THERMOFOCUS


The non-contact thermometer

THERMOFOCUS

, a technological innovation

THERMOFOCUS has been developed by Tecnimed, the Italian company specialized
in the development and manufacturing of electronic medical devices.
THERMOFOCUS is the first non-contact clinical thermometer in the world and the
only one that measures the temperature in a specific point and at an accurate and
determinate distance.
It provides instantaneous digital reading of body temperature without any contact
with the body skin or mucosa.
THERMOFOCUS does not use laser radiations, is safe and harmless.
THERMOFOCUS can also be used to measure the temperature of babies feeding
bottles, bath water etc., thanks to its accuracy over a wide rane of temperatures.
THERMOFOCUS is protected by a number of patents and patent pending filed
internationally.
THERMOFOCUS

is a trade mark registered in Italy and extended internationally.



Technical background

All objects and living beings emit infrared radiations of varying wavelength in
relation to their surface characteristics. Particularly, the human body emits infrared
radiations of wavelength between 5 and 14 micrometers.
THERMOFOCUS uses a sensor (thermopile) that, when stimulated by infrared
radiations, sends an electrical signal, which is amplified and then converted into a
digital signal which gives the correct body temperature, after the devices automatic
adjustment to room temperature.

Temperature measurement

Body temperature is measured by pointing the THERMOFOCUS thermometer at the
center of the forehead. The right distance between the skin and the thermometer
for a correct measurement is easily determined thanks to a LED system emitting
two light beams (normal light which is totally safe). As the thermometer is moved
closer to the skin and reaches the right distance (approx. cm 3), the two beams
converge to form a single spot. Release the button and keep the thermometer still
for less than one second until the lights flash and the temperature appears on the
LCD display. There is no need for concern if the aiming lights are pointed into the
eyes: the beams are absolutely harmless and are not laser lights.
In case the subject has a perspiring forehead or is crying (i.e. a child) or for elderly
people especially with wrinkled forehead (a low supplying of blood may cause a
low temperature) - or with oxygen mask, the measurement must be taken on
alternative areas such as preferably on the middle of the closed eyelid. If the
measurement on the closed eyelid is not feasible (i.e. due to inflammation, surgery
or other) the third recommended area to measure the body temperature is on the
neck in correspondence with the jugular vein.




THERMOFOCUS

helps hospitals save costs



While other thermometers require disinfection after each use or need expensive
hygienic and disposable covers, the Thermofocus thermometer does not. This
advantage can result in significant savings in hospitals. Lets try to calculate how
much it is: normally the cost of a probe cover for an ear thermometer is between 4
and 10 cents (in Euros) depending on the kind of thermometer. Assuming an
average cost of 5 cents each and assuming 2 measurements per day in a ward with
50 beds. The cost is 0.05 x 2 x 50 = 5 each day for disposable covers. At 365
days a year, the annual cost is 1,825 for each thermometer! A total waste of
money.
Besides the needing of batteries is also favorable to Thermofocus.
If you calculate the time saved by nurses in each measurement (insertion of the
cover on the probe, positioning in the ear canal, taking away the probe = about 15
seconds x 0.5 cents for each second) x (2 x 50 x 365) measurements, youll obtain
more than 2,700 each year. Hence, we can assume that each Thermofocus
allows the hospital to save 4,000/5,000 each year!!!!
Both the new version of the 01500 model and the 0800 series have a
patented quick stabilization system that makes the Thermofocus
thermometer particularly suitable for hospital use, where room temperature
can change from room to room, as well as on ambulances and in private practices.
There are two different stabilization technologies available.
In case of a very fast change of the devices temperature, the display will show,
through a countdown, the time that is necessary to wait in order to have the device
quickly stabilized. This technology is called AQCS (Automatic Quick Calibration
System) and is normally able to stabilize the device in about 5 minutes, no matter
how much the room temperature is different.
As an alternative, it is possible to use also the MQCS (Manual Quick Calibration
System) which allows the device to immediately stabilize its temperature with the
one of a reference element for example the target disc - placed in the room. In
this way the device is able to stabilize itself in just 3 seconds.


THERMOFOCUS

is multipurpose

Thermofocus can take any measurement out of the forehead on objects or other
part of the body with temperature included between 1 to 55C (33.8 to 131.0F).
For example, a mother can use this function to take the temperature of the feeding
bottle, soup, bath water, rooms window etc.; a doctor can use Thermofocus to
scan the temperature of a certain area of the skin and comparing with the area
around or with the symmetrical part of the body (the two legs, the two feet, etc.),
can find anomalies due to inflammations, blood circulation problems, diabetes.
Moreover, it is possible to use Thermofocus on open wounds and to use it on
internal organs during surgical operation, because there is no contact at all.
The device is supplied by 4 AAA batteries (included in the box) that have a life of
20,000 measurements or 3 years, according to use. The only necessary
maintenance is to protect the gilt wave-guide with its protective dust-tight cap
when the device is not in use.









THERMOFOCUS

models




Basic model:
THERMOFOCUS 0700A2
Thermofocus 0700A2 is the most advanced thermometer a family can own to
precisely measure body temperature. Totally hygienic, without touching,
Thermofocus reads infrared radiation naturally emitted by the surface of the skin
and calculates the whole body temperature. Simply hold the thermometer close to
the center of the forehead and release the activating button. It can also be used to
take the temperature of objects such as a babys feeding bottle by simply
subtracting the value written on the label overleaf of the product. The memory
button allows to recall the last 9 measurements. The 0700A2 model is provided
with the AQCS (Automatic Quick Calibration System) technology.



THERMOFOCUS
01500A3
Thermofocus 01500A3 is thought for professional users or families that require
maximum accuracy and flexibility of temperature taking. It guarantees the same
performance of model 0700, with an additional button (home) which allows the
user to easily and directly read the temperature of objects as well as other areas of
the body, in a range from 1 and 55C (33.8F to 131F). The LCD display
automatically shows room temperature when in stand-by. This model is provided
with both the AQCS and the MQCS (Manual Quick Calibration System)
technologies.



THERMOFOCUS
01500A/H1N1

Thermofocus 01500A/H1N1 is specifically studied for being used in hospitals,
airports, schools, factories, etc. in case of emergency and/or pandemic situations.
This model is particularly quick, accurate and easy to use, allowing several fast and
hygienic temperature readings. 01500A/H1N1 is provided with the MQCS
technology (mandatory procedure in the nurs setting) and with the AQCS
technology (automatic in the doct setting) comes with a special lanyard which
allows to safely carry the device, in order to use it anytime if needed.
- 3 different modes: Oral, Rectal and Axilla
- default setting "nurs": HOME button disabled; MQCS requested and mandatory
every 15 minutes (when used by nurses as well as for any intensive use, like in
airports etc.);
- optional setting "doct": HOME button enabled; AQCS automatic and optional
MQCS (for doctors use);





- air conditioning adjustment On or Off.
Thermofocus 01500A/H1N1 can take thousands of accurate readings every hour,
24/7, with no need to stop at all between consecutive measurements.
On request, it can also be available with a working range from 5 to 40C (=41 to
104F), instead of 16 to 40C.








THERMOFOCUS
0800H5

Thermofocus 0800H5 is the infrared non-contact thermometer specifically studied
for hospital use, thanks to the mandatory MQCS function, which allow doctors,
nurses and hospital staff to take temperatures precisely even if they are moving
quickly from room to room, if the device is kept in the pocket or has been handled
for a long period. The light-bulb button
permits to re-activate the backlight for 2
seconds.
Thermofocus 0800H5 is supplied with a
special lanyard with brief instructions, which
allows to safely carry the device and a
special Target Disc to facilitate the MQCS.
- amber-colored LED lights
- 3 different modes: Oral, Rectal
and Axilla
- default setting nurs: MQCS
requested and mandatory every 15 minutes (when used by nurses as
well as for any intensive use, like in airports etc.)
- optional setting "doct": if the thermometers temperature is
changing, the AQCS (Automatic Quick Calibration System) is
automatically activated and allows the rapid stabilization of the
thermometer to the room temperature: the display shows a
countdown indicating the time required. To make a
measurement is necessary to wait the end of the countdown
without touching the device or is possible to do the manual
MQCS (suggested procedure).
Using the face button it is possible to take also surface
temperatures (different from the body temperature) between 1 to
55C (33.8 to 131F). The display will show alternatively the temperature value
and Lo.3, if the value is <34C (93.2F), or Hi.2, if it is > 40C (104F).
These measurements are not entirely accurate and must be considered as
relative rather than absolute values. If taken on the skin, these values will show
the difference between two close or symmetrical areas (i.e. the two hands, the
two knees, etc.). This setting is studied for doctors use and also for the use in a
specific ward as the Emergency Room;

- air conditioning adjustment On or Off.




Manufacturing and QA
THERMOFOCUS is manufactured by Tecnimed under the ISO 9001:2008, ISO
13485:2003, ISO 13485:2003 CMDCAS, EN 60601-1 and UNI CEI EN ISO
14971:2009 Quality System.
The production, control and calibration of THERMOFOCUS are performed by
Tecnimed in a Class 100 Clean Room.
The packaging is made in an Ambient Controlled room.
The quality of the product is further certified by external independent institutes.

THERMOFOCUS is a class IIa Medical Device as measurement instrument CE
approved (CE 0051) according to 93/42 CEE Directive and TGA (Australia), FDA
(USA), Koseisho (Japan), KFDA (Korea) and SFDA (China) approved and tested
according to ASTM (American Society for Testing Materials) E 1965-98(2009).
Accordance with RoHS (Pb-free) directive.


Awards

2010
Thermofocus has been awarded First Place and Winner in the NACDS Product
Showcase competition for all Healthcare/OTC products, which took place during
NACDS Marketplace in San Diego, from 5 to 8 June 2010.
Judging was conducted by a 15 member panel of Drug Chain retail buyers and the
amazing result confirms its technological innovation, made in Italy and protected by
a number of patents worldwide.





2009
Thermofocus HUMANA has been voted by the Italian public Product of The Year
2010 thanks to its innovative, reliable and user-friendly technology.










2008
Thanks to the innovative, unique and effective technology of Thermofocus,
Tecnimed has been rewarded with the prestigious Frost & Sullivan Innovation
Award 2008 in the East European Temperature Monitoring Devices Market. The
Award is presented on an annual basis to the company that has demonstrated
technological superiority within its industry and recognizes the ability of the
company to successfully develop and introduce new technology, formulate a well-
designed product family, and make significant product performance contributions to
the industry.



TECHNICAL DATA

THERMOFOCUS 0700A2 01500A3
01500A/H1N
1
0800H5
SUGGESTED USE:



CHARACTERISTICS:
For families
that like
Thermofocus
technology but
cannot afford
the price of
01500A3
For families
that want
the best and
for private
doctors that
use the
device in
their studio
For emergency
situations
( airports,
factories, schools,
etc.)
For professional
use ( hospitals,
clinics, etc.)
No. of buttons 2 3 2
LED lights color red amber
Protective cap standard
Display backlight x
Room temperature detection
only for 2
seconds
x x x
Available modes:
Oral x x x x
Rectal x x x x
Axilla x x
Available settings ( on 0800H5 and 01500A/ H1N1 models only) :
nurs ( default)
HOME button
disabled, MQCS
mandatory every 15
MQCS mandatory
every 15
doct
HOME button
enabled, MQCS not
mandatory
AQCS automatically
activable
AQCS
automatically
activable
MQCS not
mandatory
measuring range
1/55C*
(33.8/131F)
AI R YES (air conditioning on) x x
AI R NOT (air conditioning off) x x
AQCS ( Automatic Quick Calibration
System)
x x DOCT: automatically activable
MQCS ( Manual Quick Calibration System) x
NURS: mandatory
every 15 minutes
DOCT: available but not mandatory
Mirror Target Disc 1 included
Brief additional instructions on battery
door
x x
Lanyard with additional brief instructions 1 included 1 included
Batteries ( included) 4 AAA/LR03 type (preferably alkaline)
Batteries life 3 years or 10.000 measurements
20.000
measurements
Resolution 0.1C (0.1F)
Forehead measuring range 34.0/42.5C (93.2/108.5F)
General measuring range ( apart from
forehead)
1.0/55C (33.8/131F)
Room temperature working range:
measurement on the forehead
16/40C
(60.8/104F)
16/40C (60.8/104F)
16/40C
(60.8/104F)
On request: 5/40C
(41/104F)
other measurements 5/16C (41/60.8F) please note that the accuracy is not guaranteed
Accuracy level ( in instrumental tests according to ASTM E 1965- 98 ( 2009) standard) :
from 36 to 39C (96.8/102.2F) = +/-0.2C (+/-0.4F)
from 34 to 35.9C (93.2/96.6F) = +/-0.3C (+/-0.5F)
from 39.1 to 42.5C (102.4/108.5F) = +/-0.3C (+/-0.5F)
from 1.0 to 19.9C and from 42.3 to 55.0C = +2.0/-1.0C +/-1.0C
(from 33.8 to 67.9F and from 108.1 to 131F)
=
(+3.6/-1.8F) (+/-1.8F)
from 20 to 33.9C = +2.0/-0.3C +/-0.3C
(from 68.0 to 93.0F) = (+3.6/-0.5F) (+/-0.5F)
Distance from the subject during operation: about 3 cm (1.2 inch), set through an optical signal
* The measurements are not entirely accurate and must be considered as relative rather than absolute values