. 4(a) and Fig. 4(b) with Fig.6; in this case two observed spikes of 2-3 o c determine an apparent strain of =5 J.l.rnlm. Moreover in test B it clearly appears that a positive T IDb trend causes a negative Ezs.n.i trend. However, variations in the range of 18-24C observed in test A determine a zero shift always in the range of -15 to ~ 2 2 30 Q) a.20 E Q) """ Test A 96 144 192 240 Time (hour) Fig. 5-Temperature variations as a function of ti me; test A i!' ~ ~ -T., a_20 E "' f- Test B Fig. 6-Temperature variations as a function of time; test B 298 December1993 REVERSE CURRENT METHOD ' o E ....... ..s JO 10 :.c -10 "' l e "' N (a) REVERSE CURRENT METHOD ...... ' o E ....... ~ ~ JO 10 :.c -10 (/) l e ., N Test A - Stroin goge 111 Test A - Strain goge #2 -Jof.o ~ .. ' ' ~ ' , . . . , - . ~ , . , . , , . . . ~ . . . , . , , , , ~ . , . , , ' . , . . , . . , . ~ ' . , . , 6 6 ~ ' ' Time (hour) (b) Fig. 7-Reverse current method, test A: zero-shift ver- sus time; (a) strain gage #1, (b) strain gage #2 REVERSE CURRENT METHOO Test B - Strain gage 11 (a) REVERSE CURRENT METHOO Test 8 - Stroin goge fl2 (b) Fig. 8-Reverse current method, test 8: zero-shift ver- sus time; (a) strain gage #1, (b) strain gage #2. + 20 J!rnlm; w bile in test B a temperature range of 27-30C determines differences between the maximum and the minimum of Ezs.n.i values always less than 28 J!rnlm. The results relative to RCM are reported in Fig. 7(a) and Fig. 7(b) for test A and in Fig. 8(a) and Fig. 8(b) for test B. Insignificant differences emerge both in test A and test B from the obtained zero-shift values of the two examined channels; in fact, the Ezs.n.i is always in the range of -18 to + 17 J!rnlm for test A and -7 to + 7 J!rnlm for test B, with the exception of the current value of 500 J!A which shows a nonrepetitive scatter, probably caused by the same reasons (RtJ readings) previously indicated for DRM tests. The obtained experimental results indicate that, as far as the examined system, the RCM do not improve metrologi- cal performances as accuracy and precision with respect to the DRM in spite of the adopted general-purpose low-cost mechanical switch-control unit. Conclusions The results relative to a multichannel automatic system based on the direct-resistance method indicate irrelevant zero-shift values in tests conducted over two weeks with a temperature variation ranging from l8C to 24 o c and from 2rc to 30C and confmn the effectiveness of this simple method also in system based on low-cost switch-control uni t. Almost identica! zero-shift trends and ranges of vari- ations were observed adopting the reverse-current method. Relative to the examined experimental setup, this method seems to not confirm the superiority in maintaining the zero-gage resistance readings for long-term data acquisi- tion. References l. Nelson E.J., Sikorra C.D. and Howard J.L, "Measuring Strain Gages Directly Without Signa[ Conditioning," EXPERIMENTAL TECH- NIQUES, 7, (9), 26-28 (1983). 2. Zachary L W., McConnell K.G. and Younis N. T., "Accounting for Lead Wire Resistance Changes and Loss of Zero in Long-term Strain Measurements," Proc. SEM 1990 Spring Conference on Exp. Mech., 201-204 (lune 1990). 3. Cappa P., "A Comparative Examination of Automatic Sequential Direct Systems for Strain-gage Data Readings Based on a Low-cost Switch-contro[ Unit," EXPERIMENTAL TECHNIQUES, 13-15 (Sept. 1989). 4. Cappa P., "An Experimental Analysis ofthe Zero-shift Values of Automatic and lnexpensive Strain-gage lnstrumentation Systems," EX- PERIMENTALMECHANICS, 31 (l), 88-92 (March 1991). 5. Cappa P. and Del Prete Z., "An Experimental Analysis of Accuracy and Precision of a High Speed Strain-gage System Based on the Direct- resistance Method," EXPERIMENTAL MECHANICS, 32 (l), 78-82 (March 1992). 6. Cappa P., McConnell K.G. and Zachary L W., "Zero-shift Values of Automatic and lnexpensive Strain-gage Instrumentation Systems," EXPERIMENTALMECHANICS, 31 (l), 88-92 (1991). Experimental Mechanics 299