Академический Документы
Профессиональный Документы
Культура Документы
(LPI) offers over an unaided visual inspection is that it makes defects easier to see for the
inspector. There are basically two ways that a penetrant inspection process makes flaws
more easily seen. First, LPI produces a flaw indication that is much larger and easier for
the eye to detect than the flaw itself. Many flaws are so small or narrow that they are
undetectable by the unaided eye. Due to the physical features of the eye, there is a
threshold below which objects cannot be resolved. This threshold of visual acuity is
around 0.003 inch for a person with 20/20 vision.
The second way that LPI improves the detectability of a flaw is that it produces a flaw
indication with a high level of contrast between the indication and the background which
also helps to make the indication more easily seen. When a visible dye penetrant
inspection is performed, the penetrant materials are formulated using a bright red dye that
provides for a high level of contrast between the white developer
that serves as a background as well as to pull the trapped penetrant
from the flaw. When a fluorescent penetrant inspection is
performed, the penetrant materials are formulated to glow brightly
and to give off light at a wavelength that the eye is most sensitive
to under dim lighting conditions.
Additional information on the human eye can be found by
following the links below.
The retina is a mosaic of two basic types of photoreceptors, rods, and cones. Rods are
sensitive to blue-green light with peak sensitivity at a wavelength of 498 nm, and are
used for vision under dark or dim conditions. There are three types of cones that give us
our basic color vision and they are L-cones (red) with a peak sensitivity of 564 nm, Mcones (green) with a peak sensitivity of 533 nm, and S-cones (blue) with a peak
sensitivity of 437 nm.
Cones are highly concentrated in a region near the center of the retina called the fovea
region. The maximum concentration of cones is roughly 180,000 per square mm in the
fovea region and this density decreases rapidly outside of the fovea to a value of less than
5,000 per square mm. Note the blind spot caused by the optic nerve which is void
photoreceptors.
The standard definition of normal visual acuity (20/20 vision) is the ability to resolve a
spatial pattern separated by a visual angle of one minute of arc. Since one degree contains
sixty minutes, a visual angle of one minute of arc is 1/60 of a degree. The spatial
resolution limit is derived from the fact that one degree of a scene is projected across 288
micrometers of the retina by the eye's lens.
In this 288 micrometers dimension, there are 120 color sensing cone cells packed. Thus,
if more than 120 alternating white and black lines are crowded side-by-side in a single
degree of viewing space, they will appear as a single gray mass to the human eye. With a
little trigonometry it is possible to calculate the resolution of the eye at a specific distance
away from the lens of the eye.
For the case of normal visual acuity the angle Theta is 1/60 of a degree. By bisecting this
angle we have a right triangle with angle Theta/2 that is 1/120 of a degree. Using this
right triangle it is easy to calculate the distance X/2 for a given distance d.
X/2 = d (tan Theta/2)
When visually inspecting an object for a defect such as a crack the distance (d) might be
around 12 inches. This would be a comfortable viewing distance. At 12 inches, the
normal visual acuity of the human eye is 0.00349 inch. What this means is that if you had
alternating black and white lines that were all 0.00349 inch wide, it would appear to most
people as a mass of solid gray.
Contrast Sensitivity
When conducting a visible dye penetrant inspection, the contrast sensitivity of the eye is
important. Contrast sensitivity is a measure of how faded or washed out an image can be
before it becomes indistinguishable from a uniform field. It has been experimentally
determined that the minimum discernible difference in gray scale level that the eye can
detect is about 2% of full brightness. Contrast sensitivity is a function of the size or
spatial frequency of the features in the image. However, this is not a direct relationship as
larger objects are not always easier to see than smaller objects as contrast is reduced as
demonstrated by the image below.
In the image below, the luminance of pixels is varied sinusoidally in the horizontal
direction. The spatial frequency increases exponentially from left to right. The contrast
also varies logarithmically from 100% at the bottom to about 0.5% at the top. The
luminance of peaks and troughs remains constant along a given horizontal path through
the image. If the detection of contrast was dictated solely by image contrast, the
alternating bright and dark bars should appear to have equal height everywhere in the
image. However, the bars seem to be taller in the middle of the image.
The three curves in the figure above shows the normalized response of an average human
eye to various amounts of ambient light. The shift in sensitivity occurs because two types
of photoreceptors, cones and rods, are responsible for the eye's response to light. The
curve on the right shows the eye's response under normal lighting conditions and this is
called the photopic response. The cones respond to light under these conditions. As
mentioned previously, cones are composed of three different photo pigments that enable
color perception. This curve peaks at 555 nanometers, which means that under normal
lighting conditions, the eye is most sensitive to a greenish yellow color. When the light
levels drop to near total darkness, the response of the eye changes significantly as shown
by the scotopic response curve on the left. At this level of light, the rods are most active
and the human eye is more sensitive to any amount of light that is present, but is less
sensitive to the range of color. Rods are highly sensitive to light but are comprised of a
single photo pigment, which accounts for the loss in ability to discriminate color. At this
very low light level, sensitivity to blue, violet, and ultraviolet is increased, but sensitivity
to yellow and red is reduced. The heavier curve in the middle represents the eye's
response at the ambient light level found in a typical inspection booth. This curve peaks
at 550 nanometers, which means the eye is most sensitive to yellowish green color at this
light level. Fluorescent penetrant inspection materials are designed to fluoresce at around
550 nanometers to produce optimal sensitivity under dim lighting conditions
3 Penetrant Dwell: The penetrant is left on the surface for a sufficient time to allow as
much penetrant as possible to be drawn from or to seep into a defect. Penetrant dwell
time is the total time that the penetrant is in contact with the part surface. Dwell times are
usually recommended by the penetrant producers or required by the specification being
followed. The times vary depending on the application, penetrant materials used, the
material, the form of the material being inspected, and the type of defect being
inspected. Minimum dwell times typically range from 5 to 60 minutes. Generally, there is
no harm in using a longer penetrant dwell time as long as the penetrant is not allowed to
dry.
4. Excess Penetrant Removal: This is a most delicate part of the inspection procedure
because the excess penetrant must be
removed from the surface of the sample
while removing as little penetrant as possible
from defects. Depending on the penetrant
system used, this step may involve cleaning
with a solvent, direct rinsing with water, or
first treated with an emulsifier and then
rinsing with water
8.
Clean Surface: The final step in the process is to thoroughly clean the part
surface to remove the developer from the parts that were found to be acceptable.
Rubber
Plastics
Fatigue cracks
Quench cracks
Grinding cracks
Porosity
Laps
Seams
As mentioned above, one of the major limitations of a penetrant inspection is that flaws
must be open to the surface. To learn more about the advantages and disadvantages of
LPI proceed to the next page.
Large areas and large volumes of parts/materials can be inspected rapidly and at
low cost.
Indications are produced directly on the surface of the part and constitute a visual
representation of the flaw.
Primary Disadvantages
Metal smearing from machining, grinding, and grit or vapor blasting must be
removed prior to LPI.
The inspector must have direct access to the surface being inspected.
Large areas and large volumes of parts/materials can be inspected rapidly and at
low cost.
Indications are produced directly on the surface of the part and constitute a visual
representation of the flaw.
Primary Disadvantages
Metal smearing from machining, grinding, and grit or vapor blasting must be
removed prior to LPI.
The inspector must have direct access to the surface being inspected.
remain in the defect but remove easily from the surface of the part.
remain fluid so it can be drawn back to the surface of the part through the drying
and developing steps.
All penetrant materials do not perform the same and are not designed to perform the
same. Penetrant manufactures have developed different formulations to address a variety
of inspection applications. Some applications call for the detection of the smallest defects
possible and have smooth surface where the penetrant is easy to remove. In other
applications the rejectable defect size may be larger and a penetrant formulated to find
larger flaws can be used. The penetrants that are used to detect the smallest defect will
also produce the largest amount of irrelevant indications.
Penetrant materials are classified in the various
industry and government specifications by their
physical characteristics and their performance.
Aerospace Material Specification (AMS) 2644,
Inspection Material, Penetrant, is now the
primary specification used in the USA to control
penetrant materials. Historically, Military
Standard 25135, Inspection Materials,
Penetrants, has been the primary document for
specifying penetrants but this document is
slowly being phased out and replaced by AMS
2644. Other specifications such as ASTM 1417,
Standard Practice for Liquid Penetrant Examinations, may also contain information on
the classification of penetrant materials but they are generally referred back to MIL-I25135 or AMS 2644.
Penetrant materials come in two basic types. These types are listed below:
Fluorescent penetrants contain a dye or several dyes that fluoresce when exposed to
ultraviolet radiation. Visible penetrants contain a red dye that provides high contrast
against the white developer background. Fluorescent penetrant systems are more
sensitive than visible penetrant systems because the eye is drawn to the glow of the
fluorescing indication. However, visible penetrants do not require a darkened area and an
ultraviolet light in order to make an inspection. Visible penetrants are also less vulnerable
to contamination from things such as cleaning fluid that can significantly reduce the
strength of a fluorescent indication.
Penetrants are then classified by the method used to remove the excess penetrant from the
part. The four methods are listed below:
Water washable (Method A) penetrants can be removed from the part by rinsing with
water alone. These penetrants contain some emulsifying agent (detergent) that makes it
possible to wash the penetrant from the part surface with water alone. Water washable
penetrants are sometimes referred to as self-emulsifying systems. Post emulsifiable
penetrants come in two varieties, lipophilic and hydrophilic. In post emulsifiers,
lipophilic systems (Method B), the penetrant is oil soluble and interacts with the oil-based
emulsifier to make removal possible. Post emulsifiable, hydrophilic systems (Method D),
use an emulsifier that is a water soluble detergent which lifts the excess penetrant from
the surface of the part with a water wash. Solvent removable penetrants require the use of
a solvent to remove the penetrant from the part.
Penetrants are then classified based on the strength or detectability of the indication that
is produced for a number of very small and tight fatigue cracks. The five sensitivity levels
are shown below:
The major US government and industry specifications currently rely on the US Air Force
Materials Laboratory at Wright-Patterson Air Force Base to classify penetrants into one
of the five sensitivity levels. This procedure uses titanium and Inconel specimens with
small surface cracks produced in low cycle fatigue bending to classify penetrant systems.
The brightness of the indication produced is measured using a photometer.
The sensitivity levels and the test procedure used can be found in Military Specification
MIL-I-25135 and Aerospace Material Specification 2644, Penetrant Inspection Materials.
An interesting note about the sensitivity levels is that only four levels were originally
planned but when some penetrants were judged to have sensitivities significantly less
than most others in the level 1 category, the level was created. An excellent historical
summary of the development of test specimens for evaluating the performance of
penetrant materials can be found in the following reference.
Penetrants
The industry and military specification that control
the penetrant materials and their use all stipulate
certain physical properties of the penetrant materials
that must be met. Some of these requirements address
the safe use of the materials, such as toxicity, flash
point, and corrosiveness, and other requirements
address storage and contamination issues. Still others
delineate properties that are thought to be primarily
responsible for the performance or sensitivity of the
penetrants. The properties of penetrant materials that
are controlled by AMS 2644 and MIL-I-25135E
include flash point, surface wetting capability,
viscosity, color, brightness, ultraviolet stability,
thermal stability, water tolerance, and removability.
More information on how some of these properties can affect penetrant testing can be
found by following these links.
Surface Energy
Density or Specific Gravity
Viscosity
Color and Fluorescence Brightness
Dimensional Threshold of Fluorescence
Ultraviolet and Thermal Stability
Removability
Wetting ability of a liquid is a function of the surface energies of the solid-gas interface,
the liquid-gas interface, and the solid-liquid interface. The surface energy across an
interface or the surface tension at the interface is a measure of the energy required to
form a unit area of new surface at the interface. The intermolecular bonds or cohesive
forces between the molecules of a liquid cause surface tension. When the liquid
encounters another substance, there is usually an attraction between the two materials.
The adhesive forces between the liquid and the second substance will compete against the
cohesive forces of the liquid. Liquids with weak cohesive bonds and a strong attraction to
another material (or the desire to create adhesive bonds) will tend to spread over the
second material. Liquids with strong cohesive bonds and weaker adhesive forces will
tend to bead-up or form a droplet when in contact with the second material.
In liquid penetrant testing, there are usually three surface interfaces involved, the solidgas interface, the liquid-gas interface, and the solid-liquid interface. For a liquid to spread
over the surface of a part, two conditions must be met. First, the surface energy of the
solid-gas interface must be greater than the combined surface energies of the liquid-gas
and the solid-liquid interfaces. Second, the surface energy of the solid-gas interface must
exceed the surface energy of the solid-liquid interface.
A penetrant's wetting characteristics are also largely responsible for its ability to fill a
void. Penetrant materials are often pulled into surface breaking defects by capillary
action. The capillary force driving the penetrant into the crack is a function of the surface
tension of the liquid-gas interface, the contact angle, and the size of the defect opening.
The driving force for the capillary action can be expressed as the following formula:
Force = 2 r LG cos
Where:
r = radius of the crack opening (2pr is the line of contact between the liquid and the solid
tubular surface.)
LG = liquid-gas surface tension
= contact angle
Since pressure is the force over a given area, it can be written that the pressure developed,
called the capillary pressure, is
Capillary Pressure = (2 LG cos)/ r
The above equations are for a cylindrical defect but the relationships of the variables are
the same for a flaw with a noncircular cross section. Capillary pressure equations only
apply when there is simultaneous contact of the penetrant along the entire length of the
crack opening and a liquid front forms that is an equidistant from the surface. A liquid
penetrant surface could take-on a complex shape as a consequence of the various
deviations from flat parallel walls that an actual crack could have. In this case, the
expression for pressure is
Capillary Pressure = 2( SG - s SL)/r = 2 /r
Where:
SG = the surface energy at the solid-gas interface.
SL = the surface energy at the solid-liquid interface.
r = the radius of the opening.
= the adhesion tension (SG - SL).
Therefore, at times, it is the adhesion tension that is primarily responsible for a
penetrant's movement into a flaw and not the surface energy of the liquid-gas interface.
Adhesion tension is the force acting on a unit length of the wetting line from the direction
of the solid. The wetting performance of the penetrant is degraded when adhesion tension
is the primary driving force.
It can be seen from the equations in this section, that the surface wetting characteristics
(defined by the surface energies) are important penetrant characteristics for filling the
flaw. The liquid penetrant will continue to fill the void until an opposing force balances
the capillary pressure. This force is usually the pressure of trapped gas in the void as most
flaws are open only at the surface of the part. Since the gas originally in the flaw volume
cannot escape through the layer of penetrant, the gas is compressed near the closed end of
the flaw. Since the contact angle for penetrants is very close to zero, other methods have
been devised to make relative comparisons of the wetting characteristics of these liquids.
One method is to measure the height that a liquid reaches in a capillary tube. However,
the solid interface in this method is usually glass and may not accurately represent the
surface that the penetrant inspection will be performed on. Another method of
comparative evaluation is to measure after a set time has elapsed, the radius, the diameter,
or the area of a spot formed when a drop of penetrant is placed on the test surface.
However, using this method, other factors are also acting in the comparison. These
methods include the density, viscosity, and volatility of the liquid, which do not enter into
the capillarity equations, but may have an effect on the inspection as discussed in the
related pages.
Viscosity
Viscosity has little effect on the ability of a penetrant material to enter a defect but it does
have an effect on speed at which the penetrant fills a defect. The equation for a
cylindrical void and an elliptical void are shown below:
Cylindrical Void
Fill time = (2l2)/ rcos LG
Elliptical Void
Fill time = [(2l ) / LGcos] * [a2+b2 / (a+b)ab]
2
quantized. In the animated image below, it can be seen that the electrons move to a
different energy state only when a specific amount of energy is added to or removed from
the system.
Another way of illustrating this point is with an
energy diagram as presented below. This
diagram shows the quantized energy levels for
an atom. Each energy level corresponds to a
quantum state of the atom. The lowest energy
state is called the ground state and is the E0 line
in the diagram. If energy is added to the system
an electron or electrons will jump to a higher
level and the atom is said to be at an excited
state. The upward arrow in the illustration
represents a quantum jump of the atom from the
ground state to second excited state. Depending
on the amount of energy input into the atom, the
energy jump could have been to any of the
levels. However, the jump must be to one of the
levels shown as the atom cannot have an
intermediate value of energy. Atoms will generally be in their ground state.
When considering fluorescence, energy must be
considered at a molecular level. When
molecules form, two or more atoms form an
association where the energy of the molecule is
lower than that of the constituent atoms when
they were separate. The total energy of the
molecule is the sum of the energies holding the
nuclei together and the energy of the chemical
bonds holding the molecule together. Molecules
have rotational, vibrational and electronic (due
to the electrons) energy . It is the vibrational and
electronic energy of the molecule that
contributes to fluorescence. Molecules, like
atoms, will generally be in their ground state.
Molecules can move to a greater energy state only when energy is added to their system.
One of the ways a molecule can gain energy is by absorbing light. If a molecule absorbs
light, the energy of the light must equal to the energy required to put the molecule in one
of the higher energy states. When a molecule reaches an excited state, it does not stay
there for very long but quickly returns to a lower energy state either by emitting light or
by colliding with another atomic particle. When an molecule emits light, the energy of
that light is equal to the energy difference between the quantum levels that molecules has
moved between.
Fluorescent penetrant materials usually contain several dye compounds that are especially
suited for the production of fluorescence.
I = I x e-Ct
o
Where:
I = Transmitted light intensity
Io = Incident light intensity
e = Base of natural log (2.71828)
= Absorption coefficient per unit of concentration
C = Dye concentration
t = Thickness of the absorbing layer trolled to a certain degree by the concentration of the
fluorescent tracer dye in the penetrant.
This equation states that the intensity of the transmitted energy is directly proportional to
the intensity of the incident light and varies exponentially with the thickness of the
penetrant layer and its dye concentration. Therefore, when the dye concentration is
increased, the brightness of the thin layer of penetrant generally increases. However, the
dye concentration can only be increased so much before it starts to have a negative effect
of brightness. A Meniscus-Method Apparatus can be used to measure the dimensional
threshold of fluorescence.
Ultraviolet
Indications
and
Thermal
Stability
of
Penetrant
Exposure to intense ultraviolet light and elevated temperatures can have a negative effect
on fluorescent penetrant indications. Fluorescent materials can loose their brightness after
a period of exposure to high intensity UV light. One study measured the intensity of
fluorescent penetrant indications of a sample that was subjected to multiple UV exposure
cycles. Each cycle consisted of 15 minutes of 800 microwatts/cm UV light and 2.5
minutes of 1500 microwatts/cm UV light. Two penetrants were tested in the study, water
washable, level 3 and a post emulsifiable, level 4. The results from the study showed that
the indications from both penetrants faded with increased UV exposure. After eight
exposure cycles the brightness of the indications was less than one half their original
values.
At an elevated temperature, penetrants can experience heat degradation or "heat fade."
Excessive heat
1. evaporates the more volatile constituents which increases viscosity and adversely
affects
the
rate
of
penetration.
2.
alters
wash
characteristics.
3. "boils off" chemicals that prevent separation and gelling of water soluble penetrants.
4. kills the fluorescence of tracer dyes.
This fourth degradation mechanism involves the molecules of the penetrant materials.
The phenomenon of fluorescence involves electrons that are delocalized in a molecule.
These electrons are not specifically associated with a given bond between two atoms.
When a molecule takes up sufficient energy for the excitation source, the delocalized
bonding electrons rise to a higher electronic state. After excitation, the electrons will
normally lose energy and return to the lowest energy state. This loss of energy can
involve a "radiative" process such as fluorescence or "non-radiative" processes. Nonradiative processes include relaxation by molecular collisions, thermal relaxation, and
chemical reaction. Heat causes the number of molecular collisions to increase which
results in more collision relaxation and less fluorescence.
This explanation is only valid when the part and the penetrant are at an elevated
temperature. When the materials cool, the fluorescence will return. However, while
exposed to elevated temperatures, penetrant solutions dry faster. As the molecules
become more closely packed in the dehydrated solution, collision relaxation increases
and fluorescence decreases. This effect has been called "concentration quenching" and
experimental data shows that as the dye concentration is increased, fluorescent brightness
initially increases but reaches a peak and then begins to decrease. Airflow over the
surface on the part will also speed evaporation of the liquid carrier, so it should be kept to
a minimum to prevent a loss of brightness. Generally thermal damage occurs when
fluorescent penetrant materials are heated above 71C (160F). It should be noted that the
sensitivity of an FPI inspection can be improved if a part is heated prior to applying the
penetrant material that is used, but the temperature should be kept below 71C (160F).
Some high temperature penetrants in use today are formulated with dyes with high
melting points, which improves resistance to heat damage. The penetrants also have high
boiling points and the heat related problems are greatly reduced. However, a loss of
brightness can still take place when the penetrant is exposed to elevated temperatures
over an extended period of time. When one heat resistant formulation was tested, a 20 %
reduction was measured after the material was subjected to 163C (325F) for 273 hours.
The various types of fluorescent dyes commonly employed in today's penetrant materials
begin decomposition at 71C (160F), and when the temperature approaches 94C (200F)
there is almost total attenuation of fluorescent brightness of the total composition and
sublimation of the fluorescent dyestuffs.
Removability
Removing the penetrant from the surface of the sample without removing it from the flaw
is one of the most critical operations of the penetrant inspection process. The penetrant
must be removed from the sample surface as completely as possible to limit background
fluorescence. In order for this to happen, the adhesive forces of the penetrant must not be
so strong that they can not be broken by the removal methods used. However, in order for
the a penetrant to have good surface wetting characteristics the adhesive forces, which are
the forces of attraction between the penetrant and the solid surface being inspected, must
be stronger than the cohesive forces, which are the forces holding the liquid together.
Proper formulation of the penetrant materials provides the correct balancing of these
forces.
Another consideration in the formulation of the penetrant liquid is that it should not easily
commingle and become diluted by the cleaning solution. Dilution of the penetrant liquid
will affect the concentration of the dye and reduce the dimensional threshold of
fluorescence
Emulsifiers
When removal of the penetrant from the defect due to over-washing of the part is a
concern, a post emulsifiable penetrant system can be used. Post emulsifiable penetrants
require a separate emulsifier to break the penetrant down and make it water washable.
Most penetrant inspection specifications classify penetrant systems into four methods of
excess penetrant removal. These are listed below:
1. Method A: Water-Washable
2. Method B: Post Emulsifiable, Lipophilic
3. Method C: Solvent Removable
4. Method D: Post Emulsifiable, Hydrophilic
Method C relies on a solvent cleaner to remove the penetrant from the part being
inspected. Method A has emulsifiers built into the penetrant liquid that makes it possible
to remove the excess penetrant with a simple water wash. Method B and D penetrants
require an additional processing step where a separate emulsification agent is applied to
make the excess penetrant more removable with a water wash. Lipophilic emulsification
systems are oil-based materials that are supplied in ready-to-use form. Hydrophilic
systems are water-based and supplied as a concentrate that must be diluted with water
prior to use . Lipophilic emulsifiers (Method B) were introduced in the late 1950's and
work with both a chemical and mechanical action. After the emulsifier has coated the
surface of the object, mechanical action starts to remove some of the excess penetrant as
the mixture drains from the part. During the emulsification time, the emulsifier diffuses
into the remaining penetrant and the resulting mixture is easily removed with a water
spray.
Hydrophilic emulsifiers (Method D) also remove
the excess penetrant with mechanical and chemical
action but the action is different because no
diffusion takes place. Hydrophilic emulsifiers are
basically detergents that contain solvents and
surfactants. The hydrophilic emulsifier breaks up
the penetrant into small quantities and prevents
these pieces from recombining or reattaching to the
surface of the part. The mechanical action of the
rinse water removes the displaced penetrant from the part and causes fresh remover to
contact and lift newly exposed penetrant from the surface.
The hydrophilic post emulsifiable method (Method D) was introduced in the mid 1970's
and since it is more sensitive than the lipophilic post emulsifiable method it has made the
later method virtually obsolete. The major advantage of hydrophilic emulsifiers is that
they are less sensitive to variation in the contact and removal time. While emulsification
time should be controlled as closely as possible, a variation of one minute or more in the
contact time will have little effect on flaw detectability when a hydrophilic emulsifier is
Developers
The role of the developer is to pull the
trapped penetrant material out of
defects and to spread the developer out
on the surface of the part so it can be
seen by an inspector. The fine
developer particles both reflect and
refract the incident ultraviolet light,
allowing more of it to interact with the
penetrant, causing more efficient
fluorescence. The developer also
allows more light to be emitted
through the same mechanism. This is
why indications are brighter than the
penetrant itself under UV light.
Another function that some developers performs is to create a white background so there
is a greater degree of contrast between the indication and the surrounding background.
Developer Forms
The AMS 2644 and Mil-I-25135 classify developers into six standard forms. These forms
are listed below:
1. Form a - Dry Powder
2. Form b - Water Soluble
3. Form c - Water Suspendible
4. Form d - Nonaqueous Type 1 Fluorescent (Solvent Based)
5. Form e - Nonaqueous Type 2 Visible Dye (Solvent Based)
6. Form f - Special Applications
The developer classifications are based on the method that the developer is applied. The
developer can be applied as a dry powder, or dissolved or suspended in a liquid carrier.
Each of the developer forms has advantages and disadvantages.
Dry Powder
Dry powder developer is generally considered to be the least sensitive but it is
inexpensive to use and easy to apply. Dry developers are white, fluffy powders that can
be applied to a thoroughly dry surface in a number of ways. The developer can be applied
by dipping parts in a container of developer, or by using a puffer to dust parts with the
developer. Parts can also be placed in a dust cabinet where the developer is blown around
and allowed to settle on the part. Electrostatic powder spray guns are also available to
apply the developer. The goal is to allow the developer to come in contact with the whole
inspection area.
Unless the part is electrostatically charged, the powder will only adhere to areas where
trapped penetrant has wet the surface of the part. The penetrant will try to wet the surface
of the penetrant particle and fill the voids between the particles, which brings more
penetrant to the surface of the part where it can be seen. Since dry powder developers
only stick to the part where penetrant is present, the dry developer does not provide a
uniform white background as the other forms of developers do. Having a uniform light
background is very important for a visible inspection to be effective and since dry
developers do not provide one, they are seldom used for visible inspections. When a dry
developer is used, indications tend to stay bright and sharp since the penetrant has a
limited amount of room to spread.
Water Soluble
As the name implies, water soluble developers
consist of a group of chemicals that are dissolved
in water and form a developer layer when the
water is evaporated away. The best method for
applying water soluble developers is by spraying it
on the part. The part can be wet or dry. Dipping,
pouring, or brushing the solution on to the surface
is sometimes used but these methods are less
desirable. Aqueous developers contain wetting
agents that cause the solution to function much
like dilute hydrophilic emulsifier and can lead to
additional removal of entrapped penetrant. Drying is achieved by placing the wet but well
drained part in a recalculating warm air dryer with the temperature held between 70 and
75F. If the parts are not dried quickly, the indications will will be blurred and indistinct.
Properly developed parts will have an even, pale white coating over the entire surface.
Water Suspendible
Water suspendible developers consist of insoluble developer particles suspended in water.
Water suspendible developers require frequent stirring or agitation to keep the particles
from settling out of suspension. Water suspendible developers are applied to parts in the
same manner as water soluble developers. Parts coated with a water suspendible
developer must be forced dried just as parts
coated with a water soluble developer are forced
dried. The surface of a part coated with a water
suspendible developer will have a slightly
translucent white coating.
Nonaqueous
Nonaqueous developers suspend the developer in
a volatile solvent and are typically applied with a
spray gun. Nonaqueous developers are commonly distributed in aerosol spray cans for
portability. The solvent tends to pull penetrant from the indications by solvent action.
Since the solvent is highly volatile, forced drying is not required. A nonaqueous
developer should be applied to a thoroughly dried part to form a slightly translucent white
coating.
Special Applications
Plastic or lacquer developers are a special developers that are primarily used when a
permanent record of the inspection is requi
Preparation of Part
One of the most critical steps in the penetrant
inspection process is preparing the part for inspection.
All coatings, such as paints, varnishes, plating, and
heavy oxides must be removed to ensure that defects
are open the surface of the part. If the parts have been
machined, sanded, or blasted prior to the penetrant
inspection, it is possible that a thin layer of metal may
have smeared across the surface and closed off defects.
is even possible for metal smearing to occur as a result
cleaning operations such as grit or vapor blasting. This
layer of metal smearing must be removed before
inspection.
It
of
Contaminants
Coatings, such as paint, are much more elastic than
metal and will not fracture even though a large defect
may be present just below the coating. The part must be
thoroughly cleaned as surface contaminates can prevent the penetrant from entering a
defect. Surface contaminants can also lead to a higher level of background noise since the
excess penetrant may be more difficult to remove.
Common coatings and contaminates that must be removed include: paint, dirt, flux, scale,
varnish, oil, etchant, smut, plating, grease, oxide, wax, decals, machining fluid, rust, and
residue from previous penetrant inspections.
Some of these contaminants would obviously prevent penetrant from entering defects and
it is, therefore, clear that they must be removed. However, the impact of other
contaminants such as the residue from previous penetrant inspections is less clear, but
they can have a disastrous affect on the inspection. Take the link below to review some of
the research that has been done to evaluate the effects of contaminants on LPI sensitivity.
A good cleaning procedure will remove all contamination from the part and not leave any
residue that may interfere with the inspection process. It has been found that some
alkaline cleaners can be detrimental to the penetrant inspection process if they have
silicates in concentrations above 0.5 percent. Sodium metasilicate, sodium silicate, and
related compounds can adhere to the surface of parts and form a coating that prevents
penetrant entry into cracks. Researchers in Russia have also found that some domestic
soaps and commercial detergents can clog flaw cavities and reduce the wettability of the
metal surface, thus, reducing the sensitivity of the penetrant. Conrad and Caudill found
that media from plastic media blasting was partially responsible for loss of LPI indication
strength. Microphotographs of cracks after plastic media blasting showed media
entrapment in addition to metal smearing.
It is very important that the material being inspected has not been smeared across its own
surface during machining or cleaning operations. It is well recognized that machining,
honing, lapping, hand sanding, hand scraping, shot peening, grit blasting, tumble
deburring, and peening operations can cause a small amount of the material to smear on
the surface of some materials. It is perhaps less recognized that some cleaning operations,
such as steam cleaning, can also cause metal smearing in the softer materials. Take the
link below to learn more about metal smearing and its affects on LPI
Material smearing can have a very detrimental effect on a LPI inspection as defects that
are normally open to the surface can partially or completely be covered over. Some of the
processes that can cause material to smear include machining, honing, lapping, sanding,
scraping, shot peening, grit blasting, tumble deburring, and peening operations. When
high pressure is used some cleaning operations, such as vapor and steam cleaning, can
also cause material to smear in the softer materials. Softer materials such as plastics and
aluminum alloys are most prone to smearing but many other materials, such as steel,
titanium and Inconel alloys, have also been shown to smear. To evaluate the effect of a
process on liquid penetrant inspection, cracked specimens are typically inspected before
and after performing the potentially smearing operation and a comparison between the
inspection results are made. It must be noted that under carefully controlled conditions,
material smear can be avoided. Whenever parts have been mechanically processed prior
to LPI, an evaluation should be performed to determine if flaw detectability has been
compromised. If material smearing is a problem, an etching process can be used to
remove the smeared material prior to inspection. The curves below illustrate the effect
that metal smearing can have on the probability of detection for a defect and how etching
the sample surface improves detectability.
The curve on the left shows the probability of detecting a crack versus crack length for
as-machined aluminum specimens. The curve on the right shows the POD for the same
aluminum specimens after their surfaces had been etched. Comparing the crack lengths
where the curves reach a POD level of 90%, it can be seen that in the as-machined
condition, the crack length would need to be 0.4 inches long. However, when the surface
is etched, cracks under 0.10 inch can be detected with a 90 % probability.
Removal of Material Smearing
Etching of the specimens can return the flaw to the pre-mechanical processing level of
detectability. The amount of material that must be removed by the etching process
depends on the amount of material that has been smeared and should be determined
experimentally. Volume two of the Nondestructive Testing Handbook provides a great
deal of information on material smearing and the amount of etching required to remover
the damage. The handbook includes a number of photographs such as the set below that
graphically show the effects of metal smearing.
is
The dynamic shear viscosity of the penetrant, which can vary with the diameter of
the capillary. The viscosity of a penetrant in microcapillary flaws is higher than its
viscosity in bulk, which slows the infiltration of the tight flaws.
The radius of the flaw or the distance between the flaw walls.
The ideal dwell time is often determined by experimentation and is often very specific to
a particular application. For example, AMS 2647A requires that the dwell time for all
aircraft and engine be at least 20 minutes while the ASTM E1209 only requires a 5
minute dwell time for parts made of titanium and other heat resistant alloys. Generally,
there is no harm in using a longer penetrant dwell time as long as the penetrant is not
allowed to dry.
The following tables summarize the dwell time requirements of several commonly used
specifications. The information provided below is intended for general reference and no
guarantee is made about its correctness or currentness. Please consult the specifications
for the actual dwell time requirements.
but not given time to make its way into defects to react with the trapped penetrant. The
penetrant that has reacted with the emulsifier is easily cleaned away. Controlling the
reaction time is of essential importance when using a post-emulsifiable system. If the
emulsification time is too short, an excessive amount of penetrant will be left on the
surface leading to high background levels. If the emulsification time is too long, the
emulsifier will react with the penetrant entrapped in discontinuities making it possible to
deplete the amount needed to form an indication.
The hydrophilic post emulsifiable method (Method D) is more sensitive than the
lipophilic post emulsifiable method (Method B). Since these methods are generally only
used when very high sensitivity is needed, Method D is almost always used making
Method B virtually obsolete. The major advantage of hydrophilic emulsifiers is that they
are less sensitive to variation in the contact and removal time. While emulsification time
should be controlled as closely as possible, a variation of one minute or more in the
contact time will have little effect on flaw detectability when a hydrophilic emulsifier is
used, but a variation of as little as 15 to 30 seconds can have a significant effect when a
lipophilic system is used. Using an emulsifier involves adding a couple of steps to the
penetrant process and ,therefore, slightly increases the cost of an inspection. When using
an emulsifier, the penetrant process includes the following steps (extra steps in bold): 1.
pre-clean part, 2. apply penetrant and allow to dwell, 3. pre-rinse to remove first layer
of penetrant, 4. apply hydrophilic emulsifier,and allow contact for specified time, 5.
rinse to remove excess penetrant, 6. dry part, 7. apply developer and allow part to
develop, and 8. inspect.
Rinse Method and Time for Water-Washable Penetrants
The method used to rinsing the excess penetrant from the object surface and the time of
the rinse should be controlled so as to prevent over-washing. It is generally recommended
that a coarse spray rinse or an air agitated, immersion wash tank be used. When a spray is
being used, it should be directed at a 45 angle to the part surface so as to not force water
directly into any discontinuities that may be present. The spray or immersion time should
be kept to a minimum through frequent inspections of the remaining background level.
When using
Hand Wiping of Solvent Removable Penetrants
When a solvent removable penetrant is used, care must also be taken to carefully remove
the penetrant from the part surface while removing as little as possible from the flaw. The
first step in this cleaning procedure is to dry wipe the surface of the part in one direction
using a white lint free cotton rag. One dry pass in one direction is all that should be used
to remove as much penetrant as possible. Next, the surface should be wiped with one pass
in one direction with a cleaner moistened rag. One dry pass followed by one damp pass is
all that is recommended. Additional wiping my sometimes be necessary but keep in mind
that with every additional wipe, some of the entrapped penetrant will be removed and
inspection sensitivity will be reduced.
To study the effects of the wiping process, Japanese researchers manufactured a test
specimen out of acrylic plates that allowed them to view the movement of the penetrant
in a narrow cavity. The sample consisted of two pieces of acrylic with two thin sheets of
vinyl clamped between as spaces. The plates were clamped in the corners and all but one
of the edges sealed. The unsealed edge acted as the flaw. The clearance between the
plates varied from 15 microns (0.059 inch) at the clamping points to 30 microns (0.118
inch) at the midpoint between the clamps. The distance between the clamping points is
believed to be 30 mm (1.18 inch).
Although the size of the flaw represented by this specimen is large, an interesting
observation was made. They found that when the surface of the specimen was wiped with
a dry cloth, penetrant was blotted and removed from the flaw at the corner areas were the
clearance between the plate was least. When the penetrant at the side areas was removed,
penetrant moved horizontally from the center area to the ends of the simulated crack
where capillary forces are stronger. Therefore, across the crack length, the penetrant
surface has a parabola-like shape where the liquid is at the surface in the corners but
depressed in the center. This shows that each time the cleaning cloth touches the edge of a
crack, penetrant is lost from the defect. This also explains why the bleedout of an
indication is often largest at the corners of cracks.
Developer Form
Nonaqueous Wet Solvent
Plastic Film
water-soluble
Water Suspendible
water-soluble
Water Suspendible
Dry
Dry
Dry
Dry
Method of Application
Spray
Spray
Spray
Spray
Immersion
Immersion
Dust Cloud (Electrostatic)
Fluidized Bed
Dust Cloud (Air Agitation)
Immersion (Dip)
The following TABLE lists the main advantages and disadvantages of the various
developer types.
Developer
Advantages
Dry
Soluble
Disadvantages
Does not form contrast
background so cannot be
used with visible systems
Difficult to assure entire part
surface has been coated
Coating is translucent and
provides poor contrast (not
recommended for visual
systems)
Indications for water
washable systems are dim
and blurred
Very portable
Easy to apply to readily
accessible surfaces
Nonaqueous
sit
to
is
Hydrophilic Emulsifiers
Hydrophilic emulsifiers have less tolerance for penetrant contamination. The penetrant
tolerance varies with emulsifier concentration and the type of contaminating penetrant. In
some cases, as little as 1% by volume of penetrant contamination can seriously affect the
performance of an emulsifier. One penetrant manufacture reports that 1 to 1.5% penetrant
contamination will affect solution with a 10% concentration of emulsifier. As the
emulsifier concentration increases in the solution, the penetrant contamination tolerance
also increases and a solution with a 30% emulsifier concentration can tolerate from 5 to
8.5% penetrant contamination. The percentage of added penetrant required to destroy
washability of the emulsifier can be measured and an oil tolerance index is commonly
used to compare the tolerance of different emulsifiers to contamination by penetrants.
AMS 2647A requires that the emulsification bath be discarded if penetrant is noted
floating on the surface or adhering to the sides of the tank.
Water contamination is not as much of a concern with hydrophilic emulsifiers as they are
miscible with water. However, it is very important that the emulsifier solution be kept at
the proper concentration.
It should also be noted that penetrant dragout and, thus, level of possible emulsifier
contamination by the penetrant is dependent on the type of material being processed.
Tests have shown that on both polished and grit blasted surfaces, aluminum and stainless
steel parts had a greater drag-out than titanium parts.
Emulsifier Concentration and Contact Time
The optimal emulsifier contact time is dependent on a number of variables that include
the emulsifier used, the emulsifier concentration, the surface roughness of the part being
inspected, and other factors. Usually some experimentation is required to select the
proper emulsifier contact time. The emulsifier used must be matched to the penetrant
material. For method D penetrant systems the concentration of the emulsifier should not
exceed the percentage specified by the supplier and if working to a specification should
not exceed the concentration specified. Since the emulsifier is mixed with water, which is
prone to evaporation, it is recommended that the starting concentration be less than that
recommended by the supplier. One penetrant manufacture recommends the following
starting concentrations:
increased rapidly as the lipophilic contact time was increased in steps to 5 minutes. Also
as expected, the decrease in sensitivity increased with increasing crack size.
Vaerman also looked at the effect of hydrophilic emulsifier concentration. It was found
that increasing the concentration for 5% by volume to 33 percent, decreased sensitivity
by 15% when a three minute contact time was used. When a contact time of one minute
was used the decrease in sensitivity was just over nine percent.
Ref: -- Vaerman, J., Fluorescent Penetrant Inspection, Quantified Evolution of the
Sensitivity Versus Process Deviations, Proceedings of the 4th European Conference on
Non- Destructive Testing, September 1987, Pergamon Press, Maxwell House, Fairview
Park, Elmsford, New York, Volume 4, pp. 2814-2823.
Hyam also reports on the effect of the emulsifier concentration and contact time. Both
hydrophilic and lipophilic removers were tested. The results showed that as the
concentration of the emulsifier was increased from 2.5% to 20%, sensitivity decreased.
The contact time was shown to have little effect on the hydrophilic system tested (up to
20 minutes) but to have a significant effect on the lipophilic system with sensitivity
decreasing as contact time was increased from 2 to 10 minutes.
Since fluorescent brightness is linear with respect to ultraviolet excitation, a change in the
intensity of the light (from age or damage) and a change in the distance of the light
source from the surface being inspected will have a direct impact on the inspection. For
UV lights used in component evaluations, the normally accepted intensity is 1000
microwatts per square centimeter when measured at 15 inches from the filter face
(requirements can vary from 800 to 1200). The required check should be performed when
a new bulb is installed, at startup of the inspection cycle, if a change in intensity is
noticed, or every eight hours of continuous use. Regularly checking the intensity of UV
lights is very important because bulbs loose intensity over time. In fact, a bulb that is near
the end of its operating life will often have an intensity of only 25 percent of its original
output. Black light intensity will also be affected by voltage variations. A bulb that
produces acceptable intensity at 120 volts will produce significantly less at 110 volts. For
this reason it is important to provide constant voltage to the light. Also, most UV light
must be warmed up prior to use and should be on for at least 15 minutes before beginning
an inspection.
When performing a fluorescent penetrant inspection, it is important to keep white light to
a minimum as it will significantly reduce the inspectors ability to detect fluorescent
indications. Light levels of less than 2 fc are required by most procedures with some
procedures requiring less than 0.5 fc at the inspection surface. Procedures require a check
and documentation of ambient white light in the inspection area. When checking black
light intensity at 15 inches a reading of the white light produced by the black light may be
required to verify white light is being removed by the filter.
Light Measurement
Light intensity measurements are made using a radiometer. A radiometer is an instrument
that translate light energy into an electrical current. Light striking a silicon photodiode
detector causes a charge to build up between internal layers. When an external circuit is
connected to the cell, an electrical current is produced. This current is linear with respect
to incident light. Some radiometers have the ability to measure both black and white
light, while others require a separate sensor for each measurement. Whichever type is
used, the sensing area should be clean and free of any materials that could reduce or
obstruct light reaching the sensor. Radiometers are relatively unstable instruments and
readings often change considerable over time. Therefore, they should be calibrated at
least every six months.
Ultraviolet light measurements should be taken using a fixture to maintain a minimum
distance of 15 inches from the filter face to the sensor. The sensor should be centered in
the light field to obtain and record the highest reading. UV spot lights are often focused,
so intensity readings will vary considerable over a small area. White lights are seldom
focused and depending on the wattage, will often produce in excess of the 100 fc at 15
inches. Many specifications do not require the white light intensity check to be conducted
at a specific distance.
small round defects than small linear defects. Small round defects are generally
easier to detect for several reasons. First, they are typically volumetric defects that
can trap significant amounts of penetrant. Second, round defects fill with
penetrant faster than linear defects. One research effort found that elliptical flaw
with length to width ratio of 100, will take the penetrant nearly 10 times longer to
fill than a cylindrical flaw with the same volume.
deeper flaws than shallow flaws. Deeper flaws will trap more penetrant than
shallow flaws, and they are less prone to over washing.
flaws with a narrow opening at the surface than wide open flaws. Flaws with
narrow surface openings are less prone to over washing.
flaws on smooth surfaces than on rough surfaces. The surface roughness of the
part primarily affects the removability of a penetrant. Rough surfaces tend to trap
more penetrant in the various tool marks, scratches, and pits that make up the
surface. Removing the penetrant from the surface of the part is more difficult and
a higher level of background fluorescence or over washing may occur.
flaws with rough fracture surfaces than smooth fracture surfaces. The surface
roughness that the fracture faces is a factor in the speed at which a penetrant
enters a defect. In general, the penetrant spreads faster over a surface as the
surface roughness increases. It should be noted that a particular penetrant may
spread slower than others on a smooth surface but faster than the rest on a rougher
surface.
Penetrant
When proper health and safety precautions are followed, liquid penetrant inspection
operations can be completed without harm to inspection personnel. However, there are a
number of health and safety related issues that must be addressed. Since each inspection
operation will have its own unique set of health and safety concerns that must be
addressed, only a few of the most common concerns will be discussed here.
Chemical Safety
Whenever chemicals must be handled, certain precautions must be taken as directed by
the material safety data sheets (MSDS) for the chemicals. Before working with a
chemical of any kind, it is highly recommended that the MSDS be reviewed so that
proper chemical safety and hygiene practices can be followed. Some of the penetrant
materials are flammable and, therefore, should be used and stored in small quantities.
They should only be used in a well ventilated area and ignition sources avoided. Eye
protection should always be worn to prevent contact of the chemicals with the eyes.
Many of the chemicals used contain detergents and solvents that can dermatitis. Gloves
and other protective clothing should be warn to limit contact with the chemicals.
Ultraviolet Light Safety
Ultraviolet (UV) light or "black light" as it is sometimes called, has wavelengths ranging
from 180 to 400 nanometers. These wavelengths place UV light in the invisible part of
the electromagnetic spectrum between visible light and X-rays. The most familiar source
of UV radiation is the the sun and is necessary in small doses for certain chemical
processes to occur in the body. However, too much exposure can be harmful to the skin
and eyes. Excessive UV light exposure can cause painful sunburn, accelerate wrinkling
and increase the risk of skin cancer. UV light can cause eye inflammation, cataracts, and
retinal damage.
Because of their close proximity, laboratory devices, like UV lamps, deliver UV light at a
much higher intensity than the sun and, therefore, can cause injury much more quickly.
The greatest threat with UV light exposure is that the individual is generally unaware that
the damage is occurring. There is usually no pain associated with the injury until several
hours after the exposure. Skin and eye damage occurs at wavelengths around 320 nm and
shorter which is well below the 365 nm wavelength, where penetrants are designed to
fluoresce. Therefore, UV lamps sold for use in LPI application are almost always filtered
to remove the harmful UV wavelengths. The lamps produce radiation at the harmful
wavelengths so it is essential that they be used with the proper filter in place and in good
condition.