Академический Документы
Профессиональный Документы
Культура Документы
2. Nebulization
4. Detection
F l ow o f a n a l y s i s
Fig. 1 Flow of ELSD Analysis
3 . E L SD s a n d R I D etec to rs
Like RI detectors, ELSDs are classified as general-purpose
detectors but they differ from RI detectors in the following
ways:
1. They are 5 to 10 times more sensitive than RI
detectors.
2. They support the use of the gradient elution
method.
3. They are not easily affected by changes in the
ambient temperature.
4. They are not affected by interference due to
solvent peaks.
5. Time is not required to allow for the instrument and
the baseline to stabilize.
Although the gradient elution method is effective for the batch
analysis of, for example, multiple components in natural
products, RI detectors cannot be used for this because of
fluctuations in the baseline caused by changes in the
refractive index of the mobile phase.
Previous ELSD
RI detector
Sample
Fructose, Glucose
Sucrose, Maltose
Column
Mobile Phase
Flow Rate
1mL/min
Detection(ELSD)
Temperature: 35C
Gas Pressure: 350kPa
4 . Fe a t u re s o f E L S D -L T 2
4 -1 . H i g h - Se ns itivity D etec tio n of Semi- Volat ile Subst ances
A c h i e v e d with L o w-T e mp erat ure Evaporat ion of M obile Phase
Nebulizer
Glass cell
Mobile
phase
Nebulizer gas
To the
drift tube
Quantity of droplets
Small
Size of droplets
Large
Photomultiplier
Light source
Assisting gas
Assisting gas
4 -2 . M o re U s er F rien d ly
Setting the drift tube temperature and gain is the
only preparation required to perform analysis with
the ELDS-LT2. It is also possible to turn off the
lamp and stop the nebulizer after the completion
of analysis using the auto power-down function.
The use of a long-life LED lamp
Mobile Phase
Flow Rate
1mL/min
Column Temp.
30C
Column
Mobile Phase
A) Water B) Methanol
Time Program
B Conc.40% (0min)
Flow Rate
1.0mL/min
Column Temp.
Detection (ELSD)
60% (15min)
40C
Temperature : 40C
Gas Pressure : 350kPa
Analysis of Triglycerides
Triglycerides, which are the main components of cooking oil,
exist in many different forms that vary according to the acyl
group, and the separation of these components is required
for purposes such as quality control. Although detection is
possible for triglycerides based on the absorbance of the
acyl group in the range of 200 to 210 nm (UV), with systems
that use acetone as the mobile phase, there is a high level of
background absorbance, and UV detectors cannot be used.
Fig. 8 shows an example of the results obtained by
subjecting sesame oil to gradient elution using an acetonebased mobile phase in the reverse-phase mode, and
performing detection using an ELSD. This example shows
how an ELSD can be an effective detection tool even in
cases where the mobile phase has UV absorbance.
Analysis of Oligosaccharides
In the separation of oligosaccharides according to the
glucose polymerization degree, although the normal-phase
mode (HILIC mode) demonstrates superior selectivity, the
time required for elution increases greatly as the number of
glucose units increases. With an RI detector, gradient elution
cannot be used, so analysis must be performed repeatedly
using mobile phase compositions that are suitable for the
separation of each polymerization degree. Fig. 9 shows an
example of the results obtained by subjecting
maltooligosaccharides to gradient elution using the normalphase mode, and performing detection using an ELSD. Using
an ELSD makes it possible to identify maltooligosaccharides
with glucose polymerization degrees of up to around 20.
Column
Sample
Maltooligosaccharides
Mobile Phase
A ) Acetonitrile
B ) Acetone
Column
Time Program(40min)
Mobile Phase
A ) Acetonitrile/Water=7/3 (v/v)
B ) Acetonitrile/Water=4/6 (v/v)
Flow Rate
1.0mL/min
Time Program
B Conc.0% (0min)
Column Temp.
30C
Flow Rate
1.0mL/min
Temperature : 35C
GAIN : 5
Nebulizer Gas : Air
Gas Pressure : 350KPa
Column Temp.
40C
Detection (ELSD)
Detection (ELSD)
Temperature : 40C
Gas Pressure : 350kPa
70% (40min)
100% (25min)
JQA-0376
www.shimadzu.com
URL http://www.shimadzu.com