Академический Документы
Профессиональный Документы
Культура Документы
Anesthesia Vaporizers
James B. Eisenkraft
CHAPTER OUTLINE
GENERAL PRINCIPLES
VAPOR, EVAPORATION, AND VAPOR PRESSURE
Measurement of Vapor Pressure
andSaturated Vapor Pressure
Boiling Point
Units of Vapor Concentration
Daltons Law of Partial Pressures
Minimum Alveolar Concentration
Latent Heat of Vaporization
Specific Heat
Regulating Vaporizer Output: Variable Bypass
Versus Measured Flow
Calculation of Vaporizer Output
VAPORIZER FUNCTION
Measured Flow Vaporizers
Variable Bypass Vaporizers
Control of Splitting Ratio
Efficiency and Temperature Compensation
Incorrect Filling of Vaporizers
Vaporization of Mixed Anesthetic Liquids
Filling of Vaporizers
Effect of Carrier Gas on Vaporizer Output
Effects of Changes in Barometric Pressure
Vaporizing Chamber Flow Controlled at Inlet
Hypobaric Conditions
GENERAL PRINCIPLES
The term vapor describes the gaseous phase of a substance
at a temperature at which the substance can exist in either
a liquid or solid state below a critical temperature for that
substance. If the vapor is in contact with a liquid phase,
the two phases will be in a state of equilibrium, and the
gas pressure will equal the equilibrium vapor pressure of
the liquid. The potent inhaled volatile anesthetic agents
halothane, enflurane, isoflurane, sevoflurane, and
desfluraneare mostly in the liquid state at normal room
temperature (20 C) and atmospheric pressure (760 mm
Hg).1 Anesthesia vaporizers are devices that facilitate the
change of a liquid anesthetic into its vapor phase and add
a controlled amount of this vapor to the flow of gases
entering the patients breathing circuit.
The anesthesia care provider should be familiar with
the principles of vaporization of the potent inhaled anesthetic agents and their application in both the construction
64
Hyperbaric Conditions
Vaporizer Chamber Flow Controlled at
Outlet
Arrangement of Vaporizers
Calibration and Checking of Vaporizer
Outputs
Preparation of a Standard Vapor
Concentration
3 Anesthesia Vaporizers
65
Boiling Point
The SVP exerted by the vapor phase of a potent inhaled
volatile agent is a physical property of that agent and
depends only on the agent and the ambient temperature.
The temperature at which SVP becomes equal to ambient (atmospheric) pressure and that at which all the liquid
agent changes to the vapor phase (i.e., evaporates) is the
boiling point of that liquid. Water boils at 100 C at
1atm because at 100 C, the SVP of water is 760 mm Hg.
The most volatile of the agents are those with the highest
SVPs at room temperature. At any given temperature,
these agents also have the lowest boiling points: desflurane and diethyl ether boil at 22.9 C and 35 C, respectively, at an ambient pressure of 760 mm Hg. Boiling
point decreases with decreasing ambient barometric pressure, such as occurs at increasing altitude.
Vaporized agent
Vapor pressure
760 mm
760 mm
Liquid
agent
SVP
Liquid
agent
1 atm
FIGURE 3-1 n Measurement of vapor pressures using a simple Fortin barometer. SVP, saturated vapor pressure. (From Eisenkraft JB:
Vaporizers and vaporization of volatile anesthetics. In Progress in anesthesiology, vol 2. San Antonio, 1989, Dannemiller Memorial Educational
Foundation.)
66
100 %
Desflurane
Isoflurane
Halothane
Enflurane
Sevoflurane
1400
1200
1000
Specific Heat
Specific heat is the quantity of heat (calories) required to
raise the temperature of a unit mass (grams) of a substance
by 1 C. Heat must be supplied to the liquid anesthetic in
the vaporizer to maintain the liquids temperature during
the evaporation process, when heat is being lost.
TABLE 3-1 E
xpression of MAC as a Partial
Pressure
800
Agent
MAC (vol%)
600
Halothane
Enflurane
Isoflurane
Methoxyflurane
Sevoflurane
Desflurane
0.75 760
1.68 760
1.15 760
0.16 760
2.10 760
7.25 760
400
200
0
0
10 15 20 25 30 35 40 45 50 55 60 65
Temperature C
3 Anesthesia Vaporizers
Specific heat is also important when it comes to vaporizer construction material. For the same amount of heat
lost through vaporization, temperature changes are more
gradual for materials with a high specific heat than for
those with a low specific heat. Thermal capacity, defined as
the product of specific heat and mass, represents the
quantity of heat stored in the vaporizer body.
Also of importance is the construction materials ability to conduct heat from the environment to the liquid
anesthetic. This property is called thermal conductivity,
defined as the rate at which heat is transmitted through a
substance. For the liquid anesthetic to remain at a relatively constant temperature, the vaporizer is constructed
from materials that have a high specific heat and high
thermal conductivity. In this respect, copper comes close
to the ideal; however, bronze and stainless steel have been
used more recently in vaporizer construction.
67
Contemporary anesthesia vaporizers are concentration calibrated, and most are of the variable bypass design.
In a variable bypass vaporizer, such as those made by GE
Healthcare (Tec series) and the Drger Vapor 2000
(Drger Medical, Telford, PA), the total fresh gas flow
from the anesthesia machine flowmeters passes to the
vaporizer (Fig. 3-3). The vaporizer splits the incoming
gas flow between two pathways: the smaller flow enters
the vaporizing chamber, or sump, of the vaporizer and
leaves it with the anesthetic agent at its SVC. The larger
bypass flow is eventually mixed with the outflow from the
vaporizing chamber to create the desired, or dialed in,
concentration (see Fig. 3-3).
In the now-obsolete measured flow, non-concentrationcalibrated vaporizers such as the Copper Kettle (PuritanBennett; Covidien, Mansfield, MA) or Verni-Trol (Ohio
Medical Products, Gurnee, IL), a measured flow of oxygen is set on a separate flowmeter to pass to the vaporizer,
from which vapor emerges at its SVP (Fig. 3-4). This
flow is then diluted by an additional measured flow of
gases (oxygen, nitrous oxide, air, etc.) from the main
flowmeters on the anesthesia machine. With this type of
arrangement, calculations are necessary to determine the
anesthetic vapor concentration in the emerging gas
mixture.
With both types of vaporizing systems, there must be
an efficient method to create a saturated vapor in the
vaporizing chamber. This is achieved by having a large
surface area for evaporation. Flow-over vaporizers
(Drger Vapor 2000 series, GE Tec series) increase the
surface area using wicks and baffles. In measured flow,
bubble-through vaporizers, oxygen is bubbled through
the liquid agent. To increase the surface area, tiny bubbles are created by passing the oxygen through a sintered
bronze disk in the Copper Kettle, for example, which created large areas of liquid/gas interface, over which evaporation of the liquid agent could quickly occur.
Halothane
Structure
CHBrClCF3
Molecular weight
197.4
(AMU)
Boiling point at 760
50.2
mm Hg ( C)
SVP at 20 C (mm Hg)
243
SV conc. at 20 C and
32
1ATA* (vol%)
MAC at 1 ATA*
0.75
(vol%)
PMAC1 (mm Hg)
5.7
Specific gravity of
1.86
liquid at 20 C
mL vapor per gram
123
liquid at 20 C
mL vapor per mL
226
liquid at 20 C
Enflurane
Isoflurane
Methoxyflurane
Sevoflurane
Desflurane
CHFClCF2OCHF2
184.5
CF2HOCHClCF3
184.5
CHCl2CF2OCH3
165.0
CH2FOCH(CF3)2
200
CF2HOCFHCF3
168
56.5
48.5
104.7
58.5
22.8
175
23
238
31
20.3
2.7
160
21
664
87
1.68
1.15
0.16
2.10
6-7.25
12.8
1.52
8.7
1.50
1.22
1.41
16
1.51
46-55
1.45
130
130
145
120
143
196
195
204
182
207
68
Vaporizing
chamber or sump
Vaporizer
outflow at
dialed-in
concentration
Bypass flow
Saturated vapor
Liquid agent in
vaporizer sump
FIGURE 3-3 n Schematic of a concentration-calibrated variable
bypass vaporizer. Fresh gas enters the vaporizer, where its flow
is split between a larger bypass flow and a smaller flow to the
vaporizing chamber or sump. In the sump is the agent at its saturated vapor concentration. Saturated vapor mixes with the
bypass flow, which dilutes it to the concentration dial setting.
Fresh gas from
main flowmeters
To patient
Measured
flow of
oxygen to
vaporizer
VAPORIZER FUNCTION
Liquid anesthetic agent
FIGURE 3-4 n Schematic of a measured flow vaporizing arrangement. These are commonly known as bubble-through
vaporizers.
x
69
31x = 50 69
x=
(50 69)
31
= 111 mL
3 Anesthesia Vaporizers
To patient
5000 mL/min
of 1% isoflurane
161 mL/min of
31% isoflurane in O2
Measured
flow of
oxygen to
vaporizer
111 mL/min
50 mL/min of
isoflurane vapor
isoflurane. It is important to realize that if there is oxygen flow only to the Copper Kettle vaporizer and
no bypass gas flow is set on the main machine flowmeters, lethal concentrations approaching 31% isoflurane
would be delivered to the anesthesia circuit, albeit at low
flow rates.
Because halothane and isoflurane have similar SVPs at
20 C, the Copper Kettle flows to be set for halothane
would be essentially the same as those for isoflurane when
a 1% concentration of isoflurane is to be created with a
Copper Kettle. A Copper Kettle arrangement on an older
model anesthesia machine is shown in Figure 3-6.
Because enflurane and sevoflurane have similar vapor
pressures at 20 C (175 mm Hg and 160 mm Hg, respectively), similar flow settings could be used to create approximately the same agent concentrations with a measured
flow system. In the case of sevoflurane, the measured flow
vaporizer contains 21% sevoflurane vapor (160/760 =
21%). The oxygen flow therefore represents the remaining
79% of the atmosphere in the Copper Kettle. If precisely
1% sevoflurane is required at a 5 L/min total rate of flow,
50 mL/min of sevoflurane vapor must be generated. If 50
mL represents 21% of the atmosphere in the vaporizer, the
carrier gas flow required is 188 mL/min ([50/21] 79).
If 188 mL/min of oxygen are bubbled through liquid
sevoflurane contained in a measured flow vaporizer,
238mL/min of gas will emerge, 50 mL/min of which is
sevoflurane vapor. This must be diluted by a fresh gas
flow of 4762 mL/min (5000 238) to achieve exactly 1%
sevoflurane.
Alternatively, using the formula given previously:
160
760
Isoflurane
FIGURE 3-5 n Preparation of 1% isoflurane by volume using a
measured flow vaporizing system. 1% isoflurane in a 5L/min
flow requires 50 mL/min isoflurane vapor, diluted in a total volume of 4950 mL fresh gas plus 50 mL isoflurane vapor. Isoflurane saturated vapor concentration is 31%. If 31% = 50 mL, then
69% = 111 mL, the required oxygen inflow per minute; 4839 mL/
min (4950 111) is the required bypass flow, and final dilution is
1% (50/[50 + 4839 + 111]).
69
50
50 + y
y = 188 mL/min
FIGURE 3-6 n Copper Kettle vaporizing system (Puritan-Bennett; Covidien, Mansfield, MA). A, The oxygen flowmeter knob on the
extreme left is marked C-K to indicate that it controls oxygen flow to the Copper Kettle. B, Close-up of the Copper Kettle.
70
In the preceding examples, calculations of both the oxygen flow to the measured flow vaporizer and the total
bypass gas flow needed to produce the desired output concentrations of vapor were required. This is not only inconvenient, it also predisposes to clinical errors that could
result in serious overdose (Fig. 3-7) or underdose of anesthetic. Because of the obvious potential for error with a
measured flow vaporizing system, the concurrent continuous use of an anesthetic agent analyzer with high- and lowconcentration alarms would be essential for patient safety.7
To patient
5000 mL/min of
1% sevoflurane
238 mL/min of
21% sevo in O2
O2 flow to
copper kettle
188 mL/min
50 mL/min of
sevoflurane vapor
Sevoflurane liquid
FIGURE 3-7 n Preparation of 1% sevoflurane by volume using a
measured flow vaporizing system.
2000
Bypass flow
100
79
Flow to
vaporizing
chamber
2000
O2
21
N2O
Agent
160 = 21%
760
21 = 1%
2100
21
= 1%
21 + 79 + 2000
2000/79 = 25:1
FIGURE 3-8 n Preparation of 1% (vol/vol) sevoflurane in a variable bypass vaporizer. To achieve this, a split ratio of 25:1 (bypass flow/
vaporizing chamber flow) is created for incoming gas.
3 Anesthesia Vaporizers
71
Bypass flow
100
69
Flow to
vaporizing
chamber
O2
31
N2O
Agent
Agent at SVC
31
= 1%
31 + 69 + 3000
239 = 31%
760
31 = 1%
3100
3000 = 44:1
69
FIGURE 3-9 n Preparation of 1% (vol/vol) isoflurane by a variable bypass vaporizer. To achieve this, a split ratio of 44:1 (bypass flow/
vaporizing chamber flow) is created for incoming gas.
Halothane
Enflurane
Isoflurane
Methoxyflurane
46:1
22:1
14:1
29:1
14:1
9:1
44:1
21:1
14:1
1.7:1
0.36:1
*
Sevoflurane
25:1
12:1
7:1
effective flow split for the fresh gas entering the vaporizer
but not where the split is achieved (i.e., whether by the
inlet or the outlet of the vaporizing chamber). If the split
is achieved at the outlet of the vaporizing chamber, a flow
of saturated vapor is metered into the bypass flow and an
exit splitting ratio can be calculated. In Figure 3-8, to
create 1% sevoflurane a controlled flow of 100 mL/min of
21% sevoflurane is mixed with the bypass flow of 2000
mL/min. The ratio of flows is 2000/100 = 20:1. Compare
this with the effective flow split of 25:1 for the fresh gas
entering the vaporizer from the machine flowmeters.
Similarly, in Figure 3-9, to create 1% isoflurane a controlled flow of 100 mL/min of 31% isoflurane vapor is
mixed with the bypass flow of 3000 mL/min; the exit
splitting ratio is 3000/100 = 30:1. Compare this with the
effective flow split of 44:1 for the fresh gas entering the
vaporizer.
The location of the variable orifice is of little importance at 1 atm pressure but may have an effect when the
vaporizer is used at ambient pressures that are higher or
lower than 1 atm. (see later section on Effects of Changes
in Barometric Pressure).
72
3 Anesthesia Vaporizers
Bypass
valve
Filter
Check
valve
73
Temperaturecompensating
bypass valve
Concentration
control dial
(relocated for
illustration purposes)
Vaporizing chamber
Temperaturesensing bellows
Agent
FIGURE 3-11 n Schematic of a calibrated vaporizer. Temperature compensation is achieved by a gas-filled temperature-sensing bellows
that controls the size of a temperature-compensating bypass valve. Note also the check valve in the vaporizer outlet designed to
protect against the pumping effect. (Courtesy GE Healthcare, Waukesha, WI.)
TABLE 3-4 O
utput in Volumes Percent and
Minimum Alveolar Concentration
(MAC in Oxygen) of Erroneously
Filled Vaporizers at 22 C
Vaporizers
Liquid
Halothane
Halothane
Enflurane
Isoflurane
Enflurane
Isoflurane
Halothane
Isoflurane
Haloflurane
Enflurane
Enflurane
Isoflurane
Setting
(%)
1.0
1.0
1.0
2.0
2.0
2.0
1.5
1.5
1.5
Output
(%)
Output MAC
1.00
0.62
0.96
2.00
3.09
3.21
1.50
1.56
0.97
1.25
0.37
0.84
1.19
2.69
4.01
1.30
1.95
0.57
Bruce and Linde13 reported on the output of erroneously filled vaporizers (Table 3-4). Erroneous filling
affects the output concentration and, consequently, the
potency output of the vaporizer. In their study, an enflurane vaporizer set to 2% (1.19 MAC) but filled with halothane delivered 3.21% (4.01 MAC) halothane. This is 3.3
times the anticipated anesthetic potency output.
To summarize, if a vaporizer specific for an agent with
a low SVP, such as enflurane or sevoflurane, is misfilled
74
FIGURE 3-12 n Agent-specific filling devices for sevoflurane (yellow; Quik-Fil, Abbott Laboratories, Abbott Park, IL) and isoflurane (purple; Key-Fill, Harvard Apparatus, Holliston, MA).
Filling of Vaporizers
Vaporizers should be filled only in accordance with their
manufacturers instructions. Overfilling or tilting a vaporizerthat is, either tilting a freestanding unit or tilting
the whole anesthesia machinemay result in liquid agent
entering parts of the anesthesia delivery system designed
for gases and vapor only, such as a vaporizer bypass. This
could lead to the delivery of lethal concentrations of agent
to the patient circuit. If a vaporizer has been tilted, liquid
agent may have leaked into the gas delivery system. A
patient must never be left connected to such a system.
Once the machine has been withdrawn from clinical service, the proper procedure is to purge the vaporizer with
a high flow rate of oxygen from the flowmeter of the anesthesia machine or workstationnot the oxygen flush,
which bypasses the vaporizerand with the vaporizer
concentration dial set to the maximum concentration.12,15
The maximum concentration dial setting ensures the
highest possible oxygen flow through the inflow and outflow paths of the vaporizing chamber and through the
bypass. A calibrated anesthetic agent analyzer is essential
to check the efficacy of the flush procedure before the
vaporizer is returned to clinical service. In contemporary
practice, it is probably prudent to withdraw the workstation and vaporizer from clinical use until both have been
declared safe by authorized service personnel.
TABLE 3-5 Vaporizer Output After Incorrectly Refilling from 25% Full to 100% Full
Vaporizer Output
Halothane
Vaporizer
Setting (%)
Halothane
1.0
1.0
2.0
1.5
Enflurane
Isoflurane
Refill Liquid
Enflurane
Isoflurane
Halothane
Halothane
Enflurane
Isoflurane
MAC
MAC
MAC
Total MAC
0.33
0.41
2.43
1.28
0.41
0.51
3.03
1.60
0.64
0.96
0.38
0.57
0.90
0.57
0.78
0.50
0.79
1.29
3.60
2.10
3 Anesthesia Vaporizers
3 L/min
70% N2O, 30% O2
3 L/min
70% N2, 30% O2
3 L/min
70% N2O, 30% O2
% Enflurane
75
0
0
10
20
30
40
Time (min)
50
60
70
FIGURE 3-13 n Effect of changing the carrier gas composition (nitrous oxide vs. nitrogen) on vaporizer output in a variable bypass enflurane vaporizer. (From Scheller MS, Drummond JC: Solubility of N2O in volatile anesthetics contributes to vaporizer aberrancy when changing
carrier gases. Anesth Analg 1986;65:88-90. Reproduced by permission of the International Anesthesia Research Society.)
76
Bypass flow
25:1
2500
100
Flow to
vaporizing
chamber
47
68%
160/500 = 32%
100 mL = 68%
(100/68) 32 = 47 mL
Sevo% = 47/(47 + 2600) = 1.8% by volume
Psevo = 1.8% 500 = 9 mm Hg
Potency = 9/PMAC1sevo = 9/16 = 0.6 MAC
O2
N2O
Agent
FIGURE 3-14 n Use of a concentration-calibrated variable bypass sevoflurane vaporizer under hypobaric conditions. MAC, minimum
alveolar concentration; PB, barometric pressure; PMAC1, partial pressure of a potent inhaled agent at a concentration of 1 MAC.
Hypobaric Conditions
Few reports are available concerning the use of vaporizers
under hypobaric conditions.6 This discussion therefore
focuses on theoretical considerations applying to such use.
Consider a variable bypass vaporizer set to deliver 1%
sevoflurane (0.5 MAC at 760 mm Hg atmospheric pressure) at an ambient pressure of 500 mm Hg, equivalent to
an altitude of approximately 12,000 feet above sea level,
and at a temperature of 20 C (Fig. 3-14). In the vaporizing chamber, sevoflurane has an SVP of 160 mm Hg at
a temperature of 20 C; however, this now represents
32 vol% (160/500) of the atmosphere there. Set to deliver
1% under normal conditions, for the incoming fresh gas
the vaporizer creates a splitting ratio of 25:1 between
bypass and vaporizing chamber flows.
If the total gas flow to the vaporizer is 2600 mL/min
(Fig. 3-14), 100 mL/min of carrier gas flows through the
vaporizing chamber. This now represents 68% of the volume there because sevoflurane represents the other 32
vol% (100% 68%). Emerging from the vaporizing
chamber is 100 mL/min of carrier gas plus 47mL/min of
sevoflurane vapor ([100/68] 32). When the vaporizing
chamber and bypass flows merge, the 47 mL/min of sevoflurane vapor are diluted in a total volume of 2647 mL/
min (2500 + 100 + 47 mL), giving a sevoflurane concentration of 1.8% of the atmosphere by volume. This appears
to be almost twice (1.8 times) the dialed-in concentration
in terms of volumes percent.
To determine the potency of this 1.8% sevoflurane
concentration, however, partial pressure must be considered because it is the tension of the anesthetic agent that
determines potency. If sevoflurane represents 1.8% of the
gas mixture by volume, its partial pressure in the emerging
mixture is 1.8% 500 mm Hg, or 9 mm Hg. In terms of
anesthetic potency, this represents 0.6 MAC (9/16)
because the PMAC1 of sevoflurane is 16 mm Hg (see Table
3-1). Thus, in theory, a variable bypass sevoflurane vaporizer used at an ambient pressure of 500 mm Hg (at 12,000
feet above sea level) with the dial set to 1% (vol/vol)
3 Anesthesia Vaporizers
77
Bypass flow
25:1
2500
100
Flow to
vaporizing
chamber
93%
7.5
O2
N2O
160/2280 = 7%
100 mL = 89%
(100/93) 7 = 7.5 mL
Sevo% = 7.5/(7.5 + 2600) = 0.29%
Psevo = 0.29% 2280 = 6.6 mm Hg
Potency = 6.6/PMAC1sevo = 6.6/16 = 0.41 MAC
Agent
FIGURE 3-15 n Use of a concentration-calibrated variable bypass sevoflurane vaporizer under hyperbaric conditions. ATA, absolute
atmospheric pressure; MAC, minimum alveolar concentration; PB, barometric pressure; PMAC1, partial pressure of a potent inhaled
agent at a concentration of 1 MAC.
Arrangement of Vaporizers
Very old, now-obsolete models of anesthesia machines
had up to three variable bypass vaporizers arranged in
series, making it possible for carrier gas to pass through
each vaporizer, albeit all through a bypass chamber, to
reach the common gas outlet of the anesthesia machine.
Without an interlock system, which permits only one
vaporizer to be in use at any one time, it was possible to
have all three vaporizers turned on simultaneously. Apart
from potentially delivering an anesthetic overdose to the
patient, the agent from the upstream vaporizer could contaminate agents in any downstream vaporizer.16 During
subsequent use, the output of any downstream vaporizer
would be contaminated. The resulting concentrations in
the emerging gas and vapor mixture would be indeterminate and possibly lethal.
When such a serial arrangement of vaporizers was present, it was important to ensure that a vaporizer designed
for a less volatile agentsuch as methoxyflurane, with a
low SVPwas not placed downstream relative to more
volatile anesthetics, such as halothane. In such a case, halothane upstream would dissolve in the less volatile agent
downstream. If during subsequent use the methoxyflurane
78
FIGURE 3-16 n Freestanding vaporizer mounted on a pump oxygenator. Note the warning label on the mounting bracket in A, which
reads, This bracket is to be used only in conjunction with cardiopulmonary bypass equipment. Read instructions for warnings and
caution.
3 Anesthesia Vaporizers
As previously noted, changing the composition of the carrier gas, especially by adding or removing nitrous oxide,
temporarily alters vaporizer output.17 In the case of nitrous
oxide, this effect is mainly due to solubility of nitrous
oxide, although changes in gas density and viscosity also
contribute to changes in output by affecting the flow split
between vaporizing chamber and bypass gas flows.
79
Concentration
set on Vapor 19.1
3.5
3.5
2.5
2.5
1.5
1.5
1
0.8
0.6
0.4
0.2
0
1
0.8
0.6
0.4
0.2
0
10
11
12
13
14
15
80
Effects of Temperature
As previously described, concentration-calibrated vaporizers are temperature compensated. The effects of changes
in temperature are negligible under usual conditions of
use. However, if the ambient temperature exceeds the
vaporizers specified range of performance, the vaporizers output may become high. For example, the Drger
Vapor 19.n vaporizer is specified as accurate between 15 C
and 35 C.12 This is because as temperature increases, the
vapor pressure of the anesthetic increases in a nonlinear
manner, whereas the temperature compensation is linear.
Indeed, the output may become unpredictable and uncontrolled if the boiling point of the agent is reached. Conversely, if the temperature falls below the specified range
for use, the output may be unpredictably low.
Fluctuating Backpressure
A problem with older model vaporizers was that fluctuating or intermittent backpressure, the so-called pumping
effect, may be applied to the vaporizer by changes in pressure downstream. Such pressure changes could be caused
by intermittent positive-pressure ventilation (IPPV) in
the patient circuit or operation of the oxygen flush control. These changes could produce changes in gas flow
distribution within the vaporizer, which could lead to
increased output if no compensatory mechanism existed.
The effects of intermittent pressurization are greatest at
low flow rates, low concentration settings, when small
amounts of liquid agent are present in the vaporizer, and
with large and rapid changes in pressure. Higher flow
rates, higher dial settings, larger volumes of liquid agent
in the vaporizer, and smaller, less frequent changes in
pressure minimize the pumping effect.
Of the several explanations offered for the pumping
effect, the most probable is that during pressurization, gas is
compressed in the vaporizer in both the vaporizing chamber and the bypass. When the pressure decreases, anesthetic
vapor leaves the vaporizing chamber, through both the
normal exit pathway and the vaporizing chamber inlet, to
enter the bypass flow. The intermittent addition of vapor
to the bypass flow results in the observed intermittent
increase in vaporizer output.
Contemporary vaporizers incorporate certain design
features that minimize the significance of the pumping
FIGURE 3-18 n Drger Vapor 19.1 vaporizers (Drger Medical, Telford, PA). Note the funnel-fill design, which has been superseded by an agent-specific key-fill design.
effect (discussed in the following section). Some older models, such as the Ohio Medical calibrated vaporizers, have a
check valve in the vaporizer outlet to prevent transmission
of increases in downstream pressure (see Fig. 3-11).8
Certain models of GE anesthesia machines use a check
valve just upstream of the common gas outlet to prevent
retrograde transmission of downstream pressures (see
Chapter2). However, while the check valve is closed, the
pressure upstream of itand hence, that in the vaporizers
increases because of continuous fresh gas flow from the
machine flowmeters. Thus the use of check valves can
limit, but not totally eliminate, the pumping effect. For
this reason, the newest anesthesia machine models do not
use check valves in their design (see Chapter 2).
Contemporary Vaporizers
Drger Vapor 19.n
The Drger Vapor 19.n vaporizer is used on many contemporary Draeger Medical anesthesia workstations (Fig. 3-18).
Up to three such vaporizers may be mounted on the back
bar of the workstation and are connected to an interlock
system designed to ensure that only one vaporizer or agent
is in use at any time. When one vaporizer is turned on, the
dials on the others cannot be turned from zero. Because failures have been reported,21 the interlock systems function
must be checked periodically (see Chapter 30).
A schematic of the Drger Vapor 19.1 vaporizer is
shown in Figure 3-19.11 Its operation is fairly straightforward. With the concentration knob in the zero position, the on/off switch is closed. Fresh gas enters the
vaporizer at the fresh gas inlet and leaves via the fresh
gas outlet without entering the vaporizing parts of the
unit. In this off state, the inlet and outlet ports of the
vaporizing chamber are connected and vented to the
atmosphere by a hole. The venting prevents pressure
buildup in the vaporizing chamber, thereby preventing
vapor from being driven under pressure into the fresh
gas flow. Drger specifies that venting results in a loss of
only 0.5 mL/day of anesthetic into the operating room
3 Anesthesia Vaporizers
3
10
7
9
8
5
= O2
= N2O
= Anesthetic agent
81
82
FIGURE 3-20 n Drger Vapor 2000 vaporizers (Drger Medical, Telford, PA). The white circle indicates transport mode (T) on the
concentration dial.
activated (Fig. 3-22). Once the lever is in the locked position, the dial release can be depressed. This operates the
vaporizer interlock mechanism, which causes the interlock
extension rods to extend laterally to adjacent vaporizers,
minimizing the possibility of them being turned on.
Simultaneously, the two Select-a-tec port valve actuating
spindles are activated to allow fresh gas to enter the vaporizer. Depressing the control dial release button also enables
the control dial to be turned to the desired vapor output
concentration. When the control dial is turned off and the
dial release is no longer depressed, the manifold port valves
close and the extension rods are retracted to allow selection of another vaporizer. Thus, in the Select-a-tec system, fresh gas enters a vaporizer only when the vaporizer
is turned on; otherwise, fresh gas bypasses the vaporizer(s)
via a separate channel in the manifold (Fig. 3-23).
Figure 3-24, A, shows a Tec 5 vaporizer flow diagram, and Figure 3-24, B, shows a schematic of the Tec
5. When the concentration dial is set in the zero position, all gas passages are closed except for a channel linking the vaporizer inlet and outlet. As shown in Figure
3-24, B, when the dial is turned past 0%, the carrier gas
stream is split between bypass and vaporizing chamber
flows. Bypass gas flows vertically downward from a,
across the base of the sump, through the thermostat to c,
and back up the gas transfer manifold via d to e. The
thermostat or temperature-compensating device is
located in the base of the vaporizer. It is a bimetallic
strip design, which increases bypass flow as temperature
rises and decreases bypass flow as temperature falls.
The Tec 5 incorporates an IPPV chamber to minimize
the pumping effect. The fresh gas flowing to the vaporizing
chamber flows through an IPPV assembly designed to minimize the pumping effect before it reaches the vaporizing
chamber and wick assembly system. There the gas becomes
saturated with anesthetic vapor and flows onward to combine with the bypass gas flow and exit the vaporizer.
In the Tec 5 vaporizer, temperature compensation is
achieved by a bimetallic strip valve rather than a bypass
cone and expansion element. The vaporizer incorporates
a keyed fill system, but some older models may still have
a funnel-fill design. Although the Tec 5 incorporates an
antispill mechanism, if the vaporizer is inverted, it is recommended that it be purged with carrier gas at 5 L/min
for 30 minutes with the dial set to 5%.18
The Tec 5 vaporizer has a liquid agent capacity of
300mL with dry wicks and 225 mL with wet wicks. The
vaporizer is calibrated at 22 C with oxygen at 5 L/min.
As previously discussed, changes in carrier gas composition may affect agent output concentration.
The operators manual recommends that the vaporizer
be serviced every 3 years at an authorized service center.18
Service should include complete disassembly; thorough
cleaning; inspection for wear and damage; renewal of
wicks, seals, and any worn components; replacement of
discontinued parts with more current parts; checking
ofoutput; and recalibration if necessary.
Tec 7
The Tec 7 (see Fig. 3-25) is essentially the same as the Tec
5 (see Fig. 3-21) but has an improved ergonomic design, a
300-mL capacity for liquid agent, and an improved
3 Anesthesia Vaporizers
Control dial
release
83
Percentage
control dial
Select-a-tec
locking lever
Port valve
actuating
spindle
Extension
rod
Port
valves
Port
valves
Select-a-tec
SM manifold
FIGURE 3-22 n GE Datex-Ohmeda Select-a-tec vaporizer interlock system. SM, series-mounted manifold. (Courtesy GE Healthcare, Waukesha, WI.)
Vaporizer B
(isolated)
Vaporizer A
From
flowmeters
Manifold
bypass
To common
gas outlet
FIGURE 3-23 n GE Datex-Ohmeda series-mounted manifold gas circuit. Fresh gas from the machine flowmeters enters the manifolds,
which incorporate pairs of series-connected, two-way port valves. When a Tec 5 vaporizer is locked onto the manifold and turned on
(vaporizer A), both associated port valves are opened. Fresh gas from the manifold then flows into the vaporizer through the inlet port
valves, and the gas-agent mixture exits via the outlet port valve. When the vaporizer is turned off (vaporizer B), or if no vaporizer is
fitted to the manifold, each port valve is closed to allow gas to bypass the vaporizer via the manifold bypass circuit. (Courtesy GE
Healthcare, Waukesha, WI.)
Desflurane Vaporizers
Tec 6
Desflurane (Suprane, Baxter Health Care Products, Deerfield, IL) is a potent, inhaled, volatile anesthetic approved
for use by the U.S. Food and Drug Administration (FDA)
in 1992. The physical properties differ considerably from
those of other agents in clinical use (see Table 3-2). With
an SVP of 669 mm Hg at 20 C and a boiling point of
Rotary
valve
Shut-off
open
Flow control
(vapor
channel)
10
Vaporizing
chamber
From
flowmeter
To common
gas outlet
10
1
c
Thermostat
11
A
Rotary
valve
On
Enriched
fresh
gas out
Vapor
control
channel
11
e
Combined
fresh gas
and
enriched
gas out
2
10
Fresh gas
bypass
Fresh
gas out
7
5
Wick
assembly
8
Vaporizing
chamber
Thermostat
FIGURE 3-24 n A, GE Datex-Ohmeda Tec 5 vaporizer flow diagram. B, Schematic of Tec 5 vaporizer. Gas flow enters the vaporizer (1) and
is split into two streams: the bypass circuit and the vaporizing chamber. Gas flows downward through the bypass circuit from a,
across the sump base b, through the thermostat to c, and back up the gas transfer manifold via d to e. Gas flowing to the vaporizing
chamber flows from 1 across the sump cover (2), where it is diverted via 3 through the central cavity of the rotary valve and back
through the intermittent positive pressure ventilation (IPPV) assembly via 4, 5, and 6. Gas then flows from the IPPV assembly via 7
down the tubular wick assembly, where vapor is added; it then flows across the base of the vaporizing chamber above the liquid agent
to 8. From here the gas-vapor mixture flows via 9 through the sump cover to the proportional radial drug control groove of the rotary
valve and back into the sump cover (10), where it merges with gas from the bypass circuit. The total flow then exits the vaporizer into
the outlet port of the Select-a-tec manifold. (Courtesy GE Healthcare, Waukesha, WI.)
3 Anesthesia Vaporizers
85
FIGURE 3-26 n A, GE Healthcare (Waukesha, WI) Tec 6 vaporizer for desflurane. B, Close-up of front panel shows status lights and display for
agent fill status.
86
Sump
shut-off
valve
Concentration dial
variable restrictor
FlowDes
Des vapor
1500 mm Hg
39 C
PDes
RDes
Des liquid
heated
Pressure
transducer
Control
electronics
RMain
Gas flowing
from machine
flowmeters
FlowMain
PMain
Fixed restrictor
FIGURE 3-27 n Simplified schematic of the Datex-Ohmeda Tec 6 vaporizer (GE Healthcare, Waukesha, WI). Liquid desflurane (Des) is
heated in the sump to 39 C to produce vapor under pressure (~1500 mm Hg). The variable pressure control valve is continuously
adjusted by an output from the differential pressure transducer to ensure that the pressure of the desflurane vapor upstream of the
rotary valve (concentration dial variable restrictor) is the same as the pressure of the fresh gas inflow to the fixed restrictor. The concentration dial and rotary valve control the quantity of desflurane vapor added to the fresh gas flow so that what emerges from the vaporizer
outlet is the dialed-in concentration of desflurane. No fresh gas enters the desflurane vaporizing chamber; this is in contrast to the other
Tec series vaporizers, which are of the variable bypass design. FlowDes, flow of desflurane vapor; FlowMain, gas flow from the machine
flowmeters (O2, N2O, air); PDes, pressure of desflurane vapor flowing into concentration dial variable restrictor; PMain, pressure of gas
flowing into fixed restrictor; RDes, resistance to flow of desflurane created by variable restrictor; RMain, resistance of fixed restrictor.
Flow =
( P r4 )
Resistance
Pressure
Flow
(3-1)
Pressure (main)
Flow (main)
or
Pressure (main)
Flow (main)
Resistance (main)
(3-2)
Concentration (des)
Flow (des)
Flow (main)
(3-4)
Pressure (des) K
Pressure (main) Resistance (des)
Resistance (main)
Pressure (des)
Resistance (des)
Flow (des)
(3-3)
Concentration (des)
Flow (main) Flow (des)
Flow (des)
(8 L)
Concentration (des)
Flow (des)
Flow (main)
(3-5)
1
(approximately).
Resistance (des)
3 Anesthesia Vaporizers
87
1
27
26
Gas/agent
outlet
25
Max
7
24
18 17 16
Min
19
Key
Fresh gas
15
Agent vapor
23
Gas/agent vapor
9
22
Electrical connections
14
10
21
12
11
20
13
FIGURE 3-28 n Detailed schematic of the GE Datex-Ohmeda Tec 6 desflurane vaporizer. 1, Concentration dial and rotary valve; 2, fixed
restrictor; 3, pressure transducer; 4, pressure monitor; 5, heater to prevent condensation of desflurane; 6, vapor control manifold
assembly; 7, pressure regulating valve; 8, shut-off valve; 9, sump assembly; 10, liquid desflurane; 11, level sensor; 12, sump heater to
heat desflurane to 39 C; 13, power cord; 14, backup battery for alarms; 15, power supply; 16, control electronics printed circuit board;
17, heater electronics; 18, alarm electronics circuit board; 19, liquid crystal sump agent level display; 20, alarm battery low lightemitting diode (LED); 21, warm up/operational LED; 22, low agent warning LED; 23, no output warning LED; 24, operational
LED; 25, tilt switch (shuts down vaporizer if tilted excessively); 26, solenoid dial lock; 27, heater to prevent condensation of desflurane
in valve plate (From Ohmeda Tec 6 vaporizer operation and maintenance manual. Madison, WI, 1992. Courtesy GE Healthcare, Waukesha, WI.)
Concentration (des)
Thus, at a 5 L/min flow of O2 set to deliver 10% desflurane, the Tec 6 adds 556 mL/min of desflurane vapor to
the main gas flow; in effect,
556
556
0.10 10%
5000 556 5556
(3-7)
Once the main flow, in this case 5 L/min O2, and the
desflurane flow (calculated as above) are known, the ratio
of resistances of the variable resistor, Resistance (des) in
the concentration dial, to the fixed resistor, Resistance
(main) in the main gas flow, may be calculated. Thus,
Pressure (des)
Flow (des)
Pressure (main)
Resistance (main)
Flow (main)
Resistance (des)
(3-8)
88
(3-9)
Flow
P r4
8L
(3-11)
Pressure
, or Pressure Flow Viscosity
Viscosity
(3-12)
3 Anesthesia Vaporizers
89
90
adequate desflurane delivery, at all times. In some operational environments, the temperature of the vaporizer in
general ensures that the minimum pressure is achieved in
the reservoir without assistance from the pump.
Agent Vaporization. When the vaporizer is required to
deliver desflurane after the initial start-up period, the liquid on/off valve is opened, and liquid desflurane flows into
the heater chamber and vaporizes rapidly, raising the pressure in the heater chamber to at least that of the reservoir
pressure to prevent further delivery of liquid desflurane. If
the vapor pressure in the heater chamber falls, more liquid
desflurane is forced into the heater chamber from the reservoir, allowing more vaporization of the liquid desflurane
and raising the pressure again. If pressure exceeds that in
the reservoir, the pressure is vented back to the reservoir
through the liquid desflurane delivery line, either as vapor
or as a condensed liquid.
Once running, this pressure system is self-controlling
because the heater chamber temperature, and hence
vapor pressure, is kept constant.
Agent Delivery. When in use, the required desflurane
concentration is set using the rotary control dial on the
front of the vaporizer. The user selects a desired concentration, and the vaporizer inlet and outlet fresh gas flow
sensors monitor the fresh gas flow and N2O concentration. The microprocessor then notes the pressure of the
vaporized desflurane and the fresh gas flow rate and sets
the calibrated delivery proportional valve to the correct
opening to deliver the dialed-in desflurane concentration
for that fresh gas flow. The vaporizer delivers desflurane
vapor into the fresh gas flow at the required flow rate for
Gas flow
sensor #1
Mixing
volume
Proportional
control valve
Flow
sensor
Heater and
wick device
On/off
valve (vapor)
Control dial
On/off
valve (liquid)
Agent
reservoir
Gas flow
sensor #2
Air pump
Pressure sensor
Liquid level sensor
Electronic control
PCB
(microcontroller)
Membrane
On/off
switch
Universal
power
supply
Backup
battery
LCD
display
Audible
alarm
Agent
filler
FIGURE 3-31 n Schematic of Penlon Sigma Alpha desflurane vaporizer to show principles of operation. LCD, liquid crystal display; PCB,
printed circuit board. (Courtesy Penlon, Abingdon, United Kingdom.)
3 Anesthesia Vaporizers
91
FIGURE 3-32 n Aladin vaporizer system (GE Healthcare, Waukesha, WI) used on the AS/3 Anesthesia Delivery Unit and Aisys
workstations.
92
FIGURE 3-33 n Aladin vaporizer system (GE Healthcare, Waukesha, WI) cassette for desflurane.
FIGURE 3-34 n Aladin vaporizer system (GE Healthcare, Waukesha, WI) isoflurane cassette. The workstation recognizes the
agent-specific cassette by the arrangement of identification (signature) magnets (see Fig. 3-35).
3 Anesthesia Vaporizers
93
Pins
Inlet valve
Outlet valve
Temperature
sensor
Baffles
Identification
magnet
positions
Locking
mechanism
Handle
Fill system
Liquid level
window
Vaporizing chamber
Baffles
Fill
port
Pin
Agent
FIGURE 3-35 n Schematic of Aladin vaporizer system cassette, which is the equivalent of the vaporizing chamber in a variable bypass
vaporizer. Fresh gas flows over wicks and baffles containing liquid anesthetic agent. (Courtesy GE Healthcare, Waukesha, WI.)
F=S/[R(1-S) + 1]
Bypass flow
measurement unit
Flow
restrictor
Mixing
chamber
Flow to fresh
gas outlet
Cassette flow
measurement unit
One-way
valve
Agent
proportional
valve
CPU
Cassette inflow
close valve
Cassette inflow
Cassette outflow
close valve
Cassette
pressure
sensor
Cassette
ID board
Temperature
sensor
FIGURE 3-36 n Schematic of Aladin
vaporizer system to show principles
of operation. CPU, central processing
unit; ID, identification. (Courtesy GE
Healthcare, Waukesha, WI.)
Liquid level
measurement board
Liquid
chamber
REFERENCES
1. Hill DW: Physics applied to anaesthesia ed 4, Boston, 1980, Butterworth, p 220.
2. Minimum performance and safety requirements for components and systems of anesthesia gas machines, ASTM F1161-88, Philadelphia,
1989, American Society for Testing and Materials.
3. Standard specification for particular requirements for anesthesia workstations and their components, ASTM F-1850-00, West Conshohocken,
PA, 2000, American Society for Testing and Materials.
4. Parbrook GD, Davis PD, Parbrook EO: Basic physics and measurement
in anesthesia, ed 4, Oxford, UK, 1995, Butterworth-Heinemann.
5. Fink BR: How much anesthetic? Anesthesiology 34:403404, 1971.
6. James MFM, White JF: Anesthesia considerations at moderate altitude, Anesth Analg 63:10971105, 1984.
7. Keenan RL: Volatile agent overdose is potential cause of catastrophe,
Anesthesia Patient Safety Foundation (ASPF) Newsletter 3(2):13, 1988.
8. Operations and maintenance manual, Ohio calibrated vaporizers for
ethrane, halothane and forane. Madison, WI, 1980, Ohmeda (formerly Ohio Medical Products).
9. Understanding the Tec 4 vaporizer, Steeton, UK, 1985, Ohmeda.
10. Drger Vapor 19.1 operating instructions. Lbeck, Germany, 1992,
Drgerwerk AG.
11. Drger Vapor 2000 Anaesthetic Vaporizer: instructions for use, ed 2,
Lbeck, Germany, 1998, Drger Medical.
12. Drger Vapor 19.n Anaesthetic Vaporizer, ed 16, Lbeck, Germany,
1991, Drger Medical.
13. Bruce DL, Linde HW: Vaporization of mixed anesthesia liquids,
Anesthesiology 60:342346, 1984.
14. Korman B, Ritchie IM: Chemistry of halothane-enflurane mixtures applied to anesthesia, Anesthesiology 63:152156, 1985.
Fan
Aladin
cassette