Академический Документы
Профессиональный Документы
Культура Документы
Corresponding author. Tel.: +55 81 2126 8708; fax: +55 81 2126 8708.
E-mail address: hjkhoury@globo.com (H.J. Khoury).
30 cm30 cm10 cm and covered by a Lucite plate with
the appropriate thickness, for the electron beams energy, to
provide buildup. Measurements were performed with the
sensor positioned at the center of a 15 cm15 cm eld size
and with the phantom surface at 100 cm from the source.
The response of the diode was also evaluated for
different dose rates, ranging from 80 to 240 cGy/min.
To determine depth dose percentage, the plate contain-
ing the diode was placed in the center of the electron beam,
over a phantom made of PMMA plates of variable
thickness. The measurements were made by varying the
depth of the plate containing the diode between 0
(phantom surface) and 5 cm, using a 15 cm15 cm cone.
The percentage dose at a given depth was calculated as
ratio of the reading at that point and the reading at the
point of maximum dose.
3. Results and discussion
Fig. 1 shows the photodiode response for doses in the
range of 20120 cGy. The response of the photodiode was
not studied for doses lower than 20 cGy as these doses are
not used in radiotherapy. A least square tting of the
experimental data indicated an excellent linear behavior in
this interval, with a determination coefcient of 0.99998.
Table 1 shows the results obtained with the diode for the
monitor unit set in 50 and different values of the dose rates.
The data show that, for the three electron energies, the
diode response is independent from the dose rate and the
variation in its response is lower than 2%.
Fig. 2 shows the results of the measurements carried out
with the diode XRA-50 in terms of the percent depth dose
at the center axis for the 15 cm15 cm cone for 4 and
9 MV electron beams. The depth is expressed in terms of
millimeters of water, calculated by taking into account the
densities of Lucite and water itself.
The mean energy of the electrons at the phantom surface
(E
0
), the mean range and the half-value depth (R
50
) were
estimated from the graphics of Fig. 2. The value of E
0
was
obtained through the relationship E
0
C.R
50
, where
the constant C is equal to 2.33 MeVcm
1
for water [5].
The mean range of electrons (R
p
) was calculated by
extrapolation of the dose depth percentage curve for the
eld of 15 cm15 cm. Table 2 shows the results obtained.
The results are compared with the values previously
obtained by the radiotherapy clinics, with the ionization
chamber. The results show that the values obtained with
the diode XRA-50 agree with the values obtained with the
ionization chamber.
4. Conclusions
The results indicate the linearity of the commercial diode
XRA-50 doseresponse curve, obtained with 4, 9, and
ARTICLE IN PRESS
Fig. 1. Doseresponse curves of XRA-50 for 4, 9, and 15 MV electron
beam.
Table 1
Response of the diode XRA-50 in function of the dose rates for the beam
energy of 4, 9, and 15 MV
Electron beam (MV) Dose rate (cGy/min)
80 160 240 S
a
%CV
4 2098 2118 2041 40.1 1.9
9 2086 2093 2074 9.6 0.46
15 1945 1900 1924 22.5 1.2
The monitor unite was set in 50.
a
S, standard deviation.
Fig. 2. Depth dose data obtained with XRA-50 at the central axis for the
15 cm15 cm cone eld of 4 and 9 MV electron beam.
Table 2
Mean range, mean electron energy at the phantom surface, and half-value
depth range obtained with the XRA 50 and electron beam from 4 and
9 MV
Beam energy
(MV)
Diode
XRA-50
Ionization
chamber
4 Mean range (water cm) 1.65 1.80
Mean energy E
0
(MeV) 2.80 3.0
R
50
(cm) 1.2 1.3
9 Mean range (water cm) 4.3 4.5
Mean energy E
0
(MeV) 8.3 8.4
R
50
(cm) 3.6 3.8
H.J. Khoury et al. / Nuclear Instruments and Methods in Physics Research A 580 (2007) 537539 538
15 MV electron beam. The agreement between the data
obtained with the diode and the ionization chamber shows
that the use of the XRA-50 is a reliable method for electron
beam dosimetry.
Acknowledgments
The authors would like to acknowledge the nancial
support received from the Brazilian National Research
Council (CNPq) and the Erasto Gaertner Hospital.
References
[1] I. Griessbach, M. Lapp, J. Bohsung, D. Harder, Med. Phys. 32
(2005) 12.
[2] R. Dixon, K.E. Eckstrand, Int. J. Appl. Radiat. Isot. 33 (1982) 1171.
[3] V.C. Colussi, A.S. Beddar, T.J. Kinsella, C.H. Sibata, J. Appl. Clin.
Med. Phys. 2 (2001) 210.
[4] R. Yaparpalvi, D.P. Fontenla, B. Vikram, Int. J. Radiat. Oncol. Biol.
Phys. 48 (2000) 1259.
[5] S. Klevenhagen, Physics and Dosimetry of Therapy Electron Beam,
rst ed., Medical Physics Publishing, Wisconsin, 1993.
ARTICLE IN PRESS
H.J. Khoury et al. / Nuclear Instruments and Methods in Physics Research A 580 (2007) 537539 539