19. Set the number of scans on the data acquisition menu on the computer. Then
select acquire data.
20. To interface the control panel with the computer, select SLOW button under
ACCUMULATE on the control panel. Then start the scan by pressing the button
with single forward arrow on it (red light).
21. Remove the sample from the cavity. Bring the ATTENUATOR back up to 60
dB or greater and place the instrument in STAND-BY mode. Remove the sample.
If a reference scan is desired put in a reference sample. Return the instrument to
the OPERATE MODE and bring the ATTENUATOR back down to 20 dB. Set
the GAIN to its lowest setting. REMEMBER DO NOT CHANGE THE
FREQUENCY. (this would invalidate the reference.) The iris can be adjusted if
needed.
22. Shutting down the instrument. Turn the GAIN all the way down and the
MODULATION AMPLITUDE off. Put the ATTENUATOR at 60 dB or above
and put the instrument in the STAND-BY mode. Turn off the console and the
magnet. TURN OFF THE NITROGEN!!