Envirotech Instruments Pvt.Ltd.
*High volume sampler is based on the simple design standardized by USEPA have
been in use all over the world for monitoring of TSP in ambient air.
*Health effects are caused primarily by respirable particulates in the size
range of 0.5 to 10.
*Envirotech developed Indigenous system for air quality monitoring first time
in the country, A Respirable Dust High Volume Sampler in the year 1994. The
system designated Envirotech APM451
* The APM 451 SAMPLER FIRST SEPERATES THE COURSER partcles(larger than 10
)from the air stream before filtering it on glass micro fiber filter allowing
measurement of both the TSP and the respirable fraction of the suspended
particulate matter simultaneously.
*In 1997 envirotech upgraded the APM451 sampler. The new system designated as
Envirotech APM 460 includes the following improvement.
1 The air inlet pipe and cyclone design have been modified to provide a
sharper cut-off.
2.digital timer replace with motorized timer
3. the manometer design and placement have been modified
4.the PCB design of the time totalizer circuit have been improved to reduce
field failure.
*In 460 BL Blower has been replaced by indigenous brushless blower.
* by this change not only noise generation is reduced while operation
also cost of consumbles and maintenance efforts have also been reduced
but
The
sampling rate which has been found varied with dust loading on the
filter get automatically corrected by control on input voltage frequency
through of a 3 phase drive coupled with flow controller. In 460 DXNL thus 3
phase regenerative blower is fitted however machine runs on single phase
220 v input voltage only. Single phase input voltage is first converted in
3 phase voltage with the help of driver fitted in the machine. The size
fractionating cyclone of the APM 460DXNL thus operates close to its
designed separation velocity resulting in a sharper cut-off of particles
larget than 10 . To compensate for a variation in filter characteristics
the flow controller has a provision to adjust the sampling flow rate, while
setting up the sampler, and the system will automatically maintain this
present flow rate.
DXNL
is
thus
unaffected
by
voltage