Академический Документы
Профессиональный Документы
Культура Документы
1) Make sure that the following errors / alarms are not present before
running the T1 test.
a) Power Failure
b) Dialines not conn
c) Shunt cover open
d) Connect conc.Line
e) Blood sensed by OD
f) Flow alarm
g) Water alarm
2) When T1 test fails, press Test key again and note the displayed error
code before proceeding to technical manual for further clues.
3) If more than 1 test fails select Test Service ON (PCB LP 632 DIP switch
array 1, SW 4 ON) & do the test individually to get the error codes of the
concerned test fail.
Test Equipments:
a) Hemodialysis tester
b) Pressure guage
c) Graduated Measuring jar (100ml, 1000ml)
d) Flashlight.
e) Digital multimeter
Pressure Units:
Possible Causes
S NO.
1. Water
alarm
Tips
Remarks
a) Water Inlet
Pressure(0.95 to
1.05 Bar)
Use Diagnostics
Program to
check the water
Flow Path.
Error
a) No Water
Supply/ Low
Pressure.
b) Pressure
Regulator (61)
c) Check the Inlet
Filter
d) Water Inlet
Valve
c) Check Valve 41
(41)
2.Flow alarm
a) Verify flow
a) Blocked filter
300, 500, 800
b) Defective
valve/pump(valve
resistance 50 to
80 Ohms)
c) Calibration
3.Upper Flow
Alarm
b) Calibrate if
required.
a) Create flow
from
diagnostics.
b)Verify Loading
pressure (1.2
1.45 Bar)
a) Degassing
filter
b) Degassing
pressure
c) Flow calibration
4.Temperature
i) High
temperature
ii) Low
temperature
a) Incoming
water
temperature
b) Temp Calib
a) Temp Calib
a) Incomplete hot
disinfection can be
a reason!
a) Flow errors
can lead to Low
Temp Alarm!
b) Heater Rod.
b) Heater rod
resistance
around 30 Ohms
a) Dilution Ratio.
a) Verify Dilution
Ratio in Set Up
Menu
a)Normally
1+35.83(Nacl26
);
depends on the
A & B part
b) Concentrate
or Bicarbonate
b) Wrong or
Expired
Solution??
b) Provided the
concentrate and
BIC strokes and
suction is OK
c) Calibrate if
required
5.Conductivity
i) Low
Conductivity
ii) High
Conductivity
d) Replace if
required
e) Conductivity
(Display)
Calibration
e) Temperature
and Conductivity
(display)
Calibration
f) Faulty Conc or
BIC Pump
a)Flow Errors
b) Dilution
c) Concentrate or
Bicarbonate
d) Conductivity
(Display)
Calibration
6.TMP/Dial
Pressure
7. Blood Leak
8. Pressure
Holding Test
i) Positive
a) Dialyser
membrane
defects
a) Verify flow,
Calibrate if
required
b) Verify Dilution
c) Wrong or
Expired Solution??
d) Temperature
and Conductivity
(display)
Calibration
a) Replace
Dialyser.
b) Dial Pressure
Calibration
Required.
b)Continuous
TMP Alarm /
Pressure fail
indicates Dial
Pressure
Calibration
Adjust both to 5
Volts.
a) Leakage in
a) Observe for
d) Calibrate the
Conc and BIC
pumps after
replacing the
springs. Use the
graduated
measuring Jar to
measure the
stroke volume.
a) Adjust 0
mmHg ( LP 633
Pot P3) and do
dial pressure
calibration.
a) Use a flash
Pressure Errors
Hydraulic
system.
leaks during
Positive pressure
test or Induced
pressure test.
b) Dial Pressure
Calibration
ii) Negative
Pressure Errors
b) Do Dial
Pressure
calibration only
if the error code
indicates dial
pressure
calibration.
a) UF filter
blockage
b) UF Pump
Calibration
9.Hydraulic
Test fail
i)Rinse Failure
Alarm
light.
b) Use the
graduated
measuring Jar.
c) Broken UF
Pump Spring set
c) Observe/
Replace Spring
set
d) Minute
Leakage in
Hydraulic
system
d) Check
tightness of
hoses
d) Use
Diagnostics
program
a) Rinse Filters
148 & 149
a) Replace
Filters if needed.
a) 3 minutes
before the end
of cleaning
program.
b)Rinse Valves
91,99,100
b) Most cases
pressure leak
due to loose
hose at these
valves
OR Leak from
Conc and BIC
pick up line
rubber seals.
c) Bibag
Pressure switch
134
c) Check PS 134
from
Diagnostics.
a) Clogged
a) 1 minute
Filter 210
i) Blood Pump
Errors
b) Jammed
Motor
c) Control Board
b) Replace
bearing
c) Do complete
arterial
calibration and
BP speed
calibration after
replacing the
control board
ii)Level detector
Alarm
iii) Arterial
Pressure Alarms
a) Air Detector
iv) Venous
Pressure Alarms
a) Transducer
Protector ( P
Ven)
a) Calibrate Air
detector
a) Arterial
Pressure
calibration
Partially clogged
female
connection.
b) Venous
Pressure
calibration
v) Heparin Pump
Errors
vi) Diasafe
Errors
c) Venous Line
set
a) Mechanical
problem of gear
b)F 34/ 43
c) Diasafe
Pressure
Transducer
11.
a) Manually
rotate the gear
to Top or bottom
end and retest
the pump
a) Dial Pressure
calibration
required
b) Damaged
Diasafe Filter
c) Replace
Diasafe filter.
Ultrafiltration
Errors.
i) Less fluid
removal
c) UF pump
spring set
d) UF Pump
12.Battery
Failure
a) Battery
b) Verify
Calibration
b) Verify the
stroke volume
(1ml/ stroke)
c)Replace if
required
d) Replace if
defective.
a) Battery Vol.
18 V
b) Battery Fuse
(3.15 A)
c) LP 647
c) LP 647 is in
the Power
supply module.
Diagnostics Program
1) On the service switch
2) Enter Diagnostics- Write outputs Write Digital outputs CPU 1 : Digital
outputs.