Академический Документы
Профессиональный Документы
Культура Документы
Perfecto2Series
Oxygen Concentrators without SensO2,
Oxygen Concentrators with SensO2
HomeFillCompatible
Perfecto2
Model IRC5P
Model IRC5PW
WARNING
WARNING
DO NOT USE THIS PRODUCT OR ANY AVAILABLE
OPTIONAL EQUIPMENT WITHOUT FIRST
COMPLETELY READING AND UNDERSTANDING
THESE INSTRUCTIONS AND ANY ADDITIONAL
INSTRUCTIONAL MATERIAL SUCH AS OWNERS
MANUALS, SERVICE MANUALS OR INSTRUCTION
SHEETS SUPPLIED WITH THIS PRODUCT OR
OPTIONAL EQUIPMENT. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS OR
INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL
PERSONNEL BEFORE ATTEMPTING TO USE THIS
EQUIPMENT - OTHERWISE, INJURY OR DAMAGE
MAY OCCUR.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
ACCESSORIES
Therearemanydifferenttypesofhumidifiers,oxygentubing,
cannulasandmasksthatcanbeusedwiththisdevice.You
shouldcontactyourlocalhomecareproviderfor
recommendationsonwhichofthesedeviceswillbebestfor
you.Theyshouldalsogiveyouadviceontheproperusage,
maintenance,andcleaning.
NOTE:Updatedversionsofthismanualcanbefoundat
www.invacare.com.
Perfecto2Series
Part No 1143482
Table of Contents
TABLE OF CONTENTS
ACCESSORIES .................................................. 2
SPECIAL NOTES ............................................... 4
LABEL LOCATION ............................................ 6
SECTION 1GENERAL GUIDELINES .................. 7
To Reduce the Risk of Burns, Electrocution, Fire or Injury to
Persons ......................................................................................................... 7
Operating Information ....................................................................... 8
Radio Frequency Interference.......................................................... 8
SECTION 2FEATURES .................................. 10
SECTION 3HANDLING ................................. 11
Unpacking............................................................................................ 11
Inspection ............................................................................................ 11
Storage................................................................................................. 11
SECTION 4TYPICAL PRODUCT
PARAMETERS ............................................... 12
SECTION 5OPERATING INSTRUCTIONS ....... 15
Introduction........................................................................................ 15
Select a Location ............................................................................... 15
Set Up .................................................................................................. 16
Power Switch ..................................................................................... 18
Flowrate............................................................................................... 19
Front Panel Indicators - Units with SensO2 ............................... 20
Front Panel Indicators - Units without SensO2 ......................... 21
SECTION 6MAINTENANCE .......................... 23
Routine Maintenance........................................................................ 23
SECTION 7TROUBLESHOOTING GUIDE ......... 26
LIMITED WARRANTY .................................... 31
Part No 1143482
Perfecto2Series
SPECIAL NOTES
SPECIAL NOTES
Signalwordsareusedinthismanualandapplytohazardsor
unsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Refertothefollowingtablefordefinitions
ofthesignalwords.
SIGNAL WORD
MEANING
DANGER
WARNING
CAUTION
NOTICE
Theinformationcontainedinthisdocumentissubjectto
changewithoutnotice.
DANGER
DO NOT SMOKE while using this device. Keep all
matches, lighted cigarettes or other sources of ignition
out of the room in which this product is located and away
from where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physical
injury or death.
Perfecto2Series
Part No 1143482
SPECIAL NOTES
CAUTION
Caution:Federallawrestrictsthisdevicetosaleorrentalby
oronorderofaphysician,oranyotherpractitionerlicensed
bythelawoftheStateinwhichhe/shepracticestouseor
ordertheuseofthisdevice.
Invacarerecommendsanalternatesourceofsupplemental
oxygenintheeventofapoweroutage,alarmconditionor
mechanicalfailure.Consultyourphysicianorequipment
providerforthetypeofreservesystemrequired.
Thisequipmentistobeusedasanoxygensupplementandis
notconsideredlifesupportingorlifesustaining.
Part No 1143482
Perfecto2Series
LABEL LOCATION
LABEL LOCATION
Perfecto2
FOR OPERATING INSTRUCTIONS or ALARMS (See Operators Manual)
DANGER - RISK OF FIRE
NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and
away from areas where oxygen is being delivered. Textiles, oil and other combustibles are easily
ignited and burn with great intensity in oxygen enriched air.
HomeFill
Compatible
P/N
1
147
920
Rev
.B
Specification Label is
located on the back
near the bottom
Perfecto2Series
Part No 1143482
WARNING
Thissectioncontainsimportantinformationforthesafe
operationanduseofthisproduct.
Perfecto2Series
Operating Information
Foroptimumperformance,Invacarerecommendsthateach
concentratorbeonandrunningforaminimumof30minutes
atatime.Shorterperiodsofoperationmayreducemaximum
productlife.
Keeptheoxygentubing,cord,andunitoutfromundersuch
itemsasblankets,bedcoverings,chaircushions,clothingand
awayfromheatedorhotsurfaces,includingspaceheaters,
stovesandsimilarelectricalappliances.
DONOTmoveorrelocateconcentratorbypullingonthe
powercord.
NEVERdroporinsertanyobjectorliquidintoanyopening.
InvacarerecommendsthatCrushProofoxygentubingbe
usedwiththisproductandnotexceed50ftinlength.
Therearenouserserviceableparts.Thisdoesnotinclude
normalmaintenanceitems.Seemaintenancesectionforuser
maintenanceitems.
AproductshouldNEVERbeleftunattendedwhenplugged
in.MakesurethePerfecto2isOffwhennotinuse.
Closesupervisionisnecessarywhenthisproductisusednear
childrenorphysicallychallengedindividuals.
Additionalmonitoringorattentionmayberequiredfor
patientsusingthisdevicewhoareunabletohearorsee
alarmsorcommunicatediscomfort.
DONOTconnecttheconcentratorinparallelorserieswith
otheroxygenconcentratorsoroxygentherapydevices.
Perfecto2Series
Part No 1143482
Part No 1143482
Perfecto2Series
SECTION 2FEATURES
SECTION 2FEATURES
Oxygen Outlet
Flowmeter
Circuit Breaker
Elapsed Time
Meter
Power Switch
REAR VIEW
Cabinet Filter
Power
Cord
Perfecto2Series
*Outlet
Fitting
*NOTE:Thisoutletfittingistobe
usedonlyforfillingoxygencylinders
withtheHomeFillhomeoxygen
compressor.Theoutletfittingdoesnot
affectconcentratorperformance.
RefertotheHomeFillowners
manual,partnumber1100873,for
connectionandoperating
instructions.Whennotinuse,the
plugprovidedwiththeconcentrator
shouldbeinsertedintotheoutlet
fitting.Formoreinformationabout
theHomeFill,contactyourInvacare
dealer.
10
Part No 1143482
SECTION 3HANDLING
SECTION 3HANDLING
TheconcentratorshouldALWAYSbekeptintheupright
positiontopreventcabinetdamagewhilebeingtransported.
Iftheconcentratoristobereshippedbycommoncarrier,
additionalcartonsareavailablefromInvacare.
Unpacking
NOTE:Forthisprocedure,refertoFIGURE3.1.
1. Checkforanyobvious
damagetothecartonor
itscontents.Ifdamageis
evident,notifythe
carrier,oryourlocal
dealer.
2. Removeallloosepacking
fromthecarton.
3. Carefullyremoveallthe
componentsfromthe
carton.
Inspection
Inspect/examineexterioroftheoxygenconcentratorfornicks,
dents,scratchesorotherdamages.Inspectallcomponents.
Storage
1. Storetherepackagedoxygenconcentratorinadryarea.
2. DONOTplaceobjectsontopofrepackagedconcentrator.
Part No 1143482
11
Perfecto2Series
3A
Sound Level:
Altitude:
*Oxygen Output
Concentration Levels:
Flow Range:
0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Average Power
Consumption:
Perfecto2Series
12
Part No 1143482
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient
Temperature:
Exhaust:
Oxygen Output
Temperature:
Cabinet:
Regulatory Listing:
Electrical:
No extension cords.
Part No 1143482
13
Perfecto2Series
Tubing:
Relative Humidity:
20 to 60%
Time of Operation:
Recommended Storage
and Shipping Temperature:
*NOTE:Statedconcentrationlevelsachievedafterinitialwarmup
period(approximately30minutes).
Perfecto2Series
14
Part No 1143482
SECTION 5OPERATING
INSTRUCTIONS
Introduction
Youroxygenconcentratorisintendedforindividualuse
indoors.Itisanelectronicallyoperateddevicethatseparates
oxygenfromroomair.Itprovideshighconcentrationofoxygen
directlytoyouthroughanasalcannula.Clinicalstudieshave
documentedthatoxygenconcentratorsaretherapeutically
equivalenttoothertypesofoxygendeliverysystems.
Yourproviderwillshowyouhowtouseyouroxygen
concentrator.He/Sheshouldbecontactedwithanyquestions
orproblemsregardingyouroxygenconcentrator.This
ownersmanualwilltellyouaboutyourconcentratorandwill
serveasareferenceasyouuseyourconcentrator.
Select a Location
WARNING
NEVERblocktheairopeningsoftheproductorplaceitona
softsurface,suchasabedorcouch,wheretheairopeningmay
beblocked.Keeptheopeningsfreefromlint,hairandthelike.
Keepunitatleast3inchesawayfromwalls,draperies,
furniture,andthelike.
Youmayselectaroomwhereusingyouroxygenconcentrator
wouldbemostconvenient.Yourconcentratorcanbeeasily
rolledfromroomtoroomonitscasters.
Youroxygenconcentratorwillperformbestwhenoperated
underthefollowingconditions.RefertoTypicalProduct
Parametersonpage 12.Usageinenvironmentsotherthan
thosedescribedmayresultintheneedforincreased
equipmentmaintenance.Theairintakeoftheunitshouldbe
locatedinawellventilatedareatoavoidsmoke,airborne
pollutantsand/orfumes.DONOTplaceinacloset.
Part No 1143482
15
Perfecto2Series
Set Up
Plug in Power Cord
Pluginpowercordtoanelectricaloutlet.
Humidifier
Bottle with Cap
Perfecto2Series
16
Part No 1143482
Cabinet Filter
FIGURE 5.2
5. Attachthehumidifierbottleadaptertothehumidifier
bottlebyturningthewingnutonthehumidifierbottle
counterclockwiseuntilitissecurelyfastened.SeeDetail
AinFIGURE5.3.
6. Placethehumidifierbottle/adapterassemblyinthe
humidifiercompartmentontheconcentrator.SeeDetail
BinFIGURE5.3.
7. Attachoxygentubingfromthehumidifierbottle/adapter
assemblytooxygenoutletconnectorontheconcentrator.
SeeDetailBinFIGURE5.3.
8. Attachthecannula/patientsupplytubingtothe
humidifierbottleoutlet.SeeDetailBinFIGURE5.3.
9. Afterassembly,ensurethatoxygenisflowingthroughthe
cannula.
Part No 1143482
17
Perfecto2Series
Humidifier
Bottle
Adapter
Humidifier
Bottle
DETAIL B
Oxygen Outlet
Connector
Oxygen
Tubing
Humidifier
Bottle
Outlet
Humidifier
Bottle
Humidifier
Compartment
FIGURE 5.3
Humidifier Compartment
Power Switch
NOTE:Forthisprocedure,referto
FIGURE5.4.
On/Off
(I/O)
Power
Switch
1. Presspowerswitchtoon
position.Allthepanel
lightsandtheaudible
alarmwillcomeonfor
onesecond,indicating
thattheunitisfunctioning
properly.
Perfecto2Series
18
Part No 1143482
Flowrate
NOTE:Forthisprocedure,refertoFIGURE5.5onpage 20.
1. Turntheflowrateknobtothesettingprescribedbyyour
physicianortherapist.
CAUTION
Neversettheflowgreaterthan5L/min.
WARNING
DONOTchangetheL/minsettingontheflowmeterunlessa
changehasbeenprescribedbyyourphysicianortherapist.
NOTE:Toproperlyreadtheflowmeter,locatetheprescribedflowrate
lineontheflowmeter.Next,turntheflowknobuntiltheballrisesto
theline.Now,centertheballontheL/minlineprescribed.
2. Iftheflowrateontheflowmetereverfallsbelow0.5L/min
formorethanaboutoneminute,theLOWFLOWalarm
willbetriggered.Thisisarapidbeepingoftheaudible
alarm.Checkyourtubingoraccessoriesforblockedor
kinkedtubingoradefectivehumidifierbottle.Afterrated
flowisrestored,between0.5L/minand0.75L/min,the
LOWFLOWaudiblealarmwillturnoff.
NOTE:Theuseofsomeaccessoriessuchasthepediatricflowstand
andtheHomeFillcompressorwilldeactivatetheLowFlowAlarm.
Part No 1143482
19
Perfecto2Series
Flowmeter
0.5 L/min
Ball
Front Panel
FIGURE 5.5
Flowrate
RED
Call Supplier
Below
Normal
Normal
FIGURE 5.6
Perfecto2Series
YELLOW
O2
GREEN
20
Part No 1143482
O2 PURITY
SYSTEM OKAY
O2 over 85% ( 2%)
O2
RED
Call Supplier
Normal
FIGURE 5.7
Part No 1143482
I/O
GREEN
21
Perfecto2Series
I/O
Perfecto2Series
STATUS
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
SYSTEM FAILURE
22
Part No 1143482
SECTION 6MAINTENANCE
SECTION 6MAINTENANCE
WARNING
TheInvacareconcentratorsarespecificallydesignedto
minimizeroutinepreventivemaintenance.Onlyqualified
personnelshouldperformpreventivemaintenanceonthe
concentrator.
Powershouldbedisconnectedbeforebeginningpreventive
maintenanceontheInvacareconcentrator.
NOTE:Ataminimum,preventivemaintenanceMUSTbe
performedaccordingtothemaintenancerecordguidelines.Inplaces
withhighdustorsootlevels,maintenancemayneedtobeperformed
moreoften.RefertoPreventiveMaintenanceRecordonpage 25.
Routine Maintenance
NOTE:Thefollowingroutinemaintenanceshouldbeperformed
morefrequently.Seerecommendedintervalsineachsectionand
preventivemaintenancerecord.
23
Perfecto2Series
SECTION 6MAINTENANCE
2. Cleanthecabinetfilterwithavacuumcleanerorwashin
warmsoapywaterandrinsethoroughly.
3. Drythefilterthoroughlybeforereinstallation.
Perfecto2Series
24
Part No 1143482
Part No 1143482
25
AS REQUIRED
I RC
MODEL NO.
ON EACH INSPECTION
SECTION 6MAINTENANCE
SERIAL NO.
Perfecto2Series
SECTION 7TROUBLESHOOTING
GUIDE
SYMPTOM:
Alarm:
Short beeps, long
pause
Concentrator not
operating,
power switch On.
Beep......Beep......
Alarm:
Continuous
PROBABLE CAUSE:
Main Power Loss:
1. Power cord not
plugged in.
2. No power at outlet.
3. Tripped circuit
breaker.
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required.
2. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is below
0.5 L/min, use the Pediatric
Flowmeter IRCPF16 (Only
with Perfecto2 5).
Concentrator not
operating, Power
Switch On.
Beeeeeeeep....
Alarm:
Rapid
Beep....Beep...
Beep...Beep
Perfecto2Series
SOLUTION:
26
Part No 1143482
PROBABLE CAUSE:
SOLUTION:
2. Kinked or blocked
tubing, cannula or
humidifier.*
Part No 1143482
5. Insufficient power at
outlet.
6. Internal repairs
required.
1. Internal repairs
required.
27
Perfecto2Series
NOTES
NOTES
Perfecto2Series
28
Part No 1143482
NOTES
NOTES
Part No 1143482
29
Perfecto2Series
NOTES
NOTES
Perfecto2Series
30
Part No 1143482
LIMITED WARRANTY
LIMITED WARRANTY
NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY
WITH FEDERAL LAW APPLICABLE TO PRODUCTS
MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser who purchases
this product when new and unused from Invacare Corporation or a
dealer. This warranty is not extended to any other person or entity and it
is not transferable or assignable to any subsequent purchaser or owner.
Coverage under this warranty will end upon any such subsequent sale or
other transfer of title to any other person. This warranty gives you specific
legal rights and you may also have other legal rights which may vary from
state to state.
Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when
purchased new and unused to be free from defects in materials and
workmanship for a period of three years from date of purchase from
Invacare or a dealer, with a copy of the sellers invoice required for
coverage under this warranty.
DO NOT OPEN OR ATTEMPT TO SERVICE THE 4-WAY VALVE
BODY. This will void any and all warranty applicable to the valve body. If
within such warranty periods any such product shall be proven to Invacare
Corporation's satisfaction to be defective, such product shall be repaired
or replaced, at Invacare Corporation's option. This warranty only applies
to the labor for repairs performed by the Invacare Service Department or
Invacare Authorized Service Centers. It does not apply to the labor
performed by the purchaser or user. This warranty does not include
normal wear and tear or shipping charges. Invacare Corporation's sole
obligation and your exclusive remedy under this warranty shall be limited
to such repair or replacement. Routine maintenance items, such as filters,
are excluded from this warranty.
For warranty service, please contact Invacare Corporation's Service
Department at the toll free number on the back cover. Upon receiving
notice of an alleged defect in a product, Invacare Corporation will issue a
serialized return authorization. It shall then be the responsibility of the
purchaser to return the entire unit or remove, at purchaser's cost, the
defective component part(s) identified, pack the component part(s) in a
manner to avoid shipping damage and to ship the component part(s) to
either Invacare Corporation's plant or service center as specified by
Invacare Corporation in advance. Defective component part(s) MUST be
returned for warranty inspection using the serial number as identification
within thirty days of return authorization date. DO NOT return products
to our factory without prior consent. C.O.D. shipments will be refused;
please prepay shipping charges.
Part No 1143482
31
Perfecto2Series
Invacare Corporation
www.invacare.com
USA
One Invacare Way
Elyria, Ohio USA
44036-2125
800-333-6900
Technical Services
800-832-4707
Part No 1143482
Rev E - 11/09