Академический Документы
Профессиональный Документы
Культура Документы
Enclosure
Imaging Services
CR 30-X
Type 5175/100/105/110/120/125
Referenced Documents
Document
Title
DD+DIS103.11E
DD+DIS310.09E
DD+DIS150.06E
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
05-2011 printed in Germany
DD+DIS112.11E
Enclosure
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2011 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 2 of 21
DD+DIS112.11E
Enclosure
LIST OF CONTENTS
INTRODUCTION ......................................................................................................................6
10
After Uninstallation of the Software the Message appears: You must restart your system 19
10.2
After Uninstallation of the Software the "Found New Hardware Wizard" appears .................20
10.3
After initial Installation the "Found New Hardware Wizard" does not appear.........................21
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 3 of 21
DD+DIS112.11E
Subject
Prerequisite
Enclosure
If the feature "speed class free" operation shall be used, NX workstation software NX
2.0.8000 - NX 3.0.8000 (and higher versions) is required.
For details see:
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 4 of 21
DD+DIS112.11E
Enclosure
REQUIRED TOOLS:
If software is part of a spare part kit, and distributed on CD-ROM:
REQUIRED TIME:
Approximately 60 minutes, without possibly required software upgrade of the
NX workstation
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 5 of 21
DD+DIS112.11E
Enclosure
Introduction
Software CRE_1907 is the successor of software CRE_1901.
It contains the following improvements compared to software CRE_1901:
Error 3192 does not occur anymore when the TWAIN client is closed
Polygon and Photomultiplier high tension are switched off if the scan cycle is
cancelled (e.g. by pressing "Eject" button in the image processing application)
Upgrade installation
Step
Initial
Installation
Upgrade
Installation
See
section
Perform a backup.
10
* In exceptional cases, if a software mismatch error 2712 appears after connection to the digitizer, an
upload of the digitizer software is required for initial installation, too.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 6 of 21
DD+DIS112.11E
Enclosure
Initial installation
Upgrade Installation
Figure 1
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 7 of 21
DD+DIS112.11E
Enclosure
Initial installation
Upgrade Installation
IMPORTANT:
Before starting any software installation activity at the control PC, switch off the
digitizer. Otherwise the digitizer driver on the control PC might not be installed
properly.
NOTE:
Windows 7 uses by default a driver for the FireWire interface which is not compatible
with the digitizer.
The Legacy type of FireWire interface has to be activated. Otherwise image
transmission will not work: The operating system of the Control PC will crash.
(1)
(2)
Select <START Shut down Log off> to log-off the current user.
(3)
(4)
(5)
(6)
(7)
Figure 2
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 8 of 21
DD+DIS112.11E
Enclosure
(8)
(9)
Figure 4
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 9 of 21
DD+DIS112.11E
Enclosure
Figure 5
Figure 6
Figure 7
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 10 of 21
DD+DIS112.11E
Enclosure
Initial installation
Upgrade Installation
(1)
Select <START - Log Off (current user)> to log-off the current user.
(2)
(3)
(4)
Figure 8
(5)
(6)
Deinstall the digitizer software on the control PC: Select: <Start - Settings
- Control Panel - Add or Remove Programs - CR 30-X>
(7)
If the message "You must restart your system ...." appears at the end of the
deinstallation process, continue with the steps described in section 11.1.
Otherwise continue with next section.
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 11 of 21
DD+DIS112.11E
Enclosure
Initial installation
Upgrade Installation
NOTE:
For the initial installation of the NX workstation steps (1) and (2) are already done by
production.
(1)
(2)
(3)
Press next and follow the installation wizard. See IMPORTANT note below.
IMPORTANT:
Do not change the suggested
default installation folder:
C:\Program Files\Agfa\CR 30-X\
Figure 9
(4)
(5)
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 12 of 21
DD+DIS112.11E
Enclosure
Initial installation
Upgrade Installation
(1)
(2)
(3)
Click <next>.
Figure 10
IMPORTANT:
If the Found new hardware wizard does not appear during initial installation of the
software, the driver for the digitizer possibly is not installed properly. Refer to section
11.3 for an instruction how to check and correct the problem.
(4)
Click <Finish>.
Figure 11
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 13 of 21
DD+DIS112.11E
Enclosure
Initial installation*
Upgrade Installation
* In exceptional cases, if a software mismatch error 2712 appears after connection to the digitizer, an
upload of the digitizer software is required for initial installation, too.
(1)
(2)
(Error 2712)
Figure 12
(3)
(4)
On the control PC select menu: <Start - Programs - Agfa CR 30-X Service Client>
(5)
(6)
(7)
(8)
IMPORTANT:
Do not upload the nodes individually, but use the <Select All> functionality, to mark
all three nodes for upload automatically. Reason: The software for the PMI board has
to be uploaded first before the software for the two other boards is uploaded.
(9)
Status
different
different
different
Select All
Clear Selection
517501df.cdr
Figure 13
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 14 of 21
DD+DIS112.11E
(10)
Enclosure
Wait for approximately 7 minutes until the message "Versions are current"
appears in the message box. Refer to Figure 14.
If message "SW upload required" appears instead, refer to IMPORTANT note
next page.
Node
Installed Version
PMI Board/FPGA CRE_1907/2104
CRE_1907
HDL Board
CRE_1907
SSC Board
Version on PC
CRE_1907/2104
CRE_1907
CRE_1907
Status
ok
ok
ok
Select All
Clear Selection
Activity
(11)
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 15 of 21
DD+DIS112.11E
Enclosure
IMPORTANT:
After upload of the selected boards, the PMI board is restarted. In some cases the
version check of the nodes fails like in this example for the PMI2 board:
Starting version check of nodes
PMI: installed version 'unknown/unknown'; file version 'CRE_1907/2104'
No controller with CAN-ID 0x780 found in machine
Controller HDL Board(0x780): installed version 'unknown'; file version 'CRE_1907'
No controller with CAN-ID 0x784 found in machine
Controller SSC Board(0x784): installed version 'unknown'; file version 'CRE_1907'
SW upload required.
If message "Versions are current" does not appear: Repeat the software upload:
Go back to step (8).
Initial installation*
Upgrade Installation
* during initial installation (= activation of the NX workstation software or during NX clean installation
procedure) this step only has to be performed, if the software on the digitizer was below CRE_1907.
NOTE:
Together with software CRE_1907 a new type of fixer motor was introduced with
different motor characteristics.
Software CRE_1907 in general supports all types of fixer motors. To ensure that the
previous type of fixer motor runs at optimal speed a stall calibration of the fixer motor is
required however.
If the stall calibration is not performed during software upgrade, throughput would be
slightly decreased, as some movements of the fixer motor would be slower.
(1)
(2)
(3)
(4)
Wait until the stall calibration is finished: This takes approximately 1 minute.
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 16 of 21
DD+DIS112.11E
Enclosure
Performing a Backup
This step is required for:
Initial installation
Upgrade Installation
(1)
In the CR 30-X service client select the menu <Backup & Restore>.
(2)
Figure 15
(3)
(4)
(5)
(6)
(7)
Insert a backup USB stick at the USB port of the control PC.
(8)
Copy the backup file stored on the partition <D:\> on the control PC to the
USB stick.
(9)
Unmount and remove the USB stick from the control PC and store it on a save
place.
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 17 of 21
DD+DIS112.11E
Enclosure
Figure 16
10
Initial installation*
Upgrade Installation
* These verification steps are also part of the digitizer installation instructions.
Result
(1)
(2)
(3)
(4)
(5)
(6)
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 18 of 21
DD+DIS112.11E
11
Enclosure
Symptom
See Section:
11.1
11.2
11.3
11.1
Symptom
Figure 17
Cause
The NX software or the error viewer have not been closed before software deinstallation.
Solution
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 19 of 21
DD+DIS112.11E
Enclosure
11.2
Symptom
Figure 18
Cause
Digitizer was not switched off during software uninstallation. In this case the digitizer
driver on the control PC will also be uninstalled.
Solution
(1)
Cancel the "Found new Hardware Wizard" and continue with software
installation. See section 5.
(2)
Install the driver via "found new hardware wizard" which pops up automatically
after software upload to the digitizer (and restart of the control PC).
Figure 19
Revision 0
05-2011
CR 30-X
Type 5175/100/105/110/120/125
Page 20 of 21
DD+DIS112.11E
Enclosure
11.3
After initial Installation the "Found New Hardware Wizard" does not
appear
Symptom
After initial installation the "Found New Hardware Wizard" does not appear. The
connection between digitizer and control PC cannot be established: The status LED at
the digitizer stays red blinking.
Device driver for the digitizer not installed properly.
Cause
(1)
(2)
Solution
B)
Imaging devices
PMI
(3)
? Other devices
?! PMI
PMI or Agfa or CR30 listed under
other devices. In this case the
driver is not properly installed.
Proceed with next step.
This step has only to be performed for result B), PMI or Agfa or CR30 listed
under other devices:
- Right-click on the driver named PMI or Agfa or CR30.
- Select uninstall.
- Be sure to have digitizer switched on and the FireWire cable connected.
- Reboot the control PC.
- Re-install the digitizer driver as described in Figure 19.