June 5, 2008

SECTION 3

ELECTRONICS

3.1 Whole Block Diagram....................................................... 1


3.2 PCB Layout Diagram ....................................................... 2
3.3 PCB No.3061 ................................................................... 3
3.4 PCB No. 6365T2 .............................................................. 6
3.5 PCB No. 2168 .................................................................. 8
3.6

Procedure for Changing CPU board .............................. 10

XS series S/M

June 5, 2008

3.6 Procedure for Changing CPU board


1. Copy the [SetFileViewer.exe] file (attached to TB 200819) to

4. Select ProgramFiles\sysmex\IPU\LOG\20080115\ (latest shut down

FD or memory stick etc.

date) and XS-1000i^05342311^11045.SET and open.

2. Start SetFileViewer.exe copied on XS IPUI.

5. Hard copy the following screen or write down the values.

3. Click [Read] after the following screen is displayed.

6. Change the CPU board.


7. Input the each value wrote down in step 5. as follows.
1) Select [Controller] [Service]
XS series S/M

3-10

June 5, 2008

2) Input Service Setting


3) Input Sensitivity
4) Input Optical Axis Adjustment
5) Input Aspiration Sensor
6) Input RBC Clog Level
7) Input Laser Setting
8) Start Piercer / Pipette Adjustment and click Up, Down, Front, Back
to adjust the indicated pulse.
9) Select Hand Adjustment Pulse Entry and input the pulse.
10) Select Tube Holder Adjustment Pulse Entry and input the pulse.

XS series S/M

3-11

June 5, 2008

4.3.4

(7) When the analysis is over select Sensitivity and confirm that
values are within the following range.

Adjusting Sensitivity

(1) Confirm that Aspiration Sensor adjustment is completed. (Select


Menu Controller Service Mode. Select Control 3 in
Analytical Mode.)

Target range
DIFF-X: assay value +/- 3ch
DIFF-Y: assay value +/- 2ch
FSC-X: assay value +/- 2ch
MCV: assay value +/- 1 fL
MPV: assay value +/- 1fL
HGB: assay value +/- 0.2 g/dL

(2) Run e-CHECK (level 2).


Note: For American Markets, the X-CAL product is used rather
than e-CHECK level 2 when performing Sensitivity
adjustments.
(3) Select Sensitivity in Service screen. (e-CHECK cannot be run in
the Sensitivity screen.)
(4) Enter each assay value in the Target box and click >>> and
Send.
Enter assay value

Make sure that the DIFF scattergram and FSC histogram are
discrete and displayed normally.

Click >>>

(8) Click Close when the values are within the range.
(9) Repeat step (6) and (7) until the values are within the range.
For adjusting DIFF-Y using human blood, aspirate ten normal blood
samples and Calculate mean value. Enter target value in DIFF-Y and
click Send so that the result value will be in the following target value
range.
Target Value: 43.5 +/- 3.0

Click Send

Pick up the closest sample to the mean value.


(5) Click Close.
(6) Run e-CHECK (level 2) again.

XS Series S/M

4-32

June 5,2008

SECTION 5

SERVICE PROGRAM

5.3.10 Diaphragm Pump .................................................................... 11


5.3.11 Sensitivity ................................................................................ 11

5.1 Entering Service Mode and Skip Setting (Hot Start)............. 1

5.3.12 Laser Setting........................................................................... 13

5.2 MENU Tree .......................................................................... 2

5.3.13 Laser Power............................................................................ 13

5.3

Service Function (for FSR and Production).......................... 3

5.3.1

Piercer/Pipette Adjustment ...................................................... 4

5.3.2

Optical Axis Adjustment(rough)................................................ 5

5.3.3

Optical Axis Adjustment(fine) ................................................... 7

5.3.4

Aspiration Sensor .................................................................... 8

5.3.5

Shipping1................................................................................. 8

5.3.6

Shipping2................................................................................. 9

Rinsing procedure ............................................................................... 9


This procedure is continued from Draining procedure is 5.3.5.......... 9
Shipping1. ........................................................................................... 9
1. Connect distilled water to three tubing on the rear panel. ............... 9

5.3.14 RBC Clogs .............................................................................. 14


5.3.15 Mode ....................................................................................... 15
5.3.16 Service Setting........................................................................ 16
5.3.17 rogram Restart ........................................................................ 17
5.3.18 Piercer/Pipette ........................................................................ 18
5.3.19 SV tests................................................................................... 19
5.3.20 Hand Adjustment..................................................................... 20
5.3.21 Hand ....................................................................................... 22
5.3.22 Tube Holder Adjustment.......................................................... 23
5.3.23 Permanent Counter................................................................. 24

5.4

Service Setting.................................................................... 25

Connect ethanol (not diluted) to tubing for 4DS. ................................. 9

5.5 Calibration Setting .............................................................. 29

5.3.7

Shipping3................................................................................. 9

5.6 HardwareLimit Setting ........................................................ 30

Drain Water and Ethanol ..................................................................... 9

5.7 Service Tool Specification...................................................... 32

This procedure is continued from Rinsing procedure is 5.3.6 ........... 9

5.7.1 Backup ...................................................................................... 32

Shipping2. ........................................................................................... 9

5.7.2 Restore ..................................................................................... 37

1. Disconnect three tubing on the rear panel and tubing for4DS......... 9

5.8 Installer Tool Instruction......................................................... 42

2. Run Shipping 3.(takes about 25min.) .............................................. 9


After completing above step 1 and 2, the instrument can be shipped. 9
5.3.8

Power..................................................................................... 10

5.3.9

Setting Sequence .................................................................. 10

XS seris S/M

June 5, 2008

5.7 Service Tool Specification


1

To shutdown the IPU program is running, following screen appears.


(Figure 5-7-2).
Figure 5-7-2

Setting Value Backup Tool

5.7.1 Backup
Back up XS IPU database and setting file.
(1)

Starting up
Click Start Program XSXS Tools Back up
to run the program
Figure 5-7-1

Program does not run properly when IPU is not installed.


Following screen appears when IPU is not installed.
(Figure 5-7-3)

Figure 5-7-3

XS Series S/M

5-32

Program doesnt run properly when IPU is not located.


Following screen appears when the IPU could not detected IPU ini.
(Figure 5-7-4)

Table 5-7-8 Main Screen Setting Item List


Item
Backup
folder
Selection
...

Figure 5-7-4
Database

(2)Main Screen

Setting Files

Backup for IPU is performed in accordance with the selection of back


up folder and back up option.

Contents
Input with full path the folder to be back up.
Display Folder Selection Dialog. (Figure5-7-6)
The folder selected in the folder selection dialog
will be displayed in the edit box when the dialog
exits.
Back up of the file of IPU database is performed
by checking the check box and click OK.
Default: checked
Back up of the file of IPU setting is performed by
checking the check box and click OK.
Default:checked

June 5, 2008
Figure 5-7-5

Figure 5-7-6
XS Series S/M

5-33

Subnet Mask:

(3) Run
Backup starts by clicking OK in the main screen. While backing up,
all the items in the main turn Disable,
and percentage of completion will be indicated in the progress bar
on the screen.

Default Gateway:

Setting of subnet mask of of network


card is output.
Setting in case of obtaining IP address
automatically: 0.0.0.0 fixed.
Setting of default gate way of network
card is output.
Setting in case of getting IP address
automatically: blank
If there is no specific setting: blank

When back up is completed properly, screen displays Backup


Complete Dialog (Figure5-7-7).

Backup file of the selected item is created in the folder selected in


backup folder selection.The folder name format is as follows.
(date: format is based on IPU setting)_(time: format HHMMSS
fixed)_(IPU version)_BACKUP
Setting folder is created in the back up folder and is backed up IPU
setting file.

Figure 5-7-7

"NetConfig.txt" file is created in back up folder when back up of


setting
file is performed, and the setting of Tcp/IP net work driver is output
according to the format of the table 5-7-11.

If an abnormal occurs while backing up, screen displays Backup


Failure Warning Dialog and backup is interrupted. In that case, backed
up data is deleted.

Table 5-7-11 NetConfig.txt output form


Item
Network Card Name:
Dhcp Enabled:
IP Address:

XS Series S/M

Contents
Name of the network card driver
registered in device manager is output.
DHCP setting of network card is output.
Yes: Obtain an IP address automatically
No: Assign IP address
Setting of IP address of network card is
output.
Setting in case of obtaining IP address
automatically: 0.0.0.0 fixed.

Back up was failed

Figure 5-7-8

5-34

Restore does not run properly when IPU is not installed.


Following screen appears when IPU is not installed.
(Figure 5-7-10)

5.7.2 Restore
Restore XS IPU database and setting file.

(1) Starting up
Select Start Program XSXS tools Restore
Only user administrator right can restore.

Figure 5-7-11

Table 5-7-20 User Authorization Warning Dialog Button List


If the user does not have administration right, restore is not
performed.
Following screen appears (Figure 5-7-12) when the user is not
Log-on as administrators.

Figure 5-7-9
Restore does not run properly when IPU is running.
Following Screen appears when IPU is running (Figure 5-7-10).

Figure 5-7-12

Figure 5-7-10

XS Series S/M

5-35

Setting File

(2) Main Screen


In accordance with the selection of back up folder and backup option,
perform backup for IPU is performed.

Restore is not performed by checking the


check box and click OK.
Default: unchecked
Disable is displayed if back up folder is not
in the selected Restore Folder.

Figure 5-7-13
Table 5-7-23
Item
Restore Folder
Selection

...

Data Base

XS Series S/M

Display Message

Contents
Input with full path the folder to be restore.
Select the folder (Date: format is according
to IPU setting)_(Time: format HHMMSS
fixed.)_(IPU version)_BACKUP.
Display the Folder Selection Dialog( Figure
5-7-14).
Folder selected in the Folder Selection
Dialog will be displayed when the dialog
closes.
Restore of Data Base File of IPU is
performedd by checking the check box.
Default: unchecked
Disable is displayed if back up folder is
not in the selected Restore Folder.

5-36

Figure 5-7-14

(3) Run
Restore starts by clicking OK in the main screen.
While restoring, all the items in the main screen display Disable,
and percentage of completion will be indicated in the progress bar
on the screen in percent figures.
Restore in the database file,
Database folder is in the restore folder.
Setting folder is in the restore folder.
If restore is completed, Restore Completion Dialog (Figure 5-7-15)
is displayed and inform the user about the completion of restore.

Figure 5-7-15

Restore is interrupted even if the log is displayed.


In that case, if the restore is partly completed, restore can not be
canceled.

Restore was failed

Figure 5-7-16

XS Series S/M

5-37

June 5, 2008

5.8 Installer Tool Instruction

Add Database User.exe

Add Database User exe. is used in order to add security account of user
to MSDE(Microsoft SQL Server Desktop Engine).
Log with Sysmex user to Windows as the administrative right is needed
to use this application. Access to XS IPU database is allowed to users
who registered to Windows and has administrative tight or the users who
is added the security account to MSDE. Users who have administrative
right or XS user whose security account is added to MSDE are allowed
to access to the database.
In case a limited user who is added to Windows newly and IPU is started
with this user, use Add Database User. exe and add the user to MSDE
security account.
Dialog (Figure 5-8-1-1) appears when Add Database User.exe starts
Input the desired logon user name for Windows in the edit box of dialog
and click Add User Click Close to close dialog.
IPU can be started with the added user.

AUTO_LOGON_OFF.reg

AUTO_LOGON_OFF.reg is used in order to change settings not to be


automatical logged on when Windows starts.Log in with Sysmex user to
Windows as the administrative right is needed to use this menu.
In XS standard installation, XS user is automatically logged on to
Windows OS when Windows starts. AUTO_LOGON_OFF.reg is
performed for changing settings not to be logged on automatically.

Figure 5-8-2-1 Registry Confirmation Screen


Double click AUTO_LOGON_OFF.reg file and click Yes (Y) on Registry
Confirmation Screen(Figure 5-8-2-1)
After applying this menu, Windows logon screen is displayed without
login on automatically when Windows starts.

Figure 5-8-1-1 Add Database User Dialog

XS Series S/M

5-38

June5, 2008

AUTO_LOGON_ON.reg

AUTO_LOGON_ON.reg is used in order to change settings to logg on


automatically when Windows starts.Log on with Sysmex user to
Windows OS as the admisitrative right is needed to use this menu.
Change settings to be logged on automatically with log on name XS
in case cancelling automatical log on by using AUTO_LOGON_OFF.reg.

Browser Layout Selector.exe

Browser Layout Selector.exe is used to change browser layout displayed


in XS IPU.Log in with Sysmex user to Windows OS as the administrative
right is needed to use this application.
Run this application program when IPU is not running.
In XS standard install, sample information/patient information display is
drastically changed from XT.
Sample information/ patient information displayed in the upper browser
can be changed like XT by using Browser Layout Selector.exe

Figure 5-8-3-1 Registry Confirmation Screen


Double click AUTO_LOGON_ON.reg and click Yes(Y) on Registry
Confirmation Screen (Figure 5-8-3-1).
Whindows automatically starts without displaying logon screen.
Figure 5-8-4-1 Browser Layout Dialog
Browser Layout Dialog appears when Browser Layout Selector.exe is
started.
Figure 5-8-4-1 is started if Browser Layout Selector.exe is started.
Select New layout to XS standard layout, or select XT compatible
layout to use XT compatible layout.
From next IPU program starting, browser screen will be displayed with
selected browser layout.

XS Series S/M

5-39

June5, 2008

Create Blank Database.exe

Create Blank Database.exe is used in order to initialize XS IPU


Database.Log in with Sysmex user to Windows OS as the administrative
right is needed to run this application.Run this application when IPU is
not running.
Once this application is initialized, all the data such as stored data and
QC data will be deleted. Therefore, run this application carefully.
IPU database is restructured and formatted by running.

Figure 5-8-5-2 Execute Confimation Dialog


Dialg (Figure5-8-5-2) appear and click OK to run.
XS IPU database will be initialized, and after initialization, the dialog is
automatically closed.
Click Cancel to cancel. The screen will be back to Create Blank
Database Dialog.

Figure 5-8-5-1 Blank Database Dialog


Dialog (Figure 5-8-5-1) is appears when Create Blank Database. exe is
started. Click Execute to create blank database.
Click Cancel to cancel, Create Blank Database Dialog will close
without running anything.

XS Series S/M

5-40

June 5, 2008

DBIndex.exe

DBIndex.exe is used in order to renew index key of IPU database for


Ver.00-06 and lower.Log in with Sysmex user to Windows as the
administrative right is needed to run this application.
To improve the speed of processing, index key of IPU database is
changed while upgrading from Ver.00-06 to Ver.00-07.
When the IPU database used on Ver.00-06 or lower is moved to
Ver.00-07, and higher, renewing index key executed when updating
versions is not executed. By using this application, index key of database
used on Ver.00-06 and lower is renewed for using on Version.00-07 or
higher.

Figure 5-8-6-1 DBIndex Dialog


Dialog (Figure5-8-6-1) appears only in case renewing index key is
needed,
after DBIndex. exe starts and current index key is checked.
After completing the application, dialog is automatically closed.

XS Series S/M

5-41

DistributionPrintImageEditor.exe

Distribution Print Image Editor.exe is used in order to change setting of


the color of graph or normal area on GP printing.
Log in with Sysmex user to Windows OS as the administrative right is
needed to run this application.
Run when IPU is not started.

Figure 5-8-7-2 Setting Color Dialog

Figure 5-8-7-1 Distribution Print Image Editor Dialog


When Distribution Print Image Editor.exe is started, current IPU
Distribution Print Image setting is displayed on the left side of the dialog.
Click Distribution Color to change the print setting of the graph.
Click Inside to change the color of the normal color area.

XS Series S/M

5-42

Sample display in the print image display will be renewed soon after
the setting of the color is changed.
Standard setting of XS will be read by clicking Load Default
Click OK to close dialog after setting is completed.
Click Cancel to calcel the changed setting and close the dialog.
Changed setting is reflected from next IPU starting and changed
Distribution Print Image is printed on program GP printing.

June5, 2008

Host Setting.exe

Option (Serial)
Host Format: XS, XT, RAM-1, SE-9000, XE, XE-1500, SF-3000, K-1000,
K-4500,
Format Type: Default, Flg
Date Format: YYYYMMDD
Item not in use: Updating, Space Padding

Host Setting.exe is used in order to change host format for host output
with XS IPU. (changing to ASTM format is performed in IPU setting.)
Log in with Sysmex user to Windows as the administrative right is
needed to run this application.
Host format option is displayed while XS IPU installation. Run Host
Setting.exe to change host format after IPU installation.

(Tcp / Ip)
Host Format: XS, XT
Format Type: Default, Flg, For DPS
Date Format: YYYYMMDD
Item not in use: 0 padding, Space Padding
Figure 5-8-8-1 Host Setting Dialog is displayed when Host Setting.exe
starts run.
Select the format to be output in section Tcp /IP and click OK.
Click Cancel to cancel the setting.
Changed setting is reflected from next IPU star-up.

Figure 5-8-8-1 Host Setting Dialog

XS Series S/M

5-43

ID Changer.exe
ID Changer.exe is used to change instrument ID set in XS IPU.
Log in with Sysmex user to Windows OS as the administrative right is
needed to run this application.
Run this application when IPU is not running.
Run the program carefully. If you make mistake in ID change, analysis
is not be able to be performed.

Figure 5-8-9-1 ID Change Dialog


WhenID Changer.exestarts, IPU main unit setting appears om the
dialog.Select the model from (XS-1000i / XS-800i) and input the ID to
New IDand check Sample Option if the sampler is connected.
Uncheck Sample Option when the sampler is not connected and click
OK.
Click Cancel to cancel the setting.

XS Series S/M

5-44

Changed setting is reflected from next IPU start-up and unit connection
is performed.

10

IPU Remover.exe

IPU Remover.exe is used to delete XS IPU program from the computer.


Log in with Sysmex user to Windows OS as the administrative right is
needed.
Run this application when IPU is not running.

Figure 5-8-10-1 IPU Remover Dialog


IPU Remover Dialog (Figure 5-8-9-1) is displayed when "IPU
Remover.exe starts. Check Remove Database Files check box to
delete IPU database. Remove Database Files check box to unmark.
when you do not want to delete IPU database.
Click Cancel to cancel.
IPU Remover Dialog is finished without performing anything.

Figure 5-8-11-1 Execute Confirmation Dialog

XS Series S/M

5-45

Execute Confirmation Dialog (Figure5-8-11-1) is displayed and click


Yes(Y) to execute.
IPU is deleted from the computer.
Click No (N) to cancel.
Screen is back to IPU Remover Dialog

11

Language Changer.exe

12

Language Changer.exe is performed to change XS IPU display


language.Log in with Sysmex user to Windows OS as the administrative
right is needed to run.

MSDE Remover.exe

MSDE Remover.exe is used to delete MSDE (Microsoft SQL Server


Desktop Engine).Log in with Sysmex user to Windows OS as the
administrative right is needed to run this application.
Run when IPU is not running.

Figure 5-8-12-1 MSDE Remover Dialog


MSDE Remover Dialog (Figure 5-8-12-1)appears when MSDE
Remover.exe. is started. Click OK to delete MSDE.
Click Cancel to cancel.
Figure 5-8-11-1 Language Changer Dialog
Language Changer(Figure 5-8-11-1) will show when current IPU
language setting. Select the language to be displayed on IPU and click
"OK .
Click Cancel to cancel.
Changed setting is reflected from next IPU start-up.

Figure 5-8-12-2

Execute Confirmation Dialog

When Execute Confirmation Dialog (Figure 5-8-12-2) is displayed click


Yes(Y) to execute. MSDE is deleted from the computer.
Click No (N) to cancel and screen is back to MSDE Remover Dialog.

XS Series S/M

5-46

June 5, 2008

6.3.5
6.3.5.1

2)
3)
4)

WB Aspiration Errors
Short Sample

Blood Aspiration Sensor

Description:

During sample aspiration, specified amount of sample


cannot be aspirated.
Purpose:
To ensure the amount of sample aspiration.
Method:
Before RBC sample dispensing, top end discard portion
of aspirated whole blood sample is dispensed into RBC/
HGB Chamber to dilute, and then charged into Blood
Aspiration Sensor to convert (detect) turbidity (value of
Sample Convert 1 of Blood Aspiration Sensor).
Check the difference between the results (Sample
Convert 1.) and the value of blank convert value
A=Sample Convert 1 of Blood Aspiration Sensor-Blank
Convert Value
B=Sample Convert 2 of Blood Aspiration Sensor minus
Blank
Convert Value
C=A/(A+B)*100
If A is 200 or C is 5 during operation only, system
assumes to be an error.
XSs Action:
Status: Ready Data Mask: Yes
Sampler Operation:
Depends on the setting of sampler stop condition.
Alarm Sound: Yes (only sampler analysis stop setting)
Action Message: Execute Shutdown and Restart to clean piercer.
Cause:
1) Abnormality of sample blood (volume, coagulation,
low concentration, etc.)
2) Clogging in following parts
Piercer (XS-1000i) / Pipette (XS-800i)
WB Aspiration Line Tube
3) Disconnection of WB Aspiration Line Tube
4) Defect in blood aspiration sensor
Action to be taken:
1) Check sample
(volume, coagulation, low concentration, etc.)
XS Series S/M

Rinse or replace corresponding parts


Check tube connections
Replace or adjust Blood Sensor

Figure 6-22 Blood Aspiration Sensor Check - 1

NOTE:
In the figure above shows the state of inside XS-1000i piercer
needle just after blood aspiration. XS-800i sample probe is
also monitored in the same manner.

6-20

June 5,2008

6.3.5.2

Sample Not Asp Error

Description:

Accuracy of analysis data can not be assured, as bubble


might get into due to sample shortage during WB
aspiration.
Purpose:
During WB aspiration, system checks if the top end
discard portion in the tip of piercer (pipette) is filled with
whole blood and sufficient blood volume (20L) is
secured.
Method:
After WBC sample draining, dilute tail end discard
portion and convert (detect) the turbidity with Aspiration
Sensor (value of Sample Convert 2 of Blood Aspiration
Sensor).
Compare this value with the value of blank convert to
check the difference.
If the difference between Sample Convert 2 value and
Blank Convert value is below 1700, it is considered as
an error. However, during blank checking and capillary
mode, the values are not monitored.
XSs Action:
Status: Ready
Data Mask: Yes
Sampler Operation: Follows the setting of sampler stop
condition.
Alarm Sound: Yes (only sampler analysis stop setting)
Action Message: Execute Shutdown and Restart to clean piercer.
Cause:
1) Bubble exists in Blood Aspiration Sensor Line.
2) Defect in Blood Aspiration Sensor.
Action to be taken:
1) Run Auto Rinse, check breaking and leakage in Blood
Aspiration Line.
2) Replace or adjust Blood Aspiration Sensor.

XS Series S/M

Blood Aspiration
Sensor

Figure 6-23 Blood Aspiration Sensor Check - 2

NOTE::
In the figure above shows the state of inside XS-1000i piercer
needle just after blood aspiration. XS-800i sample probe is also
monitored in the same manner.

6-21

June 5, 2008

6.3.5.3

Blood Asp Sensor Error

Description:
Purpose:
Method:

The blood aspiration sensor is not operated correctly.


To ensure the operation of Blood Aspiration Sensor
When adjusting the auto offset of Blood Aspiration
Sensor, it is considered as an error if it cannot be
adjusted within the range of adjustment.
Only during operation: Timing at the start of WB
aspiration
Blank Convert Value: < 50
Sample Convert 2 Value: > 15000
Sample Convert 2 Value Blank Convert Value:
< -50
Not monitored during capillary mode.
XSs Action:
Status: Not Ready
Data Mask: No
Sampler Operation: Interrupted Alarm Sound: Yes
Action Message:
There is a problem in the blood aspiration sensor.
Contact your Sysmex Service Representative for
assistance.
As a tentative measure, change the
setting of blood aspiration sensor to continue
measurements until the Sysmex Representative arrives.
Cause:
Dirt or Defectiveness in Blood Aspiration Sensor
Action to be taken:
Clean, replace or adjust Blood Aspiration Sensor
NOTEWhen adjusting blood aspiration sensor, change the unit to g/dL
and adjust it.

XS Series S/M

6-22

June 5, 2008

XS Series S/M

XS Series Tubing Diagram

June 5, 2008

Appendix C
C.1

SNCS

Preparations for Installation of Questra Service Agent .... 1

C.1.1

Installing the Certificate file................................................. 1

C.1.2

Confirming your environment.............................................. 7

C.1.2

Confirming your environment.............................................. 7

C.2

Installation of Questra Service Agent............................... 8

C.3

Setting for Applications .................................................. 15

C.3.1

Update Questra setting file ............................................... 15

C.3.2

Confirming Questra Operation.......................................... 18

C.3.3

Installation of UltraVNC .................................................... 19

C.3.4

Setting of INI file ............................................................... 22

C.4

Operation Manual for Remote Access ........................... 23

C4.1

C.5

Operation Procedure ......................................................... 23

Procedure when not using Questra ............................... 28

C.5.1

Setting for Outlook Express .............................................. 28

C.5.2

Editing SmtpSend.ini (File Transfer Program) .................. 28

C.5.3

Setting of INI file ............................................................... 29

C.5.3

Setting of INI file ............................................................... 30

C.5.4

Synchronizing the PC time ............................................... 30

XS-series S/M

Note: This Appendix C SNCS section is not used in the American


markets.

June 5, 2008

C.5.3

Setting of INI file

(1) Open C:\ProgramFiles\sysmex\IPU\ORG_INI\IPU.ini.


(2) Locate OnlineQC=OFF in the [SNCS] tag (bottom in the file) and
change it to OnlineQC=ON.
NOTE:
When revising INI file, use only capital letters, ON or
OFF. If you use lower-case letters, On, on, Off
or off, the changed setting will not work properly.
(3) Change the RemoteMainte=OFF to RemoteMainte=ON.
(4) Change the VNC=OFF to VNC=ON.
(5) Save the file with overwriting it.

C.5.4

Synchronizing the PC time

(1) Run [Date and Time] on the Control Panel.

(2)Click [Internet Time].


If the computer is the member of the domain, [Internet Tab] can not be
used.

(3) Uncheck [Automatically synchronize with an Internet time server].

XS Series S/M

C-30