Академический Документы
Профессиональный Документы
Культура Документы
Must be clean
The
Water
Supply
Not be mixed
chemistry with
water that has a
very high chlorine
content.
Must be reliable
Water
Flow
Test
Takes only 15 minute
to make four changes
in one hour.
Cleaning the
Tanks
At Start Up ( Daily )
Feed a 35 x 43 cm waste unprocessed film into the processor to clean any deposit in rollers.
Ti may be necessary to feed in a second film if you feel it is necessary.
Do not use processed film as this will be too hard and may contain redisual fixer.
Do not proceed with the daily workload until you are satisfied to rollers you clean.
At Shut Down
Remove or clean all crossover rollers or plates.
Wipe off chemicals deposits on or around the processor. take special care with film feedtray
Developer
It affects the density, contrast and amount of base fog of a radiograph.
Fixer
Developer/Fixer Tank
Evaluation:
a.
b.
c.
d.
e.
Place
hydrometer in
fixer tank
Note the
reading at the
surface of the
fixer
A high specific
garavity reading
means not
dilute enough.
A low specific
gravity reading
means too
dilute
1. Film Hangers.
Larger hanger sizes have loop guides to hold the bow spring and top clips in
line and maintain proper tension
Be sure that the hangers dont have any damage
Must be quality strong stainless steel construction that will last for year
Report if damage is spotted.
Reference: http://www.tedndt.com/cat/cat45.php
Reference: http://www.avishkarexport.com/x-ray-equipments.html
Developer is 64 - 66 = 7 minutes
Depending on
temperature of
developer
Reference: http://www.state.nj.us/dep/rpp/qa/qa_down/radcg.pdf
3. Darkroom Thermometer
Radiographic
Darkroom
Thermometer
Reference:
http://www.state.nj.us/dep/rpp/qa/qa_down/radcg.pdf
Reference:
http://webstag.kodak.cz/US/en/business/aim/industrial/ndt/literature/radiog
raphy/10.shtml