P UBLICATION D ATE
100-044-406 A
March 2001
Addendum
493.07 Pump Interface
! CAUTION
You must have a 493.07 Hydraulic Pump Interface designed for the
correct voltage before installation.
Connecting an interface for 24 Vdc to a 115 Vac pump causes
improper operation.
The design voltage is marked on the cover of the 493.07 Hydraulic Pump
Interface. Make sure that voltage is correct before installation.
E14 E10
E15 E11
J25
J1
To 407
Remote
E-stop and HPS
Connectors
To HPS
J2
J3
To 407
Digital I/O for Crosshead
Interlock
To Remote
Emergency
Stop
E16 E12
Compatible with:
458.05/.10/.20/.40,
490.01,
497.05
413.05
436.11
407.05
Jumpers
Standard jumper setting:
E10E11:
E12E13
E14E15
E16E17
remove
remove
remove
remove
E10E13
E14E17
install
install
remove
remove
E10E11:
E12E13
E14E15
E16E17
install
install
install
install
Remote
Emergency Stop
(Optional)
501821-XX*
J2
Crosshead
Unlock
E-stop
J3
493.07
Pump
Interface
407 Controller
HPS
J25
HPS
397087-XX (for 24 V)
397088-XX (for 115 V)
J1
HPS
501822-XX
E-stop
407 Controller
HPS
J2
Crosshead
Unlock
501822-XX
J3
493.07
Pump
Interface
E-stop
HPS
J25
HPS
397087-XX (for 24 V)
397088-XX (for 115 V)
J1
HPS
407 Controller
501821-XX*
501822-XX
407 Controller
501822-XX
J1
HPS
Load Frame
J3
493.07
Pump
Interface
E-stop
HPS
J2
Crosshead
Unlock
491091-XX*
J25
397087-XX (for 24 V)
397088-XX (for 115 V)
HPS
491091-XX*
Digital I/O
500538-XX
407 Controller
J2
Crosshead
Unlock
HPS
J3
493.07
Pump
Interface
E-stop
501822-XX
J25
500538-XX
407 Controller
J2
Crosshead
Unlock
HPS Y Cable
054078-01 (for 24 V)
054018-01 (for 115 V)
E-stop
HPS
J25
493.07
Pump
Interface
501822-XX
HPS
J1
HPS
397087-XX (for 24 V)
397088-XX (for 115 V)
Digital I/O
Load Frame
J3
491091-XX*
Load Frame
J1
HPS
*Jumper plug 493372-01
required if cable is not used on
493.07 J3 connector
Compatible Controllers
If the 407 Controller is connected to an HPS in tandem with compatible
controllers, the following configuration is used. Notice that the 493.07
jumper setting must be correct for compatibility (See Setting Jumpers on
the 493.07 earlier in this addendum).
491091-XX*
Digital I/O
500538-XX
407 Controller
J2
Crosshead
Unlock
HPS
J3
493.07
Pump
Interface
E-stop
501822-XX
J25
HPS
J1
HPS
397087-XX (for 24 V)
397088-XX (for 115 V)
Load Frame
458 or
490.01 or
497.05
HPS Y Cable
054078-01 (for 24 V)
054018-01 (for 115 V)
397087-XX (for 24 V)
397088-XX (for 115 V)
11
14
12
Definition
24 V out
HPS start in
HPS low in
HPS high in
E-stop (+)
E-stop ()
nc
nc
HPS on sense out
Common
HPS overtemp interlock out
HPS low level interlock out
nc
Shield
J2 Controller Interlock
J2
Controller
Interlock
1
3
2
4
Definition
nc
Digital input ()
nc
Digital output (+)
4
7
3
6
Definition
Low level contact ()
Crosshead unlock contact ()
Chassis ground
Crosshead unlock contact (+)
High level contact ()
Low level contact (+)
High level contact (+)
4
9
15
10
16
20
21
24
Definition
nc
Overtemperature (HPS error)
Low level
Protective earth
nc
nc
Interlock common
nc
Power from pump
Low
Start
E-stop chain
nc
nc
High
E-stop chain
Common
Defeat common
Defeat power
nc
nc
nc
nc
nc