TECHNICAL
MANUAL
1 Overview ........................................................................................................... 7
1.1 General .................................................................................................... 7
1.2 Overview diagram ................................................................................. 8
1.3 Precautions to be taken before use...................................................... 9
1.4 Internal safety features .......................................................................... 9
1.5 Technical characteristics ...................................................................... 9
1.5.1
1.5.2
1.5.3
1.5.4
PiloteCE2_A2_enTDM.fm
4 Preventive maintenance................................................................................. 37
4.1 Recommendations............................................................................... 37
4.2 Maintenance schedule......................................................................... 37
4.2.1
4.2.2
5 Diagnosis......................................................................................................... 65
5.1 Troubleshooting................................................................................... 65
5.2 Error messages.................................................................................... 67
44
PiloteCE2_A2_enTDM.fm
7 Calibration..................................................................................................... 121
7.1 Calibration procedure ........................................................................ 121
7.1.1
7.1.2
7.1.3
PiloteCE2_A2_enTDM.fm
66
PiloteCE2_A2_enTDM.fm
1 Overview
1.1 General
The Pilot A2 is a syringe pump intended for the infusion of intravenous agents at a accurate,
low flow rate. The wide choice of syringes, the use of the universally recognised control
symbols and the synoptic display of the alarms contribute to making the Pilot A2 easy to use.
The adjustable occlusion detection, the correct positioning detection and the overall syringe
protection system guarantee optimum safety.
Overview
Its technical characteristics, the flow range from 0.1 to 400 ml/hr (configured at
200 ml/hr) and its excellent accuracy (+1% on the device) contribute to makin the Pilot A2,
the ideal instrument for medical services.
01_002a_en.fm
Ext
12/15 V
15 W
Power supply
230 V
ON / OFF
DC-DC
converter
Motor
driver
Step by
step
motor
Battery
EPROM
128K x 8
RAM
8K x8
Motor speed
sensor
EEPROM
512
Syringue body
sensor
BUS
SPI
Antisiphon
sensor
CPU
Watch
dog
UART
Displacement
transducer
Opto
interface
Interface
bus
Nurse call
(option)
ADC
Occlusion
switch
Disengagement
switch
RS 232
LCD
driver
Keyboard
LCD
display
Fin switch
Master
LED
driver
LED
display
Overview
Buzzer
01_002a_en.fm
Overview
01_002a_en.fm
Overview
n In compliance with the European Directive 93/42 EEC related to Medical Equipment:
CE0459.
n In compliance with the European Directive 89/336 EEC: Electromagnetic compatibility.
n Compliant with the standards EN 60601.1 and PrEN60601-2-24.
n Protection against leakage currents: CF type.
n Protection against electric shock: Class II.
n Protection against splashing liquid: IP34.
10
01_002a_en.fm
Anti-siphon arm
Plunger disengagement
control
Syringe clamp
Upper case
Lower case
02_002a_en.fm
11
n Keyboard interface.
n Control lamps and overview diagrams.
n 7-segment display units.
J1
J3
Solder side display board.
J2
Pin
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
12
Description
SEG1 display matrix
SEG2 display matrix
SEG3 display matrix
SEG4 display matrix
SEG5 display matrix
SEG6 display matrix
SEG7 display matrix
SEG8 display matrix
COL1 display matrix
COL2 display matrix
COL3 display matrix
FAIL LED control
COL/DIG 9 LED type control
LIG1 keyboard interface
LIG2 keyboard interface
LIG3 keyboard interface
LDSECT lighting control
+5V power supply
VBAT power supply
GND power supply
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Line 7
Line 8
Column 1
Column 2
Column 3
Fail
"
Line 1
Line 2
Line 3
Mains LED
02_002a_en.fm
J2 connector to keyboard
Pin
1
2
3
4
5
6
7
8
9
10
11
12
Description
Column 1
Column 2
Column 3
Column 4
Column 5
Column 6
Line 1
Line 2
Line 3
Ton
Toff
Gnd power supply
Pin
1
2
3
4
5
6
7
8
02_002a_en.fm
13
J2
J4
J3
J5
CPU board
J2 connector: to fin de18
tection switch and syringe detection opto-electronic sensor
Pin
1
2
3
4
5
6
Description
Ground
Fin contact
Opto anode diode +5V
Common points between cathode LED , opto 1 and opto 2 transistor emitters.
Opto 1 transistor collector
Opto 2 transistor collector
Pin
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
14
Description
+5V controlled power supply
Gnd power supply
+Vbat power supply
Gnd power supply
Phase A motor control
Phase B motor control
Phase C motor control
Phase D motor control
I signal motor control
Boost signal motor control
Sopt1 opto rotation motor output
Sopt2 opto anti-siphon
Apinf nurse call independent of the buzzer signal
Cdopt1 opto rotation motor control output
Cdopt2 opto anti-siphon module control
Off signal off key pressed
Sect mains power supply on signal
Cdalim power cut signal
02_002a_en.fm
Description
Ldsect mains LED control
Cts clear to send
Deb/off disengagement signal
Rts request to send
Occ/off occlusion signal
Buz nurse call relay control
Eoc end of adc conversion
Csadc selection spi adc bus
Clk clock spi adc bus
Si data in spi adc bus
So data out spi adc bus
Cdana analogue sensor power supply control
Rx2 receive data TTL
Tx2 transmit data TTL
Txd1 transmit data TTL
Rxd1 receive data TTL
Ton ON key
Toff OFF key
+Vbat power supply
Gnd
+5V
Gnd
Line 2
Line 2
Line 2
Line 2
Line 1
Line 1
J4 connector to keyboard
02_002a_en.fm
Description
Seg1 display matrix
Seg2 display matrix
Seg3 display matrix
Seg4 display matrix
Seg5 display matrix
Seg6 display matrix
Seg7 display matrix
Seg8 display matrix
Col1 display matrix and keyboard
Col2 display matrix and keyboard
Col3 display matrix and keyboard
Fail diode fail control
Rdcrt current reduction control
Lig1 keyboard interface
Lig2 keyboard interface
Lig3 keyboard interface
Ldsect mains LED control
+5V power supply
Vbat power supply
Gnd power supply
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Line 7
Line 8
Column 1
Column 2
Column 3
Line 1
Line 2
Line 3
Pin
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
15
Pin
1
2
3
4
5
6
7
8
16
02_002a_en.fm
J8
J5
J9
F2
J3
J4
J2
J1
F1
Description
Neutral
Phase
02_002a_en.fm
Description
+5V controlled power supply
Gnd power supply
+Vbat power supply
Gnd power supply
Phase A motor control
Phase B motor control
Phase C motor control
Phase D motor control
I signal motor control
BOOST signal
Sopt1 opto rotation module output
Sopt2 opto anti-siphon module output
N.U
Cdopt1 opto rotation module control
Cdopt2 opto anti-siphon module control
Off off key pressed on the ON/OFF button
SECT mains supply presence signal
CDALIM power cut signal
LDSECT mains LED control
CTS clear to send
DEB/OFF disengagement signal active at 0
Pin
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
17
Description
RTS request to send
OCC/OFF occlusion signal active at 0
BUZ nurse call relay control
EOC end of ADC conversion
CSADC selection bus SPI ADC
CLK clock bus SPI ADC
SI data IN bus SPI ADC
SO data out bus SPI ADC
CDANA analogue sensor power supply control
RX2 receive data TTL
TX2 transmit data TTL
TXD1 transmit data TTL
RXD1 receive data TTL
Toff OFF key
Ton ON key
+Vbat power supply
Gnd
+5V
Gnd
Line 2
Line 2
Line 1
Line 1
J3 connector to potentiometer
Pin
1
2
3
Description
Vref
Centre point
Gnd
Description
+ battery
- battery
J5 connector to motor
Pin
1
2
3
4
5
6
7
8
9
10
18
Description
+Vbay
+Vbat
Phase D
Phase C
Phase B
Phase A
Opto rotation anode diode /+5V
Opto rotation cathode diode
Opto rotation transistor collector
GND/ opto rotation transistor emitter
02_002a_en.fm
Description
TX1 transmit data
+5V
RX1 receive data
Gnd
Interface validation
Nurse call relay common point
Nurse call relay normally open
Nurse call relay normally closed
CD ON external on
CD OFF external off
I-OPTON motor control output
I-SECT mains led
+Vbat external power supply plug
RX2 receive data
TX2 receive data
Gnd
CTS
RTS
BUZ
NC
Line 1
Line 1
Line 2
Line 2
Description
external power
supply
external power
supply
Pin
1
2
3
4
5
6
7
8
9
10
Description
Not used
Micro-switch input/output
Micro-switch input/output
Not used
Opto anti-siphon cathode diode
Opto anti-siphon anode diode/+5V
Opto anti-siphon transistor collector
Disengagement micro-switch on
Disengagement micro-switch off
Gnd
Do not forget to dismount the ribbon cable holder on the power supply board before
extracting the mechanical assembly from the housing (risk of breaking the ribbon
cable).
02_002a_en.fm
19
Description
GND
Position sensor output
Battery discharge control output
Amplified force sensor output
Cd coupler power supply 0-5V
Motor control opto output
Force and position sensor reference voltage
Piston head detection opto output
Control/APIN F
Control/APIN F
The plunger is fitted with a disengagement control allowing to separate this from the screwand-nut system.
20
02_002a_en.fm
02_002a_en.fm
21
22
02_002a_en.fm
It is possible to exit the configuration mode at any time by pressing the OFF key.
n
n
n
n
n
n
n
n
n
n
n
n
n
03.2_002a_en.fm
23
Function
ON, is used to switch the machine ON.
OFF, is used to switch the machine off and, when pressed for over three
seconds, to exit the configuration mode.
SILENCE ALARM, is used to access the configuration mode of the
current operation parameters.
The selection keys allow to scroll the figures and letters on the tenths,
units, tens segments etc.
PArJ
is carried out
24
03.2_002a_en.fm
may be selected.
If SEL4, when the Pilot is switched on, the user should
select the type of syringe installed.
When mode SEL3 is selected, and if there is a choice of more than one syringe, the
Pilot automatically moves onto the configuration of the list of syringes that may be
selected PAr4 when the machine is next switched on.
3.1.4 PAr3, configuration of the maximum flow rate than may be selected on the keyboard
This configuration enables users to choose the maximum flow rate that may be selected using
the keyboard for each type of syringe.
Syringe type
50/60 cc
0,1
400
20 cc
0,1
275
03.2_002a_en.fm
50c, 50 ml.
25
Press "CONFIRM"
26
03.2_002a_en.fm
PAr5 is displayed.
o
Press "CONFIRM"
s If PurG, compulsory priming, pressing "BOLUS"
during start-up is compulsory to switch to selection of
the flow rate.
s If noPu, priming is not compulsory, the flow rate
may be selected upon start-up straight after validation
of the syringe.
1,0 ml/hr (or at the same flow rate if this is under 1,0 ml/hr)
when the volume infused reaches the limit volume.
If noKV, no KVO flow, infusion stops with a limit volume
alarm when the infused volume reaches the limit volume.
03.2_002a_en.fm
27
It is possible to exit the configuration mode at any time by pressing the "OFF"
.key.
28
03.2_002a_en.fm
It is possible to exit the configuration mode at any time by pressing the "OFF"
.key.
03.2_002a_en.fm
29
If the "not memorised" mode is selected, the bolus default value must be defined.
30
03.2_002a_en.fm
It is possible to exit the configuration mode at any time by pressing the "OFF"
.key.
03.2_002a_en.fm
31
by a beep.
If noS;E, no detection.
32
Brand
Capacity (ml)
Details
BD Perfusion
50
BPf
BD Plastipak
50
BDk
BD Plastipak
20
BDk
Braun Omnifix
50
BrO
Braun Omnifix
20
BrO
Braun Perfusor
50
BrP
Braun Perfusor
20
BrP
50
DiL
Didactic Perfusion
50
DiP
Dispomed Spritze
50
Dis
Fresenius Injectomat
50
FrI
Fresenius P Spritze
50
FrP
Ivac
50
IVa
Map Gliss L L
50
MLL
50
MPL
Sherwood Monoject
50
SMJ
Sherwood Monoject
20
SMJ
Terumo
50
Trm
Terumo
20
Trm
Tutoject type T
50
TJT
03.2_002a_en.fm
03.3_002a_en.fm
35
36
03.3_002a_en.fm
The ASS tests may also be performed more easily and more quickly using a PC
with installed maintenance software (consult our After Sales Service).
The ASS test menu enables users to perform a series of 16 tests or checks:
n
n
n
n
n
n
n
n
n
n
n
n
n
n
n
03.4_002a_en.fm
37
38
03.4_002a_en.fm
4 Preventive maintenance
4.1 Recommendations
The Pilot syringe pump can only be inspected, serviced or repaired by Fresenius Vial or by
an authorised and appointed service. The qualified technicians in your establishment and our
After Sales Service should be notified of any abnormal operation of the device.
If a repair is necessary, send the instrument in its original packaging if possible with a precise
description of the observed fault, to the official dealer for Fresenius Vial.
For further information concerning troubleshooting or the usage procedure, please contact
our After Sales Service or our Sales Department.
Fresenius Vial is not liable for loss or damage to the equipment during transport to our After
Sales Service.
Name
12 months
3 years
1st inspection in
the 3rd year
Then every 2
years
3 years
Preventive maintenance
Frequency
04.1&2_001a_en.fm
41
42
04.1&2_001a_en.fm
Preventive maintenance
Preventive maintenance
4.3 Checks
In order to perform equipment follow-up by means of preventive maintenance, a regular
servicing inspection is recommended every 12 months (see "Regular servicing sheet").
To ensure the check procedure is carried out efficiently, recharge the battery
beforehand (16 hours).
Function
ON, to switch the machine ON.
OFF, to switch the machine off when pressed for over three seconds.
SILENCE ALARM, to access the test mode.
STOP, to cancel the test in progress.
CONFIRM, to validate a choice.
"PRESSURE LIMIT", is used for the backpressure test.
The selection keys allow to scroll the figures and letters on the tenths,
units, tens segments etc.
04.3_002a_en.fm
41
Preventive maintenance
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
Preventive maintenance
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
42
04.3_002a_en.fm
Selected key
SIA.L
StoP
VAL
boLu
S-..
S.-.
S..S_..
S._.
S_..
VPEr
SILENCE ALARM
STOP
CONFIRM
BOLUS
+ tens
+ units
+ tenths
- tens
- units
- tenths
VOLUME INFUSED
If two or more keys are pressed simultaneously, the display unit shows
an audio alarm sounds "beep!beep!beep!" indicating an error.
The "OFF" key is not included in the keyboard check.
"Err" and
The test may be stopped at any time by pressing the "CONFIRM" key for
over 3 seconds, and a different test may be selected.
Preventive maintenance
If the display is faulty, replace the display board (see "Display and central unit boards"
operation sheet).
04.3_002a_en.fm
43
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
Error
Description
A10
Battery alarm
E50
A11
Syringe clamp
alarm
E60
A12
End of infusion
alarm
E70
A13
E80
A14
Disengagement
alarm
E01
A15
E32
A16
Occlusion alarm
E52
A17
Fin alarm
E72
E03
Preventive maintenance
Alarm
44
04.3_002a_en.fm
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
To perform this test, use plug on which lines Rx and Tx are short-circuited (2 and 3).
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
Preventive maintenance
It is impossible to carry out this test when the PC is in communication wit the Pilot. In this
case, "OPEN" is displayed.
04.3_002a_en.fm
45
To perform this test, use plug on which lines Rx and Tx, RTS and CTS are short-circuited
(2 and 3, 7 and 8).
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
It is impossible to carry out this test when the PC is in communication wit the Pilot. In this
case, "OPEN" is displayed.
Preventive maintenance
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
46
04.3_002a_en.fm
Use the "- and +"select keys to move from one input to another.
If the value is out of limits, recalibrate the force sensor (see " EtA.9 Force sensor
calibration.").
Preventive maintenance
If the value is out of limits, recalibrate the position sensor (see "EtA.6
EtA.6 Position sensor
calibration.").
04.3_002a_en.fm
47
Display
Name
bat.1
bat.2
bat.3
HIG.H
Lou
0.G
St.G
Preventive maintenance
Capacities which are non-existent or non-configured in the EEPROM are displayed in the
form of ---.-.
o
48
04.3_002a_en.fm
n Press "ON".
o Lift the disengagement lever.
s Check for presence of the mechanical disengagement
alarm (red Led at the end of the syringe diagram).
o Fit the device with the 50 cc syringe, ensuring the fin and
plunger are in position.
o Release the disengagement lever.
s Ensure there is no mechanical lever release alarm.
s Check plunger locking.
04.3_002a_en.fm
Preventive maintenance
n Press "ON".
o Fit the device with the 50 cc syringe, ensuring the fin and
plunger are in position.
o Ensure the syringe is detected appropriately eg.
50c.c.
o Place the syringe in the device ensuring the fin is out of
the groove.
o Fit the syringe clamp and plunger correctly.
o Make sure the light alarm flashes, locating the problems
on the syringe diagram.
49
This check must be carried out for every syringe provided with the instrument.
Press "ON".
Fit the instrument with the spacer ref T301518-A and lock
it into position using the syringe clamp.
s Place the syringe plunger against the spacer with the
syringe head detection arm on the spacer.
s 20c.c is displayed without triggering the alarm.
Fit the instrument with the spacer ref T301519-A and lock
it into position using the syringe clamp.
s Place the syringe plunger against the spacer with the
syringe head detection arm on the spacer.
s 50c.c is displayed without triggering the alarm.
Press "ON".
Preventive maintenance
50
04.3_002a_en.fm
n Press "ON".
o Fit the instrument with the 50 cc syringe and lock it into
position using the syringe clamp, ensuring the fin and
plunger are correctly positioned.
o Place the manometer (or any other pressure
measurement device) at the syringe outlet.
o Select a 50 ml "B-D PLASTIPAK" syringe by pressing
"CONFIRM".
o Check the position of the backpressure adjustment
button.
o Select a maximum flow rate and initiate the infusion by
pressing the "CONFIRM" key.
s Ensure there is no acoustic and visual alarm
(backpressure Led off).
s Check that the infusion Leds are flashing.
s Ensure the alarm is triggered for a value of
0,7 bar + 0,1 bar.
o Stop the infusion cycle by pressing the "STOP
INFUSION" key.
Repeat the test setting the backpressure level to the
minimum and maximum levels.
s Min. backpressure = 0,4 bar + 0,1 bar.
s Max. backpressure = 1.2 bar + 0.2 bar.
Preventive maintenance
04.3_002a_en.fm
51
n Press "ON".
o Fit the device with the syringe, ensuring the fin and
plunger are in position.
o Ensure the syringe is detected appropriately eg.
50c.c.
o Select a "B-D PLASTIPAK" syringe filled to 20 ml.
o Select a flow rate of 120 ml/h.
s For normal flow rates, the pre-alarm is activated 5
minutes before end of infusion.
s Flow rate example: > 60 ml/hr, the pre-alarm is
activated when the remaining volume equals 10% of
the total syringe capacity.
s Ensure the end of infusion pre-alarm is present.
o Press the "SILENCE ALARM" and "TENS"
simultaneously.
s The acoustic alarm is silenced and the visual
signal is maintained.
o Measure the "hard height" at "end of infusion".
s 18,5 < x < 19,5.
Top view
For accurate checking of the "hard height", do not move the plunger when measuring.
Preventive maintenance
If the "hard height" reading is out of limits, recalibrate the position sensor (see "EtA.6
EtA.6
Position sensor calibration.").
52
04.3_002a_en.fm
n Press "ON".
o Fit the device with the syringe, ensuring the fin and
plunger are in position.
o Ensure the syringe is detected appropriately eg.
50c.c.
o Select a "B-D PLASTIPAK" syringe filled to 20 ml.
o Select a flow rate of 100 ml/hr on the display unit.
o Select a limit volume of 10 ml.
s Pre-alarm presence.
s For normal flow rates, the pre-alarm is activated 5
minutes before end of infusion.
s Flow rate example: > 60 ml/hr, the pre-alarm is
activated when the remaining volume equals 10% of
the total syringe capacity.
o Press the "SILENCE ALARM" key
s The acoustic alarm is silenced and the visual signal is
maintained.
The limit volume alarm is activated when the programmed limit volume is reached. The
device changes to KVO mode. flow rate of 1 ml/hr keeping the vein open.
o
o
o
o
o
o
Top view
Preventive maintenance
For accurate checking of the linearity do not move the plunger when measuring.
04.3_002a_en.fm
53
Press "CONFIRM"
Select a flow rate and validate.
Reduce the test power supply voltage until the battery
discharge is triggered.
s Ensure this is triggered between 5.8 V and 6 V.
If the results obtained during the tests differ from the values indicated, recalibrate the battery
voltage levels (see "EtA.4
EtA.4 Calibration of the 3 battery voltage levels.")
Preventive maintenance
54
04.3_002a_en.fm
Preventive maintenance
04.3_002a_en.fm
55
56
04.3_002a_en.fm
Preventive maintenance
Preventive maintenance
Code:
Equipment series N:
Procedure
10
11
12
fiche_002a_en.fm
Resulting value
Conformity
Yes
No
Preventive maintenance
Equipment type:
57
Procedure
Resulting value
13
14
15
16
Name:
Date:
Conformity
Yes
No
Signature:
Preventive maintenance
Comments:
58
fiche_002a_en.fm
Diagnosis
In order to purchase the ISDbit flow rate control software, contact the Fresenius Vial
customer service.
The test procedure defined below must be carried out using a 50 ml or 20 ml syringe.
Equipment required
Scales sensitivity
x < 5 ml/hr
1/10000th
1/1000th
x > 30 ml/hr
1/100th
n
n
n
n
n
n
Needle type
x < 30ml/hr
G26
x > 30 ml/hr
G18 or G21
Installation
1 cm
Cord
RS232
04.4_001a_en.fm
Diagnosis
59
The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.
Operating mode
The software works following the operating mode described in the PrEN-60-601-2-24
standard project for infusion pumps.
Diagnosis
60
04.4_001a_en.fm
n Stop clock
n Scales
Flow rate value
Scales sensitivity
x < 5 ml/hr
1/10000th
1/1000th
x > 30 ml/hr
1/100th
n
n
n
n
n
Needle type
x < 30ml/hr
G26
x > 30 ml/hr
G18 or G21
Installation
1 cm
Diagnosis
04.4_001a_en.fm
61
The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.
Operating mode
For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour. For
higher flow rates, wait for 10 to 30 minutes after infusion.
1 gram = 1 ml.
Diagnosis
62
04.4_001a_en.fm
n
n
n
n
n
n
Stop clock.
Test tube or beaker with 1 ml graduating
Liquid: distilled water and oil
Luer lock type plastic syringe (50 or 20 ml)
Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 cm).
Needle:
Needle type
x < 30ml/hr
G26
x > 30 ml/hr
G18 or G21
Installation
1 cm
Diagnosis
04.4_001a_en.fm
63
Operating mode
For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour. For
higher flow rates, wait for 10 to 30 minutes after infusion.
Diagnosis
64
04.4_001a_en.fm
Diagnosis
n Disconnect the power cable from the wall socket before cleaning.
n Do not place in an AUTOCLAVE or IMMERSE the device, and do not allow liquid to
penetrate inside the equipment case or power supply cover.
o Use a cloth soaked in DETERGENT-DISINFECTANT, diluted in water if necessary, to
eliminate micro organisms.
o Avoid excessively abrasive brushing that could scratch the case.
o Do not rinse or wipe the surfaces.
n If the equipment is used in a department with severe contamination risks, after disinfecting
by wiping with a damp cloth, equipment should be left in the room during aerial disinfection.
n TRICHLOROETHYLENE-DICHLOROETHYLENE:
o AMMONIA.
o AMMONIUM CHLORIDE
o CHLORINE AND AROMATIC HYDROCARBON.
o ETHYLENE DICHLORIDE-METHYLENE CHLORIDE
o CETONE based cleaning products.
These aggressive agents could damage the plastic parts and lead to apparatus malfunctions.
Also beware of ALCOHOL SPRAYS (20% -40% alcohol) that tarnish and crack the
plastic and fail to provide the cleaning action required prior to disinfection.
Diagnosis
For further information, please contact the competent department in your hospital for supply
of the appropriate cleaning and disinfecting products.
04.5&6_001a_en.fm
65
4.6 Storage
In case of prolonged storage, it is advisable to disconnect the battery using the battery access
door on the bottom of the device. This operation should be done by an experienced
technician.
The equipment must be stored in a dry and cool place.
n The recommended environmental temperature conditions for storage of the equipment are
between 0C and 40C.
n Relative humidity tolerated: max. 85%, no condensation.
Diagnosis
Fully recharge the battery before using the equipment to avoid any risks caused by micro
power cuts in the mains supply and to ensure maximum autonomy.
66
04.5&6_001a_en.fm
5 Diagnosis
5.1 Troubleshooting
Problem
Cause
Solution
After a fall.
Diagnosis
05.1_002a_en.fm
67
Diagnosis
68
05.1_002a_en.fm
Diagnosis
Description
Recommended action
Er10
Er20
Er30
Er40
Er50
n ADC anomaly.
Er60
n Syringe parameter
anomaly.
Er70
n Motor frequency
anomaly.
Er80
n Faulty keyboard.
n Short circuit in the
keyboard.
n Electromagnetic
disturbance exceeding
standard limits.
n EEPROM anomaly.
n EEPROM anomaly.
*: When rewriting the EEPROM, when the device is switched off, the Check Sum is
rewritten in the memory to save the parameters.
If the Hard cut-off circuit is shorter in time than the Soft circuit, the device is switched off
before the EEPROM is fully rewritten: Check Sum not compliant.
Err(-)0 or CFPc: When the device is in CFPc, reconfiguration is compulsory: Faulty WATCH
DOG.
Motor anomalies
Er01
Er52
n Anomaly during
compensation for play.
Er72
Er82
Diagnosis
Er32
05.2_001a_en.fm
69
Error code
Description
Recommended action
Er14
Er24
Er34
n Flow rate/period
calculation anomaly.
Er44
Configuration anomalies
Diagnosis
CFPc
70
05.2_001a_en.fm
6 Operation sheets
Operation sheets
06.0_001a_en.fm
71
Operation sheets
72
06.0_001a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_0010a_en.fm
73
Dismounting
When electronic components are handled, it is recommended to wear an antistatic
wriststrap linked to ground and to work on an antistatic foam mat.
n Unscrew the 5 Phillips head screws (ref. 2) located at the display board, which link this to
the upper case.
n Remove the board insulator (ref. 3) located on the left.
n Lift the display board slightly and remove the display unit flat jumper (ref. 4).
n Disconnect the black connector (ref. 5).
2
2
2
2
4
5
Remounting
A specific type of board corresponds to each Pilot "CPU and display board" ; It is
important to avoid reversing the references between each Pilot and order the part
number corresponding to your device.
Carry out the same procedures in reverse to reassemble the parts.
Operation sheets
When mounting the display board, it is important to reduce the torque in the plastic inserts
so as to avoid causing damage to these.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the Pilot (see "Calibrations") then carry out the regular servicing tests (see
"Regular servicing sheet").
74
06.0_0010a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 Posidriv Z1 flat screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_0020a_en.fm
75
Dismounting
n Remove the spring retaining ring (ref. 2) located on the syringe clamp shaft, which holds
this to the CPU support.
n Remove the obturator and the spring (ref. 3).
n Remove the syringe clamp.
2
3
Reassembly
Carry out the same procedures in reverse to remount the unit.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
76
06.0_0020a_en.fm
Operation sheets
Equipment required:
n
n
n
n
n
1 Posidriv Z1 screwdriver.
1 antistatic wriststrap.
1 soldering iron
"RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.
Silicone ref. 161249.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_0030a_en.fm
77
Dismounting
n
n
n
n
n
4
4
3
2
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and clean kept. The temperature of the iron should be
between 315C and 340C.
Operation sheets
78
06.0_0030a_en.fm
Reassembly
n Place the new switch on the support.
n Bend the mounting lugs towards the inside in order to lock them into position.
n Weld the wires of the new switch:
o Taper the 2 welding wires and the 2 lugs of the switch to be welded.
o Place the wires under the bent lugs.
o Place the welding iron tip on the welding surface.
o Place the welding wire on the pin surface of the switch to be welded.
o Remove the welding wire and then the welding iron tip.
o Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the wire.
n Screw the 2 Phillips head screws (ref. 3) which hold the "switch support" assembly in
position.
When mounting the "switch support", it is important to reduce the torque in the plastic
inserts so as to avoid causing damage to these.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
06.0_0030a_en.fm
79
Hands must not come into contact with the CPU boards.
Dismounting
n
n
n
n
5
Green
Black
White
Red
Operation sheets
80
06.0_0030a_en.fm
Reassembly
n Place the new opto assembly on the PCB support.
n Screw the 2 Phillips head screws (ref. 5) which hold the "PCB and opto" assembly in
position.
When mounting the "PCB and opto", it is important to reduce the torque in the plastic
inserts so as to avoid causing damage to these.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
06.0_0030a_en.fm
81
Operation sheets
82
06.0_0030a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_0040a_en.fm
83
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
3
n Unscrew the 2 Phillips head screws (ref. 5).
n Remove the motor, opto and disk assembly.
5
Reassembly
Operation sheets
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Perform the regular servicing tests (see "Regular servicing sheet").
84
06.0_0040a_en.fm
Operation sheets
Equipment required:
n
n
n
n
1 Posidriv Z1 screwdriver.
1 hexagon socket key (2.5).
1 soldering iron.
"RADIEL Sn60Pb RI 1" welding wire (cleaning not required for re-welding) or equivalent.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_0051a_en.fm
85
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
n Remove the plunger guide.
n Unscrew the 3 Phillips head screws (ref. 2) located at the bottom of the plunger end shield
which links this to the plunger cover.
n Remove the plunger end shield.
2
2
2
5
4
3
Operation sheets
86
06.0_0051a_en.fm
n Unscrew the 2 Phillips head screws (ref. 7 and 8) which link the plunger support to the
plunger cover.
n Unscrew and remove the Phillips head screw and the washer (ref. 9) which attach the antisiphon cam to the anti-siphon arm.
n Remove the cam/arm assembly.
n Remove the plunger cover fitted with the bonding pad.
8
n Remove the retaining ring that holds the anti-siphon opto in place (ref. 10).
n Unscrew the 6 hexagon socket screws (ref. 11).
10
Operation sheets
11
06.0_0051a_en.fm
87
12
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315C and 340C.
When dismounting the force sensor, ensure the welding pellets are not damaged.
Operation sheets
88
06.0_0051a_en.fm
Reassembly
n Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
o Taper the 3 welding points.
o Weld the 3 points.
o Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.
n Repeat the operation for the anti-siphon opto (4 welding points).
n Stick the bottom side of the ribbon cable to the double-sided board.
Anti-siphon opto
w
Backpressure switch
Lower side
(connection specific to Pilot C)
stuck to the double-sided board on the plunger support
n Impregnate with weak loctite and screw the 6 hexagon socket screws (ref. 11), which hold
the spring leaf/backpressure micro support in position.
n Position the opto and its retaining ring (ref. 10).
n Mount the plunger cover onto the plunger support. Impregnate the screw with weak loctite
(rep. 9).
n Position and screw the cam/anti-siphon arm unit using the Phillips head screw and the
washer (ref. 9).
n Screw the 2 Phillips head screws (ref. 8 and 7) which link the plunger support to the
plunger cover.
n Position the plunger cover clip (ref. 6).
Operation sheets
06.0_0051a_en.fm
89
n Cover the plunger with the protective film, mount the disengagement lever, its spring and
the adjustment button. Screw the plunger end shield back on (ref. 5, 4 and 3).
n Position the plunger cover and screw it back on (ref. 2).
Carry out the same procedures in reverse to remount the unit.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
Recalibrate the pressure sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
90
06.0_0051a_en.fm
Operation sheets
Equipment required:
n
n
n
n
1 Posidriv Z1 screwdriver.
1 flat screwdriver.
1 flat key (12).
1 potentiometer dismounting tool ref. T300869.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower housing
case, which link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_0060a_en.fm
89
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
n Unscrew the potentiometer lock nut (ref. 5).
n Remove the tooth lockwasher and the potentiometer.
5
4
Reassembly
Operation sheets
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the position sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
90
06.0_0060a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
Dismounting
n Unscrew the 3 Phillips head screws (ref. 1) located at the bottom of the plunger end shield
which links this to the plunger cover. Remove the plunger end shield (ref. 2).
n Remove the backpressure adjustment button (ref. 3), then the disengagement lever and its
spring (ref. 4) as well as the protective plunger film (ref. 5).
n Unscrew and remove the Phillips head screw and the washer (ref. 6) which attach the antisiphon cam to the anti-siphon arm.
n Remove the cam/anti-siphon arm assembly (ref. 7).
4
6-7
3
2
06.0_0071a_en.fm
Operation sheets
93
Reassembly
Carry out the same procedures in reverse to remount the unit.
n Impregnate the screw with weak loctite (ref. 6) before mounting the cam and arm
assembly.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
94
06.0_0071a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 small flat screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_0081a_en.fm
95
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the syringe pump guide.
n Disconnect the black connectors.
n Unscrew the flat jumper mounting lug (ref. 2) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 3).
n Remove the mechanical plunger unit.
n Disconnect the connectors (ref. 4, ref 5, ref.6, ref.7)
n Unscrew the 3 Phillips head screws (ref. 8) located on the power supply board, which link
this to the lower case.
n Unscrew the 2 Phillips head screws (ref. 9) located on lower case, which link this to the
power supply board.
n Remove the power supply board.
8
2
3
8
6
4
5
7
Reassembly
Carry out the same procedures in reverse to remount the unit.
Operation sheets
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the force sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
96
06.0_0081a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
Dismounting
n Unscrew the 2 Phillips head screws (ref. 1) to be found in the lower case battery holder kit,
which link this to the lower case.
n Swivel the battery holder and remove the battery.
n Disconnect this.
Reassembly
It is recommended to perform a full loading and unloading cycle to ensure the battery is in
working condition.
Perform the regular servicing tests (see "Regular servicing sheet").
06.0_0090a_en.fm
95
Operation sheets
Operation sheets
96
06.0_0090a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
06.0_00101a_en.fm
Operation sheets
99
Dismounting
n Disconnect the two black connectors (ref. 2).
n Unscrew the 2 Phillips head screws (ref. 3) located on the cabled plug support, which link
this to the lower case.
n Remove the cabled plug support.
3
2
2
Reassembly
Carry out the same procedures in reverse to remount the unit.
If the number of rear door connector points is different to the number of power
supply board points, make the connections as described below.
Operation sheets
100
06.0_00101a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_00110a_en.fm
99
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2 and ref. 3).
n
n
n
n
n
4
5
2&3
Remounting
Fitting the flexible tongue:
Operation sheets
100
06.0_00110a_en.fm
There are several types of flange tongues. If you have an older version, replace it with that
provided with the kit.
n Position the new guiding flange (ref. 4) using the two screws.
additional screw
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
06.0_00110a_en.fm
101
Operation sheets
102
06.0_00110a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 hexagon socket key (2.5).
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
06.0_00121a_en.fm
Operation sheets
105
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
n Remove the syringe pump guide.
n Unscrew the 3 Phillips head screws (ref. 2) located at the bottom of the plunger end shield
slots which link this to the plunger cover
n Remove the plunger end shield.
n Remove the backpressure adjustment button (ref. 3)
n Remove the disengagement lever and its spring as well as the protective plunger film.
n Remove the plunger cover clip.
2
2
2
n Unscrew the 2 Phillips head screws (ref. 4 and 5) which link the plunger support to the
plunger cover.
n Unscrew and remove the Phillips head screw and the washer (ref. 6) which attach the antisiphon cam to the anti-siphon arm.
n Remove the cam/arm unit as well as the plunger cover.
n Remove the plunger cover fitted with the bonding pad (ref. 7).
4
Operation sheets
106
06.0_00121a_en.fm
10
11
8
Reassembly
n Tighten the Phillips head screw which holds the stop plate (ref. 11), the spring leaf
(ref. 10), the removable stop plate (ref. 9) to the bonding pad (ref. 8).
n Clip the bonding pad assembly to the plunger cover (ref. 7)
n Position the plunger cover.
1) Do not forget to replace the anti-siphon cam with the new one supplied with the kit.
2) Weak loctite must be used to mount the new arm and the new anti-siphon arm supplied
with the kit.
n Position and screw the cam/anti-siphon arm unit using the Phillips head screw and the
washer (ref 6).
n Screw the 2 Phillips head screws (ref. 5 and 4) which link the plunger support to the
plunger cover.
n Position the plunger cover clip.
n Cover the plunger with the protective film, and mount the disengagement lever and its
spring, and screw the plunger end shield back on.
Carry out the same procedures in reverse to remount the unit.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
06.0_00121a_en.fm
107
Operation sheets
108
06.0_00121a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 cutter.
n GEB type 100% silicone grease.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_00130a_en.fm
107
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
3
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
Dismounting
n Unscrew the 2 slotted head screws (ref. 5) which link the centering ring to the mechanical
end shield.
n Using the cutter, cut the input bearing and remove it from the plunger tube.
Operation sheets
108
06.0_00130a_en.fm
Remounting
n Mount the new slotted input bearing onto the plunger tube by twisting it.
n Laying it flat on a table, cut the O-ring using a cutter.
n Place the O-ring around the tube.
n Fit the O-ring into the centering ring slot (horizontal cut, so as to avoid lining it up with the
centering ring cut).
n Mount the stainless steel slotted plate onto the pin, by twisting it.
n Place the silicone grease cord onto the plate around the passage of the pin.
n Place the plate on the input bearing.
n Position the whole assembly against the mechanical end shield and screw it on using the
two slotted screws (ref. 5).
n Check the plunger slides correctly.
It should slide uniformly across the entire centering ring and slight friction is due to the Oring scraping against the tube.
Operation sheets
06.0_00130a_en.fm
109
Operation sheets
110
06.0_00130a_en.fm
Operation sheets
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_00141a_en.fm
113
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
4
3
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315C and 340C.
n Unweld the flex circuit (ref. 5) ensuring the disengagement switch is not damaged in the
process.
n Unscrew the 2 slotted head screws (ref. 6) which link the centering ring to the mechanical
end shield.
n Unscrew the 2 Phillips head screws (ref. 7) which link the centering ring to the mechanical
end shield.
n Remove the flex circuit clip (ref. 8).
Operation sheets
6
8
114
06.0_00141a_en.fm
9
9
9
12
11
10
Operation sheets
13
06.0_00141a_en.fm
115
n Unscrew the 2 Phillips head screws (ref. 14 and 15) which link the plunger support to the
plunger cover.
n Unscrew and remove the Phillips head screw and the washer (ref. 16) which attach the
anti-siphon cam to the anti-siphon arm.
n Remove the cam/arm unit as well as the plunger cover.
14
15
16
n Remove the retaining ring that holds the anti-siphon opto in place (ref. 17).
n Unscrew the 6 hexagon socket screws (ref. 18).
18
17
Operation sheets
17
116
06.0_00141a_en.fm
19
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315C and 340C.
When dismounting the force sensor, ensure the welding blocks are not damaged and the
square is not removed in the process.
o
o
Operation sheets
n Remove the flex circuit and tube kit from the plunger support.
06.0_00141a_en.fm
117
Reassembly
n Remove the sticky tape holding the flex circuit to the tube.
n Mount the new flex circuit onto the plunger support.
n Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
o Taper the 3 welding points.
o Weld the 3 points.
o Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.
n Repeat the operation for the anti-siphon opto (4 welding points).
n Stick the bottom side of the ribbon cable to the double-sided board.
n Repeat the operation for the disengagement switch. (3 welding points)
Tube D. 12
ube
Switch
disengagement
Switch
backpressure
Anti-siphon opto
Lower side
(connection specific to Pilot C)
stuck to the double-sided board on the plunger support
Do not forget to impregnate the cam/arm clamping screw (rep. 16) with weak loctite.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
Recalibrate the force sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
118
06.0_00141a_en.fm
Operation sheets
Equipment required:
n
n
n
n
1 Posidriv Z1 screwdriver.
1 antistatic wriststrap.
1 soldering iron.
"RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
Operation sheets
06.0_00150a_en.fm
119
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315C and 340C.
o
Clean the buzzer slot surfaces using a cloth soaked in rubbing alcohol.
Display board
Operation sheets
120
06.0_00150a_en.fm
Operation sheets
06.0_00150a_en.fm
121
Operation sheets
122
06.0_00150a_en.fm
7 Calibration
7.1 Calibration procedure
The calibration menu is reserved for authorised personnel only, its access is
protected by a secret code.
Function
ON, is used to switch the machine ON.
OFF, is used to switch the machine off when pressed for over three
seconds.
SILENCE ALARM, is used to access the calibration mode.
CONFIRM, is used to validate a choice.
BOLUS, is used to access the calibration mode.
The select keys allow to scroll the figures and letters on the LCD
screen, on the units, tens, tenths segments etc.
Activate calibration
Calibration
07.1_002a_en.fm
123
o
o
o
o
o
o
Once both high and low values have been stored in the EEPROM, the device indicates
the number of LSB in decimals between the two measurement points.
This value should be 776 + 10 LSB.
o
Calibration
124
07.1_002a_en.fm
.....
.....
.....
.....
.....
.....
.....
.....
.....
.....
.....
.....
.....
Qty
Reference
Name
167246
167627
199560
167067
167296
167632
Buzzer foam
167636
99
167744
100
167476
101
1
1
1
199230
199231
199232
08.1_002a_en.fm
Mark
123
.....
.....
.....
122
111
112
104
125
123
126
127
114
128
103
178
124
102
Mark
Qty
Reference
Name
102
167361
103
167310
104
167452
111
167670
112
1
1
1
1
168101
168100
168103
168104
114
167551
Display board
167704
122
167458
123
167944
124
167462
125
126
127
124
Qty
Reference
Name
128
199618
178
167568
CPU board
Mark
08.1_002a_en.fm
125
126
08.1_002a_en.fm
117
118
107
119
Mark
Qty
Reference
Name
167059
167299
167297
Spring washer
167961
167122
167355
170416
Mains cord
107
167093
Buzzer bell
108
1
1
167967
167968
109
177201
Transformer
167432
115
170228
Fuse F2
116
1
1
167563
167565
117
167942
118
1
1
1
1
199200
199201
199202
199203
119
167249
08.2_002a_en.fm
120
121
127
109
115
116
108
117
118
107
119
120
121
Qty
Reference
Name
120
174019
121
199169
Mark
128
08.2_002a_en.fm
.....
200
201
202
.....
.....
224
225
.....
.....
.....
.....
.....
.....
.....
.....
203
204
205
206
207
208
209
.....
.....
.....
220
221
.....
.....
.....
.....
.....
.....
223
.....
222
.....
.....
.....
210
Mark
Qty
Reference
Name
167264
Anti-siphon spring
200/207
167360
201
167465
202
167469
Disengagement follower
203
167460
204
167471
205
167464
206
167472
208
167475
Mechanical flange
209
167281
210
167275
220
167271
Flex circuit
221
167317
222
167292
223
167403
177203
177204
Stainless steel plate (to be mounted between the bearing and the end
shield)
224
167291
225
199252
08.3_002a_en.fm
127
229
.....
.....
.....
230
231
.....
.....
249
260
261
.....
254
250
251
252
.....
.....
.....
.....
.....
.....
.....
270
271
232
.....
.....
.....
.....
272
255
256
257
258
259
.....
226
227
.....
.....
273
Qty
Reference
Name
226
167259
227
167442
229
167383
230
167411
231
167273
232
173408
OMRON micro-switch
249
162311
250
167492
Backpressure spring
251
167270
252
167269
254
167497
255
167460
Disengagement cam
256
168231
Disengagement finger
257
167487
258
167298
259
167272
260
190714
Retaining ring
261
167385
Anti-siphon cam
270
167260
Disengagement lever
271
167245
272
167055
273
199103
128
08.3_002a_en.fm
225
229 261
274
276
275
298
Plunger kit
Mark
Qty
225
Reference
Name
Pilot plunger cover
229
Bonding pad
261
Anti-siphon cam
274
275
276
Plate
298
Mechanical kit
199135
08.3_002a_en.fm
129
130
08.3_002a_en.fm
313
308
309
310
296
300
301
302
303
Mark
Qty
Reference
Name
190714
Retaining ring
296
167963
300
167128
301
167145
302
167111
303
167964
304
167156
Reducer Pilot A2
308
167117
309
167140
310
167158
311
167142
312/313
167143
314
167144
315
167157
Spacer
08.4_002a_en.fm
304
133
134
08.4_002a_en.fm
8.5 Labels
Qty
Reference
Name
1
1
1
1
1
1
1
1
168110
168111
168112
168113
168114
168115
168116
168117
167890
167742
1
1
1
1
1
1
1
1
167741
167979
167776
167811
167778
167858
167886
167692
167734
167756
167691
171091
Main DANGER label (inside) Fr/C, DIN, IS, NL, IT, SP, S, B, P
1
1
1
1
1
1
1
167725
167978
167814
167800
167758
167897
167690
NO
Mark
08.5_002a_en.fm
135
136
08.5_002a_en.fm
Useful addresses
SALES DEPARTMENT
Fresenius Vial
Le Grand Chemin, 38590 Brzins
CUSTOMER SERVICE
Tel.: 00 33 (0) 4 76 67 10 81
or 00 33 (0) 4 76 67 10 54
or 00 33 (0) 4 76 67 11 08
Fax: 00 33 (0) 4 76 67 11 34
SALES MANAGEMENT
Tel.: 00 33 (0) 4 76 67 10 81
Fax: 00 33 (0) 4 76 67 11 34
Fresenius Vial
Le Grand Chemin, 38590 Brzins
Tel.: 04 76 67 10 76
Fax: 04 76 67 11 22
FRANCE
Tel.: 04 76 67 10 34
Fax: 04 76 67 11 22
BELGIUM
Tel.: 32/388.07307
Fax: 32/388.05007
GERMANY
Fresenius MCM
AM-Neuen Berg
63749 ALZENAU
GERMANY
TRAINING DEPARTMENT