com - 800-732-3457
DC Power Supply
GPD-3303 Series
USER MANUAL
GW INSTEK PART NO.
Table of Contents
28H
29H
FAQ ................................................................................ 4 5
OVERVIEW....................................................................... 9
Introduction........................................... 9
Series Lineup / Main Features ............. 11
Principle of Operation.......................... 12
Front Panel Overview ........................... 14
Rear Panel Overview ............................ 17
CV/CC Crossover Characteristics ......... 18
APPENDIX ...................................................................... 4 6
Fuse Replacement ................................ 4 6
Specifications ...................................... 4 7
Declaration of Conformity .................... 4 9
30H
31H
32H
3H
34H
INDEX ............................................................................ 5 0
35H
SETUP ........................................................................... 19
Power Up ............................................. 19
Load Cable Connection ........................ 20
Output On/Off..................................... 21
Beep On/Off ........................................ 21
Front Panel Lock .................................. 22
OPERATION .................................................................. 23
CH1/CH2 Independent Mode .............. 23
CH3 Independent Mode ....................... 25
CH1/CH2 Tracking Series Mode........... 27
CH1/CH2 Tracking Parallel Mode ........ 3 2
19H
21H
2H
24H
25H
26H
Safety Guidelines
SAFETY INSTRUCTIONS
General
Guidelines
CAUTION
Safety Symbols
These safety symbols may appear in this manual or on the GPD-3303
series.
WARNING
CAUTION
WARNING
Storage
environment
Operation
Environment
Temperature: 0C to 40C
Location: Indoor
When using the GPD-3303 series in the United Kingdom, make sure
the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
Cleaning the
Disconnect the power cord before cleaning.
GPD-3303 series
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
As the colours of the wires in main leads may not correspond with the colours
marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with the letter E or by the earth symbol or coloured Green or
Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked
with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the
supplier.
This cable/appliance should be protected by a suitably rated and approved HBC
mains fuse: refer to the rating information on the equipment and/or user
instructions for details. As a guide, cable of 0.75mm2 should be protected by a 3A
or 5A fuse. Larger conductors would normally require 13A types, depending on
the connection method used.
Any moulded mains connector that requires removal /replacement must be
destroyed by removal of any fuse & fuse carrier and disposed of immediately, as
a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring
must be carried out in accordance with the information detailed on this label.
OVERVIEW
This chapter describes the GPD-3303 series in a
nutshell, including its main features and front /
rear panel introduction. After going through the
overview, follow the Setup chapter (page19) to
properly power up and set operation environment.
Introduction
Overview
Independent /
The three output modes of GPD-3303 series,
Tracking Series / independent, tracking series, and tracking parallel,
Tracking Parallel can be selected through pressing the TRACKING
key on the front panel. In the independent mode,
the output voltage and current of each channel are
controlled separately. The isolation degree, from
output terminal to chassis or from output terminal
to output terminal, is 300V. In the tracking modes,
both the CH1 and CH2 outputs are automatically
connected in series or parallel; no need to connect
output leads. In the series mode, the output
voltage is doubled; in the parallel mode, the
output current is doubled.
9
Dynamic load
10
Principle of Operation
Overview
Series Lineup
AC input circuit
Model
Tracking Error
Transformer
GPD-3303D
3 digit
3 digit
Yes
GPD-3303S
5 digit
4 digit
Yes
Main Features
Performance
Operation
Protection
Interface
Software calibration
Buzzer output
Overload protection
11
12
Main Rectifier
Current Limiter
Overload
Display
VoltMeter
AmpMeter
13
14
GPD-3303D (3 digits)
GPD-3303D (3 digits)
Control Panel
Terminals
Memory Keys
CH1/CH2/Beep
Keys
Parallel/Series
Keys
Lock Key
Default Terminals
European Terminals
GND Terminal
CH1 Output
CH1 CV/CC
Indicator
CH2 Output
CH2
CV/CC/PAR
Indicator
CH3 Output
CH3 Overload
Indicator
CH3 Voltage
Selector
42H
Output Key
Voltage Knobs
Current Knobs
Power Switch
15
16
USB Connector
Power Cord /
Fuse Socket
Cooling Fan
Power Cord /
Fuse Socket AC Selector
Background
CV mode
CC mode
Diagram
Vmax
Constant
Voltage
Constant
Current
17
Vout
Imax
18
Iout
SETUP
GTL-104
Power Up
Select AC voltage Before powering up the power
supply, select the AC input
voltage from the rear panel.
GTL-105
Connect AC
power cord
GTL-203, 204
Power On
Wire type
Power Off
19
20
20
2.5
18
16
14
10
12
16
Output On/Off
Panel operation
Panel operation
Note
Beep On/Off
Panel operation
List of beep
Power on
Output on/off
21
22
OPERATION
CH1/CH2 Independent Mode
Background /
Connection
Output rating
Panel operation
23
24
CV CC
Output rating
2.5V/3.3V/5V, 3A fixed
No Tracking
Series/Parallel
Panel operation
25
26
Connection
Output rating
0 ~ 60V/0 ~ 3A
28
Voltage level
Current level
Output rating
30
Master (CH1)
voltage level
Master (CH1)
current level
Background /
Connection
Output rating
0 ~ 30V/0 ~ 6A
Slave (CH2)
voltage level
Slave (CH2)
current level
SAVE/RECALL SETUP
Save Setup
Background
Contents
Voltage level
Current level
33
Output on/off
Panel operation
Note
34
Recall Setup
Background
Contents
Background
Panel operation
Note
REMOTE CONTROL
Output on/off
Interface
COM setting
Functionality
check
Data bit: 8
Stop bit: 1
35
36
Command Syntax
Entering the
remote control
mode
Command format
1: command header
2: output channel
3: separator
4: parameter
Output channel
1 (CH1) or 2 (CH2)
Parameter
Type
Description
Example
<Boolean>
boolean logic
0 (off), 1 (on)
<NR1>
integers
0, 1, 2, 3
<NR2>
Note
Error Messages
Leaving the
remote control
mode
Message contents
Descriptions
Program mnemonic
too long
Invalid character
Missing parameter
Undefined header
37
38
Command List
Command Details
The HELP command shows all the below commands and their
meanings, except for the HELP command itself.
ISET<X>:<NR2>
ISET<X>:<NR2>
Description
ISET<X>?
Panel operation
See page23
VSET<X>:<NR2>
Response time
Minimum 70ms
VSET<X>?
Example
ISET1:2.234
IOUT<X>?
VOUT<X>?
TRACK<NR1>
BEEP<BOOLEAN>
OUT<BOOLEAN>
STATUS?
*IDN?
RCL<NR1>
SAV<NR1>
HELP?
VSET<X>:<NR2>
ERR?
Description
Panel operation
See page23
Response time
Minimum 70ms
Example
VSET1:20.345
ISET1:2.23
ISET<X>?
Description
Panel operation
See page23
Response time
Minimum 80ms
Example
ISET1?
VSET1:20.3
39
40
VSET<X>?
BEEP<Boolean>
Description
Description
Response time
Minimum 80ms
Panel operation
See page21
Example
VSET1?
Response time
Minimum 70ms
Example
BEEP1
IOUT<X>?
Description
OUT<Boolean>
Response time
Minimum 80ms
Description
Example
IOUT1?
Panel operation
See page21
Response time
Minimum 70ms
Example
OUT1
VOUT<X>?
Description
Panel operation
See page23
STATUS?
Response time
Minimum 80ms
Description
Example
VOUT1?
Response time
Minimum 400ms
Contents
TRACK<NR1>
Description
Panel operation
See page27
NR1
0: Independent
1: Tracking series
2: Tracking parallel
Response time
Minimum 70ms
Example
TRACK0
41
42
*IDN?
Description
Response time
Minimum 300ms
Contents
RCL<NR1>
Description
Panel operation
See page35
NR1
1 4: Memory 1 to 4
Response time
Minimum 70ms
bit0:(CH1)0=CC,1=CV;bit1:(CH2)0=CC,1=CV;bit23=(TRACK)01=INDEP,1
1=SER,10=PARA;bit4:(BEEP)0=OFF,1=ON;bit6:(OUT)0=OFF,1=ON;
Example
RCL1
SAV<NR1>
NR0:1=Memory1,2=Memory2,3=Memory3,4=Memory4;
Description
Panel operation
See page34
NR1
1 4: Memory 1 to 4
Response time
Minimum 70ms
Example
SAV1
ERR?
HELP?
Description
Response time
Minimum 1000ms
43
Description
Response time
Minimum 70ms
Contents
44
FAQ
APPENDIX
Q1. I pressed the panel lock key but the output still turns on/off.
Fuse Replacement
23B
A1. The output key is not affected by the panel lock key operation,
for ensuring safety.
Steps
Rating
45
46
100V/120V:T6.3A/250V
220V/230V:T3.15A/250V
Specifications
24B
Voltage
Regulation
Current
Regulation
CH1/CH2
Independent
CH1/CH2
Series
CH1/CH2
Parallel
CH3
Line
Load
0 ~ 30V / 0 ~ 3A
A Meter
GPD-3303S
V Meter
0 ~ 60V / 0 ~ 3A
Program
Accuracy
0 ~ 30V / 0 ~ 6A
GPD-3303D
GPD-3303D
GPD-3303S
GPD-3303D
GPD-3303S
2.5V/3.3V/5.0V, 3A
0.01% + 3mV
0.01% + 3mV (rating current 3A)
0.02% + 5mV (rating current > 3A)
Ripple & Noise 1mVrms (5Hz ~ 1MHz)
Recovery Time 100s (50% load change, minimum
load 0.5A)
Temperature 300ppm/C
Coefficient
Line
0.2% + 3mA
Load
0.2% + 3mA
Ripple & Noise 3mArms
CH3 Specification Regulation
Line 5mV
Load 15mV
Ripple & Noise 2mVrms
Tracking
Tracking Error 0.5%+10mV of Master (GPD-3303S)
Operation
0.5%+50mV of Master (GPD-3303D)
Parallel
Line: 0.01% + 3mV
Regulation
Load: 0.01% + 3mV
(rating current 3A)
Load: 0.02% + 5mV
(rating current > 3A)
Series
Line: 0.01% + 5mV
Regulation
Load: 300mV
Meter Resolution GPD-3303D
Voltage: 100mV
Current: 10mA
47
Readback
Accuracy
GPD-3303D
GPD-3303S
Insulation
Operation
Environment
Storage
Environment
Power Source
Accessories
Dimensions
Weight
Voltage: 1mV
Current: 1mA
3.2A full scale, 3 digits 0.5" LED
display
3.2A full scale, 4 digits 0.4" LED
display
32V full scale, 3 digits 0.5" LED display
32V full scale, 5 digits 0.4" LED display
(0.5% of reading + 2digits)
(0.5% of reading + 2digits)
(0.03% of reading + 10mV)
(0.3% of reading + 10mA)
(0.5% of reading + 2digits)
(0.5% of reading + 2digits)
(0.03% of reading + 10mV)
(0.3% of reading + 10mA)
20M or above (DC 500V)
Chassis and
Terminal
Chassis and
30M or above (DC 500V)
AC cord
Indoor use, Altitude: 2000m
Ambient temperature 0 ~ 40C
Relative humidity 80%
Installation category: II, Pollution degree: 2
Ambient temperature 10 ~ 70C
Relative humidity 70%
AC 100V/120V/220V/230V10%, 50/60Hz
User manual x1
Test lead GTL-104 x 2, GTL-105 x 1
(Europe) Test lead GTL-203 x 1, GTL-204 x 2
210 (W) x 130 (H) x 265 (D) mm
Approx. 7kg
Options
48B
USB cable
48
GTL-246
Declaration of Conformity
25B
We
GOOD WILL INSTRUMENT CO., LTD.
(1) No.7-1, Jhongsing Rd., Tucheng City, Taipei County, Taiwan
(2) No. 69, Lu San Road, Suzhou City (Xin Qu), Jiangsu Sheng, China
declare, that the below mentioned product
Type of Product: Power Supply
Model Number: GPD-3303D/GPD-3303S
are herewith confirmed to comply with the requirements set out in the
Council Directive on the Approximation of the Law of Member States
relating to Electromagnetic Compatibility (2004/108/EC, 89/336/EEC,
92/31/EEC, 93/68/EEC) and Low Voltage Directive (73/23/EEC,
93/68/EEC).
For the evaluation regarding the Electromagnetic Compatibility and
Low Voltage Directive, the following standards were applied:
EMC
EN 61326-1: 2006 Electrical equipment for measurement, control and
laboratory use EMC requirements
Conducted Emission
Radiated Emission
Electrostatic Discharge
ClassA EN 61000-4-2: 1995 + A1:1998 +
A2:2001
EN 55011: 1998 + A1:1999 +
Radiated Immunity
A2:2002
EN 61000-4-3: 2002 + A1:2002
Current Harmonics
Electrical Fast Transients
EN 61000-3-2: 2000 + A2:2005
EN 61000-4-4: 2004
Voltage Fluctuations
Surge Immunity
EN 61000-3-3: 1995 + A1:2001
EN 61000-4-5: 1995 + A1:2001
------------------------Conducted Susceptibility
EN 61000-4-6: 1996 + A1:2001
------------------------Power Frequency Magnetic Field
EN 61000-4-8: 1993 + A1:2001
------------------------Voltage Dip/ Interruption
EN 61000-4-11: 2004
Safety
Low Voltage Equipment Directive 73/23/EEC, 93/68/EEC
Safety Requirements
IEC/EN 61010-1: 2001
49
INDEX
automatic out off ..................21
banana plug...........................20
beep setting
contents..............................21
manual ...............................21
remote control...................42
caution symbol .......................5
cleaning the instrument.........7
COM setting, remote control
............................................36
command list ........................39
recalling the list.................43
common terminal, tracking
series...................................27
cooling fan .............................17
safety instruction ................6
CV/CC
CH1/CH2 indicator .........24
CH3 indicator....................26
operation theory .........10, 18
EN61010
declaration of conformity 49
measurement category ......6
pollution degree .................7
EN61326-1 .............................49
environment
operation..............................7
specification ......................48
storage..................................8
error messages, remote
control ..........................38, 44
front panel
50
lock (manual).................... 22
overview ........................... 14
fuse
rating ................................. 46
replacing ........................... 46
safety instruction ............... 6
GPD series
block diagram................... 12
dynamic load.................... 10
lineup................................. 11
list of features ................... 11
operation theory............... 12
technology overview ......... 9
ground symbol ....................... 5
identification information... 43
load connection .................... 20
operation mode
independent...................... 23
remote control .................. 41
specifications .................... 47
tracking parallel ............... 32
tracking series................... 27
output current setting
manual............................... 24
remote control .................. 40
output on/off
FAQ ................................... 45
manual............................... 21
remote................................ 42
output voltage setting
manual............................... 24
remote control .................. 40
interface............................. 36
save settings
manual............................... 34
remote................................ 43
service operation
about disassembly ............. 6
contact ............................... 45
status, instrument ................ 42
tracking mode
operation theory................. 9
UK power cord....................... 8
USB interface ........................ 36
warning symbol ..................... 5
wire, load .............................. 20
51