Академический Документы
Профессиональный Документы
Культура Документы
Corresponding author at: Beijing Center for Disease Control and Prevention,
Beijing 100013, China. Tel.: +86 10 64407087; fax: +86 10 64407210.
C.
2.3. Preparation of calibration curve
A TGA standard stock solution (10mgmL
1
) was prepared daily
by accurately weighing 20mg of calcium thioglycolate trihydrate
into a 1.5mL of sample vial. 1mL of water was added and then
vortex-mixed. A series of standard solutions at the nal concentra-
tions of 0.006, 0.025, 0.125, 0.250, 0.500 and 1.000mgmL
1
were
made and kept in dark to avoid possible photodissociations [16].
2.4. Sample pretreatment
An appropriate amount of samples were weighed into a 15mL
centrifuge tube with a cap and then diluted with sample buffer
to the mark of 10mL. The solutions were thoroughly vortexed to
make the samples homogeneous. The mixed water-based solutions
could be injected directly, while the mixed oil-based samples must
be centrifuged at 9000rpm for 10min and then the resultant clear
supernatant was directly injected.
3. Results and discussion
3.1. Optimization of CE conditions
3.1.1. Selection of the running buffer
TGA is a weak organic acid with two pKa values, pKa
1
=4.32 and
pKa
2
=10.20 which comes fromthe dissociation of its carboxyl and
thiol groups, respectively [17]. As a rule of thumb, the pHof the run-
ning buffer should be adjusted to at least two units belowthe pKa
1
or above the pKa
2
of TGA to ensure its complete ionization [18].
Phosphate, with low UV absorption and wide pH buffer capacity
Table 1
Comparison of the results of the TGA in 16 cosmetic samples determined by CE,
HPLC and IC.
Sample CE HPLC IC
Content (%, w/v) Content (%, w/v) Content (%, w/v)
1 / /
2 / /
3 0.7
4 / / /
5 / /
6 4.1 4.0 4.3
7 1.3
8 1.1
9 8.7 8.4 8.2
10 8.1 7.7 8.0
11
a
2.7 2.6 2.8
12
a
3.4 3.3 2.9
13
a
2.3 2.4 2.3
14
a
3.7 4.0 3.7
15
a
2.7 2.7 2.6
16
a
3.4 3.4 3.4
/, Not detected; , cannot be quantied due to interferences.
a
Oil-based sample.
(pKa
1
=2.12, pKa
2
=7.20 and pKa
3
=12.36), is a popular buffer sys-
tem for CE. Considering that TGA salts at pH 7.09.5 or 12.7 and
TGA esters at pH6.09.5 are usually used in cosmetic formulations
[6] and only phosphate can provide a good buffer capacity at pH
higher than (pKa
2
(of TGA) +2), phosphate was chosen as a priority
running buffer for the subsequent optimizations.
3.1.2. Optimization of the concentration of Na
3
PO
4
and the pH of
the separation buffer
Although the analysis of only a single analyte TGA in cosmetic
samples was studied in this work, however, in such a case, the
researchers were often dismayed by the interferences coming from
the sample matrix. In the present work, a real sample 12 (listed in
Table 1) was used to optimize the following conditions of the CE
method.
The concentration of Na
3
PO
4
played a key role for the sep-
aration of TGA. Keeping the concentration of CTAB (0.5mmol/L)
unchanged, the concentration of Na
3
PO
4
(without pH adjustment)
increasing from100mmol L
1
to 350mmol L
1
, not only decreased
the EOF, but also suppressed the adsorption of TGA to the capillary
wall. As shown in Fig. 1, the detection sensitivity was enhanced and
the separation efciency was increased. However, high concentra-
tions of phosphate would produce large amount of Joule heat due
to its inherently high conductivity, which could make the TGApeak
broadened. As a result, 300mmol L
1
Na
3
PO
4
was chosen.
The pHof running buffer is another key factor for the separation
of TGA from the cosmetic sample matrix. It controls both the TGA
charge state and the level of EOF. Since the running buffer with pH
close to the pKa
3
value of H
3
PO
4
(12.36) can provide better buffer-
ing capacity [19], the effect of the running buffer at pH12.35, 12.60
(without pH adjustment) and 12.85 were investigated while the
concentration of 300mmol L
1
Na
3
PO
4
was kept constant. It was
found that under all the three pHvalues, the TGA could be baseline
separated from the interfering peaks coming from sample matrix.
Considering that the running buffer at pH12.60 could be easily pre-
paredandhadahighbuffer capacity, 300mmol L
1
Na
3
PO
4
without
pH adjustment was chosen for the following experiments.
3.1.3. The selection of the EOF-reversing agent and its
concentration
Samples are normally injected at the anode and detected at the
cathode. At pH>7, the EOF is sufcient to ensure most species
to move from the anode to the cathode in the order of cations,
N. Xie et al. / Journal of Pharmaceutical and Biomedical Analysis 88 (2014) 509512 511
Fig. 1. Effect of Na
3
PO
4
concentration on separation. Na
3
PO
4
concentration
(mmol L
1
): (a) 100, (b) 200, (c) 300, and (d) 350. (1) TGA and (2) unknown peak.
Running buffer: x mmol L
1
Na
3
PO
4
and 0.5mmol L
1
CTAB. Uncoated capillary:
40.2cm50m, i.d. (30cmto the detector). Injectiontime: 10s. Injectionpressure:
0.5psi. Separation voltage: 5kV, detection: UV 236nm, temperature: 25
C.
neutrals and anions. However, the negatively charged TGA
migrated against the EOF and moved to the anode under the opti-
mized running buffer pH 12.60, which made it difcult to migrate
out of the capillary in a reasonable time (within 30min). To ensure
that the TGA could rapidly move to the detection window, an
EOF-reversing agent was needed [20]. Long-chain alkyl trimethyl
quaternary salts are frequently used as EOF modiers for anionic
analysis. In this study, three EOF modiers, namely DTAB, TTAC and
CTAB were investigated. The migration time of TGA was reduced
along withthe increased lengthof the alkyl chainas showninFig. 2.
It might be relatedtothe surfactant structures, the longer the chain,
the stronger the adsorption and the ability to control EOF [21]. As
a result, CTAB was selected.
Once EOF is reversed and reaches a stable value, the EOF no
longer depends on the CTAB concentration when its concentration
is 0.1mmol L
1
which corresponds to 10% of its standard critical
micelle concentration. On this occasion, the surface of capillary is
a dynamic structure represented by the bilayer assembly instead
Fig. 2. Effect of the EOF-reversing agent types on migration time. (a) CTAB, (b) TTAC
and (c) DTAB. (1) TGA, other CE conditions were the same as in Fig. 1.
Fig. 3. Effect of sample buffer on separation: (a) pure water as the sample medium,
(b) one-tenth dilution of the running buffer as the sample buffer, and (c) running
buffer as the sample buffer. (1) TGA and (2) unknown peak. Other CE conditions
were the same as in Fig. 1.
of the solid backbone of silica [21]. In order to further verify this
observation, CTAB concentrations at 0.2, 0.5 and 1.0mmol L
1
were
studied, respectively, with no differences between the separation
efciency and the migration time observed. Thus, CTAB with con-
centration of 0.5mmol L
1
was chosen because the EOF is stable at
this concentration [14,21].
3.1.4. Choice of sample buffer
For efcient sample stacking, the conductivity of the sample
buffer should be relatively lower than that of the running buffer.
In this study, pure water, one-tenth dilution of the running buffer
and the running buffer were tried as the sample buffer (Fig. 3). As
expected, one-tenth dilution of the running buffer showed the best
result.
3.2. CE method validation
3.2.1. Linearity, limit of detection, limit of quantitation, precision
and recovery
The corrected peak areas (Ac) versus the concentrations (,
mgmL
1
) of TGA showed a good linear relationship (Ac =50.33
+40.174) with a correlation coefcient (r) of 0.9998. The limit
of detection (S/N=3) and limit of quantitation (S/N=10) were
0.002mgmL
1
and 0.006mgmL
1
, respectively.
Sample 12 was prepared seven times within a day to evaluate
the intra-day precision of the CE method. Samples were accurately
weighed (0.1g for each) and treated as described in Section 2.4. To
evaluate the inter-day precision of the CE method, sample 12 was
pretreated in triplicate each day on seven consecutive days. Both
the intra- and inter-day precisions of the method were 1.4% and
the average content of TGA in sample 12 was 3.4%.
The average recoveries at the three spiked levels (0.125, 0.250
and 0.500mgmL
1
) were 96.9%, 102.3% and 94.0% with relative
standard derivations 2.1%, 3.9% and 2.2%, respectively.
3.2.2. Comparison of CE, HPLC and IC results
Sixteen samples of hair-treatment products and depilatory
creams were analyzed by CE, HPLC and IC, respectively. The
results are listed in Table 1. For most of the samples, the
results of the TGA assayed by CE were in good agreement
with the values obtained by HPLC and IC. However, for the
HPLC method, six out of 16 cosmetic samples could not be
512 N. Xie et al. / Journal of Pharmaceutical and Biomedical Analysis 88 (2014) 509512
Fig. 4. Chromatogram and electropherogram of the sample 3. (A) Chromatogram, (B) electropherogram. (1) TGA and (2) interfering peak. Other CE conditions were the
same as in Fig. 2. HPLC conditions: Mobile phase: acetonitrile and 0.01mol L
1
potassium dihydrogen phosphate solution (pH 2.5) with volume ratio of 10:90. Flow rate:
1.0mL min
1
. Detection wavelength: 215nm. Column temperature: 30