Figure 1 is an actual display from the QUALILAB Plating Bath Analyzer. The
potential determines the electrochemical reaction that occurs. The potential is displayed on the X-axis with potentials becoming more positive from left to right.
A positive-going potential is becoming more strongly oxidizing, while a negative going potential generates a more strongly reducing condition. Current is shown on
the Y-axis. A positive current corresponds to an oxidation while a negative current is due to a reduction. The voltammogram has several regions of interest:
Plating region: Points 1 to 2 and 2 to 3. The current is negative in this region as the platinum electrode is plated with metal. Since the plating time is well-defined
by the scan rate, the plating rate can be accurately estimated as the amount of metal that is plated in the plating region.
Stripping Region: Points 3 to 4. As the current of the electrode becomes positive, the metal is stripped off the electrode. The area of the stripping
peak represents the charge required to strip the deposited metal. The instrument integrates the current to measure the area beneath the stripping peak. The
area is reported in milli-Coulombs (mC) and is shown in the box at the lower right of Figure 1. Since the quantity of stripped metal is the same as quantity of
plated metal, the stripping peak area is proportional to the amount of metal plated. It is more accurate to measure the amount of metal plated by integrating the
current in the Stripping Region rather that in the Plating Region. Electrochemical processes other than metal deposition, such as hydrogen reduction, may also
be taking place in the Plating Region.
Cleaning Region: Points 4 to 5 and back from 5 to 6. A positive potential on the electrode is used to desorb organics and clean the surface of the
platinum electrode. This cleaning segment is critical for accurate and reproducible results with CVS. This region is also used to monitor oxidizable organic
contaminants and chloride levels in acid copper baths.
Adsorption Region: Points 5 to 6. In this region the organic additives are adsorbed onto the electrode surface, even though no metal is plated. The adsorption
that takes place in this region enhances the sensitivity of CVS.
The influence of several organic additives on the plating rate is shown in Figure 2.
In general, additives such as levelers or wetters act as suppressors and decrease the plating rate while brighteners increase the rate of deposition. Use caution,
however, since there is no standard terminology for additives within the industry.
During electroplating, the organic additives undergo various chemical transformations such as oxidation, reduction, fragmentation, and polymerization. Most of the
products of such transformations also affect the plating process. CVS provides a measure of the effective concentration of additive (the activity of the additive in
terms of the fresh additive) which is more important than the concentration of a specific substance. Most analytical techniques (spectroscopy, HPLC) measure
absolute concentration. The transformed additive will still be electroactive, but not responsive to spectroscopy, or vice versa. Because CVS is based on the
2 of 4
6/19/2012 9:40 PM
ECI Technology | World Leading Chemical Process Control | The Fundame... http://www.ecitechnology.com/articles/fundamentals-cyclic-voltammetric-...
plating process, it is the only technique that provides a true measure of the activity of the organic additives.
DETERMINATION OF ADDITIVES AND THEIR COMPONENTS
The quantitative determination of organic additives is based on the CVS response of the production bath before and after the addition of a known amount of a
standard. The fresh additive as supplied by the manufacturer, which is used for replenishment of the bath, is used as the standard. Since the fresh additive is
used as the standard, the concentration of additive in the production bath is expressed as the equivalent activity of the fresh additive. In this way, replenishment
of the bath to the proper additive level is very convenient.
It is usually possible to quantify all of the additives in the plating bath. An example of an acid copper bath is shown in Figure 3.
The carrier acts as a suppressor and inhibits the rate of the plating process. This can be seen from the reduction in peak area as the concentration of the carrier
is increased. Notice that the activity of the carrier is very high, i.e., a small increase in concentration has a large effect on the rate of plating. The fresh carrier
as supplied by the manufacturer was used to spike the solution.
The plating rate changes dramatically as the carrier is increased from 0-1 mL/L. As the carrier concentration is increased beyond 1 mL/L, there is little additional
effect on the plating rate.
The brightener has a different effect. In the presence of carrier, higher brightener concentration causes an increase in plating rate. The activity of the brightener is
lower than the carrier. The difference in activity between the carrier and the brightener can be exploited to quantify both components.
Some manufacturers offer a pre-mixed solution containing both the carrier and the brightener. These components can be clearly distinguished in Figure 3. CVS
can be employed to insure that the carrier and the brightener are in balance in the production bath.
The analysis of plating baths using CVS benefits from experience. ECI Technology has invested over 15 years in developing CVS methods for a wide variety of
plating baths. In addition to manufacturing CVS instrumentation, ECI Technology is the leading source of CVS applications information. CVS consultation is
available to every user of ECI Technology products.
OUTGOING AND INCOMING INSPECTION OF ADDITIVES
CVS is a powerful tool for the outgoing inspection of additives by suppliers as well as incoming inspection at users sites. Figure 4 illustrates the
variations between several lots of the same brightener.
By performing a fast, easy assay, the user avoided under dosing or overdosing the brightener and losing control of the bath.